⚫ Invalid: If “no purple-red band appears in Control Line (C)” and “a blue
band appears in the Control Line (C)”, it indicates that the operation
process is incorrect or the test paper has been damaged. In this case,
please read the instruction manual carefully again and retest with a new
test paper. If the problem persists, please stop using this batch of products
immediately and contact your local supplier.
[Limitations of the testing method]
1. The test results of this product should be combined with other clinical
information and comprehensively judged by physicians, and should not be
used as the only criterion.
2. This product is only used to determine the novel coronavirus
(SARS-CoV-2) antigen in the specimen.
[Product Performance Index]
1. Description
1.1 Appearance
The test card should be clean and complete, free of burrs, damage and
contamination; the material should be firmly attached; the label should be
legible and undamaged. The specimen buffer should be clear and transparent,
free of impurities and floccules.
1.2 Liquid migration speed
The liquid migration speed should not be less than 10 mm/min.
1.3 Film strip width
The width of the film strip of the test card is ≥ 2.5mm.
1.4 Sample treatment solution volume
The volume of the sample treatment solution is not less than the labeled
value.
2. Limit of Detection
2.1 Determination of the Limit of Detection
Through gradient dilution of SARS-CoV-2 recombinant N protein, the result
showed that the positive rate was between 90% and 95% under 1/(2×106)
dilution condition, so the dilution ratio of lowest detection limit was finally
determined to be 1/(2×106), and the corresponding SARS-CoV-2
recombinant N protein concentration was about 1ng/ml through calculation.
LOD virus concentration is 200TCID50/ml.
2.2 Verification of the Limit of Detection
Detect the sensitivity reference substance and the positive rate should not be
less than 90%.
3. Accordance rate of negative reference substance
Test the enterprise negative reference substance and the negative rate should
be 100%.
4. Accordance rate of positive reference substance
Test the enterprise positive reference substance and the positive rate should
be 100%.
5. Repeatability
Test the enterprise repeatable reference substances, and the results should all
be positive and uniform in color.
6. Analysis of specificity
6.1 Cross-reaction: This product does not cross-react with endemic human
coronavirus OC43, endemic human coronavirus HKU1, endemic human
coronavirus NL63, endemic human coronavirus 229E, influenza A virus,
influenza B virus, respiratory syncytial virus, adenovirus, Epstein-Barr virus,
measles virus, cytomegalovirus, rotavirus, norovirus, mumps virus,
varicella-zoster virus, mycoplasma pneumoniae, metapneumovirus,
staphylococcus aureus, staphylococcus epidermidis, pseudomonas
aeruginosa, streptococcus pneumoniae and high concentration of N protein.
6.2 Interfering substances: When mucin concentration ≤5mg/mL and
human whole blood ≤2%, there will be no interference with the test results of
this product; the following drugs have no effect on the test results of this
product under the verification concentration, including quinine, zanamivir,
ribavirin, oseltamivir, peramivir, lopinavir, ritonavir, arbidol, acetaminophen,
acetylsalicylic acid, ibuprofen, levofloxacin, azithromycin, ceftriaxone,
meropenem, tobramycin, histamine hydrochloride, phenylephrine,
oxymetazoline, sodium chloride (containing preservatives), beclomethasone,
dexamethasone, flunisolide, triamcinolone acetonide, budesonide,
mometasone, fluticasone, strepsils (flurbiprofen 8.75mg), and Lozenges
(mint).
7. Clinical performance
A total of 508 clinical specimens based on nucleic acid assay (PCR) were
collected, including 243 positive specimens and 265 negative specimens.
After comparing this product with nucleic acid assay (PCR) through the
collected clinical samples, the results are summarized as follows:
[Precautions]
1. This test card is only suitable for professionals as an in vitro diagnostic aid.
Do not use expired products.
2. Do not refrigerate or use after the expiration date (see packaging bag for
expiration date).
3. The temperature and humidity of the experimental environment should be
avoided to be too high, the reaction temperature should be 15-30°C and the
humidity should be below 70%.
4. The packaging bag contains desiccant, do not take.
5. When testing, please wear protective clothing, gloves and goggles.
6. Please do not use the test card with damaged card bag packaging, unclear
marking or beyond the expiration date.
7. Dispose of used specimens, test cards and other waste in accordance with
relevant local laws and regulations.
8. A test card should be used within 1 hour after it is taken out from the
aluminum foil bag.
9. Users shall take samples according to the instruction manual.
10. Remove the covering layer of double-sided adhesive to prevent liquid
splashing before testing.
11. Do not drop the sample treatment solution into the wrong well
12. In the process of testing, the test card should be placed on a horizontal
table, and it should not be moved.
[Explanation of Symbols]
[Basic Information]
Beijing Lepu Medical Technology Co., Ltd.
Floor 3 Building 7-1, No.37 Chaoqian Road, Changping District,
Beijing 102200, P.R. China
Floor 5 Building 7-1, No.37 Chaoqian Road, Changping District,
Beijing 102200, P.R. China
Tel: +86-10-80123964
Email: lepuservice@lepumedical.com
Web: en.lepumedical.com
Lepu Medical (Europe) Cooperatief U.A.
Abe Lenstra Boulevard 36, 8448 JB, Heerenveen,The Netherlands
Tel: +31-515-573399 Fax: +31-515-760020
[Date of Approval and Revision of the Manual]
Approved on August 24, 2021;
Version number: CE-CG27-In01-A8