This measurement system is compliant with safety standards as outlined in:
FMRC Class Numbers 3600, 3611, and 3810 (U.S.A.)
CSA C22.2 No. 142 and C22.2 No. 213 (Canada)
EN 61010-1 (European Community)
Please read and observe the following:
•
Opening the analyzer door exposes you to line power voltage, if present, at terminals on TB2 and TB3
inside the enclosure. This may be hazardous. Always remove line power before entering this area in
the analyzer. However, the analyzer door assembly and the sensor contain only low voltages and are
completely safe to handle.
•
Wiring or repairs should only be performed by qualified personnel and only to an unpowered analyzer.
•
Whenever it appears that analyzer safety is questionable, disable the analyzer to ensure against any
unintended operation. For example, an unsafe condition is likely when:
1) The analyzer appears visibly damaged.
2) The analyzer fails to operate properly or provide the intended measurements.
3) The analyzer has been stored for long periods at temperatures above 158°F (70°C).
•
This measurement system must be installed by specially trained personnel in accordance with relevant
local codes and instructions contained in this operating instruction manual. Observe all technical
specifications of the system. If one line of the line power mains is not neutral, use a double-pole mains
switch to disconnect the analyzer.
HELPFUL IDENTIFIERS
In addition to information on installation and operation, this instruction manual may contain
WARNINGS pertaining to user safety, CAUTIONS regarding possible instrument malfunction, and
NOTES on important, useful operating guidelines.
WARNING:
A WARNING LOOKS LIKE THIS. IT WARNS YOU OF THE POTENTIAL
FOR PERSONAL INJURY.
CAUTION:
A CAUTION LOOKS LIKE THIS. IT ALERTS YOU TO POSSIBLE
INSTRUMENT MALFUNCTION OR DAMAGE.
☞
NOTE:A note looks like this. It alerts you to important operating
information.
Rev. 2-300Accu4™ Low-range Turbidimeter System
3
Definition of Equipment Symbols
This symbol
means CAUTION
and alerts you to possible danger or
instrument malfunction. Refer to this manual before proceeding.
This symbol
means that this is a protective ground terminal
alerts you to connect an earth ground to it.
This symbol
means that there is alternating current present
alerts you to be careful.
and
and
WARRANTY
GLI International, Inc. warrants the Accu4™ Low-range Turbidimeter System to be free
from defects in material or workmanship for a period of 2 years (24 months) from the
date this product was shipped from our facility. A warranty claim will not be honored if
defects are not reported within the warranty period, or if GLI International determines that
defects or damages are due to normal wear, misapplication, lack of maintenance, abuse,
improper installation, alteration, or abnormal conditions. GLI International’s obligation
under this warranty shall be limited to, at its option, replacement or repair of this product.
The product must be returned to GLI International, freight prepaid, for examination. The
product must be thoroughly cleaned and any process chemicals removed before it will be
accepted for replacement or repair. GLI International’s liability shall not exceed the cost
of the product. Under no circumstances will GLI International be liable for any incidental
or consequential damages, whether to person or property. GLI International will not be liable for any other loss, damage or expense of any kind, including loss of profits,
resulting from the installation, use, or inability to use this product.
Accu4™ Low-range Turbidimeter SystemRev. 2-300
4
CONDENSED OPERATING INSTRUCTIONS
This manual contains details for all operating aspects of the Accu4™ Low-range Turbidimeter System.
The following condensed instructions are provided to assist you in getting the system started up and operating as quickly as possible.
measurement operation.
To use specific analyzer features, refer to the appropriate sections in this
manual for instructions.
1. CONNECTING SENSOR
After the T53 analyzer is properly mounted (Part Two, Section 2.5), connect the GLI Model 8320 lowrange turbidity sensor, matching wire colors to terminals as indicated:
These condensed instructions only pertain to basic turbidity
Sensor Wire
Colors
Inner ShieldGrounding strip lugTerminal #11 on TB1
RedTerminal #12 on TB1Terminal #12 on TB1
VioletTerminal #13 on TB1Terminal #13 on TB1
GreenTerminal #14 on TB1Terminal #14 on TB1
WhiteTerminal #15 on TB1Terminal #15 on TB1
YellowTerminal #16 on TB1Terminal #16 on TB1
Inner ShieldGrounding strip lugTerminal #17 on TB1
BrownTerminal #18 on TB1Terminal #18 on TB1
GrayTerminal #19 on TB1Terminal #19 on TB1
BlueTerminal #20 on TB1Terminal #20 on TB1
BlackTerminal #21 on TB1Terminal #21 on TB1
OrangeTerminal #22 on TB1Terminal #22 on TB1
Analyzers with
“B” Prefix Serial No.
Analyzers with
No Letter Prefix Serial No.
2. CONNECTING LINE POWER
Important:
Follow the instructions in Part Two, Section 3.5 to connect line power to the analyzer.
3. ADJUSTING DISPLAY CONTRAST
Ambient lighting conditions may make it necessary to adjust display contrast to improve visibility.
With the MEASURE screen displayed, press and hold
×
Ø
×
or
until attaining the desired contrast.
key
Ø
ENTER key
the
and simultaneously press the
4. CALIBRATING SYSTEM
Always initially
calibrate
the Accu4™ Low-range Turbidimeter System to ensure accurate measurement. Any time thereafter, you can conveniently check system calibration using an optional
accessory Cal-CubeTM assembly.
•
When you need to calibrate with a
method. GLI only guarantees measurement accuracy when a 40 NTU formazin suspension is used
as the primary standard.
•
When calibration need not conform to USEPA requirements
method (requires optional GLI CAL-CUBE™ assembly) or the “SAMPLE CAL” method.
Refer to Part Three, Section 5 for instructions on all calibration methods.
5. COMPLETING ANALYZER CONFIGURATION
To further configure the analyzer to your application requirements, use the appropriate CONFIGURE
screens to make selections and “key in” values. Refer to Part Three, Section 4 for complete configuration details.
Analyzer Configuration Settings (Ranges/Choices and Defaults) ..........................................49
Relay Overfeed Timer Interactions with Other Analyzer Functions........................................70
System Diagnostic Condition Status Screens/Meanings........................................................88
TABLES
11
Accu4™ Low-range Turbidimeter SystemRev. 2-300
12
PART ONE - INTRODUCTIONSECTION 1 - GENERAL INFORMATION
PART ONE - INTRODUCTION
ECTION
1.1Capability Highlights
High Accuracy with
Minimal Maintenance
S
GENERAL INFORMATION
The Accu4™ Low-range Turbidimeter System employs a
technologically-advanced sensor to provide high measurement accuracy and stability, while reducing maintenance
requirements. It is designed to meet the International Standards For Measurement of Turbidity [ISO 7027-1984(E)]
and USEPA-approved GLI Method 2. This system has an
auto-ranging measuring scale, enabling continuous monitoring over a 0.000-100.0 NTU range with automatic
decimal point positioning.
dimeter System automatically provides increasing
display resolution as measured turbidity decreases.
further ensure high measuring accuracy, the Model 8320
sensor eliminates air or gas bubbles in the sample with its
built-in bubble trap. The system is ideal for critical monitoring and controlling applications including potable water,
filtered water, and final product clarity.
1
The Accu4™ Low-range Turbi-
To
MEASURE Screen
Passcode-protected
Access
Calibration Methods
With the display in the normal MEASURE screen mode, the
measured turbidity is always shown on the main middle line.
The bottom auxiliary display line, shown in reverse video,
can be changed by pressing the
these measurements:
• Analog Output 1 value (mA)
• Analog Output 2 value (mA)
For security, you can enable a passcode feature to restrict
access to configuration and calibration settings to authorized personnel only. See Part Three, Section 4.5 for details.
Three methods are available to calibrate the system. See
Part Three, Section 5.1 for details. With the optional Cal-
Cube™ assembly, you can conveniently check calibration any time (Section 6.2). The mA value for each analog
output can also be calibrated (Section 5.5).
ØØ and
×× keys to show
Rev. 2-300Accu4™ Low-range Turbidimeter System
13
PART ONE - INTRODUCTIONSECTION 1 - GENERAL INFORMATION
Analog Outputs
Relays
The analyzer provides two isolated analog outputs (1 and
2). Each output represents the measured turbidity, and can
be set to be 0-20 mA or 4-20 mA.
Turbidity values can be entered to define endpoints at which
the minimum and maximum analog output values are desired.
During calibration, both analog outputs can be selected to:
•
Hold their present values (HOLD OUTPUTS).
•
Transfer to preset values to operate control elements by an
amount corresponding to those values (XFER OUTPUTS).
•
Remain active to respond to the measured turbidity
(ACTIVE OUTPUTS).
For complete analog output setup details, including transfer
setup, refer to Part Three, Section 4.3.
The analyzer has four electromechanical relays with SPDT
contacts. Each relay can be set to function as a CONTROL,
ALARM (dual-alarm), or STATUS relay. CONTROL and
ALARM relays operate in response to the measured turbidity. STATUS relays operate only in response to selected
system diagnostic conditions.
☞
NOTE: Relay D is factory-set to function as a STATUS re-
lay but can be configured as a CONTROL or
ALARM relay.
When a relay is set to function as a STATUS relay, it becomes a dedicated system diagnostic-only alarm relay. A
STATUS relay can be configured for:
•
FAIL mode -- sets STATUS relay to activate when there
is a FAIL condition (analyzer, light source or detector) or
a sensor CHAMBER UNKNOWN condition.
•
WARN mode -- sets STATUS relay to activate only when
there is a sensor flow CHAMBER DIRTY condition.
•
ALL mode -- sets STATUS relay to activate when there
is any abnormal system condition (any FAIL condition,
or sensor CHAMBER DIRTY or UNKNOWN condition).
A “WARNING CHECK STATUS” message automatically
flashes on the MEASURE screen whenever the analyzer
detects any of these system diagnostic conditions:
Accu4™ Low-range Turbidimeter SystemRev. 2-300
14
PART ONE - INTRODUCTIONSECTION 1 - GENERAL INFORMATION
1.2Modular Construction
•
ANALYZER FAIL
•
CHAMBER DIRTY
•
CHAMBER UNKNOWN
•
SOURCE 1 FAIL
•
SOURCE 2 FAIL
•
DETECT 1 FAIL
•
DETECT 2 FAIL
To determine the condition causing the warning message,
display the “STATUS screens” in the TEST/MAINT menu
branch. For more details, refer to Part Three, Section 6.1.
During calibration, CONTROL and ALARM relay on/off
states are affected in the same way as the analog outputs
by the “(HOLD/XFER/ACTIVE) OUTPUTS” screen selection. These relays are also held at their present on/off
states, transferred to desired preset on/off states, or remain
active to respond to measured turbidity. For complete relay
setup details, including transfer setup, see Part Three, Section 4.4.
The modular construction of the analyzer simplifies field
servicing and provides electrical safety. The front door/
keypad assembly and the sensor use voltages no greater
than 24 VDC, and are completely safe to handle.
1.3Retained
Configuration Values
1.4Analyzer and Sensor
Serial Numbers
Opening the analyzer door accesses terminals inside the
enclosure for electrical connections. Line power must be
connected to specifically designated terminals on TB3.
WARNING:
REMOVE LINE POWER BEFORE NEARING THIS AREA
TO AVOID ELECTRICAL SHOCK.
All user-entered configuration values are retained indefinitely, even if power is lost or turned off. The non-volatile
analyzer memory does not require battery backup.
A label with the analyzer model number, serial number,
build date, and other items is affixed to the top of the analyzer enclosure. A similar label is affixed to the sensor.
Rev. 2-300Accu4™ Low-range Turbidimeter System
15
PART ONE - INTRODUCTIONSECTION 1 - GENERAL INFORMATION
1.5EMI/RFI Immunity
The analyzer is designed to provide protection from most
normally encountered electromagnetic interference. This
protection exceeds U.S. standards and meets European
IEC 801-series testing for electromagnetic and radio frequency emissions and susceptibility. Refer to Figure 1-1
and the specifications in Section 2.2 for more information.
Accu4™ Low-range Turbidimeter SystemRev. 2-300
FIGURE 1-1 EMI/RFI Immunity Diagram
16
PART ONE - INTRODUCTIONSECTION 2 - SPECIFICATIONS
ECTION
S
2
SPECIFICATIONS
2.1 Model 8320 Low-range
Turbidity Sensor
Operational
Mechanical
2.2 Model T53 Analyzer
Operational
Flow Rate................................... 0.05 to 7 GPM (0.19 to 26.5 LPM)
Cleaning Method ........................ Water rinse, wipe surfaces
Enclosure:
Standard Sensor Design ....... NEMA 4X (≅ CSA type 4; ≅ IP65), molded and
fiberglass-reinforced polyester (flame retardant)
with four integral tabs for surface mounting
High Pressure Design............ NEMA 4X (≅ CSA type 4; ≅ IP65), PPO struc-
tural foam (V-0 flammability rating per U.L. 94
test) with four integral tabs for surface mounting
Mounting Configurations............. Surface or pipe mount
Net Weight................................. 10 lbs. (4.5 kg) approximately
Display....................................... Graphic dot matrix LCD, 128 x 64 pixels with
LED backlighting; 1/2 inch (13 mm) main
character height; 1/8 inch (3mm) auxiliary
information character height; menu screens
contain up to six text lines
MeasurementAuto-ranging Scale
Turbidity................................ 0.000-100.0 NTU with auto-ranging and decimal
point shift above 1.000 NTU and 10.00 NTU
(same for other measurement units)
mA Outputs (1 and 2)............ 0.00-20.00 mA or 4.00-20.00 mA
Rev. 2-300Accu4™ Low-range Turbidimeter System
Ambient Conditions:
Operation.............................. -4 to +140°F (-20 to +60°C); 0-95% relative
humidity, non-condensing
Storage................................. -22 to +158°F (-30 to +70°C); 0-95% relative
humidity, non-condensing
Relays: Types/Outputs ...................Four electromechanical relays; SPDT (Form C)
contacts; U.L. rated 5A 115/230 VAC, 5A @ 30
VDC resistive
Operational Mode ..............Each relay (A, B, C, and D) can be assigned to be
driven by the measured turbidity or diagnostics
Function Modes:
Control...................... Settings for high/low phasing, setpoint, dead-
band, overfeed timer, off delay, and on delay
17
PART ONE - INTRODUCTIONSECTION 2 - SPECIFICATIONS
Alarm ...........................Settings for low alarm point, low alarm point
deadband, high alarm point, high alarm point
deadband, off delay, and on delay
Status....................... Setting for FAIL, WARN or ALL system diag-
nostic conditions to activate relay when
specific conditions exist (analyzer, light source
1/2 or detector 1/2 failure, sensor chamber
dirty or sensor chamber unknown)
Indicators...........................Relay annunciators (A, B, C, and D) indicate
respective relay on/off status
Sensor-to-Analyzer Distance ...... 30 ft. (9 m) maximum (consult factory if
longer distances are required)
Power Requirements .................. 90-130 VAC, 50/60 Hz. (10 VA max.) or
180-260 VAC, 50/60 Hz. (10 VA max.)
Calibration Methods:
PRIMARY............................... Enter one primary standard value (formazin
suspension is recommended).
CUBE CAL ............................. Temporarily insert an optional Cal-Cube
TM
assembly into the sensor and enter its factory-
certified standard value.
SAMPLE................................. Enter one sample value determined by labo-
ratory analysis or calibrated portable meter.
Analog Outputs .......................... Two isolated 0/4-20 mA outputs; each with
0.004 mA (12-bit) resolution and capability to
drive up to 600 ohm loads
NOTE:
Each output represents the measured turbidity. Turbidity values can
be entered to define the endpoints at which the minimum and maximum mA output values are desired. During calibration, both outputs
can be selected to hold their present values, transfer to preset values to operate control elements by an amount corresponding to
those values, or remain active to respond to the measured turbidity.
Communication: RS-232 ........... Enables configuration and retrieval of measured
data for one analyzer using an IBM-compatible
PC and optional GLI software tool kit
HART.............. Enables configuration and retrieval of measured
data for up to 15 analyzers over communication
link using appropriate hand-held terminal or data
system with HART software
Memory Backup (non-volatile) .... All user settings are retained indefinitely in
memory (EEPROM)
EMI/RFI Conformance................ Meets European standards for conducted and
radiated emissions and immunity; certified CE
compliant for applications as specified by EN 50081-1
for emissions and EN 50082-2 for immunity
Electrical Certifications:
General Purpose (pending) ..... UL, C-UL, FM, and CENELEC
Division 2 (pending)................ UL, C-UL, and FM: Groups A, B, C, D, F, and G
Zone 2 (pending) .................... CENELEC: Group IIC
Mechanical
2.3 Accu4™ Turbidimeter
System Performance
(electrical, analog outputs)
Accu4™ Low-range Turbidimeter SystemRev. 2-300
Enclosure................................... NEMA 4X; polycarbonate face panel, epoxy-
coated cast aluminum door and case with
four 1/2 inch (13 mm) conduit holes, nylon
mounting bracket, and stainless hardware
Mounting Configurations............. Panel, surface, and pipe mounting
Net Weight................................. 5 lbs. (2.3 kg) approximately
System Accuracy ....................... ± 2% of reading, all ranges
Repeatability .............................. 0.1% of span or better
Temperature Drift....................... Zero and Span: 0.01% of span per °C
18
PART TW O - INSTALLATIONSECTION 1 - UNPACKING
PART TWO - INSTALLATION
ECTION
S
UNPACKING
After unpacking, it is recommended to save the shipping
carton and packing materials in case the instrument must be
stored or re-shipped. Inspect the equipment and packing
materials for signs of shipping damage. If there is any evidence of damage, notify the transit carrier immediately.
ECTION
S
MECHANICAL REQUIREMENTS
1
2
2.1Sensor Location
Locate the Model 8320 turbidity sensor indoors or outdoors
within 30 ft. (9 m) of where the Model T53 analyzer is to be
installed. The sensor may be installed in an “in line” process
configuration using a closed loop piping arrangement (Figure 2-1) or in a sample bypass line configuration using an
open drain piping arrangement (Figure 2-2).
PART TW O - INSTALLATIONSECTION 2 - MECHANICAL REQUIREMENTS
FIGURE 2-2 Recommended Sample Bypass Line Sensor Piping Arrangement
2.2Sensor Mounting
☞
The Model 8320 low-range turbidity sensor may the standard design sensor (Figure 2-3) or the high pressure design
sensor which is housed in a different enclosure (Figure
2-4). In either case, vertically orient the sensor (inlet fitting
at bottom) and surface mount it where there is little or no
mechanical vibration. Use 1/4 inch screws to fasten the enclosure onto a flat surface. (The sensor may be mounted
onto a vertical or horizontal 1-1/2 inch or 2 inch diameter
pipe using an optional GLI pipe mount kit. Refer to Part Five
-- Spare Parts -- for the GLI part number.
NOTE:
Avoid locations where Model 8320 ambient temperature limits (32-140°F; 0-60°C) may be exceeded.
Accu4™ Low-range Turbidimeter SystemRev. 2-300
20
PART TW O - INSTALLATIONSECTION 2 - MECHANICAL REQUIREMENTS
Note:
Dimensions shown in inches (mm).
FIGURE 2-3 Standard Design Model 8320 Sensor -- Installation Dimension Details
Note:
Dimensions shown in inches (mm).
FIGURE 2-4 High Pressure Design Model 8320 Sensor -- Installation Dimension Details
Rev. 2-300Accu4™ Low-range Turbidimeter System
21
PART TW O - INSTALLATIONSECTION 2 - MECHANICAL REQUIREMENTS
2.3Sensor Plumbing
Connections
☞
Plumb the process into the Model 8320 sensor inlet fitting.
Plumb the sensor outlet fitting back into the process line
(Figure 2-1) or to an open drain (Figure 2-2). Bushings may
be used to reduce to a smaller pipe size. Barb or tube fittings may be installed.
sample tubing for applications in which sediment buildup
commonly occurs. The resultant faster flow rate helps flush
sediment through the sensor.
NOTE: Use Teflon tape to seal the inlet and outlet connec-
tions. Do not use pipe dope or other liquid sealants.
Recommendation: Use ball valves (V1 designations in Figure 2-1 and 2-2) to conveniently isolate
the sensor when calibrating or removing the sensor.
In a closed-loop piping arrangement (Figure 2-1),
3-way valves (V2 designations) are required to conveniently drain the flow chamber during calibration.
A bypass line piping arrangement (Figure 2-2) only
requires one 3-way valve.
CAUTION: WHEN USING METAL FITTINGS, DO NOT
EXCESSIVELY TIGHTEN THEM ONTO THE SENSOR’S
PLASTIC INLET AND OUTLET -- THEY MAY CRACK.
Recommendation:
Use larger
☞
NOTE: When choosing the sampling point for the process
pipe plumbing connection, avoid a top-of-pipe or
bottom-of-pipe location. Figure 2-5 illustrates the
best sampling point location: side-of-pipe with sample line inserted to the process pipe centerline.
Figure 2-5 Sampling Point Guidelines for Process Pipe Plumbing Connection
Accu4™ Low-range Turbidimeter SystemRev. 2-300
22
PART TW O - INSTALLATIONSECTION 2 - MECHANICAL REQUIREMENTS
2.4Analyzer Location
2.5Analyzer Mounting
1.Locate the analyzer as close as possible to the installed
sensor. Do not exceed a maximum distance of 30 feet
(9 m) between the sensor and analyzer. (Consult factory if longer distances are required.)
2.Mount the analyzer in a location that is:
➥
Clean and dry where there is little or no vibration.
➥
Protected from corrosive fluids.
➥
Within ambient temperature limits (-4 to +140°F or
-20 to +60°C).
CAUTION:
EXPOSING THE ANALYZER TO DIRECT
SUNLIGHT MAY INCREASE THE OPERATING
TEMPERATURE ABOVE ITS SPECIFIED LIMIT.
Figure 2-6 illustrates various ways to mount the analyzer
using the supplied bracket and hardware. Determine the
mounting method and attach the hardware as shown in the
respective illustration. Refer to Figure 2-7 for analyzer installation dimension details.
FIGURE 2-6 Analyzer Mounting Arrangements
Rev. 2-300Accu4™ Low-range Turbidimeter System
23
PART TW O - INSTALLATIONSECTION 2 - MECHANICAL REQUIREMENTS
lyzer in 1/2-inch, grounded metal conduits. If using only
shielded cables, appropriate strain reliefs or cable grips are
required. (GLI sells accessory cable grips, part number
3H1091, and watertight locknuts, part number 3H1230, for
cable entries.) Seal unused cable entry holes with appropriate plugs.
NOTE:
Use NEMA 4 (≅ CSA type 4; ≅ IP65) rated fittings
Run all wiring to the sensor and ana-
and plugs to maintain the watertight integrity of the
NEMA 4X sensor and analyzer enclosures, and to
comply with Div. 2 hazardous area requirements.
24
PART TW O - INSTALLATIONSECTION 3 - ELECTRICAL CONNECTIONS
ECTION
S
ELECTRICAL CONNECTIONS
To access terminal blocks for electrical connections, open
the left-hinged enclosure door by unscrewing the four fasteners. Figure 2-8 or 2-9 shows the terminal block
arrangement and terminal designations inside the analyzer.
3
☞
☞
☞
NOTE: All terminals are suitable for single wires up to 14
AWG (2.5 mm2).
Wiring Tip!
electromagnetic compatibility requirements, follow these
general wiring guidelines:
1. Keep all cable shields as short as possible inside
the analyzer, and connect them to the ground terminals provided. Performance may be improved by
using cable glands that enable the shield to directly
contact the analyzer chassis.
2. Use Steward ferrite 28 B0590-000 or equivalent on
the sensor cable -- two turns required.
3. In harsh conducted RF conditions, connect the earth
ground of the analyzer to a local, known earth
ground source.
NOTE: For ease of wiring, route line power and relay out-
To comply with European Community (CE)
puts through the back conduit holes, and use the
front conduit holes for all other wiring.
3.1GLI Model 8320
Low-range
Turbidity Sensor
☞
☞
Rev. 2-300Accu4™ Low-range Turbidimeter System
The Model 8320 sensor is supplied with LED light source
and detector cables connected to its terminal board. Remove its protective cover to check that these cables have
not been accidentally disconnected during shipment (refer
to wiring diagram on inside of sensor enclosure door).
Wiring Tip!
grounded metal conduit to protect it from moisture,
electrical noise, and mechanical damage.
NOTE: Do not route sensor cable in any conduit containing
AC power wiring (“electrical noise” may interfere
with sensor signal). Also, always re-calibrate the
system when the cable length between sensor and
analyzer changes.
Route the sensor cable in 1/2-inch,
25
PART TW O - INSTALLATIONSECTION 3 - ELECTRICAL CONNECTIONS
OUTPUT 2
RELAY A
NC COM
RS-232
TX RX
NO
33125476
4-20 mA
OUTPUT 1
-+-
+
HART
234567
TB2
TURBIDITY INTERCONN ECT
TTL
+
GND
GND
ORANGE
2189 221019201618
TURBIDITY ANA LYZER
RELAY B
RELAY C
COMNCNCNO
POWER
90-130 VAC
180-260 VAC
10VA 50/60 Hz
COM
8
BLUE
BROWN
F
80mA
T
WHITE
F
2
100mA
T
VIOLET
GREEN
13141512
115230N
RED
YELLOW
1
BLACK
GRAY
RELAY D
NCNONOCOM
1091211342
TB1
TB3
FIGURE 2-8 Terminal Block Designations for Analyzers with “B” Prefix Serial Number
7%785%,',7<6(1625+22.83
5('
6+,(/'
9,2/(7
:+,7(
*5((1
6+,(/'
<(//2:
%/8(
*5$<
%/$&.
%52:1
7/
25$1*(
FIGURE 2-9 Terminal Block Designations for Analyzers with No Letter Prefix Serial Number
Accu4™ Low-range Turbidimeter SystemRev. 2-300
26
PART TW O - INSTALLATIONSECTION 3 - ELECTRICAL CONNECTIONS
Refer to Figure 2-10 or 2-11 and connect the sensor (or interconnect) cable wires to appropriate terminals on TB1,
matching colors as indicated.
TURBIDITY INTERCONNECT
TB1
BLUE
BLACK
ORANGE
212219201618
GRAY
BROWN
YELLOW
VIOLET
WHITE
GREEN
RED
13141512
FIGURE 2-10 Connecting Model 8320 Low-range Turbidity Sensor
to Analyzers with “B” Prefix Serial Number
FIGURE 2-11 Connecting Model 8320 Low-range Turbidity Sensor
to Analyzers with No Letter Prefix Serial Number
Rev. 2-300Accu4™ Low-range Turbidimeter System
27
PART TW O - INSTALLATIONSECTION 3 - ELECTRICAL CONNECTIONS
3.2Analog Outputs
☞
Two isolated analog outputs (1 and 2) are provided. Each
output represents the measured turbidity, and can be set to
be 0/4-20 mA.
and earth ground, but not from each other.
configuring the outputs, refer to Part Three, Section 4.3.
Wiring Tip!
cable for connecting the analog outputs. To protect the
output signal from EMI/RFI, connect cable shields to:
The grounding strip at bottom of case (5 open holes,
Fig. 2-8) for analyzers with “B” prefix serial number.
The “ground symbol” Terminal 1 on TB1 (Figure 2-9)
for analyzers with no letter prefix serial number.
Each 0/4-20 mA output can drive a load of up to 600 ohms.
•Output 1: Connect the load to Terminals 2 and 3 on TB1,
matching polarity as indicated.
The outputs are isolated from the inputs
For details on
Use high quality, shielded instrumentation
3.3Relay Outputs
☞
•Output 2: Connect the load to Terminals 4 and 5 on TB1,
matching polarity as indicated.
NOTE: When using the HART communication option, a
digital signal is encoded onto the 4-20 mA analog
Output 1 signal. In a HART SINGLE MODE wiring
configuration, Output 1 remains available for normal
use. However, in a HART MULTI-DROP wiring configuration, Output 1 becomes dedicated to that
function and cannot be used. See Part Three, Section 8 for more HART communication information.
The analyzer is equipped with four electromechanical relays. For relay setup details, see Part Three, Section 4.4.
CAUTION:
DO NOT EXCEED THE CONTACT RATING FOR
EACH RELAY (5A 115/230 VAC). WHEN SWITCHING
LARGER CURRENTS, USE AN AUXILIARY RELAY
SWITCHED BY THE ANALYZER RELAY TO EXTEND
ANALYZER RELAY LIFE. WHEN USING RELAY
OUTPUTS, MAKE SURE THAT LINE POWER WIRING
CAN ADEQUATELY CONDUCT THE CURRENT
DRAW OF THE SWITCHED LOAD(S).
Accu4™ Low-range Turbidimeter SystemRev. 2-300
28
Loading...
+ 63 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.