Operating manual for the TG uni 1 appliance tester
Safety information
1. Safety information
Tests on the electrical safety of electrical appliances may only be carried
out by qualified electricians or under their supervision. Carefully read the
following safety information before commissioning the TG uni 1 appliance
tester (referred to as appliance tester or tester in the following).
Symbols used in the operating manual and on the tester:
Warning:hazardous situation. Follow the operating manual.
Caution! Dangerous voltage, danger of electric shock.
Note. Make sure you follow the manual.
&&
&
&&
CECE mark of conformity.
The operating manual contains information and notes which are
&&
&
&&
necessary for the safe operation and use of the tester. Before use
(commissioning / installation) of the tester the operating manual
should be carefully read through and followed in all points.
If the manual is not observed or if you fail to observe the warnings
&&
&
&&
and notes, serious injury to the user and damage to the tester can
occur.
All the technical data and quoted standards in this manual are up-
&&
&
&&
to-date at the time of going to press and have been determined to
the best of our knowledge, nevertheless this data may be subject
to errors and printing errors. Therefore no legal responsibility or
any other liability can be accepted for incorrect information or
the conseqences of this information.
The definitive document for carrying out tests is the
original version of the respective regulation or standard.
There is no intention to
other property rights with this publication.
2. Introduction
You have acquired a high-grade device from the company Gossen Müller &
Weigert (GMW), with which you can carry out repeatable measurements
over a very long period of time. The product was calibrated during the
manufacturing process in accordance with the specified operating
procedures. All quality-relevant activities and processes carried out in GMW
are permanently monitored as part of the quality management system in
accordance with DIN EN ISO 9001:2000.
infringe on any existing patents and
4
Operating manual for the TG uni 1 appliance tester
Introduction
2.1 Model and type designation / identification
A type plate and serial number sticker are situated on the inside of the
housing cover (behind the accessories bag).In the case of queries, always
give the product designation and serial number.
2.2 Product description
The appliance tester has been developed for carrying out the following
measurements for testing the safety of electrical appliances in accordance
with DIN VDE 0701 and 0702 (BGV A3):
–Measuring protective conductor resistance with compensation of the
measuring line
–Measuring insulation resistance
–Measuring substitute leakage current
–Measuring protective conductor current (direct method or as differential
current)
–Measuring touch current
–Function test with measurement of mains voltage, load current, active
power, apparent power and reactive power, power factor and
frequency
–Testing of non-heating devices and extension cables
2.3 Other equipment features
–Stable, dust-proof carrying case as housing
–Large, easy-to-read touch-screen for operating the tester (backlit)
–Good/bad display in plain text (suitable for technically trained persons)
–Connection for barcode scanner for reading-in the test object’s ID
numbers
–USB-interface
–MMC-card for saving and transporting measured data
–PC standard software for saving and printing measurement reports
2.4 Scope of delivery
1 TG uni 1 appliance tester
1 safety test line, length 100 cm
1 safety crocodile clip, clip width 30 mm
1 safety test tip
1 USB cable, length 150 cm
1 MMC memory card 128 MB
1 CD-ROM with Windows®-PC standard software
1 accessories bag with velcro fastener and quick guide
1 operating manual
2.5 Optionale accessories
(not included in delivery)
– Barcode scanner
–RS232 PC interface cable, length 200 cm
–Measuring adapter for testing devices with tree-phase connection (in
preparation for plug connectors CEE 3-pole, 16 A, 230 V; CEE
5-pole, 16 A, 400 V and CEE 5-pole, 32 A, 400 V)
– Adapter for checking extension cables
– PC software with database functions (customer master data, device
master data, test data, analysis function, data backup)
2.6 Transport and Storage
Please retain the original packaging for dispatch at a later date, e.g. for
calibration. Transport damage that occurs due to inadequate packaging is
not covered by the manufacturer’s guarantee.
The tester must be stored in a dry, closed room. If the device it transported
in extreme tenperatures, it requires at least 2 hours acclimatisation before
being switched on.
5
Operating manual for the TG uni 1 appliance tester
Safety instructions
3. Safety instructions
The appliance tester TG uni 1 was constructed and tested in accordance
with the applicable safety regulations and left the workshop in a faultless
safety condition. In order to maintain this condition and ensure safe
operation, the user must observe the instructions and warnings contained in
this operating manual.
With all work, the applicable accidend prevention regulations of the
&&
&
&&
trade associations for electrical systems and equipment must be
observed.
To avoid electric shock, the applicable safety regulations and DIN-
VDE regulations regarding high touch voltage
without fail when working with voltages greater than 120V (60 V)
DC or 50 V (25 V) eff. AC. The values in brackets apply for
restricted areas (e.g. medicine, agriculture).
Measuring at a dangerous proximity to electrical systems should only be
carried out under the instruction of a responsible electrician, and never
alone.
Check the tester and the connecting cables for external damage before
every new operation.
Make sure that the tester and the connecting cables are in faultless
condition. The tester may not be used if one or more functions fail or if
functional readiness is not evident.
must be observed
The measuring lines and the measuring accessories may only be
touched in the designated handling areas.
Touching measuring connections, e.g. test tips must be avoided
under all circumstances.
If the safety of the operator is no longer guaranteed, the tester must
&&
&
&&
be decommissioned and secured against unwanted use. This is
the case if the device:
– shows obvious signs of damage
– no longer carries out the required measurements
– has been stored in adverse conditions for too long
The tester may only be used in the operating and measuring ranges
specified under Technical Data.
Prevent overheating of the tester by exposure to direct sunlight.
Only this will guarantee faultless functioning and a long sevice life of the
device.
Never open the housing of the tester because of the danger of high
voltages. The tester contains no parts that can be replaced by the user.
Only let the tester be serviced by qualified personnel.
6
Operating manual for the TG uni 1 appliance tester
Appropriate usage
4. Appropriate usage
The tester may only be used under the conditions and for the purposes for
which it has beeen designed. It is particularly important to observe the
safety instructions and the technical data regarding ambient conditions and
usage in a dry environment.
The tester may not be used for measurement in electrical systems.
The tester may only be plugged into a properly connected earthed mains
socket. This may be protected with a maximum 16 A!
The tester is designed for operation with a rated voltage of 230 V AC 50 Hz,
it may never be connected to higher voltage.
The maximum output current of the mains socket is 16 A!
The tester may not be used for constant measuring.
Operational safety is no longer guaranteed in the case of modifications or
conversions that have not been carried out by the manufacturer.
Maintenance or calibration work may only be carried out by the
manufcturer.
If the tester has been exposed to an electromagnetic field, this can impair
the functionality of the tester.
The tester may not be used near to explosive gasses, steam or dust, or in
a damp environment.
In order to avoid damaging the testing device, the test socket and the
measurement connections may not be connected to an external voltage
source.
Only the supplied original measuring lines or equivalent safety measuring
accessories may be used!
7
Operating manual for the TG uni 1 appliance tester
Operating elements
5. Operating elements
Explanation of the tetster’s connections, operating elements and displays
Mains cable with mains plug
for connecting the T ester to
the mains power supply
230 V +/-15%, 50 Hz
Thermal fuse for the test
object and tetster F 16 A
ON / OFF switch
Card slot for an MMC memory
card
USB socket for connecting a
PC
RS232 socket for
connecting a barcode
scanner
Storage compartment for the
mains cable
Housing cover with 2 snap fasteners
Accessories bag with quick guide
Case housing
Fig. 2
Handle (fold-away)
Test / mains socket for connecting the test object with earted
plug , 230 V 16 A, 50 Hz
Front panel, Fig. 1
5.1 Meaning of the signal tones
–short single tone:confirmation of button press
–short double tone:end of a test stage with automatic
Connecting socket ‘PE’ (yellow/green), parallel
connection of the protective conductor of the
test socket, for connecting the test object
without an earthed mains plug
connecting socket ´Probe´ (red), for the test line
when measuring RPE and R
T ouch-screen (LC display and control screen)
Symbols used in the LC display:
- internal memory
- MMC memory card
INS
procedure
Operating manual for the TG uni 1 appliance Tester
Commissioning
6. Commissioning
The tester is equipped with modern touch-screen technology. That means
that the LC display and the control buttons are in one control screen. The
control buttons are always edged.
After activating the mains switch a self-test is carried out. If everything is in
order, the following display appears on the touch-screen as a switch-onmessage (Fig. 3):
Fig. 3
Type designation
Relevant standard
Firmware version
Manufacturer
Large button
Continue with large button (Fig. 3).
A display for entering the date appears (Fig. 3a).
Fig. 3a
As the tester does not have its own real-time clock, you are prompted to
confirm the displayed date or enter a new date.
Also see Point 6.4.
&&
&
&&
After confirmation with the [OK] button the Main menue appears (Fig. 4):
Fig. 4
Headline
Function buttons:
T ests according to standard
T esting of individual parameters
Basic settings of the tester
6.1 Basic settings of the tester
Press the [Configuration] button in the Main menue (Fig. 4).
The Configuration menue appears (Fig. 5):
Headline
Fig. 5
Symbol for the internal memory
or the MMC memory card
Measuring method for protective
conductor or touch current:
DIR. - direct Measurement
(test object insulated from earth)
DIF. - differential current method
(Test object not insulated from
earth)
Function buttons
9
Operating manual for the TG uni 1 appliance tester
Commissioning
6.2 Compensation of the measuring line
In order to achieve correct results when measuring the protective conductor
resistance, the resistance of the measuring line must be compensated (zero
balance).
Press the fuction button [Comp.] (Fig. 5).
The zero balance menu appears. Follow the instructions in the display
(Fig. 6):
Fig. 6
– Connect the test tip/test
clip with the measuring line
and insert the measuring
line plug in the ‚Probe’
socket (Fig. 1).
– Connect the test tip/test
clip with the protective
earthing contact of the test
socket.
If the following message appears in the display:
‚Fault: zero balance’
and a permanent warning signal sounds, the measuring line resistance is
greater than 2 Ω and cannot be compensated.
The protective conductor resistance test is locked.
In this case the measuring line should be checked or be replaced with a
low-impedance version.
If the measuring line has been successfully compensated – the following
message appears in the display
‚Zero balance OK’.
Press the [Next] button, only then disconnect the line!
6.3 Setting the measurement method for protective conductor
current measurement (also current touch measurement)
In the case of devices of protection class I, on which the insulation
resistance measurement cannot or may not be carried out, the protective
conductor current measurement is an alternative measurement for
determining the insulating capacity.
The protective conductor current can be determined either by the direct
measuring method, the differential current method or the substitute leakage
current method.
The direct measuring method or the differential current method should be
used with test objects that can only be switched on with mains voltage
(also see Point 9.8).
Caution! In this case the test object is supplied with mains
voltage during the test.
The differential current measurement determines the total leakage current of
the test object by measuring the total current of all avtive conductors (L-N).
The differential current measurement must be applied if the test object has
additional earthing connections or cannot be insulated.
Direct measurement can be applied if the test object has no additional
earthing connections or can be insulated.
In order to change the measuring method, press the [Measure] function
button in the configuration menu, see Fig. 5.
The symbol in the headline changes accordingly:
[DIR.] – direct measurement
[DIF.] – differential current method
10
Operating manual for the TG uni 1 appliance Tester
Commissioning
6.4 Setting the date
You can set the test date on the tester.
Press the function button [Date] to change to the input screen (Fig. 5). The
date is entered on the touch-sreen by means of the pictured 10-button
keypad (Fig. 7).
The date can be entered in the following formats:
|D.M.YY|, |DD.MM.YY| and |DD.MM.YYYY|.
The set date is safed along with the measuring results of every test and
also appears in the log prinout.
Fig. 7
once the date has been set, it remains unchanged in the tester’s
&&
&
&&
memory until it is reentered or deleted
- the tester is not equipped with a real-time clock!
6.5 Selection of the data memory
To save the measurement results, the tester has an internal memory (16
MB) and an MMC memory card (128 MB, supplied with the tester).
The MMC memory card can be used to backup or transfer the
measurement results.
MMC memory cards with a capacity of 128 to 256 MB can be used.
Inserting the MMC memory card:
Insert the memory card, with the label facing down, into the card slot and
slightly press until it engages.
To eject the memory card, press it down until it disengages and pull the
card upwards and out of the slot.
Caution!
Never force the memory card into the card slot of the tester.
The use of force can damage the memory card and the card
slot. If the card is not recognised by the tester, check if it
has been properly inserted.
Press the function button [Memory] in the configuration menu to switch
between the internal memory and the MMC memory card. The symbol in
the headline changes accordingly (Fig. 5):
– internal memory
– MMC memory card
The measurement results can be stored either in the internal
&&
&
&&
memory or on the MMC memory card.
Use a PC with a USB connection if you wish to completely delete or format
the MMC memory card. After the USB connection has been made the MMC
memory card appears as an external storage medium in the file manager.
From here the card can be deleted or formatted – the file system is
automatically adjusted.
Caution! Formatting a memory card deletes all the data on the
card.
To read out saved measurement results - see the operating manual of the
PC software
11
Operating manual for the TG uni 1 appliance tester
Commissioning
6.6 Changing the duration of a test stage
With automatic test procedures, the preset duration of every test stage is 5
seconds.
This setting can be changed, if for example the test object requires a longer
time to be switched on.
To do this press the function button [T.-time] in the configuration menu.
Using the 10-button keypad, you can enter a new test time in the next
screen (Fig. 8).
The possible setting range is 3 to 254 seconds.
The factory setting of the tester is 5 seconds per test stage.
Fig. 8
6.7 Entering the name of the inspector
To enter the name or designation of the inspector, press the function button
[Inspector] in the configuration menu.
The screen changes to an alphanumeric keypad, with which you can enter
the name or the designation of the inspector. The input field has 20
characters (Fig. 9).
You can switch between letter and numeric keys by using the buttons [123]
or [ABC].
The [C] button deletes the last entered character.
Press the [OK] button to confirm the entry.
The entered name is saved in the memory after the tester has been
switched off.
The entered name is assigned to every test carried out and also
&&
&
&&
appears in the log printout.
Fig. 9
6.8 Customer-specific settings
The scope of functions of the tester can be adapted to customer
specifications. This can simplify the operation of the tester.
These settings do not change the standard conformity of the tester!
You reach the setup menu by pressing the function button [Setup] in the
configuration menu (Fig. 5).
This input area is locked by a code key
&&
&
&&
(Fig. 10)
Please contact us for further information:
Tel. +49 (0)911 / 3502-0, Fax +49 (0)911 / 3502-307.
Fig. 10
12
Operating manual for the TG uni 1 appliance tester
General Information on DIN VDE 0701 and DIN VDE 0702
7. Conducting tests:
General information on DIN VDE 0701 and DIN VDE 0702
Why, What, How
7.1 Why?
Electrical systems and equipment must be safe. They have to be
manufactured and operated in such a way that they cannot endanger
people, animals and property.
The legal regulations are primarily anchored in:
–The German Ordinance on Industrial Safety and Health ( BetrSichV)
–The Accident Prevention Regulations (BGV A 3)
–The DIN-VDE standarts harmonised within the EU
The stipulation for transportable devices states:
„The manufacturer or the importer of technical equipment may only bring
this equipment onto the market if, in accordance with the generally
recognised rules of technology, the user or third parties are protected from
danger when using this equipment for its intended purpose, as far as the
form of appropriate usage allows.“
(from the German Ordinance on Industrial Safety and Health )
The operator has to ensure that the electrical systems and equipment are
operated in accordance with the rules pertaining to electrical equipment.
(from BGV A 3 Electrical systems and equipment)
Those who manufacture or modify transportable devices must comply with
the applicable standards, e.g.:
DIN VDE 0700 Part 1 „Household and similar electrical appliances - Safety“.
7.2 What?
Whoever operates, services and repairs transportable devices must
observe the following:
DIN VDE 0701„Repair, modification and inspection of electrical appliances“
Part 1 as well as the special provisions of this standard
For those commissioned with regular tests, the applicable specifications are
contained in DIN VDE 0702 „Periodic inspection on electrical appliances“.
Requirements for the appliance tester from the VDE regulations:
Safety tests on transportable electrical equipment are based on
– DIN VDE 0701
„Repair, modification and inspection of electrical appliances“
– DIN VDE 0702
„Periodic inspection of electrical appliances“
7.3 How?
Measuring and testing devices must be constructed in accordance with DIN
VDE 0404 „Testing and measuring equipment for checking the electric
safety of electric devices - Part 1: General requirements“.
Insulation and protective conductor measurements are also subject to the
standards DIN VDE 0413 Part 1 and Part 4.
In addition:
DIN VDE 0404 „Testing and measuring equipment for checking the electric
safety of electric devices - Part 2“.
13
Operating manual for the TG uni 1 appliance tester
Explanation of terminology
8. Conducting tests:
Explanation of terminology
8.1 Touch current (IT)
Current that can flow to earth via the person handling the device (test
object).
The measurement is carried out between touchable conductive parts of the
test object and the earth.
The measurement can either be made directly or with the differential current
method.
Direct measurement is applicable if the test object can be insulated from
earth.
In all other cases the differential current method should be used.
The touch current measurement is carried out on test objects in protection
class II with touchable conductive parts, or on test objects in protection
class I, which have touchable conductive parts that are not connected to the
protective conductor.
The measurement is to be carried out in both positions of the mains
&&
&
&&
plug - the reversal of polarity is automatic with the tester extracting and reversing the mains plug is not necessary.
Also see measuring principle circuits on Page 19.
&&
&
&&
8.2 Differential current (ID)
In terms of the standards DIN VDE 0701/0702, this is the total instantaneous
values of all currents flowing through all active conductors on the mains
side connection of the device (test object).
The differential current method is a measuring method for determining
protective conductor current or touch current.
The total leakage current of a test object can be measured. This measuring
method must be applied if the test object cannot be insulated.
Also see measuring principle circuit M6 on Page 18.
&&
&
&&
Caution! The test object is supplied with mains voltage
during the test.
8.3 Substitute leakage current (I
Current that would flow through the interconnected active conductors of the
device (test object) and the protective conductor or the touchable conductive
parts at nominal voltage and nominal frequency of the device.
This measuring method determines the leakage current without mains
voltage. It is an alternative measuring method for determining protective
conductor current or touch current.
Also see measuring principle circuit M4a on Page 18.
&&
&
&&
)
SL
14
Loading...
+ 30 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.