GM TS420, MANTS420 User Manual

Model TS420
r
Oxygen Deficiency Detector
The information and technical data disclosed in this document may be used and disseminated only for the purposes and to the extent specifically authorized in writing by General Monitors.
Instruction Manual 11/00
General Monitors reserves the right to change published specifications and designs without prio notice.
*MANTS420*
Part No. MANTS420 Revision B/11-00
Model TS420
WARRANTY STATEMENT
General Monitors warrants the Model TS420 to be free from defects in workmanship or material under normal use and service within two (2) years (sensor cells one (1) year) from the date of shipment. General Monitors will repair or replace without charge any equipment found to be defective during the warranty period. Full determination of the nature of, and responsibility for, defective or damaged equipment will be made by General Monitors’ personnel. Defective or damaged equipment must be shipped prepaid to General Monitors’ plant or the representative from which shipment was made. In all cases this warranty is limited to the cost of the equipment supplied by General Monitors. The customer will assume all liability for the misuse of this equipment by its employees or other personnel. All warranties are contingent upon proper use in the application for which the product was intended and do not cover products which have been modified or repaired without General Monitors’ approval or which have been subjected to neglect, accident, improper installation or application, or on which the original identification marks have been removed or altered. Except for the express warranty stated above, General Monitors disclaims all warranties with regard to the products sold, including all implied warranties of merchantability and fitness and the express warranties stated herein are in lieu of all obligations or liabilities on the part of General Monitors for damages including, but not limited to, consequential damages arising out of/or in connection with the use or performance of the product.
WARNING - THE TS420 DETECTS OXYGEN DEFICIENT ENVIRONMENTS AND EXPOSURE TO OXYGEN DEFICIENT ENVIRONMENTS MAY RESULT IN SICKNESS OR DEATH.
The TS420 Toxic Gas Detector contains components, which can be damaged by static electricity. Special care must be taken when wiring the system to ensure that only the connection points are touched.
The TS420 is rated intrinsically safe as denoted by the symbol “Ex ia” on the instrument label. Substitution of electrical components within the TS420 may impair intrinsic safety.
Damage to the TS420 housing such that any internal components or potting can be or is exposed compromises the intrinsic safety of the device and as such should not be used in a hazardous environment. Such damage includes complete fracture of the plastic housing or cracks that may be opened to expose any internal components or potting. Destruction of the sensor cap will not affect the intrinsic safety of the TS420; however, calibration accuracy and the watertight nature of the TS420 will be compromised.
WARNING - DO NOT USE A TS420 WITH A DAMAGED HOUSING IN A HAZARDOUS ENVIRONMENT.
IMPORTANT - Each oxygen sensor cell is shipped separate from the TS420. This ensures that a fresh sensor cell will be used during initial start-up. DO NOT install the cell into the TS420 until you are ready to apply power to the system. When installing this unit, an initial field calibration must be completed, since the TS420 is not factory calibrated to the specific cell.
i
System Integrity Verification
General Monitors mission is to benefit society by providing solutions through industry leading safety products, services and systems that save lives and protect capital resources from the dangers of hazardous flames, gases and vapors.
The safety products you have purchased should be handled carefully and installed, calibrated and maintained in accordance with the respective product instruction manual. Remember these products are for your safety.
To ensure operation at optimum performance General Monitors recommends that certain maintenance items are performed.
Commissioning Safety Systems
Before power up, verify wiring, terminal connections and stability of mounting for all integral safety equipment including, but not limited to:
! Power supplies ! Control modules ! Field detection devices ! Signaling / output devices ! Accessories connected to field and signaling devices
After the initial application of power (and any factory specifies warm-up period) to the safety system, verify that all signal outputs, to and from devices and modules, are within the manufacturers’ specifications. Initial calibration / calibration checking / testing should be performed per the manufacturers’ recommendations and instructions.
Proper system operation should be verified by performing a full, functional test of all component devices of the safety system, ensuring that the proper levels of alarming occur.
Proper system operation should be verifies by performing a full, functional test of all component devices of the safety system, ensuring that the proper levels of alarming occur.
Fault/Malfunction circuit operation should be verified.
Model TS420
ii
Periodic Testing/Calibration of Field Devices
Periodic testing/calibrating should be performed per the manufacturers’ recommendations and instructions. Testing/Calibrating procedures should include, but not be limited to:
! Verify zero reading ! Apply a known concentration of gas, or a simulated test device provided by the
manufacturer
! Verify integrity of all optical surfaces and devices ! For flame detectors, use the appropriate test lamp
When testing produces results outside of the manufacturers’ specifications, re-calibration or repair/replacement of the suspect device(s) should be performed as necessary. Calibration intervals should be independently established through a documented procedure, including a calibration log maintained by plant personnel or third party testing services.
Periodic System Verification
The following system verifications should be performed at least annually:
Verify wiring, terminal connections and stability of mounting for all integral safety equipment including, but not limited to:
! Power supplies ! Control modules ! Field detection devices ! Signaling / output devices ! Accessories connected to field and signaling devices
Proper system operation should be verified by performing a full, functional test of all component devices of the safety system, ensuring that the proper levels of alarming occur.
Fault/Malfunction circuit operation should be verified.
Model TS420
iii
A
1. Mount and Wire the Detector (pages 5 through 9)
There are no standard rules for detector placement since the optimum sensor location is different for each application. The customer must evaluate conditions at the facility to make this determination. Generally, the TS420 should be easily accessible for calibration checks. The TS420 should be mounted pointing down to prevent water build-up on the detector head. The detector assembly should not be placed where it may be coated by contaminating substances. Although the Model TS420 is RFI resistant, it should not be mounted in close proximity to radio transmitters or similar equipment.
Some general guidelines for sensor location:
! The TS420 should be used in locations that are tightly enclosed with poor or no
ventilation, such as utility tunnels or underground storage areas. These areas are likely to have dangerously low oxygen concentrations.
! Observe the TS420’s environmental specifications when locating the unit.
The environmental specifications covering the electrochemical cell are found in the Appendix, Section 5.1.4.
Detectors should be mounted in an area that is as free from wind, dust, water, shock and vibration as possible.
WARNING - Operation above or below temperature limits will provide unstable readings, which may cause false alarms or failure to alarm. See Section 5.1.4 for guidelines.
General Monitors discourages painting of detector assemblies. If the detector head is painted over, the gas will not be able to diffuse into the sensor. If the TS420 body is painted over, the LCD display cannot be read.
The TS420 is shipped without the sensor installed. The sensor must be installed into the detector and calibrated for proper operation. See Sections 3.3 and 3.4 for installation instructions and Section 4.3 for calibration instructions.
Once correctly installed, the TS420 requires little or no maintenance other than periodic calibration checks to ensure system integrity. General Monitors recommends a calibration check schedule be established and that the complete system, including all alarm circuitry, be tested at least annually.
The overall and mounting dimensions for the TS420 (Figure Q-A) should be used when making installation determinations. Mechanical specifications can be found in the Appendix, Section
5.1.2.
Model TS420
Quick Start Guide
4.50"
E D
Y
L
L
O
N
A
O
M
C
N E T H G
I T D N A H
S
L A
R
R
O
E
T I
N
N
E
Ø1.75"
LOOP+ (RED)
LOOP- (WHITE)
FRAME GROUND
(GREEN/YELLOW)
3/4 NPT
1 1/2-20 UN-2
Figure Q-A Outline and Dimensional Drawing
iv
Intrinsically Safe Barriers
In order for the TS420 to be used with classified equipment inside of the hazardous area, it is necessary for the customer to purchase either a Galvanic or Zener Diode Barrier, which should be mounted in the non-classified area between the TS420 and the facilities’ control room. A list of recommended manufacturers and part numbers follows for galvanic barriers:
Stahl - 9303/11-22-11 Elcon - mD323 or mD325 MTL - MTL7206 or MTL5041 Turck - mk33-Li-ExO Pepperl+Fuchs - KFD2-CR-EX1.30-300
These are recommended active type barriers for intrinsically safe applications. Each application is different and it is the ultimate decision of the user to determine the appropriate I/S barrier for their application.
NOTE - To maintain intrinsic safety, wire length and gauge are limited by the capacitance and inductance of the wire, Ci and Li of the TS420, and the barrier ratings. See Section 5.1.5 for instructions on calculating these values.
Align the TS420 so that the LCD display is easily viewed. Teflon tape may be used on the TS420’s threads.
The two wires at the base of the TS420 provide a 4-20mA loop-powered output. The red wire is the “loop +” lead and the white wire is the “loop -” lead (Figure Q-B). A third green/yellow wire is provided for frame ground.
Model TS420
Quick Start Guide
JUNCTION BOX
LOOP + (RED)
RED
CONTROL
ROOM
LOOP - (WHITE)
WHITE
GRN/YEL
TS420
Figure Q-B Wiring Diagram
It is recommended that a two wire, shielded cable be used for making power and output signal connections on the TS420.
Connect the green/yellow frame ground wire to the grounding terminal or lug inside the junction box being used with the TS420.
Ensure the junction box frame is connected to frame ground or connected to the cable shield, which is connected to frame ground at the controller.
Connect the red wire to the +24VDC terminal. Connect the white wire to the signal input terminal on the readout/relay display module, FMD Field Mounted Display, S4100E smart sensor, industrial analog to digital converter, computer-based monitor, PLC, DCS, etc.
Since the TS420 is designed to operate continuously, a power switch is not included, in order to prevent accidental system shutdown.
v
NOTE - Power must remain disconnected until all wiring connections are made.
The absolute maximum distance between the TS420 and the power supply is 8000 feet (2438 meters). Depending on the application, general purpose or intrinsically safe, the cable length will depend on the intrinsically safe barrier chosen and the cable used. See the Appendix, Section 5.1.5, for the intrinsically safe system requirements. In addition, cable length specifications can also be found in the Appendix, Section 5.1.3. In all cases, the cable run should be as short as possible.
2. Apply power to the detector.
Before applying power to the system for the first time, all wiring connections should be checked for correctness. The sensor should be allowed to stabilize prior to calibration. The stabilization time should be one (1) hour for all sensors.
Initial Power-up
The TS420 has been shipped un-calibrated, therefore, the first time power is applied, the TS420 will enter Start-up mode (3.6mA) for a short interval, then may transition to fault mode (flashing MODE LCD with the CAL LCD off and the analog output current at 3.5mA) until calibrated (See Section 4.3 for Calibration Procedure).
After calibration is complete, the TS420 will proceed to Operate mode (17.38mA = 20.9% V/V O
) where any detected oxygen will show as a proportional analog output current between 4
2
and 20mA (4mA = 0% V/V O
, 20mA = 25% V/V O2). The MODE LCD will be on steady and the
2
CAL LCD off with normal atmosphere and both the MODE and CAL LCD's will flash when oxygen deficiency is detected. (O
level < 19%).
2
General Monitors recommends the TS420 be calibrated again within the first twenty-four (24) hours after the initial calibration. Calibrations should be performed with new units and units that have been off power for more than one (1) week. See “Calibration”, Section 4.3, for more details.
Power-up After Initial Calibration
When the TS420 is powered up, after it has been previously calibrated, the unit will enter a Start-up mode (3.6mA) for approximately 50 seconds, which allows the sensor to stabilize before proceeding to the Operate mode (17.38mA). During Start-up, the “MODE” LCD will flash, the “CAL” LCD will be off and the analog output current will be 3.6mA.
OPERATING MODES INDICATOR ANALOG OUTPUT
CAL MODE
START-UP
OPERATE AMBIENT O2 = 20.9% DEFICIENT ATMOSPHERE, O2< 19%
1
■ ✹
FAULT NO POWER SYSTEM
❏ ❏
Model TS420
Quick Start Guide
3.6mA
17.38mA
4.0-16.16mA
0.0mA
3.5mA
Figure Q-C Operating Modes, Indicators and Outputs
1. Segment may flash Synchronously or alternating.
Perform a calibration and the TS420 will return to operate mode (17.38mA).
NOTE - The TS420 will not allow you to enter calibration mode during Start-up mode (3.6mA).
vi
3. Calibrate the Detector.
Activating the “CAL” switch will automatically disable the external alarm circuits by fixing the analog output current to 3.75mA.
General Monitors recommends that the TS420 be calibrated within the first twenty-four (24) hours after initial start-up, and that calibration be checked at least every ninety (90) days to ensure the integrity of the system.
NOTE: A calibration check consists of applying a 20.9% V/V O scale concentration of gas to the sensor and observing the reading on the display device being used.
General Monitors is not implying that the customer should expect problems with sensor life or stability. “Frequent” calibration checks merely ensure the integrity of the life protecting equipment, and are recommended for problem environments (i.e. sensors accidentally being painted over, etc, see Appendix 5.4).
A calibration schedule should be established and followed. A logbook should also be kept showing calibration dates and dates of sensor replacement.
Calibration Procedure:
1. If it is suspected that the air is not at “normal” levels (20.9% O calibrate the sensor with clean air. The 20.9% O
gas cylinder (GMI P/N 1400262-11) can be
2
used. Apply the clean air source prior to entering calibration mode and allow the sensor to stabilize for at least one (1) minute before proceeding.
2. Place the magnet over the GMI Logo on the body of the unit and hold it there until the “CAL” LCD indicator appears, then remove the magnet. The analog output current will drop to
3.75mA.
3. The “CAL” and “MODE” LCD indicator will flash, indicating that the unit is now calibrating.
4. After approximately 30 seconds, the display will change from a flashing “CAL” and “MODE” LCD to a solid “MODE” LCD, indicating that the calibration is complete and the unit has re­entered “OPERATE” mode. Remove any calibration gas.
6. The unit is now calibrated and the new values have been stored in the NOVRAM (non­volatile memory).
Figure Q-D shows a flow diagram of the codes that will appear in the display window during the calibration procedure.
If there is a problem and the TS420 cannot complete the calibration sequence, a fault indicator will be displayed, and the analog output current will drop to 3.5mA.
CALIBRATION INDICATOR CAL MODE Magnet Applied Magnet Recognized Cal Proceeding
Model TS420
Quick Start Guide
(“normal” ambient air) of full-
2
), it will be necessary to
2
❑ ✹
Figure Q-D Flow Diagram
vii
NOTE - If the unit fails to calibrate a flashing “MODE” LCD will be displayed. The sensor must see “clean” air for at least fifteen (15) minutes before a second calibration is attempted by reapplying the magnet.
The instrument is now ready to operate! Please consult the manual for more information on the instrument’s many features.
If you have any problems in the set-up or testing of the detector, please refer to the “Trouble Shooting Section”, or call the factory direct.
Worldwide Service is available by calling:
Lake Forest, California Phone: +1-949-581-4464 (24 hr. service) Fax: +1-949-581-1151
Houston, Texas Phone: +1-281-855-6000
Fax: +1-281-855-3290
Ireland Phone: +1-353-91-751175 Fax: +1-353-91-751317
Singapore Phone: +1-65-748-3488 Fax: +1-65-748-1911
United Arab Emirates Phone: +1-971-4-8815751 Fax: +1-971-4-8817927
United Kingdom Phone: +1-44-1625-619-583 Fax: +1-44-1625-619-098
Model TS420
Quick Start Guide
viii
Model TS420
Table of Contents
WARRANTY STATEMENT ..........................................................................................i
System Integrity Verification .........................................................................................ii
Commissioning Safety Systems......................................................................................ii
Periodic Testing/Calibration of Field Devices ...............................................................iii
Periodic System Verification ........................................................................................iii
Table of Contents ........................................................................................ix
Table of Figures .........................................................................................x
1.0 Introduction .........................................................................................1
1.1 General Description...................................................................................1
1.2 Features and Benefits ...............................................................................1
1.3 Applications .........................................................................................1
2.0 Detector Assembly .........................................................................................3
2.1 Sensor Assembly.......................................................................................3
2.2 Control Electronics ....................................................................................4
3.0 Installation .........................................................................................5
3.1 Receipt of Equipment ................................................................................5
3.2 Detector Location Considerations .............................................................5
3.3 Installation Instructions ..............................................................................6
3.3.1 Intrinsically Safe Barriers..............................................................6
3.3.2 Maintenance .................................................................................6
3.4 Wiring Connections ...................................................................................7
4.0 Operation .........................................................................................8
4.1 Applying Power .........................................................................................8
4.2 Operating Modes .......................................................................................9
4.3 Calibration .........................................................................................9
4.4 Calibration Equipment .............................................................................10
4.5 Sensor Replacement Procedure .............................................................11
4.6 Sensor Care and Maintenance................................................................13
5.0 Appendix .......................................................................................14
5.1 Specifications .......................................................................................14
5.1.1 System Specifications ................................................................14
5.1.2 Mechanical Specifications ..........................................................14
5.1.3 Electrical Specifications..............................................................14
5.1.4 Environmental Specifications .....................................................15
5.1.5 Intrinsically Safe System Requirements.....................................16
5.1.6 Sample Calculation.....................................................................16
5.2 Engineering Documentation ....................................................................17
5.2.1 Outline & Dimensional Drawing..................................................17
5.2.2 Control Drawing ..........................................................................18
5.2.3 Sample Installation .....................................................................19
5.2.4 Junction Boxes ...........................................................................20
5.2.5 Splash-Guards............................................................................21
5.2.6 Accessories ................................................................................22
5.2.7 Calibration Accessories ..............................................................22
5.3 Ordering Information................................................................................23
5.4 Calibration Schedule for Problem Environments.....................................24
5.5 Spare Parts & Accessories......................................................................25
ix
Table of Figures Page
Figure Q-A Outline and Dimensional Drawing ......................................................................iv
Figure Q-B Wiring Diagram................................................................................................... v
Figure Q-C Operating Modes, Indicators and Outputs..........................................................vi
Figure Q-D Flow Diagram .................................................................................................... vii
Figure 1 Model TS420 ...................................................................................................... 1
Figure 2 Oxygen Deficiency Sensor Assembly ................................................................ 3
Figure 3 LCD Display Location......................................................................................... 4
Figure 4 Wiring Diagram................................................................................................... 7
Figure 5 Operating Modes, Indicators and Outputs.......................................................... 8
Figure 6 Flow Diagram ................................................................................................... 10
Figure 7 Calibration Cup................................................................................................. 10
Figure 8 Calibration Plug ................................................................................................ 10
Figure 9 Change in % Oxygen in the atmosphere
with Relative Humidity at constant Temperature................................. 11
Figure 10 Sensor Alignment ............................................................................................. 12
Figure 11 Calibration Kit with Bottle.................................................................................. 13
Figure 12 Outline and Dimensional Drawing .................................................................... 17
Figure 13 Control Drawing................................................................................................ 18
Figure 14 Sample Installation ........................................................................................... 19
Figure 15 10252 Round Aluminum Junction Box ............................................................. 20
Figure 16 31305-2 Aluminum Junction Box...................................................................... 20
Figure 17 45160-1 Plastic Junction Box, NPT; 45160-2 Plastic Junction Box, M20 ........ 21
Figure 18 70631-2 Splash - Guard ................................................................................... 21
Figure 19 45170-1 Flow Block.......................................................................................... 22
Figure 20 45147-1 Flow Block Retainer ........................................................................... 22
Figure 21 914-152 Calibration Plug .................................................................................. 22
Figure 22 1400152-1 Calibration Cup .............................................................................. 22
Calibration Schedule for Problem Environments........................................................................ 24
Model TS420
x
Loading...
+ 25 hidden pages