GMW DPM 24x96-2000 S Operating Manual

Operating Manual
Digital Panel Meter DPM 24x96 - 2000 S
The DP
M 24x96-2000 S are digital panel instruments for measurement of current DC and voltage DC with 9 measuring
ranges. 3 ranges for power supply are available.
Service:
The measuring ranges and zero point selectable using jumpers at the back. The choise of many factory pre-set measuring ranges is included, with user adjustments available for the zero and full scale values.
Technical data:
Display Display range 1999
Decimal point position selectable Negative display indication „-“ Digit height 14 mm / 7 segment digits Overload indication last 3 digits blank Setting time < 1 s (0 ... 99 %)
Measuring ranges Direct current 0 ... 1 mA
(Voltage drop < 600 mV) 0 ... 20 mA
0 ... 200 mA
Direct voltage 0 ... 0.2 V (1 V) (input current < 300 µA) 0 ... 1 V (5 V)
0 ... 5 V (25 V) 0 ... 25 V (100 V) 0 ... 100 V (300 V) 0 ... 300 V
Range adjustment span from 20 % to 100 % Zero offset (suppression) up to -1000 Accuracy < 0,2 %+ 1 digit Temperature coefficient 100 ppm / K
plus 100 ppm / K zero point drift
Digital input HOLD (option) „0“ Options Brightness adjustment
Zero point adjustment Hold input
Power supply Direct voltage DC 5 V ±5 % max. 0,6 W not isolated
common mode voltage -0,5 V ... 1,5 V max.
21 ... 30 V ±15 % 4,0 W approx. Alternating voltage AC 230 V +10 %, -15 % 1,3 W approx. (47 - 60 Hz)
Environmental conditions Climatic class Class 2 to VDE / DIN 3540
Operating temperature 0 ... 50 °C Storage temperature -40 °C ... +80 °C Safety class II to IEC 348 / VDE 041 1 Protection class IP 20 to IEC EN 60 529 front IP 54 (IP 65 on request) Device safety corresp. IEC EN 61 010 EMC immunity corresp. DIN EN 61 000-4-1 to 4 EMC radiated interference corresp. DIN EN 50 081 class B
Dimensions and weights Front frame 24 mm x 96 mm to DIN 43 718
Panel cut-out 22,2 +0,3 mm x 92 +0,8 mm Overall depth 72 mm Weight approx. 90 g (210 g if power supply >5 V DC)
Sundry Connections Plug-in screw terminal blocks
DP
M24x96-2000S - Subjekt to change 09/2017
Connections Rating plate
12
3
3
4
4
55
6
6
+U -U
*
*
Current measurement
Power supply
Power supply
1
2
-I +I1 -I +I2
+U -U
I
I
(1 mA ... 100 mA) (100 mA ... 300 mA)
*) For current measurement, the user has to externally link terminals 3 (+U) and 5 (+I1), or terminals 3 (+U) and 6 (+I2).
3
4
5
6
2
+U -U
+
U
1
Measuring input
Voltage measurement
Power supply
Note:
In order to avoid errors or an unstable readout, the measuring circuits be shielded or at least twisted.
Operation
To set the basic range, first remove the safety cover over the pin connector at the rear of the instrument. To set the basic range and decimal point, plug in jumpers according to the assignment table. A zero shift is possible up to 25 % of the full-scale range.
1 5 10 1237
12 3456
Design and installation
Panel cut-out
Connections
Power supply
Measuring ranges
Decimal
point
Measuring input
DPM 48/96-2000 S 14
5V ±5% / 200mA 21 ... 30 V / 100 mA
24V 115V 230V
}
0,2........1 V 2
1........5 V 3
5......25 V 4
25......100 V 5
100....300 V 6
1,0.........4,0 mA 2
4,0.......16,0 mA 3
16,0.......64,0 mA 2;8
64,0.....100,0 mA 3,8
100,0.....300,0 mA 2
DC AC
+10%/-15%
DPM 48/96-2000 S 20
CE
DPM 24/96-2000 S
75 ±0,3
61,5 +0,3
88,5 -0,1
91,5 -0,1
24 -0,2
96 -0,2
16
22 -0,2
56 +0,3
5
8
105,5