GOSSEN METRAWATT METRA HIT 1A, METRA HIT 2A User Manual

METRA HIT 1A and METRA HIT 2A Analog Multimeters
Voltage measurement:
0...0.15, 0...0.5 V= , 0 ...1.5, 5, 15, 50, 150, 500 V=/~
Input resistance: 20 k/V=, 4 kΩ/V~
Current measurement: 0 ... 50 µA=, 0 ... 0.5, 5, 50, 500 mA, 5 A=/~
Extended current measurement, METRA HIT 2A: ... 15 A=/~
Resistance measurement: 1 ... 1 M (4 ranges)
Capacitance: 200 mF ... 2 µF (4 ranges)
Level: -15 ... +56 dB (6 ranges)
3-349-304-03
2/7.04
5 Ampere 15 Ampere / Class 2.5
A large number of measuring ranges allows for universal use of the instruments, for example in electronics, digital technology and a wide variety of general electrical applications.
Safety Features
The connector jacks and plugs included with the measurement cables which are available as accessories are protected against inadvertent contact.
Measuring Range Selector Switch
All measuring ranges are selected with a rotary switch.
Rugged Mechanical Design
The rugged plastic housing and the spring loaded bearing jewels in the moving-coil mechanism protect the instrument from damage due to mechanical stressing. Especially effective protection is provided by the protective rubber cover and the ever-ready case, both of which are available as accessories.
Applicable Regulations and Standards
IEC/EN 61010-1:2001/ VDE 0411-1:2002
EN 60529 VDE 0470, part 1
DIN EN 61326 VDE 0843, part 20
Safety requirements for electrical equipment for measurement, control and laboratory use
Test instruments and test procedures, degrees of protection provided by enclosures (IP code)
Electrical equipment for control technology and laboratory use – EMC requirements
500 V CAT II
IEC/EN 61010-1, Second Edition
Characteristic Values
Voltage Measuring Ranges
Voltage Output
0.15 V 3.15 k
0.50 V 10.00 k
1.50 V -15 ... + 6 dB 31.50 k 6.50 k
5.00 V - 5 ... + 16 dB 100.00 k 20.00 k
15.00 V + 5 ... + 26 dB 315.00 k 65.00 k
50.00 V +15 ... + 36 dB 1.00 M 200.00 k
150.00 V +25 ... + 46 dB 3.15 M 650.00 k
500.00 V +35 ... + 56 dB 10.00 M 2.00 M
1)
0 dB 0.775 V in the 1.5 V range, 0 dB 1 mW to 600
Voltage related input resistance: for : 20.0 kΩ/V
1)
METRA HIT 1A: Current Measuring Ranges
Current
50.00 µA 0.158 V
0.50 mA 1.15 V 1.00 V
5.00 mA 1.25 V 1.25 V
50.00 mA 1.25 V 1.25 V
500.00 mA 1.85 V 1.85 V
5000.00 mA 1.73 V 1.73 V
Voltage Drop, Approx.
Internal Resistance, Approx.
for : 4.0 kΩ/V
GOSSEN METRAWATT GMBH
METRA HIT 1A and METRA HIT 2A Analog Multimeters
METRA HIT 2A: Current Measuring Ranges
Current
50 µA 0.158 V
1.5 mA 1.16 V 1.21 V
15 mA 1.25 V 1.25 V
150 mA 1.25 V 1.25 V
1.5 A 1.95 V 1.95 V
15 A 0.43 V 0.49 V
Voltage Drop, Approx.
Resistance Measuring Ranges
Resistance Scale Span
Value at Scale
Middle (R
x 1 1 Ω ... 1 kΩ 18.00 83 mA
x 10 10 ... 10 k 180.00 8.3 mA
x 100 100 ... 100 k 1.8 0 k 0.83 mA
x 1000 1k ... 1 M 18.00 k 0.083 mA
2)
At a battery voltage of 1.5 V
)
i
Max. Measuring
max
Approx.
2),
Current I
Capacitance Measuring Ranges
Capacitance
3)
µF x 1000 2000 ... 200 000 µF
µF x 100 200 ... 20000 µF
µF x 10 20 ... 2000 µF
µF x 1 2 ... 200 µF
3)
Rough measurements in the resistance measuring ranges, determination of measured values with comparative scale
Scale Span
Reference Conditions
Ambient temperature
+23°C ±2K
Normal position of use Horizontal
Frequency 40 ... 60 Hz Line Voltage For : sinusoidal Relative humidity 40 ... 60% The instrument is equipped with half-wave rectification and is
RMS calibrated. It evaluates the arithmetic mean value of a half-wave and displays different values for pulsating voltage and current depending upon polarity.
Other influencing quantities: per IEC/EN 60 051
Ambient Conditions
Storage temp. range -25 to 65° C (without battery) Relative humidity
Up to 75%, no condensation allowed
METRA HIT 2A: Accuracy (under reference conditions per IEC/EN 60 051)
Class 2.5 for and , max. additional intrinsic error in the 15 A range: ±2%; in the 1.5 V~ range: +1/–2,5%, class 2.5 for (error relative to 52 mm scale length)
METRA HIT 2A: A
dditional accuracy class influence and Nominal Ranges of Use
Ambient temp. For : 0 ... +23 ... +40° C
For : +13 ... +23
... +35° C
Frequency 1.5 ... 500 V ranges:
35 ... 40 ... 60 ... 1000 Hz
1.5 mA ... 1.5 A ranges: 35 ... 40 ... 60
... 1000 Hz 15 A range: 40 ... 45 ... 60 ... 1000 Hz
Other influencing quantities: per IEC/EN 60 051
Power Supply
Battery for resistance measurement One 1.5 V mignon cell per IEC LR6/R6
(size AA), leak-proof
Electrical Safety
Safety class II per IEC/EN 61010-1:2001/
VDE 0411-1:2002 Measuring category II Nominal voltage 500 V Test voltage 3.5 kV~ Fouling factor 2 Fuse links Replaceable
METRA HIT 1A
: F1: FF630mA/700V AC (50 kA), 6.3 x 32 F2: FF6.3A/500V AC (50 kA), 6.3 x 32 METRA HIT 2A
: F1: FF1.6A/700V AC (50 kA), 6.3 x 32 F2: FF16A/500V AC (50 kA), 6.3 x 32
Soldered permanently into place
750 mA / 600 V AC
METRA HIT 1A: Overload Capacity
Range To Range To
0.15 V –
0.5 V –
1.5 V –
5.0 V –
15.0 V –
50.0 V –
150.0 V –
500.0 V –
50.0 µA –
0.5 mA –
5.0 mA –
50.0 mA –
500.0 mA –
5000.0 mA –
20 V
50 V 100 V 150 V 250 V 250 V 300 V 500 V
1.0 mA
5.0 mA
10.0 mA
70.0 mA
500.0 mA
3.0 A
5.0 A
1)
2)
2)
2)
2)
2)
2)
2)
3)
— —
1.5 V
5.0 V
15.0 V
50.0 V
150.0 V
500.0 V
0.5 mA
5.0 mA
50.0 mA
500.0 mA
5000.0 mA
500.0 mA
METRA HIT 2A: Overload Capacity
Range To Range To
0.15 V –
0.5 V –
1.5 V –
5.0 V –
15.0 V –
50.0 V –
150.0 V –
500.0 V –
50 µA –
1.5 mA – 15 mA –
150 mA –
:
1.5 A –
15 A –
1)
Fuse F1 blows in the event of overloading.
2)
Ranges are protected against overloading with PTC resistor
3)
Max. 1 minute
20 V
50 V 100 V 150 V 250 V 250 V 300 V 500 V
1.0 mA
5.0 mA
20.0 mA
150.0 mA
1.2 A
1.5 A
12.0 A
15.0 A
1)
2)
2)
2)
2)
2)
2)
2)
3)
3)
— —
1.5 V
5.0 V
15.0 V
50.0 V
150.0 V
500.0 V
1.5 mA 15 mA
150 mA
1.5 A
15 A
150.0 mA
— 25 V 50 V
150 V 250 V 300 V 500 V
5.0 mA
10.0 mA
70.0 mA
3.0 A
5.0 A
— 25 V 50 V
150 V 250 V 300 V 500 V
5.0 mA
20.0 mA
1.2 A
1.5 A
12.0 A
15.0 A
2)
2)
2)
2)
2)
2)
3)
2)
2)
2)
2)
2)
2)
3)
3)
2 GOSSEN METRAWATT GMBH
METRA HIT 1A and METRA HIT 2A
Analog Multimeters
Electromagnetic Compatibility (EMC)
Interference emission EN 61326:2002 class B Interference immunity EN 61326:2002
IEC 61000-4-2:
8 kV atmospheric discharge IEC 61000-4-3:
3 V/m
Mechanical Design
Scale length A, V – 0 ... 5.0: approx. 83 mm
A, V – 0 ...15.8: approx. 77 mm A, V~ 0 ... 5.0: approx. 67 mm A, V~ 0 ... 15.8: approx. 59 mm Ω∞... 0: approx. 52 mm dB – 15 ... + 6: approx. 42 mm
Dimensions 92 x 126 x 45 mm Weight Approx. 0.3 kg without battery Protection Housing: IP 40, terminals: IP 20 per
EN 60529/VDE 0470, part 1
Table Excerpt Regarding Significance of the IP Code
st
(1
IP XY digit, X)
Protection against pene-
tration of solid particles
0 Not protected 0 Not protected 1 50.0 mm dia. 2 12.5 mm dia. 3 2.5 mm dia. 4 1.0 mm dia. 5 Dust protected 5 Jet-water 6 Dust-proof 6 Powerful water jets
IP XY
(2nd digit, Y)
Protection against pene-
tration by water
1 Vertical dripping 2 Dripping (15° inclination) 3 Spray water 4 Splashing water
Standard Equipment
1 analog multimeter (without battery or cable set) 1 set operating instructions
Order Information
Description Type Article Number
Analog multimeter, 5 A / 500 V METRA HIT 1A M100A
Analog multimeter, 15 A / 500 V METRA HIT 2A M101A
Cable set consisting of 1 pair KS17-2 GTY 3620 034 P0002
Ever-ready case F809 GTY 3172 083 P01
Protective rubber cover GH185 GTY 3171 185 P01
Current measuring adapter for safe, trouble-free measurement of current consumption via the mains plug of connected power consumers SM16 GTM 9070190E0002
F1 fuse link (shipped in package of ten) for METRAHIT 1A FF 630 mA / 700 V Z109J
F2 fuse link (shipped in package of
10) for METRAHIT 1A FF 6.3 A / 500 V Z109K
F1 fuse link (shipped in package of
10) for METRAHIT 2A
F2 fuse link (shipped in package of
10) for METRAHIT 2A FF(UR) 16 A / 600 V AC Z109A
For additional information regarding accessories please see:
Measuring Instruments and Testers catalog
• www.gossenmetrawatt.com
FF(UR) 1.6 A / 700 V AC
Z109E
Accessories
F809 Ever-Ready Case with Cable Compartment – For tabletop operation
– The cover can be snapped to the bottom of the case when using the shoulder strap.
GOSSEN METRAWATT GMBH 3
METRA HIT 1A and METRA HIT 2A Analog Multimeters
Current Measuring Accessories Suitable for
All current sensors and transformers are equipped with a connector cable (1.2 to 1.5 m long) with 4 mm safety banana plugs
Type Designation Measuring Range Measur-
AC Current Transformer with Current Output
WZ12A Clip-on current transformer 15 ... 180 A~ 300 V /
WZ12D Clip-on current transformer 30 mA ... 150 A~ 300 V /
Z3511 Clip-on current transformer 4 ... 500 A~ 600 V /
Z3512 Clip-on current transformer 0.5 ... 1000 A~ 600 V /
Z3514 Clip-on current transformer 1 ... 2000 A ~ 600 V /
Without restriction
ing
CAT III
CAT III
CAT III
CAT III
CAT III
Cat.
Max. to Wire Dia.
15 mm 1 mA/A 45 ... 65
15 mm 1 mA/A 45 ... 65
30 x 63 mm
52 mm 1 mA/A 30...48 ... 65
64 x 150 mm
Transformation Ratio
1 mA/A 48 ... 65
1 mA/A 30...48 ... 65
Frequency Range
... 400 Hz
... 500 Hz
... 1 kHz
... 5 kHz
... 5 kHz
Intrinsic Error ±(% rdg. + ...)
3% Z219A ●●
2.5% +0.1 mA Z219D As of
3% +0.4 A GTZ 3511
0.5% ... 0.7% GTZ 3512
0.5% +0.1 A GTZ 3514
Article Number
000 R0001
000 R0001
000 R0001
METRA HIT
1A 2A
5 A
As of 50 A
As of 50 A
As of
500 A
As of
5 A
As of 50 A
As of
0 A
As of
500 A
GH185 Protective Rubber Cover WZ12 Clip-On Current Transformer SM16 Current Measuring Adapter
Edited in Germany • Subject to ch ange without notice • PDF version available on the Internet
KS17-2 Cable Set
Cable set with permanently mounted test probes and contact­protected angle plugs for safe measuring. Hook clips, alligator clips or push-on lugs can be attached to the test probes for special measuring tasks.
GOSSEN METRAWATT GMBH Thomas-Mann-Str. 16 - 20 90471 Nürnberg,