Global Water WS700 Instruction Manual

Global Water
800-876-1172 globalw.com
Global Water
Instrumentation, Inc.
Gold River, CA 95670
T: 800-876-1172
Int’l: (916) 638-3429, F: (916) 638-3270
globalw.com
WS700 Sampler
01-455 11/30/09
- 1 -
Global Water
800-876-1172 globalw.com
Table of Contents
I. WS700 Checklist Page 3 II. Inspection 3 III. Description 4 IV. Installing the Water Sampler 5 V. Installing the Pickup Hose 5 VI. The Control Panel 6 VII. Composite Sampling 8 VIII. Discrete Sampling 8 IX. External Trigger mode 9 X. Specifications 11 XI. Maintenance 12 XII. Troubleshooting 13 XIII. Warranty 14
Appendix A: Sample Size vs. Head Height 15
XIV. XV. Appendix B: Battery Life 16 XVI. Appendix C: Input and Output Connection 16 XVII. Appendix D: Accessories 17
XVIII. CE Certification 18
* Copyright © Global Water Instr umentation, Inc. 2006
- 2 -
Global Water
800-876-1172 globalw.com
Congratulations on your purchase of the Global Water WS700 Water Sampler. This instrument has been quality tested and approved for accurate and reliable operation. We are confident that you will find the WS700 to be a valuable asset for your applications. Should you require assistance, our technical staff will be happy to help.
I. WS700 Checklist
a. WS700 Water Sampler b. 12V Battery (Inside Enclosure) c. Battery Cable d. Battery Charger e. Two Shorting Plugs f. WS700 Manual
II. Inspection
The WS700 unit was carefully inspected and certified by Global Water’s Quality Assurance Team before shipping. If any damage has occurred during shipping, please notify Global Water Instrumentation, Inc. and file a claim with the carrier involved.
Use the checklist to ensure that everything needed to operate the WS700 was received.
- 3 -
III. Description
The Global Water WS700 Sampler is designed specifically to meet a wide variety of sampling requirements including industrial discharge, process control, water and wastewater treatment plants, sewers, rivers and streams.
The WS700 Sampler can take individual “time weighted” composite samples, or it can be set to take a full-bottle discrete or “Grab” sample. This is accomplished with a Sample Size control having one position being reserved for “Full” bottle. A Sample Interval control sets the time between individual composite samples and has a provision for allowing the sampler to be triggered from external sources.
A Start Delay timer can be set to start sampling after a preset time. This allows multiple samplers to be deployed in the field and have them turn on at the same time. The delay can also be used to hold off sampling to let the water source flush out debris or other contaminants.
After each sample, the pump will reverse for 15 seconds. This clears any debris from the strainer at the end of the pickup hose and also empties water from the hose so the next sample is not contaminated by the previous one.
Inputs are provided for triggering the sampler based on water level, a rain gauge sensor or a signal from an external process controller or other monitoring device. An output is also available for monitoring when the pump is taking samples, using a data logger or similar recording device.
The WS700 Sampler is easily set up near industrial discharges or streams. It can also be suspended in a manhole for sewer flow sampling. The sampler consists of a rugged, rainproof and lockable carrying enclosure. The enclosure houses a peristaltic sampling pump, sample bottle, a control panel, and a rechargeable battery. A battery charger and sample pickup hoses are also provided.
Global Water
800-876-1172 globalw.com
- 4 -
Global Water
800-876-1172 globalw.com
IV. Installing the Water Sampler
a. The sampler should be placed upright (it will not work if placed on its back or
side).
b. Open the sampler and remove the battery charger from the unit. Store the
charger for later use.
c. To secure the sample bottle:
1. Screw the bottle cap/float switch onto the sample bottle
2. Place the bottle into the sampler enclosure
3. Insert the end of the peristaltic pump’s hose in the hole at the top of the bottle cap
4. Plug the float switch lead into the jack on the front of the control panel
d. The sampling unit can be secured from vandalism and str ong winds by one of
the following methods:
1. Mount unit on post and lock closed
2. Lock closed and chain handles to a solid structure (such as a tree, post, or building)
3. Enclose and lock unit in a steel electrical box
e. Avoid drilling holes in the enclosure if possible. If holes must be drilled in the
enclosure avoid drilling through the control panel enclosure. Any holes drilled through the enclosure must be sealed with some type of silicone glue to prevent water from entering the case and causing equipment failure.
f. The unit is water resistant, not water proof. The unit must be located well
above the expected water level to continue to provide reliable service.
V. Installing the Pickup Hose
The pickup hose should be installed in a wastewater discharge channel or stream. The pickup strainer should be submerged under water and should be situated to avoid contact with the channel bottom.
- 5 -
VI. The Control Panel
1. Pump.
2. Start Delay: time. If no delay is desired, set the control to zero. Only the first sample taken by the pump is delayed, except as described later in the section titled “External Trigger Mode”.
3. Sample Interval: The “External Trigger” position is reserved for using the Trigger and Rain Sensor inputs to initiate sampling as described later in the section titled “External Trigger Mode”.
4. Sample Size: sample size settings are approximate and apply to a head height of four feet. The switch position marked “Full” causes the pump to work as a discrete sampler, filling the bottle full in one sample. Note: In the Full position, the sampler will shut off the pump after one hour, even if the bottle is not full.
5. Trigger In: selected by the Sample Interval control. A signal from an external device will cause a single sample to be taken according to the Size control. The External Trigger mode is described in more detail later in the section titled “External Trigger Mode”. Refer to the Specifications section for details about the input pulse requirements.
Global Water
800-876-1172 globalw.com
This setting delays sampling by a preset amount of
Sets the time interval between composite samples.
Sets the size for a single composite sample. The
This input works in “External Trigger” mode only, as
- 6 -