GlucoRx HCT Owner's Manual

L148*W105mm/模造紙80P/:4C,4, :,32/36/騎馬釘
機器說明書/GlucoRx HCT/英文(DiME)
311-4279300-XXX
K100
K80 K20
Emily
Color
Part no.
Product name
Spec
C40 M40 Y70 K30
C90 M60 Y10 K0
(also suitable for use in newborn babies and pregnancy)
Use with GlucoRx HCT Glucose Test Strips, GlucoRx HCT Ketone Test Strips and GlucoRx Lancets.
-6 5 15 25 35 45 55 65
-40
-20
0
40
20
60
Effect of haematocrit on accuracy
Haematocrit
% Error
Without 2+2
With 2+2
Haematocrit Correction Technology (HCT)
GlucoRx HCT Glucose test strips use advanced GDH-FAD enzyme technology with 2+2 (AC and DC) Bio Signal. AC signal works to measure the haematocrit result, and DC measures glucose. Meter then modulates glucose level according to haematocrit value.
BGM vs. YSI VALUES BEFORE AND AFTER HCT
YSI (mmol/L)
Before 2+2
BGM (mmol/L )
After 2+2
10 25 50
50
25
10
0
4 Alarm
reminders
5 sec.
TIMER
5
Software
Features
Ketone testing & warning Exceeds EN ISO 15197:2015 Accuracy criteria Haematocrit Correction Technology Measures Haematocrit (10 - 70% haematocrit range)
4 Alarm reminders 5 second glucose results
requirements Meal markers Auto QC tagging PC download facility; Diasend compatible.
3
Dear GlucoRx HCT System Owner:
Thank you for purchasing the GlucoRx HCT Blood Glucose Monitoring System. This manual provides important information to help you to use the system properly. Before using this product, please read the following contents thoroughly and carefully.
The System measures both blood glucose and β-ketone. Regular monitoring of your blood glucose and β-ketone levels can help
you and your doctor gain better control of your diabetes. Due to its compact size and easy operation, you can use the GlucoRx HCT Blood Glucose Monitoring System to easily monitor your blood
glucose and β-ketone levels by yourself anywhere, any time.
If you have other questions regarding this product, please contact
GlucoRx Customer care on 01483 755133.
Intended Use
This system is intended for use outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitoring the effectiveness of diabetes control. It is intended to be used for the quantitative measurement of glucose (sugar) in fresh whole blood
samples from the finger, and to measure β-hydroxybutyrate (ketone)
in fresh whole blood samples from the finger and palm. It should not be used for the diagnosis of or screening for diabetes mellitus.
The test for haematocrit (HCT) as part of the system, is intended
for use in the in vitro quantification of packed red blood cell volume fraction in capillary whole blood as an aid in monitoring the status of
total volume of red blood cells. The test reading of haematocrit (HCT)
is used only to determine whether the blood test sample is within the acceptable range of blood glucose system. It should not be used for the diagnosis of anaemia or erythrocytosis.
Version 8.0 2017/02 311-4279300-XXX
4
IMPORTANT SAFTY INSTRUCTIONS
READ BEFORE USE
1. Use this device ONLY for the intended use described in this manual.
2. Do NOT use accessories which are not specied by the manufacturer.
3. Do NOT use the device if it is not working properly or if it is damaged.
4. This device does NOT serve as a cure for any symptoms or diseases.
The data measured is for reference only. Always consult your doctor to
have the results interpreted.
5. The blood glucose test strip can be used for the testing of newborns;
The β-ketone test strip must not be used for the testing of newborns.
6. Before using this device to test blood glucose or β-ketone, read all
instructions thoroughly and practice the test. Carry out all the quality
control checks as directed.
7. Keep the device and testing equipment away from young children.
Small items such as the battery cover, batteries, test strips, lancets
and vial caps are choking hazards.
8. Use of this instrument in a dry environment, especially if synthetic
materials are present (synthetic clothing, carpets etc.) may cause
damaging static discharges that may cause erroneous results.
9. Do NOT use this instrument in close proximity to sources of strong
electromagnetic radiation, as these may interfere with accurate
operation.
10. Proper maintenance and periodic control testing are essential to the
longevity of your device. If you are concerned about your accuracy of
measurement, please contact GlucoRx Customer care on
01483 755133.
KEEP THESE INSTRUCTIONS IN A SAFE PLACE
5
BEFORE YOU BEGIN 09
Important Information 09
Contents of System 10
Meter Overview 11
Display Screen 12
Test Strip 12
SETTING THE METER 13
THE THREE MEASURING MODES 14
BEFORE TESTING 15
Calibration 15
How to Code Your Meter (for β-Ketone Test) 15
Checking the Code Number 15
Control Testing 16
Performing a Control Test 17
TESTING WITH BLOOD SAMPLE 18
Preparing the Lancing Device for Blood Testing 18
Preparing the Puncture Site 18
Performing a Blood Glucose or β-Ketone Test 19
Alternative Site Testing (for β-Ketone Test) 21
METER MEMORY 22
Reviewing Test Results of Blood Glucose or β-Ketone 22
Reviewing Blood Glucose Day Average Results 22
DOWNLOADING RESULTS ONTO A COMPUTER 23
MAINTENANCE 24
Battery 24
Caring for Your Meter 25
Caring for Your Test Strips 25
Caring for Your Carry Case 26
Important Control Solution Information 27
TABLE OF CONTENTS
6
SYSTEM TROUBLESHOOTING 28
Result Readings 28
Error Messages 29
Troubleshooting 30
DETAILED INFORMATION 31
SYMBOL INFORMATION 32
SPECIFICATIONS 33
7
(e) (f) (g) (h)(d)
(i)
(k) (l)
(a) (b) (c)
S
E
T
832
code
832
code
832
8
(n)
(m)
(j)
(p)
(o)
(q)
1
2
9
BEFORE YOU BEGIN
Important Information
● Severe dehydration and excessive water loss may cause readings
which are lower than actual values. If you believe you are suffering from severe dehydration, consult a healthcare professional immediately.
● If your blood glucose or β-ketone results are lower or higher than usual, and you do not have any symptoms of illness, rst repeat the
test. If you have symptoms or continue to get results which are higher or lower than usual, follow the treatment advice of your healthcare professional.
● Use only fresh whole blood samples to test your blood glucose or β-ketone. Using other substances will lead to incorrect results.
● If you are experiencing symptoms that are inconsistent with your blood glucose or β-ketone test results and you have followed all the
instructions given in this owner’s manual, contact your healthcare professional.
● We do not recommend using this product on severely hypotensive individuals or patients in shock. Please consult your healthcare
professional before use.
● The measurement unit used for indicating the concentration of blood or plasma glucose will have molarity (mmol/L). The approximate calculation rule for conversion of mmol/L in mg/dL is:
For example:
1) 120 mg/dL ÷ 18 = 6.6 mmol/L
2) 7.2 mmol/L x 18 = 129 mg/dL approximately.
mg/dL
mmol/L
Divided by 18
Times 18
= mmol/L
= mg/dL
10
Contents of System
Your new GlucoRx HCT System kit includes:
• A meter
• Owner’s manual
• Protective wallet
• Quick start user guide
• Daily log book
• Warranty card
• Lancing device with AST cap
• Batteries
• Glucose test strips vial
• Single foil ketone test strip
• Control solution
• Lancets
NOTE:
If any items are missing from your kit or opened prior to use, please contact
GlucoRx Customer care on 01483 755133.
11
Meter Overview
Front
Top
Right SideLeft Side Back
1
SET
2
6
7
4
3
5
Test Strip Slot
Insert test strip here to turn the meter on for testing.
Display Screen
Main Button (M)
Enter the meter memory
Test Strip Ejector
Eject the used strip by pushing up this button.
Data Port
Download test results with a cable connection.
SET Button (S)
Enter and conrm the meter settings.
Battery Compartment
1
2
3
5
4
6
7