Gossen Metrawatt AT3-I I S Operating Instructions

Operating Instructions
Attention!
!
Note!
!
AT3-II S
CEE Adapter 16 (20) A for Testing Protective Conductor Current in Operating Condition
3-349-337-03
5/7.14

1 Application

With the AT3-II S CEE adapter, devices which are equipped with a 16A/6h or 32A/6h 5-pin CEE plug can be quickly and efficiently connected to test instruments designed for testing portable devices in accordance with DIN VDE 0701-0702, or for testing in accordance with IEC 62353 (VDE 0751), and which are furnished with only an earthing contact outlet.
The following tests can be performed on devices with CEE plugs with the help of the AT3-II S adapter:
• Testing of protective conductor continuity
• Insulation resistance measurement
Measurement of Protective Conductor Current in measuring mode – direct (direct method via test socket) – differential (differential current measuring method)
alternative (equivalent leakage current measurement method)
Application Restrictions The AT3-II S CEE adapter may only be used for test purposes in combination with a test instrument. Other use of the adapter, in particular for continuous connection of 3-phase current devices to 3-phase electrical systems, is not permissible under any circumstances! In any case, maximum permissible conventional thermal current is 16 A AC per phase.

2 Safety Features and Precautions

The AT3-II S CEE-Adapter is manufactured and tested in accordance with safety regulations IEC 61010-1 / EN 61 010-1 / VDE 0411-1. If used for its intended purpose, the safety of the user and of the device is assured.
Carefully read the operating instructions before use, in particular the operating instructions included with the test instrument with which you intend to use the adapter. Follow all instructions contained therein.
The adapter may only be used for testing devices with 5-pin CEE plugs and current consumption of no greater than 3 x 20 A (AC-1).
The AT3-II S CEE adapter may not be used:
• With open housing
• If it demonstrates visible damage
If the CEE outlet or the connector cable is damaged
• After excessive stress, i.e. if the load capacities specified in the technical data have been exceeded
• After long periods of storage under unfavorable conditions (e.g. humidity, dust or extreme temperature)
Meanings of Symbols on the Instrument
Warning concerning a point of danger (attention: observe documentation!)
CAT II Measuring category II device
Indicates EC conformity Harmlessness with regard to interference emission and interference immunity in accordance with EN 61326:2000 / EN 61000-2 / EN 61 000-4
The device may not be disposed of with the trash. Further information regarding the WEEE mark can be accessed on the Internet at www.gossenmetrawatt.com by entering the search term WEEE.

3 Connecting the AT3-II S

Before testing can be started, the AT3-II S must be connected to the test instrument and to mains power. Proceed as follows:
Ð Connect the mains cable from the test instrument to the
earthing contact outlet labelled “Appliance tester” on the AT3 -I I S.
Ð Connect the earthing contact cable from the AT3-II S to the
test/mains outlet at the test instrument. The connection at the AT3-II S is labelled “to appliance tester”.
Prerequisite for differential current measurement
For the measurement of protective conductor current with differential current method you require a test instru­ment which is equipped with voltage inputs.
Ð Insert the 4 mm plugs on the cable from the AT3-II S, which is
labelled “tester Bu. 2 & 3”, into sockets N (2) and L (3) at the test instrument.
Ð Plug the CEE connector cable from the AT3-I I S into a suitable
16A/6h CEE outlet within your electrical installation. The connection at the AT3-II S is labelled “Mains, 3~ 400 V, 16 A
50...60 Hz”. As of this point in time, the AT3-I I S and the test instrument are supplied with electrical power.
Ð Finally, connect the device under test to the CEE outlet at the
AT3-II S. The outlet or coupling is labelled “Test-socket appliance”.
Testing can now be started (see section 4).

4 Testing with the AT3-II S

You can now execute all tests on devices with CEE plugs which are otherwise possible with the test instrument, and which are supported by the AT3-I I S adapter, just as you would for devices with earthing contact plugs. Proceed as described in the operat­ing instructions for the test instrument to this end.
The following characteristics of the AT3-I I S adapter must be taken into consideration when performing tests with the test instrument:
Testing Protective Conductor Resistance
When testing protective conductor resistance, measured protective conductor resistance is increased by the amount caused by the protective conductor system of the AT3-II S CEE adapter itself. In case of doubt, i.e. if measurement results are close to the permissible limit value, measure the adapter’s protective conductor resistance at the PE terminal of its CEE outlet, and subtract the results from the measured value obtained for the DUT-adapter system.
Insulation Test
During insulation testing, the three phase terminals (L1, L2 and L3) and the neutral terminal (N) at the DUT are short-circuited, and testing is then executed.
AT3-II S
Attention!
!
Attention!
!
Attention!
!
Attention!
!
CEE Adapter 16 (20) A for Testing Protective Conductor Current in Operating Condition
Residual Current Measurement
During 3-phase residual current measurement, the resulting residual current is transmitted via the test cable to the test instrument (sockets N (2) and L (3)) as a voltage value, and recorded.
– In the case of automatic testing in accordance with standards
Ð Initialize the standard-specific setup. Ð Select “Sequence” and deactivate the auto-test method.
During testing in accordance with standards, a menu then appears with which either equivalent leakage current or residual current can be selected.
Ð Select “residual current”. The DUT is connected to line voltage via the AT3-I I S. The DUT’s
residual current is measured, and it can be tested for correct functioning at the same time.
Function Test
Before executing the function test, or any other tests for which the
DUT is supplied with mains power and must be placed into operation, it must be assured that there are no short-circuits in the DUT between phase conductors L1, L2 and L3 and/or neutral conductor N! Nonobservance may result in damage to, or destruction of the AT3-I I S adapter, and possibly the device under test as well!
The DUT’s power and current consumption values CANNOT be deter­mined during function testing with the AT3-II S adapter. The dis­played values indicate power consumption for the AT3-II S adapter.
Always start the test with the test instrument (orange lamp on the AT3-II S lights up) before switching the device under test on! In the case of extreme initial current at the DUT, nonobservance of this start-up sequence may result in damage to, or destruction of the AT3-II S adapter!
Disabling of Protective Devices – Max. Current Consumption
As soon as the orange lamp at the AT3-I I S lights up, the terminals at the mains outlet from which power is supplied are connected to the test/mains outlet or coupling at the AT3-I I S adapter without the use of a fuse or any other protective devices. Make sure that only devices designed for use with 3-phase current with a maximum current consumption of 3 x 20 A (AC-1) are connected to this outlet or coupling.
Testing the DUT for Short-Circuits Short-circuiting of two or all three phases of the device under test cannot be detected by the test instrument until after residual current testing has been executed! In such cases, the AT3-I I S adapter may be severely damaged as a result of current surges when this test is executed!
Before connecting a device to the AT3-I I S, make sure that there are no short-circuits between the DUT’s phases by means of appropriate testing, for example with a continuity tester!
Connection of the DUT in Correct Phase Sequence
In particular after repairing 3-phase devices or connecting new CEE plugs to such devices, make sure that the phases are connected in the correct sequence.
Test Sequence – Initial Current Problem – Starting the test
Always start the test with the test instrument (orange lamp on the AT3-II S lights up) before switching the device under test on!
– Ending the test
Switch the DUT off first, and then stop testing at the test instru­ment.
In the case of extreme initial current at the DUT or inductivity within the electrical circuit, nonobservance of this sequence may result in damage to, or destruction of the AT3-II S adapter!

5 Characteristic Values

Electrical Safety
Safety class Operating voltage 300 V Test voltage 2.2 kV Current-carrying
capacity Proprietary connector cable
“mains active” 7 VA, cos ϕ ∼ 0.4 Measuring category II Contamination degree EMC EN 61326:2000/EN 61000-2/EN61000-4
Residual Current
Measuring range 0.08 to 10.0 mA AC Intrinsic error 4% rdg. ±40 μA Measuring error 6% rdg. ±60 μA
Mechanical Design
Protection Housing: IP 40, connections: IP 20 Table Excerpt Regarding Significance of the IP Code
IP XY
st
(1
digit X)
Dimensions W x H x D: 290 x 120 x 105 mm
Weight 2.4 kg
Protection against foreign
2 12.5 mm dia. 0 not protected 4 1.0 mm dia. 0 not protected
I per IEC 61010-1/EN 61010-1/VDE 0411-1
16 A / 6 h, 3-phase, 20 A (AC-1) KB 10 min.
2
IP XY
nd
object entry
(without cables and grommets)
(2
digit Y)
Protection against the
penetration by water
Edited in Germany Subject to change without notice A PDF version is available on the Internet
GMC-I Messtechnik GmbH Südwestpark 15 90449 Nürnberg •
Germany
Phone +49 911 8602-111 Fax +49 911 8602-777 E-Mail info@gossenmetrawatt.com www.gossenmetrawatt.com
Loading...