Ginevri OGB PolyCare 4 Service manual

"OGB Poly Care 4
"
( 7670 )
Infant incubator
Service Manual
OGB “Poly Care 4” – Service Manual Rev.02
7670bmt September ‘04 Ginevri S.R.L.
Rev.02 OGB “Poly Care 4” – Service Manual
Dear Customer,
Ginevri thanks you for choosing our firm and the quality of our products. The longstanding Ginevri traditions of professionalism, reliability and availability will once again prove to be your best reward for entrusting us at Ginevri with your neonatal purchases.
Giorgio Ginevri
GINEVRI s.r.l. – Registered office
Via Giacomo Boni - 00162 Roma
This manual must be carefully read by all personnel who install, use or
Warning
maintain these units.
The operation of this equipment in accordance with the instructions contained in the user and service manuals, combined with regular service maintenance - performed with Ginevri original spare parts and consumables - will assure the efficiency of our devices and the long lasting quality of their performance and reliability.
Maintenance and service must only be performed by technicians who have been trained and authorized by Ginevri.
GINEVRI S.r.l.- Customer Care Service
Via Cancelliera 25/b 00041 Cecchina (Roma) – ITALY
Tel. +39 06 93.45.93.31 Fax. +39 06 93.45.93.93 e-mail exportl@ginevri.com
www.ginevri
This manual refers to the enclosed equipment:
S/N_______________________
OGB Poly Care 4
Ginevri S.r.l. 7670bmt September ‘04
iii
Rev.02 OGB “Poly Care 4” – Service Manual
INDEX
1. CALIBRATION ....................................................................................................................................................................................................5
CALIBRATION CHECK ........................................................................................................................................................5
TEMPERATURE PROBES.....................................................................................................................................................5
OXYGEN PROBE ...................................................................................................................................................................6
HUMIDITY PROBE................................................................................................................................................................6
AIR 1 PROBE CALIBRATION ..............................................................................................................................................6
AIR 2 PROBE CALIBRATION ..............................................................................................................................................7
MAX TEMP CALIBRATION.................................................................................................................................................7
SKIN CALIBRATION ............................................................................................................................................................7
HUMIDITY CALIBRATION..................................................................................................................................................7
OXYGEN CALIBRATION.....................................................................................................................................................7
2. ELECTRICAL WIRING DIAGRAM ......................................................................................................................................................................8
3. SPARE PARTS.................................................................................................................................................................................................13
3.1. INCUBATOR OGB POLY 4 (CODE 10589A03)............................................................................................................13
3.2. INCUBATOR BODY ASSEMBLY (CODE 10226A04) .................................................................................................15
3.3. INCUBATOR HOOD ASSEMBLY (CODE 11835A08).................................................................................................16
3.3. INCUBATOR HOOD ASSEMBLY (CODE 11835A08).................................................................................................17
3.4. HOOD DOOR ASSEMBLY (CODE 10231B09).............................................................................................................18
3.5. BASE ASSEMBLY (CODE 10106A08) ..........................................................................................................................19
3.6. CONTROL PANEL ASSEMBLY (CODE 10285A04)....................................................................................................21
3.7. CONTROL PANEL - KEYBOARD BLOCK ( CODE 10445A06 ) ................................................................................23
3.8. CONTROL PANEL - POWER BLOCK (CODE 10450A06) ..........................................................................................25
3.9. SERVO STEAM HUMIDIFIER (CODE 7424)................................................................................................................27
4. MAINTENANCE SERVICE ...............................................................................................................................................................................29
5. PROGRAMMED AND PREVENTIVE MAINTENANCE....................................................................................................................................29
6. FINAL CHECK ..................................................................................................................................................................................................30
7. TECHNICAL DATA...........................................................................................................................................................................................31
Ginevri S.r.l. 7670bmt Settembre ‘04
4
Rev.02 OGB “Poly Care 4” – Service Manual
1. CALIBRATION
CALIBRATION CHECK
Check the Temperature, Humidity and Oxygen probes correct operation, at least every three months, in order to make sure of the proper operation of the device and to protect the newborn’s health. The check
of the probes reading is done by using the two Reference Simulators, produced by Ginevri s.r.l., that, in case of malfunctioning, are necessary for the re-alignment of the probes.
The Reference Simulators are optional devices that can be ordered to Ginevri s.r.l., with the part numbers 10487A60 and 10875A60. Perform the following steps to check the probes calibration.
TEMPERATURE PROBES
1. Switch the equipment off and disconnect it from the power supply
2. Disconnect the probe assembly connector from the control panel and replace it with the Reference Simulator
3. Connect the equipment to the power supply and switch it on.
4. Set the reference simulator on AIR 1 at 25,0°C (potentiometer PROBES on A1 and potentiometer TEMP °C on 25°C). The temperature display TEMP°C should show the value 25,0°C ± 0,2°C.
5. Set the reference simulator on AIR 1 at 32,0°C. The temperature display TEMP°C should show the value 32,0°C ± 0,2°C.
6. Set the reference simulator on AIR 1 at 33,0°C. The temperature display TEMP°C should show the value 33,0°C ± 0,2°C.
7. Set the reference simulator on AIR 1 at 36,0°C. The temperature display TEMP°C should show the value 36,0°C ± 0,2°C.
8. Set the reference simulator on AIR 1 at 38,0°C. The temperature display TEMP°C should show the value 38,0°C ± 0,2°C.
9. Set the reference simulator on AIR 1 at 40,0°C. The temperature display TEMP°C should show the value 40,0°C ± 0,2°C.
10. Set the reference simulator on AIR 2 at 25,0°C (potentiometer PROBES on A2 and potentiometer TEMP °C on 25°C).. The temperature display TEMP°C should show the value 25,0°C ± 0,2°C.
11. Set the reference simulator on AIR 2 at 32,0°C. The temperature display TEMP°C should show the value 32,0°C ± 0,2°C
12. Set the reference simulator on AIR 2 at 33,0°C. The temperature display TEMP°C should show the value 33,0°C ± 0,2°C
13. Set the reference simulator on AIR 2 at 36,0°C. The temperature display TEMP°C should show the value 36,0°C ± 0,2°C
14. Set the reference simulator on AIR 2 at 38,0°C. The temperature display TEMP°C should show the value 38,0°C ± 0,2°C
15. Set the reference simulator on AIR 2 at 40,0°C. The temperature display TEMP°C should show the value 40,0°C ± 0,2°C
16. Set the reference simulator on MAX TEMP (potentiometer PROBES and deviator on MAX TEMP) and on VAR (potentiometer TEMP °C on VAR), set the potentiometer VAR on 39,0°C and check that MAX TEMP alarm has not been activated. Set the reference simulator on 40,0°C and check that MAX TEMP alarm is activated.
Ginevri S.r.l. 7670bmt September ‘04
5
OGB “Poly Care 4” – Service Manual Rev.02
OXYGEN PROBE
17 During normal operation, both in AIR and in SKIN mode, connect the reference simulator
oxygen/humidity to the oxigen probe inlet. 18 Adjust the selector of the reference simulator on MEASURE mode. 19 Check that the display % OXIGEN indicates CA blinking. 20 Adjust the selector of the reference simulator on CAL mode and wait for about 30 sec. before doing
the calibration. 21 During the calibration, check that CA on the % OXIGEN display is not blinking anymore. 22 When the calibration is finished, check that the % OXIGEN display shows 21%. 23 Set the reference simulator at 98% and check that the % OXIGEN display shows 98% ± 0%. 24 Set the reference simulator at 21% and check that the % OXIGEN display shows 21% ± 0%.
HUMIDITY PROBE
25 Connect the oxygen/humidity reference simulator to the PROBE ASSEMBLY inlet. 26 Check that the led HUMIDITY probe is switched off. 27 Set the reference simulator at 90%. The % HUMIDITY display should show the value 90% ± 0%on 28 Set the reference simulator at 60%. The % HUMIDITY display should show the value 60% ± 0%. 29 Set the reference simulator at 30%. The % HUMIDITY display should show the value 30% ± 0%.
If any malfunction is detected, the probes that have shown a malfunction have to be calibrated. The calibration procedure can be done by means of the same device used for the calibration check and following the procedure below.
AIR 1 PROBE CALIBRATION
The probes calibration has to be performed as follows: 1 Switch the equipment off and disconnect it from the power supply. 2 Disconnect the probe assembly connector from the control panel and replace it with the Reference
Simulator. 3 Connect the equipment to the power supply and switch it on. 4 Set the Reference Simulator on AIR 1, MUTE, TEMP =25°C. 5 Set the multimeter on mV voltage in cc and connect the negative terminal on pin 2 of J5 and the
positive terminal on TP1 Of the microprocessor board (see related LAYOUT) 6 Adjust the trimmer R29 until the measured value is 1060mV ± 1mV. 7 Disconnect the multimeter and set the reference simulator on TEMP = 36°C. 8 Adjust the trimmer R28 until the AIR display of the display board shows the value 36,0°C ± 0,2°C. 9 Set the TEMP selector of the reference simulator on each value and check the calibration for all of
them, except for 0°C and 40°C, by reading the value shown on the AIR display. Every value must be
hold for at least 10 sec. and the following deviations are admitted:
25,0°C ± 0,2°C
29,0°C ± 0,2°C 33,0°C ± 0,2°C 34,0°C ± 0,2°C 36,0°C ± 0,2°C 38,0°C ± 0,2°C
6
7670bmt Settembre ‘04 Ginevri S.R.L.
Rev.02 OGB “Poly Care 4” – Service Manual
AIR 2 PROBE CALIBRATION
10 Switch the equipment off and disconnect it from the power supply. 11
Disconnect the probe assembly connector from the control panel and replace it with the Reference
Simulator.
12
Connect the equipment to the power supply and switch it on.
13
Set the Reference Simulator on AIR 2, MUTE, TEMP = 36°C.
14
Adjust the trimmer R31 until the AIR display of the display board shows the value 36°C ± 0,2°C.
15 Repeat step 9 with the reference simulator for both Air1 and Air2.
MAX TEMP CALIBRATION
16 Set the Reference Simulator on MAX TEMP (potentiometer PROBES and deviator on MAX TEMP),
TEMP = 40°C. 17 The MAX TEMP 18
Adjust the trimmer PT 101 of the display board until the balance point between activation and
should be active.
disactivation of the alarm is reached. Adjust the trimmer so that the alarm is activated.
19 Set the reference simulator on VAR. Rotate the potentiometer on the reference simulator until the alarm is
activated, without moving the potentiometer set the reference simulator on AIR1. Check that the Air display
shows the value 40.0°C ±0.1°C. Repeat step 18, if needed. 20 Check that at every activation of the alarm MAX TEMP the relay RL2 is disactivated and the led bar showing
the heating percentage indicates zero..
SKIN CALIBRATION
21 Set the Reference Simulator on SKIN, MUTE, TEMP = 36°C. 22
Adjust the trimmer R30 until the SKIN display of the display board shows the value 36,5°C ± 0,2°C.
23 Repeat steps 9 to 15 and steps 21 and 22 with the reference simulator for both Air1 and Air2 (the Skin reading
has to be seen on the Skin display).
HUMIDITY CALIBRATION
24 Connect the oxygen/humidity reference simulator to the proper connectors placed on the control panel. Set the
oxygen/humidity reference simulator on HUM = 0%. Connect the multimeter, select mV cc, with negative terminal on pin 2 and the positive terminal on pin 5 of U12 of the microprocessor board. Adjust the trimmer
PT1 placed on the display board until the value measured by the multimeter is 0mV ± 0,2mV. 25 Set the oxygen/humidity reference simulator on HUM% = 90%. 26 Adjust the trimmer PT1 placed on the display board until the HUM display on the display board shows the
value 90.
27 Check the calibration on the remaining set values HUM = 30% and HUM = 66%. There is no particular
deviation requirement.
OXYGEN CALIBRATION
28 Connect the oxygen reference simulator to the proper connectors placed on the control panel, and set it on
ZERO.
29 Set the multimeter on mV cc and connect the negative terminal on pin 2 of J5 and the positive terminal on pin 6
of U12 of the microprocessor board. Adjust the trimmer PT4 of the display board until the measured voltage is 0mV ± 2mV. If necessary adjust the trimmer PT3 on the display board previously.
30 Set the oxygen reference simulator on OXI% = 21% and adjust PT3 until the value measured by the multimeter
is 980mV ± 5mV.
31 Set the oxygen reference simulator on OXI% = CAL and check that after this value is hold for about 30 sec.
from the CAL indication, the OXI display shows the value 21.
32 Set the oxygen reference simulator on OXI% = 98% and adjust PT3 until the value shown on the OXI display
is 98.
33 Set the oxygen reference simulator on OXI% = 21% and check that the value shown on the OXI display is 21.
If any deviation is present, then repeat steps 28 to 33 until the values 21% e 98% are obtained.
Ginevri S.r.l. 7670bmt September ‘04
7
OGB “Poly Care 4” – Service Manual Rev.02
2. ELECTRICAL WIRING DIAGRAM
8
7670bmt Settembre ‘04 Ginevri S.R.L.
Rev.02 OGB “Poly Care 4” – Service Manual
Ginevri S.r.l. 7670bmt September ‘04
9
OGB “Poly Care 4” – Service Manual Rev.02
10
7670bmt Settembre ‘04 Ginevri S.R.L.
Loading...
+ 22 hidden pages