Global Specialties LCR-58 User Manual

SAFETY INFORMATION
EMC: Conforms to EN61326-1.
The following safety information must be observed to insure maximum personal safety during the operation of this meter:
Use the meter only as specified in this manual or the protection provided by the meter might be impaired.
Never ground yourself when taking electrical measurements. Do not touch exposed metal pipes, outlets, fixtures, etc., which might be at ground potential. Keep your body isolated from ground by using dry clothing, rubber shoes, rubber mats, or any approved insulating material.
Turn off power to the circuit under test before cutting, unsoldering, or breaking the circuit. Small amounts of current can be dangerous. When using the probes, keep your fingers behind the finger guards on the probes.
Measuring voltages which exceed the limits of the multimeter may damage the meter and expose the operator to a shock hazard.
SPECIFICATIONS*
Display: 6000 counts.
Polarity: Automatic, (-) negative polarity indication.
Overrange: " OL " mark indication.
Low battery indication:
o When a new battery, the LCD will show the battery capacity at full status ( ). If
the meter is used for a many hours, the capacity indication may show half battery status ( ). After a long time of operation, the battery may be exhausted and low battery status is shown ( ). Then “ batt ” displays accompanied by a continuous beep sound, and then the meter will shut down in 5 seconds, and no further measurement is possible.
Measurement rate: 2 times per second, nominal.
Auto power off: approx. 30 minutes.
Operating environment:
o 0 to 40 (32 104) at <70% R.H.
Storage temperature:
o -20 to 60 (-4 to 140)at < 80% R.H. with battery removed from meter.
Temperature Coefficient:
o 0.1 x(specified accuracy) per . (0 to 18, 28 to 40).
Power: 3.0V button-type lithium batteries x2, CR2032.
Battery life: 50 hours continuous operation.
Dimensions:
205mm (H) X 40mm (W) X 24.5mm (D).
Weight:
o Approx. 3.9 oz. (110g) including battery .
*(Accuracy at 23±5 , <70% R.H.)
1.1 INDUCTANCE (L)
Ranges: 600µH, 6000µH, 60mH, 600mH, 6H, 60H, 200H
Resolution: 0.1µH
Accuracy: (Q >10)
Level: Test signal amplitude 0.5V RMS
o ± (1.0% rdg + 5 dgts) on 60mH to 200H ranges o ± (2.0% rdg + 5 dgts) on 600µH, 6000µH ranges
Level: Test signal amplitude 0.1V RMS
o ± (1.5% rdg + 5 dgts) on 60mH to 60H ranges o ± (3.0% rdg + 5 dgts) on 600µH, 6000µH ranges
Test Frequency:
o (100Hz/120Hz: 60mH, 600mH, 6H, 60H, 200H ranges) o (1kHz: 6000µH, 60mH, 600mH, 6H, 60H ranges) o (10kHz: 600µH, 6000µH, 60mH, 600mH, 6H ranges)
Minimum Input Range: >1.6µH
Note: For Q >1~Q <10, accuracy is the printed specifications x 1.5
Overload protection: 10VDC or 10VAC RMS
1.2 CAPACITANCE (C)
Ranges: 600pF, 6nF, 60nF, 600nF, 6µF, 60µF, 600µF, 6mF
Resolution: 0.1PF
Accuracy: (D<0.1)
Level: Test signal amplitude 0.5V RMS
o ± (2.0% rdg + 8 dgts) on 600pF range o ± (1.0% rdg + 5 dgts) on 60nF, 600nF, 6µF ranges o ± (2.0% rdg + 5 dgts) on 6nF, 60µF, 600µF, 6mF ranges
Level: Test signal amplitude 0.1V RMS
o ± (3.0% rdg + 8 dgts) on 600pF range o ± (1.5% rdg + 5 dgts) on 60nF, 600nF, 6µF ranges o ± (3.0% rdg + 5 dgts) on 6nF, 60µF, 600µF, 6mF ranges