Gossen MetraWatt METRACLIP 63, METRACLIP 64 Operating Instructions Manual

Page 1
Operating Instructions
METRACLIP63 / 64
Milliampere Current Clamp for Measuring Fault, Leakage and Stray Current
3-349-488-03
1/10.08
Page 2
1 CT INPUT: Current clamp connector socket 2 REC OUT: Connector sockets for data logger
3 LCD panel
4 AC / DC: Selector key for direct or
alternating current
5 Range: Measuring range selector key 6
NOT AUTO POWER OFF
: Key for deactivating auto-shutdown
7 POWER: On/off key 8 0SET: Zero balancing key
9 DATA HOLD: Key for saving measured values
Scope of Delivery
1 measuring instrument 4 LR6 batteries (AA) 1 current clamp 1 carrying pouch 1 set of operating instructions
1
2
3
4
5
6
7
8
9
Measuring Instrument
1
Current Clamp
METRACLIP 63
Current Clamp
METRACLIP 64
1 Arrow must point in
direction of current
flow 2 Safety collar 3 Locking lever 4 Slider for opening
the clamp
Description in section 4.1
1 Arrow must point in
direction of current
flow 2 Safety collar 3 Key for opening the
clamp
2
3
4
3
1
2
Page 3
GMC-I Messtechnik GmbH 3
Table of contents
1 Safety Precautions .............................3
1.1 General .......................................................3
1.1.1 Meaning of Symbols on the Instrument ..........4
2 Applications ........................................4
3 Initial Start-Up ....................................4
4 Operation ............................................5
4.1 Using the METRA CLIP 63 Current Clamp .....5
4.2 Direct Current Measurement .........................5
4.3 Alternating Current Measurement .................6
4.4 Measuring Leakage and Stray Current ...........6
4.5 Using a Data Logger .....................................7
5 Characteristic Values ..........................8
6 Maintenance .....................................10
6.1 Housing .....................................................10
7
Repair and Replacement Parts Service, DKD Calibration Laboratory and Rental
Instrument Service .......................................10
8 Product Support ................................11
1 Safety Precautions
1.1 General
You have selected an instrument which provides you with high levels of safety.
This instrument fulfills all requirements of applicable European and national EC
guidelines. We confirm this with the CE mark. The relevant declaration of conformity can be obtained from GMC-I Messtechnik GmbH.
The METRACLIP 63 / 64 current clamp is manufactured and tested in accordance with safety regulations IEC 61010-1/EN 61010-1/VDE 0411-1 and IEC 61010-2-032/EN 61010-2-032/ VDE 0411-2-032
. When used for its
intended purpose, safety of the operator, as well as that of the instrument, is assured. Their safety is however not guaranteed, if the instrument is used improperly or handled carelessly.
In order to maintain flawless technical safety conditions, and to assure safe use, it is imperative that you read the operating instructions thoroughly and carefully before placing your instrument into service, and that you follow all instructions contained therein.
Observe the following safety precautions:
Avoid excessive mechanical stress such as impact and vibration, as well as high temperatures and humidity, because this may result in damage to the instrument.
Do not use this instrument in close proximity to acids, alkaline salts, organic solutions, corrosive gases etc.
Do not use this instrument in proximity to devices which generate strong magnetic and/or electrical fields, because this may result in malfunctioning.
Do not use this instrument in operating environments where dust or splash water occur.
Attention!
!
Do not perform measurements at cables with voltages of greater than 600 V AC. Before using the instrument, test the voltage of the electrical circuit to be measured.
Do not perform measurements on bare wires! Do not subject the input and output sockets to voltages of greater than
10 V AC.
The current clamp may not be used:
If the device, the housing or the cable is damaged
If the battery compartment lid has been removed
For current values which are impermissible high for the current clamp
In the rain, with wet hands, in the presence of splashing water or in moist environments
Page 4
1.1.1 Meaning of Symbols on the Instrument
Meaning of Symbols on the Instrument
Warning concerning a source of danger (attention: observe documentation)
EC label of conformity
This device may not be disposed of with the trash. Further information regarding the WEEE mark can be accessed on the Internet at www.gossenmetrawatt.com by entering the search term WEEE.
CA
TII/III
Measuring category
II
or
III
device
Closing the clamp around dangerously active conductors is permissible.
2 Applications
The instrument is capable of measuring direct and alternating current, and is equipped with a connectable current clamp.
In the case of direct current measurement, conventional current clamps are not capable of performing accurate measurement in the light current range due to influences resulting from magnetization and geomagnetism. This device is a current clamp measuring instrument which is capable of very accurate measurement of direct and alternating current thanks to the use of highly sensitive materials and a magnetization method.
Attention!
!
The specified maximum input quantities may not be exceeded in order to prevent damage to the instrument and injury to the user: voltage range: 600 V AC/DC, current range: 10 A AC/DC.
3 Initial Start-Up
Batteries
Your instrument requires four 1.5 V batteries (size AA) in accordance with IEC LR6.
Make sure that no battery leakage has occurred before initial start-up, as well as after long periods of storage. Continue to inspect the batteries for leakage at short, regular intervals.
If battery leakage has occurred, carefully and completely clean the electrolyte from the instrument with a damp cloth, and replace the batteries before using the instrument.
If the “B” symbol appears at the LCD panel, the batteries should be replaced as soon as possible.
Battery Replacement
Attention!
!
Disconnect the current clamp from the measuring circuit before opening the device in order to replace the batteries!
Remove the screw from the battery compartment lid with a screwdriver. Push the battery compartment lid in the direction indicated by the arrow and remove it. Remove the 4 batteries and replace them. Return the battery compartment lid to its original position and tighten the screw.
Do not use battery sets consisting of old and new batteries, or batteries with different brand names.
!
Page 5
GMC-I Messtechnik GmbH 5
Automatic Shutdown
After switching the device on, “APO” appears at the LCD (automatic power off). The device is switched off automatically after 10 minutes if none of its keys have been activated during this time.
Disabling Automatic Shutdown
The instrument can be set to continuous operation.
Press the NOT AUTO POWER OFF key to this
end.
4Operation
Attention!
!
For reasons of safety, use the instrument on electrical circuits with less than 500 V only. Be sure to ascertain the voltage level of the circuit to be measured before starting the instrument.
Do not activate the 0SET key with open current clamp. This may lead to erroneous measurements.
Do not apply direct currents of more than
1.5 A. This may result in measurement error and damage to the device.
4.1 Using the METRACLIP 63 Current Clamp
Opening and Closing the Clamp
Press the locking lever down in order to
unlock the device.
Press the slider down in order to open
the clamp.
Press the slider up in order to close the
clamp.
Slowly press the locking lever up in order
to securely lock the clamp.
Note!
Correct measurement is only possible when the current clamp is fully closed.
Figure 1: Operation, METRACLIP 63
4.2 Direct Current Measurement
Connect the current clamp to the
CT INPUT socket at the measuring instrument.
Press the POWER key. At first, the 100 mA
measuring range is
displayed.
Select a suitable measuring range at the
measuring instrument with the Range key.
Position the current clamp in proximity to
the conductor to be measured and fully close it.
Press the 0SET key in order to execute
zero balancing.
The “0” display may be unsteady due to geomagnetism or external magnetic fields, but this condition does not indicate malfunctioning.
If the measuring range is changed, zero balancing must be executed again.
Open the current clamp and close it
around the conductor to be measured.
Make sure that the direction of the arrow
on the current clamp corresponds with the flow direction of the current to be measured.
Locking Lever
Slider
Slider
Locking Lever
Open
Close
Page 6
Fully close the current clamp. Read the displayed value. In the event of
overloading, “OL” appears at the display.
Complete measurement and remove the
clamp from the conductor.
4.3 Alternating Current Measurement
Connect the current clamp to the
CT INPUT socket at the measuring instrument.
Press the POWER key. At first, the 100 mA
measuring range is
displayed.
Press the AC / DC key. The 100 mA
range is displayed.
Select a suitable measuring range at the
measuring instrument with the Range key.
Open the current clamp and close it
around the conductor to be measured.
Fully close the current clamp. Read the displayed value. In the event of
overloading, “OL” appears at the display.
Press the DATA HOLD key in order to
freeze the measured value at the display. “DH” appears at the display.
Press the DATA HOLD key again in order to
return to the measurement.
Complete measurement and remove the
clamp from the conductor.
4.4 Measuring Leakage and Stray Current
Measuring DC Leakage Current
Measurement is executed in the same
way as for direct current.
However, the clamp is closed around
two conductors simultaneously.
Measuring AC Leakage Current with Grounded Conductors
Measurement is executed in the same
way as for alternating current.
Measuring AC Leakage Current in Single or 3-Phase Systems, with the Exception of the Ground Conductor
Proceed in the same way as for the
alternating current measurement, except that in the case of single phase systems 2 conductors are simultaneously enclosed by the current clamp, and in the case of
3-phase systems
3 conductors are simultaneously enclosed by the current clamp.
Page 7
GMC-I Messtechnik GmbH 7
4.5 Using a Data Logger
If current measurements will be executed over a period of hours, the output signal can be redirected to a data logger (e.g. a multimeter with memory).
Connect the data logger to the REC OUT
sockets.
Connect the current clamp to the
CT INPUT socket at the measuring instrument.
Switch the measuring instrument on by
pressing the POWER key.
Press the NOT AUTO POWER OFF key in
order to deactivate automatic shutdown. “APO” is cleared from the display.
Switch the data logger on. Select the appropriate settings for the
desired measurement.
Close the current clamp around the
conductor to be measured.
Select a suitable measuring range (see
output signal under “Device Data” on page 9) and start recording.
Before changing the measuring range for a direct current measurement, first remove the current clamp from the conductor under test, press the 0SET key and close the current clamp around the conductor once again in order to perform a new measurement.
Please refer to the section entitled “Direct Current Measurement” for further details.
The output signal for the data logger corresponds to 100 mV DC full scale for all ranges.
In the event of overloading amounting to more than 110% of the selected range, the output signal generates a display of “0 mV” in order to draw attention to this situation.
Page 8
5 Characteristic Values
Direct Current (after zero balancing with the 0 SET key)
Key
rdg. = reading (measured value) d = digit(s)
Reference Conditions
Ambient temperature: +23° C ±5° C Relative humidity: Max. 85% (no condensation allowed)
Influencing Quantities
Geomagnetism influence: Less than ±0.5 mA Magnetization influence: Less than ±0.5 mA at 1.5 A DC on/off Influence caused by open/
closed current clip: Less than ±0.7 mA
Maximum input current: 1.5 A DC (if current greater than 1.5 A DC is applied,
current clamp output is reduced and “OL” does not appear at the display.
Alternating Current
Maximum input current:20 AAC
METRACLIP 63
Range Measuring Range Resolution Accuracy
100 mA 0.1 ...±50.00 mA 0.01 mA ±1% rdg. ± 5 d
±50.01 ...±99.99 mA ±1% rdg. ±10 d
1000 mA 1 ... ±300 mA 0.1 mA ±1% rdg. ±10 d
±300 ... ±700 mA ±2% rdg. ±10 d ±700 ...±999.9 mA ±3% rdg. ±10 d
METRACLIP 64
Range Measuring Range Resolution Accuracy
100 mA 0.1 ...±99.99 mA 0.01 mA ±1% rdg. ±10 d 1000 mA 1 ... ±300 mA 0.1 mA ±1% rdg. ±10 d
±300 ... ±700 mA ±2% rdg. ±10 d ±700 ...±999.9 mA ±3% rdg. ±10 d
Range Measuring Range Resolution Accuracy
100 mA 0 ... 99.99 mA 0.01 mA ±1% rdg. ±10 d (50/60 Hz) 1000 mA 0 ... 999.9 mA 0.1 mA ±1% rdg. ±10 d (50/60 Hz) 10 A 0 ... 9.999 A 0.001 mA ±1% rdg. ±10 d (50/60 Hz)
Page 9
GMC-I Messtechnik GmbH 9
Device Data
Measuring function: Alternating current / direct current Measuring method: Current clamp Measuring ranges: DC: 100 mA / 1000 mA
AC: 100 mA / 1000 mA / 10 A (45 Hz ... 65 Hz) Detection of alternating current: Average sampling A-D converter: Dual integration method Display: LCD, up to 2000 with signal generator Overload display: “OL” appears at the LCD “Data Hold” display: “DH” appears at the LCD Sampling rate: 6 per second for AC and 1 per second for DC Output signal: 100 mV DC full scale for each range; converted to an
analog signal which becomes zero if it exceeds 110% of
the measuring range in order to indicate overloading
(output impedance: less than 10 kΩ) Power supply: 4 alkaline batteries (LR6), size AA Current consumption: Approx. 9 mA (roughly 200 operating hours)
Operating voltage:
Light current range, less than 600 V AC / DC Operating temperature: 0 ... 50° C, < 85% relative humidity
(no condensation allowed) Storage temperature: 10 ... 60° C, < 70% relative humidity
(no condensation allowed) Dielectric strength: 3700 V AC / 1 minute between current clamp core and
handle Insulation value: 100 MΩ for 500 V DC insulation test between current
clamp core and handle Safety standard: EC 61010-1/ 61010-2-032,
CAT II 600 V or CAT III 300 V Dimensions and weight: W x H x D: 78 x 155 x 32 mm, approx. 280 g
Current Clamp Transformer Data for METRACLIP 63
Clamp jaw opening capacity: 5 mm diameter Dielectric strength: 3700 V AC / 1 minute between metal core and handle Cable length: 1.2 meters Safety standard: EC 61010-2-032, CAT || 600 V or CAT ||| 300 V Dimensions and weight: W x H x D: 19 x 113 x 28 mm, approx. 100 g
Current Clamp Transformer Data for METRACLIP 64
Clamp jaw opening capacity: 30 mm diameter Dielectric strength: 3700 V AC / 1 minute between metal core and handle Cable length: 1.2 meters Safety standard: EC 61010-2-032, CAT || 600 V or CAT ||| 300 V Dimensions and weight: W x H x D: 33 x 170 x 24 mm, approx. 165 g
Page 10
6 Maintenance
6.1 Housing
No special maintenance is required for the housing. Keep outside surfaces clean. Use a slightly dampened cloth for cleaning. Avoid the use of cleansers, abrasives and solvents.
Return and Environmentally Sound Disposal
The instrument is a category 9 product (monitoring and control instrument) in accordance with ElektroG (German electrical and electronic device law). This device is not subject to the RoHS directive.
We identify our electrical and electronic devices (as of August
2005) in accordance with WEEE 2002/96/EC and ElektroG using the symbol shown at the right per DIN EN 50419.
These devices may not be disposed of with the trash. Please contact our repair and replacement parts service department regarding the return of old devices.
7
Repair and Replacement Parts Service, DKD Calibration Laboratory * and Rental Instrument Service
If required please contact:
GMC-I Service GmbH Service Center Thomas-Mann-Str. 20 90471 Nürnberg, Germany Phone: +49 911 8602-0 Fax: +49 911 8602-253 e-mail: service@gossenmetrawatt.com
This address is only valid in Germany. Please contact our representatives or subsidiaries for service in other countries.
* Calibration Laboratory for Electrical
Quantities
DKD – K – 19701 accredited per DIN EN ISO/IEC 17025:2005
Accredited quantities: direct voltage, direct current value, direct current resistance, alternating voltage, alternating current value, AC active power, AC apparent power, DC power, capacitance, frequency, temperature
Competent Partner
GMC-I Messtechnik GmbH is certified in accordance with DIN EN ISO 9001:2000.
Our DKD calibration laboratory is accredited by the Physikalisch Technische Bundesan­stalt (German Federal Institute of Physics and Metrology) and the Deutscher Kalibrier­dienst (German Calibration Service) in accor­dance with DIN EN ISO/IEC 17025 under registration number DKD–K–19701.
We offer a complete range of expertise in the field of metrology: from test reports and
proprietary calibration certificates right on up to DKD calibration certificates.
Our spectrum of offerings is rounded out with free test equipment management.
Our DKD calibration laboratory is part of our service department. If errors are discovered during calibration, our specialized personnel
Page 11
GMC-I Messtechnik GmbH 11
are capable of completing repairs using original replacement parts.
As a full service calibration laboratory, we can calibrate instruments from other manufacturers as well.
8 Product Support
If required please contact:
GMC-I Messtechnik GmbH Product Support Hotline Phone +49 911 8602-0 Fax: +49 911 8602 709 E-mail: support@gossenmetrawatt.com
Page 12
Edited in Germany • Subject to change without notice • PDF version available on the Internet
GMC-I Messtechnik GmbH Südwestpark 15 90449 Nuremberg •
Germany
Phone: +49 911 8602-111 Fax: +49 911 8602-777 e-mail: info@gossenmetrawatt.com www.gossenmetrawatt.com
Loading...