Goodlife Cloudia User Manual

Page 1
User Guide
Blood Glucose Monitoring System (Cloudia)
Page 2
Contents
About GoodLife 4 About the System 5 Important Information 6 Important Safety Instruction 8 Blood Glucose Meter 10 Meter LCD Window 12 Blood Glucose Test Strip 14 Glucose Control Solution 17 Check Strip 20 Setting Meter Parameters 22 Performing Blood Test 26 Understanding Your Test Result 31 Data Transmission 33 Memory Recall 35 Replacing the battery 36 How to Clean and Disinfect 38 Storage and Handling 46 Display Messages and Problem-Solving Guide 47
Performance Evaluations
System Specication
54
FCC Regulations 66 Warranty 70
How to view transmitted reading
58
52
iglucose Diabets Management
5456
Please read this User Guide before using your GoodLife Blood Glucose Meter (Cloudia). If you have any questions or enquiries, please contact customer support at 1-855­692-3511 for assistance.
Page 3
4 5
About GoodLife™
About the System
Intended Use
Principle of the Test
GoodLife Blood Glucose Monitoring System(Cloudia) is intended to be used for the quantitative measure­ment of glucose (sugar) in fresh capillary whole blood sample drawn from the ngertips only. It is intended to be used for single patient and should not be shared. It is not indicated for the diagnosis or screening of diabetes or for neonatal use.
The GoodLife System is an electrochemical biosensor system that measures the amount of electric current produced and displays the result as a blood glucose level.
The GoodLife Blood Glucose Monitoring System (Cloudia) (GoodLife System) is designed to provide blood glucose measurements with easy and comfort­able testing. The system requires 0.5μL of blood sample and 5 seconds for the test to complete.
The GoodLife System consists of: GoodLife Blood Glucose Meter (Cloudia), GoodLife Blood Glucose Test Strips (Cloudia), Check Strip, GoodLife Glucose Control Solution(Level I & II), Safety Lancet (optional for
purchase).
These products are intended to be used together to get accurate blood glucose test results. Do not use other test strips or control solutions with your meter.
NOTE:
Not for use on patients who are dehydrated, hypo­tensive, in shock, or in a hyperosmolar state.ÿ
Page 4
The GoodLife System is intended for in vitro diagnos­tic use with capillary whole blood. The system should not be used for diagnosis and screening of diabetes or for testing newborn infant (neonatal testing).
CAUTION
The test result does not suggest any medical indication. The user should consult his or her clinician for further diagnostic and treatment. Call your doctor immediately if you experience symptoms that are not consistent with your blood glucose test results. Severe dehydration or excessive water loss may cause false, high results. Call your doctor right away if you suspect you are suering from dehy­dration. A sample with large amount of reducing substances such as triglycerides(>1000mg/dL), ascorbic acid (>2.25mg/dL), uric acid (>15mg/dL) and xylose (> 50mg/dL).
1.
2.
3.
4.
A red blood cell count (hematocrit) that is either very high (over 55%) or very low (under 30%) can cause false result. High altitudes above than 8,800 feet (2,750 meter) may aect the test results. Temperatures outside the range of 50°F to104°F (10°C to 40°C), 20%-80% RH may aect the test results. Do not use GoodLife system to test on critically ill patient. Not for use on patients who are dehydrated, hypotensive,in shock, or in a hyperosmolar state.
5.
6.
7.
8.
9.
6
7
Important Information
Triglyserides
Ascorbic acid
Substance
Highest c oncentration
without interference
( mg/dL)
Toxic
Concertration
( mg/dL)
Theraprutic
Physiological
Levels N/A N/A
N/A N/A
0.40 - 2.00 150 - 500
2.52 - 8.00
57
Uric acid
Xylose
2.25
1000
15
50
Theraprutic/
Page 5
8 9
Important Safety Instruction
The meter and lancing device are for single patient use. Do not share them with anyone including other family members. Do not use on multiple patients. Because all parts of the kit may come in contact with your blood, all parts of the kit are considered biohazardous and can potentially transmit infec­tious diseases. Please refer to "How to Clean and Disinfect" section to help mitigate biohazard. Users should wash hands thoroughly with soap and water and dry thoroughly after handling the meter, test strips and any lancing device.
For customer support contact at 1-855-692-3511. (Mon-Fri 9:00 am~4:30 pm Pacic Time)
NOTE:Please make sure that all products listed on the "contents" of the box are contained in the package before using this system. If you nd any imperfection in our products, please contact customer support at 1-855-692-3511. (Mon~Fri 9:00am~4:30pm Pacic Time)
1.
2.
3.
“FDA Public Health Notication: Use of Fingerstick Devices on More than one Person Poses Risk for Transmitting Bloodborne Pathogens: Initial Com­munication” (2010)
http://www.fda.gov/MedicalDevice/ Safety/AlertsandNotices/ucm224025.htm
“CDC Clinical Reminder: Use of Fingerstick Devices on More than One Person Poses Risk for Transmitting Bloodborne Pathogens” (2010)
http://www.cdc.gov/injectionsafety/Fingerstick­DeviceBGM.html
References
1.
2.
Page 6
10 11
Blood Glucose Meter
DISPLAY: The large, easy to read display shows the test results, mes­sages, blood glucose re­sults stored in memory, time and date.
Mem Button: Press Mem but­ton to enter memory mode to recall the information stored in meter’s memory.
Set Button: Press Set button to enter date and time setting.
STRIP SLOT: Holds a Blood Glucose Test Strip or Check Strip in place when you perform blood glucose test or perform check test.
: Holds
-
Button: For data transmission
3L-02-XXXXXX V1.0
Blood Glucose Monitoring System
Cloudia
For Self-testing Battery:AAA,6V
HMD BioMedical Inc. Hsinchu , Taiwan
serial no.:
XXXXX-0000001
SIM:
8901204043780370392
For In Vitro Diagnostic Use Only FCC ID:
AYY0000001
3L-02-XXXXXX V1
SIM
AAA AAA AAA AAA
SMM
AAAAAA
AA
3L-02-810035 V1.0
Blood Glucose Monitoring System
Cloudia
For Self-testing
Battery:AAA,6V serial no.: 65534-XXXXXXX
For In Vitro Diagnostic Use Only
FCC ID: AYY0000002
Distrubutor: Kannact Inc. / KRemedy 10260 SW. Greenburg Road Suite 525, Portland OR 97223
Battery Compartment four AAA batteries. Before us ing the meter, please install
Meter SN Label: The label shows the meter serial number.
STRIP EJECTOR: For easy ejection of the test strip.
Page 7
12 13
Meter LCD Window
DATE: Month/Year/Day and the temperature at the moment of the test.
THERMOMETER SYMBOL: Appears when ambient temperature is above or be­low the acceptable range needed for testing.
CODE: A code number should appear before me­ter is used to test sample.
Ř
Meter is at a Mem-
ory mode. Prior analysis results can be read.
Symbol:The test can be adjusted and stored into 3 types of test modes - Before Meal Test (
), After Meal Test(
) and Control Solution Test(
).
Alarm Symbol: Appears when alarm is set.
Symbol: Appears
when data transmission.
Test Result
Time: 12 and 24 hour period format.
mg/dL: The units of measurement in the U.S. are mg/dL.
Ř
Memory capacity is 999 entries. The display shows to
ensure that all digits are working properly.
Strip Symbol: Appears when the strip is inserted and meter is ready for blood testing. Replace vial cap after removing test strips.
Battery Symbol: Appears when battery is low.
Page 8
14 15
(correct) (incorrect )
Blood Glucose Test Strip
Aperture: apply the blood to the front of the strip.
Black Electrode End: Insert this end of test strip into meter.
Blood Reaction Zone: It must be lled with blood entirely.
GoodLife System measures the amount of glucose in capillary whole blood. Blood can be applied to the front of the test strip and is auto­matically drawn to the test strip through capillary action. GoodLife Blood Glucose Test Strips (Cloudia) are intend ed for in vitro diagnostic use with capillary whole blood or GoodLife Glucose Control
1.
2.
Solution. Results will not be accurate if used with plasma or serum samples. Do not use test strips beyond the expiration date indicated on the strip vial label. The discard date for test strips is 90 days after rst opening the vial. Record the discard date on the vial when you open a new vial of test strips. Blood Glucose Test Strip can be damaged by heat and light. Keep them sealed in the original vial. Store the vial in a cool, dry place between 104°F(40°C) and 50°F (10°C). Do not refrigerate. Do not use damaged test strips in any way. Use test strip immediately after taking it out from the vial or foil packet; replace the vial cap and close it tightly.
3.
4.
5.
6.
7.
Page 9
16 17
Glucose Control Solution
IMPORTANT INFORMATION
Do not transfer test strips to a new vial. Always carry test strips in their original vial. Do not place in direct heat or sunlight. Do not carry loose test strips in your carrying case. Test strips are for single use only.
8.
9.
10.
11.
GoodLife Glucose Control Solution is used to check if the GoodLife Blood Glucose Monitoring System(Cloudia) and Test Strip are working correctly as a system. Users should perform a control solution test when:
a new vial of test strip is opened. a meter is dropped whenever results are not consis­tent with patient’s symptoms.
1.
2.
3.
Do not use control solution beyond the expira­tion date indicated on the bottle label. The discard date for control solution is 90 days after rst opening. Record the discard date on the bottle when you open a new bottle of control solution.
1.
2.
Page 10
18 19
Test result will show up in 5 seconds. The result should correspond to the range printed on the label of strip vial used.
NOTE:
Repeat test if the result falls outside the control range. If you continue to get the result falling outside the control range, the potential causes are
1. the test strip is damaged or get moist
2. the test strip is expired.
3. the meter and strip may not be working properly. Do not use the system to test your blood until you get a test result falls within the control range. Contact customer support at 1-855-692-3511 or your healthcare provider for help.
5.Store the control solution closed at temperatures between 50ºF (10ºC) and 86ºF (30ºC). Please carefully read the label before use. For purchase of control solution, please contact your local distributor or customer support at 1-855-692-3511 for more information.
3.
4.
5.
Perform Control Test
Insert a new test strip into the strip slot, the meter will activate. The code number will appear on the screen. Compare the code number shown on the screen against the code number on the test strip vial. If the two numbers match, you may begin test, otherwise contact customer support at 1-855­692-3511 or your healthcare provider for help. When shows up, press Set button and choose control solution test. Gently shake the control solution and apply a drop to the aperture of strip. Make sure that the control solution has saturated the blood reaction zone.
1.
2.
3.
4.
Page 11
20 21
Check Strip
CHECK
How to check meter by check strip
How to delete memory
The Check Strip can be used in 2 ways:
1.2.To test only the function of the meter and not complete BGM test system. Please check the complete test system with Control Solution To delete all test memories.
Insert the check strip into strip slot with label side up as above. You should obtain an “OK“ reading within 10 seconds, which means your meter is working properly. Remove the check strip to exit. Meter will auto­matically turn o.
NOTE
:
If you do not get "OK" reading but appear other error message, turn o the meter by remove check strip from the meter. Then check the battery and repeat the test. If the second result persists, contact customer support at 1-855-692-3511 or your healthcare provider.
1.
2.
Insert check strip into strip slot with label side up. After “OK” displayed, press and hold the Set button until ashing “dEL” shows up with a beep sound. Press and hold the Mem button until you hear a beep sound, Meter will display “OK” before turning o and all the memory has been deleted successfully. Remove the check strip from the meter.
1.
2.
3.
4.
Page 12
22 23
Setting Meter Parameters
Setting Time & Date
Setting Temperature Unit/ 12h,24h period/ Alarm
Press the Mem button, the meter will enter to the setting of “year”. Press Set button to select the desired year. Press the Mem button to conrm and shift to the next setting. Repeat above steps to set the month, day, hour and minute by Set and Mem buttons. After minute is set, the meter will display “OK” before turning o.
Note: While setting the time and date, you can exit the setting mode anytime by pressing Mem button for 3 seconds.
1.
2.
Press the Set and Mem buttons at same time for 2 seconds to turn on the meter.
1.
When you begin to use the meter or replace a new battery, please follow the steps below to set the date and time manually.
Celsius and Fahrenheit can be set according to your preference.
Then press the Set button to choose between “C” (for Celsius) and “F” (for Fahrenheit). When Temperature Unit preference is set, press the Mem button to enter the 12h/24h period set ting. Press the Set button to choose your preference of 12h/24h period. When 12h/24h period is set, press the Mem button to enter the alarm setting. Press the Set button to turn on or turn o the alarm. The GoodLife has 5 alarms to keep your blood glucose monitoring on time.
2.
3.
4.
5.
6.
When “ON“ is choosen, press the Mem button to congure the alarm hour. Press the Set button to change hour. When alarm hour is set, Press the Mem button to congure the alarm minutes. Press the Set button to change minutes and press the Mem button to enter next setting. When “OFF“ is chosen, press the Mem button to enter next setting.
t
t
Page 13
24 25
Setting Auto Data Transmission
To congure the other 4 alarms, please repeat step 6. After setting, the meter will display “OK“ before turning o.
7.
8.
Press and hold the Set button for 2 seconds to turn on the meter when “ON“ show up, it means auto transmission. Press the Set button to choose between “ON“(auto) and “OFF“ (manual).
1.
2.
When the preference is set, press the Mem button, the meter will display “OK“ before turning o.
Page 14
26 27
Performing Blood Test
1. Wash your hands in warm, soap water. Rinse and dry completely.
!!
2.
Set
Mem
3.
9
4. When the you are ready to perform a test.
5. If the test is done within 2 hours of a meal, press the Set button to change the setting from
to . If the test is a control solution test, press the Set button again to change the setting from
to
.
6. Use safety lancet or lancing device to obtain the required blood volume. To perform the test, you need 0.5 uL of blood sample.
Remove new test strip from vial. Be sure to tightly replace vial cap after removing test strips. Insert test strip immedi­ately into strip slot as illustrated. The meter turns on automatically.
Check if the code number on the meter matches the code on the vial. If the 2 numbers match, you may begin blood testing. Otherwise insert another new one. If the code number still doesn’t match, please contact customer support at 1-855-692-3511 for help.
Caution: To avoid the blood borne pathogen transmis­sion, the lancing device is for single patient use. Do not share with anyone else including other family members. Do not use on multiple patients. NOTE: Please refer to safety lancet or lancing device user manual for further instructions on use and handling.
Page 15
28 29
7.
Mem
8. Test result will show up in 5 seconds.
9. Before removing the strip from the meter, you can press Set button to set the test as BEFORE MEALS TEST
, or AFTER MEALS TEST , or a
CONTROL SOLUTION TEST
.
Set
Mem
10. After setting complete, remove the strip from meter. Please follow the healthcare provider’s recommendations for disposal of used lancets and strips.
Set
Mem
11.
Set
Mem
WARNING: Inaccurate results may occur in severely hy ­potensive individuals or patients in shock, inaccurate low results may occur in individuals experiencing a hyperglyce­mic/hyperosmolar state, with or without ketosis, and criti­cally ill patients should not be tested with the meter.
Important Safety Information
* The meter is for single patient use. Do not share them with anyone including other family member.
Apply the blood sample to the front aperture of test strip in a way that comfortab.e for you. Make sure that the blood drop has saturated the blood reaction zone.
After removing the strip, the result will be transmitted to data­base through GSM system auto­matically. It will diaplay “OK” if the transmission succeeds. Press the “Mem” button to turn the meter o. If “Err” appears, please refer to “Display Messages and Problem-Solving Guide.
Page 16
30 31
Understanding Your Test Result
Expected range for people without diabetes :
Time of Day Expected Range, Non-Diabetes
Before Meals Less than 100 mg/dL
After Meals Less than 140 mg/dL
Source: American Diabetes, Association Position Statement, Diabetes Care Vol.37 (Suppl.1) p.s16 (2014).
Consult your healthcare professional to nd out your target blood glucose value.
If your blood glucose result seems unusually high or low, or inconsistent with your previous results, check the following:
1. Does the code number on the test strip vial match the code number on the meter?
2. pirts tset eht ot deilppa elpmas doolb eht saW immediately after the strip was removed from
Was the size of the blood sample sucient?
the vial?
3.
* All parts of the kit are considered biohazardous and canpotentially transmit infectious disease. Please refer to “How to Clean and Disinfect“ section to help mitigate biohazards. * Safety Lancet is for single use only.
Page 17
32 33
4. Was the test strip vial cap tightly sealed?
5. Was the test strip used before the expiration date?
6. Were the test strips stored away from extreme temperatures or from areas of high humidity?
Data Transmission
Then run a quality control check with your Glucose Control Solution and a new test strip. If the control test result is within the acceptable range, review testing procedure and repeat your blood glucose test with a new test strip. If your blood glucose value is still inconsistent with your previous results, contact your doctor immediately for help.
When the meter is set as “auto transmission” (refer to the section “Setting Auto Data Transmis­sion”), the test result will be transmitted to the database through GSM system after every time you test. When the meter is set as “manual transmission”, press es and holds the button for 2 seconds to turn the meter on.
The test result could be transmitted to a secure web database for further management.
NOTE:
1. It is recommended to transmit in good reception areas.
1.
2.
When the meter displays “Err”, please refer to the section “Display Messages and Problem-Solving Guide“.
3.
If there is no record in the meter, it will display “M0”, and then “OK“ before turning o. If there are records in the meter that haven’t been transmitted, it will display “OnL“ and transmit the data. After that, it will display “OK“. Press the Mem button to turn o or allow the system to auto turn o after 3 minutes of inactivity.
t
t
Page 18
34 35
Memory Recall
To recall Results Stored in Memory
2. By pressing Mem button, you will see your test record from the latest to the earliest.
3.
Press and hold the Mem button for 2 seconds to turn the system o or allow the system to auto turn o after 30 seconds of inactivity.
For example:
Pacemaker (Keep a safe distance more than 15cm), Medical telemetry or telemetry equipment (Telemetry), Physiological signal sensing or monitoring device (such as ECG, EMG, EEG, ERG, ENG, MEG, Holter, asphyxia monitors, etc.), NMR diagnosis photography devices, Electronic microscope, Life-sustaining devices with micro-processor control (such as infusion pumps, hemodialysis machines, breath­ing assistor, anesthesia machines, infant incubators, etc.)
t t t
t t t
The test records in the meter will not be removed after transmission. Please refer to the section “Check Strip” for removing records.
3.
The GoodLife Blood Glucose Meter automatically store a maximum of 999 test results including the result done before a meal or after a meal, or for a control solution test. When recalling the results, each single result appears from the latest to the earliest with time and date.
Turn meter on by pressing Mem button till you hear a beep sound. The rst result displayed on the screen is your latest test result. 12 represents the 12
th
recorded of the overall results
on this meter.
1.
Page 19
36 37
Replacing the battery
To Replace the Batteries
1. Make sure the meter is turned o.
Set
Mem
2. Place the meter face down on a at surface.
3. Slide battery compartment door open.
NOTE: Every time when the batteries are replaced, the meter will turn on. Please follow the section “Setting Meter Parameters“ to set the time and date.
The GoodLife Blood Glucose Meter comes with four AAA batteries. Please install your batteries before starting. When the battery symbol appears on the meter display, batteries level is getting low. You will still be able to test with low battery, but you should replace them as soon as possible. When battery symbol appears ashing on the display, the meter will no longer give results and you must replace the batter­ies immediately.
Remove the old batteries and insert 4 new AAA batteries into the battery compartment. Be sure to align the plus (+) and minus (-) signs correctly.
4.
Close the battery compartment door. Check to see that your meter is working. If it fails to turn on, the batteries may have been inserted incorrectly. Remove the batteries and reinsert them correctly.
5.
3L-02-810035 V1.0
Blood Glucose Monitoring System
Cloudia
For Self-testing
Battery:AAA,6V serial no.: 65534-XXXXXXX
For In Vitro Diagnostic Use Only
FCC ID: AYY0000001
Distrubutor: Kannact Inc. / KRemedy 10260 SW. Greenburg Road Suite 525, Portland OR 97223
For Self-testing Battery:AAA,6V
HMD BioMedical Inc. Hsinchu , Taiwan
3L-02-XXXXXX V1.0
IG1-0000001
8901204043780370392
Blood Glucose
Monitoring System
For Self-testing Battery:AAA,6V
HMD BioMedical Inc. Hsinchu , Taiwan
3L-02-XXXXXX V1.0
IG1-0000001
8901204043780370392
Blood Glucose Monitoring System
Page 20
38 39
How to Clean and Disinfect
1. www.cloroxprofessional.com
2. www.amazon.com
3. or call the Clorox service number at 1-800­234-7700
Dierences between Cleaning and Disinfection
t Clean: to remove visible signs of dirt or
pollution.
t Disinfect: to free from infection especially by
destroying harmful microorganisms.
NOTE:
If the meter is operated by the care giver rather
than the primary user, the meter should be disinfected by the meter operator rather than by the user.
To mitigate the risk of bloodborne pathogen trans­mission (eg. viral hepatitis), the meter must be prop­erly cleaned and disinfected. The life of the meter is 5 years for 18,250 cleaning and disinfection cycles. The meter, for single patient home use, should be cleaned and disinfected at least once per week as recommended.
The following disinfactant product with sodium hypochlorite 0.55% as the active ingredient have been shown to be safe for use with GoodLife meter, but any disinfectant product with the EPA registra­tion number of 67619-12 may be used on this device. It should be cleaned and disinfected once per week. The Clorox Germicidal Wipes can be purchased on
Page 21
40 41
Cleaning Procedures
1. Place device and wipes on a smooth surface. Be sure there is enough light.
Set
Mem
2. Wash hand with soap and water then pat dry.
3. Take a piece of premoistened wipe out of canister. Remember to recap the canister.
4. Wipe entire device until visibly clean.
Set
Mem
5. Throw away the used wipe. Please do not reuse the wipe.
CAUTION
1. The meter should always be cleaned before it is disinfected.
2. All parts of the meter should be considered to be biohazardous. Make sure that you clean and disinfect the entire meter.
Set
Mem
Page 22
42 43
Disinfection Steps
1. Prepare wipes and meter.
Set
Mem
2. Take a piece of moistened wipe out of canister. Remember to recap the canister.
3. Slowly wipe the meter back and
forth for at least 3 times. Repeat the same wiping procedure to all meter outer surfaces.
4. Make sure the meter stays wet for 1 min. Please do not get disinfectant liquid into the test strip slot.
5. Throw away the used wipes. Please do not reuse the wipe.
Warning
1. Please use disinfectant with caution.
2. For your own safety, please do not mix disinfectant with other liquid.
3. Please read the instructions carefully on the canister before starting cleaning and disinfection.
4. Please don’t use the wipes on any human part.
5. Store disinfectants out of the reach of children.
6. If any signs of deterioration are noted as below:
Cloudy Display, material cracking, brittle, softening or meter buttons malfunction,
Set
Mem
Page 23
44 45
performance of meter has changed (i.e. control fails). Stop using the meter and please contact customer support at 1-855­692-3511.
NOTE
1. The Clorox Germicidal Wipes can be purchased on www. cloroxprofessional.com, www.amazon.com or call the Clorox service number at 1-800-234-
7700.
2. If you have any question about the cleaning and disinfection or need more information, please contact your distributor or customer support at 1-855-692-3511.
References
1. “FDA Public Health Notication: Use of Fingerstick Devices on More than one Person Poses Risk for Transmitting Bloodborne Pathogens: Initial Communication” (2010)
http://www.fda.gov/MedicalDevice/ Safety/AlertsandNotices/ucm224025.htm
2. “CDC Clinical Reminder: Use of Fingerstick Devices on More than One Person Poses Risk for Transmitting Bloodborne Pathogens” (2010)
http://www.cdc.gov/injectionsafety/ Fingerstick-DeviceBGM.html
Page 24
46 47
Storage and Handling
CAUTION
Display Messages and
Problem-Solving Guide
Display check
Normal:
Abnormal:
(for example)
Keep your meter free of dust or water. Protect it from extreme temperature and humidity. Store the meter and test strips between 50ºF (10ºC) and 104ºF (40ºC)
1.
2.
If the meter will not be used for over 1 week, it is recommended to remove the battery from the meter. This will prevent the potential meter damage due to leaking battery. Please follow all the instructions on the User Guide while operating GoodLife Blood Glucose Meter. HMD BioMedical Inc. will not be responsible for any impairment occurred from NOT following the instructions.
Every time when meter is turned on, the meter will display as illustrated.
If some parts of the display are not working, contact customer support at 1-855-692-3511 for help. When any of the following messages appears, there is a problem with your GoodLife Blood Glucose Meter or perform wrong steps during a test. There messages will help you to identify certain problems. If error messages appear that are not listed on the
Page 25
48 49
Display Description Action to Take
Blood glu ­cose result may be higher than 600 mg/dL.
Review proper testing procedure and perform a quality check with control solution. Re ­peat blood test, if “HI” still appears, call your doctor immediately.
Blood glu ­cose result may be lower than 20 mg/dL.
Review proper testing procedure and perform a quality check with control solution. Repeat test, if “LO” still ap ­pears, call your doctor immediately.
Tempera ­ture is above or below the operating range of test strips.
The result you have obtained may not be accurate. Move to an area with temperature between 50ºF - 104ºF (10ºC to 40ºC ). Do not
articially heat or cool
the meter.
Battery is low.
Change battery imme ­diately.
Test strip unknown or may be damaged.
Perform the test with new test strip.
Test strip is used or
ton saw tset performed correctly.
Perform the test with a new test strip and follow the test procedure correctly.
following messages, contact customer support at 1-855-692-3511 for help.
Page 26
50 51
No re­sponses when strip is inserted into the meter.
Maybe:
1. Battery is dead.
2. Wrong strip was inserted.
3. Meter is defective.
You have to:
1. Replace battery.
2. Insert the test strip correctly.
3. Contact customer support at 1-855-692­3511 for help.
No re­sponses when blood sample is applied to the strip.
Maybe:
1. Blood sample is not suf cient.
-
2. Meter is defective.
You have to:
1. Repeat test with suf cient sample.
-
2. Perform Meter Check by inserting check strip.
Error message: Err1. Data communi
­cation module is malfunction
Err2. Failure to con
­nect to Base Station
Err5. Failure to con-
nect to the Sever
Err6. TCP/IP data
transmission failure
Err1. The meter might
be damaged. Contact custom
­er support at 1-855-692-3511 for help.
Err2. Antenna or SIM
Card might be damaged. Con­tact customer support at 1-855­692-3511 for help.
Err5, 6. Move to an
area of better reception and try again.
Page 27
52 53
1. Assay Method: Electrochemical biosensor
2.
Test Sample: Fresh Capillary Whole Blood
3.
Sample Size: 0.5 μL
4.
Measuring Time: 5 seconds
5.
Measuring Range: 20 – 600 mg/dL
6.
Acceptable Hematocrit Range: 30%-55%
7.
Operating Temp. Range: 50°F-104°F(10°C­40°C)
8.
Operating Relative Humidity: 20% - 80% RH
9.
Memory Capacity: 999 test results with time and date
10.
Result Setting: Before meal, After meal and Control solution test setting
11.
Power Supply: AAA Alkaline Batteries x 4
12. Battery Life: Approximately 1000 tests
13.
Automatic shut-o: In 3 minutes
14.
Meter Dimension: 108.9mm (L) x 59.9mm (W) x 16.5mm (H)
15.
Weight: Approximate 100g (with batteries)
Page 28
54 55
Precision
Within-run
Level No. of test Mean (mg/dL) Within-Run C.V. (%)
40.6
93.3
120
213
336
532
100
100
100
100
100
100
41.5
99.5
124.0
216.4
340.7
531.7
4.5
4.2
3.1
3.8
3.5
3.3
Between-run
Level No. of test Mean (mg/dL) Within-Run C.V. (%)
50
100
300
100
100
100
61.5
123.9
346.8
4.7
4.2
3.9
Performance Evaluations
Accuracy
Glucose concentration < 75 mg/dL
Within ± 5 mg/dL Within ± 10 mg/dL Within ± 15 mg/dL
6/13 (46%) 11/13 (85%) 13/13 (100%)
Glucose concentration ≥ 75 mg/dL
Within ± 5 % Within ± 10 % Within ± 15 % Within ± 20 %
35/87 (40%) 67/87 (77%) 78/87 (90%) 85/87 (98%)
Tests were conducted by trained technicians in the laboratory setting. The venous whole blood from one subject was adjusted to 6 dierent levels. Strips out of a single lot were tested. The results are shown in the following table.
Tests were performed at hospital by healthcare professionals and diabetic patients. Fresh capillary nger whole blood samples were tested with the GoodLife System; venous blood samples from the subjects were tested with YSI Model 2300 Glucose Analyzer as reference. The results are shown in the following table.
Page 29
56 57
Hardware and Software requirement
iglucose Diabets Management
Introduction of the system
How to use
The system is an optional software for use with the iglucose system with data management capacities. t is designed to connect to glucose meters (Cloudia) and automatically transmit blood glucose readings or control solution test results to a secure database.
Users can then utilize the iglucose Diabets manage­ment portal (web-based application) to view their blood glucose readings at home through their PC or other Internet capable devices.
The system provides an integrated table including test results of blood glucose measurements and quality controls.
The system does not recommend any medical treat­ment or dosing recommendations.
Hardware requirement: PC, NB, iPad or Cellular-Internet phone
Software requirement: Microsoft IE 8.0 , Mozilla FireFox 3.6, Google Chrome v-12, Apple Safari 5
Minimum requirements: Pentium processor, 64mb ram, 1 gb disc space
1.
2.
3.
Start the PC or other compatible-Internet connected device. Connect to the chosen Web Browser to the Internet.(e.g. Microsoft IE, Google Chrome or Apple Safari)
Enter the Website address www.iglucose.net, the user could view the record table of the blood glucose or control solution measurement.The user could view the test result according to his/her Meter ID.
1.
2.
Page 30
58 59
How to view transmitted reading
Open a web browser window on your computer and go to www.iglucose.net
1.
Click on the “login” button from the iglucose homepage.
2.
Enter the meter serial number in "Email or Serial" and "Password" and click the "login" button
3.
If you have no any account, please you apply a new account.
Page 31
60 61
The "welcome" page will appear. Click on the "View Readings" button.
4.
"Your Time zone" page will appear, choose your time zone and click on "Update Time zone" button.
5.
Page 32
62 63
Select any of the following tabular or graphic report formats to view your reading: Trend Report, Pie Chart, logbooks or Summary.
6.
Trend Report:Displays high and low glucose read­ings within a customized date and time range.
Page 33
64 65
Pie Chart: Displays percentage of pre-meal and post- meal glucose readings that fall in and out of target ranges.
logbooks: Displays daily and historical glucose readings with date and time stamp. Glucose read­ings are also displayed before and after meals with pre-determined out-of-target ranges.
Page 34
66 67
FCC Regulations
15.19(a)(3): This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired opera­tion.
15.105(b): NOTE:This equipment has been tested and found to comply with the limits for a Class B digital device, pursu­ant to part 15 of the FCC Rules. These limits are designed to pro vide reasonable protection against harmful inter­ference in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accor­dance with the instructions, may cause harmful interfer­ence to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful inter­ference to radio or television reception, which can be determined by turning the equipment o and on, the user is encouraged to try to correct the interference by
one or more of the following measures:
t3FPSJFOUPSSFMPDBUFUIFSFDFJWJOHBOUFOOB t*ODSFBTFUIFTFQBSBUJPOCFUXFFOUIFFRVJQNFOUBOE
receiver.
t$POOFDUUIFFRVJQNFOUJOUPBOPVUMFUPOBDJSDVJU
dierent from that to which the receiver is connected.
t$POTVMUUIFEFBMFSPSBOFYQFSJFODFESBEJP57UFDIOJ
cian for help. RF Exposure Information (SAR) This device meets the government’s requirements for exposure to radio waves. This device is designed and manufactured not to exceed the emission limits for exposure to radio frequency (RF) energy set by the Federal Communications Commission of the U.S. Government. The exposure standard for wireless mobile devices employs a unit of measurement known as the Specic Absorption Rate, or SAR. The SAR limit set by the FCC is
1.6W/kg. *Tests for SAR are conducted using standard operating positions accepted by the FCC with the device transmit­ting at its highest certied power level in all tested frequency bands. Although the SAR is determined at the
Page 35
68 69
highest certied power level, the actual SAR level of the device while operating can be well below the maximum value. This is because the device is designed to operate at multiple power levels so as to use only the poser required to reach the er levels so as to use only the poser required to reach thestation antenna, the lower the power output. The highest SAR value for the device as reported to the FCC when tested for use when worn on the body, as described in this user guide, is 1.26 W/kg(Body-worn SAR Value) and 3.26 W/kg(Extremity-worn SAR Value). (Body­worn measurements dier among device models, depending upon available accessories and FCC require­ments.) While there may be dierences between the SAR levels of various devices and at various positions, they all meet the government requirement. The FCC has granted an Equipment Authorization for this device with all reported SAR levels evaluated as in compliance with the FCCRF exposure guidelines. SAR information on this device is on le with the FCC and can be found under the Display Grant section of
www.fcc.gov/oet/ea/fccid/
after searching on FCC ID:
AYY0000002.
This device is compliance with SAR for general popula­tion /uncontrolled exposure limits in ANSI/IEEE C95.1­1999 and had been tested in accordance with the measurement methods and procedures specied in applicable FCC KDB guidance. For body worn operation, this device has been tested and meets the FCC RF exposure guidelines for use with an accessory that contains no metal and the positions the hand set a minimum of 1.0cm from the body. Use of other accessories may not ensure compliance with FCC RF exposure guidelines. Changes or modications not expressly approved by the party responsible for compliance could void the user‘s authority to operate the equipment.
Page 36
70
Warranty
POST- SALE SERVICE
Manufacturer: HMD BioMedical Inc. Hsinchu,Taiwan Questions? Please call toll free number at 1-855-692-3511 Mon-Fri 9:00am-4:30pm(Pacic Standard Time) Please contact a healthcare provider for assistance at all other times. E-mail: [email protected]
HMD BioMedical Inc. warrants to the original purchaser that this instrument will be free of defects in material and workmanship for 5 years from the date of original purchase. During the stated 5-year period, HMD shall, at no charge, replace a unit found to be defective with an equivalent or current version of the owner’s model. This warranty is limited to replacement due to defects in parts or workmanship. HMD shall not be required to replace any units which malfunction or are damaged due to abuse, accidents, alteration, misuse, neglect, maintenance by other than HMD, or failure to operate the instrument according to instructions.
3K-02-XXXXXX V1.3
Loading...