GMI welcomes comments on all our publications. Your
comments can be of great value in helping us to improve
our customer publications. Please send any comments
that you have to customerservice@gmiuk.com
This Licence Agreement is your proof of licence. Please treat it as valuable
property. Please note that this licence agreement supersedes the “statement of
Copyright Restrictions”.
IMPORTANT - READ CAREFULLY
This is a legal and binding contract between you (either an individual or an entity),
the end user and Gas Measurement Instruments Limited (GMI). By opening
sealed software packages or by installing, copying, or otherwise using the ‘PS500 Settings’ Auto Bump / Calibration Station application program (SOFTWARE), you
agree to be bound by the terms of this agreement. If you do not agree to the terms
of this agreement you should promptly return this package and accompanying
items* (including written materials, binders or containers) to GMI.
CALIBRATION SOFTWARE LICENCE
1. GRANT OF LICENCE: The ‘PS500 Settings’ Licence Agreement (Licence)
permits concurrent use of the SOFTWARE on the number of computers for which
this licence grants (that is: single user package licences one user; twenty user
package licences twenty users). Additional SOFTWARE licences may be purchased
for each additional concurrent user.
If you have acquired this SOFTWARE in a multiple licence pack, you may make
the number of additional copies of the SOFTWARE authorised on the printed copy
of the licence. You may also store or install a copy of the network SOFTWARE
on a storage device, such as a network server, used only to install or run the
SOFTWARE on your other computers over an internal network; however you
must acquire and dedicate a licence for each separate computer on which the
SOFTWARE is installed or run from the storage device. A licence for the software
may not be shared or used concurrently on different computers.
This licence is granted for use on the GMI products and for use on one computer.
If you require a second copy you must purchase that from GMI or its distributor.
The documentation accompanying this software is essential for the correct and
safe operation of the instruments.
This software is used on certied equipment which may be used in life critical
applications. Only trained competent personnel should use this equipment and
should be aware of the certication, calibration and Health and Safety issues
associated with the product and its application.
2. COPYRIGHT: The SOFTWARE is owned by GMI and is protected by United
Kingdom copyright laws and international treaty provisions. Therefore you must
treat the SOFTWARE like any other copyrighted material (for example, a book or
musical recording) except that you may either (a) make one copy of the SOFTWARE
solely for backup or archival purposes, or (b) transfer the SOFTWARE to a single
i
Page 4
AUTO BUMP / CALIBRATION STATION - USER HANDBOOK
PS500
hard disk provided you keep the original solely for backup and archival purposes.
You may not copy the written materials accompanying the SOFTWARE.
3. OTHER RESTRICTIONS: You may not rent or lease the SOFTWARE, but
you may transfer the SOFTWARE and accompanying written materials on a
permanent basis to another individual. If you transfer the software, you must
retain no copies and the recipient must agree to the terms of this agreement.
You must not reverse engineer, decompile, or disassemble the SOFTWARE that
is provided solely as executable programs (.EXE les) or dynamic link libraries
(.DLL les). If the SOFTWARE is an update, any transfer must include the update
and all prior versions.
30 DAY LIMITED WARRANTY
If the product was purchased in the UK , simply return the full product in resaleable
condition* within 60 days of purchase to GMI, together with your original receipt or
invoice, for a full refund. If this product was not purchased in the UK, you should
return the full product* to the place of purchase within 30 days of purchase for
a full refund.
LIMITED WARRANTY
The USB memory stick and Handbook that make up this software product are
warranted by GMI to be free from defects in materials and workmanship for a
period of 60 (sixty) days from the date you purchased this product. If you notify
GMI within the warranty period of such defects in materials or workmanship, GMI
will replace the defective USB memory stick or handbook. The sole remedy for
breach of this warranty is limited to replacement of defective USB memory stick and
handbook and / or refund of purchase price and does not include any other kinds
of damage. Apart from the foregoing limited warranty, the software programs are
provided “AS IS”, without warranty of any kind, either expressed or implied. The
entire risk as to the performance of the program is with the purchaser. GMI does
not warrant that the operation of the programs will be uninterrupted or error free.
GMI assumes no responsibility or liability of any kind for errors in the program or
documentation or for the consequences of any such errors.
INCIDENTAL AND CONSEQUENTIAL DAMAGES CAUSED BY MALFUNCTION,
DEFAULT, OR OTHERWISE WITH RESPECT TO BREACH OF THIS WARRANTY
OR ANY OTHER EXPRESS OR IMPLIED WARRANTIES, INCLUDING THOSE
OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE, ARE
NOT THE RESPONSIBILITY OF GMI AND ARE HEREBY EXCLUDED BOTH
FOR PROPERTY AND, TO THE EXTENT NOT UNCONSCIONABLE, FOR
PERSONAL INJURY DAMAGE.
The remedies described above are the exclusive remedies extended to you by GMI
for any default, malfunction or failure of the product to conform with this warranty or
otherwise for the breach of this warranty or any other warranty, whether expressed
or implied. Should you have any questions concerning this agreement please
contact the GMI sales department on +44 (0)141 812 3211.
* If the product is not in resaleable condition, you will be charged a 20% restocking
fee.
ii
Page 5
COPYRIGHT
COPYRIGHT
This document is copyright of GMI and the information contained within is for use
only with ‘PS500 Settings’ Software. Reproduction, in whole or in part, including
utilisation in machines capable of reproduction or retrieval without the written
permission of GMI is prohibited. Reverse engineering is not permitted.
TRADEMARK INFORMATION
Windows XP™ and 7™ are registered trademarks or trademarks of the Microsoft
Corporation in the United States and / or other countries.
LIABILITY
Every care has been taken in the preparation of this document, however, GMI
do not accept any responsibility for errors or omissions and their consequences.
Information in this document is subject to change without prior notice.
This document does not constitute a specication or basis for a contract.
Your statutory rights under UK law are not affected.
iii
Page 6
AUTO BUMP / CALIBRATION STATION - USER HANDBOOK
PS500
iv
Page 7
CONTENTS
LICENCE AGREEMENT ................................... i
IMPORTANT - READ CAREFULLY .......................i
INDEX.............................................................. ix
vii
Page 10
AUTO BUMP / CALIBRATION STATION - USER HANDBOOK
PS500
viii
Page 11
1
INTRODUCTION
Fig. 1-1 PS500 Auto Bump / Calibration Station
1.1 GENERAL DESCRIPTION
The PS500 Auto Bump / Calibration Station provides the
user with a quick, easy and reliable method of ensuring that
PS500 instruments are correctly calibrated where company
policy dictates.
Equipment does not have to be taken out of service for
extended time periods while testing is undertaken.
1-1
Page 12
AUTO BUMP / CALIBRATION STATION - USER HANDBOOK
PS500
The Auto Bump / Calibration Station (ABC) has 3 variants:
• Single gas unit (GMI part no. 61502) - 1 gas cylinder
can be connected.
• Multi-gas unit (GMI part no. 61504) - up to 3 gases
cylinders can be connected.
• Reactive gases (CL2 / NH3) multi-gas unit (GMI part no.
61504R) - up to 3 gases cylinders can be connected.
Note: Throughout the user handbook the multi-gas
variant (GMI part no. 61504) will be used to
describe the Auto Bump / Calibration Station.
The units can be used for:
• A Bump Test, approximately 60 seconds gas duration
per cylinder, performs a full Bump Test of alarm set
points and sensor response.
• A Calibration of the instrument, approximately 60
seconds gas duration per cylinder.
Note: Gas duration is dependent on which PS500
model is being tested.
The ABC can be free standing on a workbench, or
alternatively, secured to the workbench individually or in a
daisy chain fashion using the mounting bracket and screws
supplied with the ABC. Refer to Appendix A for tting
instructions.
Power is supplied to the rear of the ABC via a 12V Universal
Power Supply Unit (PSU), supplied with the ABC. A 12V
Vehicle Power Supply is available as an accessory.
Gases are connected to the clearly marked inlet adaptors at
the rear of the ABC. An integral pump automatically draws
gas to the instrument for bump / calibration operation.
1-2
Page 13
INTRODUCTION
The user simply inserts the PS500 into the ABC, then closes
the cover initiating the automatic testing process.
Note: The PS500 MUST be switched on before inserting
into the Auto Bump / Calibration Station.
The ABC uses a trafc light system to display progress /
result,
i.e. LED 1 (Green) = ON (power)
LED 2 (Orange) = TEST (in progress)
LED 3 (Green) = PASS
LED 4 (Red) = FAIL
This provides the user with a clear indication of whether the
PS500 meets the selected parameters.
GMI ‘PS500 Settings’ Software is a Windows™ based
‘stand alone’ application that allows the user to congure /
edit test parameters and view / print test results.
This software will run on any PC / Laptop with Windows XP™
or newer software installed. (.Net Framework 2.0 or higher
may also be required on early Windows XP™ operating
systems).
Both the ABC and the USB memory stick are supplied with
necessary ‘PS500 Settings’ software installed.
and transferred to the ABC before testing /
calibration can take place.
The software test parameters are congured via PC or
Laptop then the settings le transferred to the ABC using the
USB memory stick supplied.
1-3
Page 14
AUTO BUMP / CALIBRATION STATION - USER HANDBOOK
PS500
The software provides a full audit trail of tested instruments
and the resulting data is retained in the station memory
until downloaded to the USB memory stick, before being
transferred (refer to note on page 5-1) to any compatible PC
/ Laptop for long term storage, review and / or printing.
A Calibration Certicate can also be generated automatically
after successful completion of instrument testing.
A permanent PC connection is provided via mini-USB socket
at the rear of the ABC. An Ethernet socket is also provided
at the rear of the ABC to allow results to be gathered over a
network. Both of these options require additional software,
either the ‘exiCal Plus’ PC software package or the ‘IMS’
network software package, both available from GMI.
Note: PS500 Bump / Calibration Software is only one
part of the requirements in terms of calibration
and maintenance of a PS500 portable gas
detector. Gases, delivery and trained personnel
are all of equal or even greater importance.
1-4
Page 15
INTRODUCTION
1.2 AUTO BUMP / CALIBRATION HARDWARE
The multi-gas Auto Bump / Calibration Stations (GMI Part
No’s. 61504 / 61504R) are provided with:
• 4mm (5/32in.) ttings (3 x gas and 1 x air).
• 6mm tting (1 x exhaust).
• 12V Universal PSU.
• USB memory stick with GMI ‘PS500 Settings’
software.
The single-gas Auto Bump / Calibration Station (GMI Part
No. 61502) is provided with:
• 6mm ttings.
• 12V Universal PSU.
• USB memory stick with GMI ‘PS500 Settings’
software.
1.3 ADDITIONAL ACCESSORIES
• Test Gas Cylinders - for a full list of gases, contact GMI
or an authorised distributor.
• On Demand Gas Flow Regulator - GMI Part No.
99118.
• Tygon® tubing - GMI Part No. 66118.
• 1.0 metre Tygon® tubing complete with 6mm adaptor
- GMI Part No. 64265.
• 0.6 metres reactive gas tubing - GMI Part No. 61540.
1-5
Page 16
AUTO BUMP / CALIBRATION STATION - USER HANDBOOK
PS500
1-6
Page 17
2
CONNECT CALIBRATION GAS AND
POWER SUPPLY
2.1 CONNECT CALIBRATION GAS
Gas is supplied to the Auto Bump / Calibration Station from
Test Gas Cylinders via On Demand Gas Flow Regulators
(GMI Part No. 99118) and Tygon® tubing (GMI Part No.
12712).
Fig. 2-1 Calibration Gas Setup
Note: For a comprehensive list of calibration
accessories contact GMI or an authorised
distributor.
2-1
Page 18
AUTO BUMP / CALIBRATION STATION - USER HANDBOOK
PS500
2.1.1 Compare Gas Cylinder Label with Software
Conguration
First of all, make sure that the cylinder is not older than the
‘use by’ date on the label illustrated in Fig. 2.2.
Make sure that the cylinder contains enough gas to complete
the bump test / calibration. The pressure gauge on the
Regulator Valve, after tting, gives a clear indication of this.
When applying gas to the instrument via the Auto Bump /
Calibration Station, it is important that the gas concentrations
listed on the cylinder label are consistent with the current
software conguration downloaded from the USB memory
stick to the unit.
As illustrated in Fig. 2.2, the cylinder gas concentrations can
be easily found on the cylinder label.
2-2
Fig. 2-2 Gas Cylinder Label
Page 19
CONNECT CALIBRATION GAS AND POWER SUPPLY
The concentrations, illustrated in Fig. 2-2, should be
consistent with the installed software conguration window,
as illustrated in Fig. 2-3.
Software Configuration
- Gas Concentrations
Fig. 2-3 Software Conguration
After comparing gas values between the Cylinder Gas Label
and the Software Conguration, when consistent, proceed to
the next paragraph.
If inconsistent, refer to Chapter 3 to update software
conguration then download the updated conguration to the
ABC.
Note: The Multi-gas ABC can have up to 3 gas
inputs, therefore 3 gas cylinders labels and
congurationstocompare.
2-3
Page 20
AUTO BUMP / CALIBRATION STATION - USER HANDBOOK
PS500
2.1.2 Connect Gas Cylinder to ABC
Connect the On Demand Regulator Valve to the gas cylinder
by screwing the valve rmly into the top of the cylinder, as
illustrated in Fig. 2-4. Fitting is simplied by pressing the valve
on to the cylinder with one hand and turning the cylinder with
the other. Care must be taken not to over-tighten the valve.
Fig. 2-4 Attach Regulator Valve to Gas Cylinder
Connect one of the supplied barbed adaptors to the end of
the tubing, then push the barbed adaptor into one of the three
‘Gas’ inlet adaptors on the rear of the ABC, as illustrated in
If the ABC is located in a poorly ventilated area, it is
recommended that tubing is connected to both the ‘Air’ inlet
and ‘Exhaust’ outlet adaptors on the rear face of the ABC,
using barbed adaptors.
GMI recommend the use of an ‘Air’ in-line lter, also supplied
with the unit. The (in-line lter) barbed adaptor is a push
t in the ‘Air’ inlet adaptor on the rear face of the ABC, as
illustrated in Fig. 2-6.
Air inlet tubing, from the in-line lter, can be routed to the
nearest window for the supply of fresh air.
Alternatively, compressed air can be used to supply inlet air.
Note: If the compressed air option is adopted, the in-
linelterisnotrequired.
2-5
Page 22
AUTO BUMP / CALIBRATION STATION - USER HANDBOOK
PS500
Fig. 2-6 Connect Air Tubing Adaptor to ABC
The ‘Exhaust’ line, not illustrated, consists of a length of
tubing and a barbed adaptor. The barbed adaptor is a push
t in the ‘Exhaust’ outlet adaptor on the rear face of the ABC.
Exhaust tubing can be routed to the nearest window.
To remove the ‘Gas’, ‘Air’ and ‘Exhaust’ barbed adaptors
from the ABC, depress the yellow and blue coloured collars,
in turn to release, then withdraw the adaptor and tubing.
2-6
Page 23
CONNECT CALIBRATION GAS AND POWER SUPPLY
2.2 CONNECT POWER SUPPLY
Power is supplied to the ABC via a Universal AC Power
Adaptor (12Vdc output) supplied with the unit.
The plug adaptor is located in the 12Vdc socket on the rear
face of the ABC, as illustrated in Fig. 2-7.
Fig. 2-7 Connect Power Supply to ABC
Locate the Universal Power Adaptor in a mains supply wall
socket then switch the mains power ON.
The ‘ON’ LED, on the front face of the unit, will illuminate
green, as illustrated in Fig. 2-8.
Fig. 2-8 Unit ‘ON’ Indication
2-7
Page 24
AUTO BUMP / CALIBRATION STATION - USER HANDBOOK
PS500
After 5 seconds all four LED’s illuminate green, as illustrated
in Fig. 2-9. This indicates that the unit is loading operating
software and preparing for testing.
Fig. 2-9 Unit Loading Software
After a further 25 seconds three green LED’s extinguish
leaving only the ‘ON’ indication illuminated, as shown in Fig.
2-8. Next, three LED’s coloured orange, orange and red
illuminate briey in sequence, indicating that GMI ‘TEST’
software has started.
Finally, the ‘ON’ indication is illuminated, as illustrated in Fig.
and transferred to the ABC before testing /
calibration can take place.
To view ‘PS500 Settings’, insert the memory stick into an
available USB port on your PC / Laptop.
The following screen is displayed on a Windows 7™ PC /
Laptop.
Fig. 3-1 PS500 Settings Windows 7™ Options
3-1
Page 26
AUTO BUMP / CALIBRATION STATION - USER HANDBOOK
PS500
Note: All software instructions in the user handbook,
are based on using Windows 7™.
Select ‘Open folder to view les’ to display the contents, as
illustrated in Fig. 3-2.
Fig. 3-2 PS500 Settings Contents Folder
Note: A short-cut to both the User Handbook and the
Quick Operation Guide, can be accessed from
this window. Adobe Acrobat Reader must be
installedtoviewtheseles.
To view the test conguration settings, open the
application from the list of items in
Fig. 3-2.
3-2
Page 27
VIEW / RE-CONFIGURE SOFTWARE
The window, illustrated in Fig. 3-3, is displayed.
Fig. 3-3 Conguration Window (Multi Gas Set-up)
There are two Bump Test / Calibrate options available:
• Test (i.e. bump test)
• Calibration
PS500 instruments can have many gas congurations. To
incorporate the majority of gas congurations, there are two
ABC type’s available for test and calibration:
• Air + 1 gas. Single Gas Cylinder - Gas Input 1.
• Air + 3 gas. Multi Gas Cylinders - Gas Inputs 1, 2, 3.
Note: Each ‘Gas Input’ (cylinder) can contain up to 5
different types of gas.
3-3
Page 28
AUTO BUMP / CALIBRATION STATION - USER HANDBOOK
PS500
3.1.1 Bump Test
This feature provides the user with the ability to perform a
full bump test, testing alarm set points and sensor response.
When gas is applied, using Test Gas Cylinders, for a pre-set
time, the audible and visual alarms will activate, and the unit
will record a ‘Pass’ when the target values are attained.
Note: Gas application times, alarm set points and
target values are gas type dependent.
To select / change the instrument gas type, ensure from the
‘Bump Test / Calibrate’ section, ‘Test’ is selected. Then, from
the appropriate ‘Gas Input’ section, select the required gas
type from the drop down list. The example in Fig. 3-4 shows
Methane selected from ‘Gas Input 1’.
3-4
Fig. 3-4 Instrument Gas Type Selection
Page 29
VIEW / RE-CONFIGURE SOFTWARE
To enter / edit the gas concentration, highlight the gas value
(0.0), then type in the new value (e.g. 50.0% LEL Methane),
as illustrated in Fig. 3-5.
Fig. 3-5 Gas Concentration
Note: It is important that the ‘Gas Concentration’ value
is consistent with the value displayed on the gas
cylinder label.
Isobutylene is the recommended gas for tested or calibrating
VOC 0-100 PPM and VOC 0 - 1000 PPM ranges.
If testing or calibrating VOC ranges with a gas other than
Isobutylene, select the VOC range from the drop down list.
Isobutylene can then be edited with your calibration gas of
choice. E.g. Toluene in ‘Gas Input 2’, as illustrated in Fig. 3-6.
Fig. 3-6 Editing VOC Gas Type
3-5
Page 30
AUTO BUMP / CALIBRATION STATION - USER HANDBOOK
PS500
3.1.2 Calibration
This feature provides the user with the option of performing a
full calibration of the instrument.
When gas is applied, using Test Gas Cylinders, for a preset time, the audible and visual alarms will activate, and the
unit will record a ‘Pass’ when the recalibrated target value is
attained.
Note: Gas application times, alarm set points and
target values are gas type dependent.
To select / change the instrument gas type, ensure from
the ‘Bump Test / Calibrate’ section, ‘Calibration’ is selected.
Then, from the appropriate ‘Gas Input’ section, select the
required gas type from the drop down list. The example in
Fig. 3-7 shows Methane selected from ‘Gas Input 1’.
3-6
Fig. 3-7 Instrument Gas Type Selection
Page 31
VIEW / RE-CONFIGURE SOFTWARE
To enter / edit the gas concentration, highlight the gas value
(0.0), then type in the new value (e.g. 50.0% LEL Methane),
as illustrated in Fig. 3-8.
Fig. 3-8 Gas Concentration
Note: It is important that the ‘Gas Concentration’ value
is consistent with the value displayed on the gas
cylinder label.
3-7
Page 32
AUTO BUMP / CALIBRATION STATION - USER HANDBOOK
PS500
3.1.3 Generate Settings File
The ‘Generate Settings Files’ button, illustrated in Fig. 3-9,
allows any conguration changes to be saved to the USB
memory stick, as updated settings le, then transferred to
the ABC.
Fig. 3-9 Generate Setting Files
To save changes, select the ‘Generate Settings File’ button
to save the data to the USB memory stick. The message
illustrated in Fig. 3-10 will be displayed.
3-8
Fig. 3-10 New PS500 Settings File Generated
Page 33
VIEW / RE-CONFIGURE SOFTWARE
3.2 TRANSFER UPDATED ‘PS500 SETTINGS’ to
ABC
Make sure that the 12V power supply is connected to the
ABC and switched ON, indicated by an illuminated green
‘ON’ LED on the front of the ABC and that the ABC has
completed start-up.
Remove USB memory stick from the PC / Laptop then insert
into the USB port on the front face of the ABC, as illustrated
in Fig. 3-11.
Fig. 3-11 Auto Bump / Calibration Station USB Port
3-9
Page 34
AUTO BUMP / CALIBRATION STATION - USER HANDBOOK
PS500
When the USB is inserted in the ABC, the updated settings
le is automatically transferred to the ABC, indicated by three
ashing green LED’s, as illustrated in Fig. 3-12.
Fig. 3-12 Transferring Data to Unit
On completion of uploading the updated settings le,
the three LED’s stop ashing, as illustrated in Fig. 3-13,
indicating the upload is complete.
Fig. 3-13 Upload Complete
Next, LED’s coloured orange, orange and red illuminate,
as illustrated in Fig. 3-14 and GMI ‘TEST’ software has re-
started.
Fig. 3-14 Test Software Re-started
Next, the three (orange, orange and red) LED’s extinguish,
leaving only the green ‘ON’ LED illuminated.
The Auto Bump / Calibration Station has stored the new data
and is ready for use.
The USB memory stick can now be removed.
3-10
Page 35
VIEW / RE-CONFIGURE SOFTWARE
3.3 HELP OPTIONS
The ‘PS500 Settings’ software contains a Help menu, as
illustrated in Fig. 3-15.
Fig. 3-15 Help Menu
The menu has the following options available:
• Handbook - select to open the User Handbook (in
PDF format).
• Quick Operation Guide - select to open the Quick
Operation Guide (in PDF format).
• About - contains information relating to the owner,
name and version of the application.
Note: Adobe Acrobat Reader must be installed to view
the Handbook and Quick Operation Guide.
3-11
Page 36
AUTO BUMP / CALIBRATION STATION - USER HANDBOOK
PS500
3-12
Page 37
4
BUMP / CALIBRATION OPERATION
4.1 PS500 SWITCH ON
The PS500 must be switched ON, before inserting into the
Auto Bump / Calibration Station.
To switch ON, press and hold the blue button, as illustrated
in Fig. 4-1.
Fig. 4-1 PS500 Switch ON
The instrument will switch on, run through the warm-up
sequence and show the normal operating display, as the
example illustrates in Fig. 4-2.
Fig. 4-2 Normal Operating Display
4-1
Page 38
AUTO BUMP / CALIBRATION STATION - USER HANDBOOK
PS500
4.2 INSERT INSTRUMENT IN AUTO BUMP /
CALIBRATION STATION
Note: Before use, ensure the ABC is ready for operation
To release the front cover of the Auto Bump / Calibration
Station, press the cover latch downwards, as illustrated in
Fig. 4-3, then open the cover fully.
4-2
Fig. 4-3 Open Front Cover
Page 39
BUMP / CALIBRATION OPERATION
Locate the instrument inlet connector over the gas supply
nozzle in the unit, as illustrated in Fig. 4-4.
Fig. 4-4 Locate Instrument
4-3
Page 40
AUTO BUMP / CALIBRATION STATION - USER HANDBOOK
PS500
Push the instrument rearwards until correctly seated in the
station recess, as illustrated in Fig. 4-5.
Fig. 4-5 Instrument Correctly Seated
Note: Ensure the instrument is switched ON before
closing the front cover.
4-4
Page 41
BUMP / CALIBRATION OPERATION
Close the front cover of the Auto Bump / Calibration Station,
by using both thumbs. Press the lower centre of the cover
rmly, as illustrated in Fig. 4-6, until the latch ‘clicks’ shut.
Fig. 4-6 Close Front Cover
4-5
Page 42
AUTO BUMP / CALIBRATION STATION - USER HANDBOOK
PS500
4.3 BUMP / CALIBRATION OPERATION
The Bump / Calibration operation starts up automatically
following the locking of the unit cover latch.
After the cover latch ‘clicks’ shut, the orange ‘Test in Progress’
LED illuminates, as illustrated in Fig. 4-7.
Fig. 4-7 Test in Progress
During testing, when the alarm set points are reached, the
instrument’s audible and visual alarms will activate for a
short period of time.
On test completion the instrument will switch OFF.
A test ‘PASS’ is indicated by a green LED, as illustrated in
Fig. 4-8.
Fig. 4-8 PASS Indication
Alternatively, a test ‘FAIL’ is indicated by a red LED, as
illustrated in Fig. 4-9.
4-6
Fig. 4-9 FAIL Indication
Page 43
BUMP / CALIBRATION OPERATION
Next, the unit pump switches OFF and the instrument can be
safely removed.
The test result indication is displayed until the door is opened.
All test results are stored in the ABC memory and can be
accessed as detailed in Chapter 5.
4.4 REMOVE INSTRUMENT FROM ABC
To release the front cover of the Auto Bump / Calibration
Station, press the cover latch downwards, as illustrated in
Fig. 4-10, then open the cover fully.
Grasp the instrument, lean the top of instrument away from
the ABC, then carefully lift away from gas supply nozzle, as
illustrated in Fig. 4-11.
Close the front cover of the ABC then, using both thumbs,
press the lower centre of the cover rmly until the latch
‘clicks’ shut.
damage to the gas supply nozzle could result.
Fig. 4-11 Lift Instrument Clear of Gas Nozzle
4.5 ABC FAULT INDICATION
Fault detection indication is via the LED’s on the front face of
the unit, as follows:
4-8
Page 45
BUMP / CALIBRATION OPERATION
Test Error
A ‘Test Error’ is indicated by an orange ashing LED, as
illustrated in Fig. 4-12.
Fig. 4-12 Test Error
If this fault occurs, it is recommended that the USB memory
stick is inserted into a PC / Laptop and re-congured, as
detailed in Chapter 4.
On completion, the data should then be transferred to the
Auto Bump / Calibration Station.
Pump Fault
A ‘Pump Fault’ is indicated by orange and red ashing LED’s,
as illustrated in Fig. 4-13.
Fig. 4-13 Pump Fault
If this fault occurs, it is recommended that the tubing from
the gas cylinder to the Auto Bump / Calibration Station inlet
adaptor is inspected for possible restriction of blockage.
If this action does not clear the fault, remove the instrument
from the unit and inspect the instrument for possible ow
fault. If this does not clear the fault, contact GMI.
4-9
Page 46
AUTO BUMP / CALIBRATION STATION - USER HANDBOOK
PS500
4-10
Page 47
5
VIEW / PRINT TEST RESULTS
5.1 INTRODUCTION
The test results of all instruments that have been Bump
Tested / Calibrated in the unit are stored in the unit internal
memory.
The USB memory stick (Part No. 61251), containing ‘PS500
Settings’ software and supplied with the Auto Bump /
Calibration Station, must be used to download this data.
Fig. 5-1 Downloading Memory Stick
The memory stick can be used to download data from
several GMI Auto Bump / Calibration Stations, if required.
The data identies the serial number of the unit used to
download data from each instrument, bump tested or
calibrated.
Note: It is recommended a backup of the test results
are made, on a PC / Laptop, on a regular basis; in
case of the USB memory stick being misplaced.
5-1
Page 48
AUTO BUMP / CALIBRATION STATION - USER HANDBOOK
PS500
5.2 DOWNLOAD TEST RESULTS
Make sure that the Auto Bump / Calibration Station is
connected to mains power supply, switched ON and start-up
is completed.
This is indicated by an illuminated green ‘ON’ LED, on the
front face of the unit, as illustrated in Fig. 5-2.
Fig. 5-2 Unit ‘ON’ Indication
Insert USB memory stick into the USB port on the front face
of the unit, as illustrated in Fig. 5-3.
5-2
Fig. 5-3 Auto Bump / Calibration Station USB Port
Page 49
VIEW / PRINT TEST RESULTS
When the USB memory stick is inserted in the unit, all stored
test results are automatically transferred to the memory stick.
During this process three green LED’s ash in sequence
from left to right, as illustrated in Fig. 5-4.
Fig. 5-4 Downloading Data
This is followed by all four green LED’s illuminating, as
illustrated in Fig. 5-5, indicating that download is complete.
Fig. 5-5 Download Complete
Next, LED’s coloured orange, orange and red illuminate, as
illustrated in Fig. 5-6. This indicates that the test data has
been downloaded and GMI ‘TEST’ software has re-started.
Fig. 5-6 Test Software Re-started
Next, the three (orange, orange and red) LED’s extinguish,
leaving only the green ‘ON’ LED illuminated, as illustrated in
Fig. 5-2.
The USB memory stick can now be removed and transferred
to a PC / Laptop to view test results.
5-3
Page 50
AUTO BUMP / CALIBRATION STATION - USER HANDBOOK
PS500
5.3 VIEW / PRINT TEST RESULTS
Insert the USB memory stick into a USB port on PC / Laptop.
Select ‘Open folder to view les’ to display the contents, as
illustrated in Fig. 5-7.
Fig. 5-7 PS500 Settings Contents Folder
The contents of the folder are as follows:
• Generated Certicates:
Contains copies of all calibration certicates
generated.
• PS500 Settings:
This is a systems folder. Do not edit.#
• Test Results:
Contains copies of all test results generated.
• Upload:
Contains conguration settings used by the Auto
Bump / Calibration Station. Do not edit.#
5-4
Page 51
VIEW / PRINT TEST RESULTS
Auto Bump / Cal Station
Tested Instruments
• Autorun:
USB memory stick will start automatically when
inserted. Do not edit.#
• Handbook:
Auto Bump / Calibration Station user handbook (in
PDF format).
• Quick Operation Guide:
Auto Bump / Calibration Station poster style document
(in PDF format).
• Interop.ADODB.dll. Do not edit.#
• PS500 Settings application.
• PS500 Settings.exe. Do not edit.#
# - Depending on PC / Laptop ‘Folder Option’ settings these
items may be hidden. If visible do not edit.
5.3.1 Test Result Files
The contents of this folder includes all instrument test result
les downloaded from each Auto Bump / Calibration Station
that the USB memory stick has been used on.
Test Files
Fig. 5-8 ‘Test Results’ Folder Contents
5-5
Page 52
AUTO BUMP / CALIBRATION STATION - USER HANDBOOK
PS500
Test Result Files:
The test le names in Fig. 5-8, (e,g.
PS500_250013_21_7_2015_1415) identies the serial
number of the instrument (e.g. 250013), the date tested in
day, month and year (e.g. 21_7_2015) and the time testing
was completed (e.g. 1415hrs).
Each le can be opened individually to display a
comprehensive list of test results obtained from that
particular instrument.
The following example, in Fig. 5-9, illustrates information
available, including:
• Instrument Type (e.g. PS500)
• Test Type (e.g. CALIBRATION)
• Instrument Serial No. (e.g. 250013)
• New Cal Due Date (if Test registered as PASS)
• ABC Serial No. (e.g. 805000)
• Date Tested (e.g. 21 July 2015)
• Time Tested (e.g. 2:15:25 PM)
• Overall Result (e.g. PASS)
Followed by a listing of air tests, then gas tests which display
target / lo Limit / hi Limit / result / range / re cal val / re cal
result for the gas range.
Finally, an audible / visual alarm test result is displayed.
Test results, as illustrated in Fig. 5-9, can be printed if
required.
5-6
Page 53
VIEW / PRINT TEST RESULTS
Fig.5-9 View Test Result File
5-7
Page 54
AUTO BUMP / CALIBRATION STATION - USER HANDBOOK
PS500
Auto Bump / Calibration Station Instruments:
The tested les (e.g. Tested805000.rec) illustrated in the
Fig. 5-8, identies the serial number of each Auto Bump /
Calibration Station (e.g. 805000) that the memory stick has
been used to download data from. Each le contains a list of
all instrument serial numbers tested on that particular station.
Each ‘.rec’ le can be opened individually to display a
comprehensive list of test results obtained from that particular
Auto Bump / Calibration Station.
The following example, Fig. 5-10, illustrates a list of
instruments tested on unit serial number 805000.
Fig. 5-10 Tested Instruments on 805000
The le includes:
5-8
• A list of instruments tested
• Date Tested (day, month, year)
• Time Tested
• Test Result, Pass (P) or Fail (F)
Page 55
VIEW / PRINT TEST RESULTS
5.3.2 GeneratedCerticates
The contents of this folder includes a list of all Bump Test /
Calibration Certicates generated when selecting the ‘View
Certicate’ button in the conguration window, as illustrated
in Fig. 5-11.
Fig. 5-11 Conguration Window
5-9
Page 56
AUTO BUMP / CALIBRATION STATION - USER HANDBOOK
PS500
Certicate Content:
The certicates, as illustrated in Fig. 5-17 and 5-18, include
company details, order number, etc. These details are
completed in a ‘Certicate Printer’ window, accessed by
opening ‘View Certicate’ in the conguration window,
illustrated in Fig. 5-11.
The ‘Certicate Printer’ window is displayed, as illustrated
in Fig. 5-12.
Fig. 5-12 ‘Certicate Printer’ Window
The windows contents can now be completed to include
company name, address, etc., as illustrated in Fig. 5-13.
5-10
Page 57
VIEW / PRINT TEST RESULTS
Fig. 5-13 Certicate Completed
When ‘Certicate Printer’ window is complete, select the
‘Create Certicate’ button, illustrated in Fig. 5-13, to print
certicate as illustrated in example Fig. 5-18.
To include your company’s logo on the printed certicate,
insert the logo le, named logo.jpg into the ‘Generated
Certicates’ folder on the memory stick. The logo will
be included in the top right hand corner of the certicate,
as illustrated in the ‘GMI’ logo example in the ‘Calibration
Certicate’, illustrated in Fig. 5-18.
After selecting the ‘Create Certicate’ button, the test results
folder will be displayed, highlight a test result for certicate
generation then select ‘Open’, as illustrated in Fig. 5-14.
Fig. 5-14 Select Test Result for Certicate Generation
Next, a ‘DONE!’ window will be displayed. The certicate has
now been generated for the selected test result le. Select
‘OK’, from the window, to open the generated certicate, as
illustrated in Fig. 5-15.
5-12
Fig. 5-15 Open Certicate
Page 59
VIEW / PRINT TEST RESULTS
Generated Certicate Files
The certicate les illustrated in Fig. 5-16, identies date and
time of each certicate’s generation.
Fig. 5-16 Generated Certicate Files
Each le can be opened individually to display / print a ‘Bump
Results’ or ‘Calibration Certicate’.
Note: If background image (green tick), illustrated
in Fig. 5-15 or 5-16, is not present adjust local
printer settings.
An example of the ‘Bump Results’ and ‘Calibration Certicate’
are illustrated in Fig. 5-17 and Fig. 5-18 respectively.
5-13
Page 60
AUTO BUMP / CALIBRATION STATION - USER HANDBOOK
PS500
Fig. 5-17 Bump Results
5-14
Fig. 5-18 Calibration Certicate
Page 61
A
FIT MOUNTING BRACKET
Fig. A-1 Auto Bump / Calibration Station
(with mounting bracket tted)
1. GENERAL DESCRIPTION
The PS500 Auto Bump / Calibration Station is supplied with
a mounting bracket and necessary screws to secure the unit
to a workbench or similar.
The bracket also provides the facility to ‘daisy chain’ a series
of Auto Bump / Calibration Stations if required.
A-1
Page 62
AUTO BUMP / CALIBRATION STATION - USER HANDBOOK
PS500
2. FIT MOUNTING BRACKET
1. Make sure the workbench surface is clean and free
from dirt and grime.
2. Place clean cloth or similar on workbench to protect
Auto Bump / Calibration Station front cover surface
from scratching and / or scufng.
3. Carefully place the unit face down on workbench. Refer
to Fig. A-2.
Fig. A-2 Attach Bracket to Unit
4. Loosely attach bracket to the underside of the unit
with two pozi countersunk screws (Part No. 24722), as
illustrated in Fig. A-2.
A-2
Page 63
APPENDIX A
5. Loosely attach two double sems pozi pan screws (Part
No. 24957) through bracket and into rear panel of unit,
as illustrated in Fig. A-2.
6. Using a No.2 Pozidrive screwdriver, tighten all four
screws to secure.
3. ATTACH UNIT to WORKBENCH / DAISY CHAIN
OPTION
The bracket can be secured to a workbench or similar type
surface using four suitable screws (not supplied).
Secure to
Workbench
(4-places)
Fig. A-3 Attach Unit to Workbench
A-3
Page 64
AUTO BUMP / CALIBRATION STATION - USER HANDBOOK
PS500
(4-places)
A group of Auto Bump / Calibration Stations can be connected
in a ‘daisy chain’ fashion by using a double sems pozi pan
screw supplied (Part No. 24957) through bracket at rear of
station and into captive nut of adjoining bracket, as illustrated
in Fig. A-4.
Captive Nut
Attach to
Adjoining Bracket
Fig. A-4 Attach Adjoining Units
A-4
Page 65
INDEX
Symbols
30 DAY LIMITED
WARRANTY ii
A
About 3-11
Additional Accessories 1-5
Air + 1 gas 3-3
Air + 3 gas 3-3
Air inlet 2-5
Air in-line lter 2-5
Attach Adjoining Units A-4
Attach Bracket to Unit A-2
Attach Regulator Valve to
Gas Cylinder 2-4
Attach Unit to Workbench
A-3
Auto Bump / Calibration
Hardware 1-5
Auto Bump / Calibration
Station 1-1
Auto Bump / Calibration
Station USB Port 3-9, 5-2
(Multi Gas Set-up) 3-3
Connect Air / Exhaust
Tubing to Auto Bump /
Calibration Station 2-5
Connect Air Tubing
to ABC 2-6
Connect Calibration Gas
2-1
Connect Gas Cylinder to
Auto Bump / Calibration
Station 2-4
ix
Page 66
AUTO BUMP / CALIBRATION STATION - USER HANDBOOK
PS500
Connect Power Supply 2-7
Connect tubing to ABC 2-5
COPYRIGHT iii
D
Download Complete 5-3
Downloading Data 5-3
Download Test Results 5-2
E
Editing VOC Gas Type 3-5
Exhaust outlet 2-5
F
FAIL Indication 4-6
Fault Indication 4-8
Fit Mounting Bracket A-1
exiCal Plus 1-4
G
Gas Concentration 3-5, 3-7
Gas Cylinder Label 2-2
Gas duration 1-2
Gas inlet adaptor 2-4
General Description 1-1