Gossen Metrawatt M311F, METRACLIP 62 User guide

Operating Instructions
METRACLIP62 Clip-On Ammeter
3-349-391-15
2/3.10
Operating Overview
15
14
13
12
11
10
U: < 300 V !
I: max. 10 A !
2 GMC-I Messtechnik GmbH
1 Clip guide (+ shield)
2 Clip jaw
3 Clip clearance safety barrier:
Do not reach beyond the safety collar!
4 Button for opening clip
MAX HOLD
key: display max. current measured value
MIN HOLD
key: display min. current measured value
7 RANGE UP key: increase measuring range
8 POWER key: switch on instrument (LCD test)
/ switch off instrument
9 LCD panel
10 Eyelet for carrying strap
11 ACA key: activate AC measurement
12 DCA key: activate DC measurement
13 RANGE DOWN key: reduce measuring range
14 DATA HOLD key: save measured value
by „freezing“ display
ZERO ADJ
key:
15
DCA: zero balancing
(activate before each DC measurement
ACA: define reference value
Standard Equipment
1 Clip-on ammeter 1 Immitation leather bag 1 Set of batteries 1 Operating instructions
!)
GMC-I Messtechnik GmbH 3
Table of Contents Page
1 Safety Precautions ..............................4
1.1 Special Features of the Hall Effect Sensor ........ 6
2 Operation .............................................7
2.1 Initial Start-Up ...............................................7
2.2 Stipulations for All Measurements ................... 8
2.3 Measuring Leakage Current at Grounded
Conductors .................................................. 10
2.4 Measuring Leakage Current at
Single-Phase or 3-Phase Systems ................ 11
2.5 Measuring Line Current ................................ 12
3 Characteristic Values ......................... 13
4 Maintenance ......................................16
4.1 Battery Replacement .................................... 16
4.2 Housing ...................................................... 17

1 Safety Precautions

The METRACLIP62 clip-on ammeter has been manufactured and tested in accordance with safety regulations IEC 61010-1/EN 61010-1/ VDE 0411-1 and IEC 61010-2-032/ EN 61010-2-032/VDE 0411-2-032.
If used for its intended purpose, the safety of the user and of the device is assured. The device may only be operated by properly trained personnel who are capable of recog-
4 GMC-I Messtechnik GmbH
nizing the dangers associated with the mea-
!
surement of electrical current and voltage.
Read the operating instructions completely and carefully before using the device, and follow all instructions included therein.
Meanings of symbols on the instrument:
Warning concerning a source of dan (Attention: observe documentation!)
Continuous, doubled or reinforced insulation
Indicates EC conformity
Applying the clip to dangerous active conductors is permitted
The clip-on ammeter may not be used:
If the battery compartment lid has been removed
If visible damage is apparent
With damaged connector cables
If it no longer functions flawlessly
After lengthy periods of storage under unfavorable conditions (e.g. humidity, dust, temperature).
ger
GMC-I Messtechnik GmbH 5
Attention!
!
Do not perform measurements in the event of over-ranging!
Current which exceeds the measuring range may not be measured.
Do not perform measurements at bare wires! Do not perform measurements at busbars!
Safe Handling
The housing and the handles must be free of dust, grease and moisture.
The operator’s fingers may not be extended beyond the safety collar during measurement.
Avoid excessive mechanical stress such as impact and vibration, as well as high temperatures, moisture and strong magnetic fields.
The battery compartment lid must be secured with the appropriate Phillips head screw.

1.1 Special Features of the Hall Effect Sensor

Eliminating the influence of magnetization
The clip-on ammeter METRACLIP62 uses the hall effect for its measurements. The charac­teristics of the hall effect are subject to a hys-
6 GMC-I Messtechnik GmbH
teresis due to magnetization. As a result, the zero adjustment point for DC measurements may vary. Open and close the clip jaws several times in a row and press the ZERO ADJ key prior to each DC measurement in order to elimitate this effect. When measuring current conduct­ing cables, close and open the clip jaws slowly and firmly in order to avoid any impact on the sensitive hall effect sensor.
Remedy in the Event of Exceeding the Measuring Range
If the maximum measurable current has been exceeded by a large degree, zero balancing can no longer be effected due to the satura­tion of the current clip.
Remedy: Turn the measuring instrument on with the
POWER key.
Adjust the smallest measuring range with
the RANGE DOWN key (200 mA AC).
Close the clip around a conductor with
10 A AC and reduce the current in the conductor gradually to 0 A. This elimi­nates magnetic saturation.

2Operation

2.1 Initial Start-Up

Insert the two batteries (see chapter 4.1, page
16.
GMC-I Messtechnik GmbH 7

2.2 Stipulations for All Measurements

Attention!
!
Note
In order to prevent damage to the in­strument and the user, the maximum input quantities may not be exceeded: Voltage range 300 V AC/DC, current range 20 A AC/DC.
Exercise caution when opening and closing the clip jaws around an active conductor.
Save Measured Value Function – DATA HOLD
During the performance of measurements under conditions which impair easy reading of display values, the DATA HOLD key can be activated in order to freeze the momentary measured value at the display. The DH symbol appears at the display.
Before starting any new measurements, the DATA HOLD key must be activated once again in order to reset the data hold function.
Display of Extreme Value Function – MAX HOLD / MIN HOLD
In order to indicate the highest (MAX HOLD) or lowest current measured value (MIN HOLD) dur­ing a measurement, briefly press the corre­sponding key before starting the measure­ment; the symbol MAX-H or MIN-H appears. During this function, the automatic shutdown
8 GMC-I Messtechnik GmbH
mode is deactivated. Press the key again to cancel this function.
Selecting Current Type – ACA / DCA
The measuring instrument is always set to direct current when switched on. Press the ACA or DCA key to change to the other current type.
Zero Balancing or Reference Value Function – ZERO ADJUST
Zero Balancing for DC measurements DCA
Even if the clip jaws are not closed around a conductor, measured values are continuously displayed when the measuring instrument is set to DC measurement. It is therefore gener­ally necessary to press the ZERO ADJUST key after selecting the measuring range and before starting DC measurements. After pressing the ZERO ADJ key, the symbol ZERO ADJ is displayed and 0.0, 0 or 0.00 appears on the LCD.
After changing the measuring range or the current type, the function must be reactivated.
Reference Value for AC measurements ACA
If the ZERO ADJUST key is pressed during an AC measurement, the measured value is stored as a reference value and deducted from future measurements. After pressing the ZERO ADJ. key, the symbol ZERO ADJ is displayed and
0.0, 0 or 0.00 appears on the LCD.
GMC-I Messtechnik GmbH 9
2.3

Measuring Leakage Current at Grounded Conductors

Turn the measuring instrument on with the
POWER key.
Select the current type via the ACA or DCA
key.
Set the RANGE UP or RANGE DOWN key to the
appropriate range for the current to be measured. Otherwise, set the measuring range switch to the highest of the three ranges in order to avoid possible over-ranging.
Prior to DC measurements:
Perform zero balancing via the ZERO AD- JUST key.
Close the clip around the current conduct-
ing cable under test as well as around the current transformer.
Read the measured value from the dis-
play.
After measurement has been completed:
Remove the clip from the cable and turn the measuring instrument off by pressing the POWER key.
10 GMC-I Messtechnik GmbH

2.4 Measuring Leakage Current at Single-Phase or 3-Phase Systems

Turn the measuring instrument on with the
POWER key.
Select the current type via the ACA or DCA
key.
Set the RANGE UP or RANGE DOWN key to the
appropriate range for the current to be measured. Otherwise, set the measuring range switch to the highest of the three ranges in order to avoid possible over-ranging.
Prior to DC measurements:
Perform zero balancing via the ZERO AD- JUST key.
1-Phase System:
Close the clip around both conducting cables. 3-Phase System: Close the clip around all 3 conducting cables.
Read the measured value from the dis-
play.
After measurement has been completed:
Remove the clip from the cable and turn the measuring instrument off by pressing the POWER key.
GMC-I Messtechnik GmbH 11

2.5 Measuring Line Current

Turn the measuring instrument on with the
POWER key.
Select the current type via the ACA or DCA
key.
Set the RANGE UP or RANGE DOWN key to the
appropriate range for the current to be measured. Otherwise, set the measuring range switch to the highest of the three ranges in order to avoid possible over-ranging.
Prior to DC measurements:
Perform zero balancing via the ZERO AD- JUST key.
Close the clip around a single current con-
ducting cable only.
Read the measured value from the dis-
play.
After measurement has been completed:
Remove the clip from the cable and turn the measuring instrument off by pressing the POWER key.
12 GMC-I Messtechnik GmbH

3 Characteristic Values

Measuring sensor Hall Effect
Measuring Ranges
Meas.
Range
Input
A AC TRMS with automatic zero balancing
200 mA 0 199.9 mA 100 A 1.0 % rdg. 5 digits
2000 mA 0 1999 mA 1 mA 1.0 % rdg. 5 digits
10 A 0 9.99 A 0.01 A1.0% rdg. 10 digits
A DC with automatic zero balancing
200 mA 0 199.9 mA 100 A 1.0 % rdg. 3 digits
2000 mA 0 1999 mA 1 mA 1.0 % rdg. 3 digits
10 A 0 9.99 A 0.01 A1.0% rdg. 10 digits
1)
Frequency range: 50 60 Hz, corner frequency 2 kHz
2)
additional error for hall effect due to hysteresis: 3% rdg.
Key: rdg. = measured value (reading)
Reference Conditions
Ambient temperature +23 C 5 C Relative humidity max. 80%
Accuracy at 18 ... 28 C,
Reso-
max. 80% relative
lution
humidity
1)
1)
1)
2)
2)
2)
GMC-I Messtechnik GmbH 13
LCD
Display
7-segment characters
Number of places 3½ digits,
max. display 1999 Sampling rate Digital display: 1.6/s Overflow display OL appears
DATA HOLD display DH appears
Power Supply
Batteries 2 ea. LR6 (AA size) Service life with alkaline manga-
nese batteries:
approx. 120 hours Battery test Symbol B is dis-
played automati-
cally if battery
voltage is too low. Automatic shutdown The device turns
itself off automati-
cally 10 minutes
after it has been
switched on.
Electrical Safety
Dielectric strength 2.3 kV AC max.
1 min between the
clip jaws and the
exposed clip hous-
ing Measuring category 300 V CAT III 10 A
14 GMC-I Messtechnik GmbH
Electromagnetic Compatibility (EMC)
Interference emission/ Interference immunity IEC 61326
Ambient Conditions
Operating temperatures 0 C  +50 C
< 80% humidity (no condensation allowed)
Storage temperatures 20 C  +60 C
< 75% humidity (no condensation allowed) (without batteries)
Mechanical Design
Clip opening max. dia. 18 mm Dimensions W x H x D: 76 mm x
194 mm x 30 mm
Weight
approx. 350 g with batteries
GMC-I Messtechnik GmbH 15

4 Maintenance

Attention!
!

4.1 Battery Replacement

If the battery symbol B appears at the display, the batteries are either depleted or no longer supply sufficient operating voltage, and must be replaced. The batteries should also be removed if the instrument is placed into stor­age for a lengthy period of time, because leak­age might otherwise occur.
Remove the clip from the measuring circuit first. Only then may the battery compart­ment lid be removed in order to replace the batteries.
Set the on/off switch (POWER) to the OFF
position.
Remove the screw (size 0) which is used
to secure the battery compartment lid.
Pull the battery compartment lid off in the
direction of the arrow symbol (OPEN).
Carefully pry the batteries out of the com-
partment with the Phillips head screw­driver. Replace the depleted batteries. Make sure the batteries are inserted with correct polarity.
16 GMC-I Messtechnik GmbH
Note
Always replace both batteries at once.
Note
Slide the battery compartment lid back
into place so that the two pins engage in the recesses provided.
Reinsert and tighten the screw.
The measuring instrument does not include any replaceable fuses!

4.2 Housing

No special maintenance is required for the housing. Keep outside surfaces clean. Use a slightly dampened cloth for cleaning. Avoid the use of cleansers, abrasives and solvents.
GMC-I Messtechnik GmbH 17
Product Support
When you need support, please contact:
GMC-I Messtechnik GmbH
Product Support Hotline
Phone +49 911 8602-0 Fax +49 911 8602-709 E-Mail support@gossenmetrawatt.com
Repair and Replacement Parts Service
When you need service, please contact:
GMC-I Service GmbH Service Center Thomas-Mann-Straße 20 90471 Nürnberg • Germany Phone +49 911 817718-0 Fax +49 911 817718-253 E-Mail service@gossenmetrawatt.com
This address is only valid in Germany. Please contact our representatives or subsid­iaries for service in other countries.
GMC-I Messtechnik GmbH Südwestpark 15 90449 Nürnberg • Germany Phone +49 911 8602-111 Fax +49 911 8602-777 E-Mail info@gossenmetrawatt.com www.gossenmetrawatt.com
Loading...