Gossen MetraWatt METRAOHM 413 Quick Start Manual

METRAOHM413
Low-Resistance Ohmmeter
• Handy measuring instrument per VDE 0413 part 4 / EN 61557-1 / -4
Large measuring range from 0.01 to 1999 kΩ
• Measurement cable zero balancing
Compact and durable For rugged service calls and laboratory use
Overvoltage protection Protects the instrument in the event of inadvertent connection to mains voltage
• Indication of interference voltage
• IP 65 protection
3-348-810-03
6/7.15
Applications
The METRAOHM 413 is a battery powered instrument for the mea­surement of low-value resistances in electrical systems. Large protective conductor, earthing, equipotential bonding and lightning protection systems can also be tested quickly and safely with the METRAOHM 413. Resistance is measured between reference earth (e.g. an equipotential bonding strip) and any desired points to this end. A measuring current of 200 mA and automatic functions assure reliable measuring results.
– Adjustment of shunts in instrumentation applications
– Testing of electrical connections at busbars in open-pit
mining, in industry and in household applications
– Testing of cable resistance, wiring, shunt resistors in printed
circuit boards and thick-film circuits
– Measurement of contact resistance in relays, power
contactors and load interrupters
– Measurement of fuse resistance, as well as cable resistance in
heavy current circuits
– Testing of coil resistance in transformers, coils, small motors
etc.
Description
Easy Operation
Operation is very easy. Measurement cables of almost any length can be used thanks to the offset balancing function. Measurement polarity can be reversed, and both measured values are stored to memory for subsequent comparison. Resultant deviations in the measuring results (e.g. possible galvanic voltages) can thus be recognized.
Offset Balancing
Offset balancing compensates for measurement cable resistance. The offset value is saved to memory and is utilized for all future measurements until balancing is performed once again. The value remains in memory, even after the measuring instrument has been switched off. A variety of measurement cables with resistances of up to 3.5 Ω can be used.
Protection Against Operator Error
The METRAOHM 413 is protected against inadvertent connection to devices under test with fault voltages of up to 500 V by means of protective devices.
Interference Voltage Warning
Interference voltage is indicated visually with a red LED, and audibly with an acoustic warning signal. The voltage value is displayed at the LCD as well, as long as the test probe makes contact.
GMC-I Messtechnik GmbH
METRAOHM413
Low-Resistance Ohmmeter
Applicable Regulations and Standards
IEC 61010–1 DIN EN 61010–1 VDE 0411 Part 1
EN 61557-1 / -4 / VDE 0413 Part 4
EN 60529 VDE 0470-1
DIN EN 61326-1 VDE 0843-20-1
Safety requirements for electrical equipment for measurement, control and laboratory use Part 1: General requirements
Ohmmeters
Test instruments and test procedures Protection provided by enclosures (IP code)
Electrical equipment for measurement, control and laboratory use – EMC requirements – Part 1: General requirements
Characteristic Values
Measuring ranges 10 Ω (Lo-Ω) Auflösung: 0,01 Ω
Accuracy ±(1.5% rdg. + 4 digits) at 20 °C Measuring current 10 Ω (Lo-Ω)-range: 200 mA
Open-circuit voltage > 4 V
100, 2000 Ω Auflösung: 0,1 ... 1 Ω 20, 200, 2000 kΩ Auflösung: 0,01 ... 1 kΩ
100 Ω-Bereich: 20 mA further ranges: 1 μA ... 20 mA
Display
Electrical Safety
Safety class II per IEC 61010-1 Test voltage 3.2 kV Measuring category CAT II 600 V, CAT III 300 V Overvoltage
protection reversible by means of semiconductor at
nominal voltages of up to 400 V, from 400 to 500 V with special fuse (can only be replaced at the factory)
Electromagnetic Compatibility (EMC)
Interference emission interference immunity
EN 61326-1:2006 class B EN 61326-1:2006
Mechanical Design
Housing impact resistant ABS with unbreakable
Protection housing: IP 65 Dimensions 60 x 230 x 40 mm Weight 180 g (including battery)
display cover
Digital display LCD, 10 mm high, 3½ places,
with background illumination
Overflow indication by 0L Battery status indicated by 2 segments on the LCD External voltage
display for voltages as from approx. 15 V:
Alternately, red LED VOLT blinks and symbol – v – appears on the LCD. Numerical display of voltage level, in the case of AC voltage, symbol ~ for AC is additionally shown. Between 15 V and 400 V an intermittent acoustic signal sounds as a warning signal
Ambient Conditions
Reference temp. 20 °C Storage temperature –10 °C ... +70 °C Operating temp. –10 °C ... +50 °C
Power Supply
Battery 9 volt block battery per IEC 6 LR 61 AlMn,
automatic shutdown after 20 seconds of inactivity
Standard Equipment
1 METRAOHM 413 low-resistance ohmmeter 1 plug-in measuring cable 1 9 volt block battery per IEC 6 LR61 1 operating instructions
Battery Saving Circuit
The measuring instrument is switched off automatically if none of the keys are activated for a period of approximately 20 seconds.