GLI International Accu4 Operating Instructions Manual

OPERATING INSTRUCTION MANUAL
Accu4
Low-range
Turbidimeter System
(Model T53 Analyzer and 8320 Sensor)
Worldwide Headquarters and Sales:
9020 West Dean Road Milwaukee, Wisconsin 53224 U.S.A.
In the interest of improving and updating its equipment, GLI reserves the right to alter specifications to equipment at any time.
Rev. 2-300 Accu4™ Low-range T urbidimeter Sys tem
Phone: Fax: E-mail: Web:
Represented By:
[414] 355-3601 [414] 355-8346 info@gliint.com www.gliint.com
Viridor Instrumentation
A company
1
Accu4 Low-range Turbidimeter System Rev. 2-300
2
IMPORTANT SAFETY INFORMATION
This measurement system is compliant with safety standards as outlined in:
FMRC Class Numbers 3600, 3611, and 3810 (U.S.A.)
CSA C22.2 No. 142 and C22.2 No. 213 (Canada)
EN 61010-1 (European Community)
Please read and observe the following:
Opening the analyzer door exposes you to line power voltage, if present, at terminals on TB2 and TB3
inside the enclosure. This may be hazardous. Always remove line power before entering this area in the analyzer. However, the analyzer door assembly and the sensor contain only low voltages and are completely safe to handle.
Wiring or repairs should only be performed by qualified personnel and only to an unpowered analyzer.
Whenever it appears that analyzer safety is questionable, disable the analyzer to ensure against any
unintended operation. For example, an unsafe condition is likely when:
1) The analyzer appears visibly damaged.
2) The analyzer fails to operate properly or provide the intended measurements.
3) The analyzer has been stored for long periods at temperatures above 158°F (70°C).
This measurement system must be installed by specially trained personnel in accordance with relevant
local codes and instructions contained in this operating instruction manual. Observe all technical specifications of the system. If one line of the line power mains is not neutral, use a double-pole mains switch to disconnect the analyzer.
HELPFUL IDENTIFIERS
In addition to information on installation and operation, this instruction manual may contain WARNINGS pertaining to user safety, CAUTIONS regarding possible instrument malfunction, and NOTES on important, useful operating guidelines.
WARNING:
A WARNING LOOKS LIKE THIS. IT WARNS YOU OF THE POTENTIAL FOR PERSONAL INJURY.
CAUTION:
A CAUTION LOOKS LIKE THIS. IT ALERTS YOU TO POSSIBLE INSTRUMENT MALFUNCTION OR DAMAGE.
NOTE: A note looks like this. It alerts you to important operating
information.
Rev. 2-300 Accu4 Low-range Turbidimeter System
3
Definition of Equipment Symbols
This symbol
means CAUTION
and alerts you to possible danger or
instrument malfunction. Refer to this manual before proceeding.
This symbol
means that this is a protective ground terminal
alerts you to connect an earth ground to it.
This symbol
means that there is alternating current present
alerts you to be careful.
and
and
WARRANTY
GLI International, Inc. warrants the Accu4 Low-range Turbidimeter System to be free from defects in material or workmanship for a period of 2 years (24 months) from the date this product was shipped from our facility. A warranty claim will not be honored if defects are not reported within the warranty period, or if GLI International determines that defects or damages are due to normal wear, misapplication, lack of maintenance, abuse, improper installation, alteration, or abnormal conditions. GLI Internationals obligation under this warranty shall be limited to, at its option, replacement or repair of this product. The product must be returned to GLI International, freight prepaid, for examination. The product must be thoroughly cleaned and any process chemicals removed before it will be accepted for replacement or repair. GLI Internationals liability shall not exceed the cost of the product. Under no circumstances will GLI International be liable for any incidental or consequential damages, whether to person or property. GLI International will not be li­able for any other loss, damage or expense of any kind, including loss of profits, resulting from the installation, use, or inability to use this product.
Accu4 Low-range Turbidimeter System Rev. 2-300
4
CONDENSED OPERATING INSTRUCTIONS
This manual contains details for all operating aspects of the Accu4 Low-range Turbidimeter System. The following condensed instructions are provided to assist you in getting the system started up and op­erating as quickly as possible.
measurement operation.
To use specific analyzer features, refer to the appropriate sections in this
manual for instructions.
1. CONNECTING SENSOR
After the T53 analyzer is properly mounted (Part Two, Section 2.5), connect the GLI Model 8320 low­range turbidity sensor, matching wire colors to terminals as indicated:
These condensed instructions only pertain to basic turbidity
Sensor Wire
Colors
Inner Shield Grounding strip lug Terminal #11 on TB1
Red Terminal #12 on TB1 Terminal #12 on TB1
Violet Terminal #13 on TB1 Terminal #13 on TB1 Green Terminal #14 on TB1 Terminal #14 on TB1 White Terminal #15 on TB1 Terminal #15 on TB1
Yellow Terminal #16 on TB1 Terminal #16 on TB1
Inner Shield Grounding strip lug Terminal #17 on TB1
Brown Terminal #18 on TB1 Terminal #18 on TB1
Gray Terminal #19 on TB1 Terminal #19 on TB1
Blue Terminal #20 on TB1 Terminal #20 on TB1
Black Terminal #21 on TB1 Terminal #21 on TB1
Orange Terminal #22 on TB1 Terminal #22 on TB1
Analyzers with
B Prefix Serial No.
Analyzers with
No Letter Prefix Serial No.
2. CONNECTING LINE POWER
Important:
Follow the instructions in Part Two, Section 3.5 to connect line power to the analyzer.
3. ADJUSTING DISPLAY CONTRAST
Ambient lighting conditions may make it necessary to adjust display contrast to improve visibility. With the MEASURE screen displayed, press and hold
×
Ø
×
or
until attaining the desired contrast.
key
Ø
ENTER key
the
and simultaneously press the
4. CALIBRATING SYSTEM
Always initially
calibrate
the Accu4 Low-range Turbidimeter System to ensure accurate meas­urement. Any time thereafter, you can conveniently check system calibration using an optional accessory Cal-CubeTM assembly.
When you need to calibrate with a
method. GLI only guarantees measurement accuracy when a 40 NTU formazin suspension is used as the primary standard.
When calibration need not conform to USEPA requirements
method (requires optional GLI CAL-CUBE assembly) or the SAMPLE CAL method.
Refer to Part Three, Section 5 for instructions on all calibration methods.
5. COMPLETING ANALYZER CONFIGURATION
To further configure the analyzer to your application requirements, use the appropriate CONFIGURE screens to make selections and “key in” values. Refer to Part Three, Section 4 for complete configu­ration details.
Rev. 2-300 Accu4 Low-range Turbidimeter System
primary standard
5
, use the PRIMARY CAL calibration
, use the CUBE CAL calibration
Accu4 Low-range Turbidimeter System Rev. 2-300
6
TABLE OF CONTENTS
PART ONE - INTRODUCTION
SECTION 1 GENERAL INFORMATION
1.1 Capability Highlights ........................................................................13-15
1.2 Modular Construction ............................................................................15
1.3 Retained Configuration Values .............................................................15
1.4 Analyzer and Sensor Serial Numbers ...................................................15
1.5 EMI/RFI Immunity.................................................................................. 16
SECTION 2 SPECIFICATIONS
....................................................................................17-18
PART TWO - INSTALLATION
SECTION 1 UNPACKING
SECTION 2 MECHANICAL REQUIREMENTS
2.1 Sensor Location ...............................................................................19-20
2.2 Sensor Mounting ..............................................................................20-21
2.3 Sensor Plumbing Connections ..............................................................22
2.4 Analyzer Location..................................................................................23
2.5 Analyzer Mounting............................................................................23-24
2.6 Sensor and Analyzer Conduit Hole Requirements................................ 24
SECTION 3 ELECTRICAL CONNECTIONS
3.1 GLI Model 8320 Low-range Turbidity Sensor...................................25-27
3.2 Analog Outputs......................................................................................28
3.3 Relay Outputs...................................................................................28-29
3.4 Closed Contact TTL Input .....................................................................29
3.5 Line Power .......................................................................................30-31
.................................................................................................19
PART THREE - OPERATION
SECTION 1 USER INTERFACE
1.1 Display ..................................................................................................32
1.2 Keypad .............................................................................................32-33
1.3 MEASURE Screen (normal display mode)............................................ 34
SECTION 2 MENU STRUCTURE
2.1 Displaying Main Branch Selection Screen ............................................35
2.2 Displaying Top-level Menu Screens.................................................35-36
2.3 Displaying Submenu Screens ...............................................................36
2.4 Adjusting Edit/Selection Screen Values ...........................................36-37
2.5 Entering (Storing) Edit/Selection Screen Values/Choices.....................37
Rev. 2-300 Accu4 Low-range Turbidimeter System
7
TABLE OF CONTENTS (continued)
SECTION 3 ADJUSTING DISPLAY CONTRAST
SECTION 4 CONFIGURING THE ANALYZER
4.1 Selecting LANGUAGE to Operate Analyzer.......................................... 38
4.2 Configuring Sensor Characteristics:
SELECT UNITS ..........................................................................38-39
SET FILTER Time............................................................................ 39
Select PULSE SUPPRESS (on/off)..................................................40
ENTER NOTE (top line of MEASURE screen).................................40
4.3 Configuring Analog Outputs (1 and 2):
SET 0/4 and 20 mA VALUES......................................................41-42
SET TRANSFER Value (mA)........................................................... 42
SET FILTER Time............................................................................ 43
Select SCALE 0 mA/4 mA (low endpoint) ........................................43
4.4 Configuring Relays (A, B, C, and D):
SET FUNCTION Mode (alarm, control, or status).......................44-45
SET TRANSFER Mode (relay on or off) ..........................................46
SET ACTIVATION (configuration values) ...................................46-48
4.5 SET PASSCODE (feature enabled or disabled) ...................................48
4.6 Configuration Settings Summary (ranges/choices and defaults) ..........49
.............................................................37
SECTION 5 CALIBRATING THE ANALYZER
5.1 Things to Know About Calibration ....................................................50-51
5.2 PRIMARY CAL Method (requires formazin suspension):
Preparing Formazin Solution ......................................................51-52
Filling Sensor..............................................................................52-54
Entering Value .................................................................................54
Resuming Operation ...................................................................54-55
5.3 CUBE CAL Method (requires optional Cal-CubeTM assembly):
TM
Inserting Cal-Cube
Assembly ..................................................55-56
Entering Value ............................................................................56-57
Resuming Operation ........................................................................57
5.4 SAMPLE CAL Method (requires sample with value
determined by laboratory analysis or portable meter)......................57-59
5.5 Analog Outputs (1 and 2) Calibration...............................................59-60
SECTION 6 TEST/MAINTENANCE
6.1 STATUS Checking (analyzer, sensor, and relays)...........................61-63
6.2 CHECK CAL/HOLD Outputs ............................................................63-64
6.3 OVERFEED RESET (relay timers)........................................................ 65
6.4 OUTPUT (1 and 2) Analog Test Signals ..........................................65-66
6.5 RELAY (A, B, C, and D) Operating Test................................................66
6.6 SENSOR TYPE Checking................................................................66-67
6.7 EPROM VERSION Checking ................................................................67
Accu4 Low-range Turbidimeter System Rev. 2-300
8
TABLE OF CONTENTS (continued)
6.8 SIM SENSOR Setting.......................................................................67-68
6.9 RESET DEFAULTS............................................................................... 68
SECTION 7 RELAY OVERFEED TIMER FEATURE
7.1 Why Use an Overfeed Timer................................................................. 69
7.2 Configuring Relay Overfeed Timers......................................................69
7.3 Overfeed Timer “Timeout” Operation .................................................... 69
7.4 Resetting Overfeed Timers ...................................................................69
7.5 Interactions with Other Analyzer Functions......................................69-70
SECTION 8 HART OPTION
8.1 Introduction ...........................................................................................71
8.2 Analyzer Operating Modes for HART Network.................................72-73
8.3 SINGLE MODE (Point-to-Point) Wiring Arrangement ...........................73
8.4 MULTI-DROP Wiring Arrangement.......................................................74
8.5 HART Preferences Setup:
Changing Polling Address ...............................................................75
Viewing Number of Required Preambles ....................................75-76
8.6 Device Preferences Setup:
Viewing Final Assembly Number ..................................................... 76
Viewing Model Number...............................................................76-77
Viewing Manufacturer ......................................................................77
Assigning a Tag ...............................................................................77
Assigning a Descriptor ..................................................................... 78
Assigning a Message....................................................................... 78
Assigning User-defined Date ......................................................78-79
Viewing Identification (ID) ................................................................79
Viewing Revisions............................................................................ 79
8.7 Master Reset Function........................................................................ 80
8.8 “Refresh” Function ................................................................................80
8.9 Protocol Command Set for PC Programming........................................ 80
PART FOUR - SERVICE AND MAINTENANCE
SECTION 1 GENERAL INFORMATION
1.1 Cleaning Sensor Flow Chamber ......................................................81-82
1.2 Cleaning Optional Cal-CubeTM Assembly.............................................. 82
1.3 Replacing Sensor Light Source or Detector:
Removing Inoperative Part ..............................................................83
Installing New Part ......................................................................83-84
Re-calibrating Measurement System............................................... 84
1.4 Replacing Fuse(s) .................................................................................85
1.5 Replacing Relays ..................................................................................85
1.6 Inspecting Sensor Cable .......................................................................85
Rev. 2-300 Accu4 Low-range Turbidimeter System
9
TABLE OF CONTENTS (continued)
SECTION 2 PRESERVING MEASUREMENT ACCURACY
2.1 Eliminating Bubbles in Sensor Flow Chamber:
Restricting Sensor Outlet .................................................................86
Using an External Bubble Trap ........................................................86
2.2 Keeping Sensor Flow Chamber Clean ..................................................87
2.3 Keeping System Calibrated...................................................................87
2.4 Avoiding Electrical Interference.............................................................87
SECTION 3 TROUBLESHOOTING
3.1 System Diagnostic Condition Messages ............................................... 88
3.2 Isolating the Problem:
Checking Electrical Connections ..................................................... 89
Checking Operation Using System Diagnostics ..............................89
SECTION 4 ANALYZER REPAIR/RETURN
4.1 Customer Assistance.............................................................................90
4.2 Repair/Return Policy .............................................................................90
PART FIVE - SPARE PARTS AND ACCESSORIES
.................................................................................................................91
Accu4 Low-range Turbidimeter System Rev. 2-300
10
TABLE OF CONTENTS (continued)
ILLUSTRATIONS
Figure 1-1
Figure 2-1
Figure 2-2
Figure 2-3
Figure 2-4
Figure 2-5
Figure 2-6
Figure 2-7
Figure 2-8
Figure 2-9
Figure 2-10
Figure 2-11
Figure 2-12
Figure 2-13
Figure 2-14
Figure 2-15
Figure 2-16
Figure 2-17
Figure 2-18
Figure 3-1
Figure 3-2
Figure 3-3
Figure 3-4
Figure 4-1
Figure 4-2
Figure 4-3
EMI/RFI Immunity Diagram ..................................................................................................16
Recommended Closed-loop Sensor Piping Arrangement...................................................... 19
Recommended Sample Bypass Line Sensor Piping Arrangement ........................................ 20
Standard Design Model 8320 Sensor -- Installation Dimension Details..................................21
High Pressure Design Model 8320 Sensor -- Installation Dimension Details.......................... 21
Sampling Point Guidelines for Process Pipe Plumbing Connection.......................................22
Analyzer Mounting Arrangements ......................................................................................... 23
Analyzer Installation Dimension Details ................................................................................ 24
Terminal Block Designations for Analyzers with “B” Prefix Serial Number.............................26
Terminal Block Designations for Analyzers with No Letter Prefix Serial Number ................... 26
Connecting Model 8320 Low-range Turbidity Sensor to
Analyzers with “B” Prefix Serial Number...............................................................................27
Connecting Model 8320 Low-range Turbidity Sensor to
Analyzers with No Letter Prefix Serial Number .....................................................................27
Connecting Control/Alarm Device(s) to Electromechanical Relay(s) .....................................29
Connecting 115 V Single Phase to Analyzers with “B” Prefix Serial Number.........................30
Connecting 115 V Single Phase to Analyzers with No Letter Prefix Serial Number ...............30
Connecting 230 V Single Phase to Analyzers with “B” Prefix Serial Number.........................31
Connecting 230 V Single Phase to Analyzers with No Letter Prefix Serial Number ...............31
Connecting 230 V Split Phase to Analyzers with “B” Prefix Serial Number............................ 31
Connecting 230 V Split Phase to Analyzers with No Letter Prefix Serial Number .................. 31
Analyzer Keypad ..................................................................................................................33
Location of SINGLE MODE/MULTI-DROP Switch (HART-equipped analyzers only).............73
HART SINGLE MODE (Point-to-Point) Wiring Arrangement (for single analyzer) .................73
HART MULTI-DROP Wiring Arrangement (for multiple analyzer network) ............................ 74
Standard Design Sensor -- Component Details ..................................................................... 81
High Pressure Design Sensor -- Component Details .............................................................82
Turbidity Sensor Terminal Block Designations...................................................................... 84
Table A
Table B
Table C
Table D
Rev. 2-300 Accu4 Low-range Turbidimeter System
Relay Configuration Settings ...........................................................................................46-47
Analyzer Configuration Settings (Ranges/Choices and Defaults) ..........................................49
Relay Overfeed Timer Interactions with Other Analyzer Functions........................................70
System Diagnostic Condition Status Screens/Meanings........................................................88
TABLES
11
Accu4 Low-range Turbidimeter System Rev. 2-300
12
PART ONE - INTRODUCTION SECTION 1 - GENERAL INFORMATION
PART ONE - INTRODUCTION
ECTION
1.1 Capability Highlights
High Accuracy with
Minimal Maintenance
S
GENERAL INFORMATION
The Accu4 Low-range Turbidimeter System employs a technologically-advanced sensor to provide high measure­ment accuracy and stability, while reducing maintenance requirements. It is designed to meet the International Stan­dards For Measurement of Turbidity [ISO 7027-1984(E)] and USEPA-approved GLI Method 2. This system has an auto-ranging measuring scale, enabling continuous moni­toring over a 0.000-100.0 NTU range with automatic decimal point positioning.
dimeter System automatically provides increasing display resolution as measured turbidity decreases.
further ensure high measuring accuracy, the Model 8320 sensor eliminates air or gas bubbles in the sample with its built-in bubble trap. The system is ideal for critical monitor­ing and controlling applications including potable water, filtered water, and final product clarity.
1
The Accu4 Low-range Turbi-
To
MEASURE Screen
Passcode-protected
Access
Calibration Methods
With the display in the normal MEASURE screen mode, the measured turbidity is always shown on the main middle line. The bottom auxiliary display line, shown in reverse video, can be changed by pressing the these measurements:
Analog Output 1 value (mA)
Analog Output 2 value (mA)
For security, you can enable a passcode feature to restrict access to configuration and calibration settings to author­ized personnel only. See Part Three, Section 4.5 for details.
Three methods are available to calibrate the system. See Part Three, Section 5.1 for details. With the optional Cal-
Cube assembly, you can conveniently check calibra­tion any time (Section 6.2). The mA value for each analog
output can also be calibrated (Section 5.5).
ØØ and
×× keys to show
Rev. 2-300 Accu4 Low-range Turbidimeter System
13
PART ONE - INTRODUCTION SECTION 1 - GENERAL INFORMATION
Analog Outputs
Relays
The analyzer provides two isolated analog outputs (1 and
2). Each output represents the measured turbidity, and can be set to be 0-20 mA or 4-20 mA.
Turbidity values can be entered to define endpoints at which the minimum and maximum analog output values are desired.
During calibration, both analog outputs can be selected to:
Hold their present values (HOLD OUTPUTS).
Transfer to preset values to operate control elements by an
amount corresponding to those values (XFER OUTPUTS).
Remain active to respond to the measured turbidity
(ACTIVE OUTPUTS).
For complete analog output setup details, including transfer setup, refer to Part Three, Section 4.3.
The analyzer has four electromechanical relays with SPDT contacts. Each relay can be set to function as a CONTROL, ALARM (dual-alarm), or STATUS relay. CONTROL and ALARM relays operate in response to the measured turbid­ity. STATUS relays operate only in response to selected system diagnostic conditions.
NOTE: Relay D is factory-set to function as a STATUS re-
lay but can be configured as a CONTROL or ALARM relay.
When a relay is set to function as a STATUS relay, it be­comes a dedicated system diagnostic-only alarm relay. A STATUS relay can be configured for:
FAIL mode -- sets STATUS relay to activate when there
is a FAIL condition (analyzer, light source or detector) or a sensor CHAMBER UNKNOWN condition.
WARN mode -- sets STATUS relay to activate only when
there is a sensor flow CHAMBER DIRTY condition.
ALL mode -- sets STATUS relay to activate when there
is any abnormal system condition (any FAIL condition, or sensor CHAMBER DIRTY or UNKNOWN condition).
A WARNING CHECK STATUS message automatically flashes on the MEASURE screen whenever the analyzer detects any of these system diagnostic conditions:
Accu4 Low-range Turbidimeter System Rev. 2-300
14
PART ONE - INTRODUCTION SECTION 1 - GENERAL INFORMATION
1.2 Modular Construction
ANALYZER FAIL
CHAMBER DIRTY
CHAMBER UNKNOWN
SOURCE 1 FAIL
SOURCE 2 FAIL
DETECT 1 FAIL
DETECT 2 FAIL
To determine the condition causing the warning message, display the STATUS screens in the TEST/MAINT menu branch. For more details, refer to Part Three, Section 6.1.
During calibration, CONTROL and ALARM relay on/off states are affected in the same way as the analog outputs by the (HOLD/XFER/ACTIVE) OUTPUTS screen selec­tion. These relays are also held at their present on/off states, transferred to desired preset on/off states, or remain active to respond to measured turbidity. For complete relay setup details, including transfer setup, see Part Three, Sec­tion 4.4.
The modular construction of the analyzer simplifies field servicing and provides electrical safety. The front door/ keypad assembly and the sensor use voltages no greater than 24 VDC, and are completely safe to handle.
1.3 Retained Configuration Values
1.4 Analyzer and Sensor Serial Numbers
Opening the analyzer door accesses terminals inside the enclosure for electrical connections. Line power must be connected to specifically designated terminals on TB3.
WARNING:
REMOVE LINE POWER BEFORE NEARING THIS AREA TO AVOID ELECTRICAL SHOCK.
All user-entered configuration values are retained indefi­nitely, even if power is lost or turned off. The non-volatile analyzer memory does not require battery backup.
A label with the analyzer model number, serial number, build date, and other items is affixed to the top of the ana­lyzer enclosure. A similar label is affixed to the sensor.
Rev. 2-300 Accu4 Low-range Turbidimeter System
15
PART ONE - INTRODUCTION SECTION 1 - GENERAL INFORMATION
1.5 EMI/RFI Immunity
The analyzer is designed to provide protection from most normally encountered electromagnetic interference. This protection exceeds U.S. standards and meets European IEC 801-series testing for electromagnetic and radio fre­quency emissions and susceptibility. Refer to Figure 1-1 and the specifications in Section 2.2 for more information.
Accu4 Low-range Turbidimeter System Rev. 2-300
FIGURE 1-1 EMI/RFI Immunity Diagram
16
PART ONE - INTRODUCTION SECTION 2 - SPECIFICATIONS
ECTION
S
2
SPECIFICATIONS
2.1 Model 8320 Low-range
Turbidity Sensor
Operational
Mechanical
2.2 Model T53 Analyzer
Operational
Flow Rate................................... 0.05 to 7 GPM (0.19 to 26.5 LPM)
Ambient Conditions .................... 32-140°F (0-60°C)
Sample Temperature Range....... 32-140°F (0-60°C)
Pressure Range:
Standard Sensor Design........ 0-50 psig at 68°F (0-3.4 bar at 20°C)
High Pressure Design............ 0-150 psig at 68°F (0-10.2 bar at 20°C)
Pressure Drop:
Standard Sensor Design........ 0.0017 psig at 0.1 GPM (0.0001 bar at 0.36 LPM)
High Pressure Design............ 0.165 psig at 1.0 GPM (0.012 bar at 3.8 LPM)
Residence Time ......................... 9.5 seconds at 1 GPM (3.8 LPM)
Air Venting ................................. Integral bubble trap for 0.05 to 0.5 GPM (0.19 to
1.8 LPM) flows. Installation of restrictor valve on the sensor outlet is recommended for flows above 0.5 GPM (1.8 LPM) with air in sample.
Light Sources............................. Two near-infrared (860 nM wavelength) LEDs
Sensor Flow Configuration ......... Flow-through design
Process Connections.................. 1/2 inch NPT female standard; adaptable to
3/8 inch or 1/4 inch NPT, barb or tube fittings
Wetted Materials........................ PVC, polycarbonate, polystyrene, PPO,
nitrile, and Buna-N
Cleaning Method ........................ Water rinse, wipe surfaces
Enclosure:
Standard Sensor Design ....... NEMA 4X (≅ CSA type 4; ≅ IP65), molded and
fiberglass-reinforced polyester (flame retardant) with four integral tabs for surface mounting
High Pressure Design............ NEMA 4X (≅ CSA type 4; ≅ IP65), PPO struc-
tural foam (V-0 flammability rating per U.L. 94
test) with four integral tabs for surface mounting
Mounting Configurations............. Surface or pipe mount
Net Weight................................. 10 lbs. (4.5 kg) approximately
Display....................................... Graphic dot matrix LCD, 128 x 64 pixels with
LED backlighting; 1/2 inch (13 mm) main
character height; 1/8 inch (3mm) auxiliary
information character height; menu screens
contain up to six text lines
Measurement Auto-ranging Scale
Turbidity................................ 0.000-100.0 NTU with auto-ranging and decimal
point shift above 1.000 NTU and 10.00 NTU
(same for other measurement units)
mA Outputs (1 and 2)............ 0.00-20.00 mA or 4.00-20.00 mA
Rev. 2-300 Accu4 Low-range Turbidimeter System
Ambient Conditions:
Operation.............................. -4 to +140°F (-20 to +60°C); 0-95% relative
humidity, non-condensing
Storage................................. -22 to +158°F (-30 to +70°C); 0-95% relative
humidity, non-condensing
Relays: Types/Outputs ...................Four electromechanical relays; SPDT (Form C)
contacts; U.L. rated 5A 115/230 VAC, 5A @ 30
VDC resistive
Operational Mode ..............Each relay (A, B, C, and D) can be assigned to be
driven by the measured turbidity or diagnostics
Function Modes:
Control...................... Settings for high/low phasing, setpoint, dead-
band, overfeed timer, off delay, and on delay
17
PART ONE - INTRODUCTION SECTION 2 - SPECIFICATIONS
Alarm ...........................Settings for low alarm point, low alarm point
deadband, high alarm point, high alarm point
deadband, off delay, and on delay
Status....................... Setting for FAIL, WARN or ALL system diag-
nostic conditions to activate relay when
specific conditions exist (analyzer, light source
1/2 or detector 1/2 failure, sensor chamber
dirty or sensor chamber unknown)
Indicators...........................Relay annunciators (A, B, C, and D) indicate
respective relay on/off status
Sensor-to-Analyzer Distance ...... 30 ft. (9 m) maximum (consult factory if
longer distances are required)
Power Requirements .................. 90-130 VAC, 50/60 Hz. (10 VA max.) or
180-260 VAC, 50/60 Hz. (10 VA max.) Calibration Methods:
PRIMARY............................... Enter one primary standard value (formazin
suspension is recommended).
CUBE CAL ............................. Temporarily insert an optional Cal-Cube
TM
assembly into the sensor and enter its factory-
certified standard value.
SAMPLE................................. Enter one sample value determined by labo-
ratory analysis or calibrated portable meter.
Analog Outputs .......................... Two isolated 0/4-20 mA outputs; each with
0.004 mA (12-bit) resolution and capability to
drive up to 600 ohm loads
NOTE:
Each output represents the measured turbidity. Turbidity values can be entered to define the endpoints at which the minimum and maxi­mum mA output values are desired. During calibration, both outputs can be selected to hold their present values, transfer to preset val­ues to operate control elements by an amount corresponding to those values, or remain active to respond to the measured turbidity.
Communication: RS-232 ........... Enables configuration and retrieval of measured
data for one analyzer using an IBM-compatible
PC and optional GLI software tool kit
HART.............. Enables configuration and retrieval of measured
data for up to 15 analyzers over communication
link using appropriate hand-held terminal or data
system with HART software
Memory Backup (non-volatile) .... All user settings are retained indefinitely in
memory (EEPROM)
EMI/RFI Conformance................ Meets European standards for conducted and
radiated emissions and immunity; certified CE
compliant for applications as specified by EN 50081-1
for emissions and EN 50082-2 for immunity Electrical Certifications:
General Purpose (pending) ..... UL, C-UL, FM, and CENELEC
Division 2 (pending)................ UL, C-UL, and FM: Groups A, B, C, D, F, and G
Zone 2 (pending) .................... CENELEC: Group IIC
Mechanical
2.3 Accu4 Turbidimeter System Performance (electrical, analog outputs)
Accu4 Low-range Turbidimeter System Rev. 2-300
Enclosure................................... NEMA 4X; polycarbonate face panel, epoxy-
coated cast aluminum door and case with four 1/2 inch (13 mm) conduit holes, nylon mounting bracket, and stainless hardware
Mounting Configurations............. Panel, surface, and pipe mounting
Net Weight................................. 5 lbs. (2.3 kg) approximately
System Accuracy ....................... ± 2% of reading, all ranges
Sensitivity .................................. 0.001 NTU
Repeatability .............................. 0.1% of span or better
Temperature Drift....................... Zero and Span: 0.01% of span per °C
18
PART TW O - INSTALLATION SECTION 1 - UNPACKING
PART TWO - INSTALLATION
ECTION
S
UNPACKING
After unpacking, it is recommended to save the shipping carton and packing materials in case the instrument must be stored or re-shipped. Inspect the equipment and packing materials for signs of shipping damage. If there is any evi­dence of damage, notify the transit carrier immediately.
ECTION
S
MECHANICAL REQUIREMENTS
1
2
2.1 Sensor Location
Locate the Model 8320 turbidity sensor indoors or outdoors within 30 ft. (9 m) of where the Model T53 analyzer is to be installed. The sensor may be installed in an in line process configuration using a closed loop piping arrangement (Fig­ure 2-1) or in a sample bypass line configuration using an open drain piping arrangement (Figure 2-2).
FIGURE 2-1 Recommended Closed-loop Sensor Piping Arrangement
Rev. 2-300 Accu4 Low-range Turbidimeter System
19
PART TW O - INSTALLATION SECTION 2 - MECHANICAL REQUIREMENTS
FIGURE 2-2 Recommended Sample Bypass Line Sensor Piping Arrangement
2.2 Sensor Mounting
The Model 8320 low-range turbidity sensor may the stan­dard design sensor (Figure 2-3) or the high pressure design sensor which is housed in a different enclosure (Figure 2-4). In either case, vertically orient the sensor (inlet fitting at bottom) and surface mount it where there is little or no mechanical vibration. Use 1/4 inch screws to fasten the en­closure onto a flat surface. (The sensor may be mounted onto a vertical or horizontal 1-1/2 inch or 2 inch diameter pipe using an optional GLI pipe mount kit. Refer to Part Five
-- Spare Parts -- for the GLI part number.
NOTE:
Avoid locations where Model 8320 ambient tempera­ture limits (32-140°F; 0-60°C) may be exceeded.
Accu4 Low-range Turbidimeter System Rev. 2-300
20
PART TW O - INSTALLATION SECTION 2 - MECHANICAL REQUIREMENTS
Note:
Dimensions shown in inches (mm).
FIGURE 2-3 Standard Design Model 8320 Sensor -- Installation Dimension Details
Note:
Dimensions shown in inches (mm).
FIGURE 2-4 High Pressure Design Model 8320 Sensor -- Installation Dimension Details
Rev. 2-300 Accu4 Low-range Turbidimeter System
21
PART TW O - INSTALLATION SECTION 2 - MECHANICAL REQUIREMENTS
2.3 Sensor Plumbing
Connections
Plumb the process into the Model 8320 sensor inlet fitting. Plumb the sensor outlet fitting back into the process line (Figure 2-1) or to an open drain (Figure 2-2). Bushings may be used to reduce to a smaller pipe size. Barb or tube fit­tings may be installed. sample tubing for applications in which sediment buildup commonly occurs. The resultant faster flow rate helps flush sediment through the sensor.
NOTE: Use Teflon tape to seal the inlet and outlet connec-
tions. Do not use pipe dope or other liquid sealants. Recommendation: Use ball valves (V1 designa­tions in Figure 2-1 and 2-2) to conveniently isolate the sensor when calibrating or removing the sensor. In a closed-loop piping arrangement (Figure 2-1), 3-way valves (V2 designations) are required to con­veniently drain the flow chamber during calibration. A bypass line piping arrangement (Figure 2-2) only requires one 3-way valve.
CAUTION: WHEN USING METAL FITTINGS, DO NOT EXCESSIVELY TIGHTEN THEM ONTO THE SENSOR’S PLASTIC INLET AND OUTLET -- THEY MAY CRACK.
Recommendation:
Use larger
NOTE: When choosing the sampling point for the process
pipe plumbing connection, avoid a top-of-pipe or bottom-of-pipe location. Figure 2-5 illustrates the best sampling point location: side-of-pipe with sam­ple line inserted to the process pipe centerline.
Figure 2-5 Sampling Point Guidelines for Process Pipe Plumbing Connection
Accu4 Low-range Turbidimeter System Rev. 2-300
22
PART TW O - INSTALLATION SECTION 2 - MECHANICAL REQUIREMENTS
2.4 Analyzer Location
2.5 Analyzer Mounting
1. Locate the analyzer as close as possible to the installed sensor. Do not exceed a maximum distance of 30 feet (9 m) between the sensor and analyzer. (Consult fac­tory if longer distances are required.)
2. Mount the analyzer in a location that is:
Clean and dry where there is little or no vibration.
Protected from corrosive fluids.
Within ambient temperature limits (-4 to +140°F or
-20 to +60°C).
CAUTION:
EXPOSING THE ANALYZER TO DIRECT SUNLIGHT MAY INCREASE THE OPERATING TEMPERATURE ABOVE ITS SPECIFIED LIMIT.
Figure 2-6 illustrates various ways to mount the analyzer using the supplied bracket and hardware. Determine the mounting method and attach the hardware as shown in the respective illustration. Refer to Figure 2-7 for analyzer in­stallation dimension details.
FIGURE 2-6 Analyzer Mounting Arrangements
Rev. 2-300 Accu4 Low-range Turbidimeter System
23
PART TW O - INSTALLATION SECTION 2 - MECHANICAL REQUIREMENTS
FIGURE 2-7 Analyzer Installation Dimensions Details
2.6 Sensor and Analyzer Conduit Hole Requirements
Accu4 Low-range Turbidimeter System Rev. 2-300
Recommendation:
lyzer in 1/2-inch, grounded metal conduits. If using only shielded cables, appropriate strain reliefs or cable grips are required. (GLI sells accessory cable grips, part number 3H1091, and watertight locknuts, part number 3H1230, for cable entries.) Seal unused cable entry holes with appropri­ate plugs.
NOTE:
Use NEMA 4 (≅ CSA type 4; ≅ IP65) rated fittings
Run all wiring to the sensor and ana-
and plugs to maintain the watertight integrity of the NEMA 4X sensor and analyzer enclosures, and to comply with Div. 2 hazardous area requirements.
24
PART TW O - INSTALLATION SECTION 3 - ELECTRICAL CONNECTIONS
ECTION
S
ELECTRICAL CONNECTIONS
To access terminal blocks for electrical connections, open the left-hinged enclosure door by unscrewing the four fas­teners. Figure 2-8 or 2-9 shows the terminal block arrangement and terminal designations inside the analyzer.
3
NOTE: All terminals are suitable for single wires up to 14
AWG (2.5 mm2).
Wiring Tip!
electromagnetic compatibility requirements, follow these general wiring guidelines:
1. Keep all cable shields as short as possible inside the analyzer, and connect them to the ground termi­nals provided. Performance may be improved by using cable glands that enable the shield to directly contact the analyzer chassis.
2. Use Steward ferrite 28 B0590-000 or equivalent on the sensor cable -- two turns required.
3. In harsh conducted RF conditions, connect the earth ground of the analyzer to a local, known earth ground source.
NOTE: For ease of wiring, route line power and relay out-
To comply with European Community (CE)
puts through the back conduit holes, and use the front conduit holes for all other wiring.
3.1 GLI Model 8320 Low-range Turbidity Sensor
Rev. 2-300 Accu4 Low-range Turbidimeter System
The Model 8320 sensor is supplied with LED light source and detector cables connected to its terminal board. Re­move its protective cover to check that these cables have not been accidentally disconnected during shipment (refer to wiring diagram on inside of sensor enclosure door).
Wiring Tip!
grounded metal conduit to protect it from moisture, electrical noise, and mechanical damage.
NOTE: Do not route sensor cable in any conduit containing
AC power wiring (electrical noise may interfere with sensor signal). Also, always re-calibrate the system when the cable length between sensor and analyzer changes.
Route the sensor cable in 1/2-inch,
25
PART TW O - INSTALLATION SECTION 3 - ELECTRICAL CONNECTIONS
OUTPUT 2
RELAY A
NC COM
RS-232
TX RX
NO
3312 5476
4-20 mA
OUTPUT 1
-+-
+
HART
234567
TB2
TURBIDITY INTERCONN ECT
TTL
+
GND
GND
ORANGE
2189 2210 1920 1618
TURBIDITY ANA LYZER
RELAY B
RELAY C
COMNC NCNO
POWER
90-130 VAC 180-260 VAC 10VA 50/60 Hz
COM
8
BLUE
BROWN
F
80mA
T
WHITE
F
2
100mA
T
VIOLET
GREEN
131415 12
115230 N
RED
YELLOW
1
BLACK
GRAY
RELAY D
NCNO NOCOM
1091211 342
TB1
TB3
FIGURE 2-8 Terminal Block Designations for Analyzers with “B” Prefix Serial Number
7% 785%,',7< 6(1625 +22.83

          
5('
6+,(/'
9,2/(7
:+,7(
*5((1
6+,(/'
<(//2:
%/8(
*5$<
%/$&.
%52:1
7/ 
25$1*(
FIGURE 2-9 Terminal Block Designations for Analyzers with No Letter Prefix Serial Number
Accu4 Low-range Turbidimeter System Rev. 2-300
26
PART TW O - INSTALLATION SECTION 3 - ELECTRICAL CONNECTIONS
Refer to Figure 2-10 or 2-11 and connect the sensor (or in­terconnect) cable wires to appropriate terminals on TB1, matching colors as indicated.
TURBIDITY INTERCONNECT
TB1
BLUE
BLACK
ORANGE
2122 1920 1618
GRAY
BROWN
YELLOW
VIOLET
WHITE
GREEN
RED
131415 12
FIGURE 2-10 Connecting Model 8320 Low-range Turbidity Sensor
to Analyzers with “B” Prefix Serial Number
FIGURE 2-11 Connecting Model 8320 Low-range Turbidity Sensor
to Analyzers with No Letter Prefix Serial Number
Rev. 2-300 Accu4 Low-range Turbidimeter System
27
PART TW O - INSTALLATION SECTION 3 - ELECTRICAL CONNECTIONS
3.2 Analog Outputs
Two isolated analog outputs (1 and 2) are provided. Each output represents the measured turbidity, and can be set to be 0/4-20 mA.
and earth ground, but not from each other.
configuring the outputs, refer to Part Three, Section 4.3.
Wiring Tip!
cable for connecting the analog outputs. To protect the output signal from EMI/RFI, connect cable shields to:
The grounding strip at bottom of case (5 open holes, Fig. 2-8) for analyzers with “B” prefix serial number.
The ground symbol Terminal 1 on TB1 (Figure 2-9) for analyzers with no letter prefix serial number.
Each 0/4-20 mA output can drive a load of up to 600 ohms.
Output 1: Connect the load to Terminals 2 and 3 on TB1, matching polarity as indicated.
The outputs are isolated from the inputs
For details on
Use high quality, shielded instrumentation
3.3 Relay Outputs
Output 2: Connect the load to Terminals 4 and 5 on TB1, matching polarity as indicated.
NOTE: When using the HART communication option, a
digital signal is encoded onto the 4-20 mA analog Output 1 signal. In a HART SINGLE MODE wiring configuration, Output 1 remains available for normal use. However, in a HART MULTI-DROP wiring con­figuration, Output 1 becomes dedicated to that function and cannot be used. See Part Three, Sec­tion 8 for more HART communication information.
The analyzer is equipped with four electromechanical re­lays. For relay setup details, see Part Three, Section 4.4.
CAUTION:
DO NOT EXCEED THE CONTACT RATING FOR EACH RELAY (5A 115/230 VAC). WHEN SWITCHING LARGER CURRENTS, USE AN AUXILIARY RELAY SWITCHED BY THE ANALYZER RELAY TO EXTEND ANALYZER RELAY LIFE. WHEN USING RELAY OUTPUTS, MAKE SURE THAT LINE POWER WIRING CAN ADEQUATELY CONDUCT THE CURRENT DRAW OF THE SWITCHED LOAD(S).
Accu4 Low-range Turbidimeter System Rev. 2-300
28
Loading...
+ 63 hidden pages