Global Water WL15 User Manual

Page 1
WL15
Water Level Logger User’s Guide
Copyright © Global Water Instrumentation, Inc. 2002
Global Water (800)876-1172 www.globalw.com
6/5/2002
Page 1
Page 2
Requirements
Windows 95, 98, ME, NT, 2000 or XP*
486DX 66MHz, 16MB RAM (Windows 95 base machine; later operating systems may require
higher specifications) Available COM port
Contact Information
Global Water Instrumentation, Inc.
11257 Coloma Road Gold River, CA 95670
Toll Free: 1(800)876-1172 Local: 1(916)638-3429 Fax: 1(916)638-3270 E-mail: globalw@globalw.com Web: www.globalw.com
Terminology
In order to better understand this product manual, please be aware of the following terminology and formatting usage.
Bolded terms (i.e. puter monitor screen or a phrase to be emphasized. Context will dictate which is indicated.
Since it is assumed that users of this manual are familiar with mouse usage, many mouse opera­tions are implied. The term “click” or “clicking” means to press and release the left mouse button. Similarly, “double-click” means to press and release the left mouse button twice in quick succes­sion. For example, the phrase “Click the pointer to the task bar and click the tion, it will be specified in the documentation.
The term “select” means to move the mouse pointer over the indicated target. For example the phrase “Select after the
View
Toolbars
menu has been displayed. “Select” is normally used when the mouseover action
will cause additional menus to display.
) indicate either a keyboard key, a word or icon that appears on your com-
bold
button” should be interpreted as “Move the mouse
start
button.” If any other mouse button is used in an opera-
start
from the
menu” means to move the mouse pointer over
View
Toolbars
The symbol is used with submenu operations where the user should click the first and last items in the string separated by . For example, the phrase “Click on means click on
File
, select
and click on
New
Folder
.
File New Folder
When the documentation directs you to exit a program or window, you can generally do so by clicking the x graphic in the upper right-hand corner or by clicking
*Windows® 95, 98, ME, NT, 2000 & XP are trademarks of the Microsoft Corporation
Global Water (800)876-1172 www.globalw.com
File Exit.
x graphic
Page 3
Table of Contents
CHAPTER 1: INTRODUCTION .............................................................................. 5
THE WL15 W F
EATURES
C
HECKLIST
G
ETTING TO KNOW THE
ATER LEVEL LOGGER
.................................................................................................................................... 5
................................................................................................................................... 6
WL15 ..................................................................................................... 7
CHAPTER 2: SETUP .............................................................................................. 9
I
NSTALLING THE
S
YSTEM TIME CHECK
G
LOBAL LOGGER SOFTWARE INSTALLATION
H
ARDWARE INSTALLATION
General Installation Tips ........................................................................................................ 14
Groundwater Installation ........................................................................................................ 15
Surface Water Installation ...................................................................................................... 15
Special Installations ............................................................................................................... 16
S
TORING THE
WL15 ................................................................................................................. 9
................................................................................................................... 9
.......................................................................................................... 14
WL15 ................................................................................................................... 16
CHAPTER 3: USING GLOBAL LOGGER SOFTWARE ...................................... 17
H
ARDWARE CONNECTION
G
LOBAL LOGGER STARTUP
G
LOBAL LOGGER FUNCTIONS
Sample Continuously/Real-Time Values ............................................................................... 19
Get Settings ........................................................................................................................... 19
Get History Data .................................................................................................................... 19
Clear Memory ........................................................................................................................ 21
Synchronize Time .................................................................................................................. 22
Setup ..................................................................................................................................... 22
Datalogger Name and Sensor Warmup Time ..................................................................................................23
Calibration ...................................... ................................................................... ...............................................24
Recording Interval and Memory Management .................................................................................................26
........................................................................................................... 17
........................................................................................................ 17
CHAPTER 4: MAINTENANCE ............................................................................. 29
B
ATTERY
C
ALIBRATION
S
ENSOR SCREEN SCREENING
.................................................................................................................................... 29
............................................................................................................................. 30
..................................................................................................... 30
CHAPTER 5: TROUBLESHOOTING ................................................................... 31
C
OMM FAILURE
S
ENSOR READING INCORRECTLY
W
ATER IN THE VENT TUBE
O
THER ISSUES
.......................................................................................................................... 31
......................................................................................................... 32
........................................................................................................................... 32
CHAPTER 6: WARRANTY ................................................................................... 34
APPENDIX A: WL15 DETAILED SPECIFICATIONS ........................................... 35
D
ATALOGGER
S
ENSING ELEMENT
H
OUSING
C
....................................................................................................................................... 36
ABLE
............................................................................................................................. 35
..................................................................................................................... 36
.................................................................................................................................... 36
APPENDIX B: WORKSHEETS ............................................................................. 37
C
ALIBRATING THE BAR GRAPH TO HIGHER OR LOWER RANGE
.............................................................................................. 5
............................................................................... 10
..................................................................................................... 19
................................................................................................ 32
.................................................... 37
Global Water (800)876-1172 www.globalw.com
Page 3
Page 4
C
ALIBRATING LEVEL SENSORS FOR DEPTH TO WATER READINGS
............................................... 38
APPENDIX C: CABLE CONFIGURATION ........................................................... 39
APPENDIX D: TERMINAL PROGRAMMING ....................................................... 40
C
OMMUNICATION COMMANDS FOR WINDOWS DATALOGGER
........................................................ 40
Global Water (800)876-1172 www.globalw.com
Page 5
Introduction
Introduction
1
The WL15 Water Level Logger
Congratulations on your purchase of the Global Water WL15 Water Level Logger. This instrument has been quality tested and approved to provide accurate and reliable water level (water pressure) measure­ments. We are confident you will find the WL15 to be a valuable asset for your applications. Should you require assistance, our technical staff will be happy to help.
The WL15 provides a datalogger with a submersible pressure transducer for remote monitoring and re­cording of water level, flow or pressure data. This highly reliable and accurate water logger records 24,400 readings and is programmable from one reading per second to one reading per day. Ranges of 0-3', 0-15', 0­30', 0-60', 0-120' and 0-250' are available. A 25' cable is standard, and optional cable lengths are available up to 1000'.
The WL15’s datalogger is housed in a weather-resistant cylindrical enclosure that slips inside a 2-inch pipe. The logger is easily adapted with standard hardware for well head mounting or other installations. The WL15 includes Windows-based software, allowing easy upload of data to standard spreadsheet programs on a PC computer.
The WL15's level sensor electronics are fully encapsulated with marine-grade epoxy, providing superior moisture protection over O-ring seals. During this process, the vent tube is sealed directly to the sensing element, preventing any moisture from damaging the sensor’s electronic components.
The level sensor uses a unique silicon diaphragm to interface between the water and the sensing element. This silicon diaphragm is highly flexible and touches the sensing element, producing a sensor with excep­tional linearity and very low hysteresis. The pressure transducer utilizes a stainless steel micro-screen cap to protect the sensing element. This protective cap has hundreds of openings, making fouling the sensor with silt, mud or sludge virtually impossible.
Features
Data
Windows based Global Logger* software
Tabular display/printout
Data in standard spreadsheet format
Communication cable included (RS232C)
Logger
Weather resistant enclosure
Fits inside 2” pipe
Records 24,400 readings
12 bit or 1/4,000 resolution
Battery powered
Cable
Marine grade, heavy urethane jacket
Barometric pressure vent
Lengths up to 1000’
Sensor
Fully encapsulated submersible pressure trans-
ducer Amplified and temperature compensated
High accuracy
3/4” diameter wet-wet transducer (no vent mois-
ture problems)
*Copyright © Global Water Instrumentation, Inc. 2001
Global Water (800)876-1172 www.globalw.com
Page 5
Page 6
Chapter 1
Checklist
Your WL15 was carefully inspected and certified by our Quality Assurance Team before shipping. If any damage has occurred during shipping, please notify Global Water Instrumentation, Inc. and file a claim with the carrier involved. Use the checklist to ensure that you have received everything needed to operate the WL15. Your sensor has been inscribed with the sensor range for easy identification.
Cable
Communication Cable
Global Logger Software*
Datalogger
User Manual
Level Sensor
WL15 (Datalogger, Level Sensor and Cable)
User Manual
Communication Cable (RS232C COM Cable)
Global Logger Software (3 1/2” Floppy Diskette)
*Copyright © Global Water Instrumentation, Inc. 2001
Global Water (800)876-1172 www.globalw.com
Page 7
Getting to Know the WL15
The COM port can connect to a PC serial port using the included Communication Cable.
Introduction
COM port
lithium battery
The Communication Cable has two different connectors. Connect the DB-9 female connector to your PC and the 4-pin mini socket* to your Datalogger.
Unscrew logger cap to access lithium bat­tery.
DB-9 female connector 4-pin mini socket
Appendix C: Cable Configuration
*See
Global Water (800)876-1172 www.globalw.com
, page 39
Page 7
Page 8
Chapter 1
The water level sensor is located inside the sensor housing, near the mesh screen.
sensor location
Global Water (800)876-1172 www.globalw.com
Page 9
Setup
Setup
Installing the WL15
System Time Check
Before you install the software, check the accuracy of the time and date on your computer. The WL15 will be programmed with this information.
To verify your system time in Windows, click
tings Control Panel
from the task bar.
Start Set-
2
Global Water (800)876-1172 www.globalw.com
Double-click the
Date/Time
icon.
Page 9
Page 10
Chapter 2
Verify that your system is set to the cor­rect time. If not, make the necessary changes and click the Apply button fol­lowed by the OK button. Exit the control panel.
Global Logger Software Installation
Insert the 3 1/2” floppy diskette labeled “Global Logger Software Version x.xx” (where x.xx is a version number such as 1.36) into your computer’s floppy disk drive. Normally, the floppy will be the computer’s “A” drive.
My Computer
).
icon on your desktop.
Double-click the floppy icon (normally listed as
Double-click the
3 ½ floppy (A:)
Page 10
Global Water (800)876-1172 www.globalw.com
Page 11
Setup
Click on the file
Global Logger.exe
.
Global Water (800)876-1172 www.globalw.com
Edit Copy
Click
from the menu bar.
Page 11
Page 12
Chapter 2
Save the file to a directory of your choosing. Note, it should be a location you will easily remember. To save to the directory Global Logger located on your main “C” hard drive (although “C” is normally the main system drive, the designation can vary from system to system), start by clicking the down carrot of the address field. Click on the
Local Disk (C:)
icon or appropriate main system drive designation.
down carrot
The directory
New Folder
lected for typeover. Type
If the directory listed, click the menu bar.
should now be created and automatically se-
Global Logger
and press the
enter
key.
Global Logger
is not
File New Folder
from
Page 12
Global Water (800)876-1172 www.globalw.com
Page 13
Double-click on the new
Global Logger
directory.
Setup
Copying…
A
dialog box should now display.
Wait for the copy process to complete.
Edit Paste
Click
from the menu bar.
Global Water (800)876-1172 www.globalw.com
Page 13
Page 14
Chapter 2
To create a desktop shortcut, right-click the and click on Desktop (create shortcut).
Global Logger.exe
file you previously saved, select
Send To
You can now access your Global Logger* software directly from the desktop.
Hardware Installation
General Installation Tips
The WL15 has many applications and therefore many installation options. The WL15’s pressure sensor
is fully submersible and may be suspended by its waterproof cables in the water to be monitored.
The data logger is not waterproof. It is water-resistant. This means that condensation, which can
build up inside the enclosure during humid conditions, should not affect the performance of the WL15. However, if the data logger is submerged in water, it will corrode the electronics and cause it to stop working properly. Please ensure that the data logger will be clear from flooding and out of direct rain­fall.
Do not install the WL15 in applications that contain solvents
rate the cable and the sensing element.
The accuracy of the sensor readings can be affected by such adverse conditions as overpressure, lighten-
ing strikes, improper care/handling, and physical damage or abuse.
. Over time, many solvents can deterio-
Page 14
Global Water (800)876-1172 www.globalw.com
Page 15
Setup
The WL15’s cable contains a vent tube that allows for barometric pressure compensation.
Do not allow
the vent tube to become kinked or obstructed as this may cause erroneous readings.
Install your WL15 so that it is easily accessible for calibration purposes. You may need to remove and
reinstall it in the future, so plan ahead!
The pressure sensor will not function correctly if buried by mud, silt, leaves, or other debris. Install the
sensor in a way that avoids these conditions.
Groundwater Installation
For best results suspend the pressure sensor in a 2” monitoring well near the well screen. The sensor
may be installed in well diameters 1” and up.
For a 2” diameter pipe (or larger), the sensor can be inserted directly. Slowly lower the sensor into the
well. Warning:
sensor
. The top of the data logger will rest on the top of the well pipe. A slip cap or a locking well cap
Do not allow the sensor to “freefall” down the well, this may cause damage to the
can conceal the WL15. The cable length of the WL15 should be sized so the sensor will be positioned below the lowest expected water level, which is not necessarily the bottom of the well. The correct ca­ble length is measured from the top of the data logger to the bottom of the sensor. The sensor’s measur­ing range should be slightly larger than the expected total amount of change in water level.
To shorten the cable in a 2” diameter or larger well, coil it around a 1” diameter stick or pipe and secure
the cable at both ends with tie wraps or tape. Warning: Do not kink the cable; this may damage the internal vent tube
.
The WL15 can be adapted to larger or smaller well casings by using standard PVC pipe reducers and
fittings or Global Water’s Inside Well Option.
As factory programmed, the WL15 records depth of water above the sensor. If you want to use your
WL15 to record depth to water readings, follow the worksheet
Calibrating Level Sensors for Depth to
Water Readings*.
Surface Water Installation
The sensor may be submerged at the monitoring point and hung from its cable. It is recommended to
protect the sensor and datalogger inside a 4” PVC drainpipe that will act as a protective stilling well. PVC schedule 40, or ABS sewer drainpipe is recommended. PVC schedule 40 electrical conduit has UV protectors and pre-formed “sweeps” or bends that enable it to conform to the contours of a river­bank. The level sensor will slide down through 45° or 90° sweeps. The pipe may be buried in the river­bank, secured with rocks, or fastened to the bank with large staples made by bending pieces of concrete reinforcing steel in half and driving them into the bank. Put a cap on the bottom end of the pipe to al­low easy water flow past the sensors. Drop the sensor until it touches the bottom of the stilling well and then pull it up slightly and secure the cable. The pipe should have several ½” diameter holes drilled near the sensor location in order to eliminate any velocity effects on the sensor. In addition, a smaller ¼” hole should be drilled near the top of the pipe to allow air movement when the water goes up and down. A standard slip cap or a locking well cap can be used to secure the WL15. The pipe can also be adapted for a screw-on cap.
Appendix B: Worksheets
*See
Global Water (800)876-1172 www.globalw.com
, page 38
Page 15
Page 16
Chapter 2
In areas where vandalism is prevalent, 4” diameter galvanized steel pipe may be used for any exposed
pipe. Plastic pipe, adapted to the steel pipe above water, may be used for the underwater section. A screw-on galvanized cap can then be used to cover the data logger. Two small pipe wrenches are re­quired to access the WL15, but this installation is essentially bulletproof and vandal-proof.
Special Installations
For flumes and weirs Global Water provides the Weir Stick Water Level Logger. This logger/sensor
combination can be mounted inside a 2” diameter pipe that acts as its own stilling well, or the Weir Stick can be mounted in an existing stilling well. The weir stick is 3’ overall in length with a 2’ meas­uring range.
For measuring pressure in pipes, the level sensor may be purchased with a ¾” NPT thread on the sens-
ing end. The sensor threads can be adapted to existing ball valves with standard pipefittings.
The WL15 Sewer Pipe Flow Option includes a level sensor built into a protective “mouse” housing and
attached to a 12” stainless steel strap. When the sensor/straps are installed at the invert of a sewer or in a pipe, the data logger can be attached to a sewer’s stair rungs. If there is enough slack in the cable, the data logger can be accessed from the manhole above the stair rungs.
When installed in a pipe that is under 12” diameter, the steel straps spring into the pipe, holding the sen-
sor in place. Additional fasteners are usually not required for this type of installation.
For a sewer pipe over 12” diameter, the sensor’s steel straps must be mounted into the pipe. The straps
can be secured with molly or concrete bolts, or with marine-grade epoxy.
Storing the WL15
The WL15 may be stored without any special provisions. Place the datalogger and sensor inside a bag to keep the instrument clean and store on a shelf or hang on a wall.
Page 16
Global Water (800)876-1172 www.globalw.com
Page 17
Using Global Logger Software
Using Global Logger Software
3
Hardware Connection
Connect the DB-9 female connector of the included communication cable (RS232C) to the COM 1 serial port on your computer. If COM 1 is being used, either disconnect and/or disable the device on COM 1 or connect to COM 2*.
4-pin mini socket
Connect the 4-pin mini socket of the communication cable (RS232C) to your WL15 COM port.
Global Logger Startup
Double-click the Global Logger icon on your desktop if you created a desktop shortcut during software setup. Otherwise, access the directory in which your the stored and double-click on the file. If you have not installed the Global Logger software or if you want to create a desktop icon, see
Global Logger.exe
Global Logger Software Installation
file is
**.
* See Comm Failure, page 32.
Global Logger Software Installation
**See
Global Water (800)876-1172 www.globalw.com
, pages 10 – 14
Page 17
Page 18
Chapter 3
The Global Logger main window will display data in the following fields:
Logger Name, Date/Time, Re-
cording Interval and Recordings in Memory.
Access the Global Logger functions
ory, Synchronize Time
and
Setup
functions can also be accessed through
Sample Continuously, Get Settings, Get History Data, Clear Mem-
through the main menu check box and buttons. All Global Logger
Action
on the menu bar.
The bar graph on the left side of the screen indicates current depth of sensor. For example, the graph in the following screen shot displays a 120 foot sensor with a reading of 79.83 feet. This data indicates that the sensor is currently at a depth of 79.83 feet. Sensors are calibrated to a variety of unit specifications (i.e. me­ters), so your bar graph readings may differ slightly.
depth bar graph
Global Logger functions
Page 18
Global Water (800)876-1172 www.globalw.com
Page 19
Global Logger Functions
Sample Continuously/Real-Time Values
Using Global Logger Software
To display real-time values every second, select the
Sample Continuously
checkbox. Readings will display below the depth bar graph. To stop readings, unselect checkbox. Note, that real-time data is not recorded; therefore, you cannot retrieve it from WL15 memory.
current reading
past readings
Get Settings
To refresh information in the Global Logger main window, click the update the
Logger Name, Date/Time, Recording Interval
and
Get Settings
button. This action will
Recordings in Memory
fields.
Get History Data
To retrieve logged data from memory, click the depend on the number of data points recorded (more recordings equals longer upload time). To minimize long upload times, clear memory* after backing it up to disk**.
* See Clear Memory, pages 21 – 22
Get History Data
**See
Global Water (800)876-1172 www.globalw.com
, pages 20 – 21
Get History Data
button. The speed of your upload will
Page 19
Page 20
Chapter 3
Historical Data Viewer
A
window will display. This window lists each data point by date/time and read­ing value. If there are more data points in memory than will fit in the data view window, use the scroll bar to scroll through the list of readings.
readings listed by Date/Time and reading value (in this case Feet)
use scroll bar to scroll through list
If you want to save current readings, click the
Save As
A
window will display. If you want to save your file to a directory other than the one displayed, click the down carrot of the
Save in:
field and click on the device (i.e. hard disk drive, floppy disk drive, etc.) you want to save the file to.
Save to File
button.
down carrot
Page 20
Global Water (800)876-1172 www.globalw.com
Page 21
Double-click the directory or subdirec­tory you want to save to. If you want to change the file name click anywhere in
File name:
the
field and type the desired name. However, do not change the .csv extension. This extension ensures the file will be recognized as a comma sepa­rated value (*.CSV) file. You can open CSV files in spreadsheet programs such as Microsoft Excel*. Click
Save
to save
the file.
Using Global Logger Software
change name in file name field
To remove zero readings, click the ing. Since the pack function does not change the readings stored in memory, you will need to
Pack
button. Note, this operation will not remove the initial zero read-
Save to File
to preserve the pack operation. If you exit and then reenter the Historical Data Viewer window, all read- ings (including zero readings) stored in memory will display regardless of any prior pack operation.
Click OK to exit the
Historical Data View
window.
Clear Memory
To clear your WL15’s memory of readings click the
Clear Memory
button.
A confirmation dialog box will display. Click
No
to keep readings.
*Windows® Excel is a trademark of the Microsoft Corporation
Global Water (800)876-1172 www.globalw.com
Yes
to clear history or
Page 21
Page 22
Chapter 3
If you click
Yes
Standby
, a
graphic will display in the middle of
your monitor screen while the memory is purged.
Synchronize Time
To synchronize your WL15’s internal clock to your computer’s clock*, click the
Synchronize Time
button.
Standby
A
graphic will display in the middle of your monitor
screen while synchronizing.
Setup
Note that your Water Level Logger has come completely setup from our factory. It has been calibrated and tested for optimal accuracy and performance. If you choose to recalibrate your sensor, use the following procedures.
To begin calibration and/or to change time interval settings, click the
Setup
window will display.
Setup
button. A
Global Water
System Time Check
*See
Page 22
, pages 9 – 10
Global Water (800)876-1172 www.globalw.com
Page 23
Datalogger Name and Sensor Warmup Time
Using Global Logger Software
If you want to change the datalogger name, click in the
Global Water recommends that the
Sensor Warmup Time
Name
field and type in a new name.
field be set at 3 seconds to supply the level sen­sor with adequate power for accurate data measurements. This setting dictates the length of time power is applied to the level sensor before the datalogger takes a reading.
For example, if you set your WL15 to record a reading every 20 minutes at fixed intervals and the process begins at 12:00:00, power will be applied to the level sensor immediately and a reading taken at 12:00:03. At 12:20:00, power will again be applied to the level sensor, and the datalogger will take a reading at 12:20:03. Note that if the recording interval fields are less than or equal to the
Sensor Warmup Time
field, the sensor is powered continuously.
To change warmup time, click in the
Sensor Warmup Time
field and enter the desired time (0 –15 sec­onds). You may also click the up/down arrows to the right of the text field to increase or decrease the value by 1.
recording interval fields
Global Water (800)876-1172 www.globalw.com
Page 23
Page 24
Chapter 3
Calibration
Change the feet, etc.) by clicking in the field and typing the unit name. Note that the
Sensor Type or EU
(engineering units) field to the unit type you are calibrating to (i.e. meters,
Sensor Type or EU
field is only a
text label. Actual readings are determined by the following calibration steps.
In case you make a mistake during the calibration process, write down the existing values in the
Low Raw
and
fields before continuing (in this case
64435
and
12704
). Click the
Calibrate
button.
High Raw
Enter the maximum unit range of the level sensor (
120
for a 120’ sensor in this example) in the highest value text field and click the
Next
button.
If performing a dry calibration, insert the sensor into your water pressure meter and set to full scale range of your WL15. If performing a wet calibration, insert level sensor to a water depth of at least 10% of full scale range. For example a 120’ sensor should be calibrated at a depth of at least 12’. Calibrating at greater per­centages (up to 100%) provides greater accuracy. Calibration depth should be measured from water surface to sensor tip.
Page 24
Global Water (800)876-1172 www.globalw.com
Page 25
For both wet and dry calibrations, wait at least five seconds for the level sen­sor to adjust to current PSI level. Click the
Next
button.
Using Global Logger Software
Enter 0 in the and click the
lowest value
Next
button.
text field
For dry calibrations, release the level sensor from water pressure meter. For wet calibrations, take level sensor out of water.
For both wet and dry calibrations, wait at least five seconds for the level sen­sor to adjust to 0 PSI. Click the Finish button.
Note, that if performing a wet calibration at a fraction of total sensor range, the bar graph on the left side of the main at the calibration depth instead of actual total sensor range. All other logger operations will perform correctly. To revise calibration settings so that the graph will function cor­rectly, follow the worksheet in Appendix B*.
Appendix B: Worksheets
*See
Global Water (800)876-1172 www.globalw.com
, pages 37
Global Logger
window will interpret full scale range
Calibrating the Bar Graph to Higher or Lower Range
Page 25
Page 26
Chapter 3
Verify that
High EU
and
Low EU
fields reflect desired unit readings. Also, verify that the
High Raw
reflects a range of 60,000 – 65,000 and that the Low Raw field reflects a range of 10,000 – 15,000.
Adjust the
Decimal Places
field according to the following chart (i.e. a 3’ sensor should be adjusted to three
decimal places).
wl sensor range decimal places
3’ 3
15’ 2
30’ 2
60’ 1
120’ 1
250’ 1
field
Recording Interval and Memory Management
Select the interval type by clicking the down carrot of the specified interval, click
for fixed interval
.
Sample
down carrot
field. To repeatedly record values at a
Note, that if you choose to record at fixed intervals, memory will fill up according to chart on the right. Refer to battery maintenance chart for battery life to available memory comparisons*.
Battery
*See
Page 26
, page 29
Global Water (800)876-1172 www.globalw.com
Recording Interval Time
1 second 6 hours, 49 minutes
1 minute 17 days
5 minutes 84 days
15 minutes 254 days
30 minutes 509 days
1 hour 1018 days
Page 27
Using Global Logger Software
Time Frame Readings per time
0-19 seconds As fast as possible 1000
To record values according to a logarithmic curve, click
for logarithmic interval
. See chart for timeline speci-
fications.
20-59 seconds 1 per 1 second 40
60-599 seconds 1 per 12 seconds 45
10-99 minutes 1 per 2 minutes 45
100-999 minutes 1 per 200 minutes 45
1000-9999 min-
To take readings upon exceeding a specified unit range, click
utes
1 per 1440 minutes 45
upon exception
unit
. This setting sets the logger
Expected total readings
during Time Frame
to record a single reading after detecting a ± variance from the previously recorded exception reading. Click the long bar appearing at the bottom of the Global Logger Setup menu
Click and hold toggle slider with left mouse button (i.e. do not click and release). Move slider left or right to select desired exception range, and release mouse button. Click OK to preserve selection or
Cancel
to revert to the previous
toggle slider
setting.
The bar should change to reflect the latest exception setting. The setting in these screenshot examples sets the logger to record a single reading after water level changes (rises or falls) by 10.2 feet from the previous exception reading.
Set the recording interval digit, by clicking in the
every
text field, and typing the desired number. You may
also click the up/down arrows to the right of the text field to increase or decrease the value by 1.
up/down arrows
Set the recording interval time unit, by clicking the down carrot of the time unit field (not labeled) and clicking Seconds, Minutes, Hours or Days.
Global Water (800)876-1172 www.globalw.com
Page 27
Page 28
Chapter 3
Wrap records at end of storage
The
checkbox dictates whether or not the logger continues to record read­ings after its memory is filled. To overwrite readings in synchronous order (earliest readings are overwrit­ten first) after memory is filled, click the checkbox. To preserve readings and stop recording after memory is filled, leave box unchecked.
After calibrating and/or changing interval recording settings, click the OK button. To cancel changes, click
Cancel
.
You may also manually change WL15 settings directly using a terminal program*.
Appendix D: Terminal Programming
*See
Page 28
, page 40
Global Water (800)876-1172 www.globalw.com
Page 29
Maintenance
Maintenance
4
Battery
Global Water recommends that the battery voltage be physically checked based on your recording interval*. Refer to the following table.
Note, that the WL15 uses 60µA while “sleeping” and 3mA plus the signal current while taking readings. This table is based on the worst-case scenario where the sensor is providing the maximum signal of 20mA.
Temperature can also affect battery life. A WL15 in an environment that has a sustained temperature of 23° C will run 5% longer than a unit that has a sustained temperature of 0° C.
Recording Interval Theoretical Battery Life Recommended Time Be-
tween Battery Check
1 second 43 hours 36 hours
1 minute 34 days 1 month
30 minutes 424 days 6 months
1 hour 526 days 6 months
1 day 685 days 6 months
Checking the battery requires that the WL15 cap be unscrewed. The battery should be visible and can be gently removed from the housing with fingers or a pair of pliers.
Recording Interval and Memory Management
*See
Global Water (800)876-1172 www.globalw.com
, pages 26 – 28
Page 29
Page 30
Chapter 4
The battery must be disconnected from the battery strap.
but will not lose any recordings or settings.
Test the battery with a voltage meter. The sensor will not work reliably if the battery voltage is below
7.5V. If the battery needs to be replaced use a 9V lithium battery for reliable service. Re-attach the battery to the battery strap and gently slide it into the housing. Screw the cap on to the data logger housing, being careful not to pinch any of the wires.
Note, that the WL15 will stop recording,
Calibration
The level sensor’s calibration must be periodically verified. Global Water recommends checking the sen­sor’s calibration with a sounder or other measuring device once every 6 months.
Sensor Screen Cleaning
The screen on the end of the sensor must be periodically checked for clogging from mud and sludge. Wash the screen with clean water and/or scrub it gently with a toothbrush. Do not insert objects through the screen, as this may cause damage to the sensor.
Page 30
Global Water (800)876-1172 www.globalw.com
Page 31
Troubleshooting
Troubleshooting
5
Comm Failure
If Global Logger software displays
Verify that the communication cable is securely connected to the data logger and to your computer*.
Check the battery voltage on the logger**.
Make sure that your computer is correctly set up to use your communication channel 1 (COM 1). Refer
to the settings below for comparison.
Comm Failure
in the
Logger Name
field, perform the following steps
* See Hardware Connection, page 17
Battery
**See
Global Water (800)876-1172 www.globalw.com
, page 29
Page 31
Page 32
Chapter 5
If you only have access to COM 2, then an *.ini file must be created. Open
Notepad
and type the text exactly as it appears in the following screenshot. Change the directory to the Windows folder of your main drive if you are using Windows 95, 98, ME or XP or to the you are using Windows NT or 2000. Save the file as
Make sure that no other software or device is using COM 1, such as an infrared port or palm synching
global logger.ini
WINNT
.
folder of your main drive if
software.
Sensor Reading Incorrectly
Verify that the vent tube has not been kinked or sealed. The sensor uses this tube to compensate for
barometric pressure changes. Clean the sensor following the maintenance instructions*.
Recalibrate the sensor using the calibration methods described in the
ual**.
Calibration
section of this man-
Water in the Vent Tube
If water gets into the cable’s vent tube, place it next to a heater for 24 hours to dry the inside of the cable.
Other Issues
Refer to the manual for proper set-up.
Check all connections.
Call Global Water for tech support: 800-876-1172 or 916-638-3429 (many problems can be solved over
the phone). Fax: 916-638-3270 or Email: please have the following information ready;
1. Model number
2. Unit serial number
3. P.O. number the equipment was purchased on.
4. Our sales number or the invoice number.
5. Repair instructions and/or specific problems relating to the product.
Be prepared to describe the problem you are experiencing including specific details of the application, installation, and any additional pertinent information.
globalw@globalw.com.
When calling for tech support,
* See Sensor Screen Cleaning, page 30
Calibration
**See
Page 32
, pages 24 – 26
Global Water (800)876-1172 www.globalw.com
Page 33
In the event that the equipment needs to be returned to the factory for any reason, please call to obtain
Troubleshooting
an RMA number (Return Material Authorization). Do not return items without an RMA number dis­played on the outside of the package.
Clean and decontaminate the WL15 if necessary.
Include a written statement describing the problems.
Send the package with shipping prepaid to our factory address. Insure your shipment, since Global Water’s warranty does not cover damage incurred during transit.
Global Water (800)876-1172 www.globalw.com
Page 33
Page 34
Chapter 6
Warranty
Global Water Instrumentation, Inc. warrants that its products are free from defects in material and work­manship under normal use and service for a period of one year from date of shipment from the factory. Global Water’s obligations under this warranty are limited to, at Global Water’s option: (I) replacing or (II) repairing; any products determined to be defective. In no case shall Global Water’s liability exceed the products original purchase price. This warranty does not apply to any equipment that has been repaired or altered, except by Global Water Instrumentation, Inc., or which has been subject to misuse, negligence or accident. It is expressly agreed that this warranty will be in lieu of all warranties of fitness and in lieu of the warranty of merchantability.
The warranty begins on the date of the invoice.
6
Page 34
Global Water (800)876-1172 www.globalw.com
Page 35
WL15 Detailed Specifications
Appendix A: WL15 Detailed Specifications
Datalogger
Inputs
Channels: 1 Analog
Range: 4 – 20 mA
Storage Resolution: 1 / 65535 (16 bit)
A/D Resolution: 1 / 4095 (12 bit)
Data Storage
Memory Size: 2MB (24,000 recordings)
Type: non-volatile Flash
Intervals: 1 – 256 seconds, 1 – 256 minutes, 1 – 256 hours, or 1 – 256
days
Time Accuracy: 0.0015%
Time Format: Month / Day / Year Hour / Minute / Second
Communication
Speed: 38,400 kbs
Data Transfer: 2MB / 4 min.
Type: RS232C (Serial)
Electrical
Battery: 9V Lithium (1A / h)
Back-up Battery: 3V Lithium (950 mA / h)
Sleep Current: 60µA
Wake Current: 3 mA + Sensor (4 – 20 mA)
Communication Current: 100 mA
Environmental
Operating Temperature: -40 to 75° C (-40 to 170° F)
Storage Temperature: -50 to 90° C (-60 to 200° F)
Humidity: 0 to 95% Non-Condensing (Conformal Protection Coating)
Physical Dimensions
Body OD: 45mm (1.77”)
Cap OD: 57mm (2.24”)
Length: 315mm (12.4”)
Weight: ~450g (16 oz.)
A
Global Water (800)876-1172 www.globalw.com
Page 35
Page 36
Appendix A
Sensing Element
Output: 4 – 20 mA
Range: 0 – 3’, 0 – 15’, 0 – 30’, 0 – 60’, 0 – 150’, 0 – 250’
Accuracy: ±0.2% FS (35° F to 70° F)
Linearity and Hysteresis: ±0.1% FS
Resolution: Infinitesimal (Analog)
Operating Voltage: 10 – 36 VDC
Current Draw: 4 – 20 mA
Warm-up Time: 10 ms minimum
Operating Temperature: -40 to +85°C (-40 to +185°F)
Size of Probe: 3/4" diameter x 4 1/2" long
Weight: 0.5 lb
Overpressure: X4
Burst Pressure: X10
Material: Fluorosilicone
Type: Differential (Wet/Wet)
Compensated Temperature Range: 0 to 50°C (32 to 122°F) submerged
Housing
Material: 304L SS
Diameter: 19 mm (0.75”)
Length: 200 mm (8”)
Weight: ~115g (4 oz)
Cable
Conductors: 4 each 22 AWG
Jacket Material: 87A shore hardness Polyurethane
Optional Jacket: Fluorinated Ethylene Propylene (FEP) Teflon
Cable OD: 7.8 mm (0.307”)
Vent Tube: HD Polyethylene
Shield: Aluminum Mylar
Temperature range: -30 – 85°C (-22 – 185°F)
Weight: ~65 g / m (0.7 oz / ft)
Page 36
Global Water (800)876-1172 www.globalw.com
Page 37
Worksheets
Appendix B: Worksheets
B
Calibrating the Bar Graph to Higher or Lower Range
With some sensors it is either impractical or unfeasible to reach the full-scale or lower scale of its range. This will not affect the actual reading of the datalogger but it will cause the bar graph to work incorrectly.
Here are the equations for finding new limits on the bar graph. First take the two coordinates that were found during calibration*.
(High EU __________, High Raw __________) and (Low EU __________, Low Raw __________).
Now calculate the slope using this equation:
m = (High Raw – Low Raw) / (High EU – Low EU) m = ___________
Next find the y-intercept:
b = High Raw - (m x High EU) b = ___________
Now you can calculate the new “Raw” value for any Engineering Unit.
New Raw = (m x New EU) + b New Raw = ___________
Finally, enter these new “Raw” values into your datalogger and verify the readings are correct*.
Calibration
*See
Global Water (800)876-1172 www.globalw.com
, pages 24 – 26
Page 37
Page 38
Appendix B
Calibrating Level Sensors for Depth to Water Readings
Note, that to take depth to water readings, the bar graph must be cali­brated to the full scale of the sensor. Follow the worksheet ing the Bar Graph to Higher or Lower Range*.
The first step is to find the Total Depth of the sensor. This is the dis­tance from the end of your sensor to your datum. (Cable lengths from the factory are not exact. You may need to measure for precise dis­tances.)
Total Depth = _____________
Next, find the maximum Sensor Depth (range of the sensor). In the
Calibration
** section of the software, you can find this value by us-
ing the following equation:
Sensor Depth = | High EU______ - Low EU______ | Sensor Depth = ___________
Now you can find your new Engineering Units using these steps:
New Low EU = Total Depth New Low EU = ___________
New High EU = Total Depth – Sensor Depth New High EU = __________
Finally, enter these New EU values into your datalogger and verify the readings are correct**.
Calibrat-
* See Appendix B: Worksheets, page 37
Calibration
**See
Page 38
, pages 24 – 26
Global Water (800)876-1172 www.globalw.com
Page 39
Cable Configuration
Appendix C: Cable Configuration
The following diagram details the pin layout of the male communications cap on the WL14/15 Datalogger series. The chart explains the communications wiring.
Comm. Cap Pin Number
1 Green Dig. Ground 5
2 Blue DTR 4
3 Violet Rx 3
4 Red Tx 2
Comm. Cap
Wire Color
Function 9 pin D-sub
Pin Number (Male)
C
Global Water (800)876-1172 www.globalw.com
Page 39
Page 40
Appendix D
Appendix D: Terminal Programming
D
Communication Commands for Windows Datalogger
The following is a chart of commands that can be sent to the Windows Datalogger through a terminal pro­gram.
Command Description
D Set Date
T Set Time
? Get number of data points
h Get first historical data point
a Get historical data point again
X Abort
i Get next historical data point (increment)
E Erase
r Get real time data point
! Set description
1 Get description
@ Set Units
2 Get Units
# Set Raw Scale
3 Get Raw Scale
$ Set Engineering Unit Scale (EU)
4 Get Engineering Unit Scale (EU)
N Set Name
n Get Name
S Set Sample interval
s Get Sample interval
Page 40
Global Water (800)876-1172 www.globalw.com
Loading...