Global Specialties GNV-725 User manual

Page 1
TIME
REC
HOLD
C
F
T1
T2
POWER-UP OPTIONS
SETUP
CLOCK
INTV
C
F
0
TEMP: 20 C
HUMIDITY:
(
328
200
4
)
F
60 C
2498
1370
%RH
140
F
100
F
C
REC
GNV-720 GNV-725
Humidity Temperature Meters
User Manual
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 2
Global Specialties
Table Contents
Safety.............................................................................................................2
Compliance and Certications....................................................................2-3
Introduction....................................................................................................3
Specications ................................................................................................4
Instrument and Display Description...............................................................5
Operating Instructions.................................................................................6-7
Software Installation.......................................................................................7
Using the Software.........................................................................................8
Calibration Instructions..............................................................................9-10
Warranty/Service Information.......................................................................11
© 2019 Global Specialties. All rights reserved.
Unauthorized duplication of Global Specialties documentation materials is strictly prohibited. Customers are permitted to duplicate and distribute Global Specialties documentation for internal educational purposes only.
Safety
1
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 3
Global Specialties
Safety
To avoid personal injury and/or product damage, review and comply with the following safety precautions. These precautions apply to both operating and maintenance personnel and must be followed during all phases of operation, service, and repair of this meter.
Dry Conditions
Hands, shoes, oor, and work bench must be dry. Avoid making measurements with dampness, or other environmental conditions that might affect safety.
Cleaning
Use a soft cotton cloth lightly moistened with a mild solution of detergent and water. Do not allow any portion to be submerged at any time. Dry thoroughly before attempting to make measurements. Do not use solvents or expose to solvent fumes as they may cause deterioration or damage.
Do not operate:
In the presence of noxious, corrosive, flammable fumes, gases,
vapors, chemicals, or finely-divided particulates.
In environments where there is a danger of any liquid being spilled
on the meter.
In air temperatures exceeding the specified operating temperatures.
In atmospheric pressures outside the specified altitude limits or
where the surrounding gas is not air.
Compliance and Certications
CE Compliance
This product meets the essential requirements of the applicable European Directives as follows:
89/336: Electromagnetic Directive (EMC)
2011/65/EU: Restriction of Hazardous Substances Directive (RoHS)
2
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 4
Global Specialties
Disposal
(Applicable in the European Union and other European countries with separate collection systems). This product is subject to Directive 2012/19/EU of the European Parliament and the Council of the European Union on waste electrical and electronic equipment (WEEE), and in jurisdictions adopting that Directive, is marked as being put on the market after August 13, 2005, and should not be disposed of as unsorted municipal waste. Please utilize your local WEEE collection facilities in the disposition of this product.
Caution! Refer to the operating instructions.
Introduction
The GNV-720 and GNV-725 are digital humidty, temperature meters that measure relative humidity and ambient temperature as well as employing a “K” type thermocouple for more precise contact temperature measurements. The GNV-725 expands the capabilities with data logging functionality.
Kit Includes
This meter comes with the following:
Instruction manual
PC software (only with GNV-725)
RS-232 cable (SE-300, only with GNV-725)
9 V battery
Carrying case
Probe holder
K type bead thermocouple (CT3240*)
*The CT3240 is rated for - 50˚C to 200 ˚C. For higher temperature measurements please order the CT3241.
3
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 5
Specications
Humidity 0% to 100%
Temperature @ T1 -20 ˚C to +60 ˚C (-4 ˚F to +140 ˚F)
Temperature @ T2 -200 ˚C to +1370 ˚C ( -328 ˚F to +2498 ˚F)
Humidity 0.1 % RH
Temperature @ T1 0.1 ˚C (0.1 ˚F)
Temperature @ T2
(CT3240*)
Global Specialties
Specications
Measurement Range
Resolution
-200 ˚C to +200˚C, 0.1 ˚C 200 ˚C to 1370 ˚C, 1 ˚C
(-200 ˚F to +200 ˚F, 1 ˚F)
(200 ˚F to 2498 ˚F, 1 ˚F)
Accuracy
Humidity ± 2.5 % RH at 25 ˚C
Temperature @ T1 ± 0.7 ˚C (± 1.4 ˚F)*
Temperature @ T2 ± 0.5 % reading + 1 ˚C (± 0.5 % reading + 2 ˚F)
Response Time
Humidity 75 s
Temperature 40 s
Signal Output RS-232 data output
Display 4 digit LCD
Weight 320 g
Dimensions
Meter 186 x 64 x 30 mm
Probe 190 x 15(D) mm
Environmental
Battery Life Approx. 100 hrs w/ alkaline battery (9V)
Operating Environment 0 ˚C to 50 ˚C (32 ˚F to 122 ˚F)
Total Memory (GNV-725) 16,300 data points
4
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 6
Global Specialties
Instrument and Display Description
1
2
T1
GNV-720
TEMPERATURE METER
5
6
7
REL
C
TEMP: 20
4 F
)(
200
HUMIDITY:
POWER-UP OPTIONS
C
0
3
( )
K-TYPE
T2
HUMIDITY
4
8
HOLD
9
F
60
C
F140
1370
C
C F
2498
F328
100
%RH
11
LAC
232-SR
V9CD
9V BATTERY
NEDA 1604 6F22 00
PLEASE READ
MANUAL FOR SAFE
OPEN
1. Dust mask
2. Sensor probe
3. T2 channel, “K” type thermocouple
4. LCD display
5. Power button
6. ∆REL (GNV-720) TIME (GNV-725)
7. ˚C control (GNV-720) REC (GNV-725)
8. MAX/MIN button
9. HOLD button
10. ˚F control (GNV-720)
˚C/˚F (GNV-725)
˚C ˚F : Celsius and Fahrenheit indicators
%RH : Relative humidity indicator
MAX : The maximum value is
displayed
MIN : The minimum value is displayed
: Auto power off indicator
: Data hold function indicator
m-d : Month and day
h:m : Hour and minute
m:s : Minute and second
Y : Year
11. “K” type offset calibration screw
12. Digital output connector (RS-232)
13. AC power adapter connector
: Low battery indicator
K : Thermocouple type indicator
∆REL : Relative mode indicator
14. Tripod mounting hole
15. Battery cover
5
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 7
Global Specialties
Operating Instructions
Press the power button to turn the Humidty Temperature Meter ON or
OFF. Upon power up, the available memory will be displayed (GNV-
725).
To measure the ambient temperature and humidity place the meter
within the desired environment and press the power button.
To measure with the “K” type thermocouple, first insert the plug portion
into the channel labeled “T2” at the top of the meter. Then position the bead on the object under test and press the power button.
To switch the temperature scale, press either the ˚C or ˚F button.
To hold the current reading, press the HOLD button. To release this
value, press the HOLD button again.
To view the maximum or minimum values, press the MAX MIN button.
The maximum value is displayed while MAX is showing and the minimum value is displayed while MIN is showing. The maximum and minimum values will both be displayed while MAX and MIN are both blinking. To exit this mode, simply hold the MAX MIN button for two seconds.
To view the change in values press the ∆REL button. This uses the
current value as reference then takes the difference with all subsequent values and then displays it.
The auto power off function is the default setting and turns the meter
off after 30 minutes in the absence of button presses or RS-232 communication. This can be disabled by pressing and holding rst the HOLD button and then the power button together until you hear two beeps. The will no longer be displayed indicating the meter will not
turn off automatically anymore.
Replace the battery once the symbol is displayed.
The digital output is a 9600 bps N 81 serial interface.
Data Logging
To initiate data logging press the REC button. Pressing the REC button
again will stop recording. To clear the memory, rst power the meter off. Then, while holding the REC button, press the power button. Release both buttons and “CLR” will be displayed to show that the memory has
be cleared.
6
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 8
Global Specialties
To set up the clock, press and hold the MAX MIN button, then power
on the meter. Press the TIME button. Then press the REC or ˚C/˚F buttons to increase or decrease the value. Press TIME again to adjust the next item in the following order: year, month, day, hour, minute. Press TIME one more time to save settings. Pressing the power button at any time will cancel the process.
To set the recording interval press and hold the MAX MIN button then
power on the meter. Press the HOLD button. Then press the REC or ˚C/˚F buttons to increase or decrease the value. Press HOLD again
to adjust the next item. Press HOLD one more time to save settings. Pressing the Power button at any time will cancel the process.
To display the time press the TIME button. The time will be displayed
with the year at the top, the month and day at the bottom left, and the hour and minute on the bottom right of the LCD. Press TIME again to exit this mode. This will not interfere with recording, maximum, or minimum holding values.
Warning! Do not touch or manipulate sensor.
Do not expose the sensor to direct light, this can cause false readings.
Do not expose the sensor to static electricity.
Software Installation
System requirements:
Software download
RS-232 cable (SE-300)
Windows 10
16 MB RAM and 5 MB hard disk space
Installation:
Close all other applications.
Open TestLink file (SETUP.exe). The installer should start
automatically.
The default save location should be c:\program files\TestLink.
7
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 9
Global Specialties
Using the Software
Run the TestLink software by clicking the TestLink file.
Sample Rate
Meter Control Panel
Real Time Graph
Real Time Table
To take measurements, click on the run icon from the tool bar. The
graph and table should automatically start populating with data. The interval can be changed by typing in the sample rate box.
(GNV-725) Data logging can be selected by clicking the appropriate
menu at the top of the tool bar. Data will automatically be downloaded. Once the data is loaded succesfully a new window will pop up.
Data Sets
Real Time Graph
Real Time Table
The data sets window displays how many sets were loaded including
the details for each (start date, start time, recording rate, and data length). Clicking on any of the sets will populate the graph and table with the data.
8
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 10
Global Specialties
Calibration
Humidity
1. Turn the unit off. Pres and hold MAX MIN, HOLD and ˚C buttons. While these
buttons are pressed, turn the power on. Release the buttons and all the segments on the LCD will blink.
2. Within three seconds, press the ˚F and ∆REL buttons at the same time to enter
the calibration mode.
3. In calibration mode, the humidity reading will blink and “CAL1” will appear on the second display.
4. Insert the humidity probe into the standard humidity cavity of 32.8% RH at 25 ˚C. Let the system stabalize for 20 minutes then press the MAX MIN button to create the calibration data. If the value is out of tolerance, the meter will beep twice and remain in “CAL1” mode. If the value is within tolerance the meter will go directly into “CAL2” mode and display “CAL2” on the second display.
5. Insert the humidity probe into the standard humidity cavity of 75.3% RD at 25 ˚C and let the system again stabalize for 20 minutes. Press the MAX MIN button to create the calibration data. The meter will beep twice if the value is out of tolerance. The meter will exit calibration mode if the value is within tolerance.
T1 Temperature
1. Turn the unit off. Pres and hold MAX MIN, HOLD and ˚C buttons. While these buttons are pressed, turn the power on. Release the buttons and all the segments on the LCD will blink.
2. Within three seconds, press the ˚F and HOLD buttons at the same time to enter the calibration mode.
3. In calibration mode, the temperature reading will blink and “CAL1” will appear on the second display.
4. Insert the probe into the standard chamber of 0 ˚C (32 ˚F). Let the system stabalize for 20 minutes then press the MAX MIN button to create the calibration data. If the value is out of tolerance, the meter will beep twice and remain in “CAL1” mode. If the value is within tolerance the meter will go directly into “CAL2” mode and display “CAL2” on the second display.
5. Insert the humidity probe into the standard chamber of 40 ˚C (104˚ F) and let the system again stabalize for 20 minutes. Press the MAX MIN button to create the calibration data. The meter will beep twice if the value is out of tolerance. The meter will exit calibration mode if the value is within tolerance.
9
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 11
Global Specialties
Calibration Notes
To exit calibration mode at any point during the calibration procedure, simply press
the power button. This will cancel the calibration and will not save any data.
To reset factory conditions, press the and hold the power button and HOLD
buttons together.
It is recommended to first insert the probe into the standards for 20 minutes before
initiating the calibration procedure to ensure the best results.
During calibration, the temperature scale is fixed on the ˚F scale.
During calibration, the auto power off function is disabled and will be enabled
again after exiting calibration mode.
10
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Page 12
Limited One-Year Warranty
Global Specialties (a brand of Cal Test Electronics) warrants these products to be free from defective material or workmanship for a period of 1 year from the date of original purchase. Under this warranty, Global Specialties is limited to repairing the defective device when returned to the factory, shipping charges prepaid, within the warranty
period.
Units returned to Global Specialties that have been subject to abuse, misuse, damage, or accident, or have been connected, installed, or adjusted contrary to the instructions furnished by Global Specialties, or that have been repaired by unauthorized persons, will not be covered by this warranty.
Global Specialties reserves the right to discontinue models, change speci price, or design of this device at any time without notice and without incurring any obligation whatsoever.
The purchaser agrees to assume all liabilities for any damages and/or bodily injury which may result from the use or misuse of this device by the purchaser, his employees, or agents.
This warranty is in lieu of all other representations or warranties expressed or implied and no agent or representative of Global Specialties is authorized to assume any other obligation in connection with the sale and purchase of this device.
cations,
Service
If you have a need for calibration or repair services, technical, or sales support, please contact us:
451001-000 Rev A
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com