Gima ZIP-IQ DIGITAL CENTRIFUGE User guide

A
Zip-IQ TT Centrifuge
Instruction Manual
A Lid B Rotor C “Open” Button D Speed Adjustment E “Start/Stop” Button F
(Bottom of unit)
G Time Adjustment H Power Button I LCD Display J Power Jack
Emergency Release Lever
Includes:
• 6-place angled rotor for 7ml-15ml tubes
• 6 black inserts for 5ml-6ml tubes
• 6 white inserts for 3ml-4ml tubes
• 6 green inserts for 0.5ml-2.0ml microtubes
• Hex wrench for rotor nut
H
G
B
I
C
D
F
E
J
Introduction
The LW Scientific Zip-IQ TT centrifuge is a 6-place test tube centrifuge for spinning blood, urine, and other fluids in
0.5ml to 15ml test tubes and microtubes. The unit is very small, conserving counter space in the lab, but also sturdy with suction-cupped feet for stability. The digital controls allow speed and time to be adjusted with the touch of a finger and will show real-time speed and time remaining. The maintenance-free, brushless motor spins at 5,000rpm
(2,500g) producing serum in as little as 5 minutes. Slow the speed to 2,000rpm (400g) to spin urines in 5 minutes. The auto-brake stops the rotor in 10 seconds, and the lid opens automatically upon completion.
Warranty
LW Scientific instruments have a one (1) year limited warranty. This warranty is not valid on normal wear and tear, cosmetic damages caused by chemicals, solvents, and/or cleaning solutions, as well as acts of God.
Please register your product online at: www.LWScientific.com/warranty_form.
Important: Warranty information must be completed within 30 days of purchase.
Installation and Setup
Remove the centrifuge from the shipping container and inspect for any possible shipping damage. If the centri-
1
fuge appears to be damaged, please contact LW Scientific or your distributor immediately.
Read the instruction manual in full before operating. Store the operation instructions in a safe place, easily accessi-
2
ble by the trained staff that will be operating the centrifuge.
Place the centrifuge on a sturdy, level surface. Using the emergency release lever underneath the centrifuge,
3
open the lid. Verify that there are no loose objects or packing material in the tube chamber. DO NOT LOAD TUBES AT THIS TIME.
Using the included hex wrench, verify that the silver rotor screw in the center is tight. Verify that there are no tubes
4
or inserts in the rotor.
Close the lid, ensuring that it clicks and locks.
5
Plug the power adapter into the rear of the unit, and plug the cord into an outlet. Press the POWER button to turn
6
the unit on. The LCD display should light up.
Test the unit: Set the time at 3 minutes, and set the speed at 1,000 rpm. Press the START/STOP button. The unit
7
should come up to speed with no vibration and a smooth and quiet sound. Second, set the speed at 5,000 rpm and run again. If there are unusual vibrations or sounds, stop the unit and contact your distributor or LW Scientific.
Loading and Operation
Spin only balanced loads. Ensure that tubes of similar size and equal weight are placed opposite of each other, or
1
place 3 equal tubes in a triangle as pictured below. Use a water-filled tube as a balance tube if necessary. Proper balancing will improve sample separation and extend the life of the centrifuge. Spinning out of balance loads may break test tubes or damage the centrifuge.
Loading and Operation Continued
Ensure that test tubes are supported from the bottom and not hanging by their caps. Use inserts as follows:
2
3ml-4ml tubes (13mm x 75mm)
White Inserts
These support short, skinny tubes
5ml-6ml tubes (13mm x 100mm)
These support tall, skinny tubes
Black Inserts
7ml-15ml tubes(16mm x100mm - 120mm)
Use the openings in the rotor
No Inserts
Note:
Not hanging by cap
Green Inserts
0.5ml - 2.0ml microtubes
Microtubes ARE designed to hang
by the rim of the tube.
Note:
Hanging by cap
Set the speed and time as recommended on page 4.
3
Press the START/STOP button to run the cycle. The unit will stop on its own, beep, and open the lid upon completion.
4
Note:
Not hanging by cap
Note: Always ensure rotor is secure before each use.
Care, Maintenance, and Troubleshooting
The Zip-IQ motor is designed to be maintenance-free. With proper care, this centrifuge will provide years of service. However, if repairs should be needed, please contact LW Scientific.
Use only quality test tubes that are rated for the g-forces utilized. Lower-quality tubes may fracture and allow
1
contents to leak out of tube.
Never force tubes into the rotor. The rotor was designed to hold the most common sized test tubes.
2
Clean with common laboratory disinfectants regularly. Do not allow moisture to seep into the centrifuge and do
3
not immerse the electrical components in any liquid during the cleaning process.
Because of safety issues with high g-forces in a centrifuge, it is recommended that rotors be inspected monthly for
4
wear and fatigue. If there is any indication of wear, the rotor should be removed from service. Contact LW Scientif­ic for return instructions so the rotor can be evaluated by a technician for repair or replacement. After 2 years of service, it is recommended that rotors be returned to LW Scientific for inspection or replacement.
Following these procedures will ensure safety of lab personnel as well as extend the life of the centrifuge.
Recommended Fluid Speeds and Times
Set the speeds and times as follows, or defer to clinical protocols if different:
Fluid Tube Size Inserts Speeds Times
Blood Large (6ml-15ml) None or Black 5,000rpm 5 minutes Blood Small (3ml-5ml) White 5,000rpm 6 minutes Blood Microtubes (0.5ml-2.0ml) Green 5,000rpm 7 minutes Urine 10ml-15ml urine tubes None 2,000rpm 5 minutes Fecals 10ml-15ml fecal tubes None 1,700rpm 6 minutes
NOTE: G-Force is based on radius and speed. Using inserts for smaller tubes and microtubes reduces radius therefore reduces g-force, hence the spin times must be slightly longer in smaller tubes to achieve clear serum separations. Times and speeds can be adjusted based on visual results and clinical protocols.
Specifications
Speed Range: 300-5,000rpm (increments of 100)
Maximum RCF: 2,500g
Max. Volume: 6, 15ml test tubes
Input Voltage: 100-240v AC; 50-60Hz
Output Voltage: 24v DC, 4 amps
: 53 dBA +/- 3 dBA @ 18”-24”
dBA
Timer: 15 sec - 99 min
Display: LCD Digital
G-Force Chart
G-Force (RCF)
Speed (rpm)
500 25g
1000 101g
1500 226g
No Inserts / Black Inserts
Radius: 9cm Radius: 8cm Radius: 7cm
White Inserts Green Inserts
22g
89g
201g
Product Dimensions: Height: 6.2” (158 mm) Depth: 10.6” (270 mm) Width: 8.3” (212 mm) Weight: 9.9 lbs (4.5 kg)
Boxed Dimensions: Height: 8.75” (203.2mm) Length: 11” (279.4mm) Width: 15” (381mm) Weight: 12.25 lbs (5.56kg)
20g
78g
176g
1700
2000 402g 313g358g
2500
3000
3500
4000
4500
5000
290g
629g
906g
1233g
1610g
2038g
2516g
258g
559g
805g
1096g
1431g
1811g
2236g
226g
489g
704g
959g
1252g
1585g
1957g
Fecals - 6 minutes
Urine - 5 minutes
Blood - 5 to 7 minutes
MKT-7.5.3-L-204 | Rev 2