Gossen Metrawatt METRALINE ISOCHECK Operating Instructions

Operating Instructions
METRALINE ISO
CHECK
Insulation Resistance Tester
3-349-691-03
2/4.13
Table of Contents
!
!
1 Introduction ...................................................................... 2
1.1 Scope of Delivery ...........................................................................2
1.2 Optional Accessories ......................................................................2
1.3 Safety Precautions .........................................................................2
1.4 Applications ..................................................................................3
1.5 Applicable Standards .....................................................................3
1.6 Environment ..................................................................................3
2 Device Description ........................................................... 3
2.1 Housing ........................................................................................3
2.2 Control Panel .................................................................................3
2.3 Initial Start-Up ...............................................................................3
3 Performing Measurements .............................................. 4
3.1 Switching the Instrument On and Off, Energy-Saving Mode,
Automatic Shutdown ......................................................................4
3.2 Instructions and Principles with Validity for
All Measurements ..........................................................................4
3.3 Measuring Functions ......................................................................4
3.3.1 Voltage Measurement ....................................................................4
3.3.2 Insulation Resistance Measurement ................................................5
3.3.3 Measurements at Varistors (surge protection devices) ......................5
3.4 Further Device Functions ................................................................6
3.5 Device Reset Function ....................................................................7
4 Technical Data ................................................................. 7
4.1 Measuring Functions ......................................................................7
4.2 General Data .................................................................................7
5 Maintenance .................................................................... 7
5.1 Device Power Supply .....................................................................7
5.1.1 Inserting and Replacing the Batteries ..............................................7
5.1.2 Recharging the Batteries ................................................................8
5.2 Cleaning .......................................................................................8
5.3 Recalibration .................................................................................8
6 Repair and Replacement Parts Service,
Calibration Center and Rental
Instrument Service ........................................................... 8
7 Product Support ............................................................... 8

1 Introduction

1.1 Scope of Delivery

1 Test instrument with mobile test probe 4 Batteries (AAA) 1Pouch 1 Condensed operating instructions
1
CD ROM with operating instructions in available languages
1 Factory calibration certificate

1.2 Optional Accessories

4 ea. rechargeable AAA NiMH batteries (Z507B) 1 battery charger (Z507A)

1.3 Safety Precautions

Read the operating instructions thoroughly and carefully before using your instrument. Follow all instructions contained therein. Make sure that the operating instructions are available to all users of the instrument.
Meanings of Symbols on the Instrument
This device is equipped with double or reinforced insula­tion.
Danger of injury due to electrical current, warning regarding dangerous electrical voltage
Warning concerning a source of danger (attention, observe documentation!)
EC mark of conformity: This instrument fulfills all requirements of applicable European directives.
Before using the instrument, it must be assured that it is safe. Do not use if:
• Visible damage is apparent
• The battery compartment lid is missing
• The device has been stored under unfavorable conditions for a lengthy period of time
• The device has been handled impermissibly, e.g. has been dropped from a height of 1 meter or more
• The test instrument does not function as described in these operating instructions (if this is the case, we recommend resetting the device as described in section 3.5 on page 7)
CAUTION
• Do not touch electrically conductive parts, test probes etc. when the device is switched on and voltage might still be con­ducted by a test probe – DANGER OF INJURY!
• Only use test probes which are included with the instrument or are available as accessories.
• The device must be switched off and no voltage may be applied when exchanging accessories.
• It is absolutely essential to adhere to all safety precautions, regulations and standards when performing measurements.
• No keys may be pressed when connecting the instrument to a device under test.
• The test instrument may not be subjected to the influence of aggressive substances, gases, vapors, liquids or dust.
• The test instrument may only be used under the conditions listed in the technical data in section 5 on page 7.
• If the device is moved from a colder to a warmer room, con­densation may occur, in which case a brief period of acclima­tization is advisable.
• We recommend removing the batteries during lengths periods of storage.
• Two relatively strong magnets are integrated into the test instrument. Avoid close proximity to magnetically sensitive objects such as watches, credit cards and the like.
• The illustrations in these operating instructions are drawings and may therefore deviate from reality.
• Use original accessories only.
• Maximum permissible voltage between test probe and ground is 300 V!
• Maximum permissible (externally applied) voltage between the test probes is 600 V!
Opening of Equipment / Repair
The equipment may be opened only by authorized service per­sonnel to ensure the safe and correct operation of the equipment and to keep the warranty valid.
Even original spare parts may be installed only by authorized ser­vice personnel.
In case the equipment was opened by unauthorized personnel, no warranty regarding personal safety, measurement accuracy, conformity with applicable safety measures or any consequential damage is granted by the manufacturer.
2 GMC-I Messtechnik GmbH

1.4 Applications

Fixed Test Probe Attached to the Device
LED
OLED Display
Socket for Flexible Test Probe
Socket for Charger
P
l
u
g
-
I
n
M
o
b
i
l
e
T
e
s
t
P
r
o
b
e
Plastic
Control Panel
Housing
Screws for Battery Compartment Lid
Battery Compartment Lid
Serial Plate
3
1
2
4
5
6
The test instrument consists of a compact housing with a pat­ented means of retaining the second test probe.
The high-contrast, four-color OLED display assures excellent legi­bility. When performing measurements under unfavorable light conditions, measuring point illumination can be switched on – white LED at the front.
The test instrument is capable of the following measurements: – Insulation resistance with test voltages of 50 to 1000 V – Surge protection devices with test voltages of 50 to 1000 V – Direct and alternating voltages

1.5 Applicable Standards

Measurement EMC Safety
EN 61557-1 EN 55022 class B EN 61010-1 EN 61557-2 EN 61326-1 EN61010-031

1.6 Environment

The shipping package is made of recyclable cardboard. Batteries must be disposed of in accordance with applicable
regulations.
This device may not be disposed of with the trash. Further information regarding the WEEE mark can be accessed on the Internet at www.gossenmetrawatt.com by entering the search term “WEEE”.

2 Device Description

2.1 Housing

Figure 2.2: Detail View of the Back Panel with Battery
Compartment Lid

2.2 Control Panel

1 Graphic OLED display 2 START key:
Switch on: Press and hold until the dis­play lights up. – Start measurement: Press and hold until measure­ment starts. – Measurement point illumina- tion: Press briefly to switch illumi­nation on and off. – Switch off: Press twice briefly to switch the instrument off.
3 R
ISO/USPD KEY:
For switching between insula­tion resistance and surge pro­tection measurement
4The FUNC key allows for selec-
tion of the desired type of surge protection when USPD is activated.
5 U
TEST key for selecting
measuring voltage
TEST key for selecting
6 U
measuring voltage
Figure 2.1: Top View
For transport purposes, the movable test probe can be attached to the housing and retained by a magnet such that both metal tips are simultaneously recessed and protected.
In order to charge batteries which have been inserted into the instrument, the flexible test probe’s connector plug must be removed and the slide must be pushed to the left, so that the socket at the right is made accessible and the charger can be plugged into it.
GMC-I Messtechnik GmbH 3
Figure 2.3: Control Panel and OLED Display

2.3 Initial Start-Up

After inserting the batteries in accordance with section 5.1 on page 7, the measuring instrument is ready for operation.
Loading...
+ 5 hidden pages