Gnatus Syncrus G8 HE User manual

PRESENTATION OF MANUAL
INSTRUCTIONS FOR USE
EQUIPMENT:
Technical Name: Dental Delivery Units and Accessories Trade Name: Syncrus G8 HE Delivery Unit Brand: GNATUS
Manufacturer/ Distribuitor:
GNATUS - EQUIPAMENTOS MÉDICO-ODONTOLÓGICOS LTDA. Rod. Abrão Assed , Km 53+450m - Cx. Postal 782 CEP 14097-500 Ribeirão Preto - S.P. - Brasil Fone +55 (16) 2102-5000 - Fax +55 (16) 2102-5001 C.N.P.J. 48.015.119/0001-64 - Insc. Est. 582.329.957.115 www.gnatus.com.br - gnatus@gnatus.com.br Technical Duties: Gilberto Henrique Canesin Nomelini CREA-SP: 0600891412
Registration ANVISA #: 10229030047
ATTENTION
For greater safety:
Read and understand all the instructions contained in these
Instructions for Use before installing or operating this Equipment.
Note: These Instructions for Use must be read by all the operators
of this Equipment.
4
INDEX
PRESENTATION OF MANUAL ........................................................................02
IDENTIFICATION OF EQUIPMENT .................................................................05
Indication of Equipment ...................................................................................05
Principles and bases applied to the functioning of the product ................................06
Description of Equipment ..................................................................................06
MODULES, ACCESSORIES, OPTIONS AND MATERIALS OF CONSUMPTION .....08
TECHNICAL SPECIFICATIONS .......................................................................11
- Technical features of the Delivery Unit and its accessories ...................................11
- Electromagnetic emissions ..............................................................................13
- Dimension ....................................................................................................17
- Symbologies of packaging .............................................................................18
- Symbologies of product .................................................................................18
- Standards applied .........................................................................................20
- Content of acessible and non-accessible demarcations .......................................20
INSTALLATION OF EQUIPMENT ....................................................................20
OPERATION OF EQUIPMENT .........................................................................20
- Turning on / off the dental set .........................................................................20
- Positioning ...................................................................................................21
- Activation of Terminals ..................................................................................21
- Adjustment of Spray “TB/TM” ........................................................................21
- Adjustment of Spray “FO/MME” .......................................................................21
- Retractable rods with lock...............................................................................22
- Use of 3-Way Syringe ...................................................................................22
- Curing Light Activation ..................................................................................23
- Ultrasound Activation .....................................................................................24
- Bicarbonate Jet Activation"Jet Hand" ................................................................25
- Equipment activation by the delivery unit panel .................................................27
-
Use and applications (Digital control panel kit) ..................................................28
- How to supply the reservoirs ...........................................................................29
PRECAUTIONS, RESTRICTIONS AND WARNINGS ..........................................30
- Conditions of transport, storage and operation ..................................................30
- Sensitiveness to environmental conditions foreseeable in normal situations of use .30
- Precautions and warnings “during the installation” of equipment ..........................30
- Recommendations for preserving the equipment. ...............................................31
- Precautions and warnings “during the use” of equipment ....................................31
- Precautions and warnings “after” the use of equipment ......................................31
- Precautions and warnings during the “cleaning and disinfection” of equipment .......32
- Precautions in case of alteration in the functioning of equipment ..........................32
- Precautions to be adopted against foreseeable or uncommon risks,
related to the deactivation and abandoning of equipment ......................................32
CORRECTIVE AND PREVENTIVE MAINTENANCE AND PRESERVATION............33
- Additional procedures for reuse .......................................................................33
- Cleaning ......................................................................................................33
- Disinfection ..................................................................................................33
- Optilight LD ..................................................................................................34
- Ultrasound ...................................................................................................34
55
INDEX
- Bicarbonate Jet .............................................................................................35
- Reservoirs ....................................................................................................35
- Bio-System ..................................................................................................35
- 3-Way Syringe ..............................................................................................35
- Preventive Maintenance .................................................................................36
- Corrective Maintenance ..................................................................................36
UNFORESEEN EVENTS – SOLUTION OF PROBLEMS ........................................36
WARRANTY OF EQUIPMENT ..........................................................................38
FINAL CONSIDERATIONS ..............................................................................38
6
DESCRIPTION OF THE EQUIPMENT
Dear Customer
Congratulations. You have made a good choice when you decided to buy a GNATUS QUALITY product comparable to the best products available in the World. This manual is a general presentation of your product and it will give you important details to help you to solve possible problems.
Please, read it and keep this with you.
Identi cation
Technical Name: Dental Delivery Units and Accessories
Trade Name: Syncrus G8 HE Delivery Unit
Brand: GNATUS
Indication of Equipment
This equipment is for dental use use only. It must be operated and utilized by specialized professional (certifi ed professional, according to the legislation of the country) and following the instructions of the manual. The operation of the equipment required, for the professional, the utilization of correct instruments and it should to be in perfect conditions of the use, and to protect the professional, the patients and others, in the eventual danger situation.
77
DESCRIPTION OF THE EQUIPMENT
Principles and bases applied to the functioning of the product
It has hoses with compressed air and connectors for the supply of handpieces (high and
low rotation) and a syringe with air and water outlet.
Description of Equipment
Equipment for dental use, fl ex-type model, coupled to the chair.
It is equipped with a side panel and central control.
Flexible, removable and autoclavable handpiece support, protecting them from impact.
Movement of the handpieces by means of retractable rods with lock for alleviating the stress in the hose, which makes the movements light, allowing greater closeness in the operational fi eld.
Swivel arms with ample movement both horizontally and vertically, with pneumatic locking, activated by a button located under the body of the delivery unit, providing smoothness of movement and stopping at the position desired. Easy access bilateral frontal catcher with differentiated design.
Horizontally moving auxiliary tray bracket, designed to support the widest range of off-the-shelf trays and brackets (Support nonremovable - items not supplied by Gnatus).
Body of the delivery unit made of high impact polyethylene.
Structure made of steel with surface treatment via nanotechnology. Smooth painting with a high gloss and epoxy base, polymerized in an oven at 250ºC, resistant to corrosion and cleaning materials.
Smooth hoses, without grooves or rifl ing, rounded, light and exible.
To guarantee the safe functioning of your equipment, use only the assemble confi gurations (Dental Chair, Dental and Water Units and Dental Light) supplied by Gnatus authorized Dealers / Technical Assistance.
EN ISO 9001/2000 and EN ISO 13485/2003 Quality System, assuryng the products are manufactured under standart procedures.
Products manufactured in agreement with RDC 59/00 - ANVISA - (Sanitary Surveillance National Agency).
Curing light (optional item) – Features of the product:
The Curing Light belongs to the newest generation of LED photo-activation devices. This abbreviation stands for Light Emitting Diode, a totally different type of light emission, if compared to conventional halogen equipment.
Unlike traditional devices, which generate wide-spectrum light and heat, this technology uses a cold light of the precise wave length needed to activate various dental products.
LED technology, which was recently introduced in Dentistry, brought about several useful features to those light-curing devices used in composite resin restoration. Besides being more durable, LED technology turned devices more compact, ergonomic and easier to install and transport. The emission of cold light within a precise wave length range ensures the safe cure of camphorquinone-activated composites, preventing dental heating, pulp damage or discomfort for both patient and dentist. Although being relatively new, this technology is nowadays in its second generation. LED safety and effi ciency, now allied to high-energy emission, are available to all clinic procedures which require light-curing power, including bleaching treatments.
The light of 440nm-460nm wave length, allied to the high energy emitted by Curing Light, makes possible the multi-functionality of this device:
8
DESCRIPTION OF THE EQUIPMENT
- Direct restoration procedures: composite resins, ionomers and adhesives.
- Indirect restorations: adhesive cementation of laminates, inlays, esthetic pins and
metal-free crowns.
- Dental Bleaching: activation of bleaching gel and polymerization of gingival barriers.
Compatible with 35% hydrogen peroxide-based bleaching gels.
- Attachment of braces and orthodontic accessories.
- Activation of light-cure materials, such as sealants, surgical cements and covering
bases.
Designed and built with cutting-edge technology, it meets the highest standards specifi ed
by world’s dental authorities.
Operation control display in handpiece, sound alarm with beep every 10 seconds and
4 beeps at the end of the cycle.
Advantages offered by Curing Light:
- More spectrally-selective light than conventional lamps.*
- Cold light, it doesn’t heat up the resin nor the tooth**
• Light compact equipment that provides handling comfort.
• Low power consumption.
• Longer useful life of the light emitting diode (equivalent to 36.000.000 cycles of 10
seconds).
• It does not use optical fi lter.
• It does not require forced ventilation, thus avoiding noise emission.
We noted that the light emitted by the Curing Light is completely contained within the absorption interval of the photo starter, therefore it’s 100% used, whereas the conventional equipment running on halogen lamps has non-used wave-length regions.
The Curing Light doesn’t generate heat since it uses light emitting diodes.
The light conductor is removable, made out of high resistance polymer and of easy maintenance.
Digital Control Panel Set "electric micromotor Bien Air” (optional item) Product Features:
Refer to Owner's Manual - Digital Control Panel
Bicarbonate Jet Set – Jet Hand (optional item) - Product Features:
Refer to Owner's Manual of Jet Hand (available for viewing and downloading via www.
gnatus.com.br/manuais)
99
MODULES, ACCESSORIES, OPTIONS AND MATERIALS OF CONSUMPTION
The contents of this page are of an informative nature, the equipment being able to differ from that illustrated. So, upon acquiring the product check the technical compatibilty between equipment, coupling and accessories.
A - Power - power ultrasound adjustment
*
B - Speed - electric microengine power adjustment
*
(available on equipments set up with electric micromotor without digital control panel set). C - Light - electric microengine brightness adjustment
15
*
14
08
A B C
01
07
02
06
03
04
05
09
Basic conguration of the product (FIT Version)
1-3-Way Syringe, 2- TB/TM/FO high rotation terminals, 1-Terminal for TB/TM/FO/MME micro motor
01 - Triple syringe 02 - High-speed-motor terminals
*
03 - Micro motor terminal
*
04 - Ultrasound
*
05 - Curing Light
*
06 - Control panel 07 - Bilateral Catcher 08 - X ray view
10
12
09 - Auxiliary trays 10 - Manometer
*
11 - Digital control panel Kit (MME)
*
12 - Arm brake valve 13 - Water valve for Syringe /FO/MME/ Ultrasound 14 - Articulated arm 15 - Column arm
11
13
optional
*
:
MODULES, ACCESSORIES, OPTIONS AND MATERIALS OF CONSUMPTION
01
03
05
07
02
04
06
The drawing illustrates all optional items; therefore, your equipment will consist only of items selected during your purchase option.
;;
MODULES, ACCESSORIES, OPTIONS AND MATERIALS OF CONSUMPTION
01 - Terminals
- MME:Electrical micro motor terminal
- FO:Optical fi ber terminal
- TM:Midwest terminal
- TB:Borden terminal
02 - Curing light + tip for 3 teeth
03 - Water Heater Kit for triple Syringe
04 - Auxiliary tray / Support tools
(not supplied by Gnatus)
05 - Bicarbonate jet kit “Jet Hand”
- Bicarbonate jet
- Opener
- Covers for reservoir
- Rings for sealing
- Sachet of bicarbonate
06 - Kit Bien Air
- Digital control panel
- High torque micromotor
- Contra angle 1:1
- Contra angle 1:5
- Manual
07 - Ultrasound kit
- Ultrasound
- Tightening wrench
- Inserts “nº 1, 2, 10P and A120”
The use of any part, accessory or material not specied or foreseen in these instructions for use is entirely the user’s responsibility.
The accessories described above shall never be able to be sold separately from the product.
32
TECHNICAL SPECIFICATIONS
Technical features of the Delivery Unit and its accessories General
Model Syncrus G8 HE Delivery Unit Classication of Equipment as per ANVISA: Class II Classication of Equipment as per standard IEC 60601-1: Protection against Electric Shock - Type B and Class I Equipment (IEC 60601-1) Degree of safety of application in presence: Equipment not suited to an anesthetic mixture inammable with air, oxygen or nitrous
oxide. Mode of Operation Continuous operation with intermittent load
Power Supply
Power Supply Voltage (coming from dental chair) 127/220 V~ (Selectable) Frequency 50/60 Hz Input fuse (coming from dental chair) 8A Delayed action Voltage in equipment (coming from dental chair) 12 and 24 V~
Other speci cations
Inlet air pressure 80 PSI (5,52 BAR) Inlet air pressure - Syringe 40 PSI (2,76 BAR) Maximum consumption of air (dental set) 80 l/min Capacity of water reservoir (coming from water unit) 1000ml
3333
TECHNICAL SPECIFICATIONS
High rotation air consumption 9 l/min High rotation water consumption 0,02 l/min Syringe air consumption 17 l/min Syringe water consumption 0,1 l/min Maximum capacity of load applied to trays 1Kg Net weight (complete version) 26 Kg Gross weight (complete version) 31 Kg
Speci cations of Curring Light
Power 5,2VA Light source 1 LED Active medium Semicondutor Led (InGaN) Wavelength 440nm - 460nm Timer 90 seconds Timer alarm sound alarm with beep every 10 seconds and 4 beeps at the end of the cycle Activation Through the hand-piece button Light conductor Made out of special polymer, rotational, removable and reuse sable. Hand-piece body ABS injected