This operating instruction manual and other GLI instruction
manuals are available on GLI’s web site at gliint.com.
when viewed using Adobe’s free Acrobat reader. To get
this reader, link to Adobe through GLI’s web site or visit
Adobe’s web site at adobe.com.
Model P33 pH/ORP Analyzer (panel-mount 1/4 DIN)Rev. 6-1100
2
IMPORTANT SAFETY INFORMATION
This analyzer is compliant with safety standards as outlined in:
FMRC Class Numbers 3600, 3611, and 3810 (U.S.A.)
CSA C22.2 No. 142 and C22.2 No. 213 (Canada)
EN 61010-1 (European Community)
Please read and observe the following:
•
Line voltage may be present at terminals on TB1 at the back of the analyzer enclosure. This may be
hazardous. Always remove line power before going near this area of the analyzer. The front bezel
assembly of the analyzer, however, contains only low voltage and is completely safe to handle.
•
Wiring or repairs should only be performed by qualified personnel and only to an unpowered analyzer.
•
Whenever it appears that analyzer safety is questionable, disable the analyzer to ensure against any
unintended operation. For example, an unsafe condition is likely when:
1) The analyzer appears visibly damaged.
2) The analyzer fails to operate properly or provide the intended measurements.
3) The analyzer has been stored for long periods at temperatures above 158°F (70°C).
•
This analyzer must be installed by specially trained personnel in accordance with relevant local codes
and instructions contained in this operating instruction manual. Observe the analyzer’s technical
specifications and input ratings. If one line of the line power mains is not neutral, use a double-pole
mains switch to disconnect the analyzer.
HELPFUL IDENTIFIERS
In addition to information on installation and operation, this instruction manual may contain
WARNINGS pertaining to user safety, CAUTIONS regarding possible instrument malfunction,
and NOTES on important, useful operating guidelines.
WARNING:
A WARNING LOOKS LIKE THIS. IT WARNS YOU OF THE
POTENTIAL FOR PERSONAL INJURY.
CAUTION:
A CAUTION LOOKS LIKE THIS. IT ALERTS YOU TO POSSIBLE
INSTRUMENT MALFUNCTION OR DAMAGE.
NOTE:A note looks like this. It alerts you to important operating
and alerts the user to possible danger or instrument malfunction. Refer to this manual before
proceeding.
This symbol, which appears on the analyzer POWER terminal block
(shown in Figure 2-2),
terminal
This symbol
and alerts the user to connect an earth ground to it.
means that there is alternating current present
means that this is a protective ground
and
alerts the user to be careful.
GLI International, Inc. warrants the Model P33 to be free from defects in material
or workmanship for a period of 2 years (24 months) from the date of shipment of
this product from our facility. A warranty claim will not be honored if defects are
not reported within the warranty period, or if GLI International determines that
defects or damages are due to normal wear, misapplication, lack of maintenance, abuse, improper installation, alteration, or abnormal conditions. GLI
International’s obligation under this warranty shall be limited to, at its option, replacement or repair of this product. The product must be returned to GLI
International, freight prepaid, for examination. The product must be thoroughly
cleaned and any process chemicals removed before it will be accepted for replacement or repair. GLI International’s liability shall not exceed the cost of the
product. Under no circumstances will GLI International be liable for any incidental or consequential damages, whether to person or property. GLI International
will not be liable for any other loss, damage or expense of any kind, including
loss of profits, resulting from the installation, use, or inability to use this product.
Model P33 pH/ORP Analyzer (panel-mount 1/4 DIN)Rev. 6-1100
WARRANTY
4
CONDENSED OPERATING INSTRUCTIONS
This manual contains details for all operating aspects of the instrument. The following condensed instructions are provided to assist you in getting the instrument started up and
operating as quickly as possible.
measurement operation using a GLI Differential pH sensor.
conventional combination electrode or specific features of the instrument, refer to the appropriate sections in this manual for instructions.
A. CONNECTING SENSOR/CONFIGURING SENSOR TYPE AND TEMPERATURE ELEMENT
1.After properly mounting the analyzer (PART TWO, Section 2), connect the GLI 5-wire
Differential Technique pH sensor, matching wire colors to terminals as indicated:
Sensor Wire ColorsConnect to SENSOR Terminal
2.The analyzer is supplied factory-set for use with the GLI 5-wire Differential Technique
pH sensor. When using a conventional combination electrode, you must change the
sensor type (see PART THREE, Section 3.2, subheading “SELECT SENSOR Type”).
These condensed instructions only pertain to basic pH
To measure ORP, or use a
Yellow#4 on TB3
Shield#5 on TB3
Black#6 on TB3
White#7 on TB3
Green#8 on TB3
Red#1 on TB4
3.The analyzer is supplied factory-set for automatic temperature compensation using the
NTC 300 ohm temperature element built into all GLI Differential sensors (except Model
6006P4-2000 high purity water pH sensor systems which use Pt1000 ohm RTD).
When using a sensor with a different temperature element, or if you want fixed
MANUAL temperature compensation, you must change the temperature element type
(see PART THREE, Section 3.2, subheading “Select TEMP ELEMENT Type”).
B. CONNECTING LINE POWER
Important:
Follow instructions in PART TWO, Section 3.6 to connect line power to the analyzer.
C. CONFIGURING BUFFER TYPE/CALIBRATING THE ANALYZER
The analyzer must be calibrated so that measured values will correspond to actual process
values. Before calibrating for the
first time, select the buffer value set that will be used.
Then, calibrate using the recommended “2 POINT BUFFER” method which will provide the
most accurate pH measurements.
1.The analyzer is supplied factory-set for the common 4.00, 7.00, and 10.00 pH buffers.
When using DIN 19267 standard value buffers you must change the buffer set (see
PART THREE, Section 3.2, subheading “SELECT BUFFER Set”).
NOTE:When using buffers that are not included in either of the analyzer buffer sets,
use only the “2 POINT SAMPLE” method for calibration. Refer to that subheading in PART THREE, Section 4.2 for instructions.
2.Immerse the sensor in the first buffer (preferably pH 7).
and buffer temperatures to equalize. Depending on their temperature differences, this
may take 30 minutes or more.
Important:
Allow the sensor
NOTE:An in-progress calibration can always be aborted by pressing the ESC key.
After the “ABORT: YES?” screen appears, do one of the following:
Press ENTER key to abort. After “CONFIRM ACTIVE?” screen appears,
•
press ENTER key to display the MEASURE screen, and return the ana-
log outputs and relays to their active states.
×
•
Use
×
Ø
or
key to choose “ABORT: NO?” screen, and press ENTER key
Ø
to continue calibration.
Calibration Tip! If, at any time during calibration, the “2 POINT BUFFER:
CONFIRM FAILURE?” screen appears, press ENTER key to confirm. Then, use the
×
Ø
or
×
key to select between “CAL: EXIT” or “CAL: REPEAT” and do one of the
Ø
following:
With the “2 POINT BUFFER: CAL EXIT?” screen selected, press ENTER key.
•
Then, after the “2 POINT BUFFER: CONFIRM ACTIVE?” screen appears, press
ENTER key to return the analog outputs and relays to their active states
(MEASURE screen appears).
With the “2 POINT BUFFER: CAL REPEAT?” screen selected, press ENTER key
•
to repeat calibration of this point.
3.Press
not showing, use
4.Press
5.Press
6.Press
7.Press
during calibration. (Outputs can also be transferred to preset values or allowed to remain active.)
Model P33 pH/ORP Analyzer (panel-mount 1/4 DIN)Rev. 6-1100
MENU key
ENTER key
ENTER key
ENTER key
ENTER key
to display a “MAIN MENU” screen. If the screen is
Ø
Ø
to display
again to display
again to display
again to “hold” the analog outputs and relays at their present states
or
×
×
to display it.
key
(continued on next page)
6
.
.
.
CONDENSED OPERATING INSTRUCTIONS
C. CALIBRATING THE ANALYZER -- (continued)
8.With the sensor in the first buffer and the
ENTER key
to confirm. While the
screen is displayed, the analyzer
screen displayed, press
waits for the pH and temperature signals to stabilize, measures the buffer value, and
automatically calibrates this point. Thereafter, a screen like
appears
for 5 seconds to confirm calibration of this point.
NOTE:Any time the “PLEASE WAIT” screen is displayed during calibration, you can
manually complete calibration of the point by pressing the ENTER key. However, this is not recommended because the pH and temperature signals may
not be fully stabilized, resulting in an inaccurate calibration.
9.After the screen appears, remove the sensor from the first buffer,
rinse it with clean water, and immerse it in the second buffer (typically pH 4).
10. Press
ENTER key
to confirm. While the screen is displayed, the
analyzer waits for the pH and temperature signals to stabilize, measures the buffer
value, and automatically calibrates this point. Thereafter, a screen like
appears for 5 seconds to confirm calibration of this point.
11. A “pH SLOPE XX.X mV/pH” screen appears, indicating a slope value to measure sen-
sor performance. The slope should be within 54 to 62 mV/pH for optimal performance.
12. Press
ENTER key
to end calibration (“2 POINT BUFFER: CONFIRM CAL OK?” screen
appears).
13. Re-install the sensor into the process.
14. Press
ENTER key
to display the active measurement reading on the “2 POINT
BUFFER: CONFIRM ACTIVE?” output status screen. When the reading corresponds
to the actual typical process value, press
ENTER key
again to return the analog out-
puts and relays to their active states (MEASURE screen appears).
This completes “2 POINT BUFFER” calibration. The analyzer is now ready to measure pH.
D. COMPLETING ANALYZER CONFIGURATION
To further configure the analyzer to your application requirements, use the appropriate
CONFIGURE screens to make selections and “key in” values. Refer to PART THREE, Section 3 for complete configuration details.
Model P33 pH/ORP Analyzer (panel-mount 1/4 DIN)Rev. 6-1100
14
PART ONE - INTRODUCTIONSECTION 1 - GENERAL INFORMATION
PART ONE - INTRODUCTION
ECTION
1.1Capability Highlights
Sensor Input
S
GENERAL INFORMATION
The analyzer can be used with any GLI 5-wire Differential
Technique pH or ORP sensor, or any conventional combination electrode. The analyzer will accept an NTC 300 ohm
thermistor, Pt 1000 RTD or Pt 100 RTD temperature compensator element. Typically, one of these elements is built
into the pH or ORP sensor.
1
MEASURE Screen
Passcode-protected
Access
Calibration Methods
The MEASURE screen (normal display mode) can provide
different readouts of measured data. With the MEASURE
screen displayed, press
• Measured pH (or ORP, if selected).
• Measured temperature (°C or °F).
• Analog Output 1 and 2 values (mA).
• Measured pH (or ORP) and temperature.
For security, you can enable a passcode feature to restrict
access to configuration and calibration settings to authorized personnel only. See PART THREE, Section 3.6 for
details.
Four methods are available to calibrate the analyzer for pH.
See PART THREE, Section 4.2 for details. For ORP calibration, refer to Section 4.3. Each analog output mA value
can also be calibrated (Section 4.4).
The analyzer provides two isolated analog outputs (1 and
2). Each output can be set to be 0-20 mA or 4-20 mA, and
assigned to represent one of these measurements:
• Measured pH (or ORP).
• Measured temperature.
Parameter values can be entered to define the endpoints at
which the minimum and maximum analog output values are
desired (range expand). For analog output setup details,
refer to PART THREE, Section 3.4.
15
PART ONE - INTRODUCTIONSECTION 1 - GENERAL INFORMATION
During calibration, both analog outputs can be selected to:
•
Hold their present values (HOLD OUTPUTS).
•
Transfer to preset values to operate control elements
by an amount corresponding to those values (XFER
OUTPUTS).
•
Remain active to respond to the measured value
(ACTIVE OUTPUTS).
Relays
☞
The analyzer has two electromechanical relays with SPDT
contacts. Each relay can be set to function as a CONTROL,
ALARM or STATUS relay. CONTROL and ALARM relays
can be assigned to be driven by one of these:
•
Measured pH (or ORP).
•
Measured temperature.
NOTE: When a relay is set to function as a STATUS relay,
it is no longer configurable. Instead, it becomes a
dedicated system diagnostic-only alarm relay that
automatically energizes when the “WARNING
CHECK STATUS” message flashes on the
MEASURE screen. This occurs when the analyzer
detects a “fail” diagnostic condition. See PART
THREE, Section 5.1 for more details.
Except for STATUS relays, during calibration the relay
on/off states are affected in the same way as the analog
outputs by the “(HOLD/XFER/ACTIVE) OUTPUTS” screen
selection. These relays are also held at their present on/off
states, transferred to desired preset on/off states, or remain
active to respond to measured values.
1.2Modular Construction
Model P33 pH/ORP Analyzer (panel-mount 1/4 DIN)Rev. 6-1100
See PART THREE, Section 3.5 for relay setup details.
The modular construction of the analyzer provides electrical
safety. The front panel keypad assembly uses voltages no
greater than 24 VDC, and is completely safe to handle.
16
PART ONE - INTRODUCTIONSECTION 1 - GENERAL INFORMATION
Line power must be connected to specifically designated
terminals on TB1.
WARNING:
REMOVE LINE POWER BEFORE NEARING THIS
AREA TO PREVENT ELECTRICAL SHOCK.
1.3Retained
Configuration Values
1.4Analyzer
Serial Number
1.5EMI/RFI Immunity
All user-entered configuration values are retained indefinitely, even if power is lost or turned off. The non-volatile
analyzer memory does not require battery backup.
A label with the analyzer model number, serial number,
build date, and other items is located on top of the enclosure.
The analyzer is designed to provide protection from most
normally encountered electromagnetic interference. This
protection exceeds U.S. standards and meets European
IEC 801-series testing for electromagnetic and radio frequency emissions and susceptibility. Refer to Figure 1-1
and the specifications in Section 2.1 for more information.
Operational Mode ............Each relay (A and B) can be assigned to be driven
by the measured pH (or ORP) or tem perature
Function Modes:
Control.................... Settings for high/low phasing, setpoint, dead-
band, overfeed timer, off delay, and on delay
Alarm..........................Settings for low alarm point, low alarm point
deadband, high alarm point, high alarm point
deadband, off delay, and on delay
Status.........................Not configurable; relay only activates when a
sensor or analyzer diagnostic WARNING
condition exists
Indicators.........................Relay A and B LEDs indicate respective relay status
Temperature Compensation ....... Automatic from 14.0 to 230.0°F (-10.0 to
+110.0°C), with selection for NTC 300 ohm
thermistor, Pt 1000 ohm RTD or Pt 100 ohm
RTD temperature element, or manually fixed
at a user-entered temperature; additional
selectable temperature correction factors
(ammonia or morpholine) for pure water
automatic compensation from 0.0-50.0°C
Sensor-to-Analyzer Distance:
GLI 5-wire Differential
Technique Sensor.............. 3000 ft. (914 m) maximum
Conventional Combination
Electrode with preamp ....... 985 ft. (300 m) maximum
Conventional Combination
Electrode without preamp .. 100 ft. (30 m) maximum with electrode cable
capacitance of less than 30 pF per foot
Power Requirements .................. 90-130 VAC, 50/60 Hz. (10 VA max.) or
190-260 VAC, 50/60 Hz. (10 VA max.)
Calibration Methods:
2 POINT BUFFER................... Automatic calibration and buffer recognition
(for pH only)using two buffers from selected buffer set*.
NOTE:
When using buffers that are not included in either of the analyzer
buffer sets, use only the “2 POINT SAMPLE” method for calibration.
*Buffer Sets: 4.00, 7.00, and 10.00 pH; or
DIN 19267 standard (1.09, 4.65, 6.79, 9.23, and 12.75 pH)
Model P33 pH/ORP Analyzer (panel-mount 1/4 DIN)Rev. 6-1100
18
PART ONE - INTRODUCTIONSECTION 2 - SPECIFICATIONS
1 POINT BUFFER .................. Automatic calibration and buffer recognition
(for pH only)using one buffer from selected buffer set*.
NOTE:
When using a buffer that is not included in either of the analyzer
buffer sets, use only the “1 POINT SAMPLE” method for calibration.
2 POINT SAMPLE .................. Enter known values of two samples
(for pH only)(determined by laboratory analysis or
comparison reading) or two pH buffers
1 POINT SAMPLE .................. Enter known value of one sample (determined
(for pH or ORP)by laboratory analysis or comparison
reading), one pH buffer, or one reference
solution (for ORP measurement)
Analog Outputs .......................... Two isolated 0/4-20 mA outputs; each with
0.004 mA (12-bit) resolution and capability
to drive up to 600 ohm loads
NOTE:
Each output can be assigned to represent the measured pH (or ORP)
or temperature. Parameter values can be entered to define the endpoints at which the minimum and maximum mA output values are
desired (range expand). During calibration, both outputs can be selected to hold their present values, transfer to preset values to
operate control elements by an amount corresponding to those values, or remain active to respond to the measured value.
Communication: RS-232 ........... Enables configuration and retrieval of measured
data for one analyzer using IBM-compatible
PC and optional GLI software tool kit
HART.............. Enables configuration and retrieval of measured
data for multiple analyzers over a communication link using appropriate hand-held
terminal or data system with HART software
2.2Analyzer Performance
(Electrical, Analog Outputs)
2.3Mechanical
Memory Backup (non-volatile) .... All user settings are retained indefinitely in
memory (EEPROM)
EMI/RFI Conformance................ Exceeds U.S. and meets European standards
for conducted and radiated emissions and
immunity; certified CE compliant for applications as specified by EN 50081-2 for
emissions and EN 50082-2 for immunity
Electrical Certifications:
General Purpose (pending)..... UL, C-UL, FM, and CENELEC
Division 2 (pending)................ UL, C-UL, and FM: Groups A, B, C, D, F, and G
Zone 2 (pending) .................... CENELEC: Group IIC
Accuracy.................................... 0.1% of span
Stability...................................... 0.05% of span per 24 hours, non-cumulative
Repeatability .............................. 0.1% of span or better
Temperature Drift....................... Zero and Span: less than 0.03% of span per °C
Enclosure................................... Polycarbonate with NEMA 4X front panel;
general purpose; two brackets supplied for
panel mounting
Net Weight................................. 1.7 lbs. (0.8 kg) approximately
19
PART TW O - INSTALLATIONSECTION 1 - UNPACKING
PART TWO - INSTALLATION
ECTION
S
UNPACKING
After unpacking, it is recommended to save the shipping
carton and packing materials in case the instrument must be
stored or re-shipped. Inspect the equipment and packing
materials for signs of shipping damage. If there is any evidence of damage, notify the transit carrier immediately.
ECTION
S
MECHANICAL REQUIREMENTS
1
2
2.1Location
1. It is recommended to locate the analyzer as close as
possible to the installed sensor. Depending on the sensor type, the maximum allowable distance between the
sensor and analyzer is:
GLI 5-wire
Differential
Technique Sensor
3000 feet (914 m)985 feet (300 m)100 feet (30 m)
Recommendation:
Conventional
Combination Elec-
with
trode
Preamp
Conventional
Combination Electrode
without
Preamp
Directly connect the sensor to the
analyzer to eliminate potential problems caused by wet
environments when a junction box is used.
2. Mount the analyzer in a location that is:
➥
Clean and dry where there is little or no vibration.
➥
Protected from corrosive fluids.
➥
Within ambient temperature limits (-4 to +140°F or
-20 to +60°C).
CAUTION:
2.2Mounting
Model P33 pH/ORP Analyzer (panel-mount 1/4 DIN)Rev. 6-1100
EXPOSING THE ANALYZER TO DIRECT
SUNLIGHT MAY INCREASE THE OPERATING
TEMPERATURE ABOVE ITS SPECIFIED LIMIT,
AND DECREASE DISPLAY VISIBILITY.
Figure 2-1 or 2-2 illustrates the analyzer enclosure dimensions and panel mounting details. Using the two supplied
brackets, attach them to the analyzer case as shown to
panel mount the analyzer.
20
PART TW O - INSTALLATIONSECTION 2 - MECHANICAL REQUIREMENTS
FIGURE 2-1 Enclosure Dimension Details for Analyzers with Letter “A” Prefix Serial Number
FIGURE 2-2 Enclosure Dimension Details for Analyzers with “No Letter” Prefix Serial Number
PART TW O - INSTALLATIONSECTION 3 - ELECTRICAL CONNECTIONS
ECTION
S
ELECTRICAL CONNECTIONS
Figure 2-3 or 2-4 on the next page shows the terminal arrangement and designations on the back of the analyzer.
3
☞
☞
NOTE: For easier wiring, terminal blocks can be unplugged
from their mating connectors. All terminals are suitable for single wires up to 14 AWG (2.5 mm2).
Wiring Tip!
electromagnetic compatibility requirements, follow these
general wiring guidelines:
1.Keep all cable shields as short as possible and connect them to earth ground.
2.Use Steward ferrite 28 B0590-000 or equivalent on:
◆
◆
◆
◆
3.In harsh conducted RF conditions, connect the earth
ground of the analyzer (Terminal 4 on TB1) to a local, known earth ground source.
To comply with European Community (CE)
Mains (line power) cable -- no turns required.
Sensor cable -- one turn required.
mA analog output cables -- two turns required.
Relay cables -- no turns required.
3.1GLI 5-Wire Differential
Technique Sensor
☞
☞
All GLI 5-wire Differential Technique sensors have a built-in
NTC 300 ohm temperature element (except Model 6006P42000 high purity water pH sensor systems which use
Pt1000 ohm RTD) for automatic temperature compensation
and for measuring process temperature.
Wiring Tip!
grounded metal conduit to protect it from moisture,
electrical noise, and mechanical damage.
For installations where the distance between sensor and
analyzer exceeds the sensor cable length, indirectly
connect the sensor to the analyzer using a junction box
and interconnect cable.
NOTE: Do not route the sensor cable in any conduit con-
taining AC or DC power wiring (“electrical noise”
may interfere with the sensor signal).
Route the sensor cable in 1/2-inch,
Model P33 pH/ORP Analyzer (panel-mount 1/4 DIN)Rev. 6-1100
22
PART TW O - INSTALLATIONSECTION 3 - ELECTRICAL CONNECTIONS
S
600'
+$57
7[
5[
56
*1'
287
72P$
287
BB
12
&201&
12
&201&
5(/$<%5(/$<$
7%
1
32:(5
7%
7%
S+253
<(/
6+/'
%/.
:+7
*51
5('$&7
:$51,1*
&20%',))
7(03
7(03
*1'
5()
7%
7%
FIGURE 2-3
Terminal Designations for Analyzers with Letter “A” Prefix
Serial Number (HART switch only provided with HART option)
02'(/3
6(1625
&20%',))
8186('
7(03
7(03
8186('
8186('
8186('
8186('
8186('
*1'
5()
$&7
8186('8186('
5(029(32:(5%()25(6(59,&,1*
7%
7%7%
127
P$
5(/$<$
5(/$<%
86('
287387
287387
12
&20
12
&20
1&
1&
60
+$57
0'
32:(5
1
7%
7%
+2536(1625
8186('
8186('
8186('
<(/
6+/'
%/.
:+7
*51
8186('
8186('
5('
:$51,1*
FIGURE 2-4
Terminal Designations for Analyzers with “No Letter” Prefix
Serial Number (HART switch only provided with HART option)
PART TW O - INSTALLATIONSECTION 3 - ELECTRICAL CONNECTIONS
Refer to Figure 2-5 and connect the sensor (or interconnect) cable wires to Terminals 4 through 8 on TB3 and
Terminal 1 on TB4, matching colors as indicated.
☞
NOTE: For best immunity to electromagnetic interference,
always connect the sensor cable’s shield wire to a
terminal on TB2 instead of to Terminal 5 on TB3.
FIGURE 2-5 Connecting GLI 5-wire Differential Technique Sensor
3.2Conventional
Combination Electrode
The electrode must be within 100 ft. (30 m) of the analyzer
(985 ft./300 m for electrode with preamp). Refer to Figure
2-6 and directly connect the electrode’s coaxial cable to the
analyzer.
Model P33 pH/ORP Analyzer (panel-mount 1/4 DIN)Rev. 6-1100
24
PART TW O - INSTALLATIONSECTION 3 - ELECTRICAL CONNECTIONS
1.Connect the electrode’s reference signal -- braided
shield wire of coaxial cable (black insulated wire for GLI
electrode) -- to “REF” Terminal 10 on TB3.
2.Connect the electrode’s active signal -- center wire of
coaxial cable (clear insulated wire for GLI electrode) -to “ACT” Terminal 1 on TB4.
3.Connect a jumper between “GND” Terminal 9 and
“REF” Terminal 10 on TB3.
4.Connect the electrode’s temperature element (typically
white and red insulated wires for GLI electrode) to
“TEMP” Terminals 2 and 3 on TB3, attaching either wire
to either terminal.
3.3Conventional
Combination Electrode
with Ground Rod
Some applications require that an external ground rod be
used with the combination electrode. The electrode must be
within 100 ft. (30 m) of the analyzer (985 ft./300 m for electrode with preamp). Refer to Figure 2-7 and directly connect
the electrode’s coaxial cable to the analyzer.
Connect the electrode and temperature element wires in the
same way as described in Section 3.2 --
the jumper connecting Terminals 9 and 10 on TB3
except eliminate
. In-
stead, connect the ground rod wire to “GND” Terminal 9.
Connecting Conventional Combination Electrode with Ground Rod
25
PART TW O - INSTALLATIONSECTION 3 - ELECTRICAL CONNECTIONS
3.4Analog Outputs
☞
Two isolated analog outputs (1 and 2) are provided. Each
output can be set to be 0/4-20 mA, and assigned to represent the measured pH/ORP or temperature.
are isolated from the inputs and earth ground, but not
from each other.
PART THREE, Section 3.4.
Wiring Tip!
cable for connecting the analog outputs.
Each 0/4-20 mA output can drive a load of up to 600 ohms.
Output 1:
•
matching polarity as indicated.
Output 2:
•
matching polarity as indicated.
Connect the load to Terminals 4 and 5 on TB5,
Connect the load to Terminals 6 and 7 on TB5,
For output configuration details, see
Use high quality, shielded instrumentation
The outputs
☞
NOTE: When using the HART communication option, a
digital signal is encoded onto the 4-20 mA analog
Output 1 signal. In a HART point-to-point wiring
configuration, Output 1 remains available for normal
use. However, in a HART multi-drop wiring configuration, Output 1 becomes dedicated to that function
and cannot be used. See PART THREE, Section 7
for more HART communication information.
Model P33 pH/ORP Analyzer (panel-mount 1/4 DIN)Rev. 6-1100
26
PART TW O - INSTALLATIONSECTION 3 - ELECTRICAL CONNECTIONS
3.5Relay Outputs
The analyzer is equipped with two electromechanical relays. For relay setup details, see PART THREE, Section 3.5.
CAUTION:
DO NOT EXCEED THE CONTACT RATING FOR EACH
RELAY (5A 115/230 VAC). WHEN SWITCHING
LARGER CURRENTS, USE AN AUXILIARY RELAY
SWITCHED BY THE ANALYZER RELAY TO EXTEND
ANALYZER RELAY LIFE. WHEN USING RELAY
OUTPUTS, MAKE SURE THAT LINE POWER WIRING
CAN ADEQUATELY CONDUCT THE CURRENT DRAW
OF THE SWITCHED LOAD(S).
Two sets of SPDT relay outputs (Relays A and B) are provided at Terminals 8 through 13 on TB5.
are not powered.
The line power used to power the ana-
The relay outputs
lyzer may also be used to power control/alarm devices with
these relay contacts. Refer to Figure 2-8 for a general wiring arrangement. Always check control wiring to insure that
line power will not be shorted by the relay switching action,
and that wiring conforms to local codes.
WARNING:
MAKE SURE LINE POWER IS NOT PRESENT WHILE
CONNECTING WIRES TO TB5 RELAY TERMINALS.
FIGURE 2-8
Connecting Control/Alarm Device(s) to Electromechanical Relay(s)
PART TW O - INSTALLATIONSECTION 3 - ELECTRICAL CONNECTIONS
3.6Line Power
☞
Refer to Figure 2-9, 2-10 or 2-11 and connect line power to
appropriate terminals on TB1 using the standard three-wire
connection arrangement.
form to local codes
Use wiring practices which con-
(example: National Electric Code
Handbook in the U.S.A.).
WARNING:
REMOVE LINE POWER WHILE CONNECTING LINE
POWER WIRES TO THE TB1 TERMINALS. ALSO, USE
ONLY THE STANDARD THREE-WIRE CONNECTION
ARRANGEMENT FOR SINGLE-PHASE LINE POWER
TO PREVENT AN UNSAFE CONDITION, AND TO
ENSURE PROPER ANALYZER OPERATION.
NOTE: In all cases, connect the line power cable ground
wire (usually green) to the “ground symbol” terminal
on TB1.
FIGURE 2-9
Connecting 115 Volt
Single Phase Line Power
(90-130 VAC)
☞
The “115” and “230” voltage circuits are protected with internal, board-mounted slow-blow fuses.
NOTE: For 230 volt split phase line power, be sure to con-
form to local codes with regard to fusing the 115
volt line connected to the “N” terminal.
FIGURE 2-10
Connecting 230 Volt
Single Phase Line Power
(190-260 VAC)
FIGURE 2-11
Connecting 230 Volt
Split Phase Line Power
(190-260 VAC)
Model P33 pH/ORP Analyzer (panel-mount 1/4 DIN)Rev. 6-1100
28
Loading...
+ 63 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.