Gnatus Syncrus G4 User manual

PRESENTATION OF MANUAL
INSTRUCTIONS FOR USE
Technical Name: Odontological cuspidor
Brand: Gnatus
Trade Name: Water Unit Syncrus G4
Models: G4 / G4 A
Manufacturer / Distribuitor:
GNATUS - EQUIPAMENTOS MÉDICO ODONTOLÓGICOS S/A. Rod. Abrão Assed , Km 53+450m - Cx. Postal 782 CEP 14097-500 Ribeirão Preto - S.P. - Brasil Fone +55 (16) 2102-5000 - Fax +55 (16) 2102-5001 C.N.P.J. 48.015.119/0001-64 - Insc. Est. 582.329.957.115 www.gnatus.com.br - gnatus@gnatus.com.br Technical Duties: Ricardo J. Ravaneli CREA-SP: 5060714523
Registration ANVISA nº: 10229030067
ATTENTION
For greater safety:
Read and understand all the instructions contained in these
instructions for use before installing or operating this equipment.
Note: These instructions for use must be read by all the operators
of this equipment.
2
INDEX
PRESENTATION OF MANUAL ........................................................................02
IDENTIFICATION OF EQUIPMENT .................................................................05
Indication of Equipment ...................................................................................05
Principles and fundamentals applied to the product functioning ..............................05
Equipment Description .....................................................................................06
MODULES, ACCESSORIES, OPTIONALS AND CONSUMPTION MATERIALS .......08
TECHNICAL SPECIFICATIONS .......................................................................14
- Technical features of the Delivery Unit and its accessories ...................................14
- Electromagnetic Emissions ..............................................................................16
- List of pieces and circuit scheme .....................................................................19
- Standards applied .........................................................................................19
- Dimensions ..................................................................................................20
- Packing symbols ...........................................................................................22
- Product symbols ...........................................................................................22
- Content of accessible and non-accessible demarcations .....................................23
INSTALLATION OF EQUIPMENT ....................................................................23
EQUIPMENT OPERATION ..............................................................................24
- Control panel ................................................................................................24
- Terminal Triggering .......................................................................................22
- Adjustment of Spray of “TB/TM high and low rotation terminals” .........................25
- Ejectors operation .........................................................................................25
- Use of 3-Way Syringe ...................................................................................25
- Replacement of the standard of cannula coupling ..............................................26
- Curing Light Activation ...................................................................................27
- Activation of the “Laser Hand” .........................................................................28
- Water ow adjustment ...................................................................................29
- Regulation of the water selecting valve .............................................................29
- Master Valve .................................................................................................29
- How to provision the reservoirs .......................................................................29
PRECAUTIONS, RESTRICTIONS AND WARNINGS ..........................................30
- Transportation, storage and operation ..............................................................30
- Sensitivity to environmental conditions in normal situations of use .......................30
- Precautions and warnings “during the installation” of equipment ..........................30
- Recommendations for the dental equipment maintenance. ..................................31
- Precautions and warnings “during the use” of equipment ....................................31
- Precautions and warnings “after” the use of equipment ......................................31
- Precautions and warnings during the “cleaning and disinfection” of equipment .......32
- Precautions in case of alteration in the functioning of equipment ..........................32
- Precautions to be adopted against foreseeable or uncommon risks,
related to the deactivation and abandoning of equipment ......................................32
CORRECTIVE AND PREVENTIVE MAINTENANCE AND PRESERVATION............33
- Additional procedures for reuse .......................................................................33
- Cleaning and Disinfection ...............................................................................33
Cleaning:
- Triple syringe ................................................................................................35
- Reservoirs ....................................................................................................35
33
INDEX
- Cleaning of the drain .....................................................................................35
- Basin cleaning ..............................................................................................36
- Bio-System ..................................................................................................36
- Curring Light ................................................................................................37
- Laser Hand ...................................................................................................37
- Bicarbonate Jet "Jet Hand" .............................................................................37
- Preventive Maintenance .................................................................................37
- Corrective Maintenance ..................................................................................37
UNFORESEEN EVENTS – SOLUTION OF PROBLEMS ........................................37
WARRANTY OF EQUIPMENT ..........................................................................40
FINAL CONSIDERATIONS ..............................................................................40
4
IDENTIFICATION OF EQUIPMENT
Dear Customer
Congratulations. You have made a good choice when you decided to buy a GNATUS QUALITY product comparable to the best products available in the World. This manual is a general presentation of your product and it will give you important details to help you to solve possible problems.
Please, read it and keep this with you.
Indication of Equipment
This equipment is for dental use use only. It must be operated and utilized by specialized
professional (certied professional, according to the legislation of the country) and following
the instructions of the manual. The operation of the equipment required, for the professional, the utilization of correct instruments and it should to be in perfect conditions of the use, and to protect the professional, the patients and others, in the eventual danger situation.
Principles and fundamentals applied to the product functioning
Auxiliary waste collector unit, has suctors which suction is caused by venturi system or vacuum pump with compressed air.
Identication
Technical Name: Odontological cuspidor Trade Name: Water Unit Syncrus G4 Brand: GNATUS Models: G4/G4A
G4 G4 A
55
IDENTIFICATION OF EQUIPMENT
Equipment Description
Water unit for odontological use, for auxiliary work such as water supply for waste collection and spitting and sucking activation; ambidextrous (for right- and left-handed).
Attached to the chair, with electronic control panel which activates the following functions: water in the bowl, water in the cup holder and *water heating of the triple syringe.
Programmable timer for water trigging in the cup holder and in the spitting bowl by a
time interval set by the professional, providing bigger water savings at the ofce.
The frame is manufactured with steel structure, ABS injected body with anti-UV protection. Smooth high glossy paint, epoxy-based, cured in an oven at 250° C, with phosphate treatment resistant to rust, corrosion and cleaning chemicals.
Upper part of the unit with suitable location for the best spitting position, 180° foldable bowl providing total patient comfort.
Ceramic bowl spittoon, deep and easily removable for hygiene and asepsis, supplied with strainer drain for solids retention.
Water ux regulating system allowing ne-tuning in the bowl and the cup holder water ux.
Smooth, rounded, light and exible hoses and with quick release that is easily connected,
without the need of tools.
Has a debris lter easy to clean and disinfection.
Ejectors with automatic drive easy to use, they provide an excellent operating
performance, allow professionals to work with better visualization of the operative eld and
reduce the risk of contamination by aerosol and greater patient comfort.
*High power electric Ejectors with individual low voltage drive, provide lightness and accuracy in the drive.
*Triple syringe swivel spout, removable and autoclavable.
*Double system for water supply (network/reservoir).
*Master valve (key for water cut).
Arm Reach with Front Handle: terminal support with wide horizontal movement that
enables optimal approach to the surgical eld and excellent accessibility to the various
resources available. Optimizes work prioritizing the ergonomics and biosafety.
Automatic selection of tips through individual pneumatic valves, allowing light handling.
Translucent water tanks for syringe and spray tips and chlorinated water Bio- System.
*
Bio-System: disinfection system, which provides the internal hose and terminals cleaning
through liquid bactericide, preventing risk of cross contamination.
EN ISO 9001/2000 and EN ISO 13485/2003 Quality System, assuryng the products
are manufactured under standart procedures.
Products manufactured in agreement with RDC 59/00 - ANVISA - (Sanitary Surveillance
National Agency).
Laser Hand Kit (optional item) – Features of the product:
See the Owner’s Manual - Laser Hand
Bicarbonate jet (optional item) – Features of the product:
See the Owner’s Manual - Jet Hand
* Optional item
6
IDENTIFICATION OF EQUIPMENT
Curing light (optional item) – Features of the product:
The Curing Light belongs to the newest generation of LED photo-activation devices. This abbreviation stands for Light Emitting Diode, a totally different type of light emission, if compared to conventional halogen equipment.
Unlike traditional devices, which generate wide-spectrum light and heat, this technology uses a cold light of the precise wave length needed to activate various dental products.
LED technology, which was recently introduced in Dentistry, brought about several useful features to those light-curing devices used in composite resin restoration. Besides being more durable, LED technology turned devices more compact, ergonomic and easier to install and transport. The emission of cold light within a precise wave length range ensures the safe cure of camphorquinone-activated composites, preventing dental heating, pulp damage or discomfort for both patient and dentist. Although being relatively new, this technology
is nowadays in its second generation. LED safety and efciency, now allied to high-energy
emission, are available to all clinic procedures which require light-curing power, including bleaching treatments.
The light of 440nm-460nm wave length, allied to the high energy emitted by Curing Light, makes possible the multi-functionality of this device:
- Direct restoration procedures: composite resins, ionomers and adhesives.
- Indirect restorations: adhesive cementation of laminates, inlays, esthetic pins and
metal-free crowns.
- Dental Bleaching: activation of bleaching gel and polymerization of gingival barriers.
Compatible with 35% hydrogen peroxide-based bleaching gels.
- Attachment of braces and orthodontic accessories.
- Activation of light-cure materials, such as sealants, surgical cements and covering
bases.
Designed and built with cutting-edge technology, it meets the highest standards specied
by world’s dental authorities.
Operation control display in handpiece, sound alarm with beep every 10 seconds and 4 beeps at the end of the cycle.
Advantages offered by Curing Light:
- More spectrally-selective light than conventional lamps.
- Cold light, it doesn’t heat up the resin nor the tooth.
• Light compact equipment that provides handling comfort.
• Low power consumption.
• Longer useful life of the light emitting diode (equivalent to 36.000.000 cycles of 10
seconds).
• It does not use optical lter.
• It does not require forced ventilation, thus avoiding noise emission.
We noted that the light emitted by the Curing Light is completely contained within the absorption interval of the photo starter, therefore it’s 100% used, whereas the conventional equipment running on halogen lamps has non-used wave-length regions.
The Curing Light doesn’t generate heat since it uses light emitting diodes.
The light conductor is removable, made out of high resistance polymer and of easy maintenance.
* Optional item
77
MODULES, ACCESSORIES, OPTIONALS AND CONSUMP TION MATERIALS G4
The content on this page is informative; the equipment can be different from those illustrated. Therefore, when purchasing the product check the technical compatibility between equipment, coupling and accessories.
15 16
14
13
1211
09
Basic product conguration:
1 Venturi type suctor, 1 Vacuum pump suctor
01 - Water conductor cup holder 02 - Cup holder 03 - Water conductor bowl 04 - Bowl 05 - Drain cover 06 - Control Panel
*
07 - Vacuum pump suctor
*
08 - Venturi type suctor
09 - Suctor Filters
10 - Cabinet body 11 - Bio-System Reservoir 12 - Water Reservoir
13 - Water ux regulating cup holders 14 - Water ux Regulating bowl
*
15 - Master valve - releases/block water input
*
16 - Water selector valve - reservoir/network
*
17 - Optical Sensor
10
17
01
02
03
04
05
06
07
08
* Optional item
8
MODULES, ACCESSORIES, OPTIONALS AND CONSUMP TION MATERIALS G4A
The content on this page is informative; the equipment can be different from those illustrated. Therefore, when purchasing the product check the technical compatibility between equipment, coupling and accessories.
15 16
14
13
1211
09
Basic product conguration:
1 Venturi type suctor, 1 Vacuum pump suctor
10
18
20
01
17
02
19
08
03
04
05
06
07
01 - Water conductor cup holder 02 - Cup holder 03 - Water conductor bowl 04 - Bowl 05 - Drain cover 06 - Control Panel
*
07 - Vacuum pump suctor
*
08 - Venturi type suctor 09 - Suctor Filters 10 - Cabinet body
* Optional item
11 - Bio-System Reservoir 12 - Water Reservoir
13 - Water ux regulating cup holders 14 - Water ux Regulating bowl
*
15 - Master valve - releases/block water input
*
16 - Water selector valve - reservoir/network
17 - Cover reach
*
18 - Arm reach
19 - Central Catcher
*
20 - Optical Sensor
99
MODULES, ACCESSORIES, OPTIONALS AND CONSUMP TION MATERIALS
01
04
10
02
05 06
0807
11
09
12
03
13
14
15
10
MODULES, ACCESSORIES, OPTIONALS AND CONSUMP TION MATERIALS
16
* 01.Terminals:
- Borden terminal (TB)
- Midwest Terminal (TM)
- Fiber Optic Terminal (FO)
- Electric micro motor Terminal (MME)
* 02. Curing Light + 3 teeth tip
* 03. Intra oral Camera set
* 04. Bicarbonate spray set (Hand jet)
* 05. Coupling arm reach for up to 5 tips
* 06. Coupling arm reach for up to 3 tips
* 07. Triple syringe with fully metal body or injected thermoplastic handle
* 08. Triple syringe with fully injected thermoplastic body
* 09. Ejectors with fully metal body or injected thermoplastic:
- Venturi ejector
- High performance Venturi type ejector
- Large ejector for Vacuum Pump
- Small ejector for Vacuum Pump
- Cleaning brush ejector
- Cannula ejector
17
*Aspirators’ hoses available in smooth and corrugated versions.
* 10. Water bowl manufactured with injected material
* 11. Water selector valve - tank / network
* 12. Master Valve (system that allows cutting the ow of water and air to the ofce.)
* Optional item
1111
MODULES, ACCESSORIES, OPTIONALS AND CONSUMP TION MATERIALS
* 13. Optical sensor to actuate the water supply to the bowl
* 14. Integrated paddle "Chip Blower"
* 15. Progressive paddle
* 16. Progressive paddle with water activation/cut
* 17. Heating kit for triple syringe
*Items (5) and (6) available in versions with or without articulation.
WARNING
- The drawing illustrates all optional items (page 10 and 11). Therefore, your equipment will consist only of the chosen items selected during your purchase option.
- The use of any part, accessory or material not specied or provided
in these instructions is of entire responsibility of the user.
* Optional item
12
MODULES, ACCESSORIES, OPTIONALS AND CONSUMP
0499
CONFIGURAÇÃO CONFIGURATIONCONFIGURACIÓN
GNATUS EQUIPAMENTOS MÉDICO-ODONTOLÓGICOS S/A.
Rod. AbrãoAssed, Km 53+450m - Ribeirão Preto - SP - Brasil
EQUIPAMENTO DE CLASSE l
CLASS l EQUIPMENT
EQUIPAMIENTO DE CLASSE l
Contínuo, com carga intermitente Continuous, with intermittent load Continuo, con carga intermitente
FREQUÊNCIA FREQUENCY
FRECUENCIA
12/24 V~
50/60 Hz
NUM. REG. ANVISA
UNIDADE DE ÁGUA SYNCRUS G4A
APARELHO
EQUIPMENT
APARATO
OPERAÇÃO
OPERATION
OPERACIÓN
TENSÃO
TENSIONTENSIÓN
Wellkang Ltd
LONDON, W1G 9QR, U.K.
Suite B, 29 Harley Street
EC REP
TION MATERIALS
Units may be made composed by:
Optional Acronyms
1 BV ejector 1 Venturi ejector
1 Venturi ejector + 1 BV ejector 2 Venturi ejectors 2 BV ejectors 1 Syringe + 1 Venturi ejector 1 Syringe + 1 BV ejector 1 Syringe + 1 Venturi ejector +1 BV ejector 1 Syringe + 2 Venturi ejector 1 Syringe + 2 BV ejector 1 Syringe + 1 Photo +1 Terminal High Speed + 1 Terminal micro motor
+ 1 Venturi ejector 1 Syringe +1 Photo + 1 Terminal High Speed + Terminal micro motor +
1 BV ejector Curing Light Cup holder Intra Oral Camera Arm reach Activation by electronic panel Complete equipment
TBV TV
2T 2 TV 2T BV S TV S TBV 3T 3T V 3T BV 5T
5T BV
OPTI E CAM ALC P FULL
Identication label "responsible eld to identify the product conguration."
1313
TECHNICAL SPECIFICATIONS
Technical features of the Delivery Unit and its accessories
General
Model Water Unit Syncrus G4/G4A
Classication of Equipment as per ANVISA:
Class II
Classication of Equipment as per standard IEC 60601-1:
Protection against Electric Shock - Type B and Class I Equipment (IEC 60601-1) Degree of safety of application in presence:
Equipment not suited to an anesthetic mixture inammable with air, oxygen or nitrous
oxide. Mode of Operation Continuous operation with intermittent load
Power Supply
Inlet air pressure 80 PSI (5,52 BAR) Voltage in equipment (coming from dental chair) 12V~ and 24 V~
Other specications
Capacity of water reservoir 1000ml High rotation air consumption 9 l/min High rotation water consumption 0,02 l/min Inlet air pressure - Syringe 40 PSI (2,76 BAR) Syringe air consumption 17 l/min Syringe water consumption 0,1 l/min Net weight (complete version) 15 Kg
14
TECHNICAL SPECIFICATIONS
Gross weight (complete version) 21 Kg Venturi suction system – Maximum vacuum
220 mm/Hg
Venturi suction system – Volumetric displacement 30 l/min “Bio Vac II Vacuum Pump” suction system – Maximum vacuum
400 mm/Hg
“Bio Vac II Vacuum Pump” suction system – Volumetric displacement
120 l/min
“Bio Vac IV Vacuum Pump” suction system – Maximum vacuum
550 mm/Hg
“Bio Vac IV Vacuum Pump” suction system – Volumetric displacement
350 l/min
Specications of Curring Light
Power 5,2VA Light source 1 LED Active medium Semicondutor Led (InGaN) Wavelength 440nm - 460nm Timer 90 seconds Timer alarm sound alarm with beep every 10 seconds and 4 beeps at the end of the cycle Activation Through the hand-piece button Light conductor Made out of special polymer, rotational, removable and reuse sable. Hand-piece body ABS injected
1515
TECHNICAL SPECIFICATIONS
The materials used to produce the equipment are Biocompatible.
Electromagnetic Emissions
Use of different cables, transducers and accessories from those specied may result
in increased emissions or decreased immunity of the equipment.
16
TECHNICAL SPECIFICATIONS
Electromagnetic Emissions
1717
TECHNICAL SPECIFICATIONS
Electromagnetic Emissions
18
TECHNICAL SPECIFICATIONS
Electromagnetic Emissions
List of pieces and circuit scheme
Gnatus Company declares that the supply of the circuit scheme, list of pieces or any other information that propitiate technical attendance for the user, can be request if there is an agreement between the user and Gnatus Company.
Standards applied:
NBR 60601-1:1997 - Equipamento Eletromédico- Parte 1: Prescrições gerais para segurança; NBR ISO 14971:2004- Medical devices - application of risk management medical devices; NBR ISO 9687: 2005 - Dental equipment - graphical symbols; EN ISO 13485-2003 - Quality systems - medical devices; IEC 60601-1-2:2007 - Compatibilidade Eletromagnética.
1919
TECHNICAL SPECIFICATIONS
Dimensions (mm)
Equipment with coupling "reach arm" G4 A
20
TECHNICAL SPECIFICATIONS
Dimensions (mm)
Equipment without coupling "reach arm " G4
2121
TECHNICAL SPECIFICATIONS
Packing symbols
It determines the maximum quantity of boxes which can be stacked during transportation and storage “as per packaging”.
Packing to be transported and / or stored with the harrows up.
Packing to be transported and / or stored with care (should not suffer drop and neither receive impact).
Product symbols
Careful : It indi cates an important instruction for the operation of the product. Not following it can cause dangerous malfunctioning.
Note: It indi cates useful information for operation of the product.
Packing to be transported and / or stored avoiding humidity,
rains and wet oor.
The packing must be stored and transported away from direct sun light exposure.
Temperature limit for the packing to be stored or transported.
High-speed with FO
Curring Light
Triple syringe
Important: It indicates an instruction of safety for operation of the product. Not following it, can lead to serious danger to the patient.
Landing (in many parts of the equipment) indicates the condition of being landed.
B type equipment
Laser Hand
BV ejector
Ejector type Venturi
Bicarbonate Jet
Water heating activation
22
TECHNICAL SPECIFICATIONS
Product symbols
Bowl’s water ow
Bowl water ux regulation
Key switch water supply
Cup lling
Cup holder water ux regulation
Master valve (key for water cut)
network/reservoir
Warning - Consult the manual
Authorized representative in the European Community
Content of accessible and non-accessible demarcations
GNATUS EQUIPAMENTOS MÉDICO ODONTOLÓGICOS S/A.
Rod. Abrão Assed, Km 53+450m - Ribeirão Preto - SP - Brasil
UNIDADE DE ÁGUA SYNCRUS G4
CONFIGURAÇÃO CONFIGURATIONCONFIGURACIÓN
APARELHO EQUIPMENT APARATO
Wellkang Ltd
EC REP
Suite B, 29 Harley Street LONDON, W1G 9QR, U.K.
EQUIPAMENTO DE CLASSE l
EQUIPAMIENTO DE CLASSE l
CLASS l EQUIPMENT
TENSÃO TENSION
12/24 V~
FREQUÊNCIA
50/60 Hz
NUM. REG. ANVISA
0499
TENSIÓN
FREQUENCY
FRECUENCIA
INSTALLATION OF EQUIPMENT
The installation of this equipment requires specialized technical assistance (Gnatus).
OBS: These information also make part of the Manual of Installation and Maintenance of the equipment that can be found with the authorized Gnatus technician.
2323
INSTALLATION OF EQUIPMENT
- This equipment shall only be able to be unpacked and installed by a Gnatus authorized technician, under penalty of losing the warranty, as only (s)he has the information, suitable tools and training required to execute this task.
- Gnatus bears no responsibility for damages or accidents caused by poor installation
executed by a technician not authorized by Gnatus.
- Only after the equipment has been installed and duly tested by the authorized technician
representing Gnatus, will it be ready to start work operations.
EQUIPMENT OPERATION
Control Panel
01
A
03
02
01 – Water triggering in the Bowl
02 – Water triggering in the cup holder
03 – Water triggering heating syringe
*
(optional)
WARNING:
To set the time of water ux in the cup
holder, press the "water in the cup holder" button (02) for 3 seconds (it will produce a
long beep and the LED will ash).
Upon reaching the desired time, press
again the button "Water in the cup holder" (02). The ux time is recorded. To set the time of the water ux in the bowl, press the "Water in the bowl" button (01) for 3 seconds (it will produce one long beep and the LED will ash). Upon reaching the desired time, press again "Water in the Bowl" (01). The ux time is recorded.
The settings "Water in cup holder" and "Water in the bowl" have a water ux timeout, 1
minute for the water in the cup holder and 4 minutes for the water in the bowl.
When turning the key "triggering heating water syringe" (03), the LED will turn on (A) starting the heating of syringe water. The temperature must remain around 40º C. To turn off the "trigger water heating", press (03) again.
Terminal Triggering
For the functioning of the rotating instruments remove from the support the instrument to be used, trigger the control pedal moving the lever (A) with the feet.
B
The power (air supply) can be controlled by the operator with higher or lower pressure on the pedal lever (A).
Chip Blower System:
The "chip-blower" system allows the
release of the airow with the stopped turbine
(air function). Pressing the (B) key downward, air will trigger on the tips.
Pressing the key (B) downward and moving the lever to the right (A) together, it will
trigger the high speed air turbine and water (spray).
24
A
EQUIPMENT OPERATION
Adjustment of Spray of “TB/TM high and low rotation terminals”
- The adjustment is made via a valve positioned in the terminal. Turn it in a clockwise direction to reduce the spray and in a counter­clockwise direction to increase it.
Note: As the “TB” double terminal does not
have a spray this adjustment is not required.
Ejectors operation
The ejectors (both BV and Venturi) start
working automatically when retired from the
tips support the BV ejectors feature suction ow
adjustment , and its regulated moving the lever located at the ejector up or down.
Use of 3-Way Syringe
- Press button (A) for water to come out, (B) for air to come out or both simultaneously
to obtain a spray.
B
C
03
A
Water Heating:
When you turn on the key "hot water activation" (03), LED will light (C), starting to heat water from the syringe. Temperature should remain about 40 °C. To turn off the "water heating activation" function, press key (03) again.
2525
EQUIPMENT OPERATION
Water supply to the bowl is driven by the "optical sensor *"
Water supply is done automatically through the optical sensor by simply approaching
the patient, providing greater convenience in operation.
Optical sensor positioning for right or
lefthanded.
The water unit is designed in order to meet the left or right hand users, with ease of installation of the "optical sensor set" on both sides, without loss of functionality in any case. At the time of installation by authorized technician from Gnatus, inform desired position (right or left handed). He will make necessary adjustments.
Replacement of the standard of cannula coupling
If there is the necessity of using the 6.5mm (A) cannula in the BV suctor, make the replacement of the cannula coupling, as the procedure below:
• Remove the coupling of 11 mm (B) by
unscrewing it from.
• Screw the coupling of 6.5 mm (C) in the
aspirator BV set and attach the coupling tube.
Coupling of tube of 6.5 mm
The curve of the coupling of the cannula was designed for better handling, but can also be cut at the location indicated with the aid of a knife.
A
C
B
* Optional item
26
EQUIPMENT OPERATION
Curing Light Activation
- Select application time, press time selection button (01), which values are: 10s (standard mode), 20s, 60s, 80s and 90s.
- To initiate a polymerization cycle, press the timer trigger (02), which generates a short beep every 10 seconds and a 4 beeps at the end of cycle.
- To interrupt a polymerization cycle just activate the timer trigger again (02).
IMPORTANT:
- Keep the light conductor tip (03) at least 2mm away from the restoration.
- Keep the light conductor (03) always protected by an expendable PVC lm, which must
be changed for every patient. This procedure protects the light conductor from scratches and other residues.
- Use the polymerization time recommended by the compound resin manufacturer and
always perform restorations in incremental layers with a maximum thickness of 2mm.
WARNING
- Never aim the blue light beam towards the eyes
- Use the eyesight protection (04)
- In order to protect the eyes, the eyesight protection (04) lters only the blue light
used for the resins polimerization, and it allows ambient light to pass through.
03
04
06
02
01
2727
EQUIPMENT OPERATION
Laser Hand
The “Laser Hand Kit” is low intensity (780nm) and provides relief of acute and chronic
pain, and speeds repair of damaged tissue by means of biostimulation effects of radiation.
Eminently analgesic, anti-inammatory and biomodulation effect.
Applications:
• Inammations;
• Oral mucous lesions;
• Dental hypersensitivity;
• Analgesia;
• Paresthesia;
• Alveolitis and pericoronitis;
• Acceleration of post surgical and
injury cicatrisation;
Activation of the “Laser Hand”
Turn on the main unit power switch, which will
automatically turn on the laser.
To select application time, press the time selection button (01) with variations of: 01s to 90s. Maintain pressure on the key until desired time selection, which can be at 1-second intervals (1s, 2s, 3s, 4s, 5s, 6s, 7s...) or 10-second intervals (10s, 20s, 30s, 40s, 50s..).
To start, press timer activation button (01). A single beep will be heard, followed by 5 beeps at each conclusion.
The laser will remain active with a 10-minute program. After 10 minutes, a beep will inform that the laser is in standby mode.
To restart the cycle, press the key (02) which will sound 2 beeps and the last programmed selection will appear on the screen. To interrupt the cycle, press button (02).
• Decrease of edemas, bruising and
scabbing;
• Distension, muscular spraining and
articular pain;
• Acupuncture (optional).
02
01
Note: For a new program, in case desired time is less than the previous program, press (01) until the start of time “00”.
WARNING: Never direct the red light towards eyes.
28
EQUIPMENT OPERATION
Water ow adjustment
01 - Cup ller adjustment 02 - Bowl ush adjustment To regulate the bowl ush and cup lling water
ow, use the bowl ush adjustment (02) and the cup ller ow adjustment (01), to increase ow, turn it
anticlockwise, to decrease, turn in clockwise.
Regulation of the water selecting valve
To regulate this kind of water feeding, please use the selecting valve (04) to select the feeding through the reservoir and turn around in the clockwise sense. To select the feedign through the net, turn around counter clockwise sense.
Master Valve
The master valve is a safety device that aims to block / release the entry of water to the dental set. It is of utmost importance to have interrupted the water supply to the dental set in the end of the working day, which can be done through the key (ON/OFF - 03)
01 0402 03
A B
How to provision the reservoirs Water - Syringe/Handpieces
Remove the reservoir (B) uncoiling it on clockwise and make the replacement of water.
After the replacement put it back coiling on anticlockwise. Always use ltered water or
aseptic products.
Bio-System
Remove the reservoir (A) uncoiling it on clockwise and make the replacement. Use a
chlorinated water solution 1:500
Preparing the solution: From a solution of hypochlorite of sodium at 1%, a solution of chlorine at 500 p.p.m.
is prepared.
How to prepare the solution: Take 25ml of hypochlorite of sodium at 1% and dilute it in
500 ml of water (1 to 20). Such solution should be prepared daily.
IMPORTANT: Follow this proportion strictly to avoid damages in the equipment and to
have an efcient result in the disinfection.
Bicarbonate Jet "Jet Hand"
For further information, please see the Jet Hand manual which comes with the product.
2929
PRECAUTIONS, RESTRICTIONS AND WARNINGS
Transportation, storage and operation
This equipment must be transported and stored observing the following directions:
- Avoid falls and impacts;
- Keep it dry, do not expose it to rain, water drops or wet oor;
- Keep it away from water and direct sunlight, and in it original wrapping;
- Don’t move it over irregular surfaces, protect it from rain and observe the maximum
stack quantity specied in the packaging;
- Transportation and storage temperature range: -12°C to 50°C.
- Ambient temperature range recommended by Gnatus +10 ° C to +35 ° C.
The Equipment maintains its condition of safety and efcacy, provided that it is
maintained (stored) as mentioned in this instruction of use. Thus, the equipment will not lose or alter its physical and dimensional features.
Sensitivity to environmental conditions in normal situations of use
The equipment has been planned not to be sensitive to interference such as magnetic
elds, external electrical factors, electrostatic discharge, pressure or variance of pressure,
provided that the equipment is installed, maintained, clean, preserved, transported and operated as per this instruction for use.
Precautions and warnings “during the installation” of equipment
- The equipment should only be installed by Gnatus authorized technical assistance or
technicians.
- Position the unit in a place where it will not get wet.
- Install the unit in a place where it will not be damaged by the pressure, temperature,
humidity, direct sunlight, dust, salts, or sulfur compounds.
- The unit should not be submitted to inclination, excessive vibrations, or blows (including
during transportation and handling).
- This equipment was not planned for use in an environment where vapors, anesthetic
mixtures inammable with air, or oxygen and nitrous oxide can be detected.
- Before the rst use and/or after long interruptions from work such as vacations, clean
and disinfect the equipment; eliminate air and water deposited in the internal hoses.
These information also make part of the Manual of Installation and Maintenance of the equipment that can be found with the authorized Gnatus technician.
30
PRECAUTIONS, RESTRICTIONS AND WARNINGS
Recommendations for the dental equipment maintenance
Your Gnatus equipment has been designed and developed according to the standards
of modern techology. Similarly to other kinds of equipment, it requires special care, which is many times neglected due to several reasons and circunstances.
Therefore, here are some important reminders for your daily routine. Try to follow these
simple rules, which will save you a lot of time and will avoid unnecessary expenses once they start making part of your working procedure.
Precautions and warnings “during the use” of equipment
- The equipment should only be operated by duly enabled and trained technicians (Dental
Surgeons, Capacitated Professionals)
- If any maintenance should be required, only use services of the Gnatus Authorized
Technical Assistance.
- The equipment has been manufactured to handle both continuous and intermittent
operation; so follow the cycles described in these Instructions for Use.
- Although this equipment has been planned in accordance with the standards of electromagnetic compatibility, it can, in very extreme conditions, cause interference with other equipment. Do not use this equipment together with other devices very sensitive to interference or with devices which create high electromagnetic disturbance.
- Do not expose the plastic parts to contact with chemical substances, use in the routines of dental treatment, such as: acids, mercury, acrylic liquids, amalgams, etc.
Bicarbonate Jet:
- It is not advisable to use this equipment in patients who have serious renal or respiratory alterations, or who undergo hemodialysis. These cases should be followed be followed by a doctor.
- We recommend the use of a mask and goggles for applying the bicarbonate jet.
- Avoid leaving sodium bicarbonate in the container for long periods without use.
The effect of residual humidity in the air may alter the properties of the powder and
cause blocking.
Gnatus shall not be responsible for:
- Use of the equipment differing from that for which it is intended.
- Damages caused to the equipment, the professional and/or the patient by the incorrect installation and erroneous procedures of maintenance, differing from those described in these Instructions for use which come with the equipment or by the incorrect operation of it.
Precautions and warnings “after” the use of equipment
- Turn off the main switch of the dental set when it is not in use for an extended period
of time.
- Always maintain the equipment clean for the next operation.
- Do not modify any part of the equipment. Do not disconnect the cable or other
connections without need.
- After using the equipment, clean and disinfect all the parts which may be in contact
with the patient.
- Upon noticing irremovable stains, splits or cracks in the light conductor or in the eye
protector, replace the damaged components.
3131
PRECAUTIONS, RESTRICTIONS AND WARNINGS
Precautions and warnings during the “cleaning and disinfection” of equipment
Unidad:
- Before cleaning the equipment, turn off the main switch.
- Avoid spilling water, even accidentally, or other liquids inside the equipment, which
could cause short circuits.
- Do not use microabrasive material or steel wool when cleaning, or employ organic
solvents or detergents which contain solvents such as ether, stain remover, gasoline etc.
Amalgam collecting vessel, lters and drains:
- To prevent infection risks, use protective gloves during amalgam collecting vessel
replacement and when handling lters and drains. Dispose wastes and contaminated
products in biological waste.
Curring Light:
- The equipment and the light conductor cannot be placed in the oven or autoclaves.
- The conductor can’t be immersed in solvents or substances that contain acetone in its
composition.
- Avoid the light conductor to terminal to touch the resin to be polymerized.
- When using the Curring Light check if the light conductor output doesn’t have residues
that might obstruct the light beam.
Bicarbonate Jet:
For further information, please see the Jet Hand manual which comes with the product.
Laser Hand:
For further information, please see the Laser Hand manual which comes with the product.
Precautions in case of alteration in the functioning of equipment
- If the equipment has any abnormality, check if the problem is related to any item listed in the topic of unforeseen events (failures, causes and solutions). If it is not possible to resolve the problem, turn off the equipment, remove the power supply cable from the socket and contact your representative (Gnatus).
Precautions to be adopted against foreseeable or uncommon risks, related to the deactivation and abandoning of equipment
In order to avoid environmental contamination or undue use of the Equipment after it has become useless, it should be discarded in the suitable place (as per the local legislation of the country).
- Pay attention to the local legislation of the country for the conditions of installation
and disposal of residue.
32
CORRECTIVE AND PREVENTIVE MAINTENANCE AND PRESERVATION
Additional procedures for reuse
The equipment can be reused in undetermined, i.e. unlimited, quantities, only needing
to be cleaned and disinfected.
Cleaning and Disinfection
Important: In order to execute cleaning or any type of maintenance, ensure that the
equipment is disconnected from the electrical network.
The cleaning procedure below should be executed at the start of the working day and after each patient. Always turn off the main switch before executing the procedures of daily maintenance.
To clean the equipment, we recommend the use of “BactSpray
(Reg nº MS: 3.2079.0041.001-5) or any other similar product:
Active component: Benzalkonium chloride (tri-quaternary
ammonium)
Solution 50%................................................. 0.329%
Chemical composition: Butyl Glycol, Decyl polyglucose, Sodium Benzoate, Sodium Nitrate, Essence, Deodorized Propane / Butane, demineralized Water.
For more information concerning cleaning procedures, see manufacturer’s instructions.
WARNING:
- In order to prevent risks and damages to equipment, make sure
that the liquid does not enter into the unit.
- The application of other solvent-based cleaning products or sodium hypochloride isn’t recommended, because they may damage the equipment.
NOTE: The registration at the Ministry of Health of the “BactSpray” is executed separately from the product described in this manual, as the “BactSpray” is not manufactured by Gnatus.
Disinfection
Use clean and soft cloth dampened in alcohol 70% to disinfection of the equipment. Never use corrosive disinfectants or solvents.
Note: Use gloves and other systems of protection, during the disinfection.
3333
CORRECTIVE AND PREVENTIVE MAINTENANCE AND PRESERVATION
Clearing the suction system
Using Vacuum Pump (BioVac II or IV):
Gnatus suggests performing a daily suction of the clearance and disinfectant solution, avoiding the risk of cross contamination and increasing equipment service life. To perform the disinfection of your equipment we recommend the use of the “Sugclean” (MS Reg. No.:
31.080.003-2) product.
Indication: It is indicated for clearance of Ejector and hose suction system. It is important to perform the suction solution in all suction terminals, which it is also important to be open. Then, remove Ejectors from hose for asepsis.
Preparing the Solution: Add “Sugclean” 30mL in 1 liter of water. Aspirate the solution with maximum power of the Ejectors, and also put the liquid in the water unit bowl.
In the rst use of “Sugclean” product, we suggest adding
60mL of concentrated product in 1 liter of water during the rst
5 days in order to remove accumulated residues.
Composition:
• Active Drug: Phosphoric Acid 13.6%
• Excipients: Isopropyl Alcohol, Acidulant, Dye and Thickener.
Warning: do not use foaming product. NOTE: The registration at the Ministry of Health of the
“Sugclean” is executed separately from the product described in this manual, as the “Sugclean” is not manufactured by Gnatus.
Fig.A
Using Vacuum Pump (BioVac Sec):
CAUTION: For internal cleaning of Suction pipelines of the Pump Vacuum BioVac Sec, the use of "Sugclean" product or any similar product is not allowed; use only the recommended mixture below:
• After each patient leaves, perform the suction of 250 ml of clean water in each totally
open suctor;
• At the end of the working Day, perform the suction of the 250 ml mixture of bleach
mixed with 250 ml of clean water (proportionally divided in each used suctor); Then, remove Ejectors from hose for asepsis.
WARNING:
• Never use foamy products in the suction (deep cleaner, detergents, oaters, etc), this
procedure may damage the internal parts of the dry vacuum pump’s engine;
• Never use the bleach solution for external cleaning of any equipment, because this
mixture is highly corrosive and may damage metal parts.
34
CORRECTIVE AND PREVENTIVE MAINTENANCE AND PRESERVATION
Triple syringe
Only the syringe tip is autoclavable (01). The other pieces must be cleaned using a piece of cotton wool and alcohol 70% vol. Never use a hot air sterilizer.
Reservoirs
It’s highly recommended the cleaning of the water reservoirs, using chlorinated water solution 1:500 (as described previously).
Cleaning of the Ejector and lters
After the suction of the solution through the
suctor, take the lid (01) and the lter (02) and
wash them in running water.
01
0201
03
Cleaning of the drain
Pull the drain (03) with a tweezer, clean
and disinfect it.
Warning: Always use protection gloves
when manipulating lters and drainages.
Discard waste and contaminated products
in biological waste containers.
3535
CORRECTIVE AND PREVENTIVE MAINTENANCE AND PRESERVATION
Basin cleaning
In order to obtain better results in the clean the
bowl in your water unit, we strong recommedn the use of the product “Easy-Off Bang” or similar, nº Reg. MS: 3.00227-0.
Chemical composition: Glycolic Acid, Maleic
Anhydride, Citric Acid, Ethoxyled Fatty Alcohol, Essence and Water.
Apply the “Easy-Off Bang” in the water unit bowl
along with a smooth cloth up to clean it.
If you want additional information regarding
cleaness, please look at the instructions from the product.
Bio-System
Remove the terminal hand parts. Take the hand parts of
terminals to the sink or tank of water unit.
Completely open the spray register of the terminals.
Press the Bio-System triggering button for some seconds, located under the coupling panel reaching arm of the water unit, to perform the disinfection of internal components with bactericidal liquid. Then, trigger the control pedal for a few seconds to perform rinsing in order to remove chemical residues of bactericidal liquid internally retained in the components of water unit.
Important: This procedure should be done in the
beginning of the workday and after each patient.
36
CORRECTIVE AND PREVENTIVE MAINTENANCE AND PRESERVATION
Curring Light
The light conductor cleaning and the optical protector must be done using only neutral
soap and cotton. To the exterior of the pen use neutral soap or alcohol 70% vol.
Never use any other chemical based product than previous mentioned, because along
the time these products attack the surface of the instrument.
Never immerse the instrument in disinfection baths.
Laser Hand
For further information, please see the Laser Hand manual which comes with the product.
Bicarbonate Jet "Jet Hand"
For further information, please see the Jet Hand manual which comes with the product.
Preventive Maintenance
The equipment should be calibrated routinely, as per the legislation in force in the country.
But never with a period exceeding 3 years.
In order to protect your equipment, seek Gnatus technical assistance for periodic revisions
of preventive maintenance.
Corrective Maintenance
If the equipment has any abnormality, check if the problem is related to any of the items listed in the item Unforeseen Events (situation, cause and solution).
If it is not possible to solve the problem, turn off the equipment, and request Gnatus technical assistance.
UNFORESEEN EVENTS  SOLUTION OF PROBLEMS
Upon coming across any problem in operation, follow the instructions below to check
and repair the problem, and/or get in touch with your representative.
Problem Probable cause Solution
Water Unit
- Ejector without suction.
-Handpiece with low speed. -Inlet pressure below speci-
- Insufficient air pressure from compressor.
- Filter clogged with particles.
- Filter lid misplaced.
ed (80 PSI).
-Adjust air ow.
-Remove and clean lter.
-Remove lid and place it correctly.
-Adjust inlet pressure (80 PSI).
3737
UNFORESEEN EVENTS  SOLUTION OF PROBLEMS
Problem Probable cause
-No water from handpiece spray.
-Handpiece is not working. - Compressor disconnected. -Plug the compressor in.
-No water from syringe. -Reservoir run out of water.
- Bowl’s water ow and cup lling are not operating.
- There is no water ow in the
tank when the patient gets closer (optical sensor).
- Insufficient air pressure from compressor.
-Reservoir run out of water.
-Closed terminal.
-Compressor disconnected.
- Lack of water
- Water valve is closed
- Power cut
- Chair fuse burned.
-Chair’s main switch is off or terminal box is disconnected.
- Damaged-sensor.
- Distance between the pa­tient to the upper sensor larger than 300mm.
- Dirt on the lens sensor.
- Chair fuse is burnt.
-Adjust air ow.
- Put ltered water in reser- voir.
- Open terminal.
-Put ltered water in reser­voir.
-Plug compressor in.
- Check the water supply
- Open the water valve
- Check the energy supply
- Turn off the chair from mains power and request a Technician presence.
- Switch the main switch on or connect the terminal box.
- Request assistance by GNA­TUS Technician.
- Get closer to the sensor (less than 300 mm).
- Clean the sensor lens.
- Turn off the power supply to the chair and request asistance by the Technician.
Solution
-When Bio-system is ope­rated no disinfectant come from handpiece terminals.
Curring Light
-Equipment’s not working.
-Bio-system reservoir run out of water.
- Chair fuse burned.
-Main or chair switch is off.
-Power cut.
-Chair’s fuse burned.
38
-Put disinfectant in the re­servoir.
-Turn off the chair from mains power and request a Technician presence.
-Switch main/chair switch on.
-Check power supply.
-Turn off the chair from mains power and request a Techni­cian presence.
UNFORESEEN EVENTS  SOLUTION OF PROBLEMS
-Equipment is not polymeri­zing resins.
Bicarbonate Jet - For further information, please see the Bicarbonate Jet
Laser Hand - For further information, please see the Laser Hand manual
-Resin is not appropriate for LED’s photopolymerizer wave length range.
"Jet Hand" manual which comes with the product.
which comes with the product.
-Get the indicated resin for the photopolymerizer’s wave length range, one with con­tains photoinitiators based on camphorquinone.
3939
WARRANTY OF EQUIPMENT
This equipment is covered by the warranty terms and norms contained in the Warranty
Certicate that accompany the product.
FINAL CONSIDERATIONS
The most important aspect related to equipment care is that concerning spare parts. To guarantee the life span of your equipment, use only original Gnatus spare parts.
They are sure to follow the technical specications and standards required by Gnatus.
We must also point out to you our chain of authorized dealers. Only dealers that make part
of this chain will be able to keep your equipment constantly new for they count on technical
assistants who have been trained and on spedic tools for the correct maintenance of your
equipment.
Doubts and information: GNATUS Call center (55-16) 2102-5000 / SAC: 0800-7015-054.
40
4141
42
4343
Loading...