DSC-5300 Series Digital Oscilloscope is mini-type and portable bench type instruments,
which could be used for measuring as the GND voltage.
This Chapter shows you how to operate following tasks:
It is important for you to understand the DSO’s front panel before operating it. The
following contents are the brief introduction for the front panel function, which is useful
to be familiar with the operation of the DSC-5300 Digital Storage Oscilloscope in short
time.
The oscilloscopes provides an easy-to-use front panel to convenience users to operate
them,the panel contains knobs and buttons. Thereisa list of five ashen buttons as menu
operational buttons on the right of display screen. You can set different options of the
current menu in virtue of them. Other buttons are function buttons; you can enter
different function menus or obtain given function application in virtue of them.
Global Specialties is the registered trademark of our company.
2. Trigger status
Armed. The oscilloscope is acquiring pre-trigger data. All triggers are ignored in this
state.
Ready. All pre-trigger data has been acquired and the oscilloscope is ready to accept
a trigger.
Trig’d. The oscilloscope has seen a trigger and is acquiring the posttrigger data.
Stop. The oscilloscope has stopped acquiring waveform data.
Auto. The oscilloscope is in auto mode and is acquiring waveforms in the absence of
2. Set the switch to 1X on the probe and connect the probe to channel 1 on the
oscilloscope. To do this, align the slot in the probe connector with the key on the CH 1
BNC, push to connect, and twist to the right to lock the probe in place. Connect the probe
tip and reference lead to the PROBE COMP connectors
Picture 1.2-2
3Press “AUTO” to show the 1 KHz frequency and about 3V peak-peak square wave
4. Press “CH1” two times to cancel the channel 1, Press“CH2” to change screen into
channel 2, reset the channel 2 as step 2 and step 3.
1.3 Probe
1.3.1 Probe Safety
A guard around the probe body provides a finger barrier for protection from electric
shock.
Picture 1.3-1
Connect the probe to the oscilloscope and connect the ground terminal to ground before
you take any measurements.
Note ● To avoid electric shock when using the probe, keep fingers behind the guard
on the probe body.
● To avoid electric shock whil e using the probe, do not touch metallic po rtions of
the probe head while it is connected to a voltage source. Connect the probe to
the oscilloscope and connect the ground terminal to ground before you take any
measurements.
1.3.2 Probe Attenuation Setting
Probes are available with various attenuation factors which affect the vertical scale of
the signal. The Probe Check function verifies that the Probe attenuation option matches
the attenuation of the probe.
You can push a vertical menu button (such as the CH 1 MENU button), and select the
Probe option that matches the attenuation factor of your probe.
Note: The default setting fo r the Probe option is 1X.
Be sure that the attenuation switch on the probe matches the Probe option in the
oscilloscope. Switch settings are 1X and 10X.
Note: When the attenuation switch is set to 1X, the probe limits the bandwidth of the
oscilloscope to 6MHz (according to Probe spec). To use the full bandwidth of the
oscilloscope, be sure to set the switch to 10X
1.3.3 Probe Compensation
As an alternative method to Probe Check, you can manually perform this adjustment to
match your probe to the input channel.
Picture1.3-2
1. Set the Probe option attenuation in the channel menu to 10X. Set the switch to 10X
on the probe and connect the probe to channel 1 on the oscilloscope. If you use the
probe hook-tip, ensure a proper connection by firmly inserting the tip onto the probe.
2. Attach the probe tip to the PROBE COMP 3V connector and the reference lead to the
PROBE COMP Ground connector. Display the channel and then push the “AUTO”
button.
■ HORI MENU: Press to display the Horizontal menu. You can use the Horizontal menu
to display the waveform and zoom in a segment of a waveform.
■ TRIG MENU: Press to display the Trigger menu. You can use the Trigger menu to set
the trigger type (Edge. Pulse, Video, Slope, Alternative) and trigger settings.
■ SET TO 50%: Press to stabilize a waveform quickly. The oscilloscope can
set the trigger level to be halfway between the minimum and maximum
voltage level automatically. This is useful when you connect a signal to the
EXT TRIG connector and set the trigger source to Ext or Ext/5.
■ FORCE: Use the FORCE button to complete the current waveform acquisition
whether the oscilloscope detects a trigger or not. This is useful for Single acquisitions
and Normal trigger mode.
■ SAVE/RECALL: Press to display the Save/Recall menu. You can use the Save/Recall
menu to save and recall up to 20 oscilloscope setups or waveforms in internal memory
(up to 20 waveforms) or on a USB memory device (limited by memory capacity of USB
device). You can also use it to recall the default factory settings, to save waveform data
as a comma-delimited file (.CSV), and to save or print the displayed waveform image.
■ ACQUIRE: Press to display Acquire menu. You can use the Acquire menu to set the
■ MEASURE: Press to display a menu of measurement parameters.
■ CURSORS: Display the Cursor Menu. Vertical Position controls adjust cursor position
while displaying the Cursor Menu and the cursors are activated. Cursors remain
displayed (unless the “Type” option is set to “Off”) after leaving the Cursor Menu but
are not adjustable.
■ DISPLAY: Press to open the Display menu. You can use the Display menu to set grid
and waveform display styles, and persistence.
■ UTILITY: Press to open the Utility menu. You can use the Utility menu to configure
oscilloscope features, such as sound, language, counter, etc. You can also view
system status and update software.
■ DEFAULT SETUP: Press to reset the oscilloscope’s settings to the default
■ AUTO: Automatically sets the oscilloscope controls to produce a usable display of the
input signals.
■ RUN/STOP: Continuously acquires waveforms or stops the acquisition.
NoteIf waveform acquisition is stopped (using the RUN/STOP or SINGLE
button), the SEC/DIV control expands or compresses the waveform.
■ SINGLE: Acquire a single waveform and then stops.
2.2 Connector
Picture 2-2
■ Channel Connector (CH1, CH2): Input connectors for waveforms display.
■ EXT TRIG: Input connector for an external trigger source. Use the Trigger
Menu to select the “Ext” or “Ext/5” trigger source.
■ Probe Component: Voltage probe compensation output and ground. Use
to electrically match the probe to the oscilloscope input circuit.
Note: If you connect a voltage source to a ground terminal, you may
damage the oscilloscope or the circuit under test. To avoid this, do not
connect a voltage source to any ground terminals.
2.3 Auto Setup
The DSC-5300 Digital Storage Oscilloscopes have a Auto Setup function that identifies
the waveform types and automatically adjusts controls to produce a usable display of
the input signal.
Press the AUTO button, and then press the menu option button adjacent to the desired
waveform as follows:
Picture 2-3
Table 2-1 Auto Set function Menu:
Option Description
Multi-cycle sine
Single-cycle sine
Auto set the screen and display several cyc
signal.
Set the screen and auto display single cyc
signal.
Rising edge Auto set and show the rising time.
Falling edge Auto set and show the falling time.
Undo Setup
Causes the oscilloscope to recall the
previous setup.
Auto set determines the trigger source based on the following conditions:
● If multiple channels have signals, channel with the lowest frequency signal.
● No signals found, the lowest-numbered channel displayed when Auto set was invoked
● No signals found and no channels displayed, oscilloscope displays and uses channel
Acquire Mode Adjusted to Sampling
Display Format Y-T
Display Type Set to Dots for a video signal, set to Vectors for
an FFT spectrum; otherwise, unchanged
Vertical Coupling Adjusted to DC or AC according to the input
signal
Bandwidth Limit Off(full)
V/div Adjusted
VOLTS/DIV adjustability Coarse
Signal inverted Off
Horizontal position Center
S/div Adjusted
Trigger type Edge
Trigger source Auto detect the channel which has the input
signal
Trigger slope Rising
Trigger mode Auto
Trigger coupling DC
Trigger holdoff Minimum
Trigger level Set to 50%
2.4 Default Setup
The oscilloscope is set up for normal operation when it is shipped from the factory. This
is the default setup. To recall this setup, press the DEFAULT SETUP button. The options,
buttons and controls that change settings when you press the DEFAULT SETUP button,
refer to appendix B.
The DEFAULT SETUP button does not reset the following settings:
You can use the Universal knob with many functions, such as adjusting the holdoff
time, moving cursors, setting the pulse width, Setting the Video Linage, adjusting
the upper and lower frequency limit, adjust X and Y masks when using the pass/fail
function etc. You can also turn the “Universal” knob to adjust the storage position of
setups, waveforms, pictures when saving/recalling and to select menu options.
2.6 Vertical System
The vertical control could be used for displaying waveform, rectify scale and position.
Next Page Page 1/3 Enter the second page of the menu.
AC
GND
On
Off
Coarse
Fine
1X,5X
10X,50X
100X,
500X,1000X
DC passes both AC and DC components of
the input signal.
AC blocks the DC component of the input
signal and attenuates signals below 10 Hz.
GND disconnects the input signal.
Limits the bandwidth to reduce display noise;
filters the signal to reduce noise and other
unwanted high frequency components.
Selects the resolution of the Volts/Div knob
Coarse defines a 1-2-5 sequence.
Fine changes the resolution to small steps
between the coarse settings.
Set to match the type of probe you are using
to ensure correct vertical readouts.
Table 2-4 CH1, CH2 function menu 2:
Option Setting Instruction
Invert
Digital Filter
on
off
Press this button to enter the “Digital Filter
Turn on invert function.
Turn off invert function.
menu”.(See table 2-5)
Next Page Page 2/3 Enter the second page of the menu.
Table 2-5 Digital Filter function menu:
Option Setting Introduction
Digital Filter
Type
On
Off
Turn on the digital filter.
Turn off the digital filter.
Setup as LPF (Low Pass Filter).
Setup as HPF (High Pass Filter).
Setup as BPF (Band Pass Filter).
Setup as BRF (Band Reject Filter).
Upper_limit Turn the “Universal” knob to set upper limit.
Lower_limit Turn the “Universal” knob to set lower limit.
In order to assort the attenuation coefficient, you need to response in the channel
operation Menu. If the attenuation coefficient is 10:1, the input coefficient should be set
to 10X, so that the mistake of the Volts/div information and measure testing should be
forbidden.
Take the CH1 for example, when you use the 100:1 probe:
Press “CH1”→“Next Page page1/3”→ “Filter”, display the digital filter menu. Select
“Filter Type”, then select “Upper Limit” or “Lower Limit” and turn the “Universal” knob to
adjust them.
● Press “CH1”→“Next Page page1/3”→ “Filter” →“Off”. Turn off the Digital Filter
function.
Picture 2.6-7
● Press “CH1”→ “Next Page page1/3”→ “Filter” → “On”. Turn on the Digital Filter
2.6.2 Using Vertical “Position” Knob and “Volt/div” Knob
■Vertical “POSITION” Knob
1. Use the Vertical “POSITION” knobs to move the channel waveforms up or down on
the screen. This button’s resolution is variety as per the vertical scale.
2. When you adjust the vertical position of channels waveforms, the vertical position
information will display on the left bottom of screen. For example “Volts
Pos=24.6mV”.
3. Press the vertical “POSITION” knob to set the vertical position to zero.
■“Volts/div” Knob
1. Use the “Volts/div” knobs to control how the oscilloscope amplifies or attenuates the
source signal of channel waveforms. When you turn the “volts/div” knob, the
oscilloscope increases or decreases the vertical size of the waveform on the screen with
respect to the ground level;
2. When you press the “Volt/div” Knob, you can switch “Volt/div” option between “Coarse”
and “Fine”. The vertical scale is made sure by the 1-2-5 step in the Coarse. Increase in
the clockwise, reduce in the anticlockwise. In the fine mode, the knob changes the
Volts/Div scale in small steps between the coarse settings. Increase in the clockwise,
reduce in the anticlockwise.
2.6.3 Math Functions
Math shows the results after +,-,*, / and FFT operations of the CH1 and CH2.
Press the MATH button to display the waveform math operations. Press the MATH
button again to remove the math waveform display.
Table 2-6 MATH function menu:
Function Setting
Operation + - * / FFT
Invert
on
off
Introduction
Math operates between signal source
CH1 and CH2.
Invert the MATH waveform.
Turn off MATH Invert function.
Table 2-7 MATH function instruction
Operation Setting Introduction
+ CH1+CH1
CH1-CH2
CH2-CH1
* CH1*CH2
/
CH1/CH2
CH2/CH1
FFT Fast Fourier Transform.
Adjust the vertical position of the math
waveform by using universal knob.
Adjust the range of the math waveform
by using universal knob.
CH1 waveform adds CH2 waveform.
The channel 2 waveform is subtracted
from the channel 1 waveform.
The channel 1 waveform is subtracted
from the channel 2 waveform.
Channel 1 multiply Channel 2.
The FFT process mathematically converts a time-domain signal into its frequency
components. You can use the Math FFT mode to view the following types of signals:
● Analyze the Humorous wave in the Power cable.
● Test the Humorous content and distortion in the system
● Show the Noise in the DC Power supply
● Test the filter and pulse response in the system
Changes the horizontal magnification of
the FFT display.
10X
Next Page Page 1/2
Enter the second page of FFT menu.
Table 2-9 FFT function menu 2:
FFT Option Setting
Scale
Display
Vrms
dBVrms
Split
Full screen
Next Page Page 2/2
Introduction
Set Vrms to be the Vertical Scale unit.
Set dBVrms to be the vertical Scale unit.
Display FFT waveform on half screen.
Display FFT waveform on full screen.
Return the first page of FFT menu.
To use the Math FFT mode, you need to perform the following tasks:
1. Set up the source (time-domain) waveform.
● Press the AUTO button to display a YT waveform.
● Turn the vertical “POSITION” knob to move the YT waveform to the center vertically
(zero divisions).
● Turn the horizontal “POSITION” knob to position the part of the YT waveform that you
want to analyze in the center eight divisions of the screen.
The oscilloscope calculates the FFT spectrum using the center 1024 points of the
time-domain waveform.
● Turn the “Volts/div” knob to ensure that the entire waveform remains on the screen.
● Turn the “S/div” knob to provide the resolution you want in the FFT spectrum.
● If possible, set the oscilloscope to display many signal cycles.
To display FFT correctly, follow these steps:
1. Push the “MATH” button.
2. Set the “Operation” option to FFT.
3. Press “Source” button to select “CH1” or “CH2” according to input signal channel.
4. According to Nyquist law, turn the “S/div” knob to adjust the sampling rate (This
parameter is displayed behind the time base parameter) is at least double than input
signal frequency.
Press the MATH button to display the Math Menu. Use the options to select the Source
channel, Window algorithm, and FFT Zoom factor. You can display only one FFT
spectrum at a time. You can select “Full screen” or “Split” in “Display” option to display
FFT waveform on full screen or display channel waveform and its FFT waveform on half
screen at a time.
Picture2.6-10 FFT function
3. Select FFT window
Windows reduce spectral leakage in the FFT spectrum. The FFT assumes that the YT
waveform repeats forever. With an integral number of cycles, the YT waveform starts
and ends at the me amplitude and there are no discontinuities in the signal shape A
non-integral number of cycles in the YT waveform causes the signal start and end points
to be at different amplitudes. The transitions between the start and end points cause
discontinuities in the signal that introduce high-frequency transients.
According to the tested options and source speciality, make sure the window you
need to use.
Table 2-10 FFT window instruction
Window Speciality Satisfied Test content
Rectangulr
Best frequency
resolution, worst
magnitude resolution.
This is essentially the
Symmetric transients or bursts.
Equal-amplitude sine waves
with fixed frequencies.
Broadband random noise with
Better frequency, poorer
magnitude accuracy
than Rectangular.
Hamming has slightly
better frequency
resolution than Hanning.
Best magnitude, worst
frequency resolution.
Sine, periodic, and
narrow-band random noise.
Asymmetic transients or
bursts.
Single frequency waveforms,
to find higher order harmonics.
4.Magnifying and Positioning an FFT Spectrum
You can magnify and use cursors to take measurements on the FFT spectrum. The
oscilloscope includes an “FFT Zoom” option to magnify horizontally, press this option
button to select “1X”, “2X”, “5X” or “10X”. Moreover, you also can turn the “Universal”
knob to magnify FFT waveform horizontally in a 1-2-5 step. To magnify vertically; you
can turn the “Volts/div” knob.
5. Measuring an FFT Spectrum Using Cursors
You can take two measurements on FFT spectrums: magnitude (in dB) and frequency
(in Hz). Magnitude is referenced to 0 dB, where 0 dB equals 1 VRMS. You can use the
cursors to take measurements at any zoom factor. (Refer to cursor measure2.11.2)
Use horizontal cursors to measure amplitude and vertical cursors to measure
frequency.
If you input a sine signal to channel 1, follow these steps:
1. Measure FFT Amplitude
1) Input a sine signal to channel 1, and press the “AUTO” button.
2) Press the “Cursor Mode” button to select “Manual”.
3) Press the “Type” option button to select “Time”.
4) Press the “Source” option button to select “MATH”.
5) Press the “CurA” option button, turn the “Universal” button to move Cursor A to the
highest position of the FFT waveform.
6) The value of CurA displaying on the top of the left screen is FFT highs frequency.
This frequency should be the same as input signal frequency.
Picture 2.6-12
2.6.4 Using Ref
The reference control saves waveforms to a nonvolatile waveform memory. The
reference function becomes available after a waveform has been saved.
Table 2-11 REF function menu
Option Setting
CH1,CH2,
Source
CH1 off
CH2 off
REFA
REFB
Save
Introduction
Choose the waveform display to store.
Choose the reference location to store or recall
a waveform.
Stores source waveform to the chosen
reference location.
As follow Picture, there are one button and two knobs in the HORIZONTAL area.
S/div knob
Horizontal
position knob
Picture 2.7-1
Table 2- 12 SDS1000CML Horizontal system function menu:
Option Setting Description
Delayed
On
Off
Turn on this function that main timebase
waveform display on the top half screen and
window timebase waveform display on the
below half screen at the same time.
Turn off this function that only display main
timebase waveform on the screen.
Set memory depth to normal.
Set memory depth to long Memory depth in
order to get more waveform dots.
MemDepth
Normal
Long Mem
Note: If Set memory depth to normal mode, the memory depth is 40Kpts
If Set memory depth to Long Mem mode, the memory depth is 2Mpts
MemDepth Detailed Specification
Memory
Depth
Channel Mode
Samping
Rate
Short
Memory
Long Memory
Single Channel 1Gsa/s 40kpts No Support
Single Channel
Double
Channels
500MSa/s
or lower
500MSa/s
or lower
20kpts 2Mpts
20kpts 1Mpts
2.7.1 Horizontal Control Knob
You can use the horizontal controls to change the horizontal scale and position of
waveforms. The horizontal position readout shows the time represented by the center
of the screen, using the time of the trigger as zero. Changing the horizontal scale causes
the waveform to expand or contract around the screen center.
■ Horizontal “POSITION” Knob
1. Adjust the horizontal position of all channels and math waveforms (the position of the
trigger relative to the center of the screen). The resolution of this control varies with the
time base setting.
2. When you press the horizontal “POSITION” Knob, you can set the horizontal position
to zero.
■“S/div” Knob
1. Using to change the horizontal time scale to magnify or compress the waveform. If
waveform acquisition is stopped (using the RUN/STOP or SINGLE button), turn the
S/div knob to expand or compress the waveform.
2. Select the horizontal time/div (scale factor) for the main or the window time base.
When Window Zone is enabled, it changes the width of the window zone by changing
the window time base.
■ Display Scan Mode
When the SEC/DIV control is set to100 ms/div or slower and the trigger mode is set to
Auto, the oscilloscope enters the scan acquisition mode. In this mode, the waveform
display updates from left to right. There is no trigger or horizontal position control of
waveforms during scan mode.
2.7.2 Window Zone
Use the Delayed option to define a segment of a waveform to see more detail. The
Window time base setting cannot be set slower than the Main time base setting.
You can turn the Horizontal Position and SEC/DIV controls to enlarge or diminish
waveforms in the Window Zone.
“M” Mean main time base, “W” mean window time base.. It is also a arrow on the scale
top to show the vertical position.
If you want to see a section of the waveform in detail, follow these steps:
(1) Press the “HORI MENU” button to enter the “Horizontal menu”.
(2) Turn the “S/div” knob to change the main timebase scale.
(3) Press the “Delayed” option button to select “On”.
(4) Turn the “Horizontal Position” knob (adjust window’s position) to select the
window that your need and expanded window waveform display on the below
half screen at the same time.
2.8 Trigger System
The trigger determines when the oscilloscope starts to acquire data and display a
waveform. When a trigger is set up properly, the oscilloscope converts unstable displays
or blank screens into meaningful waveforms.
There are three buttons and one knob in the Trigger area. See picture.2-29:
LEVEL knob
Picture 2.8-1
■“TRIG MENU” Button: Press the “TRIG MENU” button to display “Trigger Menu”.
■“LEVEL” Knob: The LEVEL knob is to set the corresponding signal voltage of trigger
point in order to sample. Press the “LEVEL” knob can set trigger level to zero.
■“SET TO 50
The oscilloscope can set the Trigger Level to be about halfway between the minimum
” Button: Use the “SET TO 50%” button to stabilize a waveform quickly.
and maximum voltage levels automatically. This is useful when you connect a signal to
the EXT TRIG BNC and set the trigger source to Ext or Ext/5.
■“FORCE” Button: Use the FORCE button to complete the current waveform
acquisition whether the oscilloscope detects a trigger or not. This is useful for SINGLE
acquisitions and Normal trigger mode.
■ Pre-trigger/Delayed trigger: The data before and after trigger the trigger position is
typically set at the horizontal center of the screen, in the full-screen display the 6div data
of pre-trigger and delayed trigger can be surveyed. More data of pre-trigger and 1s
delayed trigger can be surveyed by adjusting the horizontal position.
The feature is very useful because you can see the events that led up the trigger point
everything to the right of the trigger point is called posttrigger information the amount of
delay range (pre-trigger and posttrigger information) available is dependent on the
sweep speed selected.
2.8.1 Signal Source
You can use the Trigger Source options to select the signal that the oscilloscope uses
as a trigger. The source can be any signal connected to a channel BNC, to the EXT TRIG
BNC or the AC power line (available only with Edge triggers).
2.8.2 Trigger Type
The scopes have five trigger types: Edge, Video, Pulse, Slope, and Alternative
Edge Trigger
Use Edge triggering to trigger on the edge of the oscilloscope input signal at the trigger
threshold.
Table 2-13 Edge Trigger function Menu:
Option Setting
Type Edge
Source CH1
CH2
Explain
With Edge highlighted, the rising or falling edge of the
input signal is used for the trigger.
Triggers on a channel whether or not the waveform is
displayed.
Does not display the trigger signal; the Ext option uses
the signal connected to the EXT TRIG front-panel
BNC and allows a trigger level range of -1.2V to +1.2V.
Same as Ext option, but attenuates the signal by a
factor of five, and allows a trigger level range of +6V to
-6V.This extends the trigger level range.
This selection uses a signal derived from the power
line as the trigger source; trigger coupling is set to DC
and the trigger level to 0 volts.
Trigger on Rising edge of the trigger signal.
Trigger on Falling edge of the trigger signal.
Trigger on Rising edge and Falling edge of the trigger
signal.
Use this mode to let the acquisition free-run in the
absence of a valid trigger; This mode allows an
untriggered, scanning waveform at 100 ms/div or
slower time base settings.
Use this mode when you want to see only valid
triggered waveforms; when you use this mode, the
oscilloscope does not display a waveform until after
the first trigger.
When you want the oscilloscope to acquire a single
waveform, press the “SINGLE ”button.
Enter the “Trigger Setup Menu” (See table 2-14).
Table 2-14 Trigger Setup function menu
Option Setting Explain
Coupling
Holdoff
Holdoff
DC Passes all components of the signal
AC
HF Reject
LF Reject
Blocks DC components attenuates signals
below 50 Hz.
Attenuates the high-frequency components
above 150 kHz.
Blocks the DC component , attenuates the
low-frequency components below kHz.
Using the “universal” knob to adjust holdoff
time(sec)the holdoff value is displayed.
Reset holdoff time to 100ns.
Reset
Return Return the first page of “Trigger main menu”.
1) Press the “TRIG MENU” button to display “Trigger” menu.
2) Press the “Type” option button to select “Edge”.
2. Set up Source
According to input signal, press the “Source” option button to select “CH1”,
“CH2”,“EXT”, “EXT/5”or “AC Line”.
3. Set up Slope
The image
Press the “Slope” option button to select “”, “
cannot be
” or “”.
The image
cannot be
4. Set up Trigger mode
Press the “Trigger mode” option button to select “Auto”, “Normal”, “Single”.
Auto: The waveform refresh at a high speed whether the trigger condition is
satisfied or not.
Normal: The waveform refresh when the trigger condition is satisfied and waits for
next trigger event occurring when the trigger condition is not satisfied.
Single: The oscilloscope acquire a waveform when the trigger condition is
satisfied and then stops.
5. Set up Trigger coupling
a. Press the “Set up” button to enter the “Trigger Setup Menu”.
b. Press the “Coupling” option button to select “DC”, “AC”, “HF Reject” or “LF
Select the pulse to trigger the pulse match the
trigger condition.
Select the type of triggering; Normal mode is
best for most Pulse Width trigger applications.
single
Set up Enter the “Trigger setup menu”.
Next Page Page 2/2
Press this button to return the first page.
Picture 2.8-4
Operate Instruction:
1. Set up Type
1) Press the “TRIG MENU” button to display “Trigger menu”.
2) Press the “Type” option button to select “Pulse”.
Video TriggerTrigger on fields or lines of standard video signals.
Table 2-17 Functional Manu of Video Trigger 1:
Option Setting Instruction
When you select the video type, put the
Type Video
couple set to the AC, then you could trigger
the NTSC,PAL and SECAM video signal.
Select the input source to be the trigge
signal.
Ext and Ext/5 use the signal applied to the
EXT TRIG connector as the source.
Normal triggers on the negative edge of
the sync pulse.
Source
Polarity
CH1
CH2
EXT
EXT/5
(Normal)
(Inverted) Inverted triggers on the positive edge of
the sync pulse.
Sync
Line Num
All lines
Odd field
Select appropriate video sync.
Even Field
Next Page Page 1/2
Enter the second page of “Video
trigger menu”.
Table 2-18 Functional Manu of Video Trigger 2:
Option Setting Instruction
When you select the video type, put
Type Video
the couple set to the AC, then you
could trigger the NTSC,PAL and
SECAM video signal.
Standard
NTSC
Pal/Secam
Select the video standard for sync and
line number count.
Use this mode to let the acquisition
free-run in the absence of a valid
Mode Auto
trigger; This mode allows an
untriggered, scanning waveform at
100 ms/div or slower time base
settings.
Trigger on positive slope of negative slope
according to setup time of the oscilloscope.
Select the trigger level that can be adjusted by
“LEVEL” knob. You can adjust “LEVEL A”,
“LEVEL B” or adjust them at the same time.
Use this mode to let the acquisition free-run in the
Auto
Mode
Normal
absence of a valid trigger; This mode allows an
untriggered, scanning waveform at 100 ms/div or
slower time base settings.
Use this mode when you want to see only valid
triggered waveforms; when you use this mode,
the oscilloscope does not display a waveform
until after the first trigger.
Single
When you want the oscilloscope to acquire a
single waveform, press the “SINGLE” button.
Set up Enter “Trigger setup menu” (See table 2-14).
The trigger signal comes from two vertical channels when you use alternative trigger.
In this mode, you can observe two irrelative signals at the same time. You can select
different trigger types for two vertical signals, and selected types cover edge, pulse,
video and slope trigger. Trigger information of two channel signals display on the bottom
right of the screen.
Picture 2.8-8
Table 2-21 Set trigger mode to edge trigger function menu 1:
Option Setting Instruction
When use alternative trigger, the trigger signal
Type Alternative
comes from two vertical channels. In the mode, you
can observe two irrelative signals at a time.
Chann
els
Source
CH1-CH2 Set the trigger channels
CH1
CH2
Set trigger type information for CH1 signal
Set trigger type information for CH2 signal
Mode Edge Set trigger type of vertical channel signal to edge
The image
Slope
cannot be
The image
cannot be
Triggering on rising edge.
Triggering on falling edge.
Triggering on rising edge and falling edge.
Set up Enter “Trigger setup menu” (See table 2-14).
Normal triggers on the negative edge of the
sync pulse.
Inverted triggers on the positive edge of the
sync pulse.
Enter the second page of Alternative trigger
menu.
Table 2-25 Set trigger mode to video trigger function menu 2:
Option Setting Instruction
Sync
StandardNTSC
Next
Page
Line Num
All lines
Odd field
Even Field
Pal/Secam
Page 1/2
Select appropriate video sync.
Select the video standard for sync and line
number count.
Enter the second page of Alternative trigger
menu.
Table 2-26 Set trigger mode to slope trigger function menu 1
Option Setting Instruction
Type Alternative The trigger signal comes from two vertical
channels when you use alternative trigger. In
this mode, you can observe two irrelative
signals at the same time.
Source
CH1
CH2
Mode Slope
The image
cannot be
The image
cannot be
The image
cannot be
When
The image
cannot be
The image
cannot be
The image
cannot be
Next Page Page 1/2
Set trigger type informa
tion for CH1 signal
Set trigger type information for CH2signal
Set trigger type of the vertical channel signal to
slope trigger.
Select slope trigger condition.
Enter the second page of the alternative
trigger.
Table 2-27 Set trigger mode to slope trigger function menu 2
Option Setting Instruction
Time
Set time
Turn the “Universal” knob to set the slope
time. Time setup range is 20ns-10s.
Vertical Select the trigger level that can be adjusted by
“LEVEL” knob. You can adjust “LEVEL A”,
“LEVEL B” or adjust them at the same time.
Set up Enter “Trigger setup menu”(see table 2-14).
Next Page Page 2/2 Return to the first page of “Alternative trigger
menu”.
Operate Instruction:
Observe two irrelative channel signals, follow these steps
1. Input two irrelative signals to channel 1and channel 2.
2. Press the AUTO button.
3. Press the TRIG MENU button to enter “trigger menu”.
4. Press the “Type” option button to select “Alternative”.
5. Press the “Channels” option button to select “CH1-CH2”
6. Press the “Source” option button to select “CH1”.
7. Press the CH1 button and turn the “S/div” knob to optimize waveform display.
8. Press “Mode” option button to select “Edge”, “Pulse”, “Slope” or “Video”.
9.Set the trigger according to trigger edge.
10. Press the “Source” option button to select “CH2”.
11. Press the CH2 button and turn the “S/div” knob to optimize waveform display.
12. Repeat steps 8 and 9.
2.8.3 Coupling
Use the “Coupling” to make sure the signal that passes the trigger circuit. It is useful for
us to gather a steady wave form.
If you use the trigger coupling, you should press the “TRIGGER” button and then select
“edge”, “Pulse”, “Video” or “Slope” trigger. Then select the “coupling” option in “Set up
menu”.
The horizontal position control establishes the time between the trigger position and the
screen center. You can adjust the horizontal “POSITION” knob control to view waveform
data before the trigger, after the trigger, or some of each. When you change the
horizontal position of a waveform, you are changing the time between the trigger and the
center of the display actually. (This appears to move the waveform to the right or left on
the display.)
2.8.5 Slope & Level
The Slope and Level controls help to define the trigger. The Slope option (Edge trigger
type only) determines whether the oscilloscope finds the trigger point on the rising or the
falling edge of a signal.
The TRIGGER LEVEL knob controls where on the edge the trigger point occurs.
T
r
i
g
g
e
r
l
e
v
e
l
c
a
n
b
e
a
d
j
u
s
t
e
d
v
e
r
t
i
c
a
l
l
y
Note: ● Press the SINGLE button when you want the oscilloscope to
acquire a single wave form .
● Trigger coupling affects on ly the signal passed to the trigger
system. It d oes no t a ffec t t he bandwidt h o r cou plin g o f th e si gna l
displayed on the screen.
● Normal Polarity Sync triggers always occur on negativ e-going
horizontal sync pulses. If the video waveform has positive-going
horizontal sync pulses, use the Inverted Polarity selecti on.
www.GlobalTestSupply.com
Page 51
2.8.6 Trigger Holdoff
You can use the Trigger Holdoff function to produce a stable display of complex
waveforms. Holdoff is time between when the oscilloscope detects one trigger and when
it is ready to detect another. The oscilloscope will not trigger during the holdoff time. For
a pulse train, you can adjust the holdoff time so the oscilloscope triggers only on the first
pulse in the train.
Trigger position
Trigger level
Holdoff time
Picture 2.8-10
If you want to change holdoff time, please follow next operations:
1. Press the “TRIG MENU” button to show the “TRIG Menu”.
2. Press the “Type” option button to select trigger type.
3. Press the “Set up” option button to enter the “Trigger setup menu”.
4. Press the “Holdoff” option buttonturn the “Universal” knob to change the holdoff
time until the waveform trigger steadily.
Note
Use trigger holdoff to help stabilize the display of aperiodic waveforms.
2.9 Acquiring Signals System
Showing as the follow picturethe “ACQUIRE” button for Acquiring Signals system is at
the menu.
Table 2-28 The Function manual of Acquiring Signals:
Option Setting Introduction
Sampling
Acquisition
Peak
Detect
Average
Averages
4, 16,32
,64,128,
256
Sinx/x
Mode
On
off
Equ time
Real time
Sa Rate
Use for sampling and accurately display most
of the waveform.
Detect the noise and decrease the possibility
of aliasing.
Use to reduce random or uncorrelated noise in
the signal display.
Select number of averages.
Use sin interpolation
Use linear interpolation
Set the Sampling mode to Equ time.
Set the Sampling mode to Real time.
Display system sampling rate.
When you acquire a signal, the oscilloscope converts it into a digital form and displays
a waveform. The acquisition mode defines how the signal is digitized and the time base
setting affects the time span and level of detail in the acquisition.
■ Sampling: In this acquisition mode, the oscilloscope samples the signal in evenly
spaced intervals to construct the waveform. This mode accurately represents signals
most of the time.
Advantage: You can use this mode to reduce random noise.
Disadvantage: This mode does not acquire rapid variations in the signal that may occur
between samples. This can result in aliasing may cause narrow pulses to be missed. In
these cases, you should use the Peak Detect mode to acquire data.
■ Peak Detect: Peak Detect mode capture the maximum and minimum values of a
signal Finds highest and lowest record points over many acquisitions.
Advantage: In this way, the oscilloscope can acquire and display narrow pulses, which
may have otherwise been missed in Sample mode.
Disadvantage: Noise will appear to be higher in this mode.
Picture 2.9-3 Peak Detect mode
Average: The oscilloscope acquires several waveforms, averages them, and
Advantage: You can use this mode to reduce random noise.
Picture 2.9-4 Average Mode
■ Equivalent Time Sampling:
The equivalent time sampling mode can achieve up to 20 ps of horizontal resolution
(equivalent to 50GSa/s). This mode is good for observing repetitive waveforms.
■ Real Time Sampling: The scope has the highest Real-time sampling rate
up to 1GSa/s.
■ “RUN/STOP” Button: Press the RUN/STOP button when you want the
oscilloscope to acquire waveforms continuously. Press the button again to stop the
acquisition.
■ “SINGLE” Button: Press the SINGLE button to acquire a single waveform. Each time
you push the SINGLE button, the oscilloscope begins to acquire another waveform.
After the oscilloscope detects a trigger it completes the acquisition and stops.
When you push the RUN/STOP or SINGLE buttons to start an acquisition, the
oscilloscope goes through the following steps:
1. Acquire enough data to fill the portion of the waveform record to the left of the trigger
point. This is also called the pre-trigger.
2. Continue to acquire data while waiting for the trigger condition to occur.
3. Detect the trigger condition.
4. Continue to acquire data until the waveform record is full.
5. Display the newly-acquired waveform.
■ Time Base: The oscilloscope digitizes waveforms by acquiring the value of an input
signal at discrete points. The time base allows you to control how often the values are
digitized.
Use S/div knob to adjust the time base to a horizontal scale suits your purpose,
■ Time Domain Aliasing
Aliasing occurs when the oscilloscope does not sample the signal fast enough to
construct an accurate waveform record. When this happens, the oscilloscope displays
a waveform with a frequency lower than the actual input waveform, or triggers and
displays an unstable waveform.
You can press the “Acquisition” option button or turn the “Universal” knob to select
“Sampling” mode, “Peak Detect” mode or “Average” mode..
Set up Averages
When you select “Average” format, you can press the “Averages” option button to select
“4”, “16”, “32”, “64”, “128”or “256”.
Set up function interpolation
You can also turn Sins/x interpolation on or off. Press the corresponding option button
to turn Sins/s on or off.
Press the “Mode” option button to select “Real Time” or “Equ Time”.
Set up Sampling Rate
Adjust the sampling rate by pressing the “Sa Rate” option button and turning the
Time/div front panel knob. The sampling rate is shown at the corresponding timebase
scale.
2.10 Display System
The display function could be expressed by the “DISPLAY” Button.
Use the XY format to analyze phase differences, such as those represented by
Lissajous patterns. The format plots the voltage on channel 1 against the voltage on
channel 2, where channel 1 is the horizontal axis and channel 2 is the vertical axis. The
oscilloscope uses the untriggered Sample acquisition mode and displays data as dots.
XY mode has a breakthrough that traditional oscilloscopes restrict sampling rate at
1MSa/s
support 25KSa/s250MSa/s (1-2.5-5 sequence) adjusted ;
Operation steps
Channel 1 “Volt/div” and vertical “POSITION” set up the horizontal scale and
position.
Channel 2 “Volt/div” and vertical “POSITION” set up the horizontal scale and
position.
Turn the “S/div” knob to adjust the sampling rate
The following functions are forbidden at XY display form:
Benchmark wave form and wave mathematic
Cursor
Autoresets display format to YT
Trigger Control
Horizontal Position Knob
Vector Display Type
Scan Display
2.11 Measure System
The Oscilloscope displays the voltage in relation to time and test the wave form
displayed. There are scale, Cursor and auto measure modes.
2.11.1 Scale Measurement
This method allows you to make a quick, visual estimate. For example, you might look
at waveform amplitude and determine that it is a little more than 100 mV. You can take
simple measurements by counting the major and minor graticule divisions involved and
multiplying by the scale factor. For example, if you counted five major vertical graticule
divisions between the minimum and maximum values of a waveform and knew you had
a scale factor of 100mV/div, then you could easily calculate your peak-to-peak voltage
as follows:
5 divisions x 100 mV/division = 500 mV.
2.11.2 Cursor Measurement
Picture 2-50 displays the cursor button on the front-panel for this menu.
Picture 2.11-1
Press the “CURSORS” button to display the “Cursor” Menu.
The cursor measurement has three modes: Manual, Track and Auto Measure.
Cursor Mode Manual In this menu, set the manual cursor measure.
Type
Voltage
Time
Use cursor to measure voltage parameters.
Use cursor to measure time parameters.
CH1
CH2
Source
MATH
Select input signal channel.
REFA
REFB
Cur A
Cur B
Select this option use “Universal” knob to
adjust cursor A.
Select this option use “Universal” knob to
adjust cursor B.
In this mode, the screen displays two horizontal parallel cursors or vertical parallel
cursors to measure voltage or time. You can move the cursor by turning the “Universal”
knob. Before using cursors, you should make sure that you have set the signal source
as the channel for measuring.
■ Voltage Cursor: Voltage cursors appear as horizontal lines on the display and
measure the vertical parameters.
■ Time Cursor: Time cursors appear as vertical lines on the display and measure the
horizontal parameters.
■ Cursor Moving: Use the “universal” knob to move cursor1 and cursor2 .They could
be moved when the corresponding cursor option are selected, and cursor value will
display on the bottom left and top left of the screen when you move the cursor.
To do manual cursor measurements, follow these steps:
1. Press CURSOR button to enter the cursor function menu.
2. Press the “Cursor Mode” option button to select “Manual”.
3. Press the “Type” option button to select “Voltage” or “Time”.
4. Press the “Source” option button to select “CH1”, “CH2”, “MATH”,
“REFA” ,“REFB” according to input signal channel.
5. Select “Cur A”, turn the “Universal” knob to adjust Cursor A.
6. Select “Cur B”, turn the “Universal” knob to adjust Cursor B.
7. The measurement values are displayed on the top of the left corner:
If the measurement type is set to “Voltage”, the values are:
The value of Cur A: CurA
The value of Cur B: Cur B
The voltage increment between Cursor A and Cursor B: V
If the measurement type is set to “Time”, the values are:
The value of Cur A: CurA
The value of Cur B: Cur B
The time increment between Cursor A and Cursor B: T
The reciprocal of time increment between Cursor A and Cursor B: 1/T
Cursor Mode Track In this mode, set track cursor measure.
Cursor A
Cursor B
Cur A
Cur B
CH1
CH2
NONE
CH1
CH2
NONE
Set the input signal channel that the Cursor A
will measure.
Set the input signal channel that the Cursor B
will measure.
Select this option, turn the “Universal” knob to
adjust horizontal coordinate of Cursor A.
Select this option, turn the “Universal” knob to
adjust horizontal coordinate of Cursor B.
In this mode, the screen displays two cross cursors. The cross cursor sets the position
on the waveform automatically. You could adjust cursor’s horizontal position on the
waveform by turning the “Universal” knob. The oscilloscope displays the values on the
top of the right screen.
To do track cursor measurement, follow these steps:
1. Press CURSOR button to enter the cursor measure function menu.
2. Press the “Cursor Mode” option button to select “Track”.
3. Press the “Cursor A” option button to select the input signal channel.
4. Press the “Cursor B” option button to select the input signal channel.
5. Select “Cur A”, turn the “Universal” knob to move Cursor A horizontally.
6. Select “Cur B”, turn the “Universal” knob to move Cursor B horizontally.
7. The measurement values are displayed on the left of the top corner:
A T: The horizontal position of Cursor A (Time cursor centered around the
midpoint of screen).
AV: The Vertical position of Cursor A (Voltage cursor centered around channel
This mode will take effect with automatic measurements. The instruments will display
cursors while measuring parameters automatically. These cursors demonstrate the
physical meanings of these measurements.
To do auto cursor measurements, follow these steps:
1. Press the CURSOR button to enter “Cursor measure menu”.
2. Press the “Cursor Mode” option button to select “Auto”.
3. Press the “MEASURE” button to enter “Auto cursor measure mode menu”
to select the parameter that you want to measure.
When you take automatic measurements, the oscilloscope does all the calculating for
you. Because the measurements use the waveform record points, they are more
accurate than the graticule or cursor measurements.
Press the ‘MEASURE’ for the Automatic Test.
There are three auto measurement types: Voltage Measure, Time Measure and Delay
Measure. There are thirty two measure parameters in all.
Table 2-35 Auto measure function menu 1:
Option Instruction
Voltage Press this button to enter the Voltage measure menu.
Time Press this button to enter the Time measure menu.
Delay Press this button to enter the Delay measure menu.
All Mea Press this button to enter the All Measurement menu.
Clear
Press this button to clear all the measurement that you chose.
Return to the first page of auto
measurement menu.
Table 2-38 Auto measure function4-Delay measure menu:
Option Setting Instruction
Source CH1-CH2 Select any two input signal
source for Delay measure.
Type PhaseFRRFRFFFRFFF
LRRLRFLFRLFF
Press the “Type” button or turn
the “Universal” knob to select
Delay measure parameter.
Choose the measurement, after
pressing the button, the
Add
measurement value will be
shown in the lower
left of the screen
Return Return to the first page of auto
measurement menu.
Table 2-39 All Measurement function menu:
Option Setting Instruction
Source
CH1
CH2
Select input signal channel.
Turn on the all measurement function to
Voltage
On
Off
measure voltage parameters.
Turn of the all measurement function to
measure voltage parameters.
Turn on the all measurement function to
Time
On
Off
measure Time parameters.
Turn of the all measurement function to
measure Time parameters.
Return Return to the “All Measure main menu”.
Table 2-40 Introduction of the Measure Type:
Measure Type Introduction
Vmax
Vmin
Vpp
The most positive peak voltage measured over the entire
waveform.
The most negative peak voltage measured over the
entire waveform.
Measures the absolute difference between the
maximum and minimum peaks of the entire waveform.
Measures the highest voltage over the entire waveform.
Measures the lowest voltage over the entire waveform.
Voltage between Vhig and Vlow of a waveform.
The arithmetic mean over the first cycle in the waveform.
The arithmetic mean over the entire waveform.
The true Root Mean Square voltage over the first cycle
in the waveform.
The true Root Mean Square voltage over the entire
waveform.
Defined as (Vmax-Vhig)/Vamp after the waveform
rising.
Defined as (Vmin-Vlow)/Vamp after the waveform
falling.
Defined as (Vmin-Vlow)/Vamp before the waveform
rising.
Defined as (Vmax-Vhig)/Vamp before the waveform
falling.
Rise Time Measures the time between 10% and 90% of
the first rising edge of the waveform.
Fall Time Measures the time between 90% and 10% of
the first falling edge of the waveform.
The duration of a burst.
Measured over the entire waveform.
+ Width Measures the time between the first rising edge
and the next falling edge at the waveform 50% level.
- Wid
+ Duty
Duty
The image
cannot be
Phase
FRR
FRF
and the next rising edge at the waveform 50% level.
Measures the first cycle waveform.
Positive Duty Cycle is the ratio between positive pulse
width and period.
-Width Measures the time between the first falling edge
Measures the first cycle waveform. Negative Duty Cycle
is the ratio between negative pulse width and period.
The amount one waveform leads or lags another in time.
Expressed in degrees, where 360 degrees comprise
one waveform cycle.
The time between the first rising edge of source X and
the first rising edge of source Y.
The time between the first rising edge of source X and
the first falling edge of source Y.
The time between the first falling edge of source X and
the first rising edge of source Y.
The time between the first falling edge of source X and
the first falling edge of source Y.
The time between the first rising edge of source 1 and the
last rising edge of source 2.
The time between the first rising edge of source X and
the last falling edge of source Y.
The time between the first falling edge of source X and
the last rising edge of source Y.
The time between the first falling edge of source X and
the last falling edge of source Y.
If you want to measure voltage parameters, please follow next steps:
1. Press the “MEASURE” button to enter the “Auto measurement” menu.
2. Press the top first option button to enter the “second measurement menu”.
3 . Select measure type. If you press the “Voltage” option button, “Voltage
measurement” menu will display on the screen.
4. Press the “Source” option button to select “CH1”, “CH2” according to input signal
channel.
5. Press the “Type” option button to select the parameter type that you want to
measure, then press the “add” option button to add the measurement . The
corresponding icon and value will displayed in the lower left of the
screen.
6. Press the “Return” option button to return the home page of “Auto Measurement”
menu. The selected parameter and the corresponding value will display on the top
first position of the home page.
You can display the other parameters and its’ value on the corresponding position
using the same way, and the screen can display five parameters at one time.
If you want to measure time parameters using all measure function, please
follow next steps:
1. Press the “MEASURE” button to enter the “Auto Measure menu”.
2.Press the top option button to enter the second page of “Auto Measure menu”.
3. Press “All Mea” option button to enter the “All Measure menu”.
4. Press “Source” option button to select input signal channel.
5. Press the “Time” option button to select “On”. Now all of time parameters value
will display on the screen at the same time.See Picture
Showing as following picture, The SAVE/RECALL is the Storage System Function
Button.
Picture 2.12-1
You can quickly save and recall up to 20 oscilloscope panel settings, 10 groups’
waveforms in internal memory. There is a USB Host interface in the front panel of the
oscilloscope and you can save setup data, waveform data, waveform interface image,
CSV file to a USB flash drive furthest at a time. Setup data and waveform data can be
recalled to the current oscilloscope or the other same model oscilloscopes. Picture data
and CSV file can’t be recalled to the oscilloscope, but it can be opened on the correlative
computer software.
■ Using SAVE /RECALL Screen
The SAVE ALL screen is divided into functions based on Files or Directory.
Files shows option button for New File, Del File and Load.
Picture 2.12-2
While Directory shows option buttons for New Dir., Del Folder.
Picture 2.12-3
Recalling Files
The Load button is used to recall your setup files. Once you’ve navigated to the desired
file and it’s highlighted in the main screen area, press the Load option button and the
setup is recalled from the USB flash drive.
Note: The Load button option is disabled when BMP or CSV file types are selected.
Both Directory and Files have Rename and Return option buttons on Page2/2. Use
these buttons to give an existing folder of file a new name or to exit the SAVE/REC ALL
screen, respectively.
Picture 2.12-4
Creating Directory and File
Create new directory and file by pressing the New Dir or New File option button. The
following screen is shown.
● The New Filemenu choices and behavior is the same as the New Folder menu. It just
has a different heading. The InputChar option button adds the selected character to the
courser position in Name field.
● Move the cursor position in the name field using the “”and “” option buttons.
● Turn the Universal knob to move through character selections. When the desired
character is highlighted, press the Universal knob or press he “InputChar” option button
to add it to the specific position in the Name field.
● Additional selections for BackSpace, DeleteCharacter, and CleanName are available
for convenience and are also accessed using the “Universal” knob in the same manner.
● Press the “Confirm” option button (when you have the Name field completed as
desired) to save the file onto the memory device. After the “Confirm” option button is
pressed, a “Data Store Success!” messages briefly shown and the new folder or file is
shown on your USB flash drive.
SAVE/RECALL SETUP
■ Save Setups to Device
you recall the setup, the oscilloscope will be in the mode from which the setup was
The complete setup is stored in nonvolatile memory. When
saved.
The oscilloscope saves the current setup if you wait three seconds after the last change
before you power off the oscilloscope. The oscilloscope recalls this setup the next time
you apply power.
Menu for the Storage/Recall setting in the
oscilloscope
Save setup to the oscilloscope’s internal
75
Page 78
memorizer.
Setup No.1 to No.20
Save
Recall
Press the “Setup” option button or turn the
“universal” knob to select storage position.
Accomplish the storage
Recall the storage in the “Setup” operation
Picture 2.12-6
Save setups to the oscilloscope’s internal memorizer, follow next steps:
For example: Save setup that set waveform display type to “Dots” to internal memory.
1Press the “SAVE/RECALL” button to enter the “SAVE/RECALL” menu.
2Press the “Type” option button to select “setups”.
3Press the “Save to” option button to select “Device”.
4Press the “Setup” option button to select “No.1”.
5Press the “DISPLAY” button to enter the “Display” menu.
6Press the “Type” option button to select “Dots”.
7Press the “SAVE/RECALL” button to enter the “SAVE/RECALL” menu.
8Press the “Save” option button.
Recalling setup, follow next steps:
1Press the “SAVE/RECALL” button to enter the “SAVE/RECALL” display menu.
2Press the “Type” option button to select “Setups”.
3Press the “Save to” option button to select “Device”.
4Press the “Setup” option button or turn the “Universal” knob to select “No.1”.
Save waveforms in internal memory, follow next steps:
1. Input a sine signal to channel 1 and press the “Auto” button.
2. Press the “SAVE/RECALL” button to enter “SAVE/RECALL” display menu.
3. Press the “Type” option button to select “waveforms”.
4. Press the “Save to” option button to select “Device”.
5. Press the “waveform” option button or turn the “Universal” knob to select “No.1”.
6.Turn the “Volts/div” knob or “S/div” knob to adjust the waveform that you want to need.
7. Press the “Save” option button.
Recall waveforms, follow next steps:
1. Press the “SAVE/RECALL” button to enter “SAVE/RECALL” display menu.
2. Press the “Type” option button to select “waveforms”.
3. Press the “Save to” option button to select “Device”.
4. Press the “waveform” option button or turn the “Universal” knob to select “No.1”.
5. Press the “Recall” option button.
Table 2-45 Save waveforms to USB flash drive function menu
Option Setup
Type Waveforms
Save to File
Save
Introduction
Menu for the Storage/Recall waveforms.
Save waveforms to USB flash drive.
Accomplish the storage.
Picture 2.12-10
Save waveforms to USB flash drive, follow next steps:
1Input a sine signal to channel 1, press the “AUTO” button.
2Press the “SAVE/RECALL” button to enter the “SAVE/RECALL” display menu.
3Press the “Type” option button to select “Waveforms”.
4Insert USB flash drive to USB host port of the oscilloscope and wait that the
oscilloscope has initialized USB flash drive (about five seconds).
5Press the “Save to” option button to select “File”.
6Press the “Save” option button then you’ll go into the Save/Recall interface.
7Create a file then press the “Confirm” button (about five seconds, there is a
prompting message that “Save data success” displaying on the screen), now
waveform data have been saved to USB flash drive
Recall waveforms form USB flash drive, follow next steps:
1. Press the “SAVE/RECALL” button.
2. Press the “Type” button to select “Waveforms”.
3 . Insert USB flash drive to USB host port of the oscilloscope and wait that the
oscilloscope has initialized USB flash drive (about five seconds).
4. Press the “Save to” option button to select “File”.
5. Press the “Save” option button then go into the Save/Recall interface.
6. Choose the file you want then press the “Load” option button (about five seconds,
there is a prompting message that “Recall data success” displaying on the screen),
now waveform data have been recalled from USB flash drive.
■ Save Picture
Waveform interface image can be saved to USB flash drive, but they can’t be recalled.
You can open them on correlative computer software.
Table 2-46 Picture storage function menu:
Option Setting
Type Pictures
Save
Introduction
Menu for the Storage/Recall waveform
interface image.
Go to the Save/Recall interface.
Self Calibration procedure can optimize the signal path at the most measurement
precision. You can run this procedure at any time. If the operating temperature changes
by more than 5° C or the units run more than thirty minutes, you should do the self
calibration.
When you run the self Calibration, you should cut off all the probes and lead. Then
press the “Utility” button to choose the “Do self cal” to show the self Calibration Menu,
and execute self calibration program according to the prompting on the screen.
Screen Test Run Screen Test Program.
Keyboard Test Run Keyboard Test Program.
LED Test Run LED Test Program.
Operation Steps:
1. Screen Test:
Select “Screen Test” to enter the screen test interface. The clew words “Press ‘SINGLE’
Key to continuePress ‘RUN/STOP’ Key to exit” is displayed, you could press the “Single”
for test.
Picture 2.13-9
2. Keyboard Test
Select “keyboard Test” to enter the key test interface, the on-screen lathy rectangle
shapes represent the front panel keys. The shapes with two arrows beside them
represent the front panel knobs. The squares represent the knob presses for scale
knobs. Test all keys and knobs and you should also verify that all the backlit buttons
When you operate, the screen would display the white (Color LCD) and black (single
LCD) repose.
The tested button or knobs corresponding area would display green (Color LCD) or
white (single LCD).
At the bottom of the screen display “Press‘ RUN/STOP’ Key Three Times to exit”
information prompt to show that press “RUN/STOP” three times for quitting the test .
Picture 2.13-10
3. LED test
Select “LED Test” to enter the lighten interface, the on-screen lathy rectangle shapes
represent the front panel keys; the lathy rectangle shapes represent the front panel
knobs. The squares represent the knob presses for scale knobs. The clew words “Press
‘SINGLE’ Key to continue , Press ‘RUN/STOP’ Key to exit” is displayed, you could press
the “Single” button continuously for testing, when buttons are lighted ,the corresponding
area on the screen would display green(Color LCD) or white(Single LCD ).
Turn the “Universal” knob to set the range of
horizontal clearance to the waveform.
<0.04div-4.00div>
Turn the “Universal” knob to set the range of
vertical clearance to the waveform.
<0.04div-4.00div>
Create a test mask according to the above
clearance.
Select position to store created mask.
Enter the second page of “Mask Setting
11) Press the “Create Mask” button to create mask and you also can enter the next
page of “Mask menu” to recall the stored mask.
12) Enter the second page of “Pass/Fail function menu”, press the “Output” option
button to set the output option.
13) Enter the first page of “Pass/Fail function menu”, press the “Operate” option button
to select “”to run the pass/fail test.
Picture 2.13-16
2.13.7 Waveform Record
Waveform record can record input waveform from CH1 and CH2 with a maximum record
length of 2500 frames. This record behavior can also be activated by the pass/fail test
output, which makes this function especially useful to capture abnormal signals in long
term without keeping an eye watching it.
Waveform recorder: Record the waveforms with specified interval.
Table 2-59 Waveform recorder function menu:
Option Setting Instruction
Mode
Record
Play Back
Storage
Off
Set Record function menu.
Set Play Back function menu.
Set Storage function menu.
Turn off waveform recorder menu.