Gmi Shipsurveyor User Manual

Page 1
User Handbook
Gas Measurement Instruments Ltd
Page 2
Issue 8
10/05/17
Part Number: 48160
GMI welcomes comments on all our publications. Your comments can be of great value in helping us to improve our customer publications. Please send any comments that you have to customerservice@gmiuk.com
Copyright © Gas Measurement Instruments Ltd 2008
Page 3

COPYRIGHT

COPYRIGHT
This User Handbook is copyright of Gas Measurement Instruments Ltd (GMI) and the information contained within, is for use only with
SHIPSURVEYOR instruments. Reproduction, in whole or in part,
including utilisation in machines capable of reproduction or retrieval without written permission of GMI is prohibited. Reverse engineering is not permitted.

LIABILITY

Every care has been taken in the preparation of this document, but GMI do not accept any responsibility for errors or omissions and their consequences. Information in this document is subject to change
without notice. This document does not constitute a specication or
basis for a contract. Your statutory rights under law are not affected.

MODIFICATION NOTICES

GMI aim to notify customers of relevant changes in the product operation and maintain this manual up to date. In view of the policy of continuous product improvement there may be operational differences between the latest product and this manual.
This Handbook is an important part of the SHIPSURVEYOR product. Please note the following points:
• It should be kept with the instrument for the life of the product.
• Amendments should be attached.
• This Handbook should be passed on to any subsequent owner / user of the instrument.
• Although every care is taken in the preparation of this Handbook, it does not constitute a specification for the instrument.

SOFTWARE

Software supplied on EPROM or similar device for use in a particular product, may only be used in that product and may not be copied without the written permission of GMI. Reproduction or disassembly of such embodied programmes or algorithms is prohibited. Ownership of such software is not transferable and GMI does not warrant that the operation of the software will be error free or that the software will meet the customer’s requirements.
i
Page 4
USER HANDBOOK

DISPOSAL ADVICE

When no longer in use, dispose of the instrument carefully and with respect for the environment. GMI will dispose of the instrument without charge if returned to the factory.

SAFETY

• The instrument must be regularly serviced and calibrated by fully trained personnel in a safe area.
Batteries: Alkaline or *Rechargeable batteries must be exchanged (*and recharged) in a safe area and fitted correctly before use.
Never use damaged batteries or expose to extreme heat.
See Chapter 5 : OPERATOR MAINTENANCE.
• Only GMI replacement parts should be used.
• If the instrument detects gas, follow your own organisation’s procedures and operational guidelines.
• Shipsurveyor instruments are certied as
ATEX II 2 G Exd ia IIB T3 Gb (-20oC < Tamb < 50oC).
IEC Ex d ia IIB T3 (-20oC < Ta < +50oC)
• This equipment conforms to standard EN 50104.
• This equipment is also designed and manufactured to comply
with MED Directive 0038/YY (Module B&E). Further detail
of the Marine Equipment Directive, is located on the MED declaration of conformity, supplied with the instrument.
Any right of claim relating to product liability or consequential damage to any third party against GMI is removed if the warnings are not observed.
WARNING: To prevent ignition of flammable or combustible atmospheres, remove batteries before servicing.
WARNING: To prevent ignition of flammable or combustible atmospheres, read, understand and adhere to the manufacturer’s live maintenance procedures.
WARNING: To reduce the risk of ignition of a flammable or explosive atmosphere, batteries must be changed only in a location known to be non-hazardous.
ii
Page 5
COPYRIGHT
WARNING: To reduce the risk of explosion, do not mix old batteries with used batteries or mix batteries from different manufacturers.
WARNING: Never attempt to recharge non rechargeable cells.
CAUTION: Not for use in oxygen enriched atmospheres.
CAUTION: Replace instrument batteries only with approved batteries, as follows:
ATEX APPROVED INSTRUMENTS: Alkaline (LR20 ‘D’ size): Duracell Procell; Duracell
Industrial; Duracell Plus; Energizer Ultra; Energizer Industrial.
Rechargeable (LR20 ‘D’ size): Uniross 2600mAh NiMH.

AREAS OF USE

Do not use instrument in potentially hazardous atmospheres containing greater than 21% Oxygen. The enclosure material is polypropylene and must not be exposed to environments which are liable to result in mechanical or thermal degradation or to damage caused by contact with aggressive substances. Additional protection may be required in environments where the instrument enclosure is liable to damage.

STORAGE, HANDLING AND TRANSIT

Rechargeable batteries contain considerable energy and care should be taken in their handling and disposal. Batteries should be removed if the instrument is stored for longer than 3 months. The instrument is designed to handle harsh environments and is IP54 rated. If not subjected to misuse or malicious damage, the instrument will provide many years of reliable service. The instrument contains electrochemical sensors with a life of 2 years. Under conditions of prolonged storage the sensors should be removed. The sensor contains potentially corrosive liquid and care should be taken when handling or disposing of the sensor, particularly when a leak is suspected.

WARRANTY

The SHIPSURVEYOR instrument has a warranty against faulty goods or workmanship of 2 years. Consumable and Mechanical parts are not included in this. These are covered under GMI standard warranty conditions. For details, please contact GMI Ltd (UK).
iii
Page 6
USER HANDBOOK
iv
Page 7
CONTENTS
COPYRIGHT ..................................................................i
LIABILITY .......................................................................i
MODIFICATION NOTICES .............................................i
SOFTWARE ...................................................................i
DISPOSAL ADVICE ...................................................... ii
SAFETY ......................................................................... ii
AREAS OF USE ........................................................... iii
STORAGE, HANDLING AND TRANSIT ....................... iii
WARRANTY ................................................................. iii
INTRODUCTION ................................................ 1-1
GENERAL INFORMATION .................................. 2-1
2.1 Instrument Models / Ranges ................................ 2-1
2.2 Details of Operating Ranges................................ 2-2
2.2.1 LEL, 0 - 100 % ............................................... 2-2
2.2.2 Volume Gas, 0 - 100 % (CGI mode only) ....... 2-3
2.2.3 Oxygen (O2), 0 - 25 % .................................... 2-3
2.2.4 Hydrogen Sulphide (H2S), 0 - 100 ppm .......... 2-4
2.2.5 Carbon Dioxide (CO2), 0 - 20 % ..................... 2-4
2.2.6 Carbon Monoxide (CO), 0 - 1000 ppm ........... 2-5
v
Page 8
USER HANDBOOK
2.3 Operating Modes ................................................. 2-6
Conned Space Monitor (CSM) Operation ............. 2-6
Combustible Gas Indicator (CGI) Operation ........... 2-6
2.4 Alarms .................................................................. 2-7
2.5 Datalogging.......................................................... 2-7
2.5.1 Location (LOC) Selector ................................. 2-8
2.6 Max / Min Values ................................................. 2-8
2.7 Construction......................................................... 2-8
2.8 Batteries............................................................... 2-8
2.9 Filters ................................................................... 2-9
2.10 Liquid Crystal Display (LCD).............................. 2-9
2.11 Before Use Checks .......................................... 2-10
OPERATION ........................................................ 3-1
3.1 Instrument Features............................................. 3-1
3.2 Operating Buttons ................................................ 3-2
3.3 Probe Handle / Quick Connect ............................ 3-3
3.4 Operating Modes - Switch ON ............................. 3-3
3.5 Language Option ................................................. 3-4
3.6 Instrument Identication ...................................... 3-5
3.7 Time / Date .......................................................... 3-6
3.8 Calibration Due Date ........................................... 3-6
3.9 Service Due Date............................................... 3-12
vi
Page 9
CONTENTS
3.10 Sensor Zeroing ................................................ 3-18
3.10.1 Instrument Will Not Zero - Zero Fault ......... 3-18
3.10.2 Instrument Will Not Zero - Gas Present ..... 3-20
3.11 Alarm Warning (CGI mode only) ...................... 3-21
3.12 Warm-up Complete .......................................... 3-22
3.13 Alarms (CSM mode only) ................................. 3-22
3.14 LEL / Vol Gas Range Selection (CGI mode only) ....
................................................................................. 3-23
3.15 Switch Pump OFF / ON (CGI mode only) ........ 3-23
3.16 Zero All Ranges (CGI mode only) .................... 3-23
3.17 Datalogging...................................................... 3-24
3.17.1 CSM Mode ................................................. 3-24
3.17.2 CGI Mode ................................................... 3-25
3.18 Max / Min Values ............................................. 3-25
3.19 Location (LOC) Selector .................................. 3-28
3.19.1 To select a location (LOC): ........................ 3-28
3.20 Acknowledge Alarms (CSM mode only)........... 3-30
3.21 Switch the instrument OFF .............................. 3-31
3.22 Button Operation Summary ............................. 3-32
ALARMS .............................................................. 4-1
4.1 Gas Alarms .......................................................... 4-1
4.1.1 FLAMMABLE (LEL) ALARMS ....................... 4-2
4.1.2 OXYGEN (O2) ALARMS ................................ 4-2
4.1.3 TOXIC ALARMS (e.g. CO) ........................... 4-3
vii
Page 10
USER HANDBOOK
4.2 Alarm Types ......................................................... 4-4
4.2.1 LATCHING / NON-LATCHING ...................... 4-4
4.2.2 MUTING / ACKNOWLEDGING ..................... 4-4
4.3 Default Alarm Settings ......................................... 4-4
4.3.1 Default Alarms Table ...................................... 4-5
4.4 Alarm Examples ................................................... 4-6
4.4.1 EXAMPLE 1 (LEL HiHi ALARM)................... 4-6
4.4.2 EXAMPLE 2 (O2 LoLo ALARM) .................... 4-7
4.4.3 EXAMPLE 3 (CO STEL ALARM) ................. 4-8
4.5 Fault Alarms ......................................................... 4-9
4.5.1 Fault Alarms - CSM Mode ............................ 4-10
4.5.2 Fault Alarms - CGI Mode ...............................4-11
4.5.3 ZERO FAULT .............................................. 4-12
4.5.4 SENSOR FAULT ......................................... 4-13
4.5.5 FLOW FAULT .............................................. 4-14
4.5.6 MEMORY FAULT (Calibration / Conguration) ...
.............................................................................. 4-15
4.5.7 MEMORY FAULT (General Instrument) ...... 4-16
4.5.8 LOW BATTERY ........................................... 4-16
4.5.9 BATTERY EXHAUSTED ............................. 4-17
OPERATOR MAINTENANCE ............................. 5-1
5.1 Batteries............................................................... 5-1
5.1.1 Replace Alkaline / Rechargeable Cells .......... 5-2
viii
Page 11
CONTENTS
5.2 Filter Replacement............................................... 5-5
5.2.1 Internal Filter .................................................. 5-5
5.2.2 Probe Handle (Accessory) - Part No. 48120 .. 5-5
CALIBRATION .................................................... 6-1
6.1 Calibration Validity ............................................... 6-2
ACCESSORIES................................................... 7-1
ADDITIONAL INFORMATION ............................ 8-1
8.1 Training ................................................................ 8-1
8.2 World Wide Web .................................................. 8-1
TYPICAL OPERATING PARAMETERS ..............A-1
Dimensions ............................................................. A-2
Weight ..................................................................... A-2
Operating Temperature ........................................... A-2
Humidity .................................................................. A-2
Construction ............................................................ A-2
Display .................................................................... A-2
Warm-Up / Stabilization Time .................................. A-2
Typical Flow Rate Information .................................A-2
Response Time (T90) ..............................................A-2
Power Source .......................................................... A-2
ix
Page 12
USER HANDBOOK
QUICK OPERATING INSTRUCTIONS ...............B-1
CHECKLIST ...............................................................B-2
SAFETY ..................................................................... B-2
AREAS OF USE ........................................................ B-3
OPERATION ..............................................................B-4
Switch ON ............................................................... B-4
Range Selection (CGI only) .................................... B-5
Switch Pump OFF / ON (CGI only) ......................... B-5
Zero All Ranges (CGI only) in Fresh Air .................. B-5
Manual Datalogging ................................................ B-5
Max / Min Values .....................................................B-5
Enter (LOC) Location List ....................................... B-6
Acknowledge Alarms (CSM only) ............................B-6
Switch OFF ............................................................. B-6
INDEX...................................................................... I
x
Page 13
1

INTRODUCTION

The GMI - SHIPSURVEYOR instrument is a single instrument solution for all your safety monitoring and inerting applications.
This extremely versatile instrument complies with international
marine regulations for Conned Space Monitoring, Inerting, Pre-entry Testing, Tank Cleaning and many other marine
applications.
Fig. 1-1 Shipsurveyor Instrument
As a minimum, the instrument provides measurement of Lower Explosive Level (LEL) and Volume ammable gas for Shipping Industry surveying and leak detection. Other gas ranges are available, as detailed in Chapter 2 of this handbook.
The SHIPSURVEYOR has a variety of user congurable options. This handbook details the default conguration, with possible options detailed in italic text.
1-1
Page 14
USER HANDBOOK
The instrument is supplied in a sturdy carrying case that is
able to accommodate a selection of accessories.
For a comprehensive list of accessories that are supplied in
the carrying case and additional accessories available, refer
to Chapter 7 ‘ ACCESSORIES’.
The main features of the SHIPSURVEYOR are:
• A gas detection instrument that is simple to operate with easy maintenance procedures.
• Two button operation allowing the user access to all
features.
• Two operating modes: Confined Space Monitor
(CSM) mode and Combustible Gas Indicator (CGI)
mode.
• LCD with backlighting which displays the current
gas readings together with operational and status information.
• Clear battery life indication.
• Audible and Visual alarms enabled in CSM mode
only. Alarm levels are pre-set and an optional periodic confidence signal is emitted. The audible and visual condence signals will provide assurance that the instrument is sensing for gas without the need for the operator to constantly
view the display. The signals, which consist of an audible ‘beep’ and a visual pulse of the LED approximately every eight seconds, are active in
CSM mode.
• Manual and automatic datalogging is available in
CSM mode.
Manual datalogging is available in CGI mode.
1-2
Page 15
Manual Datalogging
Live Reading
View Max Values (or Min if option displayed)
Location Selector
Alarms Activated
Gas Range
Battery Status
INTRODUCTION
• Rugged polypropylene case, sealed to IP54 rating and suitable for outdoor use.
• Directly interfaces with GMI Auto Calibration Unit.
During normal operation, the top and / or bottom line of the instrument display indicates button press options.
If the option is not highlighted, a single press of the adjacent button selects that option.
If the option is highlighted, a press and hold of the adjacent button selects that option.
An audible ‘beep’ will be heard for a single button press, and two ‘beeps’ will be heard for a press and hold, of either buttons.
Fig. 1-2 illustrates a typical CSM mode LEL display for a
Shipsurveyor 1.
Note: A congurable option allows the instrument to display this range type as LFL (Lower Flammable Limit).
Fig. 1-2 CSM Button Press Options
1-3
Page 16
USER HANDBOOK
Manual Datalogging
Toggle Pump OFF / ON
Zero All Ranges
Live Reading
Location Selector
View Max Values (or Min if option displayed)
Toggle Flammable Range (LEL to VOL)
Gas Range
or
Battery Status
Fig. 1-3 illustrates a typical CGI mode LEL display for a Shipsurveyor 1.
Note: A congurable option allows the instrument to display this range type as LFL (Lower Flammable Limit).
Fig. 1-3 CGI Button Press Options
Note: Figs. 1-2 and 1-3 illustrate instrument displays
with datalogging enabled (including LOC panel
and STORE option). If instrument LOC selector is
disabled, the display does not show the LOC panel.
1-4
Page 17
SENSORS FITTED
(* IR = infrared)
SHIPSURVEYOR MODELS
1 2 3 4 5 6 7 8
0-100% LEL (IR)
X X X X X X X
0-100% VOL GAS (IR)
X X X X X X X
0-25% Oxygen
X X X
X X X X
0-100ppm Hydrogen Sulphide
X X X
0-20% Carbon Dioxide (IR)
X X X X
0-1000ppm Carbon Monoxide
X X
2

GENERAL INFORMATION

2.1 Instrument Models / Ranges

There are eight (8) models in the SHIPSURVEYOR range of instruments. Not all detection ranges may be included in the version of instrument selected by your company.
The table below illustrates each model and corresponding gas sensor ranges included:
Fig. 2-1 Gas Ranges / Instrument Model
2-1
Page 18
USER HANDBOOK

2.2 Details of Operating Ranges

The basic instrument can measure 0-100% LEL and 0-100% VOL GAS.
Other models can measure the following gas types:
• Oxygen (O2)
• Hydrogen Sulphide (H2S)
• Carbon Dioxide (CO2)
• Carbon Monoxide (CO)

2.2.1 LEL, 0 - 100 %

The LEL range indicates the explosivity of the ammable
gas in the sample. This is displayed as a percentage of the lower explosive limit (LEL) of the gas.
% LEL is displayed clearly in the LCD. From 0 - 100% LEL, the digital display resolves to 1% LEL.
A congurable option allows the displayed LEL value to have a resolution of 0.1% LEL from 0 - 9.9%, and 1% LEL from 10 - 100%.
When the reading reaches 100% LEL, EEE (over-range) is displayed. The user can, by using the top button, change range to % Volume Gas in CGI mode only.
An example of a 10% LEL gas detection reading is illustrated in Fig. 2-2 (example from Shipsurveyor 1 model).
2-2
Fig. 2-2 LEL Example (CGI mode)
Page 19
GENERAL INFORMATION

2.2.2 Volume Gas, 0 - 100 % (CGI mode only)

This range displays the total volume of a ammable gas.
% VOL is displayed clearly in the LCD. The displayed VOL gas value has a resolution of 0.1% VOL from 0 - 9.9%, and 1% VOL from 10 - 100%.
A congurable option allows the displayed VOL gas value to have a resolution of 1% VOL from 0 - 100%.
Fig. 2-3 illustrates a typical VOL Gas display (example from Shipsurveyor 1 model).
Fig. 2-3 Volume Gas Example

2.2.3 Oxygen (O2), 0 - 25 %

This range displays the % of oxygen in the sample.
% O2 is clearly displayed in the LCD. The displayed oxygen value has a resolution of 0.1% O2 from 0 - 20.9%, and 1% from 21 - 25%. Fig. 2-4 illustrates a typical display including Oxygen (example from Shipsurveyor 2 model).
Fig. 2-4 Oxygen Example
2-3
Page 20
USER HANDBOOK

2.2.4 Hydrogen Sulphide (H2S), 0 - 100 ppm

This range displays the parts per million (ppm) of Hydrogen Sulphide in the sample.
H2S is clearly displayed in the LCD. The displayed hydrogen sulphide value has a resolution of 1 ppm. Fig. 2-5 illustrates a typical display including Hydrogen Sulphide (example from Shipsurveyor 3 model).
Fig. 2-5 Hydrogen Sulphide Example

2.2.5 Carbon Dioxide (CO2), 0 - 20 %

This range displays the percentage of Carbon Dioxide in the sample.
CO2 is clearly displayed in the LCD. The displayed carbon dioxide value has a resolution of 0.1% CO2 from 0 - 5%, and 0.5% from 5 - 20%. Fig. 2-6 illustrates a typical display including Carbon Dioxide (example from Shipsurveyor 4 model).
2-4
Fig. 2-6 Carbon Dioxide Example
Page 21
GENERAL INFORMATION

2.2.6 Carbon Monoxide (CO), 0 - 1000 ppm

This range displays the parts per million (ppm) of Carbon Monoxide in the sample.
CO is clearly displayed in the LCD. The displayed carbon monoxide value has a resolution of 1 ppm. Fig. 2-7 illustrates a typical display including Carbon Monoxide (example from Shipsurveyor 5 model).
Fig. 2-7 Carbon Monoxide Example
2-5
Page 22
USER HANDBOOK

2.3 Operating Modes

Conned Space Monitor (CSM) Operation
In this mode the instrument operates as a safety monitor for use when entering confined spaces which may
contain hazardous gas mixtures (ammable, toxic and/or
asphyxiant). The instrument continually samples the ambient atmosphere to which the operator is exposed and generates
alarms for low oxygen, high oxygen, high ammable gas,
high toxic gas, Short Term TWA (STEL) and Long Term TWA (LTEL).
During normal operation in CSM mode, the instrument
emits a condence beep and illuminates a pair of red LED’s briey every 8 seconds. This function is programmable in
the instrument setup software.
The condence signal function makes the user aware that
the instrument is sensing for gas without the need for the operator to constantly view the display:
The condence beep and / or LED indication can be disabled.

Combustible Gas Indicator (CGI) Operation

In this mode the instrument operates as a gas indicator drawing a sample via a probe from points where gas is suspected to be present. Alarms are disabled in CGI mode as the instrument is generally used for measuring gas levels rather than monitoring for the presence of gas.
2-6
Page 23
GENERAL INFORMATION

2.4 Alarms

The SHIPSURVEYOR alarms are disabled in Combustible
Gas Indicator (CGI) mode. When operating in Conned
Space Monitor (CSM) mode, both audible and visual alarms
are active, (refer to Conned Space Monitor Operation in
Chapter 3). The audible alarm is rated at 85 dB(A). The
visual alarm consists of two (2) pairs of red LED’s ashing
alternately in the display panel.
The instrument, by default, will have alarm levels pre-set,
(refer to ‘Default Alarms’ table in Chapter 4).
It is the responsibility of the user to ensure that the alarm levels, where set in the instrument, are appropriate for
the safe operation and legal requirements for the country / industry in which the unit is being used.
Latching alarms can only be cleared manually after the detected gas level has fallen below the alarm limit.
An option allows the alarms to be non-latching. These clear automatically when the gas levels fall below the alarm limit. Another option allows the audible alarm to be muted.
TWA alarms, by default, are non-latching.

2.5 Datalogging

The SHIPSURVEYOR memory can store over 1000 logged entries in date order. The memory will overwrite entries in
the event of an overow.
The instrument, by default, has automatic and manual datalogging enabled when operating in CSM mode. Manual datalogging allows the user to store a snapshot of the gas readings at any time.
When the instrument is operating in CGI mode, only manual datalogging is enabled. There is no automatic datalogging available in this mode.
2-7
Page 24
USER HANDBOOK
Alkaline Rechargeable
Two (2) Cells
> 20 hrs > 8 hrs

2.5.1 Location (LOC) Selector

The SHIPSURVEYOR instrument features a ‘LOC’ location selector, allowing the user to allocate datalogged readings to a location (LOC) of a vessel. Refer to Chapter 3.18 for more details.

2.6 Max / Min Values

The SHIPSURVEYOR instrument records the maximum / minimum gas values for each sensor, since switch-on.

2.7 Construction

The instrument is housed in a tough, impact resistant, moulded case made of polypropylene.
The top panel features a carbon loaded polycarbonate LCD cover.
The battery cover is attached to the main instrument body by means of two stainless steel hexagonal screws.
The instrument is sealed against dust and water to IP54 standard. The instrument sensors are protected from dust
and water by an integral hydrophobic lter.

2.8 Batteries

The SHIPSURVEYOR instrument can operate on either two
(2) alkaline or rechargeable cells. Typical operating times
are as follows:
Fig. 2-8 Typical Battery Operating Times
A battery symbol, as illustrated in Fig. 2-9, provides an indication of remaining battery life. The symbol is displayed in
2-8
Page 25
GENERAL INFORMATION
100%
75%
50%
25%
the bottom RH corner of the display during normal operation.
Fig. 2-9 Remaining Battery Life
Refer to Chapters 4.5.7 and 4.5.8 for detailed battery warnings.
To replace batteries, refer to Chapter 5, ‘OPERATOR
MAINTENANCE’.

2.9 Filters

Note: The SHIPSURVEYOR instrument has an integral
hydrophobic lter tted. To remove / install integral hydrophobic filter, refer to
‘MAINTENANCE MANUAL’ (GMI Part No. 48166). Where possible an external lter (accessory) should always
be used to prevent ingress of dust / water.

2.10 Liquid Crystal Display (LCD)

The dot matrix LCD (128 x 64) shows the current gas readings in a clear, digital form. Operational and status
information is also displayed. Backlighting is controlled by an
optical sensor that illuminates automatically during warm-up, when an alarm is active, and also in low lighting conditions.
2-9
Page 26
USER HANDBOOK

2.11 Before Use Checks

The following checks should be carried out before every
operation:
• The instrument is clean and in good condition.
• The sample line and any accessories are in good condition and leak free. Make sure that the instrument displays a flow fault when the
instrument inlet / sample line is blocked. If a ow fault is not displayed, check tightness of all ttings.
• Any accessory lters used are clean and in good
condition.
• The batteries have sufficient power left for the intended use of the instrument.
• The instrument is still within the calibration period and is calibrated for the gas that you are intending to sample.
• All gas ranges are operational and zeroed correctly.
2-10
Page 27
SHIPSURVEYOR
ALARM CONDITION (4 RED) LED's
and CONFIDENCE SIGNAL in CSM mode (outer 2 RED LED's)
POWER ON
(GREEN) LED
FAULT LED (YELLOW)
3

OPERATION

3.1 Instrument Features

The SHIPSURVEYOR front panel features two operating buttons, a Liquid Crystal Display (LCD) illuminated in low lighting conditions, and a series of status LED’s as illustrated in Fig. 3-1,
Fig. 3-1 Instrument Front Panel
3-1
Page 28
USER HANDBOOK
The POWER ON LED is green in colour and indicates that the instrument is powered ON.
The ALARM LED’s are red in colour and, when ashing,
indicate that the instrument is in an alarm condition. Refer to Chapter 4 ‘ALARMS’ for details.
The CONFIDENCE SIGNAL LED’s (CSM mode only) are also red in colour and provide assurance that the instrument is sensing for gas without the need for the operator to constantly view the display.
The FAULT LED is yellow in colour and when illuminated, indicates that a fault has been detected. Refer to Chapter
4.2 ‘Fault Alarms’, for details.

3.2 Operating Buttons

The SHIPSURVEYOR features two operating buttons, as illustrated in Figure 3-2.
All functions are easily accessible by a single press or a press and hold of the relevant button.
Each button press is acknowledged with a ‘beep’ sound. Each press and hold is acknowledged by two ‘beeps’.
Fig. 3-2 Switch Panel
Refer also to paragraph 3.21 ‘Button Operation Summary’ later in this chapter.
3-2
Page 29
OPERATION

3.3 Probe Handle / Quick Connect

The SHIPSURVEYOR instrument is equipped to accept probe types listed in Chapter 7 ‘Accessories’.
Fig. 3-3 Probe Connection

3.4 Operating Modes - Switch ON

The SHIPSURVEYOR is a dual conguration instrument. It operates as a Conned Space Monitor (CSM) or as a
Combustible Gas Indicator (CGI) depending on the button used to switch it on.
For Confined Space Monitor (CSM) operation, the instrument must be switched on by pressing and holding the
Bottom Button for approximately one second in fresh air.
For Combustible Gas Indicator (CGI) operation, the instrument must be switched on by pressing and holding
the Top Button for approximately one second in fresh air.
3-3
Page 30
USER HANDBOOK
The title screen is displayed, as illustrated in Fig. 3-4, the pump is activated and the instrument begins its warm-up routine that lasts approximately 60 seconds. During the warm-up cycle, a countdown timer appears in the top right hand corner of the display.
The display backlight illuminates and remains ON during warm-up.
Fig. 3-4 Title Screen

3.5 Language Option

The language setup has three congurable options:
• English (Default)
• Other (single) language
• Choice of language
During the warm-up cycle, the language screen, as illustrated in Fig. 3-5, is only displayed if ‘choice of language’ is
congured. This allows the user to select a pre-programmed
language from the menu. The previously selected language is highlighted in the display.
3-4
Fig. 3-5 Language Selection
Page 31
OPERATION
Note: If choice of language is not congured, the warm- up cycle will automatically continue with the
congured language.
To select the required language option
• Single press the Top Button , or single press
Bottom Button , to highlight required language.
• Press and hold OK (the Bottom Button) to
conrm selection.
3.6 Instrument Identication
During the warm-up cycle, the instrument display identies
the model, serial number, software version, datalogging, battery status and calibration gas information as illustrated in Fig. 3-6:
Fig. 3-6 Instrument Identication
Note: The battery symbol is indicated during the warm­ up cycle, then on the bottom RH corner of the display during normal operation.
3-5
Page 32
USER HANDBOOK

3.7 Time / Date

The current time and date is displayed during the warm-up cycle, as illustrated in Fig. 3-7.
Fig. 3-7 Time and Date

3.8 Calibration Due Date

The CAL DUE period can be set from 1 to 400 days and is set to 365 days by default.
The CAL DUE date is automatically reset when the instrument is successfully calibrated.
The CAL DUE feature has ve (5) congurable options:
1. CAL DUE DATE MESSAGE IS NOT DISPLAYED (OPTION)
2. CAL DUE DATE MESSAGE IS DISPLAYED (OPTION)
Cal Due date message is displayed, as illustrated in
Fig. 3-8.
3-6
Page 33
OPERATION
Fig. 3-8 Calibration Due Date
If the Cal Due date has expired, i.e. overdue, the
screen illustrated in Fig. 3-9 is displayed.
Fig. 3-9 Calibration Expired
After approximately ve (5) seconds, the
instrument warm-up continues.
3. CAL DUE DATE MESSAGE IS DISPLAYED WITH USER ACKNOWLEDGE IF EXPIRED (DEFAULT)
Cal Due date message is displayed, as illustrated
in Fig. 3-10.
3-7
Page 34
USER HANDBOOK
Fig. 3-10 Calibration Due Date
If the Cal Due date has expired, i.e. overdue, the
screen illustrated in Fig. 3-11 is displayed.
Fig. 3-11 Calibration Overdue
The user must acknowledge that Cal Due date has
expired.
To accept expiry (continue):
• Press YES (Top Button) to continue the warm-up cycle.
To acknowledge expiry:
• Press NO (Bottom Button) to abort.
The screen, illustrated in Fig. 3-12 is
displayed.
3-8
Page 35
OPERATION
Fig. 3-12 Switch OFF
• Press and hold the Top Button to switch the instrument OFF.
4. CAL DUE DATE MESSAGE IS DISPLAYED
- USER ACKNOWLEDGE FOR EXTENDED PERIOD IF EXPIRED (OPTION)
Cal Due date message is displayed, as illustrated in
Fig. 3-13.
Fig. 3-13 Calibration Due Date
If the Cal Due date has expired, i.e. overdue, but
within the ‘extended period’, the screen illustrated in Fig. 3-14 is displayed.
3-9
Page 36
USER HANDBOOK
Fig. 3-14 Calibration Overdue
The user must acknowledge that Cal Due date has
expired.
Note: The extended period can be set from 1 to 31
days.
To accept extended expiry (continue):
Press YES (Top Button) to continue the warm-up cycle.
Note: When the extended period option expires, the
user will be forced to switch the instrument OFF.
To reject the extended expiry:
Press NO (Bottom Button) .
The screen, illustrated in Fig. 3-15, is displayed.
3-10
Fig. 3.15 Calibration Expired
Page 37
OPERATION
To proceed with the shut-down sequence:
Press and hold the Top Button to switch the instrument OFF.
5. CAL DUE DATE MESSAGE IS DISPLAYED ­USER SHUT-DOWN IF EXPIRED (OPTION)
Cal Due date message is displayed as illustrated in
Fig. 3-16.
Fig. 3-16 Calibration Due Date
If Cal Due date has expired, the screen illustrated in
Fig. 3-17 is displayed.
Fig. 3.17 Calibration Expired
To proceed with the shut-down sequence:
Press and hold the Top Button to switch the instrument OFF.
3-11
Page 38
USER HANDBOOK

3.9 Service Due Date

The Service due date can be set by the workshop and is set to two (2) years by default from last service date. The date can be set over a period of 1 to 36 months in 1 month steps.
Note : The service due date, if enabled, will only
be displayed if less than 90 days from the current date.
The SERVICE DUE feature has ve (5) congurable options:
1. NOT USED (DEFAULT)
Service Due date message is not displayed during
warm-up (Default).
2. SERVICE DUE DATE MESSAGE IS DISPLAYED (OPTION)
Service Due date message is displayed, as illustrated
in Fig. 3-18.
3-12
Fig. 3-18 Service Due Date
Page 39
OPERATION
If the Service Due date has expired, i.e. overdue, the
screen illustrated in Fig. 3-19 is displayed.
Fig. 3-19 Service Expired
After approximately ve (5) seconds, the
instrument warm-up continues.
3. SERVICE DUE DATE MESSAGE IS DISPLAYED ­USER ACKNOWLEDGE IF EXPIRED (OPTION)
Service Due date message is displayed, as
illustrated in Fig. 3-20.
Fig. 3-20 Service Due Date
If the Service Due date has expired, i.e. overdue,
the screen illustrated in Fig. 3-21 is displayed.
3-13
Page 40
USER HANDBOOK
Fig. 3-21 Service Overdue
The user must acknowledge that Service Due date
has expired.
To accept expiry (continue):
• Press YES (Top Button) to continue the warm-up cycle.
To acknowledge expiry:
• Press NO (Bottom Button) to abort.
The screen illustrated in Fig. 3-22 is
displayed.
Fig. 3-22 Switch OFF
• Press an hold the Top Button to switch the
instrument OFF.
3-14
Page 41
OPERATION
4. SERVICE DUE DATE MESSAGE IS DISPLAYED
- USER ACKNOWLEDGE FOR EXTENDED PERIOD IF EXPIRED (OPTION)
Service Due date message is displayed, as
illustrated in Fig. 3-23.
Fig. 3-23 Service Due Date
If the Service Due date has expired, i.e. overdue, but
within the ‘extended period’, the screen illustrated in Fig. 3-24 is displayed.
Fig. 3-24 Service Overdue
The user must acknowledge that Service Due date
has expired.
Note: The extended period can be set from 1 to 31
days.
3-15
Page 42
USER HANDBOOK
To accept extended expiry (continue):
• Press YES (Top Button) to continue the warm-up cycle.
Note: When the extended period option expires,
the user will be forced to switch the instrument OFF.
To reject the extended expiry:
• Press NO (Bottom Button) .
The screen, illustrated in Fig. 3-25, is displayed.
Fig. 3.25 Service Expired
To proceed with the shut-down sequence:
Press and hold the Top Button to switch the instrument OFF.
3-16
Page 43
OPERATION
5. SERVICE DUE DATE MESSAGE IS DISPLAYED ­USER SHUT-DOWN IF EXPIRED (OPTION)
Service Due date message is displayed as illustrated
in Fig. 3-26.
Fig. 3-26 Service Due Date
If Service Due date has expired, the screen illustrated
in Fig. 3-27 is displayed.
Fig. 3.27 Calibration Expired
To proceed with the shut-down sequence:
Press and hold the Top Button to switch the instrument OFF.
3-17
Page 44
USER HANDBOOK

3.10 Sensor Zeroing

The nal feature in the instrument warm-up cycle is sensor
zeroing.
During this check, the screen illustrated in Fig. 3-28 is displayed.
Fig. 3-28 Sensor Check Display
When all sensors have been correctly zeroed, the instrument is ready for use.

3.10.1 Instrument Will Not Zero - Zero Fault

If a sensor fails to zero because of a fault, the screen illustrated in Fig. 3-29 is displayed, providing the user with the option to continue working and use the remaining operational sensor(s). (Details of the failed sensor will be datalogged).
3-18
Fig. 3-29 Failed Sensor Display
Page 45
OPERATION
To accept sensor fault and continue operation:
• Press and hold YES (Top Button) to continue using operational sensor(s).
Note: A spanner symbol ashes in the faulty range
(LEL in example), as illustrated in Fig. 3-30.
alternates with
Fig. 3-30 Faulty Range Display
To reject sensor fault and proceed with the shut-down
sequence:
• Press and hold NO (Bottom Button) to switch the instrument OFF.
3-19
Page 46
USER HANDBOOK

3.10.2 Instrument Will Not Zero - Gas Present

If a sensor fails to zero because of gas in the atmosphere,
the gas reading ashes as illustrated in Fig. 3-31.
alternates with
Fig. 3-31 Gas Present During Warm-up
To correct this fault, switch the instrument OFF then ON again in fresh air.
3-20
Page 47
OPERATION

3.11 Alarm Warning (CGI mode only)

When the instrument has been switched ON in CGI mode
and if congured, a ‘WARNING’ ashes in the display as
illustrated in Fig. 3-32, to alert the user that alarms are disabled in this mode.
alternates with
Fig. 3-32 Alarm Disabled Warning
To continue operation:
• Press and hold ‘YES’ (Top Button) for one second.
To abort instrument warm-up:
• Press and hold ‘NO’ (Bottom Button) for one
second. The instrument automatically switches OFF.
3-21
Page 48
USER HANDBOOK

3.12 Warm-up Complete

The instrument will now display the current gas readings for the selected mode. Fig. 3-33 illustrates the display for a Shipsurveyor 1 in CGI mode.
Fig. 3-33 LEL Display

3.13 Alarms (CSM mode only)

If alarms are enabled, an ‘A’ character is displayed in the top RH corner of the screen as illustrated in Fig. 3-34.
Fig. 3-34 Alarm Enabled Identier
Alarm levels are set as per the configuration of the instrument.
3-22
Page 49
OPERATION

3.14 LEL / Vol Gas Range Selection (CGI mode only)

When the warm-up cycle is complete, the instrument automatically selects the LEL range as default.
To select Vol Gas range:
• Single press ‘RANGE’ (Top Button) .

3.15 Switch Pump OFF / ON (CGI mode only)

When the warm-up cycle is complete, the instrument pump is automatically set to ON.
To switch the pump OFF / ON:
• Press and hold ‘PUMP’ (Bottom Button) for one second.

3.16 Zero All Ranges (CGI mode only)

To zero all ranges, e.g. If the instrument displays a gas reading when operating in fresh air:
To zero all ranges in fresh air:
Press and hold the Bottom Button for
approximately ve (5) seconds The following screen
is displayed:
Fig. 3-35 Zero All Ranges
3-23
Page 50
USER HANDBOOK
• Press and hold ‘YES’ (Bottom Button) for one
second to zero all ranges and return to live readings.
Alternatively,
• Press and hold ‘NO’ (Top Button) for one
second to return to live readings without zeroing.

3.17 Datalogging

The SHIPSURVEYOR can store over 1000 logged entries in date order. In addition, it can store up to 300 location entries. Location (LOC) entries allows logged data to be associated with particular areas of the vessel.

3.17.1 CSM Mode

The instrument, by default, has automatic and manual datalogging enabled when operating in CSM mode.
During automatic datalogging the instrument stores the values of all gas ranges together with the current time and date at 60 second intervals. This interval can be set between
1 second and 10 minutes.
Manual datalogging allows the user to store a snapshot reading at any time.
To manually store a reading:
• Press and hold ‘STORE’ (Top Button) for one second.
3-24
Page 51
OPERATION

3.17.2 CGI Mode

When the instrument is operating in CGI mode, only manual datalogging is available:
To manually store a reading:
• Press and hold ‘STORE’ (Top Button) for one second.

3.18 Max / Min Values

The SHIPSURVEYOR instrument records the maximum and minimum gas values for each sensor, since switch-on.
Note: Only the Oxygen (O2) sensor will display a minimum value. All other sensors will indicate zero.
To view max values:
• Start from the normal operating display, illustrated in Fig. 3.36.
Fig. 3-36 Normal Operating Display
3-25
Page 52
USER HANDBOOK
• Press ‘MAX’ (Bottom Button) to display
each recorded maximum gas reading since switch­on. ‘MAX’ alternates with the range name as illustrated in Fig. 3-37.
alternates with
Fig. 3-37 Max Displayed Values
Note: The display will automatically return to live readings if no buttons are pressed within ten (10) seconds. The stored values are retained in the instrument memory unless they have been reset
by a press and hold of ‘CLEAR’ (Top Button) .
To view min values:
Note: ‘MIN’ button press option is only displayed in instruments with an Oxygen (O2) sensor tted.
3-26
Page 53
OPERATION
• Press ‘MIN’ (Bottom Button) to display the
recorded minimum oxygen reading. ‘MIN’ alternates with the range name as illustrated in Fig. 3-38.
alternates with
Fig. 3-38 Min Displayed Values
Note: The display will automatically return to live readings if
no buttons are pressed within ten (10) seconds. The stored value is retained in the instrument memory unless it has been reset by a press and hold of
‘CLEAR’ (Top Button) .
To return to live readings:
• Press and hold ‘LIVE’ (Bottom Button) for one second to return to live readings.
3-27
Page 54
USER HANDBOOK

3.19 Location (LOC) Selector

The SHIPSURVEYOR instrument features a ‘LOC’ location selector allowing the user to allocate datalogged readings to a location (LOC) of a vessel.
For example, if ‘HOLD1’ is selected, all subsequent datalogging (manual and automatic) will be referenced to ‘HOLD1’.
300 Location (LOC) names are available and can be customised to meet individual user requirements.
Location (LOC) names can only be customised and stored in the instrument using the Shipsurveyor Data Downloading Software (GMI Part No. 48150).

3.19.1 To select a location (LOC):

The location list can be accessed by simultaneously pressing
the Top Button and Bottom Button for two (2) seconds.
1. Location data already stored in the instrument:
If location data has previously been stored in the instrument, it will now be displayed as illustrated in Fig. 3-39.
3-28
Fig. 3-39 ‘LOC’ Display
Page 55
OPERATION
To select next entry in ‘LOC’ list :
• Press Bottom Button .
To quickly scroll down:
• Press and hold Bottom Button .
To select previous entry in ‘LOC’ list:
• Press Top Button .
To quickly scroll up:
• Press and hold Top Button .
If a location is highlighted and the user exits the location (LOC) list, then that location will be associated with all subsequent datalogging.
To exit ‘LOC’ list:
• Press and hold Top Button and Bottom Button
simultaneously for three seconds.
The instrument will return to displaying the gas readings.
2. No Location data stored in the instrument:
If no location data has been stored in the instrument, the display will be as illustrated in Fig. 3-40.
The instrument will still datalog, but no location data will be associated with the stored readings.
3-29
Page 56
USER HANDBOOK
Fig. 3-40 ‘LOC’ Display

3.20 Acknowledge Alarms (CSM mode only)

The SHIPSURVEYOR default alarm function means that instantaneous alarms will be latching and both the audible and visual functions will operate. (Where an alarm level
is exceeded, the instrument LED’s ash and the sounder
pulses rapidly). Latching alarms must be cleared manually.
To clear a latching alarm:
• Press and hold the Bottom Button , for one
second, after the detected gas level has fallen below the alarm limit. Refer to Chapter 4.4.1.
The option for the alarms is non-latching. These clear automatically when the gas levels fall below the alarm limit. The audible alarm can be muted.
To mute a non-latching alarm:
• Press and hold the Bottom Button for one second. Refer to Chapter 4.4.3.
Note: An alarm remains muted until a new alarm
event occurs.
3-30
Page 57
OPERATION

3.21 Switch the instrument OFF

Before switching OFF from the normal operating mode, make sure that the instrument is in fresh air in order to purge any remaining gas from the instrument.
To initiate the shutdown sequence:
• Press and hold the Top Button .
The OFF sequence countdown begins and the user
must keep the buttons depressed for a further three (3) seconds to switch the instrument OFF.
The countdown sequence is illustrated in Fig. 3-41.
Fig. 3-41 Countdown Sequence to OFF
Note 1: The OFF sequence can be aborted by
releasing the Top Button, providing access to the previous operating display.
Note 2: The countdown sequence will commence at
‘9’ if an alarm is active when the shutdown sequence is initiated.
3-31
Page 58
USER HANDBOOK

3.22 Button Operation Summary

3-32
Page 59
SHIPSURVEYOR
ALARM CONDITION (4 RED) LED's
and CONFIDENCE SIGNAL in CSM mode (outer 2 RED LED's)
POWER ON
(GREEN) LED
FAULT LED
(YELLOW)
4

ALARMS

4.1 Gas Alarms

The SHIPSURVEYOR alarms, by default, are enabled
when operating in Conned Space Monitor (CSM) mode as indicated by an ‘A’ symbol in the top right corner of the display. Both audible and visual alarms are active. The audible alarm is rated up to 85 dB(A). The visual alarm consists of two (2) pairs of red LED’s in the display panel as illustrated in Fig. 4-1. When ashing, the red LED’s indicate that the instrument is in alarm condition.
Fig. 4-1 Instrument Display Panel
4-1
Page 60
USER HANDBOOK
Alarm levels are set at the time of instrument manufacture.
It is important that the user ensures that the levels are in
accordance with their company’s alarm levels and with health
and safety legislation. The alarm levels are user congurable.
All gas ranges can have alarm limits that trigger the alarm if the measured gas reading exceeds the set level. If a preset alarm level is exceeded, the audible alarm sounds, the LED’s ash, and the gas range name alternates with the alarm type.
Note: Alarms are disabled in Combustible Gas Indicator
(CGI) mode.

4.1.1 FLAMMABLE (LEL) ALARMS

Note: A congurable option allows the instrument to display this range type as LFL (Lower Flammable Limit).
Two (2) instantaneous alarm levels (Hi and HiHi) are congurable, each with different alarm indications, refer to Fig. 4-1. Both are rising alarms, i.e. if the detected gas concentration rises above the specic alarm level, the alarm is triggered. All alarms are user congurable to meet the needs of different companies.

4.1.2 OXYGEN (O2) ALARMS

Three (3) instantaneous alarm levels (HiHi, Lo and LoLo) are congurable, each with different alarm indications, refer to Fig. 4-1. These consist of one rising, and two falling alarms (necessary to trigger alarms in oxygen decient scenarios). All alarms are user congurable to meet the needs of different companies.
4-2
Page 61
ALARMS

4.1.3 TOXIC ALARMS (e.g. CO)

Two (2) instantaneous alarm levels (Hi and HiHi) are congurable, each with different alarm indications, refer to Fig. 4-1. Both are rising alarms, i.e. if the detected gas concentration rises above the specic alarm level, the alarm is triggered. All alarms are user congurable to meet the needs of different companies.
Additionally, two (2) Time Weighted Average (TWA) alarm levels (STEL and LTEL) are configurable with alarm indications, refer to Fig. 4-1.
Two (2) TWA alarms are congurable for each toxic range
within the instrument.
Note: A Time Weighted Average (TWA) value is the mean
average gas level over a specic period. The Short Term Exposure Limit (STEL) is 15 minutes and the Long Term Exposure Limit (LTEL) is 8 hours. The time weighted averages are averaged over a full 24 hour period whether the instrument is ON or OFF. Such
averaging essentially makes the instrument single
user applicable.
The option is available to restart the averaging after
each instrument switch-off, thus allowing for multiple
user application.
4-3
Page 62
USER HANDBOOK

4.2 Alarm Types

4.2.1 LATCHING / NON-LATCHING

Each alarm can be latching or non-latching.
Latching alarms can only be cleared by the user when the gas level returns to within the alarm limits. A latching alarm cannot be muted.
Non-latching alarms clear automatically when the gas level returns to within the alarm limits. A non-latching alarm can
be muted.

4.2.2 MUTING / ACKNOWLEDGING

Muting of an alarm is achieved by a press and hold of the Bottom Button. Muting means that the audible alarm is cancelled. The audible alarm is reactivated if a new alarm occurs.
Acknowledging is only applicable to latching alarms and allows audible / visual alarms to be cancelled after the gas readings have returned to safe concentrations.

4.3 Default Alarm Settings

Fig. 4-2, shows the GMI default selections. Latching or non­latching options exist in all allowable alarms.
If an alarm is disabled (Dis), it will not function.
4-4
Page 63

4.3.1 Default Alarms Table

Alarm
Type
Level Latch D isplay Audible Vis ual
(RED LED's)
LE L Hi - Dis Hi 1 beep
per second
4 - slo w
flas hing
LEL Hi Hi 20% Y Hi Hi C ontinuo us
warble
Alternate pairs
flas hing
O
2
HiHi 23% Y Hi Hi C ontinuo us
warble
Alternate pairs
flas hing
O
2
Lo - Dis Lo 1 beep
per second
4 - slo w
flas hing
O
2
LoLo 19.5% Y LoLo C ontinuo us
warble
Alternate pairs
flas hing
H2S Hi - Dis Hi 1 beep
per second
4 - slo w
flas hing
H2S HiHi 15ppm Y HiHi Cont inuous
warble
Alternate pairs
flas hing
H2S STEL 10ppm N STEL 1 beep
per second
4 - slo w
flas hing
H2S LTEL 5ppm N LTEL 1 beep
per second
4 - slo w
flas hing
CO Hi - Dis Hi 1 beep
per second
4 - slo w
flas hing
CO Hi Hi 300ppm Y Hi Hi C ontinuo us
warble
Alternate pairs
flas hing
CO STEL 200ppm N STEL 1 beep
per second
4 - slo w
flas hing
CO LTE L 35ppm N LTEL 1 beep
per second
4 - slo w
flas hing
CO2 Hi - Dis Hi 1 beep
per second
4 - slo w
flas hing
CO2 HiHi 2.5% Y Hi Hi C ontinuo us
warble
Alternate pairs
flas hing
CO2 STEL 1.5% N STEL 1 beep
per second
4 - slo w
flas hing
CO2 LTEL 0.5% N LTE L 1 beep
per second
4 - slo w
flas hing
ALARMS
Fig. 4-2 Alarm Indication
4-5
Page 64
USER HANDBOOK

4.4 Alarm Examples

The following examples, in CSM mode, illustrate the instrument’s display in various alarm conditions.
Examples are based on Default Alarm settings as detailed in Fig. 4-2.

4.4.1 EXAMPLE 1 (LEL HiHi ALARM)

Fig. 4-3, illustrates a Shipsurveyor 1 instrument following a ‘HiHi’ alarm condition. The audible alarm emits a continuous warble and the red LED’s ash in alternate pairs.
alternates with
Fig. 4-3 LEL HiHi Alarm
The alarm is latching, and therefore, can only be acknowledged (cleared) by the user when the gas level returns below the 20% LEL alarm limit.
4-6
Page 65
ALARMS
To clear the latching alarm:
• Press and hold ‘ACK’ (Bottom Button) for one
second.

4.4.2 EXAMPLE 2 (O2 LoLo ALARM)

This example, Fig. 4-4, illustrates a Shipsurveyor 2 instrument following a ‘LoLo’ alarm condition. The audible alarm emits a continuous warble and the red LED’s ash in
alternate pairs.
alternates with
Fig. 4-4 O2 LoLo Alarm
The alarm is latching, and therefore, can only be cleared by the user when the oxygen level rises above the 19.5%
O2 alarm limit.
4-7
Page 66
USER HANDBOOK
To clear the latching alarm:
• Press and hold ‘ACK’ (Bottom Button) for one second.

4.4.3 EXAMPLE 3 (CO STEL ALARM)

This example, Fig. 4-5, illustrates a Shipsurveyor 5 instrument following an ‘STEL’ alarm condition. The audible alarm beeps every second and all four (4) red LED’s ash
slowly.
alternates with
Fig. 4-5 CO STEL Alarm
The alarm is non-latching and therefore can be muted by the user, and is reset automatically when the STEL level returns below the 200ppm CO alarm limit.
4-8
Page 67
ALARMS
SHIPSURVEYOR
4 RED LED's - USED TO INDICATE FAULTS
POWER ON
(GREEN) LED
FAULT LED
(YELLOW)
To mute the non-latching alarm:
• Press and hold ‘MUTE’ (Bottom Button) for one
second.

4.5 Fault Alarms

The SHIPSURVEYOR instrument features a series of fault
alarms to alert the user that the instrument is not functioning correctly. The audible and visual indications used are both fault / type and mode (CSM or CGI) dependent.
Visual alarms are indicated by a series of LED’s, as illustrated in Fig. 4-6. The power ON LED will remain illuminated during all fault alarm indications.
Fig. 4-6 Instrument LED Panel
4-9
Page 68
USER HANDBOOK
Alarm Type
(CSM Mode)
Mute Display Audible V isual
(LED )
LOW
BATTE RY
No LO<>BAT
Co ntinuo us
To ne
Yellow &
4 Red
BATTE RY
EXHAUSTED
No BAT<>B AT
Co ntinuo us
To ne
Yellow &
4 Red
ZERO FAULT
Ye s
Fla shi ng Spanner
Co ntinuo us
To ne
Yellow &
4 Red
SENSOR
FAULT
No
Follo wing
Ranges Failed
Co ntinuo us
To ne
Yellow &
4 Red
FLOW FAULT
No 'F LOW FA ULT'
Co ntinuo us
To ne
Yellow &
4 Red
IR SUPPLY
FAULT
No ' P '
Co ntinuo us
To ne
Yellow &
4 Red
COMMS.
FAULT
No ' I '
Co ntinuo us
To ne
Yellow &
4 Red
MEMORY
FAULT
No ' C '
Co ntinuo us
To ne
Yellow

4.5.1 Fault Alarms - CSM Mode

The following table, Fig. 4-7, illustrates the fault alarms in
CSM mode. Muting options exist in selected alarms.
4-10
Fig. 4-7 Fault Alarms in CSM Mode
Page 69
ALARMS
Alarm Type (CGI Mo de)
Mute Display Audible Vis ual
(LED )
LOW
BATTE RY
No LO<>BAT No Yellow
BATTE RY
EXHAUSTED
No BAT<>B AT No Yellow
ZERO FAULT
N/A
Fla shi ng Spanner
No Yellow
SENSOR
FAULT
N/A
Follo wing
Ranges Failed
No Yellow
FLOW FAULT
N/A 'F LOW FAULT' No Yellow
IR SUPPLY
FAULT
No ' P '
Co ntinuo us
To ne
Yellow &
4 Red
COMMS.
FAULT
No ' I '
Co ntinuo us
To ne
Yellow &
4 Red
MEMORY
FAULT
No ' C '
Co ntinuo us
To ne
Yellow

4.5.2 Fault Alarms - CGI Mode

The following table, Fig. 4-8, illustrates the fault alarms in
CGI mode. Muting options exist in selected alarms.
Fig. 4-8 CGI Mode Fault Alarms
4-11
Page 70
USER HANDBOOK

4.5.3 ZERO FAULT

If a zero fault occurs in CSM mode, the instrument’s audible alarm will activate. The visual alarm (yellow and four red LED’s) will also activate.
If a zero fault occurs in CGI mode, only the visual alarm (yellow LED) will activate.
In both modes, a spanner symbol ashes in the display adjacent to the corresponding gas type (H2S), as illustrated in Shipsurveyor 3 example, Fig. 4-9.
alternates with
Fig. 4-9 Zero Fault
To correct the zero fault, switch the instrument OFF then ON again in fresh air. If this does not correct the fault, return the instrument for service.
4-12
Page 71
ALARMS

4.5.4 SENSOR FAULT

If a sensor fault occurs in CSM mode, the instrument’s audible alarm will activate. The visual alarm (yellow and four red LED’s) will also activate.
If a sensor fault occurs in CGI mode, only the visual alarm (yellow LED) will activate.
In both modes, a warning display is followed by a display
showing a spanner symbol adjacent to the faulty range, as illustrated in Shipsurveyor 3 example, Fig. 4-10.
Fig. 4-10 H2S Sensor Fault
The instrument should be returned for service.
4-13
Page 72
USER HANDBOOK

4.5.5 FLOW FAULT

A ow fault will occur if the ow path is restricted.
If a ow fault occurs in CSM mode, the instrument’s audible alarm will activate. The visual alarm (yellow and four red LED’s) will also activate.
If a ow fault occurs in CGI mode, only the visual alarm (yellow LED) will activate and the instrument’s pump will
stop running.
In both modes, the FLOW FAULT display will alternate between the two screens, illustrated in Fig. 4-11.
alternates with
Fig. 4-11 Flow Fault
4-14
Page 73
ALARMS
Follow the instructions as displayed, i.e.
- Check the ow path
- Remove the blockage
- Acknowledge the alarm
To acknowledge ow fault alarm:
• Press and hold ‘ACK’ (Bottom Button) for one
second.
4.5.6 MEMORY FAULT (Calibration /
Conguration)
During warm-up, if the ‘Calibration Required’ screen is displayed as illustrated in Fig. 4-12, the instrument has detected a calibration / conguration memory fault and is unable to continue without re-calibration and / or conguration set-up.
Fig. 4-12 Calibration / Conguration Memory Fault
The instrument must be returned for service.
4-15
Page 74
USER HANDBOOK

4.5.7 MEMORY FAULT (General Instrument)

At any time during operation, if a ‘C’ character is ashing in the top RH corner of the display as illustrated in Fig. 4-13, the instrument has detected a memory fault and is unable to continue. The fault activates the audible alarm and the (yellow) fault LED is illuminated.
Fig. 4-13 General Instrument Memory Fault
The instrument must be returned for service.

4.5.8 LOW BATTERY

During operation, ‘LO’ alternating with ‘BAT’ is displayed, as illustrated in Fig. 4-14, when typically 30 minutes operating
time remains at normal temperature and depending on battery type.
4-16
Fig. 4-14 LO-BAT
Page 75
ALARMS
In CSM mode, the instrument audible alarm will activate. The visual alarm (yellow and four red LED’s) will also activate.
In CGI mode, only the visual alarm (yellow LED) will activate.

4.5.9 BATTERY EXHAUSTED

When ‘BAT’ ashes in the display, as shown in Fig. 4-15, the
instrument batteries are almost exhausted.
Fig. 4-15 BAT-BAT
In CSM mode, the instrument audible alarm will activate. The visual alarm (yellow and four red LED’s) will also activate.
In CGI mode, only the visual alarm (yellow LED) will activate.
In both modes, the instrument automatically switches OFF.
The instrument batteries must be replaced. (Refer to Chapter 5, ‘Operator Maintenance’).
4-17
Page 76
USER HANDBOOK
4-18
Page 77
5

OPERATOR MAINTENANCE

5.1 Batteries

The SHIPSURVEYOR contains two alkaline or rechargeable batteries that provides the power required to operate the instrument.
The alkaline batteries should be replaced, or the rechargeable batteries recharged in the following situations:
• The‘LO-BAT’agisashinginthedisplay.
Indicates that the remaining operating time of the
instrument is low.
• The‘BAT’agisashinginthedisplay.
Indicates exhausted batteries and automatic switch-
off.
The procedure to remove / discard alkaline cells or to remove rechargeable cells for charging is detailed in paragraph
5.1.1.Re-ttingofthecellsisalsoincludedinthisparagraph.
The rechargeable cells should be charged using a
commercialtypecharger.
5-1
Page 78
USER HANDBOOK

5.1.1 Replace Alkaline / Rechargeable Cells

WARNING 1: To prevent ignition of flammable or
combustible atmospheres, remove batteries before servicing.
WARNING 2: To prevent ignition of flammable
or combustible atmospheres, read, understand and adhere to the manufacturer’s live maintenance procedures.
WARNING 3: To reduce the risk of ignition of a
flammable or explosive atmosphere, batteries must be changed only in a location known to be non-hazardous (safe area).
WARNING 4: To reduce the risk of explosion,
do not mix old batteries with new batteries or mix batteries from different manufacturers.
WARNING 5: Never attempt to recharge non
rechargeable cells.
5-2
Page 79
OPERATOR MAINTENANCE
The following procedure shouldonlybecarriedoutin a
safe area:
1) Using the Hex Driver, loosen the two instrument base
screwsthenremovethebatterycover.
2) Removebatterycover.
5-3
Page 80
USER HANDBOOK
3) Remove the alkaline / rechargeable batteries.
4) Charge rechargeable batteries using a commercial
type charger.Charging duration is dependent on chargertypeandconditionofrechargeablebatteries.
5) Check battery compartmentfordamagetospring
contacts or corrosion on springs.
6) Insert new alkaline / fully charged rechargeable batteries, observing correct polarity indicationin batterycompartmentbase.
7) Replacebatterycoverandfastenbothbasescrews.
8) Check that the instrument switches on and works to
specication.
5-4
Page 81

5.2 Filter Replacement

HYDROPHOBIC FILTER
PROBE HANDLE ASSEMBLY
COTTON PARTICULATE FILTER

5.2.1 Internal Filter

The SHIPSURVEYORinstrumenthasanintegralhydrophobic
ltertted.Toreplacethislter,refertoMaintenanceManual (GMIPartNo.48166).

5.2.2 Probe Handle (Accessory) - Part No. 48120

Hydrophobicandcottonparticulateltersintheprobehandle
minimise the danger of water and dust ingress.
Toreplacethelter(s),proceedasfollows:
OPERATOR MAINTENANCE
1) Unscrewtheprobehandleassembly.
2) Removethecottonparticulatelteranddiscard.
3) Removethehydrophobiclter.
4) Clean the probe handle to make sure that it is free from dirt and water.
5) Fitanewcottonparticulatelter.
6) Fit the hydrophobic lter. The yellowlabelonthe filter fits againsttheyellowlabelon the probe
handle.
5-5
Page 82
USER HANDBOOK
7) Reassembletheprobehandleassembly.
5-6
Page 83
6

CALIBRATION

The instrument has been calibrated for a particular ammable
gas mixture. Where any doubt exists, the instrument should be returned to GMI or an authorised distributor for calibration.
Three methods of calibration are possible:
• Field Calibration.
• Manual or Automatic Calibration using exiCal Plus software CD-ROM (GMI Part No. 99553). CD-ROM
includes user instructions.
• The GMI Instrument Management System (IMS) provides all the facilities of Automatic Calibration with the added feature of instrument database management.
Note: The calibration systems above (hardware and
software) are manufactured by GMI. For more details
contact GMI or an authorised distributor.
6-1
Page 84
USER HANDBOOK

6.1 Calibration Validity

Calibration validity is the responsibility of the user. Under normal operating conditions a 12 month period can be expected. This is no guarantee however, as the precise application of the product is unknown to GMI. Individual codes of practice may dictate shorter periods.
Regular checking establishes a pattern of reliability and
enables the calibration check period to be modied in line
with operational experience. The higher the risk, the more frequently calibration should be checked.
6-2
Page 85
7

ACCESSORIES

Instrument complete with (c/w) Hard Case & Accessories:
48021 Shipsurveyor 1 (LEL ; Vol Gas) 48022 Shipsurveyor 2 (LEL ; Vol Gas ; O2) 48023 Shipsurveyor 3 (LEL ; Vol Gas ; O2 ; H2S) 48024 Shipsurveyor 4 (LEL ; Vol Gas ; O2 ; CO2) 48025 Shipsurveyor 5 (LEL ; Vol Gas ; O2 ; CO2 ; CO) 48026 Shipsurveyor 6 (LEL ; Vol Gas ; O2 ; H2S ; CO) 48027 Shipsurveyor 7 (O2 ; CO2) 48028 Shipsurveyor 8 (LEL ; Vol Gas ; O2 ; H2S ; CO2)
7-1
Page 86
USER HANDBOOK
Note: All Shipsurveyor instruments are supplied in a hard Carry Case c/w: Neck / Shoulder Harness; 4mm Hex Driver; Probe Handle c/w 1m (39ins.) Conductive Tubing; 15cm (6ins.) Open End Probe; Box (10) Cotton Filters x 2; User Handbook; Quick Operating Instructions; Calibration
Certicate; Conguration Report; Datalogging & exiCal Plus (Calibration) Software & IrDA Adaptor.
Accessories available for SHIPSURVEYOR:
Part Number Description
48100 Hard Carry Case 48101 Leather Field Case 12370 Neck / Shoulder Harness 12371 Waist Harness 48120 Probe Handle c/w 1m (39in.) Conductive
Tubing 12358 Hydrophobic Filter 10077 Cotton Filters (Box of 10) 12688 Sample Line Adaptor 12712 Clear Sample Line - per metre 48102 Ball Float 48142 Sample Line Adaptor c/w Waist
Harness, Quick Connect, and Conductive Tubing x 10m (32ft.)
48145 Sample Line Adaptor c/w Waist Harness, Quick Connect, and Conductive Tubing x 25m (80ft.)
48147 Sample Line Adaptor c/w Waist Harness, Quick Connect, and Conductive Tubing x 35m (114ft.)
13427 35cm (14in.) Open End Probe 13413 35cm (14in.) S/Steel Open End Probe 12287 50cm (20in.) S/Steel Open End Probe
7-2
Page 87
ACCESSORIES
48110 Rechargeable Option Kit
(Charger, Universal Power Supply,
2 Rechargeable Batteries)
47280 Rechargeable Batteries x 2
48150 Data Downloading Package
99553 exiCal Plus Calibration Software
CD-ROM
48151 exiCal Plus Calibration Package
48160 Shipsurveyor User Handbook
48164 Shipsurveyor Conguration &. Field
Calibration User Handbook
48166 Shipsurveyor Maintenance
Manual
Note: For other sampling probes and accessories, contact
GMI Ltd.
7-3
Page 88
USER HANDBOOK
7-4
Page 89
8

ADDITIONAL INFORMATION

8.1 Training

Training courses are available on all GMI products. Contact GMI Marketing Department for further details:
Tel: +44 (0) 141 812 3211
Fax: +44 (0) 141 812 7820
e-mail: sales@gmiuk.com

8.2 World Wide Web

Visit GMI web site at www.gmiuk.com
8-1
Page 90
USER HANDBOOK
8-2
Page 91
A
TYPICAL OPERATING
PARAMETERS
Typical operating parameters for the SHIPSURVEYOR
instrument are as follows:
Gas Range Resolution Accuracy
Range
LEL 0 to 9.9% 0.1% *
10 to 100% 1%
VOL 0 to 9.9% 0.1%
10 to 100% 1%
O2 0 to 20.9% 0.1%
21 to 25% 1%
H2S 0 to 100ppm 1ppm
CO2 0 to 5% 0.1%
5 to 20% 0.5%
CO 0 to 1000ppm 1ppm
* Congurable option, see page 2-2.
Notes:
All the values above are at normal temperature and pressure.
Humidity is between 0% and 100% RH (non-condensing).
Pressure changes at the inlet and exhaust should be minimised as they
may cause transient changes in reading.
+
2% / + 5% of reading
+
1% / + 5% of reading
+
0.1% / + 5% of reading
+
5% of reading
+
1ppm / + 5% of reading
+
0.1% / + 5% of reading
+
5% of reading
+
5ppm / + 5% of reading
A-1
Page 92
USER HANDBOOK

Dimensions

180mm (7.1in.) x 95mm (3.7in.) x 100mm (3.9in.)

Weight

1.35kg (3.0lb.) with two alkaline batteries

Operating Temperature

-20 oC to 50 oC (-4 oF to 122 oF)

Humidity

0 – 100% RH, non-condensing.

Construction

Moulded polypropylene case protected to IP54

Display

Dot Matrix LCD (128 x 64)

Warm-Up / Stabilization Time

~ 60 seconds

Typical Flow Rate Information

Nominal pump ow rate is 0.5 to 0.7 litres per minute. Typical ow fail rate is 0.1 to 0.2 litres per minute.

Response Time (T90)

Typical O
response time (0.0 to 18.8% O
2
) > 15 seconds.
2

Power Source

Two ‘D’ size alkaline cells typically providing in excess of 20 hours runtime at 20oC (68oF).
or
Two ‘D’ size rechargeable cells typically providing in excess of 8 hours runtime at 20oC (68oF).
A-2
Page 93
B

QUICK OPERATING INSTRUCTIONS

The following multi-language instructions provide the user with a quick guide to the operation of the . . .
SHIPSURVEYOR instrument.
Each language and page reference is as follows:
English - pages B-2 to B-6
B-1
Page 94
USER HANDBOOK

CHECKLIST

1. Check the instrument has no obvious faults.
2. Read and understand handbook before use.
3. Switch ON
4. Check battery levels.
5. Check “ZERO” in fresh air.
6. Check that the sample line and any accessories are in good condition and leak free. Make sure that the instrument displays
a ow fault when the instrument inlet / sample line is blocked. If fault is not displayed, check tightness of all ttings.

SAFETY

• The instrument must be regularly serviced and calibrated by fully trained personnel in a safe area.
Batteries: Alkaline or *Rechargeable batteries must be exchanged
(*and recharged) in a safe area and tted correctly before use.
Never use damaged batteries or expose to extreme heat.
See Chapter 5: OPERATOR MAINTENANCE.
• Only GMI replacement parts should be used.
If the instrument detects gas, follow your own organisation’s
procedures and operational guidelines.
• The combustion chamber is a ameproof assembly and must not be opened in the presence of a ammable atmosphere.
• Shipsurveyor instruments are certied as:
ATEX II 2 G Exd ia IIB T3 Gb (-20oC < Tamb < 50oC).
IEC Ex d ia IIB T3 (-20oC < Ta < +50oC)
• This equipment conforms to standard EN 50104.
• This equipment is also designed and manufactured to comply with
MED Directive 0038/YY (Module B&E). Further detail of the
Marine Equipment Directive, is located on the MED declaration of conformity, supplied with the instrument.
Any right of claim relating to product liability or consequential damage to any third party against GMI is removed if the warnings are not observed.
B-2
Page 95
QUICK OPERATING INSTRUCTIONS
WARNING: To prevent ignition of ammable or combustible atmospheres,
remove batteries before servicing.
WARNING: To prevent ignition of ammable or combustible atmospheres,
read, understand and adhere to the manufacturer’s live maintenance procedures.
WARNING: To reduce the risk of ignition of a ammable or explosive
atmosphere, batteries must be changed only in a location known to be non-hazardous.
WARNING: To reduce the risk of explosion, do not mix old batteries with used batteries or mix batteries from different manufacturers.
WARNING: Never attempt to recharge non rechargeable cells.
CAUTION: Not for use in oxygen enriched atmospheres.
CAUTION: Replace batteries only with approved batteries, as follows:
ATEX Approved Instruments: Alkaline (LR20 ‘D’ size): Duracell Procell; Duracell Industrial;
Duracell Plus; Energizer Ultra; Energizer Industrial.
Rechargeable (LR20 ‘D’ size): Uniross 2600mAh NiMH.

AREAS OF USE

Exposure to certain chemicals can result in a loss of sensitivity of the
ammable sensor (not applicable to infrared ammable sensor). Where such
environments are known or suspected it is recommended that more frequent response checks are carried out. The chemical compounds that can cause loss of sensitivity include Silicones, Lead, Halogens and Sulphur. Do not use instrument in potentially hazardous atmospheres containing greater than 21% Oxygen. The enclosure material is carbon loaded polypropylene and must not be exposed to environments which are liable to result in mechanical or thermal degradation or to damage caused by contact with aggressive substances. Additional protection may be required in environments where
the instrument enclosure is liable to damage.
English
B-3
Page 96
USER HANDBOOK

OPERATION

Switch ON

Press and hold either, Bottom Button
(CSM mode), or Top Button (CGI mode),
for one second, to switch the instrument and pump ON in fresh air.
The instrument now begins its 60 second warm-up cycle, during which, a countdown timer appears in the top right hand corner of the display. The display backlight illuminates and remains ON until the warm-up cycle
is complete and then automatically switches OFF. During the warm-up cycle, if congured, the instrument displays language
option then model, serial number, software version, datalogging, battery status and calgas information as shown below.
The battery capacity level is displayed for approximately ve seconds
during the warm-up cycle, then on the bottom RH corner of the display during normal operation.
Next, the current time and date is displayed. This is followed by the Calibration Due Date then the
Service Due Date (if congured).
Each feature has ve congurable options:
1. Cal Due / Service Due message is not displayed.
2. Cal Due / Service Due date and overdue date messages are displayed.
3. Cal Due / Service Due message is displayed with user acknowledge
if expired.
4. Cal Due / Service Due message is displayed with user acknowledge for
extended period, if expired.
5. Cal Due / Service Due message is displayed with user shut-down, if
expired.
The next feature in the instrument warm-up cycle is sensor zeroing.
Finally, in CGI mode only and if congured, a ‘WARNING’ display alerts
the user that alarms are disabled in this mode and provides the option of continuing operation or automatically switching the instrument off .
B-4
Page 97
QUICK OPERATING INSTRUCTIONS
On completion of the warm-up cycle, the instrument will automatically select, as default, the LEL range and display the current gas detection reading, as shown opposite.

Range Selection (CGI only)

To change range LEL-VOL GAS-LEL, single press the ‘RANGE’ Top Btn .

Switch Pump OFF / ON (CGI only)

Press and hold ‘PUMP’ Bottom Button to switch the pump OFF / ON.

Zero All Ranges (CGI only) in Fresh Air

Press and hold Bottom Button for approximately ve (5) seconds.
Then, press and hold ‘YES’ (Top Button) for one second to zero all ranges
and return to live readings.

Manual Datalogging

To store a snapshot reading, press and hold ‘STORE’ (Top Button) .

Max / Min Values

To display max. value since switch-on, press ‘MAX’ (Bottom Button) .
To display min. value since switch-on, press ‘MIN’ (Bottom Button) .
Note: ‘MIN’ Only applicable to instruments with Oxygen (O2) sensor tted.
English
To erase Max / Min values, press and hold ‘CLEAR’ (Top Button) .
To return to live readings, press and hold ‘LIVE’ (Bottom Button) .
B-5
Page 98
USER HANDBOOK

Enter (LOC) Location List

Press and hold Top Button and Bottom Button simultaneously.
To select next entry in LOC list, press ‘DOWN’ (Bottom Button) .
To scroll through next entries in LOC list, press and hold ‘DOWN’
(Bottom Button)
To select previous entry in LOC list, press ‘UP’ (Top Button) .
To scroll through previous entries in LOC list, press and hold ‘UP’
(Top Button) .
To select location and exit LOC list, press and hold Top Button
and Bottom Button simultaneously. Note that ‘LOC’ names must
be pre-loaded.
.

Acknowledge Alarms (CSM only)

To clear a latching alarm, press and hold ‘ACK’ (Bottom Button) after
the detected gas level has fallen below the alarm limit.
To mute a non-latching alarm, press and hold ‘MUTE’ (Bottom Button) .

Switch OFF

Press and hold the Top Button . Keep button depressed while OFF
sequence countdown completes.
B-6
Page 99

INDEX

A
ACCESSORIES 7-1 Acknowledge Alarms
(CSM mode only) 3-30 Acknowledging 4-4 ADDITIONAL INFORMATION
8-1 Alarms 2-7 ALARMS 4-1 Alarms, Acknowledge 3-30 Alarms (CSM mode only) 3-22 Alarms, Default 4-5 Alarm Settings 4-4 Alarms, Fault 4-9 Alarms in CGI Mode 4-11 Alarms in CSM Mode 4-10 Alarm Warning (CGI mode only)
3-21 alkaline A-2 AREAS OF USE iii
B
Batteries 2-8 BATTERY EXHAUSTED 4-17 BATTERY, LOW 4-16 Battery Replacement
4-4, 4-6, 5-1 Before Use Checks 2-10 Button Operation Summary 3-32 Buttons, Operating 3-2
C
CAL DUE 3-6 CALIBRATION 6-1 Calibration Due Date 3-6, 3-12 Calibration Validity 6-2 Carbon Dioxide (CO2),
0 to 20% 2-4
Carbon Monoxide (CO),
0 to 1000ppm 2-5 CGI 3-3 CGI Mode 3-25, 4-11 Combustible Gas Indicator
(CGI) Operation 2-6
Conned Space Monitor
(CSM) Operation 2-6 Conrmation, Sensor 3-18 Construction 2-8, A-2 COPYRIGHT i CO STEL ALARM 4-8 CSM 3-3 CSM Mode 4-10
D
Datalogging 2-7, 3-24 Date 3-6 Default Alarms 4-5 Default Alarms Table 4-5 Details of Operating Ranges
2-2 Display A-2 DISPOSAL ADVICE ii DUE, CAL 3-6 DUE, SERVICE 3-12
I
Page 100
USER HANDBOOK
E
English B-1 EXHAUSTED, BATTERY 4-17
F
FAIL, FLOW 4-14 Fault Alarms 4-9 FAULT, MEMORY 4-15, 4-16 FAULT, SENSOR 4-12, 4-13 Fault, Zero 3-18 features 1-2 Features, Instrument 3-1 Filter, Internal 5-5 Filters 2-9 Filter(s) Replacement 5-5 FLAMMABLE (LEL) ALARMS
4-2 FLOW FAIL 4-14 Flow Rate Information A-2
G
GENERAL INFORMATION 2-1
H
Handle, Probe 3-3, 5-5 HANDLING iii Humidity A-2 Hydrogen Sulphide (H2S),
0 to 100ppm 2-4
I
Identication, Instrument 3-5 IMS 6-1 INFORMATION, ADDITIONAL
8-1
INFORMATION, GENERAL
2-1
INSTRUCTIONS, OPERATING
B-1 Instrument Features 3-1 Instrument Identication 3-5 Instrument Models / Ranges
2-1 Instrument Will Not Zero -
Gas Present 3-20 Instrument Will Not Zero -
Zero Fault 3-18 Internal Filter 5-5 INTRODUCTION 1-1
L
Language Option 3-4 Latching 4-4 LCD A-2 LEL 2-2 LEL HiHi ALARM 4-6 LIABILITY i Liquid Crystal Display (LCD)
2-9 Location (LOC) Selector 3-28 LOW BATTERY 4-16
M
MAINTENANCE, OPERATOR
5-1 Max / Min Values 2-8, 3-25
II
Loading...