A Member of the Environmental Instrumentation Group
1
HELPFUL IDENTIFIERS
In addition to information on installation and operation, this instruction manual may contain
WARNINGS pertaining to user safety, CAUTIONS regarding possible instrument malfunction,
and NOTES on important, useful operating guidelines.
A WARNING LOOKS LIKE THIS. ITS PURPOSE IS TO WARN YOU
OF THE POTENTIAL FOR PERSONAL INJURY.
A CAUTION LOOKS LIKE THIS. ITS PURPOSE IS TO ALERT YOU
TO POSSIBLE INSTRUMENT MALFUNCTION OR DAMAGE.
☞NOTE: A note looks like this. Its purpose is to alert you to im-
PART ONE - INTRODUCTIONSECTION 1 - GENERAL INFORMATION
PART ONE - INTRODUCTION
SECTION 1
GENERAL INFORMATION
1.1 Instrument Capability
1.2 Product Identification
The Model 677F remote-mount analog flowmeter/transmitter
can be used with any GLI impeller flow sensor. With standard brackets, the Model 677F can be panel, surface, or
horizontal pipe mounted. When ordered with special, optional mounting brackets, the Model 677F can be vertical
pipe mounted.
The Model 677F provides flow rate measurements, and may
be equipped with a 6-digit resetting totalizer or a 7-digit
non-resetting totalizer. In this case, the Model 677F also
provides a reading of total volumetric flow.
The Model 677F transmits a 4-20 mA signal, corresponding
to the measured flow rate, on a simple two-wire hookup.
Any of the following (or similar type) devices can be connected to the Model 677F:
• Data logger
• Recorder
• Remote readout
The instrument serial number is located on the backside of
the display board. Refer to this serial number for convenient
identification should technical assistance be required.
Rev. 1-398Model 677F Remote-mount Analog Flowmeter/Transmitter
5
PART ONE - INTRODUCTIONSECTION 2 - SPECIFICATIONS
SECTION 2
SPECIFICATIONS
2.1 Operational
Display:
Flow Rate............................. 5-1/2 inch (140 mm) dual-scaled meter with
250° meter movement
Total Flow (optional)............. 7-digit non-resetting totalizer or 6-digit reset-
ting totalizer, each with 0.15 inch (4 mm) high
digits
Measuring Range....................... From 0-5 to 0-80,000 GPM (0-18 to
0-300,000 LPM) with ±2% of full scale
accuracy
2.2 Performance
(Electrical, 4-20 mA Output)
2.3 Mechanical
Ambient Conditions .................... 14 to 122°F (-10 to +50°C); 0-90% relative
humidity
Sensor-to-Transmitter
Distance ............................... 2000 ft. (609 m) maximum
Output Transmission Distance ... Limited only by wire resistance and power
supply voltage
Power Requirements.................. 18-36 VDC, 6 watts maximum
Stability...................................... 0.06% of span per 24 hours, non-cumulative
Repeatability.............................. 0.05% of span
Non-linearity ..............................0.06% of span
Temperature Drift ....................... Zero: 0.01% of span per °C;
Span: 0.02% of span per °C
Response Time.......................... Selectable, 2 or 10 seconds (approximately)
to 90% of value upon step change
Load Driving Capacity................ Maximum load (in ohms) is dependent on
power supply voltage. Use the following
equation to calculate the maximum allowable
load:
Load (ohms) = (Power Supply VDC - 5) x 50
Enclosure (optional)................... NEMA 4X; ½ DIN, polycarbonate; surface
mount
Net Weight................................. 1 lb. (0.46 kg) approximately
6