Global Specialties GNV-102 User Manual

Page 1
GNV-102
Sound Level Calibrator
Class 2
User Manual
1 globalspecialties.com
Page 2
Global Specialties
Table of Contents
Compliance and Certications...............................................3
Features................................................................................4
Specications........................................................................4
Instrument Description..........................................................5
Operating Instructions...........................................................6
Warranty................................................................................7
Service..................................................................................8
© 2018 Global Specialties (a brand of Cal Test Electronics). All rights reserved.
Unauthorized duplication of Global Specialties documentation materials is
strictly prohibited. Customers are permitted to duplicate and distribute Global
documentation for internal educational purposes only.
1 globalspecialties.com
Page 3
Global Specialties
Safety
To avoid personal injury and/or product damage, review and comply with the following safety precautions. These precautions apply to both operating and maintenance personnel and must be followed during all phases of operation, service, and repair of this probe.
Dry Conditions
Hands, shoes, oor, and work bench must be dry. Avoid making measurements under humidity, dampness, or other environmental conditions that might affect safety.
Cleaning
Use a soft cotton cloth lightly moistened with a mild solution of detergent and water. Do not allow any portion to be submerged at any time. Dry thoroughly before attempting to make voltage measurements. Do not use solvents or expose to solvent fumes as they may cause deterioration or damage
Do not operate:
In the presence of noxious, corrosive,
flammable fumes, gases, vapors, chemicals, or finely-divided particulates.
In environments where there is a danger of any
liquid being spilled on the probe.
In air temperatures exceeding the specified
operating temperatures.
In atmospheric pressures outside the specified
altitude limits or where the surrounding gas is
not air.
Page 4
Global Specialties
Compliance and Certications
CE Compliance
This product meets the essential requirements of the applicable European Directives as follows:
2014/30/EU: Electromagnetic Directive (EMC)
2011/65/EU: Restriction of Hazardous Substances Directive (RoHS)
Safety
This product meets the follow standard(s) of safety for electrical equipment for measurement, control and laboratory use:
IEC/EN 60942-1
Disposal
your local WEEE collection facilities in
(Applicable in the European Union and other European countries with separate collection systems). This product is subject to Directive 2012/19/EU of the European Parliament and the Council of the European Union on waste electrical and electronic equipment (WEEE), and in jurisdictions adopting that Directive, is marked as being put on the market after August 13, 2005, and should not be disposed of as unsorted municipal waste. Please utilize
the disposition of this product
3 globalspecialties.com
Page 5
Global Specialties
Features
Capable of calibrating 1/2 inch and 1 inch microphones
Conforms to IEC 60942 Class 2
Calibrates at 94 dB and 114 dB
Specications
Measuring Level Ranges 94 dB at 20 uPa
Output Frequency 1000 Hz ± 2 %
Accuracy ± 0.5 dB Total Harmonic Distortion <3%
Temperature 23 °C (74 °F) Relative Humidity 50 % Atmospheric Pressure 1013 hPa
Temperature Coefficient 0.005 dB / °C Hummidity Coefficient 0.005 dB / % Relative humidity
Operating Temperature 0 to 40 °C (32 to 104 °F) Operating Humidity 10 to 90% Relative humidity Storage Temperature -10 to 50 °C (14 to 122 °F) Storage Humidity 0 to 70% Relative humidity Pressure 650 - 1080 hPa
Power One 9V battery: 006P, IEC 6F22, NEDA 1604 Battery Life Approx. 40 hours Dimensions 113 x 63 x 44 mm (4.4 x 2.4 x 1.7”) Weight Approx. 170g (8.6 oz)
Operating Limits and Functions
114 dB at 20 uPa
Reference Conditions
Influence of Ambient Conditions
Environmental Specifications
Mechanical and Electrical Specifications
Page 6
Global Specialties
Instrument Description
1. Power and low battery LED indicator
2. Power and output level switch
3. 1 inch calibration slot
4. 1/2 inch calibration slot adapter
5. Battery cover and compartment
5 globalspecialties.com
Page 7
Global Specialties
Operating Instructions
Power
1. Remove the battery cover and install a 9V battery in the battery compartment
2. Slide the switch from off to 94 dB or 114 dB
If the LED shines green, the device is ready to be used. If the LED shines red, the battery needs to be
replaced.
Calibration Procedure
The instrument may be calibrated on the Fast or Slow setting, as well as either A or C weighting.
1. The calibration slot will accomodate either a 1 or 1/2 inch sound level meter
2. Place the tip of the sound level meter into the calibration slot as depicted below and hold the devices still
3. Set the range on the sound level meter to the appropriate value for calibration at 94 dB or 114 dB
4. Adjust the sound level meter until it matches the selected calibration volume
Page 8
Global Specialties
Limited One-Year Warranty
Global Specialties warrants these products to be free from defective material or workmanship for a period of 1 year from the date of original purchase. Under this warranty, Global Specialties is limited to repairing the defective device when returned to the factory, shipping charges prepaid, within the warranty period.
Units returned to Global Specialties that have been subject to abuse, misuse, damage, or accident, or have been connected, installed, or adjusted contrary to the instructions furnished by Global Specialties, or that have been repaired by unauthorized persons, will not be covered by this warranty.
Global Specialties reserves the right to discontinue models, change specications, price, or design of this device at any time without notice and without incurring any obligation whatsoever.
The purchaser agrees to assume all liabilities for any damages and/or bodily injury which may result from the use or misuse of this device by the purchaser, his employees, or agents.
This warranty is in lieu of all other representations or warranties expressed or implied and no agent or representative of Global Specialties is authorized to assume any other obligation in connection with the sale and purchase of this device.
7 globalspecialties.com
Page 9
Global Specialties
Service
If you have a need for calibration or repair services, technical, or sales support, please contact us:
22820 Savi Ranch Parkway Yorba Linda, CA 92887
Global Specialties
800-572-1028 or 714-221-9330
globalspecialties.com
Page 10
Global Specialties
Notes:
9 globalspecialties.com
Page 11
Notes:
Global Specialties
Page 12
Global Specialties
Loading...