The accessories available for your instrument are checked for
compliance with currently valid safety regulations at regular
intervals, and are amended as required for new applications.
Currently up-to-date accessories which are suitable for your measuring instrument are listed at the following web address along
with photo, order number, description and, depending upon the
scope of the respective accessory, data sheet and
operating instructions:
www.gossenmetrawatt.com
We calibrate and recalibrate all instruments supplied by GMC-I
Messtechnik GmbH, as well as other manufacturers, at our service center, for example after one year within the framework of
your test equipment monitoring program, as well as prior to use
etc. and offer you test equipment management free of charge.
See also chapter 9.6 on page 71.
Repair and Replacement Parts Service
Calibration Center* and Rental Instrument Service
This address is only valid in Germany. Please contact our
representatives or subsidiaries for service in other countries.
* DAkkS Calibration Laboratory for Measured Electrical Quantities
D-K-15080-01-01 accredited per DIN EN ISO/IEC 17025
Accredited quantities: direct voltage, direct current value, direct current
resistance, alternating voltage, alternating current value, AC active
power, AC apparent power, DC power, capacitance, frequency,
temperature
4GMC-I Messtechnik GmbH
Competent Partner
GMC-I Messtechnik GmbH is certified in accordance with
DIN EN ISO 9001.
Our DAkkS calibration laboratory is accredited by the Deutsche
Akkreditierungsstelle GmbH (National accreditation body for the
Federal Republic of Germany) in accordance with DIN EN ISO/IEC
17025 under registration number D-K-15080-01-01.
We offer a complete range of expertise in the field of metrology:
from test reports and factory calibration certificates, right on up to
DAkkS calibration certificates.
Our spectrum of offerings is rounded out with free test equipment management.
As a full service calibration lab, we can calibrate instruments from
other manufacturers as well.
Contact Persons
GMC-I Messtechnik GmbH5
Table of Contents
ContentsPageContentsPage
1Safety Features and Precautions ....................................... 8
1.1Use for Intended Purpose ................................................................10
1.2Meanings of Danger Symbols ..........................................................10
1.3Meanings of Acoustic Warning Signals ............................................10
You have selected an instrument which provides you with a high
level of safety.
This instrument fulfills the requirements of the applicable European guidelines and national regulations. This is confirmed by
means of the CE mark. A corresponding declaration of conformity
can be
requested from GMC-I Messtechnik GmbH.
The TRMS digital multimeter has been manufactured and tested
in accordance with the following safety regulations:
IEC 61010–1:2010/DIN EN 61 010–1:2011/VDE 0411–1:2011.
When used for its intended purpose (see page 10), safety of the
operator, as well as that of the instrument, is assured. Their safety
is however not guaranteed, if the instrument is used improperly or
handled carelessly.
In order to maintain flawless technical safety conditions, and to assure
safe use, it is imperative that you read the operating instructions
thoroughly and carefully before placing your instrument into service, and
that you follow all instructions contained therein.
The multimeter is equipped with an automatic socket blocking
mechanism for your safety, and in order to safeguard your
instrument. This mechanism is linked to the rotary switch and only
allows access to those jacks which are actually required for the
selected function. It also prevents the user from turning the rotary
switch to impermissible functions after the measurement cables
have already been plugged in.
Measuring Categories and their Significance per IEC 61010-1
CATDefinition
Measurements in electrical circuits which are not directly connected to the
I
mains: e.g. electrical systems in motor vehicles and aircraft, batteries etc.
Measurements in electrical circuits which are electrically connected to the
II
low-voltage mains: via plug, e.g. in household, office and laboratory applications
Measurements in building installations: stationary consumers, distributor
III
terminals, devices connected permanently to the distributor
The measuring category and the maximum rated voltage which
are printed on the device apply to your measuring instrument, e.g.
600 V CAT II I or 1000 V CAT II.
For the application of measuring cables see page 72.
Observe the following safety precautions:
• The multimeter may not be used in potentially explosive
atmospheres.
• The multimeter may only be operated by persons who are
capable of recognizing contact hazards and taking the
appropriate safety precautions. Contact hazards according to
the standards exist anywhere, where voltages of greater than
33 V RMS or 70 V DC may occur. Avoid working alone when
taking measurements which involve contact hazards. Be
certain that a second person is present.
• Maximum allowable voltage between the voltage measuring
sockets or all connector sockets and ground is 1000 V for
measuring category II, and 600 V for measuring category I I I.
• Be prepared for the occurrence of unexpected voltages at
devices under test (e.g. defective devices). For example,
capacitors may be dangerously charged.
8GMC-I Messtechnik GmbH
Safety Warnings
• Make certain that the measurement cables are in flawless
condition, e.g. no damage to insulation, no interruptions in
cables or plugs etc.
• No measurements may be made with this instrument in
electrical circuits with corona discharge (high-voltage).
• Special care is required when measurements are made in HF
electrical circuits. Dangerous pulsating voltages may be present.
• Measurements under moist ambient conditions are not
permitted.
• Be absolutely certain that the measuring ranges are not
overloaded beyond their allowable capacities. Limit values are
included in chapter 8; “Technical Data” in the table entitled
“Measuring Functions and Measuring Ranges” in the
“Overload Capacity” column.
• The multimeter may only be operated with installed batteries or
rechargeable batteries. Dangerous currents and voltages are
otherwise not indicated, and the instrument may be damaged.
• The instrument may not be operated if the fuse cover or the
battery compartment lid has been removed, or if its housing is
open.
• The input for the current measuring range is equipped with a
fuse link.
Maximum permissible voltage for the measuring circuit (= rated
voltage of the fuse) is 1000 V AC/DC.
Use specified fuses only (see page 65)! The fuse must have a
breaking capacity of at least 30 kA.
Opening of Equipment / Repair
The equipment may be opened only by authorized service personnel to ensure the safe and correct operation of the equipment
and to keep the warranty valid.
Even original spare parts may be installed only by authorized service personnel.
In case the equipment was opened by unauthorized personnel,
no warranty regarding personal safety, measurement accuracy,
conformity with applicable safety measures or any consequential
damage is granted by the manufacturer.
Repair and Parts Replacement by Authorized Service Personnel
When the instrument is opened, voltage conducting parts may be
exposed. The instrument must be disconnected from the measuring circuit before the performance of repairs or the replacement of
parts. If repair of a live open instrument is required, it may only be
carried out by trained personnel who are familiar with the dangers
involved.
Defects and Extraordinary Strains
If it may be assumed that the instrument can no longer be
operated safely, it must be removed from service and secured
against unintentional use.
Safe operation can no longer be relied upon:
• If the device demonstrates visible damage,
• If the instrument no longer functions, or if malfunctioning
occurs,
• After long periods of storage under unfavorable conditions,
e.g. humidity, dust or extreme temperature (see “Ambient Conditions” on page 64).
GMC-I Messtechnik GmbH9
Safety Warnings
!
1.1 Use for Intended Purpose
• The respective multimeter is a portable device which can be
held in the hand during the performance of measurements.
• Only those types of measurements described in chapter 5 may
be performed with the measuring instrument.
• The measuring instrument, including measurement cables and
plug-on test probes, may only be utilized within the specified
measuring category (see page 65 and the table on page 8
regarding significance).
• Overload limits may not be exceeded. See technical data on
page 60 for overload values and overload limits.
• Measurements may only be performed under the specified
ambient conditions. See page 64 regarding operating
temperature range and relative humidity.
• The measuring instrument may only be used in accordance
with the specified degree of protection (IP code) (see page 66).
1.2 Meanings of Danger Symbols
Warning concerning a point of danger
(attention: observe documentation!)
Warning concerning dangerous voltage at the measurement
input: U > 15 V AC or U > 25 V DC
1.3 Meanings of Acoustic Warning Signals
Voltage warning: > 1000 V (intermittent acoustic signal)
Current warning: > 11 A (continuous acoustic signal)
1 Display (LCD) (see page 13 for significance of symbols)
2
MAN / AUTO
Increase parameter values
shift key for manual/automatic measuring range selection
“Operating Mode” menu:
Selection of individual menu entries against
the direction of flow
3
ON / OFF | LIGHT
key for switching device and display illumination on and off
4 FUNC | ENTER multifunction key
“Operating Mode” menu: Acknowledge entry (ENTER)
ISO ON / OFF Insulation resistance measurement
U
Key for switching insulation resistance measurement on and off
5 Increase measuring range or move decimal point to the right (MAN function)
6
Rotary switch
for measuring functions (see page 14 for significance of symbols)
7 DAkkS calibration mark
8 Connector socket for ground / connected to ground
9 Connector socket for current measurement with automatic blocking
10 Connector socket for voltage, resistance, temperature, diode
capacitance measurement
with automatic blocking
11 DATA / MIN / MAX
Key for freezing, comparing, deleting the measured value, and for Min/Max function
Decrease values
“Operating Mode” menu:
12
MEASURE | SETUP
Select individual menu entries in flow direction
Key for switching between measuring and menu functions
13 ZERO | ESC
12GMC-I Messtechnik GmbH
Key for zero balancing
“Operating Mode” menu: Exit menu level and return to a higher level, exit
parameters entry function without saving
METRAHIT COIL
only – Function COIL:
UCOIL ON / OFFInterturn short-circuit measurement as long as
key is being pressed, see chapter 5.8.
14 Decrease measuring range or move decimal point to the left (MAN function)
15 Power pack connector jack
16 Infrared interface
and
Symbols used in the Digital Display
Battery full
Battery OK
Battery weak
Battery (almost) dead, U < 1.8 V
Data transmission to /
from the multimeter is active
IR interface in standby mode
(ready to receive starting commands)
1 Battery level indicator
2 ON: continuous operation (automatic shutdown deactivated)
3 MAN: manual measuring range selection active
4 Digital display with decimal point and polarity display
5 max/min: Min/Max value storage
6 DATA: display memory, “freeze measured value”
7 STORE: memory mode active
8 ISO: insulation resistance measurement active / selected test voltage
9 1: x current clip factor (transformation ratio)
10 IR: infrared interface display
11 Scale for analog display
12 Pointer for analog display (bar graph – pointer)
triangle appears: indicates overranging
13 Diode measurement selected
14 Selected type of current
15 TRMS measurement
16 Pt100(0): selected platinum resistance thermometer with
automatic recognition of Pt100/Pt1000
17 Type K: temperature measurement with type K (NiCr-Ni) thermocouple
18 sec (seconds): unit of time
19 REL: relative measurement with reference to offset
20 Unit of measure
21 ZERO: zero balancing active
22 Duty Cycle Measurement
(this function only available with customer-specific variant)
23 Warning regarding dangerous voltage: U > 15 V AC or U > 25 V DC
24 Continuity test with acoustic signal active
25 h (hours): unit of time
Position of the power pack adapter socket,
see also chapter 3.1.
Fuse for current measuring ranges (see chapter 9.3)
The device may not be disposed of with the trash.
Further information regarding the WEEE mark can be
accessed on the Internet at
www.gossenmetrawatt.com under the search term
WEEE (see also chapter 9.5).
Calibration mark (blue seal):
CAT II / III Measuring category II or I II device, see also
See also “Recalibration” on page 71.
“Measuring Categories and their Significance per IEC
61010-1” on page 8.
Continuous, doubled or reinforced insulation
Indicates EC conformity
▲
IR
▼
GMC-I Messtechnik GmbH15
Position of the infrared interface, window on the top of the
instrument
Initial Start-Up – Setup
Attention!
!
Note
3Initial Start-Up
3.1 Batteries
Be certain to refer to chapter 9.2 regarding correct battery
installation!
Momentary battery voltage can be queried in the “Info” menu (see
chapter 6.3).
Disconnect the instrument from the measuring circuit before opening the battery compartment lid in order to replace the batteries.
Operation with Power Pack (not included, see chapter 10.3)
Installed batteries are disconnected electronically if the NA X-TRA
power pack is used, and need not be removed from the
instrument. If rechargeable batteries are used, they must be
recharged externally.
If the external power supply is switched off, the device is switched
to battery operation without interruption.
3.2 Activation
Switching the Instrument On Manually
➭ Press the ON / OFF | LIGHT key until the display appears.
Power-up is acknowledged with a brief acoustic signal. As
long as the key is held depressed, all of the segments at the
liquid crystal display (LCD) are illuminated.
The LCD is depicted on page 13.
The instrument is ready for use as soon as the key is released.
Display Illumination
After the instrument has been switched on, background illumination can be activated by briefly pressing the ON / OFF | LIGHT key. Illumination is switched back off by once again pressing the same
key, or automatically after approximately 1 minute.
Switching the Instrument On with a PC
The multimeter is switched on after transmission of a data block
from the PC, assuming the “irStb” parameter has been set to “ir on” (see chapter 6.4).
However, we recommend using the power saving mode: “ir off”.
Electrical discharge and high frequency interference may
cause incorrect displays to appear, and may disable the
measuring sequence.
Disconnect the device from the measuring circuit. Switch the
instrument off and back on again in order to reset. If the
problem persists, briefly dislodge the battery from the
connector contacts (see also chapter 9.2).
3.3 Setting the Operating Parameters
Setting Time and Date
See “t iME” and “dAtE” parameters in chapter 6.4.
Display Modes for the Digital Display
Selection can be made from two different display modes
(see “0.diSP” parameter in chapter 6.4).
16GMC-I Messtechnik GmbH
ON / OFF
LIGHT
FUNC
ENTER
ON / OFF
LIGHT
Auto-Range
88.8.8.8
Illumination On
800.00
ON/ OFF
LIGHT
Illumination Off
800.00
ON/ OFF
LIGHT
0ff
ON/ OFF
LIGHT
long
(1 s)
3.4 Switching the Instrument Off
Switching the Instrument Off Manually
➭ Press the ON / OFF | LIGHT key until 0FF appears at the display.
Shutdown is acknowledged with a brief acoustic signal.
Automatic Shutdown
The instrument is switched off automatically if the measured value
remains unchanged for a long period of time (maximum measured
value fluctuation of approximately 0.8% of the measuring range
per minute or 1 C or 1 F per minute), and if none of the keys or
the rotary switch have been activated before a selected period of
time in minutes has elapsed (see “APoFF” parameter on page 53).
Shutdown is acknowledged with a brief acoustic signal.
Exceptions include:
Transmission and memory mode operation, continuous operation
and whenever a dangerous voltage is applied to the input
(U > 15 V AC or U > 25 V DC).
Disabling Automatic Shutdown
The instrument can be set to continuous operation.
➭ Simultaneously press the
The “Continuous On” function is indicated at by means of the
ON display to the right of the battery symbol.
The “Continuous On (on)” setting can be cancelled by changing the respective parameter or by switching the instrument off manually.
In this case the parameter will be reset to 10 min (see “APoFF” on
page 56).
and keys to this end.
Initial Start-Up – Setup
GMC-I Messtechnik GmbH17
Control Functions
4Control Functions
4.1 Selecting Measuring Functions and Measuring Ranges
4.1.1 Automatic Range Selection
The multimeter is equipped with auto-ranging for all measuring
functions, except for temperature measurement, and diode and
continuity testing. Auto-ranging is active as soon as the
instrument is switched on. The instrument automatically selects
the measuring range which allows for highest possible resolution
of the applied quantity. When the instrument is switched to
frequency measurement, the previously selected voltage
measuring range remains active.
AUTO-Range Function
The multimeter is switched automatically to the next higher range
at (3099 d + 1 d 0310 d), and to the next lower range at
(280 d - 1 d 2799 d).
With high resolution the multimeter is switched automatically to
the next higher range at (30999 d + 1 d 03100 d), and to the
next lower range at (2800 d - 1 d 27999 d).
4.1.2 Manual Measuring Range Selection
Auto-ranging can be deactivated and measuring ranges can be
selected manually in accordance with the following table by
pressing the MAN / AUTO button.
The desired measuring range can then be selected with the
or
key.
The instrument is automatically returned to range selection when
the MAN / AUTO key is pressed, the rotary switch is activated or the
instrument is switched off and back on again.
Overview: Auto-Ranging and Manual Range Selection
FunctionDisplay
MAN / AUTO
Range switching sequence for:
V:300 mV*
Hz: 300 Hz
:
300
or
MAN / AUTO
* Via manual measuring range selection only
A:
300A 3 mA 30 mA 300 mA 3A 10A
A : See chapter 5.9.3 und chapter 5.9.4
F:30 nF 300 nF 3F 30F 300F
M
:
300 k
@ISO
Return to automatic measuring range selection—
Manual mode active:
utilized measuring range is fixed
3 V 30 V 300 V 1000 V
3 kHz 30 kHz 300 kHz (Hz(U))
3 k
30 k
300 k
3 M
30 M
300 M
3 M
3000 M
30 M
MAN
MAN
The multimeter is held in the selected measuring range. If the
range limit is exceeded, 0L appears at the display. You should then
switch to the next higher measuring range with the help of the
key.
18GMC-I Messtechnik GmbH
Control Functions
Note
4.1.3 Quick Measurements
Measurements performed using a suitable fixed measuring range
are executed more quickly than those which utilize automatic
range selection. Quick measurement is made possible with the
following two functions:
• Manual measuring range selection, i.e. selection of the measuring
range with the best resolution (see chapter 4.1.2)
or
•With the DATA function (see chapter 4.4). In this way, the
appropriate measuring range is selected automatically after the
first measurement and the second measurement is executed
more quickly.
The selected measuring range remains active for the subsequent
series of measurements with these two functions.
4.2 Zero Offset / Relative Measurements
Zero balancing or a reference value for relative measurements can
be stored to memory depending upon deviation from the zero
point:
Deviation from zero
– with short-circuited measurement cables for V, , A
– with open input for capacitance unit of measure: F
0 to 200 digitsZERO REL
> 200 to 1500 digitsREL
Display
The relevant reference or correction value is deducted individually
for the respective measuring function as an offset from all future
measurements and remains in memory until deleted, or until the
multimeter is switched off.
Zero balancing and reference value adjustment can be used for
auto-ranging, as well as for manual measuring range selection.
Note: Zero balancing is not possibIe in switch position
M@UISO
.
Zero Balancing
➭ Plug the measuring cables into the instrument and connect the
free ends to each other, except for capacitance measurement
in which case the ends of the cables are not connected to
each other.
➭ Briefly press the ZERO | ESC key.
The instrument acknowledges zero balancing with an acoustic
signal, and the “ZERO REL” symbol appears at the LCD. The
value measured at the moment the key is pressed serves as a
reference value.
➭ Zero balancing can be cleared by once again pressing the
ZERO | ESC key.
As a result of TRMS measurement, the multimeter displays a
residual value of 1 to 10/35 digits with short-circuited
measurement cables as the zero point for V AC / I AC or
V(AC+DC) / I (AC+DC) measurements (non-linearity of the
TRMS converter). This has no influence on specified
accuracy above 1% of the measuring range (or 3% in the
mV, V(AC+DC) ranges).
Setting the Reference Value
➭ Plug the measuring cables into the instrument and measure a
reference value (max. 1500 digits).
GMC-I Messtechnik GmbH19
Control Functions
➭ Briefly press the ZERO | ESC key.
The instrument acknowledges storage of the reference value
with an acoustic signal, and the “ZERO REL” or the “REL”
symbol appears at the LCD. The value measured at the
moment the key is pressed serves as a reference value.
➭ The reference value can be cleared by once again pressing the
ZERO | ESC key.
Notes Regarding Relative Measurement
• Relative measurement effects the digital display only. The
analog display continues to read out the original measured value.
• In the case of relative measurement, F or AC quantities may
also appear as negative values.
4.3 Display (LCD)
4.3.1 Digital Display
4.3.2 Analog Display
Measured Value, Polarity
The analog display demonstrates the dynamic performance of a
moving-coil mechanism. This display is especially advantageous
for observing measured value fluctuation, and for balancing
procedures.
Display mode Pointer: the current measured value is tracked in
real-time.
The analog scale displays a negative range of 5 scale divisions for
the measurement of zero-frequency quantities, allowing for
precise observation of measured value fluctuation around zero. If
the measured value exceeds the negative range of 5 scale
divisions, polarity is reversed at the analog display.
Scaling of the analog scale is automatic. This is very helpful for
manual measuring range selection.
Measured Value, Unit of Measure, Type of Current, Polarity
The measured value with decimal and plus or minus sign appears
at the digital display. The selected unit of measure and current
type are displayed as well. A minus sign appears to the left of the
value during the measurement of zero-frequency quantities, if the
Overranging
Overranging in the positive range is displayed by means of the
right triangle symbol.
plus pole of the measured quantity is applied to the “” input.
The “0.d iSP” parameter can be used to determine whether lead-
ing zeros will be appear or be suppressed at the measured value
display (see chapter 6.4).
Refresh Rate
In the bar graph and pointer modes, the analog display is
refreshed 40 times per second.
Overranging
If the upper range limit of 3100 digits is exceeded “0L” (overload)
appears at the display.
Exceptions: “0L” appears at the display as of 1000.0 V in the case
of voltage measurement in the 1000 V range, as of 5.100 V for
diode testing, and as of 11.00 A in the 10 A range.
20GMC-I Messtechnik GmbH
Note
V, A, Hz
t [s]
100%
10%
Activated
Reactivated
Save
Save
3000 digits
of Measuring Range
F, %
Control Functions
4.4 Measured Value Storage: DATA (auto-hold / compare)
An individual measured value can be automatically “frozen” with
the DATA function (auto-hold). This is useful, for example, when
contacting the measuring points with the test probes requires
your full attention. After the measuring signal has been applied
and the measured value has settled in in accordance with the
“condition” listed in the table below, the measured value is frozen
at the digital display and an acoustic signal is generated. The test
probes can now be removed from the measuring points, and the
measured value can be read from the digital display. If the
measuring signal falls below the value specified in the table, the
function is reactivated for storage of the next value.
Measured Value Comparison (DATA Compare)
If the currently frozen value deviates from the first saved value by
less than 100 digits, the acoustic signal is generated twice. If
deviation is greater than 100 digits, only one brief acoustic signal
is generated.
The DATA function has no effect on the analog display, at
which the current measured value continues to appear.
DATA
Function
activatebriefblinksonce
save
(stabilized
measured
value)
reactivate
1)
Press
DATA /
Min/Max
However, when the digital display is “frozen”, the decimal
point is fixed as well (fixed measuring range, symbol: MAN).
The selected measuring range should not be manually
changed as long as the DATA function is active.
The DATA function is deactivated by pressing and holding the
DATA/MIN/MAX key (approx. 1 second), when the measuring
function is changed or when the instrument is switched off and
back on again.
GMC-I Messtechnik GmbH21
change to
Min/Max
exitlong
1)
Reactivation results from falling short of specified measured value limits.
2)
Two acoustic signals are generated the first time a measured value is saved as
a reference value. For subsequent data hold, two acoustic signals are only
generated if the currently frozen value deviates from the
than 100 digits.
Key: MV = measured value, R = measuring range
briefSee table in chapter 4.4.2
Condition
Measuring
Function
V, A, F, Hz, %> 10%
V, A, F, Hz, %< 10%
Measuring
Signal
of R
0L
of R
= 0L
Response from Instrument
Display
MV
Digital
is displayed
stored
MV
is cleared is cleared
first
saved value by less
DATA
static
blinks
Acou-
stic
once
twice
twice
2)
Control Functions
Example
The voltage measuring range is set manually to 30 V.
The first measured value is 5 V and is stored to memory because
it is greater than 10% of the measuring range (= 3 V), and is thus
reliable above the background noise level. As soon as the
measured values drops to less than 10% of the measuring range,
i.e. amounts to less than 3 V which corresponds to removal of the
test probes from the measuring point, the instrument is ready to
store a new value.
4.4.1 Function DATA during insulation measurement *
This DATA function differs from the standard DATA-Compare
function.
In the interference voltage measurement function V(ac+dc)1M
the special DATA function for insulation measurement is activated
by pressing the DATA key – DATA blinks in the display. During inter-
ference voltage measurement, no DATA values are determined or
stored yet. By briefly pressing the FUNC | ENTER key, the insulation
measurement is activated and the test voltage is applied.An intermittent acoustic signal indicates to the user that the test voltage is
applied (accompanied by a blinking of the ISO symbol in the display). After applying the test probes, a check is being made as to
whether a valid measured value is present.As soon as a stable
measured value appears in the display, the DATA value is stored
(frozen in the display). A long acoustic signal indicates that the
measurement is completed, the test voltage is switched off again,
DATA no longer blinks and the user can read and note down the
measured value. Byrepeatedlypressing the FUNC | ENTER key, the
test voltage is applied again and DATA is activated (DATA blinks in
the display).
If no DATA value is determined or if OL is shown in the display, the
test voltage is automatically switched off after approx. 10 seconds. In order to deactivate the DATA function, the FUNC | ENTER
key must be pressed and held.
* This function is especially available for variant METRAHIT ISO AERO,
for other variants as from firmware version 1.18.
22GMC-I Messtechnik GmbH
4.4.2 Saving Minimum and Maximum Values – MIN/MAX Function
Note
Minimum and maximum measured values applied to the
measuring instrument’s input after the Min/Max function has been
activated can be “frozen” at the display. The most important use
of this function is the determination of minimum and maximum
values during long-term measured value observation.
The Min/Max function can be activated in all measuring functions.
The Min/Max function has no effect on the analog display, at
which the current measured value continues to appear.
Apply the measured quantity to the instrument and set the
measuring range with the MAN / AUTO key before activating the
Min/Max function.
The Min/Max function is deactivated by pressing and holding the
DATA/MIN/MAX key (approx. 1 second), when the measuring
function is changed or when the instrument is switched off and
back on again.
As opposed to the DATA function, the Min/Max function can
also be used for temperature measurement.
Function
Min/Max
activate
and save
save and
display
return to 1
stoplongare deleted
Press
DATA /
Min/Max
1
2 x briefare saved
brief
2
brief
3
brief
Min. and Max.
Measured Values
storage continues
in background,
new min. and max.
values are displayed.
same as 1,
stored values are not
deleted
Control Functions
Response from Instrument
Display
Measured
Valu e
Digital
current
measured
value
saved min.
value
saved max.
value
same as 1
current
measured
value
Max.
Min.
max
and
min
min.
max.
same
as 1
is
deleted
Acous-
Signal
tic
twice
once
once
once
twice
GMC-I Messtechnik GmbH23
Loading...
+ 53 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.