Gossen MetraWatt MAVOSPEC BASE Operating Instructions Manual

Page 1
Operating Instructions
MAVOSPEC BASE 15579
Spectrometer 1/07.16
Page 2
2 Gossen Foto- und Lichtmesstechnik GmbH
Thank you for selecting a GOSSEN product.
You’ll enjoy easy operation, top quality and precise measurement, as well as an extensive range of applications. Please check to makes sure that all of the parts listed below have been included in the scope of delivery.
If anything is missing, please contact your dealer.
MAVOSPEC BASE Cover cap, micro SDHC memory card (installed in the battery compartment), SD adapter V070A rechargeable battery, power pack and USB interface cable Aluminum case, sheath, carrying strap Calibration report Operating instructions (German/English)
Outstanding MAVOSPEC BASE Measuring Functions and Features
Spectral power distribution within a range of 380 to 780 nm (VIS)  Chromaticity, color coordinates per CIE 1931 [x,y], CIE 1960 [u,v], CIE 1976 [u',v'], display of the CIE standard color table  Correlated color temperature CCT and color rendering index Ra, as well as individual indices R1 through R15 Flicker value as index, as percentage [%] and with frequency [Hz]  Illuminance with cosine-corrected measuring probe in accordance with class B per DIN 5032-7  Large dynamic range from 10 to 100,000 lx, highest resolution: 0.01 lx  Automatic or manual measured value storage to an interchangeable micro SDHC memory card Easy data transfer thanks to CSV format  Documentation of measured values via included Excel templates for evaluation Universal USB port for convenient data exchange, device control, firmware updates and battery charging  Individualized system integration via open USB interface protocol Sustainable device concept thanks to update option via USB port  Ecological power supply with rechargeable Li-ion battery, recharging via USB port and display of charging status and charge level  Extremely long rechargeable battery operating time of roughly 8 hours continuous operation, which can be extended thanks to adjustable
display and device shutdown
Outstanding measured value stability thanks to automatic temperature compensation and zero-point correction Individualized calibration (photometric and radiometric), calibration report included in scope of delivery Convenient daily use, intuitive operation via ring controller, readily legible high-resolution color display, compact design
Page 3
Gossen Foto- und Lichtmesstechnik GmbH 3
Table of Contents
Page
Safety Precautions, Warnings and Other Notes
5
1
Initial Startup
1.1
Inserting the Memory Card
8
1.2
Inserting the Rechargeable Battery
8
1.3
Charging the Battery
9
1.4
Default Settings
9
2
The Meter and its Controls
2.1
Device Overview
10
2.2
Controls
11
2.3
Display
12
3
Functions
3.1
Switching the Meter On
13
3.2
Measurements
13
3.3
Measured Value Window
14
3.3.1
Spectrum
15
3.3.2
Report
15
3.3.3
CIE
16
3.3.4
CRI
16
3.3.5
Data
16
3.4
Measured Value Memory
17
3.4.1
Saving Measured Values to Memory
17
3.4.2
Loading Stored Measured Values
17
4
MENU
4.1
Overview
18
4.2
Navigation – Setup
19
4.3
MENU – General
20
4.4
MENU – Settings
21
4.5
MENU – Measurements
22
4.6
MENU – Memory
23
4.7
MENU – Presentation
24
4.8
MENU – Information
25
Page 4
4 Gossen Foto- und Lichtmesstechnik GmbH
5
USB Port
25
6
Firmware Update
25
7
Documenting Measured Values
26
8
Practical Tips
26
9
Factory Calibration
27
10
Service
27
11
Error Messages
28
12
Technical Data
29
EU Declaration of Conformity
31
Page 5
Gossen Foto- und Lichtmesstechnik GmbH 5
Safety Precautions
Please read these safety precautions carefully before using your meter. This will help you to avoid damaging the product and prevent personal injury.
This symbol identifies important warnings which should be read
in any case before initial startup of your GOSSEN product.
Warnings
In the event of malfunctioning, switch the meter off immediately.
If the event that smoke develops or unusual odors become apparent, which are caused by either the meter or the power pack, disconnect from mains power immediately and remove the rechargeable battery from the device in order to prevent possible fire. Continuing to operate the meter or the power pack after such malfunctions have occurred may result in severe injury. Please contact your local dealer or GOSSEN Service in order to eliminate malfunctioning. If you bring or send the meter in for repairs, make sure that the rechargeable battery has first been removed.
Do not use the meter in proximity to flammable gases. Electronic devices should never be used in proximity to flammable gases. Danger of explosion and fire is otherwise immanent.
Never hang the device from children with the carrying strap. Danger of strangulation exists if the carrying strap is hung around the neck of a child.
Store the meter at a location which cannot be accessed by children.
The meter and its accessories include parts which can be swallowed. Make sure that these parts (e.g. housing covers, rechargeable batteries etc.) do not fall into the hands of children who might swallow them. Otherwise, danger of suffocation prevails.
Use suitable cables only. Use included, original GOSSEN cables only for connection to external devices. GOSSEN assumes no liability if other cables are used.
Page 6
6 Gossen Foto- und Lichtmesstechnik GmbH
Do not dismantle the meter.
Never touch any parts inside the housing. Injury may result. Do not repair the meter yourself. Repairs may only be conducted by appropriately trained personnel. If the meter’s housing is damaged due to dropping or other external influences, remove the rechargeable battery or power pack and contact your local dealer or GOSSEN Service for repair.
Avoid any and all contact with the liquid crystals. If the display is damaged (e.g. broken), danger of injury due to contact with glass shards or discharge of liquid crystals exists. Make sure that skin, eyes and mouth do not come into contact with the liquid crystals.
Be careful when handling rechargeable batteries.
Rechargeable batteries may leak or explode if handled incorrectly. Please adhere to the following safety precautions:
• Make sure that the meter is switched off before removing or inserting rechargeable batteries. If the meter is used with a power pack, supply power must first be disconnected (pull the mains plug out of the electrical outlet).
• Only use rechargeable batteries which are recommended for this meter.
• Make sure that the rechargeable battery is correctly inserted.
• Never short-circuit rechargeable batteries, and never attempt to open a rechargeable battery.
• Do not expose rechargeable batteries to excessive heat or open fire.
• Do not expose rechargeable batteries to moisture and never immerse rechargeable batteries in water.
• After removing the rechargeable battery from the meter, close the battery compartment with the lid
(e.g. if the meter will not be used for a lengthy period of time).
• Never store rechargeable batteries together with metallic objects which might cause short-circuiting.
• Danger of leakage exists, especially in the case of empty rechargeable batteries. In order to prevent damage to the meter,
rechargeable batteries should be removed when fully depleted or in the case of lengthy periods of non-use.
• When not in use, rechargeable batteries should be stored in a cool place.
• Rechargeable batteries heat up during use and may become hot. Be careful not to burn yourself when removing rechargeable
batteries. Switch the meter off or wait until it has shut itself down, and then wait a bit longer until the rechargeable battery has cooled down.
• Do not use rechargeable batteries which show any signs of damage such as discoloration or deformation of the housing.
Page 7
Gossen Foto- und Lichtmesstechnik GmbH 7
Other Notes
Reproduction of product documentation or duplication of any excerpts therefrom necessitates the express consent of GOSSEN Foto- und
Lichtmesstechnik GmbH. This applies as well to duplication in any electronic format and translation into other languages.
GOSSEN Foto- und Lichtmesstechnik GmbH reserves the right to make changes of any type without providing advanced notice.  GOSSEN assumes no liability for damages resulting from incorrect use of the product.  Documentation for your GOSSEN meter was prepared with the greatest of care. If you should nevertheless discover errors, or if you would
like to suggest any improvements, GOSSEN would be very pleased to hear from you.
Symbol for separate collection of recyclable materials / hazardous waste in European countries
This symbol indicates that this product must be disposed of separately. The following must be observed by users in European countries:
• This product may only be disposed of separately at a designated collection point. It may not be disposed of with household trash.
• For further information contact your local dealer or waste disposal authorities.
The following symbols are used in order to make it easier to find additional information.
!!
Important safety precautions: Please read these safety precautions before using the meter in order to avoid damaging your
MAVOSPEC BASE.
!
Important information which you should also read before using your MAVOSPEC BASE
i
Tips: additional, useful information regarding use of your MAVOSPEC BASE
Reference to other information included in these operating instructions
M
Individual functions which can be configured in the menu
Page 8
8 Gossen Foto- und Lichtmesstechnik GmbH
1 Initial Startup
The MAVOSPEC BASE works with an interchangeable micro SD card for measured value storage and a device-specific, rechargeable lithium-ion battery. Use only the original V070A GOSSEN rechargeable battery included with the meter, or available as an optional accessory, and the USB charger.
Wait until your MAVOSPEC BASE has been shut down.  Unlock the lid at the back of the meter with a Phillips head screwdriver, remove it
and pull it down and away.
1.1 Inserting the Memory card
Remove the rechargeable battery from the battery compartment.  Pull the micro SDHC memory card out of the included adapter.  Insert the micro SDHC memory card into your MAVOSPEC BASE in
the recess provided for this purpose in the battery compartment.
Push the card into the slot in the meter in the indicated direction.
1.2 Inserting the Rechargeable Battery
Insert the battery into the compartment as shown in the figure. Make sure
that the + and – poles are correctly connected!
Close the battery compartment lid and secure it with the screw.
Remove the lid.
Unlock.
Insert battery.
!
The MAVOSPEC BASE can also be operated without a memory card – all measuring and display functions are still available and only measured value storage at the meter is excluded.
Pull Micro SDHC
card from adapter.
Insert micro SDHC card und push into the slot.
Lock.
Page 9
Gossen Foto- und Lichtmesstechnik GmbH 9
1.3 Charging the Battery
Connect the included USB cable first of all to the USB port on the meter and then to the USB socket on the power pack, and then connect the power pack to a mains outlet.
The LED in the housing top indicates the charging status of the rechargeable battery. It lights up red as long as the battery is charging, and changes to green when the battery has been fully charged. If the meter is switched on, a symbol appears at the battery display indicating that the device is being externally supplied with electrical power. Charging time for a fully depleted battery is about 1½ hours.
1.4 Default Settings
The MAVOSPEC BASE is shipped with default settings which, based on our own experience, fulfill the basic requirements of most users. A complete summary of the default settings and instructions on how to adapt them to your individual requirements are included in section 4 of these operating instructions. Entered settings are retained until they have been changed again, or until the meter is reset to its default settings as described in section 4.4.
!
When connected to a PC, the device is switched to continuous operation and is supplied with electrical power from the PC’s USB port.
The meter can be operated with or without an inserted battery when connected to a PC or to the power pack.
An additional or replacement battery (3.7 V / 890 mAh) can be purchased from GOSSEN under order number V070A.
Page 10
10 Gossen Foto- und Lichtmesstechnik GmbH
2 The Meter and its Controls
2.1 View of the Meter
Cosine-Corrected Measuring Probe
Charging Status LED
Display Panel
DATA Key
Ring Controller
Carrying Strap Holder
USB Port for Charging Battery or for Data Transmission
MENU Key
Measurement Key
Page 11
Gossen Foto- und Lichtmesstechnik GmbH 11
Ring Controller
Up/down
Measured value view:
Report, data: scroll through the measured value list.
CIE: Switch amongst 1931 - 1960 – 1976.
CRI: Switch amongst table - bar – spider.
Menu view:
Scroll through the parameters list.
Left/right Measured value view:
Switch amongst report - CIE - CRI – data.
Menu view:
Menu item selection, parameter display, exit parameter display
2.2 Controls
DATA Key
Press key briefly:
Storage of the last measurement, inactive in case of automatic storage
Press and hold: Retrieve stored measured values, stop current retrieval operation
MENU Key
Open menu,
exit menu
Measurement Key
Measurement:
Briefly press the measurement key.
Settings: Acknowledge selection.
Device in sleep mode: Switch device on.
!
After switching the device on, the settings used for the last work cycle are displayed.
DATA + MENU Key
Pressing and holding both of these keys simultaneously switches the device off.
Page 12
12 Gossen Foto- und Lichtmesstechnik GmbH
2.3 Displays
The battery icon in the information bar shows the charging status of the battery, or indicates operation with mains power or power from the PC. Even when the device is switched off, the charging status LED indicates whether the battery is being charged or has already been fully charged.
Icons in the Information Bar and the Charging Status LED
Battery full
Operation with mains power or at PC
Battery partially charged
LED blinks red – battery is being charged
Battery nearly depleted – charge now
LED lights up green – battery is fully charged
LCD brightness can be adjusted to meet your individual needs within a range of 50% to 100% in 10% steps. In order to further extend rechargeable battery operating time, display shutdown time can be adjusted after which LCD brightness is automatically reduced to 50% of the selected value. Unless the device has been fully shut down, LCD brightness is returned to its originally selected value each time a key is pressed. A complete summary of the default settings and instructions on how to adapt them to your individual requirements are included in section 4 of these operating instructions.
Status Bar
Shows the name of the active measured value window or the measured value memory’s directory, and time of day
Information Bar
Shows currently running actions such as measurement, storage, loading, the file name and the battery’s charging status
Measured Value Window
Shows the active measured value window or the measured value memory’s directories and files
Page 13
Gossen Foto- und Lichtmesstechnik GmbH 13
3 Functions
3.1 Switching the Meter On
Use the
measurement
key to switch the
meter on.
The meter
is initialized ...
… and the initial screen
appears with an empty
display field.
3.2 Measurements
Press the
measurement
key to start.
Possible from any measured value window.
A measurement
is triggered and
the measured values appear
in the display
field.
Configurable Individual Functions in the MENU
M
Units of measure
lx / °C - fc / °F
Decimal separator
Comma/point
Auto int. time
On/off
Int. time
10 to 3000 ms in 10 ms steps
Averaging
1 to 5
Auto save
On/off Report
Display presetting
Spectrum
Color/simple
CIE standard
1931 / 1960 / 1976
CRI standard
Bar / table / spider
Page 14
14 Gossen Foto- und Lichtmesstechnik GmbH
3.3 Measured Value Windows
After the measurement has been performed, its results are displayed in the selected measured value window.
The ring controller is used to switch amongst the individual measured value windows.
The availability of the individual measured value windows und the parameters they contain can be adapted in a customer-specific fashion in the presentation menu (see section 4). Individual measured value windows or measured values which are not displayed can be activated there.
A new measurement can be triggered in each measured value window.
Spectrum
CCT E vis
Report
E vis
CCT
Duv
CRI
ʎpeak
ʎdominant
Purity
Flicker index
Flicker [%]
Flicker [Hz]
CIE1931
x, y
CRI
Bar graph Ra, R9
Data
380 to 780 nm
CIE1960
u, v
CRI
Ra, Ra1 - Ra15
CIE1976
u', v'
CRI
Spider
Page 15
Gossen Foto- und Lichtmesstechnik GmbH 15
3.3.1 Spectrum
This measured value window displays the spectral power distribution of a light source and also provides information concerning color temperature and illuminance. The window as well as the colored background for the spectral power distribution can be shown or hidden with the help of the corresponding function in the presentation menu (see section 4).
CCT Correlated color temperature in Kelvin (K) E vis Illuminance (lx/fc) Graphic Standardized spectral power distribution (mW/m²/nm) over the entire measured waveband
3.3.2 Report
This measured value window lists all measured values calculated by the device from the measured spectral power distribution. The window as well as the individual values can be shown or hidden with the help of the corresponding function in the presentation menu (see section 4).
E vis Illuminance (lx/fc) CCT Correlated color temperature in Kelvin (K) Duv Delta uv – color temperature difference relative to the Planckian locus CRI Color rendering index Ra
ʎ peak Wavelength (nm) of the maximum spectral radiant intensity (ʎp) – peak intensity ʎ dominant Dominant wavelength
Purity Color purity Flicker index Flicker index Flicker [%] Flicker as a percentage Flicker [Hz] Flicker frequency
Page 16
16 Gossen Foto- und Lichtmesstechnik GmbH
3.3.3 CIE
These measured value windows display the color coordinates in the respective CIE standard color system and depict chromaticity graphically in the associated standard color diagram. The window as well as all or one preferred standard color system can be shown or hidden with the help of the corresponding function in the presentation menu (see section 4).
CIE 1931 Standard color system, graphic – chromaticity in the standard color diagram, color coordinates x, y CIE 1960 Standard color system, graphic – chromaticity in the standard color diagram, color coordinates u, v CIE 1976 Standard color system, graphic – chromaticity in the standard color diagram, color coordinates u', v'
3.3.4 CRI
These measured value windows display color rendering index Ra and individual indices R1 through R15 in
various formats. The window as well as all or one preferred display format can be shown or hidden with the help of the corresponding function in the presentation menu (see section 4).
Bar graph Color rendering indices R1 through R15 as a bar graph as well as Ra and R9 as numeric values Table Color rendering index Ra and color rendering indices R1 through R15 as numeric values Spider Color rendering index Ra and color rendering indices R1 through R15 as a spider graph
3.3.5 Data
This measured value window displays irradiance either as original data from the sensor or as data interpolated to even 5 nm steps. The window as well as the preferred display format can be shown or hidden with the help of the corresponding function in the presentation menu (see section 4).
Wavelength (nm), associated irradiance (mW/m²/nm)
Page 17
Gossen Foto- und Lichtmesstechnik GmbH 17
3.4 Measured Value Memory
The MAVOSPEC BASE has an interchangeable micro SDHC memory card in the battery compartment to which measured values can be saved either automatically or manually. Manual saving is selected as a default setting. The respective memory modes can be selected in the memory menu (see section 4.6). Measured values are saved in the CSV data format, which can be opened with Excel and other programs for convenient further processing.
3.4.1 Saving Measured Values
Manual saving
Briefly pressing the DATA key saves the values from the last measurement.
Automatic saving
The value is saved automatically after measurement has been completed.
A CSV file using the time of day as a file name 10-58-22.CSV is saved to the respective day file (yy-mm-dd) 15-12-23.
3.4.2 Loading and Deleting Saved Measured Values
The memory menu is opened or exited by pressing and holding the DATA key.
micro SDHC card
yy-mm-dd
Navigate to the folder with the
ring controller.
..
hh-mm-ss.CSV
hh-mm-ss.CSV
hh-mm-ss.CSV
Navigate to the file
with the ring
controller.
Press the
measurement key.
The menu for
loading or deleting
the file appears.
Navigate to the desired command
with the ring controller ...
Load
Delete
… and acknowledge by pressing the
measurement key.
Load: The measured value appears in
the measured value windows.
Delete: The measurement is deleted.
Briefly pressing the
measurement key
opens the folder.
Pressing and holding the
measurement key accesses
the menu for opening or
deleting the folder.
Page 18
18 Gossen Foto- und Lichtmesstechnik GmbH
4 The Menu
This chapter expands upon the above described functions and explains the great variety of options for default settings and presets offered by the MAVOSPEC BASE.
4.1 Overview
General
Default Setting
Options
Information
LCD brightness (%)
100
100 - 90 - 80 - 70 - 60 - 50
Sensor no.
Display off (min.)
2
5 - 4 - 3 - 2 -1 - 0.5 - off
Serial no.
Device off (min.)
4
10 - 5 - 4 - 3 - 2 - 1 - off
Software rev.
Language
German
English - German
Hardware rev.
Time/date
00:00 01.01.2015
Product ID
Settings
Temperature
Units of measure
lx / °C
fc / °F - lx / °C
Batt. voltage
Decimal separator
, (comma)
. (point) - , (comma)
Hour format
24h
24h - 12h
Date format
dd
dd.mm.yyyy - mm/dd/yyyy - yyyy-/mm/dd
Default settings
Cancel - Reset
Measurement
Auto int. time
On
Off - On
Int. time
10 to 3000 ms in 10 ms steps
Averaging
1
1 - 2 - 3 - 4 - 5
Notification
Sound
Graphic - Sound
Dark adjustment
Cancel - Execute
Memory
Auto save
Off
Off - On
Spectrum
5 nm
Off – 5 nm - Original
Presentation
Spectrum
Color
Off - Color - Simple
Report
E vis - CCT - CRI ­Flicker Index - % - Hz
E vis – CCT – Duv – CRI - ʎpeak - ʎdominant – Purity – Flicker Index – Flicker [%] – Flicker [Hz]
CIE
All
Off - All – 1931 – 1960 - 1976
CRI
All
Off - All - Bar – Table - Spider
Data
5 nm
Off – 5 nm - Original
Page 19
Gossen Foto- und Lichtmesstechnik GmbH 19
4.2 Navigation – Setup
Press the MENU key to open the main menu.
Use the ring
controller ...
to select the desired
menu.
Use the ring
controller ...
to navigate through
the user settings.
Use the ring
controller ...
to select a submenu.
Use the ring
controller ...
to display values.
Use the ring
controller ...
to select the desired
value.
Use the ring
controller ...
to navigate through
the value settings.
Use the ring
controller ...
to select the desired
setting.
Use the
measurement key ...
to acknowledge.
Use the ring
controller ...
to return to the
submenu.
Press the menu key to exit the main menu at any time.
Page 20
20 Gossen Foto- und Lichtmesstechnik GmbH
4.3 MENU – General
LCD Brightness (%)
Display brightness can be adjusted to suit the respective requirements. Minimal display brightness results in low current consumption and increases availability of the meter during battery operation.
LCD brightness (%) 100 - 90 - 80 - 70 - 60 - 50
Display Off (min.)
In order to further extend battery operating time, a display shutdown time can be selected. If none of the meter’s controls are activated during this time, LCD brightness is automatically reduced to 50%. Activation of any key returns LCD brightness to the level selected as described in section 4.3.1. All measured values and settings are retained.
Display off (min.) 5 - 4 - 3 - 2 - 1 - 0.5 - Off
Device Off (min.)
The meter is switched off if none of the controls are activated during the selected period of time. All measured values and settings are fist saved and are retained until the meter is switched back on again by pressing the measurement key, M.
Device off (min.) 10 - 5 - 4 - 3 - 2 - 1 - Off
Language
The meter’s menu can be set to the desired language.
Language English - German
Time/Date
The meter displays time of day in the status line and uses it as a file name for the measured value memory, which additionally uses the date as a designation for the day file. If the battery is depleted or removed for the purpose of replacement, the integrated real-time clock continues to run for approximately 12 hours before it stops. In order to assure that date and time are retained for as long as possible, it’s advisable to recharge the battery after lengthy periods of use.
Time, date 00:00 01.01.2016 to 23:59 31.12.2099
Page 21
Gossen Foto- und Lichtmesstechnik GmbH 21
4.4 MENU – Settings
All settings apply to the display at the device as well as the stored measured value file. Where the decimal separator and date are concerned, country specific adaptation may be required in order to assure that the CSV file is read in correctly by the respective software program.
Units of Measure
Units of measure can be switched from the metric system to the imperial system. Illuminance in Lux (lx) is then changed to foot-candles (fc) and temperature is changed from degrees Celsius (°C) to degrees Fahrenheit (°F).
Units of measure fc / °F - lx / °C
Decimal Separator
The decimal comma (,) is used in some countries as the decimal separator, and the decimal point (.) in others.
Decimal separator . (point) - , (comma)
Hour Format
Some countries use a 24 hour clock (24h), and others use a 12 hour clock (12h) together with the abbreviations a.m. (ante meridiem) for before noon and p.m. (post meridiem) for afternoon.
Hours 24h - 12h
Date Format
The date is written differently in the various countries. The following settings are possible:
Date format dd.mm.yyyy (day.month.year) - mm/dd/yyyy (month/day/year) - yyyy/mm/dd (year/month/day)
Default Settings
The meter is reset to the default settings described in section 4.1 but date and time are retained.
Default settings Cancel - Reset
Page 22
22 Gossen Foto- und Lichtmesstechnik GmbH
4.5 MENU – Measurement
The measuring range, and thus the sensitivity of the spectrometer, are controlled by means of integration time, i.e. the time during which the sensor collects light. The brighter the light source, the shorter integration time must be, and vice versa. In this respect it must be assured that the sensor is not saturated even during the shortest integration time. It may be necessary to increase distance to the light source.
Auto Int. Time
The meter can automatically adjust integration time, and thus the measuring range as well, to prevailing measuring conditions. Automatic determination of integration time should only be deactivated by experienced users, after which integration time must selected manually.
Auto int. time Off - On
Int. Time
The integration time used for the last measurement is displayed in this menu, and is continuously overwritten if automatic selection of integration time has been activated. When automatic selection of integration time is deactivated, integration time is selected manually in this menu.
Int. time (ms) 10 to 3000 ms in 10 ms steps
Averaging
In the case of fluctuating light sources, averaging can be activated in order to obtain more stable measured values. The meter conducts the specified number of measurements and generates a mean value from all of the results. Measurement takes longer when averaging is activated.
Averaging 1 - 2 - 3 - 4 - 5
Notification
The beginning of each measurement can be indicated by a brief acoustic signal, or by the appearance of a red hourglass at the display. A long acoustic signal is generated at the end of the measurement, or a green hourglass is displayed briefly. Sound or graphic can be selected depending on the respective requirements.
Notification Graphic - Sound
Page 23
Gossen Foto- und Lichtmesstechnik GmbH 23
Dark Adjustment
The meter is equipped with automatic dark stream compensation with temperature sensor, which is preset during setup at the factory. Dark adjustment is usually unnecessary within the specified operating temperature range.
Dark adjustment can be conducted as follows, but the meter is reset to the factory default value when it’s switched off in order to prevent erroneous measurements.
Dark adjustment Cancel - Execute Please cover the sensor! Cancel - Adjust
4.6 MENU – Memory
Data storage to the micro SDHC memory card offers the following setting options.
Auto Save
The measured values for the last measurement are saved either manually by briefly pressing the DATA key, or automatically after every measurement. The auto save function can be activated in this menu.
Auto save On - Off
Spectrum
Storage of the spectral power distribution to the file can be either deactivated, or it can be interpolated to 5 nm steps or with original sensor resolution.
Spectrum Off – 5 nm - Original
Page 24
24 Gossen Foto- und Lichtmesstechnik GmbH
4.7 MENU – Presentation
The content and the availability of the individual measured value windows can be adapted to the user’s own specific requirements.
Spectrum
The following settings are available for the “Spectrum” measured value window:
Spectrum Off - Color - Simple
Report
The individual measured quantities can be shown or hidden for the “Report” measured value window.
E vis, CCT, Duv, CRI, ʎ peak, ʎ dominant, Purity, Flicker Index, Flicker [%], Flicker [Hz] On - Off
CIE
Various display formats can be selected for the CIE measured value window. The measured value window can be hidden, all CIE standard color systems can be shown, or just one selected CIE standard color system can be shown.
CIE Off - All - 1931 - 1960 - 1976
CRI
Various display formats can be selected for the CRI measured value window. The measured value window can be hidden, all display formats can be shown, or just one selected display format can be shown.
CRI Off - All - Bar - Table - Spider
Data
The “Data” measured value window can be hidden, or spectral power distribution can be displayed either interpolated to 5 nm steps or with original sensor resolution.
Data Off – 5 nm - Original
Page 25
Gossen Foto- und Lichtmesstechnik GmbH 25
4.8 MENU – Information
Important device information is summarized in this menu. It includes the sensor number, the serial number, the software version and the hardware version. This information is necessary if any questions regarding the product should arise, or if a malfunction should occur. Battery voltage and the temperature of the measuring probe appear in this menu as well.
5 USB Port
The meter’s USB port is located on the bottom of the housing. The meter is connected to a PC via the USB interface cable, and the PC detects the integrated micro SDHC memory card as a removable data storage medium. The measurement files stored in CSV format can then be easily opened, copied, moved or deleted. As long as the meter is connected to the PC, it’s supplied with electrical power via the USB port and is not switched off.
If the meter is connected to the power pack via the USB interface cable, the integrated battery is charged as described in section 1.3. Charging via the
PC’s USB port takes a long time and is not recommended. The open protocol for device control and data communication permits incorporation into the user’s own applications. The interface description and an
associated demo application can be downloaded from the MAVOSPEC BASE product page at www.gossen-photo.de.
6 Firmware Update
The sustainable device concept is open to future function expansions and amendments to the applicable standards. If necessary, new firmware versions will be made available by GOSSEN, which the customer can install to the device himself. After updating the firmware, the meter is then fully up to date. User settings remain unchanged. Updating instructions and new firmware can be downloaded from the MAVOSPEC BASE product page at
www.gossen-photo.de.
Page 26
26 Gossen Foto- und Lichtmesstechnik GmbH
7 Documenting Measured Values
A Microsoft Excel file with various sample templates for reports is included in the integrated micro SDHC memory card. The individual elements can be adapted as desired, and new templates can be created. They access the “Data” spreadsheet, into which the stored measured value files can be read in automatically after clicking a button, or a measurement can be started with the connected meter. The report can then be saved as a PDF file by clicking a button.
The current version of the Microsoft Excel file is available for download from the MAVOSPEC BASE product page at www.gossen-photo.de.
8 Practical Tips
A broad range of information regarding measured quantities, measuring methods, applications and photometry standards, as well as support in selecting a suitable meter, is included in the Photometry Compendium. The compendium can be downloaded under Hints & Guides under the heading Light Measuring at www.gossen-photo.de, or requested from GOSSEN as a printed version.
Page 27
Gossen Foto- und Lichtmesstechnik GmbH 27
9 Factory Calibration
The MAVOSPEC BASE with intuitive user interface is one of the most accurate and reliable spectrometers in its class, and reflects the most up-to-date technology available on the market. Like all other precision light meters, this product also requires regular maintenance, recalibration and firmware updates in order to continuously fulfil performance capabilities within the tolerances and specifications stipulated by the manufacturer. Depending on conditions of use, a calibration interval of 12 to 24 months is recommended.
10 Service
The device does not require any special maintenance if used in accordance with the operating instructions. If the outside of the device becomes contaminated during use, clean the surface of the housing with a slightly moistened cloth. Avoid the use of cleansers, abrasives or solvents.
If at any time your meter does not function to you full satisfaction, please contact us or send it to us at: GOSSEN Foto- und Lichtmesstechnik GmbH I Lina-Ammon-Str.22 I D-90471 Nuremberg I Germany
Phone: +49 911 8602-181 I Fax: +49 911 8602-142 I e-mail: info@gossen-photo.de I www.gossen-photo.de Customers outside of Germany are requested to contact their authorized dealer, whose address can be found on our website.
Page 28
28 Gossen Foto- und Lichtmesstechnik GmbH
11 Error Messages
The error messages listed below may appear during operation and are displayed in the information bar.
Error Message
Situation
Cause
Elimination
Faulty calibration data
Initial window
Calibration data memory illegible or faulty
If this error occurs several times after a restart, please send the meter for repair.
Device not calibrated
Measured value window
No calibration data available
SD card error
Initial window or save/load measurement
SD card cannot be accessed
Insert the SD card or examine it at the PC and reformat it if necessary (FAT16).
File error
Save/load measurement File cannot be accessed
Folder error
Folder cannot be accessed
Saving not permitted
Save measurements
The measurement has already been saved or is outside of the valid measuring range.
Signal too weak
After performing a measurement
The measurement signal is too weak.
Reduce distance to the light source.
Signal too strong
The measurement signal is too strong.
Increase distance to the light source.
Battery almost dead
Measured value window
Battery has not been charged on time.
Charge the battery or supply the device with electrical power from the power pack of the PC.
3 brief acoustic signals
Any time
Battery dead
Page 29
Gossen Foto- und Lichtmesstechnik GmbH 29
12 Technical Data
Sensor Technology, Measuring Uncertainty
Sensor
CMOS image sensor, 256 pixels
Diffusor light-entry surface
7 mm diameter
Distance, diffuser to surface to be measured
25 mm
Error limit, cosine correction
< 3% (comparable with the f2 failure of a class B device according DIN 5032-7)
Spectral range
380 … 780 nm
Full width at half maximum (FWHM)
< 15 nm (typically 12 nm)
Physical resolution
~ 1.72 nm
Digital resolution
16 bit
Wavelength reproducibility
+ 0.5 nm
Integration time
Automatic, manual, 10 … 3000 ms in 10 ms steps
Signal-to-noise ratio
1000:1
Spurious light
-25 dB
Dark stream compensation
Automatic with integrated temperature sensor
Measurement uncertainty, illuminance
Standard light type A, 2856 K
@ 1000 lx
+ 3 %
Measurement uncertainty, chromaticity
Reproducibility, chromaticity
+ 0.0005
Measurement uncertainty, CCT
+ 2%
Measurement uncertainty, CRI
+ 1.5%
Measurement uncertainty, flicker
+ 1.5%
Photometry
Measuring functions
Illuminance
10 … 100,000 lx
Color temperature (K)
Duv – color temperature difference relative to the Planckian locus
Color rendering index Ra, R1 to R15
CIE 13.3
ʎpeak – peak wavelength
ʎdominant – dominant wavelength
CIE 15
Purity – color purity
CIE 15
Chromaticity coordinates
CIE 1931 [x,y], CIE 1960 [u,v], CIE 1976 [u',v']
Flicker index
0.00 ... 1.00
Flicker as percentage (%)
0% ... 100%
Flicker frequency (Hz)
2 ... 1000 Hz
Units of Measure
lx / °C - fc / °F, selectable
Page 30
30 Gossen Foto- und Lichtmesstechnik GmbH
Operation, Interfaces, Memory
Display
2.1" TFT color display, 320 x 240 pixels, brightness adjustable from 50% to 100%, shutdown adjustable from 0.5 … 5 min. or off
Controls
3 keys, ring controller
Languages
German, English
Interface
USB 2.0, data transmission, battery charging, open interface protocol
Data storage
4 GB micro SDHC memory card, manual and automatic memory modes, CSV data format, memory occupancy per measurement: 2 kB, per thousand measurements: 2 MB
Power Supply
Mains power pack
100 … 240 V AC (50/60 Hz) 0.15 A, USB socket: 5 V DC, 1 A
Rechargeable battery
Li-ion, 3.7 V, 890 mAh, replaceable
Charging time with power pack
1.5 hours
Operating time with rechargeable battery
> 8 hours continuous operation
General
Dimensions
139 x 60 x 30 mm
Weight
150 g (meter with battery and memory card)
Operating temperature
+5 to +40° C
Storage temperature
-20 to +70° C
Scope of delivery
Meter, sensor cover cap, 4 GB micro SDHC memory card with SD adapter, V070A rechargeable battery, power pack, USB interface cable, aluminum case, neoprene sheath, carrying strap, calibration report, operating instructions in German and English
Optional Accessories
Replacement battery (V070A)
Li-ion, 3.7 V, 890 mAh
Page 31
Gossen Foto- und Lichtmesstechnik GmbH 31
EU KONFORMITÄTSERKLÄRUNG
EU DECLARATION OF CONFORMITY
Hersteller / Manufacturer: GOSSEN Foto- und Lichtmesstechnik GmbH Anschrift / Address: Lina-Ammon-Str. 22
D-90471 Nuremberg, Germany
Produktbezeichnung / Product designation: Spektrometer / Spectrometer
Typ / Type: MAVOSPEC BASE
Der oben beschriebene Gegenstand der Erklärung erfüllt die ein­schlägigen Harmonisierungsrechtsvorschriften der Union, nach­gewiesen durch die vollständige Einhaltung folgender Normen:
The object of the declaration described above is in conformity with the relevant Union harmonization legislation, proven through complete compliance with the following standards:
Nr. / No.
Richtlinie
Directive
2014/35/EU
Bereitstellung elektrischer Betriebsmittel zur Verwendung innerhalb bestimmter Spannungsgrenzen auf dem Markt
Making available on the market of electrical equipment designed for use within certain voltage limits
Norm / Standard: EN 60950-1 : 2006 + A11:2009 + A1:2010 + A12:2011 + AC:2011
Nr. / No.
Richtlinie
Directive
2014/30/EU
Elektromagnetische Verträglichkeit
Electromagnetic compatibility
Norm / Standard: EN 61326-1 : 2013-07
Nürnberg, 15 Juli 2016 ___________________________________________________ _____________________________________
Ort, Datum / Place, Date Geschäftsführer / Managing Director
Diese Erklärung bescheinigt die Übereinstimmung mit den genannten Richtlinien, beinhaltet jedoch keine Zusicherung von Eigenschaften. Die Sicherheitshinweise der mitgelieferten Produktdokumentationen sind zu beachten.
This declaration certifies compliance with the above mentioned directives, but does not imply any warranted characteristics. The safety precautions included in the product documentation furnished with the product must be adhered to.
Page 32
32 Gossen Foto- und Lichtmesstechnik GmbH
Printed in Germany - Subject to change without notice
GOSSEN Foto- und Lichtmesstechnik GmbH I Lina-Ammon-Str.22 I D-90471 Nuremberg I Germany Phone: +49 911 8602-181 I Fax: +49 911 8602-142 I e-mail: info@gossen-photo.de I www.gossen-photo.de
Loading...