Gossen MetraWatt MetraHit 28s, MetraHit 29s User Manual

METRA 28S and 29S Precision Digital Multimeter, Pow er Meter
3-348-866-03
5/8.00
GOSSEN-METRAWATT GMBH
-Calibration Certificate
DQSCertifiedper
QUALITY MANAGEMENT SYSTEM
Applications
METRAHitâ28S and 29S multimeters are high performance pre­cision instruments for the lab, service and training. With a display range of 300,000 digits as well high level accuracy and long-term stability, they fulfill all calibration and R&D labora­tory demands. Battery operation allows for on-site precision ser­vicing and calibration.
Features
Convenient Triple Display
The current measurement value, as well as two other values, are updated and displayed, e.g. minimum and maximum value. The usual switching back and forth between display values is no longer necessary. Display resolution for the main display value can be varied from 4¾ places for AC to 5¾ places for DC (METRAHit
â
29S: from 2¾ to 5¾ places). The measurement rate, adjustable from 1 ms to 0.5 s, allows for high-speed recorder, precision data logger and mean value recorder functions with the help of METRAwin
â
10/METRAHitâ (optional PC software).
High Resolution and Precision
5¾ places, or 309,999 digits, allow for cost effective calibration of tested devices and components, as well as precise reference measurements.
Effective Value for Distorted Waveforms
The implemented measurement process provides for effective value measurement (TRMS) independent of the waveform up to 100 kHz.
Additional Functions
Continuity testing with acoustic signal, event counting (number of events and event durations), counts zero crossings, stopwatch, data comparison and wide-range capacitance measurement. The integrated temperature measurement function allows for the con­nection of thermocouples and platinum resistance thermometers.
Overload Protection
The instrument is protected from overload in the voltage range. Acoustic signalling occurs if measuring range lim its are exceeded. The FUSE display indicates a defective fuse in the current range.
Calibration
METRAHit
â
28S and 29S multimeters are shipped with a DKD calibration certificate. They can be automatically calibrated with the CS92E calibration system via the RS232 data interface.
Infrared Measurement Data Transmission *
These devices include a serial, duplex data interface for remote control and measurement value transmission via infrared light.
METRAHit
â
28S and 29S
• 300,000 digits and triple display
• Precision multimeter (V, dB, A, Ω, F, Hz, F, °C/°F)
• Current measurement with (clip-on) current transformer: a transformation ratio of 1000:1 or 10000:1 is utilized for display purposes
• Integrated quartz movement for recording of MIN/MAX values relative to real-time
•AUTOSELECT automatic measured quantity recognition (V, and F)
METRAHit
â
29S
• Power and energy measurements (W, VAr, VA, Wh, PF) with analog signals or energy measurement with pulses, current measurement value, mean values and peak power values are displayed
• Power disturbances recording
• Large, 128 kB measurement value memory
METRA 28S and 29S Precision Digital Multimeter, Pow er Meter
2 GOSSEN-METRAWATT GMBH
Automatic Blocking System (ABS) *
Automatic blocking of connector sockets prevents incorrect con­nection of measurement cables, and selection of incorrect mea­suring quantities. Thus hazards for the operator, the instrument and the system are substantially reduced, and in many cases entirely eliminated.
Automatic Measurement Value Storage *
The DATA HOLD function allows for storage of the digitally dis­played measurement value. A patented process assures that the actual measurement value, and not random values, are stored when rapid measurement magnitude changes occur. The stored measurement value appears at the digital display.
* Protected by patent
Sampling Rate
The sampling rate determines the time interval after which the respective measurement value is transmitted to the interface, or saved to measurement value memory. Depending upon the measured quantity, sampling rates can be set in steps of 1, 2 or 5 from 0.05 s to 10 min.
Additional Functions, METRAHitâ29S
Power Measurement
The METRAHitâ29S is a compact high performance measure­ment instrument for direct and alternating current for single and three-phase current measurements. The curr ent path can be con­nected directly, or with a clip-on current transformer. If a current transformer is connected to the multimeter (mA or A input), all current and power displays appear with the correct value in accordance with the selected transformation ratio. The utilized current transformer mu st have a tran sforma tion ratio of 1000:1 or 10,000:1. Universal power measuring functions include the mea surement of active, reactive, and apparent power, as well as power factor and energy . Furthermore, mean and peak values (e.g. a 15 minute max. value) can be measured during a predefined period of time. Type WZ12D, Z3511, Z3512 and Z3514 current clips are suitable for use with the instrument.
Power Disturbance Recording
The instrument is capable of continuously recording line voltage and power disturbances. The following power disturbances can be acquired: violation of predetermined limit values, power failure and positive or negative pulses. Each of these events is stored to memory and can be queried at a later point in time. The type of event, the time of its occurrence and its duration (except for pulses) are displayed.
If memory mode is deactivated, approximately 250 events can be stored to volatile memory and read out at the display. These data remain in storage until the instrument is switched to a different operating mode. If memory mode is activated, the events are recorded at the instru­ment as a non-compressed data file which is uploaded to a PC at a later point in time.
Memory Mode
The instrument is equipped with a 128 kB measurement value memory which is synchronized with a quartz movement, and which has room for 13,000 to 60,000 measurement values depending upon configuration. Data can be stored to intermedi­ate memory or uploaded directly to a PC. The system records data relative to real-time which allows for use as a real time data logger. Depending upon the measured quantity, sampling rates can be adjusted in steps of 1, 2 or 5 within a range of 0.5 ms to 10 min. Selected measurement values can also be stored to memory by pressing a key.
The contents of the memory can be read out with the help of a PC which has been connected to the multimeter via the METRAHit
â
BD232 IR adapter, and with METRAwinâ10/
METRAHitâ analysis software.
Applicable Regulations and Standards
Standard Equipment
1 Multimeter 1 Protective case for operation under adverse conditions.
A cover made of pliable rubber including support bar protects the device from damage due to impacts and falls. The rubber material assures a secure stance for the instrument even on
vibrating surfaces. 1 KS17 cable set 2 Batteries 1 Operating Instructions 1 DKD calibration certificate
Guarantee
3 years material and workmanship 1 ... 3 years (depending on application) for calibration
IEC 61010-1 DIN EN 61010 Part 1 VDE 0411-1
Safety regulations for electrical measurement, control, regulating and laboratory devices
DIN 43751 Digital measuring instruments DIN EN 50081 Part 1
Generic standard for interference emission; residential, business and light industry
DIN EN 50082 Part 1
Generic standard for interference immunity; residential, business and light industry
VDI/VDE 3540 Reliability of me asurement, control and reg ulating devices DINEN60529
DIN VDE 0470 Part 1
Test instruments and procedures – level of protection provided by enclosures (IP code)
GOSSEN-METRAWATT GMBH 3
METRA 28S and 29S
Precision Digital Multimeter, Power Meter
METRAHitâ28S and 29S Characteristic Values
1)
Display: 5¾ places for DC, 4¾ places for AC, other resolution and sampling rates can be selected for storage and transmission in the rAtE menu.
2)
Stopwatch; format: mm:ss:hh, where m=minute, s=second and h=hundredth second, max.: 99:59:59
3)
Lowest measurable frequency with sinusoidal measurement signal symmetric to zero point
4)
at 0 ° ... + 40 °C
5)
Values < 100 digits are suppressed, 16 ... 45 ... 65 Hz ... 100 kHz sine. See page 4 for influences.
6)
12 A – 5 min, 16 A – 30 s
7)
ZERO is displayed when the “Zero Balancing” function is active.
8)
Range 300mV : UE=50mV
eff/rms
. . . 300mV
eff/rms
3V :UE=0,3V
eff/rms
. . . 3 V
eff/rms
30 V : UE=3V
eff/rms
. . . 30 V
eff/rms
300 V : UE=30V
eff/rms
. . . 300 V
eff/rms
1000 V : UE=300 V
eff/rms
. . . 1000 V
eff/rms
for voltages > 100 V: power limiting at 3 · 106V · Hz
9)
Plus sensor deviation
10)
Without integrated reference junction: inherent deviation
±2K
11)
For U > 10% of the measuring range
Key: D = Digit
Meas.
Function
Measuring Range
Resolution at
Measuring Range Upper Limit
Input Impedance
Inherent Deviation at highest resolution
under reference conditions
Overload Capacity
4)
Meas.
Function
±(...% of rdg. + ... %
of range + ... D)
±(...% of rdg. . + ... D)
300000
1)
30000 1)3000
1)
5)
Value Duration
V
300 mV 1 µV10µV>20M 5MΩ // < 50 pF 0.02 + 0.005 + 5
10)
0.5 + 30 1050 V
DC AC eff
sine
cont.
V
3V 10µV100µV11M 5MΩ // < 50 pF 0.02 + 0.005 + 5 0.2 + 30
30 V 100
µV1mV 10M 5MΩ // < 50 pF 0.02+ 0.005 + 5 0.2 + 30
300 V 1mV 10mV 10M
5MΩ // < 50 pF 0.02+ 0.005 + 5 0.2 + 30
1000 V 10 mV 100 mV 10 M
5MΩ // < 50 pF 0.02+ 0.005 + 5 0.2 + 30
dB
see table following page same as V ± 0.1 dB
11)
dB
Volt. Drop, approx. for upper range limit B
5)
A
300 µA 1 nA 10 nA 160 mV 160 mV 0.05 + 0.02 +5 0.5 + 30
0.36 A cont.
A
3 mA 10 nA 100 nA 160 mV 160 mV 0.05 + 0.01 + 5 0.5 + 30
20 mA 30 mA
100 nA 1
µA 170 mV 170 mV
0.02 + 0.01 +5
0.5 + 30
0.05 + 0.01 +5
300 mA 1
µA1A 300 mV 300 mV 0.1 + 0.01 +5 0.5+ 30
3A 100
µA 110 mV 110 mV 0.2 + 0.05 + 5 0.5+ 30
10 A
6)
cont.
10 A 1 mA 350 mV 350 mV 0.2 + 0.05 + 5 0.5+ 30
Open-circuit voltage
Measuring current
at upper ranger limit
±(...% of rdg. + ... % of range + ... D)
300 1m 0.6 V max. 250 µA 0.05 + 0.01 + 5
7)
500 V
DC AC eff
sine
10 min
3k 10mΩ 0.6 V max. 45 µA 0.05 + 0.01 +5
7)
30 k 100m 0.6 V max. 4.5 µA 0.05 + 0.01 + 5
300 k
1 0.6 V max. 1.5 µA 0.05 + 0.02+ 5
3M
10 0.6 V max. 150 nA 0.1 + 0.02 + 5
30 M
100 0.6 V max. 15 nA 1 + 0.2 + 5
300 0.1 max. 3 V max. 1 mA 1 + 0 + 3
300 mV 100 µV max. 3 V max. 1 mA 0.2 + 0 + 3
3V 100µV max. 3 V max. 1 mA 0.2 + 0 + 3
Discharge resistor U
0 max
±(...% of rdg. + ... % of range)
F
3nF 1pF 10M 3 V 1.0 + 0.2
7)
500 V
DC AC eff
sine
10 min
F
30 nF 10 pF 10 M 3 V 1.0 + 0.2
7)
300 nF 100 pF 1 M 3 V 1.0 + 0.2
3
µF 1 nF 100 k 3 V 1.0 + 0.2
30
µF10nF11k 3 V 1.0 +0.2
300
µF 100 nF 2 k 3 V 5.0 + 1
3000
µF1µF2k 3 V 5.0 + 1
30000
µF1µF2k 3 V 5.0 + 1
f
min
3)
±(...% of rdg. + ... D)
Hz
300.000 Hz 0.001Hz 1 Hz 0.05 + 1
8)
1000 V
cont.
Hz
3.00000 kHz 0.01 Hz 1 Hz 0.05 + 1
8)
1000 V
300.000 kHz 1 Hz 1 Hz 0.05 + 1
8)
300 V
30 V
100 min
2)
10 ms ±15 D
DC1000V
AC 750V
±(...% of rdg. + ... D)
°C/°F
Pt 100/
Pt 1000
– 200.0 ... +100.0
°C
0.1
°C
0.5K+ 3
9)
500 V
DC eff
sine
10 min
°C/°F
+ 100.0 ... +850.0
°C
0.2% + 3
9)
K
NiCr-Ni
– 270.0 ...
+1372.0
°C
0.1
°C 0.7 + 3
9. 10)
1050 V
DC eff
sine
J
Fe-CuNi
– 210.0 ...
+1200.0
°C
0.1
°C 0.8 + 3
9. 10)
Loading...