GLI International P33 Operating Instructions Manual

OPERATING INSTRUCTION MANUAL
Model P33
pH/ORP Analyzer
(Panel-mount 1/4 DIN style;
selectable for pH or ORP measurement)
Worldwide Headquarters and Sales:
9020 West Dean Road Milwaukee, Wisconsin 53224 U.S.A.
In the interest of improving and updating its equipment, GLI reserves the right to alter specifications to equipment at any time.
Rev. 6-1100 Model P33 pH/ORP Analyzer (panel-mount 1/4 DIN)
Phone: Fax: E-mail: Web:
Represented By:
[414] 355-3601 [414] 355-8346 info@gliint.com www.gliint.com
Viridor Instrumentation
A company
1
This operating instruction manual and other GLI instruction manuals are available on GLI’s web site at gliint.com. when viewed using Adobe’s free Acrobat reader. To get this reader, link to Adobe through GLI’s web site or visit Adobe’s web site at adobe.com.
Model P33 pH/ORP Analyzer (panel-mount 1/4 DIN) Rev. 6-1100
2
IMPORTANT SAFETY INFORMATION
This analyzer is compliant with safety standards as outlined in:
FMRC Class Numbers 3600, 3611, and 3810 (U.S.A.)
CSA C22.2 No. 142 and C22.2 No. 213 (Canada)
EN 61010-1 (European Community)
Please read and observe the following:
Line voltage may be present at terminals on TB1 at the back of the analyzer enclosure. This may be
hazardous. Always remove line power before going near this area of the analyzer. The front bezel assembly of the analyzer, however, contains only low voltage and is completely safe to handle.
Wiring or repairs should only be performed by qualified personnel and only to an unpowered analyzer.
Whenever it appears that analyzer safety is questionable, disable the analyzer to ensure against any
unintended operation. For example, an unsafe condition is likely when:
1) The analyzer appears visibly damaged.
2) The analyzer fails to operate properly or provide the intended measurements.
3) The analyzer has been stored for long periods at temperatures above 158°F (70°C).
This analyzer must be installed by specially trained personnel in accordance with relevant local codes
and instructions contained in this operating instruction manual. Observe the analyzer’s technical specifications and input ratings. If one line of the line power mains is not neutral, use a double-pole mains switch to disconnect the analyzer.
HELPFUL IDENTIFIERS
In addition to information on installation and operation, this instruction manual may contain WARNINGS pertaining to user safety, CAUTIONS regarding possible instrument malfunction, and NOTES on important, useful operating guidelines.
WARNING:
A WARNING LOOKS LIKE THIS. IT WARNS YOU OF THE POTENTIAL FOR PERSONAL INJURY.
CAUTION:
A CAUTION LOOKS LIKE THIS. IT ALERTS YOU TO POSSIBLE INSTRUMENT MALFUNCTION OR DAMAGE.
NOTE: A note looks like this. It alerts you to important operating
information.
Rev. 6-1100 Model P33 pH/ORP Analyzer (panel-mount 1/4 DIN)
3
Definition of Equipment Symbols
This symbol
means CAUTION
and alerts the user to possible dan­ger or instrument malfunction. Refer to this manual before proceeding.
This symbol, which appears on the analyzer POWER terminal block (shown in Figure 2-2),
terminal
This symbol
and alerts the user to connect an earth ground to it.
means that there is alternating current present
means that this is a protective ground
and
alerts the user to be careful.
GLI International, Inc. warrants the Model P33 to be free from defects in material or workmanship for a period of 2 years (24 months) from the date of shipment of this product from our facility. A warranty claim will not be honored if defects are not reported within the warranty period, or if GLI International determines that defects or damages are due to normal wear, misapplication, lack of mainte­nance, abuse, improper installation, alteration, or abnormal conditions. GLI Internationals obligation under this warranty shall be limited to, at its option, re­placement or repair of this product. The product must be returned to GLI International, freight prepaid, for examination. The product must be thoroughly cleaned and any process chemicals removed before it will be accepted for re­placement or repair. GLI Internationals liability shall not exceed the cost of the product. Under no circumstances will GLI International be liable for any inciden­tal or consequential damages, whether to person or property. GLI International will not be liable for any other loss, damage or expense of any kind, including loss of profits, resulting from the installation, use, or inability to use this product.
Model P33 pH/ORP Analyzer (panel-mount 1/4 DIN) Rev. 6-1100
WARRANTY
4
CONDENSED OPERATING INSTRUCTIONS
This manual contains details for all operating aspects of the instrument. The following con­densed instructions are provided to assist you in getting the instrument started up and operating as quickly as possible.
measurement operation using a GLI Differential pH sensor.
conventional combination electrode or specific features of the instrument, refer to the appro­priate sections in this manual for instructions.
A. CONNECTING SENSOR/CONFIGURING SENSOR TYPE AND TEMPERATURE ELEMENT
1. After properly mounting the analyzer (PART TWO, Section 2), connect the GLI 5-wire Differential Technique pH sensor, matching wire colors to terminals as indicated:
Sensor Wire Colors Connect to SENSOR Terminal
2. The analyzer is supplied factory-set for use with the GLI 5-wire Differential Technique pH sensor. When using a conventional combination electrode, you must change the sensor type (see PART THREE, Section 3.2, subheading SELECT SENSOR Type”).
These condensed instructions only pertain to basic pH
To measure ORP, or use a
Yellow #4 on TB3
Shield #5 on TB3
Black #6 on TB3
White #7 on TB3
Green #8 on TB3
Red #1 on TB4
3. The analyzer is supplied factory-set for automatic temperature compensation using the NTC 300 ohm temperature element built into all GLI Differential sensors (except Model 6006P4-2000 high purity water pH sensor systems which use Pt1000 ohm RTD). When using a sensor with a different temperature element, or if you want fixed MANUAL temperature compensation, you must change the temperature element type (see PART THREE, Section 3.2, subheading Select TEMP ELEMENT Type”).
B. CONNECTING LINE POWER
Important:
Follow instructions in PART TWO, Section 3.6 to connect line power to the analyzer.
C. CONFIGURING BUFFER TYPE/CALIBRATING THE ANALYZER
The analyzer must be calibrated so that measured values will correspond to actual process values. Before calibrating for the
first time, select the buffer value set that will be used. Then, calibrate using the recommended 2 POINT BUFFER” method which will provide the most accurate pH measurements.
1. The analyzer is supplied factory-set for the common 4.00, 7.00, and 10.00 pH buffers.
When using DIN 19267 standard value buffers you must change the buffer set (see PART THREE, Section 3.2, subheading SELECT BUFFER Set”).
(continued on next page)
Rev. 6-1100 Model P33 pH/ORP Analyzer (panel-mount 1/4 DIN)
5
CONDENSED OPERATING INSTRUCTIONS
C. CALIBRATING THE ANALYZER -- (continued)
NOTE: When using buffers that are not included in either of the analyzer buffer sets,
use only the 2 POINT SAMPLE method for calibration. Refer to that sub­heading in PART THREE, Section 4.2 for instructions.
2. Immerse the sensor in the first buffer (preferably pH 7).
and buffer temperatures to equalize. Depending on their temperature differences, this may take 30 minutes or more.
Important:
Allow the sensor
NOTE: An in-progress calibration can always be aborted by pressing the ESC key.
After the ABORT: YES? screen appears, do one of the following:
Press ENTER key to abort. After CONFIRM ACTIVE? screen appears,
press ENTER key to display the MEASURE screen, and return the ana- log outputs and relays to their active states.
×
Use
×
Ø
or
key to choose ABORT: NO? screen, and press ENTER key
Ø
to continue calibration.
Calibration Tip! If, at any time during calibration, the 2 POINT BUFFER: CONFIRM FAILURE? screen appears, press ENTER key to confirm. Then, use the
×
Ø
or
×
key to select between CAL: EXIT or CAL: REPEAT and do one of the
Ø
following:
With the 2 POINT BUFFER: CAL EXIT? screen selected, press ENTER key.
Then, after the 2 POINT BUFFER: CONFIRM ACTIVE? screen appears, press ENTER key to return the analog outputs and relays to their active states (MEASURE screen appears).
With the 2 POINT BUFFER: CAL REPEAT? screen selected, press ENTER key
to repeat calibration of this point.
3. Press
not showing, use
4. Press
5. Press
6. Press
7. Press
during calibration. (Outputs can also be transferred to preset values or allowed to re­main active.)
Model P33 pH/ORP Analyzer (panel-mount 1/4 DIN) Rev. 6-1100
MENU key
ENTER key
ENTER key
ENTER key
ENTER key
to display a MAIN MENU screen. If the screen is
Ø
Ø
to display
again to display
again to display
again to “hold” the analog outputs and relays at their present states
or
×
×
to display it.
key
(continued on next page)
6
.
.
.
CONDENSED OPERATING INSTRUCTIONS
C. CALIBRATING THE ANALYZER -- (continued)
8. With the sensor in the first buffer and the
ENTER key
to confirm. While the
screen is displayed, the analyzer
screen displayed, press
waits for the pH and temperature signals to stabilize, measures the buffer value, and
automatically calibrates this point. Thereafter, a screen like
appears
for 5 seconds to confirm calibration of this point.
NOTE: Any time the PLEASE WAIT screen is displayed during calibration, you can
manually complete calibration of the point by pressing the ENTER key. How­ever, this is not recommended because the pH and temperature signals may not be fully stabilized, resulting in an inaccurate calibration.
9. After the screen appears, remove the sensor from the first buffer,
rinse it with clean water, and immerse it in the second buffer (typically pH 4).
10. Press
ENTER key
to confirm. While the screen is displayed, the analyzer waits for the pH and temperature signals to stabilize, measures the buffer value, and automatically calibrates this point. Thereafter, a screen like
appears for 5 seconds to confirm calibration of this point.
11. A pH SLOPE XX.X mV/pH” screen appears, indicating a slope value to measure sen- sor performance. The slope should be within 54 to 62 mV/pH for optimal performance.
12. Press
ENTER key
to end calibration (“2 POINT BUFFER: CONFIRM CAL OK?” screen
appears).
13. Re-install the sensor into the process.
14. Press
ENTER key
to display the active measurement reading on the 2 POINT BUFFER: CONFIRM ACTIVE? output status screen. When the reading corresponds to the actual typical process value, press
ENTER key
again to return the analog out-
puts and relays to their active states (MEASURE screen appears).
This completes 2 POINT BUFFER calibration. The analyzer is now ready to measure pH.
D. COMPLETING ANALYZER CONFIGURATION
To further configure the analyzer to your application requirements, use the appropriate CONFIGURE screens to make selections and key in values. Refer to PART THREE, Sec­tion 3 for complete configuration details.
Rev. 6-1100 Model P33 pH/ORP Analyzer (panel-mount 1/4 DIN)
7
Model P33 pH/ORP Analyzer (panel-mount 1/4 DIN) Rev. 6-1100
8
TABLE OF CONTENTS
PART ONE - INTRODUCTION
SECTION 1 GENERAL INFORMATION
1.1 Capability Highlights ........................................................................15-16
1.2 Modular Construction .......................................................................16-17
1.3 Retained Configuration Values .............................................................17
1.4 Analyzer Serial Number ........................................................................17
1.5 EMI/RFI Immunity.................................................................................. 17
SECTION 2 SPECIFICATIONS
....................................................................................18-19
PART TWO - INSTALLATION
SECTION 1 UNPACKING
SECTION 2 MECHANICAL REQUIREMENTS
2.1 Location.................................................................................................20
2.2 Mounting...........................................................................................20-21
SECTION 3 ELECTRICAL CONNECTIONS
3.1 GLI 5-wire Differential Technique Sensor ........................................22-24
3.2 Conventional Combination Electrode ...............................................24-25
3.3 Conventional Combination Electrode with Ground Rod ........................25
3.4 Analog Outputs......................................................................................26
3.5 Relay Outputs........................................................................................27
3.6 Line Power ............................................................................................28
.................................................................................................20
PART THREE - OPERATION
SECTION 1 USER INTERFACE
1.1 Display ..................................................................................................29
1.2 Relay A and B Indicators.......................................................................29
1.3 Keypad .............................................................................................29-30
1.4 MEASURE Screen (normal display mode)............................................ 31
SECTION 2 MENU STRUCTURE
2.1 Displaying Main Branch Selection Screen ............................................32
2.2 Displaying Top-level Menu Screens......................................................33
2.3 Displaying Submenu Screens ...............................................................34
2.4 Adjusting Edit/Selection Screen Values ................................................34
2.5 Entering (Storing) Edit/Selection Screen Values/Choices.....................34
Rev. 6-1100 Model P33 pH/ORP Analyzer (panel-mount 1/4 DIN)
9
TABLE OF CONTENTS (continued)
SECTION 3 ANALYZER CONFIGURATION
3.1 Selecting LANGUAGE to Operate Analyzer.......................................... 35
3.2 Configuring Sensor Characteristics:
SELECT SENSOR Type.............................................................35-36
Select DISPLAY FORMAT...............................................................36
SELECT BUFFER Set for pH Calibration ........................................37
Select PURE H2O COMP (for special applications) ...................37-38
SET ISO POINT (isopotential for special Differential pH sensor) ........38
SET FILTER Time.......................................................................38-39
Select PULSE SUPPRESS (on/off)..................................................39
ENTER NOTE (top line of MEASURE screen)............................39-40
Select TEMP ELEMENT Type ....................................................40-41
3.3 SET °C OR °F (temperature display format) .........................................41
3.4 Configuring Analog Outputs (1 and 2):
SET PARAMETER (representation) ................................................42
SET 0/4 and 20 mA VALUES......................................................42-43
SET TRANSFER Value (mA)...........................................................43
SET FILTER Time............................................................................ 44
Select SCALE 0 mA/4 mA (low endpoint) ........................................ 44
3.5 Configuring Relays (A and B):
SET PARAMETER (representation) ................................................45
SET FUNCTION Mode (alarm, control or status)........................45-46
SET TRANSFER Mode (relay on or off) ..........................................46
ACTIVATION (configuration values) ...........................................47-48
3.6 SET PASSCODE (feature enabled or disabled) ...................................49
3.7 Configuration Setting Summary (ranges/choices and defaults) .......50-51
SECTION 4 ANALYZER CALIBRATION
4.1 Important Information:
Calibrate Periodically .......................................................................52
Temperature-corrected pH Measurement........................................ 52
4.2 pH Calibration:
2 POINT BUFFER Method..........................................................53-55
1 POINT BUFFER Method..........................................................55-57
2 POINT SAMPLE Method..........................................................57-59
1 POINT SAMPLE Method..........................................................60-61
4.3 ORP Calibration ...............................................................................62-63
4.4 Analog Outputs (1 and 2) Calibration...............................................63-64
Model P33 pH/ORP Analyzer (panel-mount 1/4 DIN) Rev. 6-1100
10
TABLE OF CONTENTS (continued)
SECTION 5 TEST/MAINTENANCE
5.1 STATUS Checking (analyzer, sensor, and relays)...........................65-66
5.2 HOLD OUTPUTS .................................................................................. 67
5.3 OVERFEED RESET (relay timers)........................................................ 67
5.4 OUTPUT (1 and 2) Analog Test Signals ...............................................68
5.5 RELAY (A and B) Operating Test.....................................................68-69
5.6 ALARM LEDS Operating Test............................................................... 69
5.7 EPROM VERSION Checking ................................................................69
5.8 SELECT SIM Measurement .................................................................. 70
5.9 SIM Setting............................................................................................70
5.10 RESET CONFIGURE Values to Factory Defaults................................. 71
5.11 RESET CALIBRATE Values to Factory Defaults ..................................71
SECTION 6 RELAY OVERFEED TIMER FEATURE
6.1 Why Use an Overfeed Timer................................................................. 72
6.2 Configuring Relay Overfeed Timers......................................................72
6.3 Overfeed Timer “Timeout” Operation .................................................... 72
6.4 Resetting Overfeed Timers ...................................................................72
6.5 Interactions with Other Analyzer Functions......................................72-73
SECTION 7 HART OPTION
7.1 Introduction ...........................................................................................74
7.2 Analyzer Operating Modes for HART Network.................................75-76
7.3 SINGLE MODE (Point-to-Point) Wiring Arrangement ...........................76
7.4 MULTI-DROP Wiring Arrangement....................................................... 77
7.5 HART Preferences Setup:
Changing Polling Address ...............................................................78
Viewing Number of Required Preambles ....................................78-79
7.6 Device Preferences Setup:
Viewing Final Assembly Number ..................................................... 79
Viewing Model Number...............................................................79-80
Viewing Manufacturer ......................................................................80
Assigning a Tag ...............................................................................80
Assigning a Descriptor ..................................................................... 81
Assigning a Message....................................................................... 81
Assigning User-defined Date ......................................................81-82
Viewing Identification (ID) ................................................................82
Viewing Revisions............................................................................ 82
7.7 Master Reset Function........................................................................ 83
7.8 “Refresh” Function ................................................................................83
7.9 Protocol Command Set for PC Programming........................................ 83
Rev. 6-1100 Model P33 pH/ORP Analyzer (panel-mount 1/4 DIN)
11
TABLE OF CONTENTS (continued)
PART FOUR - SERVICE AND MAINTENANCE
SECTION 1 GENERAL INFORMATION
1.1 Inspecting Sensor Cable .......................................................................84
1.2 Replacing Fuse(s) ............................................................................84-85
1.3 Replacing Relays ..................................................................................85
SECTION 2 PRESERVING MEASUREMENT ACCURACY
2.1 Keeping Sensor Clean ..........................................................................86
2.2 Keeping Analyzer Calibrated.................................................................86
2.3 Avoiding Electrical Interference.............................................................86
SECTION 3 TROUBLESHOOTING
3.1 Ground Loops:
Determining if Ground Loop Exists ..................................................87
Finding Source of Ground Loop.......................................................88
3.2 Isolating Measuring System Problem:
Checking Electrical Connections ..................................................... 88
Verifying Sensor Operation..............................................................88
Verifying Analyzer Operation ......................................................88-90
Verifying Interconnect Cable Integrity ..............................................90
SECTION 4 ANALYZER REPAIR/RETURN
4.1 Customer Assistance.............................................................................91
4.2 Repair/Return Policy .............................................................................91
Model P33 pH/ORP Analyzer (panel-mount 1/4 DIN) Rev. 6-1100
12
TABLE OF CONTENTS (continued)
ILLUSTRATIONS
Figure 1-1
Figure 2-1
Figure 2-2
Figure 2-3
Figure 2-4
Figure 2-5
Figure 2-6
Figure 2-7
Figure 2-8
Figure 2-9
Figure 2-10
Figure 2-11
Figure 3-1
Figure 3-2
EMI/RFI Immunity Diagram..............................................................................................17
Enclosure Dimension Details for Analyzers with Letter “A” Prefix Serial Number..............21
Enclosure Dimension Details for Analyzers with “No Letter” Prefix Serial Number............21
Terminal Designations for Analyzers with Letter “A” Prefix Serial Number
(HART switch only provided with HART option) ...........................................................23
Terminal Designations for Analyzers with “No Letter” Prefix Serial Number
(HART switch only provided with HART option) ...........................................................23
Connecting GLI 5-wire Differential Technique Sensor ......................................................24
Connecting Conventional Combination Electrode.............................................................24
Connecting Conventional Combination Electrode with Ground Rod..................................25
Connecting Control/Alarm Device(s) to Electromechanical Relay(s).................................27
Connecting 115 Volt Single Phase Line Power (90-130 VAC)...........................................28
Connecting 230 Volt Single Phase Line Power (190-260 VAC).........................................28
Connecting 230 Volt Split Phase Line Power (190-260 VAC)............................................28
Analyzer Keypad..............................................................................................................30
Location of SINGLE MODE/MULTI-DROP Switch (HART-equipped analyzers only) ........76
Figure 3-3
Figure 3-4
Figure 4-1
Table A
Table B
Table C
HART SINGLE MODE (Point-to-Point) Wiring Arrangement (for single analyzer).............76
HART MULTI-DROP Wiring Arrangement (for multiple analyzer network)........................77
Removing Analyzer Bezel (only analyzers with letter “A” prefix serial number).................84
TABLES
Relay Configuration Settings............................................................................................47
Analyzer Configuration Settings (Ranges/Choices and Defaults) .................................50-51
Relay Overfeed Timer Interactions with Other Analyzer Functions...................................73
Rev. 6-1100 Model P33 pH/ORP Analyzer (panel-mount 1/4 DIN)
13
Model P33 pH/ORP Analyzer (panel-mount 1/4 DIN) Rev. 6-1100
14
PART ONE - INTRODUCTION SECTION 1 - GENERAL INFORMATION
PART ONE - INTRODUCTION
ECTION
1.1 Capability Highlights
Sensor Input
S
GENERAL INFORMATION
The analyzer can be used with any GLI 5-wire Differential Technique pH or ORP sensor, or any conventional combi­nation electrode. The analyzer will accept an NTC 300 ohm thermistor, Pt 1000 RTD or Pt 100 RTD temperature com­pensator element. Typically, one of these elements is built into the pH or ORP sensor.
1
MEASURE Screen
Passcode-protected
Access
Calibration Methods
The MEASURE screen (normal display mode) can provide different readouts of measured data. With the MEASURE screen displayed, press
Measured pH (or ORP, if selected).
Measured temperature (°C or °F).
Analog Output 1 and 2 values (mA).
Measured pH (or ORP) and temperature.
For security, you can enable a passcode feature to restrict access to configuration and calibration settings to author­ized personnel only. See PART THREE, Section 3.6 for details.
Four methods are available to calibrate the analyzer for pH. See PART THREE, Section 4.2 for details. For ORP cali­bration, refer to Section 4.3. Each analog output mA value can also be calibrated (Section 4.4).
ÕÕ and
ÖÖ key to show:
Analog Outputs
Rev. 6-1100 Model P33 pH/ORP Analyzer (panel-mount 1/4 DIN)
The analyzer provides two isolated analog outputs (1 and
2). Each output can be set to be 0-20 mA or 4-20 mA, and assigned to represent one of these measurements:
Measured pH (or ORP).
Measured temperature.
Parameter values can be entered to define the endpoints at which the minimum and maximum analog output values are desired (range expand). For analog output setup details, refer to PART THREE, Section 3.4.
15
PART ONE - INTRODUCTION SECTION 1 - GENERAL INFORMATION
During calibration, both analog outputs can be selected to:
Hold their present values (HOLD OUTPUTS).
Transfer to preset values to operate control elements
by an amount corresponding to those values (XFER OUTPUTS).
Remain active to respond to the measured value
(ACTIVE OUTPUTS).
Relays
The analyzer has two electromechanical relays with SPDT contacts. Each relay can be set to function as a CONTROL, ALARM or STATUS relay. CONTROL and ALARM relays can be assigned to be driven by one of these:
Measured pH (or ORP).
Measured temperature.
NOTE: When a relay is set to function as a STATUS relay,
it is no longer configurable. Instead, it becomes a dedicated system diagnostic-only alarm relay that automatically energizes when the “WARNING CHECK STATUS message flashes on the MEASURE screen. This occurs when the analyzer detects a “fail” diagnostic condition. See PART THREE, Section 5.1 for more details.
Except for STATUS relays, during calibration the relay on/off states are affected in the same way as the analog outputs by the (HOLD/XFER/ACTIVE) OUTPUTS screen selection. These relays are also held at their present on/off states, transferred to desired preset on/off states, or remain active to respond to measured values.
1.2 Modular Construction
Model P33 pH/ORP Analyzer (panel-mount 1/4 DIN) Rev. 6-1100
See PART THREE, Section 3.5 for relay setup details.
The modular construction of the analyzer provides electrical safety. The front panel keypad assembly uses voltages no greater than 24 VDC, and is completely safe to handle.
16
PART ONE - INTRODUCTION SECTION 1 - GENERAL INFORMATION
Line power must be connected to specifically designated terminals on TB1.
WARNING:
REMOVE LINE POWER BEFORE NEARING THIS AREA TO PREVENT ELECTRICAL SHOCK.
1.3 Retained Configuration Values
1.4 Analyzer Serial Number
1.5 EMI/RFI Immunity
All user-entered configuration values are retained indefi­nitely, even if power is lost or turned off. The non-volatile analyzer memory does not require battery backup.
A label with the analyzer model number, serial number, build date, and other items is located on top of the enclo­sure.
The analyzer is designed to provide protection from most normally encountered electromagnetic interference. This protection exceeds U.S. standards and meets European IEC 801-series testing for electromagnetic and radio fre­quency emissions and susceptibility. Refer to Figure 1-1 and the specifications in Section 2.1 for more information.
Rev. 6-1100 Model P33 pH/ORP Analyzer (panel-mount 1/4 DIN)
FIGURE 1-1 EMI/RFI Immunity Diagram
17
PART ONE - INTRODUCTION SECTION 2 - SPECIFICATIONS
ECTION
S
2
SPECIFICATIONS
2.1 Operational
Display....................................... Two-line by 16 character backlit LCD
NOTE:
Ambient Conditions:
Relays: Types/Outputs .................Two electromechanical relays; SPDT (Form C)
The measured pH (or ORP) or temperature can be shown separately, or both measurements can be displayed together.
Measurement Selectable Ranges
pH......................................... -2.0 to 14.0 pH or -2.00 to 14.00 pH
ORP...................................... -2100 to +2100 mV (fixed)
Temperature ......................... -20.0 to +200.0°C or -4.0 to +392.0°F
Analog Outputs (1 and 2) ...... 0.00-20.00 mA or 4.00-20.00 mA
Operation.............................. -4 to +140°F (-20 to +60°C); 0-95% relative
humidity, non-condensing
Storage................................. -22 to +158°F (-30 to +70°C); 0-95% relative
humidity, non-condensing
contacts; U.L. rated 5A 115/230 VAC, 5A @ 30 VDC resistive
Operational Mode ............Each relay (A and B) can be assigned to be driven
by the measured pH (or ORP) or tem perature
Function Modes:
Control.................... Settings for high/low phasing, setpoint, dead-
band, overfeed timer, off delay, and on delay
Alarm..........................Settings for low alarm point, low alarm point
deadband, high alarm point, high alarm point deadband, off delay, and on delay
Status.........................Not configurable; relay only activates when a
sensor or analyzer diagnostic WARNING condition exists
Indicators.........................Relay A and B LEDs indicate respective relay status
Temperature Compensation ....... Automatic from 14.0 to 230.0°F (-10.0 to
+110.0°C), with selection for NTC 300 ohm thermistor, Pt 1000 ohm RTD or Pt 100 ohm RTD temperature element, or manually fixed at a user-entered temperature; additional selectable temperature correction factors (ammonia or morpholine) for pure water automatic compensation from 0.0-50.0°C
Sensor-to-Analyzer Distance:
GLI 5-wire Differential
Technique Sensor.............. 3000 ft. (914 m) maximum
Conventional Combination
Electrode with preamp ....... 985 ft. (300 m) maximum
Conventional Combination
Electrode without preamp .. 100 ft. (30 m) maximum with electrode cable
capacitance of less than 30 pF per foot
Power Requirements .................. 90-130 VAC, 50/60 Hz. (10 VA max.) or
190-260 VAC, 50/60 Hz. (10 VA max.)
Calibration Methods:
2 POINT BUFFER................... Automatic calibration and buffer recognition
(for pH only) using two buffers from selected buffer set*.
NOTE:
When using buffers that are not included in either of the analyzer buffer sets, use only the 2 POINT SAMPLE method for calibration.
*Buffer Sets: 4.00, 7.00, and 10.00 pH; or
DIN 19267 standard (1.09, 4.65, 6.79, 9.23, and 12.75 pH)
Model P33 pH/ORP Analyzer (panel-mount 1/4 DIN) Rev. 6-1100
18
PART ONE - INTRODUCTION SECTION 2 - SPECIFICATIONS
1 POINT BUFFER .................. Automatic calibration and buffer recognition
(for pH only) using one buffer from selected buffer set*.
NOTE:
When using a buffer that is not included in either of the analyzer buffer sets, use only the 1 POINT SAMPLE method for calibration.
2 POINT SAMPLE .................. Enter known values of two samples
(for pH only) (determined by laboratory analysis or
comparison reading) or two pH buffers
1 POINT SAMPLE .................. Enter known value of one sample (determined
(for pH or ORP) by laboratory analysis or comparison
reading), one pH buffer, or one reference solution (for ORP measurement)
Analog Outputs .......................... Two isolated 0/4-20 mA outputs; each with
0.004 mA (12-bit) resolution and capability to drive up to 600 ohm loads
NOTE:
Each output can be assigned to represent the measured pH (or ORP) or temperature. Parameter values can be entered to define the end­points at which the minimum and maximum mA output values are desired (range expand). During calibration, both outputs can be se­lected to hold their present values, transfer to preset values to operate control elements by an amount corresponding to those val­ues, or remain active to respond to the measured value.
Communication: RS-232 ........... Enables configuration and retrieval of measured
data for one analyzer using IBM-compatible PC and optional GLI software tool kit
HART.............. Enables configuration and retrieval of measured
data for multiple analyzers over a communi­cation link using appropriate hand-held terminal or data system with HART software
2.2 Analyzer Performance
(Electrical, Analog Outputs)
2.3 Mechanical
Memory Backup (non-volatile) .... All user settings are retained indefinitely in
memory (EEPROM)
EMI/RFI Conformance................ Exceeds U.S. and meets European standards
for conducted and radiated emissions and immunity; certified CE compliant for appli­cations as specified by EN 50081-2 for emissions and EN 50082-2 for immunity
Electrical Certifications:
General Purpose (pending)..... UL, C-UL, FM, and CENELEC
Division 2 (pending)................ UL, C-UL, and FM: Groups A, B, C, D, F, and G
Zone 2 (pending) .................... CENELEC: Group IIC
Accuracy.................................... 0.1% of span
Stability...................................... 0.05% of span per 24 hours, non-cumulative
Repeatability .............................. 0.1% of span or better
Temperature Drift....................... Zero and Span: less than 0.03% of span per °C
Enclosure................................... Polycarbonate with NEMA 4X front panel;
general purpose; two brackets supplied for panel mounting
Mounting Configuration ..............Panel mounting
Rev. 6-1100 Model P33 pH/ORP Analyzer (panel-mount 1/4 DIN)
Net Weight................................. 1.7 lbs. (0.8 kg) approximately
19
PART TW O - INSTALLATION SECTION 1 - UNPACKING
PART TWO - INSTALLATION
ECTION
S
UNPACKING
After unpacking, it is recommended to save the shipping carton and packing materials in case the instrument must be stored or re-shipped. Inspect the equipment and packing materials for signs of shipping damage. If there is any evi­dence of damage, notify the transit carrier immediately.
ECTION
S
MECHANICAL REQUIREMENTS
1
2
2.1 Location
1. It is recommended to locate the analyzer as close as possible to the installed sensor. Depending on the sen­sor type, the maximum allowable distance between the sensor and analyzer is:
GLI 5-wire
Differential
Technique Sensor
3000 feet (914 m) 985 feet (300 m) 100 feet (30 m)
Recommendation:
Conventional
Combination Elec-
with
trode
Preamp
Conventional
Combination Electrode
without
Preamp
Directly connect the sensor to the analyzer to eliminate potential problems caused by wet environments when a junction box is used.
2. Mount the analyzer in a location that is:
Clean and dry where there is little or no vibration.
Protected from corrosive fluids.
Within ambient temperature limits (-4 to +140°F or
-20 to +60°C).
CAUTION:
2.2 Mounting
Model P33 pH/ORP Analyzer (panel-mount 1/4 DIN) Rev. 6-1100
EXPOSING THE ANALYZER TO DIRECT SUNLIGHT MAY INCREASE THE OPERATING TEMPERATURE ABOVE ITS SPECIFIED LIMIT, AND DECREASE DISPLAY VISIBILITY.
Figure 2-1 or 2-2 illustrates the analyzer enclosure dimen­sions and panel mounting details. Using the two supplied brackets, attach them to the analyzer case as shown to panel mount the analyzer.
20
PART TW O - INSTALLATION SECTION 2 - MECHANICAL REQUIREMENTS
FIGURE 2-1 Enclosure Dimension Details for Analyzers with Letter “A” Prefix Serial Number
FIGURE 2-2 Enclosure Dimension Details for Analyzers with “No Letter” Prefix Serial Number
Rev. 6-1100 Model P33 pH/ORP Analyzer (panel-mount 1/4 DIN)
21
PART TW O - INSTALLATION SECTION 3 - ELECTRICAL CONNECTIONS
ECTION
S
ELECTRICAL CONNECTIONS
Figure 2-3 or 2-4 on the next page shows the terminal ar­rangement and designations on the back of the analyzer.
3
NOTE: For easier wiring, terminal blocks can be unplugged
from their mating connectors. All terminals are suit­able for single wires up to 14 AWG (2.5 mm2).
Wiring Tip!
electromagnetic compatibility requirements, follow these general wiring guidelines:
1. Keep all cable shields as short as possible and con­nect them to earth ground.
2. Use Steward ferrite 28 B0590-000 or equivalent on:
3. In harsh conducted RF conditions, connect the earth ground of the analyzer (Terminal 4 on TB1) to a lo­cal, known earth ground source.
To comply with European Community (CE)
Mains (line power) cable -- no turns required. Sensor cable -- one turn required. mA analog output cables -- two turns required. Relay cables -- no turns required.
3.1 GLI 5-Wire Differential Technique Sensor
All GLI 5-wire Differential Technique sensors have a built-in NTC 300 ohm temperature element (except Model 6006P4­2000 high purity water pH sensor systems which use Pt1000 ohm RTD) for automatic temperature compensation and for measuring process temperature.
Wiring Tip!
grounded metal conduit to protect it from moisture, electrical noise, and mechanical damage.
For installations where the distance between sensor and analyzer exceeds the sensor cable length, indirectly connect the sensor to the analyzer using a junction box and interconnect cable.
NOTE: Do not route the sensor cable in any conduit con-
taining AC or DC power wiring (electrical noise may interfere with the sensor signal).
Route the sensor cable in 1/2-inch,
Model P33 pH/ORP Analyzer (panel-mount 1/4 DIN) Rev. 6-1100
22
PART TW O - INSTALLATION SECTION 3 - ELECTRICAL CONNECTIONS
S
60 0'
+$57
7[
5[
56
*1'
287
 72  P$
287
BB
12
&201&
12
&201&
5(/$< % 5(/$< $
7%



1
32:(5
7%
7%
S+253
<(/
6+/'
%/.
:+7
*51
5(' $&7
:$51,1*
&20%',))
7(03
7(03
*1'
5()
7%

7%
FIGURE 2-3
Terminal Designations for Analyzers with Letter “A” Prefix
Serial Number (HART switch only provided with HART option)
02'(/ 3
6(1625
&20%',))
8186('
7(03
7(03
8186('
8186('
8186('
8186('
8186('
*1'
5()
$&7
8186('8186('
5(029(32:(5%()25(6(59,&,1*
7%

7% 7%
127
 P$
5(/$< $
5(/$< %
86('
287387
287387
12
&20
12
&20
1&
1&
60
+$57
0'

32:(5


1
7%
7%
+253 6(1625
8186('
8186('
8186('
<(/
6+/'
%/.
:+7
*51
8186('
8186('
5('
:$51,1*
FIGURE 2-4
Terminal Designations for Analyzers with “No Letter” Prefix
Serial Number (HART switch only provided with HART option)
Rev. 6-1100 Model P33 pH/ORP Analyzer (panel-mount 1/4 DIN)
23
PART TW O - INSTALLATION SECTION 3 - ELECTRICAL CONNECTIONS
Refer to Figure 2-5 and connect the sensor (or intercon­nect) cable wires to Terminals 4 through 8 on TB3 and Terminal 1 on TB4, matching colors as indicated.
NOTE: For best immunity to electromagnetic interference,
always connect the sensor cables shield wire to a terminal on TB2 instead of to Terminal 5 on TB3.
FIGURE 2-5 Connecting GLI 5-wire Differential Technique Sensor
3.2 Conventional Combination Electrode
The electrode must be within 100 ft. (30 m) of the analyzer (985 ft./300 m for electrode with preamp). Refer to Figure 2-6 and directly connect the electrodes coaxial cable to the analyzer.
FIGURE 2-6 Connecting Conventional Combination Electrode
Model P33 pH/ORP Analyzer (panel-mount 1/4 DIN) Rev. 6-1100
24
PART TW O - INSTALLATION SECTION 3 - ELECTRICAL CONNECTIONS
1. Connect the electrodes reference signal -- braided shield wire of coaxial cable (black insulated wire for GLI electrode) -- to “REF” Terminal 10 on TB3.
2. Connect the electrodes active signal -- center wire of coaxial cable (clear insulated wire for GLI electrode) -­to ACT Terminal 1 on TB4.
3. Connect a jumper between “GND” Terminal 9 and REF Terminal 10 on TB3.
4. Connect the electrodes temperature element (typically white and red insulated wires for GLI electrode) to TEMP Terminals 2 and 3 on TB3, attaching either wire to either terminal.
3.3 Conventional Combination Electrode with Ground Rod
Some applications require that an external ground rod be used with the combination electrode. The electrode must be within 100 ft. (30 m) of the analyzer (985 ft./300 m for elec­trode with preamp). Refer to Figure 2-7 and directly connect the electrodes coaxial cable to the analyzer.
Connect the electrode and temperature element wires in the same way as described in Section 3.2 --
the jumper connecting Terminals 9 and 10 on TB3
except eliminate
. In-
stead, connect the ground rod wire to “GND” Terminal 9.
Rev. 6-1100 Model P33 pH/ORP Analyzer (panel-mount 1/4 DIN)
FIGURE 2-7
Connecting Conventional Combination Electrode with Ground Rod
25
PART TW O - INSTALLATION SECTION 3 - ELECTRICAL CONNECTIONS
3.4 Analog Outputs
Two isolated analog outputs (1 and 2) are provided. Each output can be set to be 0/4-20 mA, and assigned to repre­sent the measured pH/ORP or temperature.
are isolated from the inputs and earth ground, but not from each other.
PART THREE, Section 3.4.
Wiring Tip!
cable for connecting the analog outputs.
Each 0/4-20 mA output can drive a load of up to 600 ohms.
Output 1:
matching polarity as indicated.
Output 2:
matching polarity as indicated.
Connect the load to Terminals 4 and 5 on TB5,
Connect the load to Terminals 6 and 7 on TB5,
For output configuration details, see
Use high quality, shielded instrumentation
The outputs
NOTE: When using the HART communication option, a
digital signal is encoded onto the 4-20 mA analog Output 1 signal. In a HART point-to-point wiring configuration, Output 1 remains available for normal use. However, in a HART multi-drop wiring configu­ration, Output 1 becomes dedicated to that function and cannot be used. See PART THREE, Section 7 for more HART communication information.
Model P33 pH/ORP Analyzer (panel-mount 1/4 DIN) Rev. 6-1100
26
PART TW O - INSTALLATION SECTION 3 - ELECTRICAL CONNECTIONS
3.5 Relay Outputs
The analyzer is equipped with two electromechanical re­lays. For relay setup details, see PART THREE, Section 3.5.
CAUTION:
DO NOT EXCEED THE CONTACT RATING FOR EACH RELAY (5A 115/230 VAC). WHEN SWITCHING LARGER CURRENTS, USE AN AUXILIARY RELAY SWITCHED BY THE ANALYZER RELAY TO EXTEND ANALYZER RELAY LIFE. WHEN USING RELAY OUTPUTS, MAKE SURE THAT LINE POWER WIRING CAN ADEQUATELY CONDUCT THE CURRENT DRAW OF THE SWITCHED LOAD(S).
Two sets of SPDT relay outputs (Relays A and B) are pro­vided at Terminals 8 through 13 on TB5.
are not powered.
The line power used to power the ana-
The relay outputs
lyzer may also be used to power control/alarm devices with these relay contacts. Refer to Figure 2-8 for a general wir­ing arrangement. Always check control wiring to insure that line power will not be shorted by the relay switching action, and that wiring conforms to local codes.
WARNING:
MAKE SURE LINE POWER IS NOT PRESENT WHILE CONNECTING WIRES TO TB5 RELAY TERMINALS.
FIGURE 2-8
Connecting Control/Alarm Device(s) to Electromechanical Relay(s)
Rev. 6-1100 Model P33 pH/ORP Analyzer (panel-mount 1/4 DIN)
27
PART TW O - INSTALLATION SECTION 3 - ELECTRICAL CONNECTIONS
3.6 Line Power
Refer to Figure 2-9, 2-10 or 2-11 and connect line power to appropriate terminals on TB1 using the standard three-wire connection arrangement.
form to local codes
Use wiring practices which con-
(example: National Electric Code
Handbook in the U.S.A.).
WARNING:
REMOVE LINE POWER WHILE CONNECTING LINE POWER WIRES TO THE TB1 TERMINALS. ALSO, USE ONLY THE STANDARD THREE-WIRE CONNECTION ARRANGEMENT FOR SINGLE-PHASE LINE POWER TO PREVENT AN UNSAFE CONDITION, AND TO ENSURE PROPER ANALYZER OPERATION.
NOTE: In all cases, connect the line power cable ground
wire (usually green) to the ground symbol” terminal on TB1.
FIGURE 2-9
Connecting 115 Volt
Single Phase Line Power
(90-130 VAC)
The 115 and 230 voltage circuits are protected with in­ternal, board-mounted slow-blow fuses.
NOTE: For 230 volt split phase line power, be sure to con-
form to local codes with regard to fusing the 115 volt line connected to the “N” terminal.
FIGURE 2-10
Connecting 230 Volt
Single Phase Line Power
(190-260 VAC)
FIGURE 2-11
Connecting 230 Volt
Split Phase Line Power
(190-260 VAC)
Model P33 pH/ORP Analyzer (panel-mount 1/4 DIN) Rev. 6-1100
28
Loading...
+ 63 hidden pages