Global Water SP200 Instruction Manual

Page 1
Global Water
800-876-1172 globalw.com
Global Water
Instrumentation, Inc.
Gold River, CA 95670
T: 800-876-1172
Int’l: (916) 638-3429, F: (916) 638-3270
Variable Speed Portable Sample Pump: SP200
06/14/06
- 1 -
Page 2
Global Water
800-876-1172 globalw.com
Congratulations on your purchase of the Global Water SP200 Variable Speed Portable Sample Pump. This instrument has been quality tested and approved for providing accurate and reliable measurements. We are confident that you will find the SP200 to be a valuable asset for your application. Should you require assistance, our technical staff will be happy to help.
Table of Contents
I. Checklist
Page 3
II. Inspection
3
III. Description
4
IV. Using the Variable Speed Portable Sample Pump
4
V. Specifications 5 VI. Maintenance
5
VII. Troubleshooting
6
VIII. Warranty
7
IX. Appendix A: Accessories
8
* Copyright © Global Water Instrumentation, Inc. 2003
- 2 -
Page 3
I. SP200 Checklist
a. Variable Speed Portable Sample Pump b. 15’ pickup hose c. 10’ Battery Cable (Inside Enclosure) d. SP200 Manual e. 12 volt battery (Not included)
II. Inspection
a. Your SP200 unit was carefully inspected and certified by our Quality
Assurance Team before shipping. If any damage has occurred during shipping, please notify Global Water Instrumentation, Inc. and file a claim with the carrier involved.
Use the checklist to ensure that you have received everything needed to operate the SP200.
Global Water
800-876-1172 globalw.com
- 3 -
Page 4
Global Water
800-876-1172 globalw.com
III. Description
a. The Global Variable Speed Portable Sample Pump, SP200, is
designed specifically for the user that needs to take samples in areas where it is not reasonable to have a permanent sampler. The sampler has a speed adjustment knob that allows the user to determine what the sampling rate is. The pump is also reversible, so the pickup hose can be purged prior to moving the pump or taking the next sample.
b. The Global Variable Speed Portable Sample Pump consists of a
rugged, water resistant, lockable carrying enclosure. The enclosure contains an external peristaltic sampling pump, a 15’ pickup hose with strainer, and a 10’ battery cable.
IV. Using the Variable Speed Portable Sample Pump
Pick up hose set up
a. The pickup strainer should be submerged under water and should be
situated to avoid contact with the bottom.
Sampling unit
b. The sampler should be placed upright. c. Secure the battery cable to the battery connector on the enclosure’s
front panel.
d. Ensure the toggle switch is in the off position. Attach the battery
cable’s leads to a 12V battery. e. Place the sample bottle under the output spout. f. Push the toggle switch into the forward position to begin pumping.
Adjust the pump speed with the control knob. g. Push the toggle switch to the off position to stop the pump or turn the
speed control knob to zero velocity. Once the sample is taken, the
toggle switch can be used to reverse the pump’s direction to purge the
pick up hose.
- 4 -
Page 5
V. Specifications
Sample Size: Variable
Operating Temperature: 0° to +70°C
Size of unit: 9" x 7.5" x 4.5"
Weight: 4.5 lb
Operating Voltage: 12 V
Current Draw: 2A max
Materials:
Enclosure: Expanded UV protected PVC Sample Tubing: 15' nylon reinforced 1/4" ID
Sample Pump:
Flow Rate: Variable (500 ml per minute max at Purge Rate Same
Type: Peristaltic Pump Tubing: Norprene 1/4” ID
Maximum Lift: 22'
VI. Maintenance
Sampler
Global Water
800-876-1172 globalw.com
polyethylene flexible tubing with intake strainer
4 ft. head)
a. The Global Water SP200 Variable Speed Portable Sample Pump
requires minimal maintenance. The sampler enclosure is rain resistant
and rugged. Avoid exposure to extremely rough usage. Routinely
wipe down the carrying case, rinse the pickup hose and debris strainer
with mild soap and warm water. Additional norprene tubing for the
sampler pump, removable debris strainers, and pickup hose can be
purchased from Global Water (see Appendix A section).
- 5 -
Page 6
VII. Trouble Shooting
Issue: The pump does not work a. Test the battery voltage and ensure that it has a full charge. b. Test the battery cables for continuity. c. Open enclosure verify that battery and motor connections are attached
to circuit board. Other issues
d. Call us for tech support: 800-876-1172 or 916-638-3429 (many
problems can be solved over the phone). Fax: 916-638-3270 or Email:
globalw@globalw.com.
Be prepared to describe the problem you are experiencing including
specific details of the application and installation and any additional
pertinent information. e. In the event that the equipment needs to be returned to the factory for
any reason, please call to obtain an RMA# (Return Material
Authorization). Do not return items without an RMA# on the outside of
the package.
Include a written statement describing the problems.
Send the package with shipping prepaid to our factory address. Insure
your shipment, as the warranty does not cover damage incurred during
transit. f. When calling for tech support, please have the following information
ready;
1. Model #.
2. Unit serial number.
3. P.O.# the equipment was purchased on.
4. Our sales number or the invoice number.
5. Repair instructions and/or specific problems relating to the product.
Global Water
800-876-1172 globalw.com
- 6 -
Page 7
VIII. Warranty
a. Global Water Instrumentation, Inc. warrants that its products are free
from defects in material and workmanship under normal use and service for a period of one year from date of shipment from factory. Global Water’s obligations under this warranty are limited to, at Global Water’s option: (I) replacing or (II) repairing; any products determined to be defective. In no case shall Global Water’s liability exceed the products original purchase price. This warranty does not apply to any equipment that has been repaired or altered, except by Global Water Instrumentation, Inc., or which has been subject to misuse, negligence or accident. It is expressly agreed that this warranty will be in lieu of all warranties of fitness and in lieu of the warranty of merchantability.
b. The warranty begins on the date of your invoice.
Global Water
800-876-1172 globalw.com
- 7 -
Page 8
Global Water
800-876-1172 globalw.com
IX. Appendix A: Accessories
Part Description Part Number Unit
12V 5Ah Battery 00-010 Each Battery Cable CJ0100 Each Battery Charger FE0400 Each Pickup Hose 00-546 Feet Pickup Strainer CA0300 Each Pump tubing 00-744 Feet
- 8 -
Loading...