alarm sounds, and panel shows word“E3” or “E4”. Total unit shutdown.
Switch immediately to backup oxygen supply. Call supplier immediately.
Ⅵ.SETTING TIMING UP
User can press the timing button to setting the operating time in 0~2 Hours.
When it start working, the display shows “----H elapsed time H ”, it
means that timing function is closed. It will work continually until power
supply cuts off.
Press the “+” button once, operation time increases 1 minute, holding the
button more than 1.5 seconds it will increase continually.
Press the “-” button once, operation time decrease 1 minute, holding the
button more than 1.5 seconds it will decrease continually.
The device will auto turn off and the display will show “00:00H” when
finishing timing. Reset timing function.
Ⅶ.TURNNING OFF
During using the device, user can press down the on/off button, mounted in
control panel, to stop/start oxygen supply.
Take off nasal oxygen cannula from oxygen outlet first, turn off power
switch, and then cut off the power source.
1415
OPERATING & INSTALLATION
MAINTENANCE
Ⅷ.SYMBOLS
Symbol
Description
Alternating current
Class Ⅱ Equipment
OFF (power
disconnection from
the mains)
Breaker
No Smoking
Keep Dry
Symbol
ON (power connection
to the mains)
4
Ⅸ.Remote Control
Use infrared remote controller to operate the device. (Figure 9)
Transmit Window
Plus Timing
On/Off
Description
Consult the manual
Type BF Applied par
Keep Up
Fragile
Stacking Limitation
NOTE: Power should be disconnected before beginning preventive
maintenance on the concentrator. The concentrator is specifically designed
to minimize routine preventive maintenance at intervals of once per year. In
places with high dust or soot levels, maintenance may need to be performed
more often. The following must be performed at a minimum of one year in
service to assure years of additional reliability.
Ⅰ.CLEAN CABINET
WARNING:
Clean the cabinet with a mild household cleaner and non-abrasive cloth or
sponge at least one time a month. Do Not drop any liquid into the seam of
the device.
CUT OFF the power supply first to avoid electrical shock.
DO NOT remove device cabinet.
Ⅱ.CLEAN OR REPLACE FILTER
Please clean or replace the filters in time, it's very important to protect
compressor and to extend the device life.
Disassembly Filter
Remove the filter cover and take out the filter.
Minus Timing
Figure 9: Remote Controller
Note: During remote control operating, the remote controller must be toward
the receiver which is located in the control panel.
Filter Cover
Figure 10: Remove Filter
Filter
1617
MAINTENANCE
TROUBLESHOOTING
Clean Filter
1) Clean the filter with a soft cleaner or wash in warm soapy water and rinse
thoroughly.
2) DRY the filter thoroughly before reinstallation.
3) The filter must clean or replace.
CAUTION:
DO NOT operate the concentrator without installing filter,
or while filter is wet. These actions could permanently
damage the concentrator.
Clean Humidifier
1) Remove the humidifier bottle from the humidifier cap then clean the bottle.
2) Remove the humidifier tube and diffuser then clean them.
Figure 11 Figure 12
*Clean Neublizer
Note: Must clean the neublizer after using.
1) After nebulizing, remove the neublizer from oxygen concentrator. Turn
off the oxygen concentrator, disconnect the hose, remove the cap,
disassembly the neublizer as shown in Figure 8.
2) Put all the nebulizer components to warm water for 15 min. (Add some
vinegar to the warm water, if necessary.)
DO NOT cook or use boiling water to clean the nebulizer components.
3) DRY all the components thoroughly before storage.
TROUBLESHOOTING GUIDE
故障现象
Symptom
Power Interruption
Alarm:
Press the power on
switch on the
control panel,
buzzer alarming,
display not lit,
device not work.
Concentrator is
working, the
operating sound is
normal, the
flowmeter knob
can be adjusted but
no oxygen output
or weak output.
Probable cause
1) Power cord and wall
outlet have poor contact.
2) Wall outlet no power
output.
3) Low power of wall
outlet.
4) Breaker reset button
raised.
5) If the device is still not working, please contact the
supplier.
1) Leakage between
humidifier bottle and cap.
2) The safety valve of
humidifier is opened.
The connection between
oxygen output and
humidifier is leaking.
4) The accessory (nasal
cannual, mask, humidifier,
tubing…., etc.) is leaking.
5) If the phenomenon is still appeared, please contact
the supplier.
Solution
1) Re-insert the power cord
into wall outlet securely.
2) Change to another wall
outlet.
3) Change to another wall
outlet. Do not use any
extension cord.
4) Press down the breaker
reset button.
1) Re-insert the power cord
into wall outlet securely.
2) Change to another wall
outlet.
3) Re-install the humidifier
tubing.
4) Replace the leaked
accessory.
18
19
TROUBLESHOOTINGTROUBLESHOOTING
故障现象故障现象
SymptomSymptom
Concentrator is
working, but green
light, yellow light
illuminate at the
same time.
Concentrator is not
working, the red
light illuminate,
buzzer alarm, the
display show
“LO”.
Concentrator is not
working, the red
light illuminate,
buzzer alarm, the
display show “E1”.
Concentrator is not
working, the red
light illuminate,
buzzer alarm, the
display show “E2”.
Concentrator is not
working, the red
light illuminate,
buzzer alarm, the
display show “E3”.
Probable causeProbable cause
1) 50% ≤ Oxygen
Concentration ≤ 82%.
2) Oxygen flow rate over
the maximum recommend
flow rate: 3L/min(8F-3A,
8F-3AW);
5L/min(8F-5A, 8F-5AW).
3) If the phenomenon is still appeared, the device can be
used but please contact the supplier.
1) Oxygen Concentration
≤ 50%.
2) Oxygen flow rate over
the maximum recommend
flow rate: 3L/min(8F-3A,
8F-3AW);
5L/min(8F-5A, 8F-5AW).
3) If the phenomenon is still appeared, stop using the
device, please contact the supplier immediately.
1) System cycle pressure
is too low.
) If the phenomenon is still appeared, stop using the
device, please contact the supplier immediately.
1) System cycle pressure
is too high.
1) Compressor circuit is
opened.
SolutionSolution
1) Clean or replace the filter.
2) Adjust the flow rate only
under the advice of a
physician.
1) Clean or replace the filter.
2) Turn off and re-turn on
the device. Adjust the flow
rate only under the advice
of a physician.
1) Clean or replace the filter.
1) Stop using the device,
please contact the supplier
immediately.
1) Stop using the device,
please contact the supplier
immediately.
Concentrator is not
working, the red
light illuminate,
buzzer alarm, the
display show “E4”.
The nasal cannula
is fogged with
vapor or full of
liquid drops.
(Remove the vapor
or liquid drops
method: Turn on
the device and
connect the tubing
to the oxygen
outlet. Use your
finger to block and
release the end of
the tubing several
times to remove
the moisture and
liquid.)
1) Compressor circuit is
closed.
1) Poor ventilation around
the around the device, so
operating temperature is
too high.
2) The water which added
to humidifier is too hot.
3) Add too many water
into humidifier.
4) The device stops
suddenly during operating.
5) Accessory is connected
to the device. The tubing
is kinked and the device
stops suddenly.
6) Cooling fan which
mounted inside the device
stop running or rotate
slowly, make the operating
temperature is too hot.
7) If the phenomenon is still appeared, the device can
be used but please contact the supplier.
1) Stop using the device,
please contact the supplier
immediately.
1) Make sure the device is
at least 10 cm away from the
walls, other jam or heater.
2) Add cold water to
humidifier.
3) Water added should
between the maximum
and minimum levels in
humidifier.
4) Turn off the device
immediately. Restart the
device to remove the
moisture and liquid.
5) Smooth the nasal
cannula and tubing.
6) Replace the cooling fan.
Note: Cut off the power
first. And perform by
qualified personnel only.
2021
TROUBLESHOOTING
OTHER ATTENTION ITEMS
故障现象
Symptom
*During nebulizing
operating, the
nebulizing
rate is too small
or not working.
(8F-3AW, 8F-5AW)
Remote control
malfunction.
CAUTION:
If you have all other problems, TURN OFF the concentrator
first, use your reserve oxygen supply, and contact the
supplier immediately.
Probable cause
1) Neublizer components
installed incorrect.
2) Neublizer deformed due
to washing or disinfection.
3) Neublizer output is
blocked.
4) Add too much medicinal
liquid into the nebulizer.
5) If the phenomenon is still appeared, stop using
nebulizer, please contact the supplier.
1) Do not use the remote
controller in the receiving
area.
2) Low battery.
Solution
1) Install the nebulizer
correctly. (Figure 8).
2) Replace the nebulizer.
3) Clean or replace the
nebulizer.
4) Add medicinal liquid
into the nebulizer
according to physician
prescription and not over
the maximum level.
1) Operate the remote
controller in the front
side of the device.
2) Replace the battery in
the remote controller.
Ⅰ. GAS PASS OPERATION SKETCH MAP
Air
Oxygen/Nitrogen
Separation
Oxygen
Product
Tank
Nasal Oxygen
Cannula
Filter
Nitrogen
Pressure
Regulator
Humidifier
Compressor
Molecular
Sieve
Commute Valve
Bacilli
Filter
Oxygen
Outlet
Heat Exchanger
Valve Assembly
Muffler
Flowmeter
Check Valve
2223
OTHER ATTENTION ITEMS
Ⅱ. ELECTRICAL RATIONALE
Display Board
Breaker
~
Fan
Main Board
Compressor
Oxygen Sensor
Exchange
ValveⅠ
Ⅲ.PACKING LIST
1. Oxygen Concentrator: 1 unit
2. User Manual: 1 piece
3. Filter: 1 piece
*4. Nebulizer (8F-3AW, 8F-5AW): 1 set
Ⅳ.DEVICE DISPOSAL
Contact the local city or town offices for proper disposal instructions for
oxygen concentrator.