How to reach us .......................................................................................................................................... 10
Warranty Period ....................................................................................................................................... 11
Main features of the product ....................................................................................................................... 16
Program composition .................................................................................................................................. 16
System Requirement ................................................................................................................................... 19
1.4 System Preference ................................................................................................................... 19
1.4 System Preference ................................................................................................................... 20
Basic Design of FC Central II .................................................................................................................... 45
Main Menu ................................................................................................................................................. 45
Layout and Display ..................................................................................................................................... 46
Main Menu & Client Window .................................................................................................................... 46
1.3. Operation of Main Menu ....................................................................................................... 47
Names of Menu .......................................................................................................................................... 47
Previous / Next ........................................................................................................................................... 47
Date/Time Information ............................................................................................................................... 48
1.4. Channel layout and display ................................................................................................... 48
Previous / Next ........................................................................................................................................... 51
Basic structure of Client Window............................................................................................................... 51
Device Number ........................................................................................................................................... 52
Patient information ..................................................................................................................................... 52
Record Time ............................................................................................................................................... 53
Client Window Menu ................................................................................................................................. 53
Measured Time ........................................................................................................................................... 53
Color Tab .................................................................................................................................................... 59
System Tab ................................................................................................................................................. 61
User Setting ................................................................ ................................................................ ................ 63
2.3. How to use Client Window .................................................................................................... 66
Client Window Preference and synchronization ........................................................................................ 78
Client Window Display information .......................................................................................................... 78
Client Window Patient information ............................................................................................................ 79
Client Window Alarm information ............................................................................................................. 80
Client Window Equipment information ..................................................................................................... 81
Close Client Window ................................................................................................................................. 82
2.4. Parameter Trend Screen structure & how to use. ............................................................... 82
Adjusting screen output of Parameter Trend .............................................................................................. 82
Indicating the whole Parameter Trend ........................................................................................................ 83
Hide the indication of Parameter Trend Data ............................................................................................. 84
Parameter Value Screen area ...................................................................................................................... 86
Client Window - Parameter Value .............................................................................................................. 87
FC Reviewer II ..................................................................................................................... 88
1. Execute and Close FC Reviewer II .................................................................................. 89
1.1. Execute FC Reviewer II ......................................................................................................... 89
Use of FC Reviewer II ................................................................................................................................ 89
Execute FC Reviewer II ............................................................................................................................. 89
1.2. Close FC Reviewer II ............................................................................................................. 90
Close FC Reviewer II ................................................................................................................................. 90
2. FC Reviewer II Program Structure ................................................................................. 91
2.1. FC Reviewer II Program Structure ...................................................................................... 91
FC Reviewer II Structure ............................................................................................................................ 91
Main Menu ................................................................................................................................................. 91
Study View ................................................................................................................................................. 93
3. How to use FC Reviewer II Menu ................................................................................... 95
3.1. FC Reviewer II Menu ............................................................................................................. 95
Menu of FC Reviewer II ............................................................................................................................. 95
Close FC Reviewer II ................................................................................................................................. 95
3.2. FC Reviewer II Print preview ............................................................................................... 96
FC Reviewer II Preview settings ................................................................................................................ 96
FC Reviewer II Preview result ................................................................................................................... 97
3.3. Print FC Reviewer II .............................................................................................................. 98
FC Reviewer II Print settings ..................................................................................................................... 98
FC Reviewer II Print result ......................................................................................................................... 99
3.4. Export FC Reviewer II image.............................................................................................. 102
FC Reviewer II image .............................................................................................................................. 102
File contents of FC Reviewer II image ..................................................................................................... 103
3.5. FC Reviewer II Preference .................................................................................................. 104
FC Reviewer II Preference ....................................................................................................................... 104
Color Tab .................................................................................................................................................. 105
System tab ................................................................................................................................................ 107
3.6. Close FC Reviewer II ........................................................................................................... 109
Execute FC Reviewer II ........................................................................................................................... 109
1. Broker System .................................................................................................................. 111
1.1. What is a Broker system? ................................................................................................ 111
1.2. Broker System Structure ................................................................................................. 111
FC Central Server ..................................................................................................................................... 112
FC Central Client(Broker) ........................................................................................................................ 112
2. How to use Broker System .............................................................................................. 112
2.1. Installation / Uninstallation of FC Central II Client ..................................................... 112
2.2. How to use FC Central II Client ..................................................................................... 112
2.3. Difference between FC Central II Server and FC Central II Client ........................... 115
Ver. 1.2
7
FC Central II Manual
NOTE
The specifications and functions described in this instruction manual can be modified without
1. FC Central II Menu Tree ................................................................................................ 118
1.1. FC Central II Main Menu Tree ....................................................................................... 118
1.2. FC Central II Client Menu Tree ..................................................................................... 119
1.3. FC Reviewer II Menu Tree .............................................................................................. 120
2. Administrator authority settings in Windows 7 & 8 ...................................................... 121
3. Q & A .............................................................................................................................. 123
Ver. 1.2
FC Central II Manual
General
1.1 Checkpoints before use
Registered Trademark
How to reach us
Warranty Period
Warning, Caution, Note
General Precaution for service environment
General Precaution on Electrical Safety
Equipment connection, Maintaining and cleaning equipment connection
1.2 Outline
Introduction of FC Central II
Main functions of FC Central II
1.3 Product components
Product components and Optional products
Main product components
1.4 System composition
Wired Environment LAN composition
Wired Environment (Wi-Fi) LAN composition
Wireless Bluetooth composition
9
FC Central II Manual
Product standard or the contents of the manual may be modified without notice, due
to the product improvements.
Copying the whole or the part of the manual is prohibited unless authorized.
Use the product in the country where you purchased.
For Wi-Fi Module, use the product provided by our company.
For Central System, use the product provided by our company or the one with
recommended specification.
1.1 Checkpoints before use
Registered Trademarks
- Microsoft Windows and Windows Logo are trademarks of Microsoft Corporation.
- Mac is a trademark of Apple Corporation.
- HDMI, HDMI Logo and High Definition Multimedia Interface are trademarks of HDMI Licensing
LLC.
- Wi-Fi and Wi-Fi logo are trademarks of Wi-Fi association.
- Fetal Monitor in the manual is a Bionet product(FC700, FC1400, FetalXP, etc.)
- FC700 is a trademark of Bionet.
- FC1400 is a trademark of Bionet.
- FetalXP is a trademark of Bionet.
- FC Central is a trademark of Bionet.
- Trademarks and product names indicated in the manual are the trademarks of each company.
Overseas sales dept. TEL : 82-2-6300-6418 (International)
82-2-6300-6477 (Domestic)
FAX : 82-2-6499-7789
E-mail : sales@ebionet.com
URL : http://www.ebionet.com
Consumables Order
Bionet Co., Ltd. – Sales Department
TEL : 82-2-6300-6418 (International)
82-2-6300-6477 (Domestic)
FAX : 82-2-6499-7789
E-mail : sales@ebionet.com
URL : http://www.ebionet.com
Any Authorized Bionet Distributor
http://www.bionetus.com/locate-a-dealer
A/S and Technical Support
Bionet Co., Ltd. – Customer Service Department
TEL : 82-80-801-6410
FAX : 82-2-6499-7789
E-mail : service@ebionet.com
Internet
URL : http://www.bio2net.com
http://www.ebionet.com
How to reach us
BIONET services are always available to you.
The followings are address and phone number for contacting information, services, and product
supplies.
However, the service is limited for genuine products only. Service may not be provided for
other cases.
* In the event of malfunction or failure, contact us along with the model name, serial number,
and product name of the product.
Ver. 1.2
11
FC Central II Manual
Warranty Period
· This product is manufactured and passed through strict quality control and through inspection.
· Main System Warranty period is 1 year and accessory warranty period is 6 months.
· Other than the product of our company, the product provided by cooperative company is
under the service policy and warranty period of the corporate company.
· We will repair or replace any part of the product that fails due to manufacturer defects.
· This warranty does not apply to any defect caused by improper abuse, misuse or exposure to
poor management.
Ver. 1.2
12
FC Central II Manual
Warning
To inform that it may cause serious injury or death to the patient, property damage, material
losses against the “warning” sign
Caution
To inform that it may cause no harm in life but lead to injury against the “caution” sign
Note
To inform that it is not dangerous but important “note” sign for proper installation, operation, and
maintenance of the equipment.
Warning, Caution, Note
For special emphasis on agreement, terms are defined as listed below in user’s manual. Users
should operate the equipment according to all the warnings and cautions.
Ver. 1.2
13
FC Central II Manual
Avoid placing in an area
exposed to moist.
Do not touch the equipment
with wet hand.
Avoid exposure to direct
sunlight
Avoid placing in an area
where there is a high
variation of temperature.
Operating temperature
ranges from 10(C to
40(C. Operating humidity
ranges from 30% to 85%.
Avoid in the vicinity of
Electric heater.
Avoid placing in an area
where there is an excessive
humidity rise or ventilation
problem.
Avoid placing in an area
where there is an
excessive shock or
vibration.
Avoid placing in an area
where chemicals are
stored or where there is
danger of gas leakage.
Avoid being inserted dust
and especially metal
material into the
equipment
Do not disjoint or
disassemble the equipment.
We take no responsibility for
it.
Power off when the
equipment is not fully
installed.
Otherwise, equipment
could be damaged.
General Precaution on Environment
Do not keep or operate the equipment in the environment listed below.
Ver. 1.2
14
FC Central II Manual
Warning
For the equipment other than FC Central System, please refer to the manual provided and
make sure to be informed of relevant details. This document explains the details related to FC
Central.
Warning
FC Central System uses fixed IP band. When using Router or firewall, Wi-Fi, the connection
status and monitoring function cannot be guaranteed if the relevant AP has a lack of security or
the network system is operated with DHCP. In this case, let the technician for medical
equipment or the administrator make a security setting for the relevant device and separate the
network area used by the device from other devices.
Caution
In hospitals, doctors and patients are exposed to the danger of uncontrollable electric
currents. These electric currents are generated by potential differences between the
equipment and conductive objects that may come into contact with the equipment. Please
make sure that the auxiliary equipment connected to the equipment to solve this problem
satisfies EN60601-1-1:1996.
Equipment connection
Ver. 1.2
15
FC Central II Manual
Functions
Description
Monitoring units
(max : 16 units)
Serial : 1~8 units
Lan : 1~16 units
Supported Device Model
FC700, FC1400 , Fetal XP
Network
Wired : Serial, LAN
Wireless : Bluetooth, Wireless LAN
View Mode
1, 2, 4, 8 and 16 Channel View Mode
Data Storage
Max 72 hours
Alarm
Sound and Display
Communication type
Bidirectional communication with patient monitor
Reviewer
Parameter Trend, 10sec Wave
Print and Report
Parameter Trend
1.2 Outline
Introduction of FC Central II
FCCentral is a software only for fetal monitors which are manufactured by BIONET Co.,Ltd.
FCCentral can be connected to up to 16 units of fetal monitors. This system is able to
monitor and save parameters and waveforms. You can retrieve and register patient
information, you can have various data by using this system also.
* Note : FC Central lite version is limited to connect up to 2 units of fetal monitors.
Specification
Ver. 1.2
16
FC Central II Manual
Warning
When using Central System which is not provided by our company and the accessory that is not
standardized, the system performance cannot be guaranteed. In this case, Central System may
not function properly. Please use the system we have provided and the recommended product.
For accessories that are not standardized, they can be a reason for signal distortion or noise.
Make sure to use genuine accessory provided by us.
1.3 Product Component
Main features of the product
FC Central II Program is software produced by Bionet, and it is used as fetal monitor,
providing 16 fetal monitors maximum. For Maternal version, it provides not only fetal
monitoring but also monitoring mother’s condition. Also, it provides a variety of functions
other than monitoring, such as saving files and images, and review functions such as
measurement data.
Program composition
FC Central II Program is consisted of three programs. They are, FC Central II, which
implements server function, FC Reviewer II, which can analyze and make a judgment on
measurement data, and FC Central II Client(Broker), in which the measurement data of server
may be checked in each work area.
FC Central II Client(Broker) is an optional product.
Ver. 1.2
17
FC Central II Manual
Type
Central system
Description
Common
PC with FC Central II installed.
Common
FC Central II Software CD
Common
Access Point(AP)
FetalXP
Wireless Network USB Adaptor
FC700
FC1400
Serial to WLAN Converter
FC700
FC1400
Serial to WLAN Converter
Power Supply Wired Environment
FC700
FC1400
Serial to WLAN Converter Bracket
* 2 screws required
( BWH 3X12 Tapping Screw )
Hardware Preference
Ver. 1.2
18
FC Central II Manual
Type
Central System
Description
FC700
FC1400
Serial To WLAN Adaptor
* Setup for Serial to WLAN converter
FC700
FC1400
Serial Wired Environment
* Setup for Serial to WLAN converter
Common
use
USB Lock Key
Caution
For product composition, the components may be different according to the direction of
construction for FC Central II System. According to Wired Environment LAN and wireless Wi-Fi,
Bluetooth environment, components may vary as mentioned. Final composition of the product
should be purchased and constructed from the list above, according to the system construction.
Hardware Preference
Ver. 1.2
19
FC Central II Manual
Minimum Requirement
Recommended Requirement
OS
Windows XP
Windows XP (32bit) or
Windows 7 (32/64bit)
CPU
Core2Duo 1.86GHz
Core2Duo 3.0 GHz or more
RAM
2GB
3GB or more
Monitor
19 inches
24 inches or more
Graph Card
(VGA Card)
VGA RAM 256 MB
1280 x 1024
VGA RAM 512MB or more
1920 x 1080
CD-ROM drive
24X
48X
HDD
320 GB
640 GB more
LAN Speed
10Mbps
100Mbps
USB Port
3
3 or more
System Requirement
Minimum requirement for FCCentral is Windows XP
Installation is available for FCCentral CD or USB memory.
The minimum requirements of installation of FCCentral are as below.
Ver. 1.2
20
FC Central II Manual
FC Central II
Fetal XP
HUB
1.4 System Preference
Wired Environment(LAN)
For Wired Environment LAN, it is supported only in FetalXP equipment. For FC700 and
FC1400, Wired LAN is not supported.
Ver. 1.2
21
FC Central II Manual
FC Central II
AP(Access Point)
Fetal XP
FC1400
FC700
Wi-Fi Module
Wi-Fi Module
Wireless Environment (Wi-fi)
For wireless LAN Environment, FC700, FC1400 and FetalXP are all supported.
Ver. 1.2
22
FC Central II Manual
Installation
1. Installation of FC Central II
1.1 Installation in Window XP & Windows 7
1.2 Installation in Windows 8
2. Uninstallation of FC Central II
2.1 Uninstallation in Window XP & Windows 7
2.2 Uninstallation in Windows 8
3. Execute / Close FC Central II
3.1 Execute FC Central II Program
3.2 Close FC Central II Program
Ver. 1.2
23
FC Central II Manual
1. Installation of FC Central II
1.1. Installation in Window XP & Windows 7
1) If you were using FC Central, you should back up measurement data and DB data
files which manage the data in order to secure the safety of data. After that, stat the
installation
2) The folder where DB files managing the measurement data are located is in
“C:\Program Files\BIONET\FC Central II”. Find “FCCentralII.mdb” here in this
folder or in the subfolder and back it up.
3) For patient measurement data, it is located in the folder designated in user settings.
Check the relevant location and back up the folder where data is saved. This backup
files and DB files can be placed back to their relevant location, after reinstallation.
4) Insert FCCentral Installation CD to the CD Drive.
5) Open CD-ROM drive and run “FCCentral II Install.msi”.
6) You should use the management mode when you install or operate the
software in Windows 7.
7)If window-browser pops up as follows, click “Next”
Ver. 1.2
24
FC Central II Manual
8) Select ‘folder’ where you want to install FCCentral. If you want to install
the program to another folder, click “Browse” and select folder.
9)Install the relevant program in the folder which is set above. Click “Next” button.
Ver. 1.2
25
FC Central II Manual
10) If installation is completed, click “Finish”.
11) If installation is successful, you may see the “FCCentral” and “FCReviewer
II” Icon.
Ver. 1.2
26
FC Central II Manual
12) After finishing the installation, access right should be given in order to make network
communication work in Windows 7 and Windows 8. You can register the program to
be allowed in Windows Firewall.
13) Enter “System and security in Control Panel.
14) Select ‘Windows Firewall in system and security.
15) Select “Allowed programs” in Windows Firewall.
16) Click “Allow other programs” in allowed programs.
17) In ‘Allow other programs’ dialogue, select ‘search’
18) Move to the folder “C:\Program Files\BIONET\FC Central II” where program is
located, and select “FCCentralII“
Ver. 1.2
27
FC Central II Manual
19) Select the program and click “add”
20) If “FCCentral II Application” is on the allowed program list, it is properly registered. If
it is not properly registered, wave may not be seen even if the program is executed
and equipment is interlocked.
Ver. 1.2
28
FC Central II Manual
Caution
Please check for the remaining space, for the data generated while executing the program.
While executing FC Central II program, measured data and image saving process for
interlocking with EMR will occur. In “preference”, “system” of FC Central II program, select the
hard disk folder which has a sufficient amount of remaining space.
21) Register installed program in allowed program list. If you have backup files before
installation, copy it to the same folder and use the system.
1.2. Install in Window 8
1) If you were using FC Central, you should back up measurement data and DB data
files which manage the data in order to secure the safety of data. After that, stat the
installation
2) The folder where DB files managing the measurement data are located is in
“C:\Program Files\BIONET\FC Central II”. Find “FCCentralII.mdb” here in this
folder or in the subfolder and back it up.
3) For patient measurement data, it is located in the folder designated in user settings.
Check the relevant location and back up the folder where data is saved. This backup
files and DB files can be placed back to their relevant location, after reinstallation.
4) Insert FCCentral Installation CD to the CD Drive.
5) When the screen below is shown, click the marked “desktop” area or press “Window
key() + D” and convert the screen into the main screen(wallpaper).
Ver. 1.2
29
FC Central II Manual
6) When the screen is converted into Desktop screen, click “Windows Explorer” marked
in the picture below.
7) Open CD-ROM drive and run “FCCentral Install.msi”.
8) You should use the management mode when you install or operate the
software in Windows 8.
Ver. 1.2
30
FC Central II Manual
9) If window-browser pops up as follows, click “Next”.
Ver. 1.2
31
FC Central II Manual
10) Select ‘folder’ where you want to install FCCentral. If you want to install
the program to another folder, click “Browse” and select folder .
11)Click "Next", then installation will start.
Ver. 1.2
32
FC Central II Manual
12) If installation is completed, click “Finish”.
13) If installation is successful, you may see the “FCCentral” and “FCReviewer
II” Icon.
14) After finishing the installation, access right should be given in order to make network
communication work in Windows 7 and Windows 8. You can register the program to
be allowed in Windows Firewall.
15) Enter “System and security in Control Panel.
16) Select ‘Windows Firewall in system and security.
17) Select “Allowed programs” in Windows Firewall.
18) Click “Allow other programs” in allowed programs.
Ver. 1.2
33
FC Central II Manual
19) In ‘Allow other programs’ dialogue, select ‘search’
20) Move to the folder “C:\Program Files\BIONET\FC Central II” where program is
located, and select “FCCentralII“
Ver. 1.2
34
FC Central II Manual
Caution
Please check for the remaining space, for the data generated while executing the program.
While executing FC Central II program, measured data and image saving process for
interlocking with EMR will occur. In “preference”, “system” of FC Central II program, select the
hard disk folder which has a sufficient amount of remaining space.
21) Select the program and click “add”
22) If “FCCentral II Application” is on the allowed program list, it is properly registered. If
it is not properly registered, wave may not be seen even if the program is executed
and equipment is interlocked.
23) Register installed program in allowed program list. If you have backup files before
installation, copy it to the same folder and use the system.
Ver. 1.2
35
FC Central II Manual
2. Uninstallation of FC Central II
2.1. Uninstallation in Window XP & Windows 7
1) Move to Start->Control Panel->Program addition/deletion
2) Then, select “FCcentral II”.
3) Then, select “Remove” from “installShield Wizard”.
Ver. 1.2
36
FC Central II Manual
4) Click “OK” to delete program completely.
5) Click “Finish”.
Ver. 1.2
37
FC Central II Manual
Caution
Measured data and images saved for interlocking with EMR, which are occurred while
executing FC Central II program, are not deleted. Decide whether you are going to use the data
and remove it manually.
Ver. 1.2
38
FC Central II Manual
2.2. Uninstallation in Window 8
1) For uninstalling the program, press “Window key() + X” to convert into the menu
screen. When the screen below is shown, select “control panel”
2) Move to Start->Control Panel->Program addition/deletion.
5) Then, select “Remove” from “installShield Wizard”.
6) Click “OK” to delete program completely.
Ver. 1.2
40
FC Central II Manual
Caution
Measured data and images saved for interlocking with EMR, which are occurred while
executing FC Central II program, are not deleted. Decide whether you are going to use the data
and remove it manually.
7)Click “Finish”.
Ver. 1.2
41
FC Central II Manual
3. Execute / Close FC Central II
3.1. Execute FC central II
1) Double-click “FC Central II” Icon on Desktop.
2) When starting the program for the first time, the dialogue below might appear due to
the authority problem. Click “Yes”, or “Confirm”
3) When starting the program for the first time, the dialogue might appear with a
message that blocks some features in the firewall for security purposes. Click “Allow
access” button to acquire the authority.
Ver. 1.2
42
FC Central II Manual
Note
FC Central II provides two types of connecting method, Fetal Monitor, and LAN & Serial. Make
a setting for FC Central II and Fetal Monitor by referring to the manual. If the preference is
properly set, the connection will be automatically made and ClientWindow will be formed.
When FC Central II and Fetal Monitor are connected, you can monitor the patient data in real
time. When you want to save data, set the patient information and you should admit it.
4)This is basic design of FC Central II.
Ver. 1.2
43
FC Central II Manual
Caution
Before closing FC Central II program, it is recommended that you should close the data of realtime measurement for the patient. Data being measured might be deleted.
3.2. Close FC Central II
1) If you want to close FC Central II program, click the close button indicated below.
2) When the dialogue for closing the program appears, click “Yes(Y)” button.
3) Program is then closed.
Ver. 1.2
44
FC Central II Manual
How to use
1. Basic functions of the program
1.1 Client Window structure
1.2 How to use Main Menu
1.3 Channel layout and channel scrolling
1.4 Explanation of Client Window Structure
2. How to use Program
2.1 Patient Manager
2.2 Preference
2.3 Patient information measurement management
2.4 Review Program
Ver. 1.2
45
FC Central II Manual
1. Basic functions of the program
1.1. Client Window Structure
Basic Design of FC Central II
Main Menu
The main menu consists of Patient Manager, User Manager, Channel Display for
connected clients, Client Window Setting, and Run “Review Program”.
Client Window
The client window automatically opens when the fetal monitor is connected to
FCCentral . One client window displays one fetal monitor. You can check the
monitor’s parameter and waveform in real time, remotely control the monitor
settings, and print the information in real time
Ver. 1.2
46
FC Central II Manual
1.2. Client & Parameter Window Structure
Layout and Display
Main Menu & Client Window
This indicates the window presented when the connection between the menu of FC Central
and Fetal Monitor is set. For details, refer to “1.1 Client Window structure”
Parameter Window.
When FC Central and Fetal Monitor equipment is connected more than one in 1Ch window
structure, Parameter Window will be presented as many as the number of connected
equipment, providing the function that can monitor connected equipment. When you double-
click each parameter window, relevant window will be opened in Client Window and details
can be seen.
Ver. 1.2
47
FC Central II Manual
Patient manager
Run Review program
8CH
2CH
Next
Date/Time information
Exit program
Configuration setting
16CH
4CH
1CH
Previous
1.3. Operation of Main Menu
Names of Menu
Previous / Next
When more than 2 of equipment are connected in Client Window, it provides the function that
shows you the window indicating each device. In the case of 1ch window, it gives a function
of switching to the window of next equipment. In 2ch window, more than 2 devices should be
connected to use the function. That is, the first window shows the screen of two devices, and
the next window shows the other two. It is the same in 4ch and 8ch.
1, 2, 4, 8, 16 Channel window
To select number of clients (patient monitor) on screen
Patient Manager
To register “patient information”.
Preference Setting
To set up the environment of system.
Run “Review Program”
To call up and open the saved file regarding patient information.
Main View of FC Central II can store data which is an hour after the measurement time, and
can analyze the status of the patient. When you saved the measurement data, the data before
1 hour can be analyzed by opening the saved data through Review Program.
Exit
To quit the program.
Ver. 1.2
48
FC Central II Manual
Date/Time Information
To show the present date and time.
1.4. Channel layout and display
1 CH
The 1-channel view displays up to 16 monitors simultaneously. To check the
details (parameter, waveform, and parameter trend) of one of the monitors, select
the client window of the monitor by double-clicking
2, 4, 8 CH
The 2-channel, 4-channel, and 8-channel views display 2, 4, and 8 monitors,
respectively. The information displayed includes the parameter and parameter
trend but excludes the waveform
Ver. 1.2
49
FC Central II Manual
Ver. 1.2
50
FC Central II Manual
16 CH
The 16-channel view displays up to 16 monitors. The information displayed
includes the parameter but excludes the waveform and parameter trend .
Ver. 1.2
51
FC Central II Manual
Device Number: 1~4
Device Number: 5~8
Next
Previous
Previous / Next
The Previous and Next buttons appear only when the number of fetal monitors
connected to FCCentral exceeds the number of available client windows. By
clicking these buttons, you can check the information of the monitors that are not
displayed on the client window.
[Example : 8 monitors connected in 4-CH display]
1.5. Client Window Structure
Basic structure of Client Window
Client Window is formed one by one, when equipment is connected to FC Central II. The basic
structure of connected window is divided into 4 parts. They are the caption bar area which
indicates equipment number, patient information, alarm message and Client menu, the area
indicating the parameter trend as graphs, the wave graph area which indicates ECG and SpO2,
and the parameter area which shows the figures of parameter.
Ver. 1.2
52
FC Central II Manual
Recording time
Client Window Menu
Parameter Trend
Wave display
Parameter information
Patient Information
Device Number
Alarm message
Recording period
Device Number
In the upper part of Client Window, “Device Number” is presented on the top left side of the
image below. Device Number is an equipment number (or patient’s Bed information) set in the
Fetal Monitor. For equipment number, use 1~16( ~ ) so that it is identical with Bed
number. If the equipment number is not set or out of the range, it is indicated as question
mark.( )
Patient information
On the top of Client Window, “Patient information” is presented on the top left side of the
image below. When it is connected for the first time, patient information is presented as
underlined blanks (------.------), as indicated in the first image. Set the patient
information and connect through Admit( ) button, then the patient information will be
indicated.
Alarm Message
The alarm message is indicated after receiving the message from Fetal Monitor. It helps us to
check the current status of patients as well as equipment.
Ver. 1.2
53
FC Central II Manual
Record Time
When saving connected data of Client Window, it records the time it is saved, that is the
elapsed time.
Client Window Menu
It is the set of menus used in Client Window. It provides various functions such as connection
between equipment and patient, sound volume, image saving, printing and Client Window
preferences.
Parameter Trend
The parameter data received from Fetal Monitor is indicated as graphs and symbols in real
time.
Measured Time
It indicates the measured time of parameter data received from Fetal Monitor.
Graph
The data of ECG and SpO2 received from Fetal Monitor is indicated as graphs in real time.
Parameter
The parameter data received from Fetal Monitor is indicated as figures in real time.
Ver. 1.2
54
FC Central II Manual
2. How to use program
2.1. Patient Manager
Patient Manager structure
When you click “Patient Manager” icon on the Main Menu, a relevant dialogue will appear. The
dialogue for Patient Manager is consisted of searching and releasing conditions for existing
patients in order to manage patients, registering new patients, modifying patient information
and deleting patient information.
Search Condition
The search condition fields include the registered patient’s ID, name, or
registration date and you can search the registered patients using them. (Click the
“Clear” button to clear all search conditions.).
Registering patients
It is used for registering new patient information and managing patients by interlocking with
equipment. Registered information is activated when the patient is called to the hospital and
the measured information is gathered together and managed. You can register patient
Ver. 1.2
55
FC Central II Manual
information by clicking “Add” in the “Patient Manager” dialogue.
If you click add button, “Add patient information” dialogue will appear as indicated in the
image below. You can insert patient’s ID, name, gender, birthday, menstruation date, the
number of pregnancy, the number of childbirth, height, and weight. ID and name are
compulsory, so they must be inserted. Then, click “save” button to register patient information.
Ver. 1.2
56
FC Central II Manual
Modifying patient information
Select the patient you want to modify the patient information in the Patient Manager
and click the Edit button.
In “Modify patient information” dialogue, select what you want to modify and apply addition or
change. But, the ID cannot be modified. Click “save” button and apply the modification.
Deleting a registered patient
It deletes the information in the patient information list when there is unnecessary information
Ver. 1.2
57
FC Central II Manual
or a need to delete it. You can select which patient you want to delete, and click “delete”. It
will delete all of the patient information. Deleted patient information cannot be restored, so be
careful in deleting it.
2.2. Preferencesetting
Preference setting
Preference setting consists of default, color and system.
Default Tab
It is the function where you can set basic units (height, weight, pressure, temperature) used in
the program, alarm sound, how to indicate graphs, and printed area.
Ver. 1.2
58
FC Central II Manual
- Fixing the location of equipment : It is a function that fixes the printed area from
number 1 to 16, for 16 equipment connected to the Central system. Generally, this
function is not used quite often and since it follows the normally connected order. In
order to use this function, you have to set a BED number for each device, and then
you can have a proper operation.
- Basic units setting: It sets up the basic units of height, weight, pressure and
temperature that are used by each country.
- Alarm sound: It is a function that controls Alarm sound that rings out according to the
status of patient as well as equipment. It is used when the Alarm sound function for
the whole system is not used.
- Graph printing control: It controls the location and type of Wave graphs for ECG,
FECG, SpO2. General location is set in this function, but if it is interlocked with
equipment, the window of each patient will be changed according to the measured
Ver. 1.2
59
FC Central II Manual
data from the equipment.
- Automatic recording: If you want to save measured data of patients, the basic
function registers the patient information in Central system patient management
dialogue and you have to select Admit patient information icon, select the patient, and
click “Admit” button. Then you have to click save button in order to save. Automatic
recording function allows you to save the information by clicking Admit patient
information icon in patient measurement data window, selecting the patient, and
clicking “Admit” again.
- Measurement window maintenance: Other than FC700 and FC1400, FetalXP is
designed to share the signal for window management with Central. Ifyou close FetalXP,
Central closes the window of the relevant equipment. If you want to maintain the
measurement window even if you stop the equipment, this option can be used. If you
check on it, it would not close but maintain the window.
If you click “Save changes” button after setting each area, relevant settings will be applied.
Color Tab
Specify the color for display including parameter and waveform screen.
Ver. 1.2
60
FC Central II Manual
- Color Skin : This function allows you to set all the color settings applied in graphs,
grid and events. For user-defined mode, there are “basic colors” and “paper colors”.
For basic colors, it provides the color for basic window appeared first on the screen,
and for paper colors, it provides an identical color with the paper color printed from
the equipment.
- Graph: It provides you to set each color for parameter, wave and background that are
indicated on the Client Window.
- Grid: It sets up the shape and color of grid that is indicated on the Client Window
- Event: It sets up the color of marker and stimulator that are indicated on the Client
Window.
If you click “Save changes” button after setting each area, relevant settings will be applied.
Ver. 1.2
61
FC Central II Manual
System Tab
Specify the system information including COM port, EMR options, language, and
data storage path.
1) Serial setting: If the connection method of FC Central II and Fetal Monitor is Serial,
select relevant COM Port. For Serial, it only provides support for 8 devices. If you click
“Save changes” button after setting each area, relevant settings will be applied.
2) EMR setting: It sets up the data path interlocked with EMR, path for executing the
program, and the type of image. When you save images in Client Window, a Main
program in FC Central II, and in Review Program, they are saved here. Please select
the data path where there is a sufficient amount of remaining space in the hard disk.
Ver. 1.2
62
FC Central II Manual
3) If you click “Save changes” button after setting each area, relevant settings will be
applied..
4) User setting: It provides a function for adding and editing the user information (doctor,
nurse) used in the program. <Details on the next page>
5) System: It sets up where the data measured in the program and patient information
DB are to be saved. When the dialogue appears as indicated on the image below,
select the folder in the hard disk where there is an enough space.
Data path sets up the folder that patient information and datas are saved. If this
Setting is changed, FC Central II program can't get the patient information and data
files. Please set up "Data path" only once at the time that FC Central II is installed. If
this setting will be changed, please copy all data in the old folder into new folder
Ver. 1.2
63
FC Central II Manual
Caution
While operating FC Central II program, saving images for measured data and interlocking with
EMR will occur. In the “System” of “preference” in FC Central II Program, please select the
folder with a sufficient amount of remaining space. If the space is not sufficient, the data may be
lost or the program may not be operated properly.
System group provides a function for setting languages presented in the Program UI
Select your language in the “Language”. For the languages not provided, the
additional work will be needed.
The settings of System group will be applied after restarting the program.
User Setting
“User setting”, one of the items in the system tab, provides a function for adding and editing
the user information (doctor, nurse) used in the program.
1) Search condition: It provides a function for searching for registered users with ID and
name. Insert the relevant information and click “search”, then the result will be
presented on the list below. If you click “Release condition”, it will be initialized to
the default.
2) Register user(addition) : Click “Add” in the dialogue.
Ver. 1.2
64
FC Central II Manual
Insert the information you want to add. The ID cannot be modified, so be careful in
inserting it. Then, click “Save” button.
If you are asked to add user information, click “Yes”
Ver. 1.2
65
FC Central II Manual
You can see that the user information is added on the list.
3) Modifying & deleting User information: It provides a function for modifying or deleting
registered user information. As indicated in the image below, select the user and
click “modify” or “delete”. But, when you modify, the ID cannot be changed.
Ver. 1.2
66
FC Central II Manual
2.3. How to use Client Window
Client Window
Device Number & Patient Information
In the upper part of Client Window, “Device Number” is presented on the top left side of
the image below. Device Number is an equipment number (or patient’s Bed information)
set in the Fetal Monitor. For equipment number, use 1~16( ~ ) so that it is
identical with Bed number. If the equipment number is not set or out of the range, it is
indicated as question mark.( ) When it is connected for the first time, patient
information is presented as underlined blanks (------.------), as indicated in the first
image. Set the patient information and connect through Admit( ) button, then the
patient information will be indicated.
Ver. 1.2
67
FC Central II Manual
Client Window Parameter
Trend Classification
For Trend classification of Client Window, it is classified as indicated in the first image. These
graphs are drawn with the same colors of the parameter area in the second image.
Ver. 1.2
68
FC Central II Manual
View settings of Client Window Parameter trend
If you double-click parameter trend area in Client Window, the dialogue box below appears.
You can check or remove parameters you want to see on the window.
Previous data of Client Window Parameter
If you drag Client Window area with your mouse, the screen presented in real time will stop.
You can drag and check the previous data in this situation.
The mode for checking previous data of parameter trend will be changed into a real time
mode, if the user does not use it for 10 seconds.
Ver. 1.2
69
FC Central II Manual
Icon
Description
File saving is OFF.
(Click the button to turn it ON.)
File saving is ON.
(Click the button to turn it OFF.)
Admit/Discharge button
Alarm is ON.
(Click the button to turn it OFF.)
Alarm is OFF.
(Click the button to turn it ON.)
Button to create image file
Print button
Preference setting/Synchronize button
Exit button
(Activated only when the device is disconnected)
Alarm Message
It receives the alarm occurred in Fetal Monitor, and indicates the relevant data as shown in the
image below. For details, please refer to the equipment manual.
Client WindowMenu configuration
Menu of Client Window performs the functions in the following list. Details will be explained in
the following. Please refer to the description below.
Admit Client Window Patient information
If you want to manage the data you are going to measure after the connection is made between FC
Central and Fetal Monitor, you have to admit with the database. In order to make DB connection,
patient information should exist in DB. If not, click “Patient information” icon as indicated in the
image below, and click “add” to register patient information.
Ver. 1.2
70
FC Central II Manual
Select the patient you want to admit with real time measurement data, and click “Admit” button.
Once it is admitted with the database, Client Window will be changed as the following. On the
top left, the patient information will be printed, and in the Client Window menu on the top right,
“save” button will be made.
Ver. 1.2
71
FC Central II Manual
Discharge Client Window Patient information
If you want to discharge the connection with real time measurement data, select the patient
and click “Discharge” button. Connection with DB will be disconnected, “Save” button will
disappear and only monitoring function will be performed.
Ver. 1.2
72
FC Central II Manual
Caution
When the real time measurement data of Client Window is saved, it will be saved in the “data
path” of system group. Sufficient amount of remaining space in the path is necessary. If the
space is not sufficient, the data may be lost or the program may not be operated properly.
Client Window Measurement data saving On / Off
If you want to save real time measurement data of Client Window, click “Save” button, which
is marked on the image below. In order to save real time measurement data, you have to
register patient information first and admit it. Once it is admitted, “Save” button will be
activated.
Once you click “save” button, data will be saved and the saving time will be indicated, as
shown on the image below. Next to the saving time, the save button is changed into “quit”
button. If you want to stop saving, click the button. The saving will be finished and turned into
the previous mode.
Client Window Alarm sound On / Off
If the alarm figure is received from Fetal Monitor, the relevant data is presented on the caption
bar, and FC Central rings the alarm sound. The function of turning the alarm on and off is
performed by the speaker button, which is marked on the image below. If you click speaker
image button, Alarm sound is turned off.
If you click the speaker button, Alarm sound is turned on.
Ver. 1.2
73
FC Central II Manual
Note
JPG image saved for interlocking with EMR is saved in the designated folder in “Preference
System EMR settings Data Path” EMR interlock is provided to the relevant business in a
JPG form. If the storage is not enough, data may be lost or program may cause malfunction.
Save Client Window Image
It provides a function for saving the measurement data of parameter and trend information on
the current window, and patient information in a JPG form, and it allows you to interlock it with
EMR. If you click the save button marked on the image below, the current window will be
saved.
After it is saved, the dialogue indicates the path where the JPG file is saved.
The file name is formed the patient’s ID and the time when it is saved.
Ver. 1.2
74
FC Central II Manual
< The file saved in a JPG form>
Ver. 1.2
75
FC Central II Manual
Print Client Window
It provides a function for printing measurement data of parameter and trend information on the
current window, and patient information. If you click printer button marked on the image below,
dialogue for printing will appear. According to the selected option, the current window will be
printed.
Options in the printing dialogue
- Cover printing: It decides whether the cover including measurement information and
patient information will be printed.
- Trend graph printing: It decides whether the parameter trend information will be
printed.
- Whole: All measured data(1 hour buffering data) will be printed.
- Range setting: It allows users to set a range of printing. Printing range may be set to
the minute.
- Trend speed: The speed of trend data is set and the degree of precision and amount
of data can be adjusted. The trend speed is set with 10, 20, 30mm/minute.
- Printed direction: You can choose to print either in horizontal or vertical direction.
Ver. 1.2
76
FC Central II Manual
< Cover printing >
Ver. 1.2
77
FC Central II Manual
< Trend graph printing >
Ver. 1.2
78
FC Central II Manual
Client Window Preference and synchronization
It provides a function for adjusting preference as well as synchronizing patient information,
alarm information and equipment information with connected Fetal Monitor. You can use this
by clicking “Client preference” icon marked on the image below.
The dialogue is consisted of “Display information”, “Patient information”, and “Alarm
information” tab.
Client Window Display information
- Connecting information: It indicates the type, version and connection method of
connected equipment.
Ver. 1.2
79
FC Central II Manual
- Alarm sound: It turns the alarm sound On / Off in Client Window.
- Graph: It sets up the wave selection, Gain and trend speed that are to be indicated
on the graph area on the lower part of Client Window.
- Automatic recording time: It reserves the recording time and controls the saving
time. It selects the time range between 5~120 minutes then it is saved until its
relevant time range. This option is initialized to default after reconnection or restart.
- After clicking “Apply”, changes will be applied.
Client Window Patient information
- Patient information: It presents the patient information of Client Window. For this
patient information, the figures will be presented if it is connected with DB through
“Admit patient information” icon. Patient information cannot be modified. It can only
be changed in “Patient Manager” But, the modified items are applied after
disconnecting with Client Window and reconnecting it again.
- User information: Select user information (doctor and nurse) used for saving
measurement data. Please refer to “Preference System User preference” for
registering users.
Ver. 1.2
80
FC Central II Manual
Client Window Alarm information
It provides a function for setting whether you are going to use the alarm for each parameter. It
also sets up the range and level of the alarm. As indicated in the image below, you can check
on parameters you are going to use. It activates the limit and the level which were deactivated.
Ver. 1.2
81
FC Central II Manual
.
Adjust the limiting and level of the alarm, click “Apply” and the figures will be applied.
Client Window Equipment information
It provides a function for numbering equipment used for Fetal Monitor, as well as setting
sound volume information, parameter filter and Gain information. Apply changes according to
the patient’s status and click “Apply”, then changes will be applied.
Ver. 1.2
82
FC Central II Manual
Caution
If you intend to close Client Window, Fetal Monitor equipment should be closed in
advance. If you close the equipment, all the real time measurement data for patients is no
longer saved. If you close equipment and Client Window, please be careful to do so.
Close Client Window
This function is used for closing connection with Fetal Monitor equipment. Equipment and FC
Central automatically sets up the connection, so Fetal Monitor should be ended before this
function is used. Close Fetal Monitor, and click the close button marked on the image below.
2.4. Parameter Trend Screen structure & how to use.
Adjusting screen output of Parameter Trend
Client Window indicates many types of parameter trend. If you want to see carefully by
classifying specific parameters, FC Central provides a function for limiting an indication of
other parameters. You can use this function by double-clicking parameter trend area.
Ver. 1.2
83
FC Central II Manual
Indicating the whole Parameter Trend
When you double click on the status above, the “Trend view settings” will appear. The whole
contents shown on the screen indicate that everything is checked.
Ver. 1.2
84
FC Central II Manual
Hide the indication of Parameter Trend Data
In “Trend view settings”, remove all checked items in parameter. If you click “Everything”, all
parameters will be selected and removed again. The images below indicate that selected
items area applied.
Ver. 1.2
85
FC Central II Manual
Indicating Parameter Trend wave
In FC Central Maternal version, ECG and SpO2 waves are measured and indicated.
By turning On/Off the “Show waves”, it provides a function for controlling what is shown in
the window.
Ver. 1.2
86
FC Central II Manual
Parameter Value Screen area
This is an area where the figures of Trend wave indicated in Client Window area are actually
presented. Indicate the figures of real time measurement for each parameter.
Ver. 1.2
87
FC Central II Manual
Note
The color of Parameter Value and the color of Parameter Trend graph are identical. If you want
to look at the changes of relevant Parameter figure, you can refer to trend graphs of relevant
color.
Note
For detailed implications for Parameter Value, please refer to Fetal Monitor equipment manual.
The meanings of parameter figures and trend graphs are explained in detail.
Client Window - Parameter Value
When more than 8 devices are connected with FC Central and if you want to have a look at
the data in the Client Window at a time, 16ch Window mode should be selected. 16h window
mode does not indicate waves due to the problem in presenting them, and it only indicates
parameter figures.
Ver. 1.2
FC Central II Manual
FC Reviewer II
1. Execute / Close FC Reviewer II
1.1 Execute FC Reviewer II
1.2 Close FC Reviewer II
2. FC Reviewer II Program structure
2.1 FC Reviewer II Program structure
3. How to use FC Reviewer II Menu
3.1 FC Reviewer II Menu
3.2 FC Reviewer II Print preview
3.3 FC Reviewer II Print
3.4 FC Reviewer II Export images
3.5 FC Reviewer II Preferences
3.6 Close FC Reviewer II
89
FC Central II Manual
1. Execute and Close FC Reviewer II
1.1. Execute FC Reviewer II
Use of FC Reviewer II
FC Reviewer is a program used for reexamining and analyzing the patient data measured by
FC Central. In FC Central, data of only 1 hour can be buffered and the previous status of
patient may be analyzed. However, FC Reviewer provides a function for analyzing saved data
in FC Central, which is measured more than an hour. Also, it can save images or print
analyzed data in various options and file them in a chart.
Execute FC Reviewer II
FC Reviewer can be executed in two ways. First is to double-click the image con on Desktop.
Second is to click “Reviewer” Icon in Main Menu of FC Central II. You can click the icon
marked on the image below.
Once the program is executed, the window below appears.
Ver. 1.2
90
FC Central II Manual
1.2. Close FC Reviewer II
Close FC Reviewer II
To close FC Reviewer II program, click “Close” icon marked on the image below, and the
program will be closed.
Ver. 1.2
91
FC Central II Manual
2. FC Reviewer II Program Structure
2.1. FC Reviewer II Program Structure
FC Reviewer II Structure
FC Reviewer II consists of 4 main parts, as indicated in the image below.
Main Menu
Main Menu is a major menu in FC Reviewer II, and it provides various functions such as print
preview, image export, preference, and program closing.
Search View
It presents the database list of patient saved by FC Central II. You can search for the patient
by inserting ID and name into the space indicated in the image below and clicking “Search”
button.
Ver. 1.2
92
FC Central II Manual
In Search View list, if you select the patient and double-click, Patient information and History
information will appear on “Patient View”, which is located on the right side of the image
below.
Ver. 1.2
93
FC Central II Manual
Patient View
It is an area where the database information for the patient selected in Search View is
presented. It indicates patient’s ID, name, birthday, last menstruation date, birth due date,
and the number of pregnancy. If there is a need for modifying the information, you can make
some changes and click the “Modify” button to apply. If you want to know the measured
result for the patient, double-click the list of designated date in the History list indicated on
the image below. Selected list will show its trend data and wave information in Study View
area below.
Study View
Study View indicates the record time for selected items in the History list of Patient View, as
well as the information of doctor and nurse. If necessary, it also provides a function for
taking a memo and the trend data information and wave information will be indicated below.
If you drag from left to right while left-clicking the mouse in Trend information, it allows you
to move from previous data to present data, providing a function for you to analyze.
Ver. 1.2
94
FC Central II Manual
The image below is about the wave information. It indicates ECG and SpO2.
Ver. 1.2
95
FC Central II Manual
3. How to use FC Reviewer II Menu
Print preview
Print
Export images
Preferences
Close program
3.1. FC Reviewer II Menu
Menu of FC Reviewer II
Main Menu is a major menu in FC Reviewer II, and it provides various functions such as print
preview, image export, preference, and program closing.
Print preview
It provides a function for previewing various conditions on printing and setting options on
them.
Print
It provides a function for setting various options on printing and printing documents.
Export images
It provides a function for turning Study information into images and exporting them in a JPG
form, according to the selected options.
Preference
It makes a setting for system preference. You can set various options such as view settings,
language, options and colors for waves of measured data and saving folder for measured
data.
Close FC Reviewer II
It closes the FC Reviewer II program.
Ver. 1.2
96
FC Central II Manual
3.2. FC Reviewer II Print preview
FC Reviewer II Preview settings
It provides a function for previewing measurement data of parameter and trend information on
the current window, and patient information. If you click “print preview” button marked on the
image below, dialogue will appear.
The image below is a dialogue for print preview. Each option can be used as followings.
Options in the printing dialogue
- Cover printing: It decides whether the cover including measurement information and
patient information will be printed.
- Trend graph printing: It decides whether the parameter trend information will be
printed.
- Whole: All loaded data of the patient will be printed.
Ver. 1.2
97
FC Central II Manual
- Range setting: It allows users to set a range of printing. Printing range may be set to
the minute.
- Trend speed: The speed of trend data is set and the degree of precision and amount
of data can be adjusted. The trend speed is set with 10, 20, 30mm/minute.
- Printed direction: You can choose to print either in horizontal or vertical direction.
FC Reviewer II Preview result
The image below is an indication when the printing direction is set as horizontal. The cover
page is printed first and the pages related with Trend will be printed in the next pages. For
vertical direction, only the printing direction is different and the printed content is the same.
In the following section, there is a sample of vertical printing. Please refer to the relevant part.
< Cover printing >
Ver. 1.2
98
FC Central II Manual
< Trend graph printing >
3.3. Print FC Reviewer II
FC Reviewer II Print settings
It provides a function for printing measurement data of parameter and trend information on the
current window, and patient information. If you click printer button marked on the image below,
dialogue for printing will appear. According to the selected option, the current window will be
printed.
Ver. 1.2
99
FC Central II Manual
Options in the printing dialogue
- Cover printing: It decides whether the cover including measurement information and
patient information will be printed.
- Trend graph printing: It decides whether the parameter trend information will be
printed.
- Whole: All loaded data of the patient will be printed.
- Range setting: It allows users to set a range of printing. Printing range may be set to
the minute.
- Trend speed: The speed of trend data is set and the degree of precision and amount
of data can be adjusted. The trend speed is set with 10, 20, 30mm/minute.
- Printed direction: You can choose to print either in horizontal or vertical direction.
FC Reviewer II Print result
The image below is an indication when the printing direction is set as vertical. The cover
page is printed first and the pages related with Trend will be printed in the next pages. For
horizontal direction, only the printing direction is different and the printed content is the same.
In the above section, there is a sample of horizontal printing. Please refer to the relevant part.
Ver. 1.2