GMI PPM Gasurveyor 500 User Handbook Manual

PPM Gasurveyor 500
User Handbook
Issue 6 28/07/04 Part Number: 42142
GMI welcomes comments on all our publications. Your comments can be of great value in helping us to improve our customer publications. Please send any comments that you have to our Sales Department at GMI. Contact details are provided inside the back cover of this handbook.
Instrument Service / Repair contact details are also provided inside the back cover of this handbook.

COPYRIGHT

This User Handbook is copyright of Gas Measurement Instruments Ltd (GMI) and the information contained within is for use only with the PPM Gasurveyor 500 instruments. Reproduction, in whole or in part, including utilisation in machines capable of reproduction or retrieval without written permission of Gas Measurement Instruments Ltd is prohibited. Reverse engineering is not permitted.

LIABILITY

Every care has been taken in the preparation of this document, but GMI Ltd do not accept any responsibility for errors or omissions and their consequences. Information in this document is subject to change without notice. This document does not constitute a specification or basis for a contract. Your statutory rights under law are not affected.

MODIFICATION NOTICES

GMI aim to notify customers of relevant changes in the product operation and maintain this manual up to date. In view of the policy of continuous product improvement there may be operational differences between the latest product and this manual.
This Handbook is an important part of the PPM Gasurveyor 500 product. Please note the following points:
It should be kept with the instrument for the life of the product.
Amendments should be attached.
This Handbook should be passed on to any subsequent owner/user of the instrument.
Although every care is taken in the preparation of this Handbook it does not constitute a specification for the instrument.

SOFTWARE

Software supplied on EPROM or similar device for use in a particular product, may only be used in that product and may not be copied without the written permission of Gas Measurement Instruments Ltd. Reproduction or disassembly of such embodied programmes or algorithms is prohibited. Ownership of such software is not transferable and GMI Ltd does not warrant that the operation of the software will be error free or that the software will meet the customer’s requirements.

DISPOSAL ADVICE

When no longer in use, dispose of the instrument carefully and with respect for the environment. GMI will dispose of the instrument without charge if returned to the factory.
i
PPM GASURVEYOR 500 USER HANDBOOK

SAFETY

The instrument must be regularly serviced and calibrated by fully trained personnel in a safe area.
Batteries: Alkaline batteries or *Rechargeable battery pack must be exchanged (*and recharged) in a safe area and fitted correctly before use. Never use damaged batteries or expose to extreme heat. See Section 4 : OPERATOR MAINTENANCE.
Only GMI replacement parts should be used.
If the instrument detects gas, follow your own organisation’s procedures and operational guidelines.
The combustion chamber is a flameproof assembly and must not be opened in the presence of a flammable atmosphere.
PPM Gasurveyor 500 instruments are certified as EEx iad IIC T4
(-20oC< Tamb < 50oC). BAS01ATEX2292 II 2 G.
UL Class 1 Groups A, B, C and D.
This equipment is designed and manufactured to protect against other hazards as defined in paragraph 1.2.7 of Annex II of the ATEX Directive 94/9/EC
Any right of claim relating to product liability or consequential damage to any third party against GMI is removed if the warnings are not observed.

AREAS OF USE

Exposure to certain chemicals can result in a loss of sensitivity of the flammable sensor. Where such environments are known or suspected it is recommended that more frequent response checks are carried out. The chemical compounds that can cause loss of sensitivity include Silicones, Lead, Halogens and Sulphur. Do not use instrument in potentially hazardous atmospheres containing greater than 21% Oxygen. The enclosure material is polypropylene and must not be exposed to environments which are liable to result in mechanical or thermal degradation or to damage caused by contact with aggressive substances. Additional protection may be required in environments where the instrument enclosure is liable to damage.

STORAGE, HANDLING AND TRANSIT

The batteries in the rechargeable pack contain considerable energy and care should be taken in their handling and disposal. Battery packs should be removed if the instrument is stored for longer than 3 months. The instrument is designed to handle harsh environments. The sensing elements are sealed to IP54 and the rest of the instrument to IP64. If not subject to misuse or malicious damage, the instrument will provide many years of reliable service. The instrument contains electrochemical sensors with a life of 2 years. Under conditions of prolonged storage the sensors should be removed. The sensor contains potentially corrosive liquid and care should be taken when handling or disposing of the sensor, particularly when a leak is suspected.
ii

REVISION RECORD

Date Pages Description Of Change
Issue 1 All New Handbook
03.06.99
Issue 2 All Handbook revised to
18.05.01 reflect updated
Issue 3 All Handbook revised to
22.03.02 incorporate effect of
Issue 4 All Handbook revised to
01.05.02 incorporate effect of
Issue 5 All Handbook revised to
17.07.03 incorporate effect of
Issue 6 All Handbook revised to 28/07/04 include Appendix C
instrument options and features.
CR 2272 and CR 2288.
CR 2305
CR 2291
(translations)
iii
PPM GASURVEYOR 500 USER HANDBOOK
iv

CONTENTS

COPYRIGHT ................................................................. i
LIABILITY ...................................................................... i
MODIFICATION NOTICES ........................................... i
SOFTWARE .................................................................. i
DISPOSAL ADVICE ...................................................... i
SAFETY ........................................................................ ii
AREAS OF USE ........................................................... ii
STORAGE, HANDLING AND TRANSIT ...................... ii
REVISION RECORD ..................................... iii
INTRODUCTION ........................................... 1
GENERAL INFORMATION ........................... 3
Ranges of Operation .................................................... 3
LEL, 0 to 100% .......................................................... 3
Volume Gas, 0 to 100 % ............................................ 4
PPM Flammable 0 – 1000 ppm ................................. 5
Threshold Display Options ......................................... 6
Search Mode .............................................................. 7
Geiger Indication ........................................................ 8
v
PPM GASURVEYOR 500 USER HANDBOOK
Construction .................................................................. 8
Batteries ........................................................................ 9
Disposable Alkaline (LR20) Dry Cell Batteries .......... 9
Rechargeable Battery Pack ....................................... 9
Filters .......................................................................... 10
Liquid Crystal Display (LCD) ...................................... 10
Before Use Checks ..................................................... 11
OPERATING INSTRUCTIONS.................... 13
Modes of Operation .................................................... 13
Mode 1 ..................................................................... 14
Mode 2 ..................................................................... 14
Calibration Date Features ........................................ 15
Switching Off the Instrument Pump ......................... 15
Switching Off ............................................................ 16
Changing Range ...................................................... 16
Activate the Audible Geiger Indication (PPM or
Search Range) ......................................................... 16
Zeroing the ppm or Search Range .......................... 17
Summary of Button Operation ................................. 18
Operator Messages / Fault Flags ............................... 18
OPERATOR MAINTENANCE ..................... 21
Rechargeable Battery Pack ........................................ 21
Standard Charger ..................................................... 21
vi
CONTENTS
Flatbed Charger ....................................................... 22
Smart Charger .......................................................... 22
Replacing the Battery Pack ..................................... 23
Recharging the Battery Pack ................................... 24
Replacing Alkaline (LR20) Dry Cell Batteries ............ 24
Filter Replacement ..................................................... 26
Probe Handle Assembly - Part No. 13561 .............. 27
In-line Dust Filter (Accessory) ................................. 28
In-line Water Filter (Accessory) ............................... 29
CALIBRATION ............................................ 31
Calibration Validity ...................................................... 32
ACCESSORIES........................................... 33
ADDITIONAL INFORMATION ..................... 37
Training ....................................................................... 37
World Wide Web ......................................................... 37
TYPICAL OPERATING PARAMETERS .....A-1
Size ........................................................................ A-2
Weight .................................................................... A-2
Operating Temperature .......................................... A-2
Humidity ................................................................. A-2
vii
PPM GASURVEYOR 500 USER HANDBOOK
Construction ........................................................... A-2
Display .................................................................... A-2
Sampling System ................................................... A-2
Power Source ......................................................... A-3
FIELD CALIBRATION ................................ B-1
Selectable Ranges in FCM ....................................... B-3
Entering FCM ............................................................ B-3
Zeroing the Instrument ............................................. B-4
Field Calibration Procedure ...................................... B-5
Quitting FCM ............................................................. B-9
Quit And Save Changes ........................................ B-9
Quit Without Saving Changes ................................ B-9
OPERATING INSTRUCTIONS.................. C-1
English ..................................................................... C-2
Deutsch (German) ................................................... C-6
Svensk (Swedish) .................................................. C-10
Dansk (Danish) ...................................................... C-14
Nederlands (Dutch) ............................................... C-18
INDEX ............................................................ I
viii
1

INTRODUCTION

The PPM Gasurveyor 500 is a two button instrument designed to measure Parts Per Million, Lower Explosive Level and Volume flammable gas. The instrument is designed to the latest standards and is certified for use in Hazardous Areas.
The PPM Gasurveyor 500 contains the following ranges:
$ LEL, 0 to 100%
$ Volume gas, 0 to 100%
$ PPM Flammable
$ Search mode (enhanced ppm flammable range)
1
PPM GASURVEYOR 500 USER HANDBOOK
$ As the PPM Gasurveyor 500 has a wide variety of
available user selected options, it is not possible to provide an operator handbook specific to each possible variation, therefore, what we have pro­vided in the following pages is the standard de­fault of how the instrument would generally be configured, with the possible options detailed in italic text.
GMI recommend that you take the time to study your instrument and, where practical and with advice from your company’s Purchasing / Management departments, highlight your particular instrument configuration.
The main features of the instrument are:
$ Rugged polypropylene case, sealed to IP54 rating
and suitable for outdoor use.
$ Two button operation allowing the user access to
all features.
$ LCD with backlighting which displays the current
gas readings (in both digital and analogue forms) together with operational and status information.
$ Audible and Visual Geiger indication, configured
in PPM Flammable and Search ranges, provides indication of gas concentration increase via the in­creased rate of sounder and LED pulse.
$ Directly interfaces with the GMI Auto Test Cali-
bration Systems.
2
2

GENERAL INFORMATION

Ranges of Operation

LEL, 0 to 100%

The LEL range indicates the explosibility of the flammable gas in the sample. This is displayed as a percentage of the lower explosive limit (LEL) of the gas.
The instrument range is displayed in the top right corner of the display as % LEL. From 0 to 10% LEL the digital display resolves to 0.1% LEL. From 10 to 100% LEL the digital display resolves to 1% LEL. The analogue bar graph follows in 4% steps. An example of the LEL display is shown in Figure 2.1. Autoranging will automatically switch the range to Volume Gas when 100% LEL is reached.
The detection principle for this range is a catalytic reaction.
Figure 2.1 LEL Range
3
PPM GASURVEYOR 500 USER HANDBOOK

Volume Gas, 0 to 100 %

This range displays the total volume of a specific flammable gas with respect to air. The calibration gas is shown on the service label and for the purpose of this handbook is assumed to be methane. Instruments calibrated for methane in air should only be used for measuring such mixtures. To change the calibration gas, e.g. from methane to propane, the instrument must be recalibrated by suitably trained personnel.
On the Volume Gas range the instrument range is displayed in the top right corner of the LCD as GAS. The digital display resolves the signal to 1% GAS with the analogue bar graph following in steps of 4%. Figure 2.2 shows the Volume Gas display. The detection principle for the Volume Gas range is thermal conductivity.
Figure 2.2 Volume Gas Range
4
GENERAL INFORMATION
PPM Flammable 0 – 1000 ppm
This range is used to measure very low levels of gas and indicates the parts per million concentration of the gas in air (1000 ppm is equivalent to 0.1% Volume Gas).
The digital display reading resolves to 5 ppm with the analogue bargraph following in steps of 40 ppm. The ppm range is more sensitive than LEL and Volume ranges and takes longer to stabilise. The detection principle is a catalytic reaction with the sensitivity of the sensors greatly enhanced compared to the LEL range. Digital signal correction techniques are used to minimise drift. The ppm range includes a manual zeroing procedure.
The ppm range has the option of using a ‘threshold’. By this, we mean that we can display either a blank screen or simply a zero (0) on the display, up to a pre-set level.
This allows small readings, which may not be due to the gas we wish to find or very small amounts of gas, to be ignored if they fall below any pre-set action levels. This allows the operator to react only to significant levels.
The ppm range has no threshold set by default. If your instrument has a threshold, write the level here:
.......................................................................................
The ppm range, by default, has the audible and visual Geiger indication configured which provides indication of gas concentration increase via the increased rate of sounder and LED pulse.
When the instrument is switched on, the visual Geiger indication is enabled and the audible Geiger indication is
5
PPM GASURVEYOR 500 USER HANDBOOK
disabled. To enable audible Geiger indication, press and
hold Button Two .
The display changes between PPM and Aud (if enabled) as shown in Figure 2.3.
Figure 2.3 PPM Range

Threshold Display Options

The ppm range will by default, if a threshold is selected, display below the threshold as zero (0) on the screen.
The option is to display a blank screen for levels below the threshold, as shown in Figure 2.4.
Figure 2.4 Threshold Display Option
6
GENERAL INFORMATION

Search Mode

Search mode is an enhanced variant of the ppm flammable range with increased resolution and faster response. Search mode can be used for detecting leaks which can then be classified using the more accurate ppm, LEL and Volume Gas ranges.
The Search range provides an indication of gas only, and it should not be used to carry out any assessment of gas levels.
The pump speed in this range is higher, which is apparent from the increased sound from the pump.
The Search range, by default, has the audible and visual Geiger indication configured which provides indication of gas concentration increase via the increased rate of sounder and LED pulse.
When the instrument is switched on, the visual Geiger indication is enabled and the audible Geiger indication is disabled. To enable audible Geiger indication, press and
hold Button Two
.
The display changes between search symbol and Aud (if enabled) as shown in Figure 2.5.

Figure 2.5 Search Mode

7
PPM GASURVEYOR 500 USER HANDBOOK

Geiger Indication

The Geiger provides indication that as the gas concentration increases, the sounder (if activated) and the LED, pulse at an increasing rate. When the PPM Gasurveyor 500 has PPM Flammable or Search range selected, or is switched on in full autorange mode then, by default, the visual Geiger indication is enabled and the audible Geiger indication is disabled.
To enable the audible Geiger indication, press and hold
Button Two
.
A further press and hold of Button Two
will increase
the audible level. A further press and hold of Button Two
will return to visual only Geiger indication.
Note: When Search mode is selected, the pump operates at an increased speed. If Search mode is selected when the pump is switched off, the pump then reactivates.

Construction

The instrument is housed in a tough, impact resistant, moulded case made of polypropylene.
The top panel is protected by a stainless steel top plate covering a toughened glass LCD cover. The battery pack is sealed and attached to the main in­strument body by means of 2 stainless steel hexagonal screws. The instrument is sealed against dust and water to IP54 standard. The sensors are protected from dust and water by membrane and cotton filters.
8
GENERAL INFORMATION

Batteries

Disposable Alkaline (LR20) Dry Cell Batteries

Alkaline batteries provide approximately 15 hours operational life at ambient temperature of 20 When the batteries are low or exhausted it is necessary to fit 4 new batteries to reset the battery indicator to 100%. Do not mix old and new batteries.
An indication of the battery condition is displayed after power on and during warm-up, with status shown as either OK or LO. If LO condition is displayed, a maximum battery operational life of 120 minutes remains. During operation the ‘BAT’ alarm flag is displayed when approximately 60 minutes of operating time remains at normal temperature. The instrument may continue to be used until it switches off automatically.

Rechargeable Battery Pack

The GMI rechargeable battery pack provides approximately nine hours operational life, from fully charged, at ambient temperature of 20 indication of the battery condition is displayed after power on and during warm-up, with status shown as either OK or LO. If LO condition is displayed, a maximum battery operational life of 90 minutes remains. During operation the ‘BAT’ flag is displayed when approximately 30 minutes operating time is left at normal temperatures. The instrument will then turn off.
There are three GMI Battery Chargers: a Standard Charger, a Flatbed Charger and a Smart Charger. The
o
C (68oF).
o
C (68oF). An
9
PPM GASURVEYOR 500 USER HANDBOOK
Smart Charger has both slow and fast charge options as well as a serial link for communications with the instrument. See Rechargeable Battery Pack in Section 4
OPERATOR MAINTENANCE.

Filters

The PPM Gasurveyor 500 has a cotton particulate filter located behind the instrument inlet nozzle. The filter should be checked at frequent intervals or when the SAMPLE FAULT flag appears in the instrument display, and where appropriate, changed to ensure a clean sample path. In the event of the SAMPLE FAULT flag becoming activated, the pump should be stopped as soon as possible and the filter replaced. See Filter Replacement in Section 4 OPERATOR MAINTENANCE.
If SAMPLE FAULT still shows, after filter replacement, the instrument will require to be returned for servicing.

Liquid Crystal Display (LCD)

The LCD shows the current gas readings in both analogue and digital form together with operational and status information. The display is protected by a toughened glass cover. Backlighting is provided to enable the display to be seen under low ambient light conditions.
10
GENERAL INFORMATION

Before Use Checks

The following checks should be carried out before using the instrument on site:
$ The instrument is clean and in good condition.
$ The batteries have sufficient power left in them
for the intended use of the instrument.
$ Any filters are clean and in good condition.
$ The sample line and any other accessories are in
good condition and leak free.
$ All gas ranges are operational and zeroed correctly.
$ The calibration is still valid.
11
PPM GASURVEYOR 500 USER HANDBOOK
12
3

OPERATING INSTRUCTIONS

Modes of Operation

The instrument has two modes of operation which are accessed by switching the instrument ON with either
Button One or Button Two , as shown in Figure
3.1:
ON
OFF
INST.
ON OFF
PUMP
RANGE
PPM ZERO
PPM GASURVEYOR 500
Figure 3.1 Instrument Front Plate
13
PPM GASURVEYOR 500 USER HANDBOOK

Mode 1

Switching ON with Button One
provides three gas
ranges, PPM, LEL / Gas, and Search. Note: Geiger feature is not operational in Search or PPM
range when instrument is switched on in normal mode.

Mode 2

Switching ON with Button Two
provides full
autorange mode which, following warm up cycle, automatically changes range between PPM / LEL / GAS
Figure 3.2 displays the warm up cycle for the PPM Gasurveyor 500.
.
14
Note: Battery status is displayed as either OK or LO at start-up.
Figure 3.2 Warm Up
OPERATION

Calibration Date Features

At the end of warm-up and before the PPM Gasurveyor 500 instrument is ready for measuring, the instrument will indicate on the display when the next calibration is due. This will be displayed as month and year, as shown in Figure 3.3:
Calibration Due
February 2002
(Alternate flash)
Figure 3.3 Calibration Date
The re-calibration interval pre-set for all PPM Gasurveyor 500 instruments is twelve (12) months.
This period can be altered as an option, however, you should ensure that the instrument is always within its calibration period prior to use.
An option which ensures that an ‘out of calibration instrument’ is not used, is the automatic switch-off when overdue.

Switching Off the Instrument Pump

A single press of Button One
when the pump is
running turns the pump off and stops sampling. Pressing button one again turns the pump back on.
15
PPM GASURVEYOR 500 USER HANDBOOK

Switching Off

A double press of Button One turns the instrument
off immediately.
The instrument switches off automatically after 30 minutes. ‘OFF’ is shown in the top right hand corner of the display. Pressing any button cancels this automatic switch-off. The instrument will then allow another 30 minutes of use.
This 30 minute period is set as a default.
The switch off time may be altered to any time between 1 and 1000 minutes

Changing Range

Each single press of Button Two changes the range
selected. The display cycles through the available ranges in the order LEL/GAS – PPM – LEL/GAS, etc.
Note: The instrument, by default, auto-ranges between
LEL and Volume Gas.

Activate the Audible Geiger Indication (PPM or Search Range)

The audible Geiger indication can be activated by pressing
and holding Button Two .
A further press and hold of Button Two
the audible level.
A further press and hold of Button Two
visual only Geiger indication.
16
will increase
will return to
OPERATION

Zeroing the ppm or Search Range

A double press of Button Two
zeroes the ppm range.
This should be carried out in fresh air and may take a few minutes to reach the optimum zero stability. The display in Figure 3.4 occurs if a large adjustment is needed to reach the zero point.
Figure 3.4 Zeroing Range
17
Loading...
+ 63 hidden pages