Gossen Metrawatt M360E, M360D User manual

Operating Instructions
PROFITESTPVSUN and PVSUN MEMO
Test Instruments for Testing PV Modules and Strings per DIN EN 62446 (VDE 0126-23)
3-349-672-15
11/3.16
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Contents Page
Caution! Dangerous Voltage
Attention!
!
Warning!
Attention!
!
Note
!

2 Significance of Symbols

1 Scope of Delivery ............................................................. 2
2 Significance of Symbols .................................................. 2
3 Introduction ...................................................................... 3
4 Transport and Storage ..................................................... 3
5 Safety Precautions ........................................................... 3
6 Use for Intended Purpose ................................................. 3
7 Operating and Display Elements ..................................... 3
8 Initial Start-Up .................................................................. 4
9 Measurements ................................................................. 4
9.1 Testing Protective Conductor Continuity
per EN 62446 (VDE 0126-23) (> 200 mA) ......................................4
9.2 Polarity Test / Measuring Open Circuit Voltage U0
per EN 62446 (VDE 0126-23) ........................................................4
9.3 Measuring Short-Circuit Current I
9.4 Testing Insulation Resistance per EN 62446 ....................................5
9.5 General Insulation Resistance Measurement ....................................6
9.6 Ground Fault Test ..........................................................................6
per EN 62446 ..........................5
SC
10 Data Storage with PROFITEST PVsun memo .................... 7
10.1 Operating and Display Elements .....................................................7
10.2 Function Description ......................................................................7
Meanings of Symbols on the Instrument
Warning concerning a point of danger (attention: observe documentation!)
Protection class II device (continuous, doubled or reinforced insulation)
This device may not be disposed of with the trash. Fur-
EC mark of conformity
Meanings of Symbols in the Operating Instructions
While performing measurements in the presence of dan­gerous voltages, the PROFITEST PV
PROFITEST PV
safety gloves as it involves the risk of hazardous contact.
SUN MEMO should only be operated with PPE
SUN/
11 Application of PC Software .............................................. 8
11.1 System Requirements ....................................................................8
11.2 Installation ....................................................................................8
11.3 Initial Startup of the Device .............................................................8
11.4 Creation of Structures and Transfer to the Test Instrument ...............8
11.5 Reading Out and Exporting Test Data ..............................................8
12 Measurement of irradiation, temperature and inclination
with PROFITEST PVsun-sor (optional) .............................. 9
13 Characteristic Values ....................................................... 9
14 Maintenance .................................................................. 10
14.1 Battery Replacement ....................................................................10
14.2 Housing ......................................................................................10
15 Repair and Replacement Parts Service
Calibration Center* and Rental Instrument Service .............10
16 Product Support ............................................................. 10

1 Scope of Delivery

1 PROFITEST PVSUN or PROFITEST PVSUN MEMO 4 Batteries, 1.5 V IEC LR6 (AA)
3 Safety measurement cables, 1.5 m, red, blue and yellow:
banana plug – banana plug 1 Solar plug adapter, red: MC3 socket – banana socket 1 Solar plug adapter, red: MC4 socket – banana socket 1 Solar plug adapter, blue: MC3 plug – banana socket 1 Solar plug adapter, blue: MC4 socket – banana socket 1 Plug-on safety test probe with socket, red 1 Plug-on safety alligator clip with socket, yellow-gray
1 Carrying case with foam insert 1 Set of operating instructions
Warning concerning a source of danger
Non-observance may result in distorted measurement results.
Please observe!
Opening of Equipment / Repair
The equipment may be opened only by authorized service per­sonnel to ensure the safe and correct operation of the equipment and to keep the warranty valid.
Even original spare parts may be installed only by authorized ser­vice personnel.
In case the equipment was opened by unauthorized personnel, no warranty regarding personal safety, measurement accuracy, conformity with applicable safety measures or any consequential damage is granted by the manufacturer.
Measuring Categories and their Significance per IEC 61010-1
CAT Definition
Measurements in other electrical circuits
0
which are not directly connected to the mains:
e.g. electrical systems in motor vehicles and aircraft, batteries etc.
Measurements in electrical circuits
II
which are electrically connected to the low-voltage mains:
via plug, e.g. in household, office and laboratory applications
Measurements in building installations:
III
Stationary consumers, distributor terminals, devices connected permanently to the distributor
Measurements at power sources for low-voltage installations:
IV
meters, mains terminals, primary overvoltage protection devices
U
SB interface cable (PROFITEST PV
1
Find Quality Products Online at: sales@GlobalTestSupply.com
SUN MEMO only)
www.GlobalTestSupply.com

3 Introduction

Caution! Dangerous Voltage
Attention!
!
Warning!
2
1
7
8
9
34 56
10
11
Battery full
Battery weak
These operating instructions contain information and directions which are necessary for safe operation and use of the instrument. Before using the instrument, the operating instructions must be carefully read and all points contained therein must be followed correctly. Non-observance of the operating instructions, as well as warnings and directions included therein, may result in severe injury to the user or damage to the instrument.
The
PROFITEST PV
modules. The terms used in these operating instructions to desig­nate systems for generating electrical power from sunlight include: solar system, photovoltaic system, PV system and the like.
Scope of Functions
• Voltage measurement to 1000 V DC
• Short circuit current measurement to 20 A DC
• Selection of insulation test voltage: 250, 500 or 1000 V DC
• Insulation measurement to 20 M with display of limit value
• Testing of protective conductor continuity
• Earth fault measurement
• Polarity check
• Backlit LCD panel
• Optional temperature measurement
SUN
is used for simple testing of solar cells and
The instrument may not be opened, dismantled or modified in any way. The instrument may only be used with recommended accessories. The use of unsuitable accessories is impermissible.
Trade association accident prevention regulations for electrical systems and operating equipment must be adhered to during all work with the instrument.
Do not allow the instrument to warm up due to exposure to direct sunlight. Flawless functioning and a long service life cannot other­wise be assured.
The internal warming of the test instrument is measured electronically. An inpermissible overheating is signalled by ISC and °C in the display while further measurements are being disabled. Allow the test instrument to cool down before continuing measurements.

6 Use for Intended Purpose

The instrument may only be used under the conditions and for the purposes for which it has been designed. In this respect, special attention must be given to safety precautions, technical data regarding ambient conditions and use in a dry environment.

4 Transport and Storage

Please retain the original packaging for subsequent shipment of the instrument. Transport damages due to inadequate packaging are excluded from the guarantee.
The instrument may only be stored in dry, closed rooms. If the instrument is subjected to extreme temperature during
transport, it requires at least 2 hours for acclimatization before it is switched on.

5 Safety Precautions

The instrument has been manufactured and tested in accordance with IEC/EN 61010-1, “Safety requirements for electrical equip­ment for measurement, control and laboratory use”, and has been shipped from the factory in flawless technical safety condition. In order to assure that this condition is retained, the user must observe the safety precautions included in these operating instructions.
In order to prevent electrical shock, corresponding safety precautions must be implemented when working with voltages of greater than 120 V (60 V) DC or 50 V (25 V) AC TRMS. In accordance with DIN VDE, these values specify the limit for exposed (touchable) voltages (the val­ues in parentheses apply to, for example, the fields of medicine and agriculture).
Before each measurement, make sure that the measurement cables and the measuring instrument are in flawless condition. The measuring instrument may only be used within the specified measuring ranges.
If the safety of the user is no longer assured, the instrument must be removed from service and secured against unintentional use.
Safety of the user is no longer assured if the instrument:
• Demonstrates apparent damage
• No longer performs the desired measurements
• Has been stored for too long under unfavorable conditions
• Has been subjected to mechanical stressing during transport
Operating safety is no longer assured in the event of modification or alteration. The instrument may only be opened by authorized service technicians.
Never operate the instrument at a higher voltage than specified in the technical data! The instrument may otherwise be destroyed or permanently damaged.

7 Operating and Display Elements

1 LCD panel with background illumination 2 Sensor socket for temperature measurement (optional) 3 On/Off switch / background illumination 4 Low-resistance measurement R
5 Short-circuit measurement ISC /
6 General insulation resistance measurement RISO /
7 Measurement input + 8 Measurement input – / insulation measurement / continuity test 9
10 Back: battery compartment 11 Battery level indicator
The instrument may not be operated at measuring cate­gory II, III and IV mains power circuits. In accordance with EN 61010-1:2010, the instrument has been designed without a rated measuring category, and may only be used at direct voltage circuits with up to 1000 V without transient overvoltages.
/
measurement cable compensation COMP
insulation resistance measurement R
selection of insulation test voltage
Measurement input ground terminal / insulation measurement / continuity test
LOW
ISO
per EN 62446
Find Quality Products Online at: sales@GlobalTestSupply.com
GMC-I Messtechnik GmbH 3
www.GlobalTestSupply.com

8 Initial Start-Up

Attention!
!
Warning!
Caution! Dangerous Voltage
Warning!
Attention!
!
Warning!
0 0
U0: 0 V R
LOW
: 0.1
R
ISO
: ---
The PROFITEST PVSUN is supplied with power from four AA batteries (1.5 V IEC LR6). The batteries must be inserted before the instru­ment can be used.
Remove the battery compartment lid on the back of the in-
strument by loosening the two Phillips head screws. Insert the batteries. Be sure to observe the polarity symbols
which are embossed into the floor of the battery compart-
ment. Replace the battery compartment lid and retighten the
screws.
The test instrument may only be operated with installed battery compartment lid.

9 Measurements

The photovoltaic system’s plug connectors may not be disconnected under load. Please observe warnings and instructions provided by the plug manufacturer.
Solar modules continuously generate current, even when shaded. Regulations for working with live voltage must be adhered to during all work.
Disconnect all solar modules from the system before performing the measurement.
Compensation for Measurement Cable Resistance
Ohmic measurement cable resistance can be subtracted from the measurement results automatically. Proceed as follows in order to determine the resistance offset value:
Switch the PROFITEST PVConnect the measurement cables to the (–) and () sockets
and short-circuit the plug-on test probes.
Press and hold the COMP key for at least 2 seconds in order to
trigger offset resistance measurement. Symbol briefly appears. The following display appears after completion of the mea­surement: R
The measured internally and is subtracted from the actual result of all subsequent measurements, until a new offset resistance value is acquired.
If the measured value of the connected cable is > 5 the symbol
briefly appears. Compensation of measurement cable resis-
tance is not possible.
R
OFFSET
LOW
: 0.0 .
SUN test instrument on.
value is not displayed, but is permanently stored
R
LOW
Protective Conductor Resistance Measurement
Connect the (–) socket to a protective conductor terminal at
the solar module and the () socket to the equipotential bond­ing busbar.
Briefly press the R Measurement is started. The R
the protective conductor system appears at the display: Example
LOW
key.
resistance value measured for
LOW
Switching the PROFITEST PVSUN On
The test instrument is switched on by pressing the key.After the test instrument has been switched on, background
illumination can be switched on by briefly pressing the same
key once again. Briefly pressing the key a third time switches background illu-
mination back off. The test instrument can be switched off by pressing and hold-
ing the key for longer than 2 seconds.
9.1 Testing Protective Conductor Continuity
per EN 62446 (VDE 0126-23) (> 200 mA)
If a protective conductor or an equipotential bonding conductor is included at the DC side of the photovoltaic system, its electrical connection must be tested for continuity.
The connection to the main grounding terminal must also be tested.
Measurement results may be distorted due parallel con­nected impedances at load current circuits, as well as equalizing current.

9.2 Polarity Test / Measuring Open Circuit Voltage U0 per EN 62446 (VDE 0126-23)

Disconnect the solar module under test from the system before performing measurement.
Connect the PROFITEST PV
Find Quality Products Online at: sales@GlobalTestSupply.com
4 GMC-I Messtechnik GmbH
www.GlobalTestSupply.com
with the help of suitable solar plugs.
SUN test instrument to the module
Switch the PROFITEST PVSUN test instrument on.
Note
Warning!
Warning!
Attention!
!
U0: 995 V I
SC
: ---
R
ISO
: ---
U0: Pol I
SC
: ---
R
ISO
: ---
!
U0: 985 V I
SC
: 7.6 A
R
ISO
: > 20 M
U0: 870 V I
SC
: 0.0 A
R
ISO
: ---
Uiso = 1000 V
The module’s open circuit voltage U Example
In the case of incorrect polarity, a warning symbol appears to the right of the voltage display.
Example
Reverse polarity at the solar module and start testing over
again.
appears at the display.
0

9.3 Measuring Short-Circuit Current ISC per EN 62446

Disconnect the solar module under test from the system before performing the measurement.
Disconnect the solar module under test from the system before performing measurement.
Insulation measurement per EN 62446 is executed be­tween short circuited (+) / (–) and () terminals.
Connect the PROFITEST PV
with the help of suitable solar plugs.
Connect the ground terminal at the PROFITEST PV
module frame.
Switch the PROFITEST PVSUN test instrument on. The module’s open circuit voltage U
SUN test instrument to the module
SUN to the
appears at the display.
0
Selecting the Test Voltage
Press and hold the RISO/250V/500V/1000V key for at least 2
seconds.
The currently selected test voltage appears the display. Select the required insulation test voltage by briefly and re-
peatedly pressing the R
The last selected value is saved after a short period of time, and is then cleared from the display.
Example
/250V/500V/1000V key.
ISO
Connect the PROFITEST PV
with the help of suitable solar plugs.
Switch the PROFITEST PV The module’s open circuit voltage U
Press the I Open circuit voltage U
display: Example
The (+) and (–) terminals at the solar module are short circuited during the measurement, and momentary short-circuit current is measured. Insulation resistance is measured at the same time (see section 9.4).
SC
key.
SUN test instrument to the module
SUN test instrument on.
appears at the display.
0
and short-circuit current ISC appear at the
0
Executing the Insulation Resistance Measurement
Press the ISC/R
display.
The (+) and (–) terminals at the solar module are short circuited during the measurement, and momentary short-circuit current is measured. Insulation resistance is measured at the same time.
key and read the measured value from the
ISO

9.4 Testing Insulation Resistance per EN 62446

Find Quality Products Online at: sales@GlobalTestSupply.com
GMC-I Messtechnik GmbH 5
www.GlobalTestSupply.com
Minimum Permissible Limit Values for Insulation Resistance
Note
Warning!
Note
Note
Warning!
U0: 0 V I
SC
: 0.0 A
R
ISO
: < 1 M
U0: 0 V I
SC
: 0.0 A
R
ISO
: ---
Uiso = 1000 V
U0: 0 V I
SC
: 0.0 A
R
ISO
: > 20 M
U0: 144 V I
SC
: 0.0 A
Ue: 48 V
Relative to Insulation Test Voltage per EN 62446
System Voltage (U0
x 1.25)
< 120 V 250 V 0.5 M
120 to 500 V 500 V 1 M
> 500 V 1000 V 1 M
If the limit value is fallen short of, <1M or < 0.5 M appears for R
Example
Test Voltage Limit Value
ISO
.
Example
Executing the Insulation Resistance Measurement
Press the R
value from the display.
Example
/250V/500V/1000V key and read the measured
ISO

9.5 General Insulation Resistance Measurement

The PROFITEST PVSUN test instrument is additionally equipped with system-independent insulation testing. This is helpful, for exam­ple, when testing systems or supply lines when no modules have yet been installed. All insulations tests known to date can be per­formed.
The test object must be potential-free. Assure absence of voltage with the help of a suitable measuring instru­ment,
for example the METRAVOLT 12D+L voltage tester.
If the minimum permissible limit value is fallen short of, <1M or < 0.5 M appears for R
ISO
.

9.6 Ground Fault Test

If the test instrument recognizes an inadmissible, low insulation value during I sible ground fault.
If a ground fault occurs within the system, the error can be nar­rowed down on the basis of the voltage ratio.
Disconnect all solar modules from the system before performing the measurement.
measurement, the instrument tests for pos-
SC/RISO
General insulation measurement is executed between short circuited (–) and () terminals.
Connect the (–) and () terminals at the PROFITEST PV
device under test.
Switch the PROFITEST PV
SUN test instrument on.
Selecting the Test Voltage
Press and hold the RISO/250V/500V/1000V key for at least 2
seconds.
The currently selected test voltage appears at the display. Select the required insulation test voltage by briefly and re-
peatedly pressing the R
The last selected value is saved after a short period of time, and is then cleared from the display.
Find Quality Products Online at: sales@GlobalTestSupply.com
6 GMC-I Messtechnik GmbH
/250V/500V/1000V key.
ISO
www.GlobalTestSupply.com
SUN to the
Connect the PROFITEST PV
suitable solar plugs.
Connect the ground terminal at the PROFITEST PV
module frame.
Switch the PROFITEST PVSUN test instrument on.
Momentary open circuit voltage is displayed. At the same time, a second voltage (Ue) appears in the event of a
ground fault. Example
SUN to the module with the help of
SUN to the
10 Data Storage with PROFITEST PV
Note
Note
2
1
16
15
14
45 67
1383910 1112
Battery full
Battery weak
UOC=0V I
SC
=---
R
ISO
=---
UOC=0V I
SC
=---
R
ISO
=---
? OBJ:Objekt1 (1/1) STR:String1 (1/1)
20.02.2012 10:00:00 U
OC
=0V ISC=0.0A
R
ISO
=>20M R
low
>10 E=--- T=--­=--- Ib=---
STR:String West (1/1)
20.02.2012 10:00:00 U
OC
=0V ISC=0.0A
R
ISO
=>20M R
low
>10 E=--- T=--­=--- Ib=---
SUN MEMO
The PROFITEST PVSUN MEMO test instrument is equipped with an internal memory for the storage of 10,240 data records. The stored data can be read out via a USB port for subsequent evalu­ation from a PC. Furthermore, the included PC software allows for the creation of individual test structures.

10.1 Operating and Display Elements

1 LCD panel with background illumination 2USB interface 3 Sensor socket for temperature measurement (optional) 4 ON/OFF switch / background illumination 5 Low-resistance measurement R
compensation COMP
6 Short-circuit measurement I
measurement R General insulation resistance measurement
7
per EN 62446
ISO
selection of insulation test voltage
/ measurement cable
LOW
/ insulation resistance
SC
R
/
ISO
8 Battery level indicator
9 Selection of objects or strings 10 Switch between display / activation of storage function 11 Switch between objects and strings 12 Selection of objects or strings 13 Back: battery compartment 14 Measuring input ground terminal () /
insulation measurement / continuity test 15 Measuring input – / insulation measurement / continuity test 16 Measuring input +

10.2 Function Description

Switch between standard measurement display and storage display
Standard measurement display without memory function
The storage function is not activated before switch-over to storage display, see screenshot below.
Standard measurement display with activated storage function.
If a measurement is per­formed, the measured values are stored under the object and string selected before.
Stored values are overwritten in the activated object/ string.
Storage display
OBJ:Object1 / STR:String1 are active. Presently, one object (1/1) and one string (1/1) have been created.
Date / time
Measured values currently stored (meaning of abbreviations): U
Open-circuit voltage
OC
ISC Short-circuit voltage
Insulation resistance
R
ISO
Low-resistance measurement
R
low
E Solar irradiation (optional with PROFITEST PV SUN-SOR) T Module temperature (optional with PROFITEST PV
SUN-SOR)
Inclination angle of module (optional with
PROFITEST PV
Operating current (optional with current clamp)
I
b
? OBJ:Testkunde (1/1)
SUN-SOR)
(Alphanumeric) names can be assigned to objects and strings with the PC software (e. g. OBJ: test client).
Switch to previous object or string
If objects and strings are extended in the test instrument, it is not possible to assign alphanumeric characters. There is only an incrementation instead (gradual increase by one character each).
Switch to next object or string. If the last object or string is reached, a new object or a new string can be created by pushing this key (> 2 s).
Switch between objects and strings. Symbol indicates whether objects or strings are active (e. g.: OBJ:Object1).
Find Quality Products Online at: sales@GlobalTestSupply.com
GMC-I Messtechnik GmbH 7
www.GlobalTestSupply.com

11 Application of PC Software

Attention!
!
Individual test structures can be created at the PC with the included software. This structure makes it possible to assign measurements to the objectes and strings of different units and buildings. For use at the construction site, the respective test structures are first transferred to the test instrument via the USB interface. After performing the measurements, the test instrument is read out and the structures are represented together with the measured values in tabular form.

11.1 System Requirements Hardware

• Processor and RAM see minimum requirements of the respective operating system
• VGA monitor with a resolution of at least 1024 x 768 pixels
• Hard disc with at least 12 MB available memory
• Both input devices (mouse / keyboard)
• One or several USB interfaces
Operating System – Software
• Windows XP, Windows Vista, Windows 7 or 8

11.2 Installation

The latest version of the PC software PV SUN MEMO Control Center is available for free download as a zip file from our homep­age:
Products Electrical Testing Testing of Electr. Installations PROFITEST PV
The zip file must be unpacked and stored in a directory. The setup file has been placed in this directory. Double-click on the setup to launch the installation routine. An assistant guides you through the complete installation.
The installation directory and the program file can be selected at your own discretion.
SUN MEMO

11.4 Creation of Structures and Transfer to the Test Instrument

Create a new structure by clicking on icon „New“ under „Measur­ing Data“.
Clicking on „New“ deletes the structures and measuring data indicated before.
You can freely extend a structure by clicking on icon „+Object“ or „+String“. Repeat these steps until you have created the requested structure. By clicking on the object or string denomina­tion you can change the name at your discretion (alphanumeric).
The number of possible data records = objects which can be transferred to the test instrument, amounts to approx. 10,000.
Click on the „Write“ icon for transferring the data records to the test instrument.
Setting of Date/Time
Date/time in the test instrument are synchronized automatically upon importing the file (structure) from the PC.

11.5 Reading Out and Exporting Test Data

After performing the measurements, click on icon „Read“ for reading out the data records (structures and measured values) from the test instrument. The structures are presented in tabular form together with the measuring data. The column width of the header elements can be adjusted by dragging the column border lines to the right or to the left.

11.3 Initial Startup of the Device

Switch on the test instrument and connect it with your PC. Windows announces that a device has been connected and that
the driver must be installed. Select manual installation „Install software from a list or a
specific source“.
Click on „Search“ in the next menu and select directory
„C:\Programme\PVSUN MEMO Control Center\USB Driver“ (unless you have changed the installation path).
Ignore the message regarding unknown provenance of the
software.
Complete the driver installation.Please check: „PVSUN MEMO Memory (COMxx)“ should be
displayed in the device manager under Connections.
Start the installed program „PVSUN MEMO Control Center“. Click on icon „Search device“.
For storage and subsequent evaluation of the measuring data, press icon „Export“. The data are written into a csv file so that they can be exported to other programs, such as Microsoft Excel.
Deleting of Structures
The structural contents are automatically deleted when an empty structure is transmitted to the test instrument.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
12 Measurement of irradiation, temperature and
Warning!
U0: 0 V I
SC
: 0.0 A
R
ISO
: ---
: 1.0
E: 1000W/m² T:25°C
inclination with PROFITEST PVSUN-SOR (optional)
Hot surfaces may cause burns.
Connect the PROFITEST PVPlace the PROFITEST PV
voltaic module.
Switch the PROFITEST PVSUN / PROFITEST PVSUN MEMO on. The display shows the monentary irradiation, the surface tem-
perature and the inclination angle of the modules.
Example:
SUN-SOR to the sensor socket.
SUN-SOR onto the surface of the photo-
Characteristic Values PROFITEST PVSUN-SOR
Irradiance Measuring range 0 … 2000 W/m²
Resolution 1 W/m² Cell monocrystalline, approx. 27 x 16 mm Accuracy 15% (T = 25 C)
Temperature Measurement Measuring range –55…99 C
Inclination Angle Measuring range 0 … 90
Cable length approx. 10 m Power Supply via PROFITEST PV
Resolution 1 C Accuracy 1 C
Resolution 0,1 Accuracy 1
SUN /
PROFITEST PVSUN MEMO
approx. 3,3 V / 3,1 mA Dimensions approx. 120 x 65 x 27 mm Weight approx. 150 g

13 Characteristic Values

Voltage Measurement U0
Measuring range 0 to 100 V to 1000 V DC
(no transient voltages) Resolution 1 V Accuracy (2% rdg. + 1 d)
Current (direct)
Measuring range 0 to 20 A DC Voltage range 2 to 1000 V DC Resolution 0.1 A Accuracy (1% rdg. + 1 d) Overcurrent protection
Max. 24 A (shutdown of internal circuit)
Insulation Resistance Measurement RISO
Test voltage 250 V DC 500 V DC 1000 V DC Measuring range 0.5 M Resolution 0.1 M 1 M 1 M Accuracy (1% rdg. + 1 d) (1% rdg. + 2 d) (1% rdg. + 2 d) Limit value > 0.5 M No. of measurements Approx. 1000 (with battery set per IEC LR6)
to 1 M1 M to 20 M1 M to 20 M
> 1 M > 1 M
Earth Fault Measurement
Measuring range 0 to 1000 V DC Resolution 1 V Accuracy (5% rdg. + 1 d)
Low-Resistance Measurement
Measuring range 0 to 10 Test current > 200 mA Resolution 0.1 Accuracy (1% rdg. + 1 d) No. of measurements Approx. 500 low-resistance measurements
(batteries: 1.5 V per IEC LR6)
Display
LCD Backlit multiple display
Dot matrix: 128 x 64 pixels
Reference Conditions
Ambient temperature +23 C 2K Relative humidity 40 to 75% Battery voltage 6 V 1V
Ambient Conditions
Operating temperature 0 to +40 C
Storage temperature –10 C to 60 C Relative humidity < 80%, no condensation allowed Elevation Max. 2000 m above sea level
Power Supply
Batteries 4 ea. 1.5 V IEC LR6, AA, AM3, MN1500 Consumption Approx. 20 μA when switched off
Approx. < 30 mA during normal operation Approx. 190 mA with backgr. illumination
Electrical Safety
Voltage range 1000 V
Electromagnetic Compatibility (EMC)
EMC directive EMC 2004/108/EC Basic standard EN 61326-1:2006
Data Memory (PROFITEST PVSUN MEMO only)
Storage capacity max. 10,240 data records Memory type non-volatile flash memory (data are pre-
served even if batteries are depleted)
Data Interface (PROFITEST PVSUN MEMO only)
Type USB 2.0 (USB 2.0 full speed (12 Mbps)
Connection mini-USB type B (5-pin) Cable USB 2.0 cable (A plug | Mini-B-)
compatible)
Mechanical Design
Protection Housing: IP 42
Dimensions 209 x 98 x 35 mm Weight Approx. 500 g with batteries
per DIN VDE 0470 part 1/EN 60 529
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
GMC-I Messtechnik GmbH 9