This User Handbook is copyright of Gas Measurement Instruments Ltd (GMI) and
the information contained within, is for use only with
Reproduction, in whole or in part, including utilisation in machines capable of
reproduction or retrieval without written permission of GMI is prohibited. Reverse
engineering is not permitted.
LIABILITY
Every care has been taken in the preparation of this document, but GMI do
not accept any responsibility for errors or omissions and their consequences.
Information in this document is subject to change without notice. This document
does not constitute a specication or basis for a contract. Your statutory rights
under law are not affected.
MODIFICATION NOTICES
GMI aim to notify customers of relevant changes in the product operation and
maintain this manual up to date. In view of the policy of continuous product
improvement there may be operational differences between the latest product
and this manual.
This Handbook is an important part of the
the following points:
• It should be kept with the instrument for the life of the product.
• Amendments should be attached.
• This Handbook should be passed on to any subsequent owner/user of
the instrument.
• Although every care is taken in the preparation of this Handbook it does
not constitute a specication for the instrument.
SOFTWARE
Software supplied on EPROM or similar device for use in a particular product,
may only be used in that product and may not be copied without the written
permission of GMI. Reproduction or disassembly of such embodied programmes
or algorithms is prohibited. Ownership of such software is not transferable and
GMI does not warrant that the operation of the software will be error free or that
the software will meet the customer’s requirements.
DISPOSAL ADVICE
When no longer in use, dispose of the instrument carefully and with respect for
the environment. GMI will dispose of the instrument without charge if returned
to the factory.
GT series
GT series
instruments.
product. Please note
i
Page 4
SAFETY
• The instrument must be regularly serviced and calibrated by fully trained
personnel in a safe area.
• Batteries: Alkaline or *Rechargeable batteries must be exchanged
(*and recharged) in a safe area and tted correctly before use.
Never use damaged batteries or expose to extreme heat.
See Chapter 13 : OPERATOR MAINTENANCE.
• Only GMI replacement parts should be used.
• If the instrument detects gas, follow your own organisation’s procedures
and operational guidelines.
• The combustion chamber is a ameproof assembly and must not be
opened in the presence of a ammable atmosphere.
•
GT series
SIRA 05ATEX II 2 G EEx iad IIB T3
• This equipment is designed and manufactured to protect against other
hazards as dened in paragraph 1.2.7 of Annex II of the ATEX Directive
94/9/EC.
Any right of claim relating to product liability or consequential damage to any
third party against GMI is removed if the warnings are not observed.
WARNING: To prevent ignition of ammable or combustible atmospheres,
remove batteries before servicing.
WARNING: To prevent ignition of ammable or combustible atmospheres, read,
understand and adhere to the manufacturer’s live maintenance procedures.
WARNING: To reduce the risk of ignition of a ammable or explosive atmosphere,
batteries must be changed only in a location known to be non-hazardous.
WARNING: To reduce the risk of explosion, do not mix old batteries with used
batteries or mix batteries from different manufacturers.
WARNING: Never attempt to recharge non rechargeable cells.
CAUTION: Not for use in oxygen enriched atmospheres.
CAUTION: Replace instrument batteries only with approved batteries, as follows:
ATEX / IECEx APPROVED INSTRUMENTS:
• See ‘SPECIAL CONDITIONS FOR SAFE USE’ on the following page.
UL APPROVED INSTRUMENTS:
• Alkaline: Any ‘LR14’ type
• Rechargeable NiMH: Any ‘C’ type
instruments are certied as:
UL 913 Class I, Div 1 Groups C and D.
ii
Page 5
COPYRIGHTUSER HANDBOOK
SPECIAL CONDITIONS FOR SAFE USE
Note: For ATEX / IECEx approved instruments only.
1. When the batteries are replaced, only the following non-rechargeable
cells are permitted:
2. When the batteries are replaced, only the following rechargeable cells
are permitted:
Panasonic 3000mAh Sub-C size, Nickel Metal Hydride cells type
HHR300SCP
AREAS OF USE
Exposure to certain chemicals can result in a loss of sensitivity of the ammable
sensor. Where such environments are known or suspected it is recommended
that more frequent response checks are carried out. The chemical compounds
that can cause loss of sensitivity include Silicones, Lead, Halogens and Sulphur.
Do not use instrument in potentially hazardous atmospheres containing greater
than 21% Oxygen. Do not use in areas where explosive concentrations of
Hydrogen may be present. The enclosure material is polypropylene and must not
be exposed to environments which are liable to result in mechanical or thermal
degradation or to damage caused by contact with aggressive substances.
Additional protection may be required in environments where the instrument
enclosure is liable to damage.
STORAGE, HANDLING AND TRANSIT
Rechargeable batteries contain considerable energy and care should be taken
in their handling and disposal. Batteries should be removed if the instrument
is stored for longer than 3 months. The instrument is designed to handle harsh
environments. The instrument is sealed to IP54. If not subject to misuse or
malicious damage, the instrument will provide many years of reliable service.
The instrument contains electrochemical sensors with a life of 2 years. Under
conditions of prolonged storage the sensors should be removed. The sensor
contains potentially corrosive liquid and care should be taken when handling or
disposing of the sensor, particularly when a leak is suspected.
WARRANTY
The
GT series
of 5 years. Consumable and mechanical parts are not included in this. These
are covered under GMI standard warranty conditions. For details, please
contact GMI Ltd (UK).
instrument has a warranty against faulty goods or workmanship
iii
Page 6
iv
Page 7
CONTENTS
COPYRIGHT ............................................................................ i
LIABILITY ................................................................................. i
Sampling System .................................................... A-2
Power Source .......................................................... A-2
QUICK OPERATING INSTRUCTIONS ...............B-1
INDEX...................................................................... I
x
Page 13
Up
LH
RH
Centre
Down
INTRODUCTION
The GMI
GT series
multi-application gas detectors to suit all the needs of a Gas Industry
Service Technician.
In the following procedure,
the five instrument buttons
shown in Fig 1.2 are referred
to as detailed on following
page:
instruments are designed to be multifunction,
Fig. 1.1
GT series
Instrument
Fig. 1.2 Button References
1
1-1
Page 14
Button Text Reference
LH (INVERT): LH button
Centre (RANGE): Centre button
RH (PUMP): RH button
Up: UP button
Down: DOWN button
The bottom line of the
instrument display may
indicate button press options.
For example, to select ‘Yes’
from the following screen
option, you would press the
RH button. To select ‘No’, the
LH button would be pressed.
The principle of selecting any one of three options on the bottom
line of the instrument display, by pressing either LH, CENTRE or RH
buttons, is used throughout the operation of this instrument.
1.1 INSTRUMENT RANGES
The detection ranges, available but which may not all be included in
the version of GT selected by your company, are:
• 0-10,000 ppm Methane (CH4) Flammable - Sniffer Range
• 0-100% LEL Methane (CH4) Flammable
• 0-100% VOL Methane (CH4) Flammable
• 0-25% Oxygen (O2)
• 0-2000 ppm Carbon Monoxide (CO)
• 0-100 ppm Hydrogen Sulphide (H2S)
• 0 to 60 in. Water Gauge (0 to 150mBar)
Fig 1.3 Option Select
1-2
Page 15
INTRODUCTIONUSER HANDBOOK
Note 1: This instrument is calibrated for Methane (CH4)
PPM & LEL ranges, although calibrated for CH4, will
Avoid exposing the VOL gas range to ammable gases
Note 2: Some sensors respond to gases other than the target
ammable gas.
respond to other ammable gases but will only detect
CH4 accurately.
other than CH4 as this can cause a zero shift on all
ammable ranges that may eventually result in a fault
alarm.
gas. Typically, this cross response is not enough to result
in operational problems, however, should you have any
concerns, please contact GMI.
1-3
Page 16
1-4
Page 17
2
GENERAL FEATURES
• An integral pump draws the required sample over all of the
sensors
• The pump flow is monitored by means of a pressure
transducer
• Datalogging is available for some applications
• A ‘Bleep’ sounds for each button press
• A green LED indicates ‘Power ON’
To simplify use in its many
applications, the instrument
utilises a menu system similar to
a (cell) mobile phone, as shown
in Fig 2.1.
Fig. 2.1 Mode Menu
The bottom line of the LCD can also be used to indicate press and
hold button functions available.
By simple selection of the application you need, the appropriate
gas(es) are shown on the display and only the necessary button
functions and alarms are activated.
Note: Your company may have decided to have its instruments
congured only for certain of the aforementioned applications.
2-1
Page 18
2.1 MODES OF OPERATION
Leak Test: This mode is for the technician to investigate odour or
See Chapter 4 for ‘Leak Test’ mode operation.
Conned Space: Used for conned space pre-entry testing and for
See Chapter 5 for ‘Conned Space’ mode operation.
Barhole: Used to locate underground leaks. This can be timed or
See Chapter 6 for ‘Barhole’ mode operation.
CO: The CO (Carbon Monoxide) mode allows the user to make
See Chapter 7 for ‘CO’ mode operation.
Purge: This mode is used in gas and air purging applications.
See Chapter 8 for ‘Purge’ mode operation.
Sniffer: This mode is used to nd small tting leaks. Very fast
See Chapter 9 for ‘Sniffer’ mode operation.
Pressure: In this mode, the instrument can be used as a manometer
See Chapter 10 for ‘Pressure’ mode operation.
Bump Test: This mode allows the user to set up gases, apply gases, log
See Chapter 11 for ‘Bump Test’ mode operation.
leak complaints and to pinpoint the leak.
personal monitoring in areas such as basements etc.
non-timed. Your company may have selected only one of
these options.
either CO Direct, Differential CO, or Air Free CO readings (if
an O2 cell is tted). Note: Your company may have chosen
only certain of these options.
detection rates are achieved using the semiconductor
sensor in the probe.
to measure appliance and regulator pressure and to check
the system for leaks.
the readings automatically and view the logged information.
By default, this mode is disabled.
2-2
Page 19
OPERATING PROCEDURE
Check the following:
• The instrument is clean and in good condition.
• The batteries are in good condition, fully charged and tted
correctly.
• The hydrophobic lter is clean and in good condition.
• The sample line and any other accessories used are in good
condition.
• The ‘clear bulb’ at probe end is screwed tightly.
• Switch instrument ON in fresh air.
• The battery indication provides sufcient capacity for the
operation.
• The instrument is within the calibration period you have
decided is necessary for your application.
• All applicable ranges are operational.
• There are no fault indications.
• Attach optional accessories, as required.
• The instrument displays a sample / ow fault when inlet is
blocked (with a nger for example). If fault is not displayed,
check tightness of all ttings.
• After use, allow the instrument to run for 1 – 2 minutes in
fresh air before switching the instrument OFF.
3
3-1
Page 20
Each time you use the instrument, carry out the following procedure:
3.1 SWITCH THE INSTRUMENT ON
To switch the instrument ON in fresh air:
• Press and hold the RH button , for one second.
Fig. 3.1 Switch ON
The instrument begins its warm-up routine, which lasts approximately
30 seconds. During the warm-up cycle, a countdown timer appears
in the top right hand corner of the display.
The Power ON LED illuminates when the instrument is switched
ON and also during operation. The display backlight illuminates and
remains ON during warm-up. When the warm-up cycle is complete,
the screen light automatically switches OFF.
The Fault LED illuminates briey, for approximately ve (5) seconds,
during the warm-up cycle.
3.2 INSTRUMENT IDENTIFICATION
During the warm-up cycle, the
instrument display identies the
model, serial number, software
version and battery status
information as shown in Fig. 3.2:
3-2
Fig. 3.2 Instrument I/d
Page 21
OPERATING PROCEDUREUSER HANDBOOK
100%
75%
50%
25%
3.3 BATTERY STATUS
This feature provides the user
with a battery capacity level
indicator that displays instrument
battery power remaining, as
shown in Fig. 3.3.
This battery symbol will be
indicated for approximately ve
(5) seconds during the warmup cycle, then on the top of the
display during normal operation.
Fig. 3.3 Battery Capacity
3.4 FILTER CHECK / FLOW FAULT TEST
Checking lters are in place and in good condition and performing a
regular leak check are pertinent to effective use of the GT instrument,
therefore, the user will be asked to conrm these checks before the
instrument completes the warm-up sequence. By default, these
checks should be performed on a daily basis.
3.4.1 Filter Check
If filters are installed, and in
good condition, press ‘YES’ for
instrument warm-up to continue.
If ‘NO’ is selected, instrument
will automatically enter ‘switch
off’ sequence.
3.4.2 Flow Fault Test
To perform a ow fault test, block probe tip inlet with nger for
approximately 5 seconds. Pump should ow fault and display screen
illustrated in Fig. 3.6.
If ‘YES’ is selected, instrument
warm-up will continue without
performing test.
If ‘NO’ is selected, instrument
will automatically enter ‘switch
off’ sequence.
Fig. 3.4 Filter Check
Fig. 3.5 Flow Fault Test
3-3
Page 22
A successful ow fault test is
confirmed, as shown in Fig.
3.6, by ‘successful’ ashing in
the display.
If the instrument fails the ow
fault test, the screen shown in
Fig. 3.5 is displayed.
The instrument stores a log of
a successful test. Likewise, if
‘YES’ was selected, in Fig. 3.5,
without performing a ow fault
test, a log is also recorded.
On completion, select ‘YES’
to switch the pump ON and
continue warm-up cycle.
alternating with
Fig. 3.6 Successful Test
3.5 TIME AND DATE
The time and date from the
instrument’s built-in clock is
displayed on the screen during
warm-up, as shown in Fig. 3.7.
If datalogging is being used, the
time and date is set from this
clock. This may be important
when viewing the logged data.
Fig. 3.7 Time and Date
3.6 CALIBRATION DUE DATE
The CAL DUE date can be set by the workshop and is set to 365
days by default. The date can be set from 1 to 400 days.
The CAL DUE is reset when the instrument is successfully calibrated.
The CAL DUE feature has ve (5) congurable options:
1. Cal Due Date message is not displayed.
2. Cal Due Date and overdue date messages are displayed.
3-4
Page 23
OPERATING PROCEDUREUSER HANDBOOK
Calibration due date is
displayed, as shown in
Fig. 3.8.
Fig. 3.8 Calibration Due Date
The screen, shown in Fig.
3.9, is displayed when
the Calibration date has
expired. i.e. overdue.
After approximately five
seconds, the instrument
warm-up continues.
3. Cal Due Date message is displayed with user acknowledge
if overdue (Default).
Calibration due date is
displayed as shown in Fig.
3.10.
The screen, shown in Fig.
3.11, is displayed when
the Calibration date has
expired. i.e. overdue.
Fig. 3.9 Calibration Overdue
Fig. 3.10 Calibration Due Date
Fig. 3.11 Calibration Overdue
3-5
Page 24
The user must acknowledge that Calibration has expired.
To continue:
• Press and hold YES for instrument warm-up to continue.
To switch OFF:
• Press and hold NO .
The screen, shown in Fig.
3.12, is displayed.
Fig. 3.12 Switch OFF
• Press and hold both the LH button and the RH button
simultaneously to proceed with the switch-off sequence.
4. Cal Due Date message is displayed with user acknowledge
for extended period option, if overdue.
Calibration Due Date
message is displayed, as
shown in Fig. 3.13.
Fig. 3.13 Calibration Due Date
If overdue but within the
‘extended period’, the
screen, shown in Fig. 3.14,
is displayed.
3-6
Fig. 3.14 Calibration Overdue
Page 25
OPERATING PROCEDUREUSER HANDBOOK
The user must acknowledge that Calibration has expired.
Note: The extended period can be set from 1 to 31 days
To accept ‘extended period’ option:
• Press and hold YES and the instrument warm-up continues.
Note: When the extended period option expires, the user will be
forced to switch the instrument OFF.
To reject ‘extended period’ option:
• Press and hold NO .
The screen, shown in Fig.
3.15, is displayed.
Fig. 3.15 Switch OFF
• Press and hold both the LH button and the RH button
simultaneously to proceed with the switch-off sequence.
5. Cal Due Date message is displayed with user shut-down
if overdue.
Calibration due date is
displayed, as shown in Fig.
3.16.
Fig. 3.16 Calibration Due Date
The screen, shown in Fig.
3.17, is displayed when the
Calibration date has expired.
i.e. overdue.
Fig. 3.17 Switch OFF
3-7
Page 26
To proceed with the switch-off sequence.
• Press and hold both the LH button and the RH button
simultaneously.
3.7 SERVICE DUE DATE
The Service due date can be set by the workshop and is set to two
(2) years by default from last service date. The date can be set over
a period of 1 to 36 months in 1 month steps.
Note : The service due date will not normally be displayed at
start up but if display is selected, it will be shown at 90
days prior to the preset date, as shown in Fig. 3.15.
The SERVICE DUE feature has ve (5) congurable options:
1. Service Due Date message is not displayed (Default).
2. Service Due Date and overdue date messages are displayed,
as shown in Fig. 3.18.
Fig. 3.18 Service Due Date
The screen, shown in Fig.
3.19, is displayed when the
Service date has expired.
i.e. overdue.
After approximately five
seconds, the instrument
warm-up continues.
Fig. 3.19 Service Overdue
3-8
Page 27
OPERATING PROCEDUREUSER HANDBOOK
3. Service Due Date message is displayed with user
acknowledge if overdue.
Service due date is
displayed, as shown in
Fig. 3.20.
Fig. 3.20 Service Due Date
The screen, shown in
Fig.3.21, is displayed
when the Service date has
expired, i.e. overdue.
Fig. 3.21 Service Overdue
The user must acknowledge that Service has expired.
To continue:
• Press and hold YES for instrument warm-up to continue.
To switch OFF:
• Press and hold NO .
The screen, shown in Fig.
3.22, is displayed.
Fig. 3.22 Switch OFF
• Press and hold both the LH button and RH button
simultaneously to proceed with the switch-off sequence.
3-9
Page 28
4. Service Due Date message is displayed with user
acknowledge for extended period option, if overdue.
Service Due Date message
is displayed, as shown in
Fig. 3.23.
Fig. 3.23 Service Due Date
If overdue but within the
‘extended period’, the
screen, shown in Fig. 3.24,
is displayed.
Fig. 3.24 Service Overdue
The user must acknowledge that Service has expired.
Note: The extended period can be set from 1 to 31 days
To accept ‘extended period’ option:
• Press and hold YES and the instrument warm-up continues.
Note: When the extended period option expires, the user will
be forced to switch the instrument OFF.
To reject ‘extended period’ option:
• Press and hold NO .
The screen shown in Fig.
3.25 is displayed.
3-10
Fig. 3.25 Switch OFF
Page 29
OPERATING PROCEDUREUSER HANDBOOK
• Press and hold both the LH button and the RH button
simultaneously to proceed with the switch-off sequence.
5. Service Due Date message is displayed with user shut-
down if overdue.
Service due date is
displayed, as shown in
Fig. 3.26.
Fig. 3.26 Service Due Date
The screen, shown in Fig.
3.27, is displayed when the
Service date has expired,
i.e. overdue.
Fig. 3.27 Switch OFF
To proceed with the switch-off sequence:
• Press and hold both the LH button and the RH button
simultaneously.
3.8 SENSORS ZEROING
Before warm-up is completed, the instrument automatically ensures
that the sensors are zeroed before operation.
The screen shown in Fig.
3.28 is displayed during this
process.
Fig. 3.28 Zero Sensors
3-11
Page 30
Warm-up Complete
The instrument will now automatically select the Leak Test Mode,
as default.
The following congurable options are available:
a) To start up in specic operational mode.
b) To start up in the mode last used.
3.9 SWITCH THE INSTRUMENT OFF OR RE-ENTER
THE MODE MENU
To initiate the shut down sequence:
• Press and hold both the LH and RH buttons
simultaneously.
Fig. 3.29 Switch OFF
While both buttons are depressed, the instrument display will step
through previous menus, each displayed for approximately two (2)
seconds.
3-12
Page 31
OPERATING PROCEDUREUSER HANDBOOK
After this time, the OFF sequence countdown begins and the user
will have to keep the buttons pressed for a further three (3) seconds
to switch the instrument OFF. The countdown sequence is shown
in Fig. 3.30.
Note: In Conned Space Mode, to prevent inadvertently switching
instrument OFF or changing mode while alarms are active,
the user must press and hold both the LH and RH
buttons simultaneously for an additional ve (5) seconds
before the mode menu appears.
Fig. 3.30 OFF sequence
The switching OFF sequence
can be aborted at any time
by releasing the held buttons
which provides access to the
Mode menu, shown in Fig.
3.31:
Fig. 3.31 Mode Menu
3-13
Page 32
To highlight the required option:
• Press the UP or DOWN buttons, then . .
To select the highlighted option:
• Press and hold OK .
3-14
Page 33
4
LEAK TEST MODE
This mode is for the technician to investigate odour or leak complaints
and to pinpoint the leak.
Note: The instrument pump must be switched ON to measure
in Leak Test mode.
4.1 LEAK TEST RANGES
Leak Test mode will have the following ranges available for normal
operation:
• 0 – 2000 ppm Methane (CH4) Flammable with following
options:
(to display as LEL with 0.1% resolution up to 9.9%).
(to display without numerical indication).
(to display zero until gas reading exceeds 100 ppm).
• 0 – 100% LEL Methane (CH4) Flammable with option:
(to display as gas in air).
• 0 – 100% VOL Methane (CH4) Flammable.
• 0 – 2000 ppm Carbon Monoxide (CO) if tted.
4.2 LEAK TEST FEATURES
Leak Test mode will have the following features available:
• Audible / Visual PPM Ticker (Geiger) Indication
• Audible / Visual Gas Alarms
• Backlight
• Flashlight
• Pump - can be toggled ON / OFF and will stop if ow fault
is detected
• Max Display
• Autoranging from ppm-LEL-Gas
• Manual Range selection between Flam and Flam plus CO
(Dual Display)
• Threshold adjustment
• Display Invert
• Automatic datalogging
4-1
Page 34
4.3 LEAK TEST DISPLAYS
During the warm-up period,
all applicable sensors will be
checked and the display will
indicate any sensor faults.
A positive zero fault is indicated
by a flashing gas reading, as
illustrated in Fig. 4.1.
A negative zero fault is indicated
by a spanner alternating with
a zero reading (not illustrated).
After the sensors have been
checked, the normal display will
be as shown in Fig. 4.2.
Note 1: Any faulty sensor will
continue to display a spanner
(wrench) symbol alternating with
the gas reading and the fault LED
will illuminate.
Note 2: Continuous display of
both flammable and CO is a
congurable option, as shown
in Fig. 4.3.
alternating with
Fig. 4.1 Sensor Check
Fig. 4.2 Normal Display
Example
4-2
Fig. 4.3 Flammable / CO
Display
Page 35
LEAK TEST MODEUSER HANDBOOK
4.3.1 Ticker (Geiger)
To enable / disable the audible Ticker (Geiger):
(in the PPM range)
• Press and hold the UP and DOWN buttons
simultaneously, as follows:
1. First press and hold - to enable visual only.
2. Second press and hold - to disable both audible and visual.
3. Third press and hold - to enable both audible and visual.
When the PPM autoranges to
LEL, the display changes to the
example shown in Fig. 4.4.
Fig. 4.4 LEL Display
If displaying the maximum
reading, the display will be as
the example shown in Fig. 4.5.
Note: Maximum display only
for LEL and VOL.
Fig. 4.5 Max Reading
To return the display to normal (live) operation:
• Press and hold LIVE .
Note: If the LH button is not pressed, the display will return to
normal after 30 seconds.
4-3
Page 36
If a CO sensor is included in
your instrument and the CO
concentration rises above a preset level, the display will change
together with an audible alarm
to attract your attention. If alarm
is enabled, CO alternates with
HIHI until cleared.
To acknowledge the alarm:
• Press and hold ACK .
After acknowledgement of the
alarm, the screen, shown in Fig.
4.7, is displayed.
When CO levels are below alarm concentration, the Centre button
enables return to ammable only. This button can be pressed
at any time to change between Flammable and Flammable + CO.
alternating with
Fig. 4.6 CO Alarm
Fig. 4.7 Alarm Acknowledge
4-4
Page 37
ACTION
PRESS
PRESS
and
HOLD
LH
(INVERT)
FLAM PPM
ZERO
BACKLIGHT /
FLASHLIGHT
ACKNOWLEDGE
ALARM
MAX / LIVE
RANGE
FLAM
FLAM + CO
INVERT
DISPLAY
RH
(PUMP)
CENTRE
(RANGE)
PUMP
ON / OFF
CLEAR FLOW
FAULT
LEAK TEST MODEUSER HANDBOOK
4.4 LEAK TEST BUTTON OPERATION
A summary of the button operation is detailed in Table 4.1:
Table. 4.1 Button Operation in Leak Test Mode
4.5 DESCRIPTION OF LEAK TEST BUTTON
OPERATION
4.5.1 Ticker (Geiger) ON / OFF
To enable and disable the audible Ticker (Geiger):
(in the PPM range)
• Press and hold the UP and DOWN buttons
simultaneously, as follows:
1. First press and hold - to enable visual only.
2. Second press and hold - to disable both audible and visual.
3. Third press and hold - to enable both audible and visual.
When PPM autoranges to LEL, the Ticker (Geiger) audible/visual
continues until an LEL alarm is reached. The Ticker (Geiger) is then
switched off and the LEL alarm will be activated. If no LEL alarm is
set, the Ticker (Geiger) will remain activated.
4-5
Page 38
4.5.2 Ticker (Geiger) Adjust
To adjust the Ticker (Geiger) threshold:
To increase
• Press and hold the UP button.
To decrease
• Press and hold the DOWN button.
The threshold setting is
displayed briey beneath the
PPM CH4. In the example
opposite, Fig. 4.8, the threshold
is 500.
Fig. 4.8 Ticker (Geiger)
4.5.3 Range Selection
The instrument initially selects the PPM range which autoranges to
LEL/VOL.
Note: When CO range is selected the display reverts to the 2-gas
display.
A congurable option in Leak Test mode enables user to manually
select between PPM, and autoranging between LEL & Volume.
By default, the reading will show LEL (autoranging to volume).
• Press the Centre button , to toggle to PPM.
If a CO range is present in the GT, and this option is enabled, it is no
longer possible to toggle between one and two ranges on the display.
It will always be the ammable range and the CO range.
4.5.4 Lights
To switch ON the backlight:
• Press and hold the Centre button .
Threshold
4-6
Page 39
LEAK TEST MODEUSER HANDBOOK
To switch ON the ashlight:
• Press and hold the Centre button again.
The backlight and ashlight are both timed to switch OFF after two
minutes. Both can be switched OFF by a third press and hold of the
Centre button .
4.5.5 Max / Live
To show the Maximum readings:
(since the mode was selected)
• Press and hold the LH button .
To return display to the live reading.
• Press and hold the LH button again.
Note 1: If the instrument is in alarm, the Max function cannot be
used until the alarm is Acknowledged.
Note 2: If no button is pressed, after 30 seconds the display reverts
to live reading.
4.5.6 Alarm Acknowledge
Non-latching can be muted for one (1) minute or, when the
concentration drops below the alarm level, the alarm will automatically
cancel.
Latched Alarms can be muted temporarily for one minute or cancelled
permanently when the display level is below alarm level.
To acknowledge / mute latched alarms:
• Press and hold the LH button .
4-7
Page 40
4.5.7 Pump
To switch the pump ON and OFF:
• Press the RH button .
4.5.8 Zero
To zero the Flammable PPM range when displayed:
• Press and hold the RH button .
Note: The pump must be switched ON to zero the ammable ppm
range.
4.5.9 Flow Fault
If ow fault is detected, the pump stops automatically. The instrument
should be checked for water ingress or blockage and the pump reset.
To clear the ow fault:
(Once the blockage has been cleared):
• Press the RH button .
4.5.10 Menu / Off
To re-select the Mode Menu:
• Press and hold both the LH button and the RH button
simultaneously
When the menu appears on the screen, release the buttons,
otherwise the instrument will proceed into the switch OFF
process and will switch OFF after a further three (3) seconds.
.
4-8
Page 41
LEAK TEST MODEUSER HANDBOOK
4.6 LEAK TEST TICKER (GEIGER) FUNCTION
The Ticker (Geiger) function is available on ammable ppm range.
The Ticker (Geiger) range at start-up is 0-1000 PPM.
For any subsequent PPM concentration, the Ticker (Geiger) audible
/ visual can be ‘backed off’ using the UP or DOWN buttons.
e.g. 600 PPM Ticker (Geiger) can be ‘backed off’ to enable tracing
to higher concentrations, in which case the Ticker (Geiger) range
becomes 600 to 1600 PPM etc, etc.
Note: The display range remains always as 0 to 2000 PPM.
The Ticker (Geiger) visual is such that the lights illuminate in pairs
according to Fig. 4.9 over any 1000 PPM range as explained above.
The Ticker (Geiger) function can be set for audible and visual
combination options as follows: Both ON; Audible OFF / Visual ON;
Both OFF; Both ON; etc.
To select required audible / visual combination option:
• Press and hold both UP and DOWN buttons.
• Repeat operation to cycle through combination options.
Fig. 4.9 PPM Level / LED Sequence
4-9
Page 42
4.7 LEAK TEST ALARMS
Refer to Chapter 11.
4.8 LEAK TEST LOGGING
Automatic datalogging is active for CO and CH4 ranges (except ppm
CH4). Timed logs will be recorded every minute (default) or as per
conguration.
4-10
Page 43
5
CONFINED SPACE MODE
This mode is used for conned space pre-entry testing and for
personal monitoring in areas such as basements etc.
5.1 CSM RANGES
Conned Space Mode (CSM) will have the following ranges available:
• 0 – 100% LEL Methane (CH4) Flammable
(EEE Over-range) and option to display up to 9.9% LEL with
0.1% resolution
• 0 – 25% Oxygen (O2) if tted
• 0 – 2000 ppm Carbon Monoxide (CO) if tted
• 0 – 100 ppm Hydrogen Sulphide (H2S) if tted
5.2 CSM FEATURES
Conned Space mode will have the following features available:
• Audible / Visual Alarms.
• Min / Max Display.
• Backlight.
• Flashlight.
• Pump – ON continuously - will not stop if ow fault is detected
- ow fault alarm will be activated.
• Datalogging – automatic every 60 seconds.
• Manual Logging is also available. This will log all gas
concentrations at that instant.
• Time weighted averaging for Alarms.
• Condence signal every 15 seconds.
• Display invert.
5-1
Page 44
5.3 CSM DISPLAYS
During the warm-up period,
all applicable sensors will be
checked and the display will
indicate any sensor faults.
A positive zero fault is indicated
by a flashing gas reading, as
illustrated in Fig. 5.1.
A negative zero fault is indicated
by a spanner alternating with
a zero reading (not illustrated).
When an instrument has an
alarm active, the bottom line of
the display changes. A sample
fault will change both top and
bottom lines, as shown in
Fig. 5.2:
alternating with
5.1 Four Gas Sensor Check
alternating with
5-2
Fig. 5.2 Four-gas Display with
Alarm and Sample Fault
Page 45
CONFINED SPACE MODEUSER HANDBOOK
When viewing Max in conned
space mode, Confined Max
identifier is in the top of the
display, with the option of
selecting Min on the bottom line,
as shown in Fig. 5.3:
Fig. 5.3 Conned Max Identier
When viewing Min in conned
space mode, Confined Min
identifier is in the top of the
display, with the option of
selecting Live on the bottom
line, as shown in Fig. 5.4:
Fig. 5.4 Conned Min Identier
5.4 CSM BUTTON OPERATION
A summary of the button operation is detailed in Table 5.1:
Table. 5.1 Button Operation in Conned Space Mode
5-3
Page 46
5.5 DESCRIPTION OF CSM BUTTON OPERATION
5.5.1 Lights
To switch ON the backlight:
• Press and hold the Centre button .
To switch ON the ashlight:
• Press and hold the Centre button again.
The backlight and ashlight are both timed to switch OFF after
two minutes or can be switched OFF by a third press and hold the
Centre button .
5.5.2 Min / Max
To show the Maximum gas readings:
(since the current mode was selected)
• Press and hold the LH button .
To show the Minimum gas readings:
(since the current mode was selected)
• Press and hold the LH button again.
A third press and hold the LH button returns to the live reading.
Note1: If the instrument is in alarm, the Min / Max function cannot
be used until the alarm is Acknowledged.
Note 2: If no button is pressed, after 30 seconds the display reverts
to live reading.
5-4
Page 47
CONFINED SPACE MODEUSER HANDBOOK
5.5.3 Alarm Acknowledge
HIHI, LOLO for O2 (if tted), and Time Weighted Average (TWA)
alarms are latching by default. The AL / ACK will only cancel the
alarms if all gases have returned to “safe” levels.
An option of HI, or LO for O2 (if tted), non latching alarms are
available, as follows:
When the reading drops below the alarm level, the alarm will
automatically cancel.
Alarms can also be muted for one minute.
To acknowledge for one minute:
• Press and hold the LH button .
If an alarm occurs in CSM, the user should go into fresh air until
the gas concentration falls below the alarm level and is cancelled.
Menu mode can still be entered while an alarm is active in CSM.
To enter menu mode:
• Press and hold the LH button and the RH button
simultaneously, for 5 seconds.
5.5.4 Flow Fault
Before clearing a ow fault alarm the instrument must be checked
for water ingress or blockage.
To clear a Flow Fault alarm:
• Press the RH button .
5.5.5 Manual Log
A manual log can be taken at any time.
To capture a manual log:
• Press and hold the RH button .
5-5
Page 48
Note: When a manual log is taken, the term ‘LOG’ is inversed on
the display for one (1) second to provide a visual conrmation
that the log has been captured.
5.5.6 Menu / Off
To re-select the Mode Menu:
• Press and hold both the LH button and RH button .
When the menu appears on the screen, release the buttons,
otherwise the instrument will proceed into the switch OFF
process and will switch OFF after a further three (3) seconds.
Note: In Conned Space Mode, to prevent inadvertently switching
instrument OFF or changing mode while alarms are active,
the user must press and hold both the LH and RH
buttons simultaneously for an additional ve (5) seconds
before the mode menu appears.
5.6 CSM PUMP
The pump will run continuously for safety reasons.
Flow Fault will be agged and fault LED illuminated.
To clear the ow fault:
• Press and hold the RH button .
5.7 CSM ALARMS
Refer to Chapter 11.
5.8 CSM LOGGING
Automatic datalogging is active for all ranges available. Timed logs
will be recorded every minute (default) or as per conguration.
Manual Logging is also available. This will log all gas concentrations
at that instant.
5-6
Page 49
CONFINED SPACE MODEUSER HANDBOOK
5.9 CSM CONFIDENCE SIGNAL
During normal operation, the instrument sounds a condence beep
and illuminates the bottom pair of red LED’s briey every 15 seconds.
This function is programmable in the instrument setup software.
The condence signal function makes the user aware that the
instrument is operating correctly:
Note: The condence beep and / or LED indication can be disabled.
Refer to the ‘SET-UP SOFTWARE USER HANDBOOK’ for further
information.
5-7
Page 50
5-8
Page 51
6
BARHOLE TESTING MODE
This mode is used to locate underground leaks and on start-up will
be ranged to 0-100% LEL Methane (CH4) Flammable which will
autorange to VOL Methane (CH4) Flammable at 100% LEL.
6.1 BARHOLE RANGES
Barhole testing will have the following ranges available:
• 0 – 100% LEL Methane (CH4) Flammable with the
following option:
(to display with 0.1% resolution up to 9.9%)
• 0 – 100% VOL Methane (CH4) Flammable
6.2 BARHOLE FEATURES
Barhole mode will have the following features:
• Timed or Non-Timed Sampling.
• Backlight.
• Flashlight.
• Pump control.
• Invert Display.
• Six sets of barhole readings with Viewing / Overwriting.
6-1
Page 52
6.3 BARHOLE DISPLAYS
During the warm-up period,
all applicable sensors will be
checked and the display will
indicate any sensor faults.
A positive zero fault is
indicated by a flashing gas
reading, as illustrated in Fig.
6.1.
A negative zero fault is
indicated by a spanner
alternating with a zero reading
(not illustrated).
Both timed and non-timed
modes are available to the
user, as shown in display
Fig. 6.2.
USER HANDBOOK
alternating with
Fig. 6.1 Sensor Check
Fig. 6.2 Timed / Non-Timed Mode
To highlight the required option:
• Press the UP or DOWN buttons, then . .
To select the highlighted option:
• Press and hold OK .
Note: The option initially highlighted will be that previously
selected.
Selection
6-2
Page 53
BARHOLE TESTING MODE
6.3.1 Timed Mode
If TIMED is selected, and the instrument conguration allows user
selectable barhole sample time, the screen shown in Fig. 6.3 is
displayed. The sample time previously set is displayed.
If the instrument conguration
does not allow user selectable
barhole sample time, the screen
shown in Fig. 6.4 is displayed.
Fig. 6.3 Timed Mode
To change the sample time:
• Press the UP or DOWN buttons.
Sample time range = 10 to 300 seconds.
Holding the button will perform a fast ramped change. When
300 is reached, the range rolls over to 10.
To accept:
• Press and hold OK .
When the time is set or has been
pre-selected at 30 seconds
for example, the instrument
performs a purge, as shown
in Fig 6.4.
Fig. 6.4 Purge
The minimum purge time is ten (10) seconds, therefore, the ‘stop’
option is not displayed for the rst ten (10) seconds.
Note: The timer starts at 0 and when 999 is reached, the timer
rolls over to 0.
6-3
Page 54
USER HANDBOOK
Stop purge when the live readings reach zero:
To stop purge:
• Press and hold OK .
The first barhole
screen is displayed
as shown in Fig. 6.5.
Fig. 6.5 Barhole Timer
Up to six barhole readings can be stored. These are identied as
‘Barhole 1’ to ‘Barhole 6’.
After the rst reading is stored (as Barhole 1), the second reading
(Barhole 2) will be automatically selected, however, by using the
UP or DOWN buttons the user can specify where the next
reading will be stored. This may be useful if there was a problem
with the reading, e.g. sample fault.
The sequence of events will be as follows:
• Press and hold START to initiate timer and pump. Once
started, the user cannot stop a sample.
• Counter decrements from initial value.
• At ‘zero-time’, the
pump switches OFF
and the peak and
final sustained
(actual) readings are
displayed as shown in
Fig. 6.6.
Fig. 6.6 Barhole Numbering
6-4
Page 55
BARHOLE TESTING MODE
• Between Barhole tests there is a mandatory ‘Purge’ mode
to ensure that any gas in the instrument is cleared before
the next barhole is sampled.
• During purge, the peak
reading is reset to zero
and blanked from
display, as shown in
Fig. 6.7.
Fig. 6.7 Purge Complete
6.3.2 Non Timed Mode
If NON TIMED is selected
or has been pre-selected
in configuration then the
instrument performs a purge,
as shown in Fig. 6.8:
Fig. 6.8 Purge
The minimum purge time is ten (10) seconds, therefore, the ‘stop’
option is not displayed for the rst ten (10) seconds.
Note: The timer starts at 0 and when 999 is reached, the timer
rolls over to 0.
Stop purge when the live readings reach zero:
To stop purge:
• Press and hold OK .
6-5
Page 56
USER HANDBOOK
The rst barhole screen
is then displayed, as
shown in Fig. 6.9
Fig. 6.9 Non Timed Mode of
Operation
Up to six barhole readings can be stored. These are identied as
‘Barhole 1’ to ‘Barhole 6’.
After the rst reading is stored (as Barhole 1), the second reading
(Barhole 2) will be automatically selected, however, by using the UP
or DOWN buttons the user can specify where the next
reading will be stored. This may be useful if there was a problem
with the reading, e.g. sample fault.
The sequence of events will be as follows:
• Press and hold START to initiate. This will switch the
pump ON. (START on bottom line of LCD will change to
STOP).
• Counter counts up from zero.
• Press and hold STOP when sampling is complete. The
pump will be switched OFF.
• View as per TIMED MODE.
• Purge mode as per TIMED MODE.
• From switch-on, each barhole is consecutively numbered
automatically.
6-6
Page 57
6.4 VIEW BARHOLE RESULTS
To view previous barhole results:
(to a maximum of six barhole tests)
• Press and hold VIEW
.
Fig. 6.10 View Barhole Results
To return the display to normal operation:
• Press and hold LIVE .
If a *Flow Fault or **Bead Fault
is detected during a barhole
test, the pump will stop and
the test will be aborted with
indication as shown in Fig. 6.11
or Fig. 6.12 respectively.
BARHOLE TESTING MODE
Fig. 6.11 Flow Fault
Fig. 6.12 Bead Fault
* Refer to paragraph 6.6.3 for explanation of Flow Fault.
** Refer to paragraph 6.6.4 for explanation of Bead Fault.
6-7
Page 58
USER HANDBOOK
LH
(INVERT)
RH
(PUMP)
CENTRE
(RANGE)
ACTION
PRESS
INVERT
DISPLAY
START / STOP /
PURGE
VIEW
BACKLIGHT /
FLASHLIGHT
PRESS
and
HOLD
CLEAR FLOW
FAULT
6.5 BARHOLE OPERATION
A summary of the button operation is detailed in the following table:
Table. 6.1 Button Operation in Barhole Mode
6.6 DESCRIPTION OF BARHOLE BUTTON
OPERATION
6.6.1 View
To view previous barhole results:
(to a maximum of six (6) barhole tests)
• Press and hold VIEW .
6.6.2 Lights
To switch ON the backlight:
• Press and hold the Centre button .
To switch ON the ashlight:
• Press and hold the Centre button again.
The backlight and ashlight are both timed to switch OFF after one
minute. Both can be switched OFF by a third press and hold the
Centre button .
6-8
Page 59
BARHOLE TESTING MODE
6.6.3 Flow Fault
If a ow fault is detected, the pump stops automatically.
If a sample was in progress then current sample is halted and ‘Purge’
cycle will be the next part of the sequence.
If ow fault is indicated, the instrument should be checked for water
ingress or blockage.
To clear the ow fault:
(Once the blockage has been cleared)
• Press the RH button .
6.6.4 Bead Fault
If a bead fault is detected, the pump stops automatically.
If a sample was in progress then current sample is halted and ‘Purge’
cycle will be the next part of the sequence.
A persistent bead fault indicates a faulty sensor and therefore the
instrument should be returned for service / repair.
6.6.5 Menu / Off
To re-select the Mode Menu:
• Press and hold both the LH button and the RH
button .
While both buttons are depressed, the instrument display
will step through previous menus, each displayed for
approximately two (2) seconds.
When the menu appears on the screen, release the buttons,
otherwise the instrument will proceed into the switch OFF
process and will switch OFF after a further three (3) seconds.
6-9
Page 60
6.7 BARHOLE ALARMS
There are no Gas Alarms in this mode.
USER HANDBOOK
6-10
Page 61
7
CO (CARBON MONOXIDE) MODE
The CO mode is used to check the interior of premises and appliances
for CO leakage.
7.1 CO RANGES
The CO mode can have the following ranges available:
• 0 – 2000 ppm Carbon Monoxide (CO)
• 0 – 25% Oxygen (O2) if tted
7.2 CO FEATURES
CO mode has the following features available:
Four possible display options (CO Direct / Differential / Air Free /
Viewing).
• Backlight
• Flashlight
• Datalogging (automatic every 60 seconds)
• Invert Display
• Manual Logging is also available. This will log all gas
concentrations at that instant
• Pump ON / OFF
7.3 CO MENU
On selection of CO mode, all four
of the CO sub-modes shown in
Fig. 7.1 can be available to the
user (See note 3).
Fig.7.1 CO Menu
7-1
Page 62
To highlight the required option:
• Use the UP and DOWN buttons.
To select the highlighted option:
• Press and hold OK .
Note 1: The highlighted option will be that previously used.
Note 2: If an O2 sensor is not tted then the Airfree Option is
automatically unavailable.
Note 3: Your company may have decided to have its instruments
congured only for certain of the above options. If only one option is
selected, then after selecting CO mode from the main menu, either
7.4.1, 7.4.2 or 7.4.3 will be entered directly.
7.4 CO DISPLAYS
During the warm-up period,
all applicable sensors will be
checked and the display will
indicate any sensor faults.
A positive zero fault is indicated
by a ashing gas reading, as
illustrated in Fig. 7.2.
A negative zero fault
is indicated by a spanner
alternating with a zero reading
(not illustrated).
alternating with
7-2
Fig. 7.2 Sensor Fault
Page 63
CO (CARBON MONOXIDE MODE)USER HANDBOOK
7.4.1 CO Direct
In this option, normal
atmospheric air is checked for
CO content.
The typical display for this
mode is shown in Fig. 7.3:
Fig. 7.3 CO Direct
7.4.2 Differential CO
This option enables the user to zero out ambient CO backgrounds and
display the differential measurement. This is useful in areas where
the ambient CO is high from other sources such as trafc pollution.
Note: It is possible to have negative readings on the display in this
mode.
The typical display for this
option is shown opposite, in
Fig. 7.4
Fig. 7.4 Differential CO
To zero the displayed CO reading:
• Press and hold ZERO .
7.4.3 Air-Free CO
Air-free CO is the CO reading modied by the O2 reading, therefore
this measurement will only be available when an O2 sensor is tted.
The purpose is to determine whether or not the emissions from an unventilated appliance, principally an oven or a stove / cooker, are safe.
7-3
Page 64
This is the CO reading modified by the O2 reading
(only applicable if O2 sensor is tted).
CO Air-free = (20.9 x CO) / ( 20.9 – O2).
The typical display for this
mode is shown opposite, in
Fig. 7.5:
A sample is considered
invalid until the Oxygen (O2)
concentration is below 19%
Vol.
Changes to Fig 7.6 when valid
sample is taken.
7.4.4 CO Viewing
This option allows the user to view the six (6) most recent manually
logged readings.
If this mode is enabled, a
results screen will be
displayed, as shown in Fig.
7.7, when entering any of the
other three modes.
Fig. 7.5 Air-Free CO
Fig. 7.6 Valid Sample
Fig. 7.7 CO Viewing Display
Note 1: The display briey ickers each time that a timed log is
captured (normally every minute).
7-4
Page 65
CO (CARBON MONOXIDE MODE)USER HANDBOOK
Note 2: A proper ‘DIF’ reading will only be obtained if there is a
preceding DIFZ (zeroing).
Note 3: Before viewing logs from the differential mode, a manual
‘zero’ must be performed as a reference point. If this has
not been done, since the last clear, a ‘zzz’ fault indication
will result.
If this option is enabled, the viewing display (Fig. 7.7) will be shown
before either CO Direct, Differential CO or Air Free CO.
Once viewed, the user can view the actual CO reading as per mode
previously selected.
To view actual CO reading:
• Press and hold OK .
Note: If CO Viewing is selected from the CO menu
(Fig. 7.1), the display has no OK option.
To exit from this option:
• Press and hold LH and RH buttons together.
To CLEAR (remove) existing readings from the display:
• Press and hold CLEAR .
Note: The values are stored in the datalogging memory for
future reference.
7-5
Page 66
7.5 CO BUTTON OPERATION
LH
(INVERT)
RH
(PUMP)
CENTRE
(RANGE)
ACTION
PRESS
INVERT
DISPLAY
MANUAL
LOG
CLEAR
VIEWING
OK (CONTINUE)
VIEWING
BACKLIGHT /
FLASHLIGHT
PUMP
ON / OFF
CLEAR FLOW
FAULT
ZERO
PRESS
and
HOLD
A summary of the button operation is detailed in table 7.1:
Table. 7.1 Button Operation in CO Mode
7.6 DESCRIPTION OF CO BUTTON OPERATION
7.6.1 Lights
To switch ON the backlight:
• Press and hold the Centre button .
To switch ON the ashlight:
• Press and hold the Centre button again.
The backlight and ashlight are both timed to switch OFF after two
minutes. Both can be switched OFF by a third press and hold of the
Centre button .
7.6.2 Manual Zero
Only available in differential CO mode.
To zero differential CO:
• Press and hold ZERO .
7-6
Page 67
CO (CARBON MONOXIDE MODE)USER HANDBOOK
7.6.3 Manual Log
A manual log can be taken at any time provided that the pump is
running.
To capture a manual log:
• Press and hold LOG .
Note: When a manual log is taken, the term ‘LOG’ is inversed on
the display for one (1) second to provide a visual conrmation
that the log has been captured.
7.6.4 Flow Fault
If ow fault is detected, the pump stops automatically. The instrument
should be checked for water ingress or blockage.
To clear the ow fault:
(Once the blockage has been cleared):
• Press the RH button .
7.6.5 Menu / Off
To re-select the Mode Menu:
• Press and hold both the LH button and the RH button
.
While both buttons are depressed, the instrument display
will step through previous menus, each displayed for
approximately two (2) seconds.
When the menu appears on the screen, release the buttons,
otherwise the instrument will proceed into the switch
OFF process and will switch OFF after a further three (3)
seconds.
7-7
Page 68
7.7 CO ALARMS
There are no alarms in this mode.
7.8 CO LOGGING
Automatic datalogging is active and manual logs can also be
captured. Timed logs will be taken every minute (default) or as
per conguration. For direct CO and differential CO, the direct CO
reading will be logged.
For the Air Free CO, the calculated reading will be stored.
A manual zero will also be stored.
7-8
Page 69
PURGE MODE
The purge mode is used in gas and air purging applications.
8.1 PURGE RANGES
Purge mode will have the following ranges available:
• 0 – 100% VOL Methane (CH4) Flammable
• 0 – 25% Oxygen (O2) if tted
8.2 PURGE FEATURES
Purge mode has the following features available:
• Pump ON / OFF
• Manual zero
• Backlight
• Flashlight
• Display Invert
8.3 PURGE DISPLAYS
During the warm-up period,
all applicable sensors will be
checked and the display will
indicate any sensor faults.
A positive zero fault is indicated
by a flashing gas reading, as
illustrated in Fig. 8.1.
A negative zero fault is indicated
by a spanner alternating with
a zero reading (not illustrated).
alternating with
8
Fig. 8.1 Sensor Check
8-1
Page 70
After the sensor check the
LH
(INVERT)
RH
(PUMP)
CENTRE
(RANGE)
ACTION
PRESS
INVERT
DISPLAY
RANGE
ZERO
BACKLIGHT /
FLASHLIGHT
PUMP
ON / OFF
CLEAR FLOW
FAULT
PRESS
and
HOLD
normal display will be shown,
as in Fig. 8.2:
Fig. 8.2 Normal Display
To display both % Gas and % O2:
(as shown in Fig. 8.3)
• Press the Centre
button
Fig. 8.3 Gas and Oxygen
Display
8.4 PURGE BUTTON OPERATION
A summary of the button operation is detailed in Table 8.1:
8-2
Table. 8.1 Button Operation in Purge Mode
Page 71
PURGE MODEUSER HANDBOOK
8.5 DESCRIPTION OF PURGE BUTTON OPERATION
8.5.1 Lights
To switch ON the backlight:
• Press and hold the Centre button .
To switch ON the ashlight:
• Press and hold the Centre button again.
The backlight and ashlight are both timed to switch OFF after two
minutes. Both can be switched OFF by a third press and hold the
Centre button .
8.5.2 Pump
To switch the pump ON and OFF:
• Press the RH button .
8.5.3 Manual Zero
To zero the reading (in fresh air):
• Press and hold ZERO .
Note: Pump must be switched ON to zero reading.
8.5.4 Menu / Off
To re-select the Mode Menu:
• Press and hold both the LH button and RH button
When the menu appears on the screen, release the buttons,
otherwise the instrument will proceed into the switch OFF
process and will switch OFF after a further three (3) seconds.
8-3
Page 72
8.5.5 Range
The range button allows the selection of the % Gas only display or
the % Gas plus % O2 display.
8.5.6 Flow Fault
If ow fault is detected, the pump stops automatically. The instrument
should be checked for water ingress or blockage.
To clear the ow fault:
(Once the blockage has been cleared)
• Press the RH button .
8.6 PURGE ALARMS
There are no alarms in this mode.
8-4
Page 73
9
SNIFFER MODE
This mode is used to nd small tting leaks. Fast detection rates are
achieved using a semiconductor sensor in the probe.
9.1 SNIFFER RANGES
Sniffer mode will have the following ranges available:
• 0 – 10,000 ppm Methane (CH4) Flammable
9.2 SNIFFER FEATURES
Sniffer mode will have the following features:
• Audible / Visual Ticker (Geiger) indication
• Ticker (Geiger) back-off
• Manual zero
• Pump ON / OFF
• Display invert
• Backlight
• Flashlight
9-1
Page 74
9.3 SNIFFER DISPLAYS
During the warm-up period,
all applicable sensors will be
checked and the display will
indicate any sensor faults.
A positive zero fault is indicated
by a ashing gas reading, as
illustrated in Fig. 9.1.
A negative zero fault
is indicated by a spanner
alternating with a zero reading
(not illustrated).
To zero the reading (in fresh air)
• Press and hold ZERO .
Note: Pump must be switched ON to zero reading.
After the sensors have been
checked, the normal display
will be as shown in Fig. 9.2:
alternating with
Fig. 9.1 Sensor Check
Fig. 9.2 Normal Display
9-2
Page 75
SNIFFER MODEUSER HANDBOOK
LH
(INVERT)
RH
(PUMP)
CENTRE
(RANGE)
ACTION
PRESS
INVERT
DISPLAY
BACKLIGHT /
FLASHLIGHT
PUMP
ON / OFF
CLEAR FLOW
FAULT
ZERO
PRESS
and
HOLD
9.4 SNIFFER BUTTON OPERATION
A summary of the button operation is detailed in Table 9.1:
Table 9.1 Button Operation in Sniffer Mode
9.5 DESCRIPTION OF SNIFFER BUTTON OPERATION
9.5.1 Lights
To switch ON the backlight:
• Press and hold the Centre button .
To switch ON the ashlight:
• Press and hold the Centre button again.
The backlight and ashlight are both timed to switch OFF after two
minutes. Both can be switched OFF by a third press and hold the
Centre button .
9.5.2 Pump
To switch pump ON and OFF:
• Press the RH button .
9-3
Page 76
9.5.3 Zero
To zero the ppm range
• Press and hold the RH button .
Note: The pump must be switched ON to zero the ppm range.
9.5.4 Flow Fault
If ow fault is detected, the pump stops automatically. The instrument
should be checked for water ingress or blockage.
To clear the ow fault:
(Once the blockage has been cleared):
• Press the RH button .
9.5.5 Ticker (Geiger) On / Off
To enable / disable the Geiger (Ticker):
• Press and hold the UP and DOWN buttons
simultaneously, as follows:
1. Press and hold - to enable visual only.
2. Press and hold again - to disable both audible and visual.
3. Press and hold again - to enable both audible and visual.
9.5.6 Ticker (Geiger) Adjust
To adjust the Ticker (Geiger) threshold:
(after the Ticker (Geiger) feature has been enabled)
• Press the UP or DOWN buttons.
The threshold setting is displayed briey beneath the PPM CH4.
9-4
Page 77
SNIFFER MODEUSER HANDBOOK
9.5.7 Menu / Off
To re-select the Mode Menu:
• Press and hold both the LH button and RH button .
When the menu appears on the screen, release the buttons,
otherwise the instrument will proceed into the switch OFF
process and will switch OFF after a further three (3) seconds.
9.6 SNIFFER TICKER (GEIGER) INDICATION
The Ticker (Geiger) range at start up is 0-1000 ppm.
For any subsequent PPM concentration, the Ticker (Geiger) audible
/ visual can be “zeroed” with the DOWN button , e.g. 600ppm
Ticker (Geiger) can be ‘backed off’ with the DOWN button , for
example, 600 ppm Ticker (Geiger) can be ‘backed off’ to enable
tracing to higher concentrations, in which case the Ticker (Geiger)
range becomes 600-1600 ppm etc, etc.
Note: The display range always remains as 0-10,000ppm.
The Ticker (Geiger) visual is such that the LED’s illuminate in pairs,
according to the table (Fig. 9.3), over any 1000 ppm range as
explained above.
9-5
Page 78
9.6.1 Select Audible / Visual Ticker (Geiger)
Red LED's - Pair 1
Red LED's
PPM Level
Pair 7
Pair 6
Pair 5
Pair 4
Pair 3
Pair 2
Pair 1100
Pair 8
Pair 2200
Pair 3300
Pair 4400
Red LED's
PPM Level
Pair 5500
Pair 6600
Pair 7700
Pair 8
800 - 1000
Combination Option
The Ticker (Geiger) function can be set for audible and visual
combination options as follows: Both ON; Audible OFF / Visual ON;
Both OFF; Both ON; etc.
To select required audible / visual combination option:
• Press and hold both UP and DOWN buttons.
• Repeat operation to cycle through combination options.
Fig. 9.3 Ticker (Geiger) LED Illumination
9-6
Page 79
10
PRESSURE MODE
In this mode, the instrument can be used as a manometer to measure
appliance and regulator pressure and to check the system for leaks.
Note 1: It is important to zero the pressure mode in ambient
atmosphere before taking measurements.
Note 2: Pump is OFF in pressure mode.
10.1 PRESSURE RANGES
Pressure mode will have either of the following congurable ranges
available:
• 0 to 60in Water Gauge (Resolution 0.1in)
• 0 to 150mBar (Resolution 1mBar)
Note: EEE (Over-range) is displayed if the pressure exceeds
60in. / 150mBar.
10.2 PRESSURE FEATURES
Pressure mode will have the following features available:
• Backlight
• Display Invert
10.3 PRESSURE DISPLAYS
When pressure test mode is
selected from the menu, the
normal operating display is
shown. See Fig. 10.1.
Fig. 10.1 Normal Display
10-1
Page 80
Note: A congurable option is
available to measure in mBar
as shown in Fig. 10.2.
Fig. 10.2 mBar Display
10.4 PREPARE INSTRUMENT (PRESSURE MODE)
Zero the instrument before tubing is attached.
To zero the instrument:
• Press and hold ZERO .
Connect the required length of tubing from the appliance to the
pressure port on the rear of the instrument, as shown in Fig. 10.3.
Fig. 10.3 Tubing Connected to Pressure Port
You can now utilise the instrument as a manometer in accordance
with your company practice.
10-2
Page 81
LH
(INVERT)
RH
(PUMP)
CENTRE
(RANGE)
ACTION
PRESS
BACKLIGHT
INVERT
DISPLAY
ZERO
PRESS
and
HOLD
PRESSURE MODEUSER HANDBOOK
10.5 PRESSURE BUTTON OPERATION
A summary of the button operation is detailed in Table 10.1:
Table. 10.1 Button Operation in Pressure Mode
10.6 DESCRIPTION OF PRESSURE BUTTON
OPERATION
10.6.1 Lights
To switch ON the backlight:
• Press and hold the Centre button .
The backlighting is timed to switch OFF after two minutes, or
alternatively, it can be switched OFF by a further press and hold the
Centre button .
10.6.2 Zero
To zero the instrument:
(before tubing is attached)
• Press and hold the RH button .
10-3
Page 82
10.6.3 Menu / Off
To re-select the Mode Menu:
• Press and hold both the LH button and RH button .
When menu appears on the screen, release the buttons,
otherwise the instrument will proceed into the switch
OFF process and will switch OFF after a further three (3)
seconds.
10.7 PRESSURE ALARMS
There are no alarms in this mode.
10-4
Page 83
11
BUMP TEST MODE
This mode is to allow the user to periodically test the instrument
against known gas concentrations. This mode is disabled by default.
11.1 BUMP TEST RANGES
All the gas ranges that are present in the GT are available in this
mode:
• 0-10000 ppm Methane (CH4) Flammable
• 0-100 % LEL Methane (CH4) Flammable
• 0-100 % VOL Methane (CH4) Flammable
• 0-25 % Oxygen (O2) if tted
• 0-2000 ppm Carbon Monoxide (CO) if tted
• 0-100 ppm Hydrogen Sulphide (H2S) if tted
11.2 BUMP TEST FEATURES
If enabled, Bump Test mode will have the following features available:
• Set-up of gases according to gas cylinder values.
• Five (5) Cal gases can be dened.
• Date stored when gas set-up changed.
• Peak values stored for applied gases.
• Ability to view details of previous 32 bump tests, including
time and date.
11-1
Page 84
11.3 BUMP TEST DISPLAYS
On entering bump test mode,
the rst screen is shown:
Each gas type / concentration
is stored using a CalGas
number, e.g. PPM may be
stored as CalGas 1; LEL as
CalGas 2; etc.
The CalGas selection number is highlighted. A different gas type
(number) can be selected.
To select a different calgas number:
• Press the UP or DOWN buttons as necessary.
In the example shown in Fig.
11-2, CalGas 2 is selected
and has a 50% LEL value
previously stored.
The concentration of a gas
range can be entered / edited
to suit the gas value shown
on the calibration gas bottle
to be used.
To enter / edit the gas range concentration:
• Press and hold NEXT
to highlight the
gas value, if previously
stored. Refer to Fig.
11-3. (50% LEL in
example).
Fig. 11-1 CalGas Selection
Fig. 11-2 Enter / View CalGas
Value
Fig. 11-3 Enter / Edit CalGas
Value
11-2
Page 85
BUMP TEST MODEUSER HANDBOOK
• Press the UP or DOWN buttons as necessary to
enter / edit value.
Note 1: As soon as either the UP or DOWN buttons are touched
and when there is no value displayed, a value one unit away
from the GMI default value, is automatically displayed. This
assists the user to enter the required gas concentration
more quickly.
Note 2: Pressing and holding either of the buttons will perform a
fast ramped increase / decrease in value.
At this point, the user can accept the highlighted value and apply the
gas, or alternatively, continue dening the other gases.
To accept the highlighted gas range concentration value:
• Press and hold OK .
If OK was selected, then the
screen, shown opposite, is
displayed:
On the left, the gas ranges
are displayed. In the middle
column, the target for the
applicable ranges are
displayed and on the right,
the (applied gas) peak value.
Gas range values that were
not entered in the applied CalGas will display the peak value as a
dash (-).
Apply the gas manually, using a balanced ow regulator.
Once the peak reading is high enough, or stabilised, disconnect the
gas. At this point the user can EXIT and all values will be stored
automatically, or go back to the CalGas set-up to dene and apply
another gas.
Fig. 11-4 Applied Gas Peak
Values
11-3
Page 86
To exit current gas range and return to main menu:
• Press and hold EXIT .
To return to CalGas selection display:
• Press and hold OK .
From CalGas set-up screen,
Fig. 11-5, the user can also
view the previous 32 bump
tests.
Fig. 11-5 CalGas Display
To view the previous 32 bump tests:
• Press and hold LAST .
The screen, shown opposite,
is displayed:
Fig. 11-6 Previously Stored Bump
To highlight required bump test:
• Press the UP or DOWN buttons as necessary
to highlight date and time of test.
Tests
11-4
Page 87
BUMP TEST MODEUSER HANDBOOK
LH
(INVERT)
RH
(PUMP)
CENTRE
(RANGE)
ACTION
PRESS
INVERT
DISPLAY
NEXTOK
EXIT
VIEW PREVIOUS
32 BUMP TESTS
PRESS
and
HOLD
To select highlighted option:
• Press and hold NEXT .
The screen, shown opposite,
displays the previously stored
bump test data:
Fig. 11-7 Stored Bump Test Data
To return to CalGas Display:
• Press and hold OK .
11.4 BUMP TEST BUTTON OPERATION
A summary of the button operation is detailed in Table 11.1.
Table 11.1 Button Operation in Bump Test Mode
11-5
Page 88
11.5 BUMP TEST LOGGING
All bump tests performed will be automatically stored in the
datalogging memory if the mode is exited via the EXIT (Centre) button.
Up to 32 different bump tests are stored in chronological order. If any
more are done, the oldest will be over-written.
The data can be extracted from the GT using the GT Data
Downloading software package (GMI Part no. 67164).
11-6
Page 89
12
ALARMS
12.1 GAS ALARMS
The following gas alarms are available according to gas type and are
programmable according to application and / or customer preference.
(See Tables 12.1 and 12.2).
12.2 FLAMMABLE (LEL) ALARMS
Up to three (3) instantaneous alarm levels are programmable. All
three are rising alarms, i.e. if the concentration is above the specic
alarm level, the alarm is triggered.
12.3 OXYGEN (O2) ALARMS
Up to three (3) instantaneous alarm levels are programmable, one (1)
rising and two (2) falling (to trigger alarms in O2 decient scenarios).
12.4 TOXIC GAS (e.g. CO) ALARMS
When operating normally, the instrument records minimum and
maximum readings for each gas. It also calculates the Short Term
Exposure Limit (STEL) and Long Term Exposure Limit (LTEL), known
as Time Weighed Average (TWA) readings, for each toxic gas range
as appropriate. Up to two (2) instantaneous rising and two (2) TWA
alarms are programmable for each toxic range tted to the instrument.
Note: A Time Weighted Average (TWA) value is the mean average
gas level over a specic period. The STEL is 15 minutes
and the LTEL is 8 hours. In accordance with legislation, this
requires the time weighted averages to be averaged over
a full period whether the instrument is ON or OFF. Such
12-1
Page 90
averaging essentially makes the instrument single user
applicable. The option is available to restart the averaging
after each instrument switch-off, thus allowing for multiple
user application.
A TWA alarm is intermittent with the actual gas reading,
therefore it is possible to get a value of zero (0) and an LTEL
or STEL alarm.
All alarms are user congurable to meet the specic needs of
different companies.
Note: The gas alarm levels, instantaneous STEL and LTEL, are
set at the time of instrument manufacture. It is important that
the user ensures that the levels are in accordance with their
company’s alarm levels and with health and safety legislation.
The alarm levels may be changed, if required, as detailed in
the ‘SET-UP SOFTWARE USER HANDBOOK’.
Example 1:
In the following examples,
example 1 (Fig. 12.1) shows
an instrument in confined
space mode signalling a
‘LOLO’ Oxygen alarm. The
audible alarm warbles and the
red LED’s ramp.
Toggles to
12-2
Fig. 12.1 LOLO Oxygen Alarm
Page 91
Example 2 (Fig. 12.2) shows
the instrument signalling a
‘HIHI’ LEL alarm. Again
the alarm warbles and all
eight (8) LED’s ramp. If more
than one gas alarm level is
exceeded, the gas value will
ash for each type in alarm.
Example 3 (Fig. 12.3) shows
an instrument in Leak
Test Mode with a Carbon
Monoxide warning alarm.
There is no audible alarm and
the LED’s do not ramp. The
gas range (ppm CO in this
example) will ash.
ALARMSUSER HANDBOOK
Example 2:
Toggles to
Fig. 12.2 HIHI LEL Alarm
Example 3:
Toggles to
Fig. 12.3 CO Warning Alarm
12-3
Page 92
Example 4 (Fig. 12.4) shows
an instrument in Leak Test
Mode with a ‘HI’ LEL alarm.
The audible alarm indication is
a high pitch tone together with
four ashing LED’s.
Example 5 (Fig. 12.5) shows
an instrument in Confined
Space Mode with a ‘LO’ O2
alarm. The audible alarm
indication is a high pitch tone
together with four flashing
LED’s.
Example 4:
Toggles to
Fig. 12.4 HI LEL Alarm
Example 5:
Toggles to
12-4
Fig. 12.5 LO O2 Alarm
Page 93
Examples 6 and 7 (Fig. 12.6
and Fig. 12.7) shows an
instrument in Conned Space
Mode with Time Weighted
Average STEL and LTEL H2S
alarms for toxic sensors. In
both cases, the audible alarm
warbles and eight (8) LED’s
ramp.
ALARMSUSER HANDBOOK
Example 6:
Toggles to
ACTUAL GAS READING
Fig. 12.6 STEL H2S Alarm
Example 7:
Toggles to
ACTUAL GAS READING
Fig. 12.7 LTEL H2S Alarm
12-5
Page 94
12.5 ALARM TYPES
12.5.1 Latching / Non Latching
Each alarm can be latching or non latching. Latching alarms must
be, and can only be cleared by the user when the gas level returns
to within the alarm limits.
Non-latching alarms clear automatically when the gas level returns
to within the preset alarm limits.
12.5.2 Muting / Acknowledging
Muting:
Muting of an alarm is achieved by a press and hold of the
LH button.
Muting means that the audible alarm will cancel for 60
seconds, after which if the gas concentration is still above
the set level (or below for Oxygen) the alarm will reactivate.
Acknowledging:
Only applicable to latching alarms. The audible/visual alarms
can only be cancelled after the atmosphere has returned to
safe concentrations.
The following Table 12.1 shows the GMI default selections. Latching
or non-latching options exist in all allowable alarms.
Alarms are allowed in Leak Test Mode and CSM Mode only.
Any single alarm e.g. LEL HI, Toxic Warning etc., can only have one
alarm concentration level and either latching/non latching and mute
or non mute and the output , i.e. LO Pitch / HI Pitch with LED’s
ashing/ramping are preset according to the following Table 12.1.
This means that if a CO HIHI Alarm is set at 35ppm and latching
and non-muting is selected for leak and/or CSM. The level will be the
same in each mode and the type, e.g. latching, will also be common.
If the alarm is disabled, it will not function in any mode.
12-6
Page 95
ALARMSUSER HANDBOOK
ALARM TYPE LATCHING
Yes / No
MUTE
Yes / No
AUDIBLE
INDICATION
VISUAL LED
INDICATION
DISPLAY
LEL WarningNN / AN / AN / AFlash Range
LEL (
HI
)NY
High Pitch
Tone
(4) Flashing
Toggle
Hi / Conc
LEL (
HIHI
)YN
High Pitch
Warble
(8) Ramping
Toggle
Hi Hi / Conc
O
2
(LO)
NY
High Pitch
Tone
(4) Flashing
Toggle
Lo / Conc
O
2
(
LOLO
)YN
High Pitch
Warble
(8) Ramping
Toggle
Lo Lo / Conc
O
2
(
HIHI
)YN
High Pitch
Warble
(8) Ramping
Toggle
Hi Hi / Conc
H
2
S WarningNN / AN / AN / AFlash Range
H
2
S (
HIHI
)YN
High Pitch
Warble
(8) Ramping
Toggle
Hi Hi / Conc
H
2
S (STEL)YN
High Pitch
Warble
(8) Ramping
Toggle
STEL / Conc
H
2
S (LTEL / TWA)YN
High Pitch
Warble
(8) Ramping
Toggle
LTEL / Conc
CO WarningNN / AN / AN / AFlash Range
CO (
HIHI
)YN
High Pitch
Warble
(8) Ramping
Toggle
Hi Hi / Conc
CO (STEL)YN
High Pitch
Warble
(8) Ramping
Toggle
STEL / Conc
CO (LTEL / TWA)YN
High Pitch
Warble
(8) Ramping
Toggle
LTEL / Conc
LO
YN / A
Low Pitch
Tone
Fault LED
Flashing
Fault 'Term'
Low BatteryYN / A
Low Pitch
Tone
Fault LED
On
Fault 'Term'
Zero FaultYN / A
Low Pitch
Tone
Fault LED
Flashing
Sensor FaultYN / A
Low Pitch
Tone
Fault LED
Flashing
Flow FaultYN / A
Low Pitch
Tone
Fault LED
Flashing
Calibration
Expired
YN / A
Low Pitch
Tone
Fault LED
Flashing
Service ExpiredYN / A
Low Pitch
Tone
Fault LED
Flashing
Fault 'Term'
Fault 'Term'
Fault 'Term'
Fault 'Term'
Fault 'Term'
Table. 12.1 Alarm Indication
12-7
Page 96
The following Table 12.2 shows the alarm options available in the
LEAK TESTCSM
FLAM COH2S O2FLAM CO H2S O2
WARNING
20
Hi
Hi Hi20351523
STEL200 10
LTEL305
Lo
Lo Lo
19.5
Leak Test and Conned Space Modes.
Table. 12.2 Alarm Options
Note:
1. Where there is a square with a number, it means that the GMI
default is to have these alarms active at the gas concentration
shown by the number.
2. Where there is a square without a number, it means that
this alarm option is available should the user wish to have it
activated.
3. The alarm concentrations must, if active, be the same level
in both modes.
12-8
Page 97
ALARMSUSER HANDBOOK
12.6 FAULT ALARMS
Refer to Alarms Table 12.1 to identify the audible / visual indication
for any of the following faults.
12.6.1 Low Battery
The battery symbol is displayed on the screen intermittently
with LO. When the instrument’s battery power is low, i.e.
approximately 30 minutes operating time remaining, the audible
alarm sounds once every two seconds and the Orange LED ashes.
Recharge the battery or replace the alkaline batteries if using alkaline
batteries.
The Low Battery ag ashes when approximately three (3)
minutes operating time remains. The audible alarm sounds once
every second and the Orange LED illuminates constantly. After three
(3) minutes the instrument automatically switches off.
Note: Both audible and visual gas alarms continue to operate after
the low battery warning message appears.
12.6.2 Zero Fault
A “ZERO FAULT” ag and a ashing spanner symbol appear after
warm-up and after entering a mode if the instrument is switched on
in the presence of gas or the instrument has been unable to zero
all sensors correctly.
The audible alarm sounds, and the orange fault LED ashes. If the
user is in a mode where the faulty sensor is used the orange LED
is on, continuously.
It is strongly recommended the instrument is returned to a gas free
area. Switch the instrument off and then switch on again in clean air.
If the fault persists, return the instrument for service.
The instrument can however still be used to detect and alarm on the
other sensor(s) tted.
12-9
Page 98
The faulty sensor will cause
the instrument to display a
flashing spanner symbol to
warn the user that this sensor
is not working correctly, as
shown in Fig. 12.8:
alternating with
Fig. 12.8 Zero Fault
12.6.3 Sensor Fault
There are two types of sensor fault as illustrated in the following
displays:
1) If a “ZERO FAULT”
flag and a flashing
spanner symbol
appear, alternating
with a zero reading
as shown in Fig. 12.9,
apply relevant test
gas (LEL in example)
for two minutes and
allow the display to
return to zero, then
switch instrument Off
and On again.
If fault remains,
return instrument to
an approved Service
/ Repair facility.
alternating with
Fig. 12.9 Sensor Fault
12-10
Page 99
2) If a “ZERO FAULT”
flag and a flashing
spanner symbol
appear, alternating
with a gas value as
shown in Fig. 12.10,
leave instrument
on for 30 to 60
minutes then switch
instrument Off and
On again.
If fault remains,
return instrument to
an approved Service
/ Repair facility.
12.6.4 Sample / Flow Fault
If a ‘FLOW FAULT’ flag
appears, alternating with
mode as shown in Fig. 12.11,
a ‘ow fault’ exists. The pump
symbol will be extinguished
(except in Confined Space
mode). The fault LED will
also be ON.
Check sample line, sample
lter or probe for blockage,
if applicable. Clear blockage
then restart the pump.
ALARMSUSER HANDBOOK
alternating with
Fig. 12.10 Sensor Fault
alternating with
Fig. 12.11 Sample / Flow Fault
12-11
Page 100
To clear the ow fault:
• Press the RH button .
or
• Press and hold FLOW ACK .
(in Conned Space Mode only)
Note: In Conned Space mode, the pump will not switch off if a
sample fault exists.
12.6.5 Calibration Expired
During the warm up of the instrument, a check is done to verify if
the calibration date has expired. If the instrument is congured to
pause if calibration is due, the user is asked to continue or not. If yes
is selected, the user can continue using the instrument as normal.
If no is selected, the user is forced to switch the instrument off and
return the instrument for calibration. The calibration due date is only
displayed during the warm up of the instrument and there will be no
indication during normal operation.
12.6.6 Calibration Required
During warm-up, if the
‘CALIBRATION REQUIRED’
flag is displayed and an
audible alarm and Red LED’s
are activated, the instrument
has detected a fault in the
calibration memory during
start-up and is unable to
continue without re-calibration.
The instrument must be switched off immediately. Follow appropriate
action required by your company for calibration.
An extended period is possible, allowing the user to continue to
use the instrument for a set amount of time after the calibration has
expired. When this extended period is over, the user will be forced
to switch off the instrument and return the instrument for calibration.
Fig. 12.12 Calibration Required
12-12
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.