1. Wichtige Informationen zur Beachtung vor Inbetriebnahme
2. Sicherheitshinweise und Elektromagnetische Verträglichkeit
3. Zweckbestimmung
4. Befestigung
5. Inbetriebnahme und Funktion
6. Ein und Ausschalten
7. Reinigung und Desinfektion
8. Technische Daten
Table of Contents
1. Important information to be observed before operation
2. Safety information and electromagnetic compatibility
3. Intended use
4. Attachment
5. Operation and function
6. Switching on and off
7. Cleaning and disinfection
8. Technical data
Sommaire
1. Informations importantes à lire attentivement avant la mise en
service
2. Consignes de sécurité et compatibilité électromagnétique
3. Usage prévu
4. Fixation
5. Mise en service et fonctionnement
6. Allumage et mise hors tension
7. Nettoyage et désinfection
8. Fiche technique
3
Page 4
Índice
1. Informaciones importantes que deben tenerse en cuenta antes
del uso
2. Indicaciones sobre la seguridad y la compatibilidad
electromagnética
3. Finalidad de uso
4. Fijación
5. Puesta en marcha y funcionamiento
6. Encendido y apagado
7. Limpieza y desinfección
8. Datos técnicos
одержание
1. ажная информация – читать перед вводом в эксплуатацию
2. нформация по безопасности и электромагнитной совместимости
3. азначение
4. репление
5. вод в эксплуатацию и работа
6. ключение и выключение
7. истка и дезинфекция
8. ехнические данные
Indice
1. Importanti avvertenze da rispettare prima dell'uso
2. Avvertenze di sicurezza e compatibilità elettromagnetica
3. Uso previsto
4. Fissaggio
5. Messa in esercizio e funzionamento
6. Accensione e spegnimento
7. Pulizia e disinfezione
8. Dati tecnici
4
Page 5
1. Wichtige Informationen zur Beachtung vor
Inbetriebnahme
Sie haben ein hochwertiges RIESTER Wandgerät erworben, welches
entsprechend der Richtlinie 93/42 EWG hergestellt wurde und ständig
strengsten Qualitätskontrollen unterliegt.
Bitte lesen Sie diese Gebrauchsanweisung vor Inbetriebnahme sorgfältig durch, und bewahren Sie sie gut auf.
Sollten Sie Fragen haben, stehen wir Ihnen jederzeit gerne zur
Verfügung. Unsere Adresse finden Sie in dieser Gebrauchsanweisung.
Die Adresse unseres Partners erhalten Sie gerne auf Anfrage.
Bitte beachten Sie, das alle in dieser Gebrauchsanweisung beschriebenen Instrumente ausschließlich für die Anwendung durch entsprechend ausgebildete Personen geeignet sind.
Bitte beachten Sie, dass die einwandfreie und sichere Funktion dieses
Gerätes nur mit original Zubehör von Riester gewährleistet wird.
2. Sicherheitshinweise und
Elektromagnetische Verträglichkeit
Bedeutung des Symbols auf dem Typenschild
Achtung, Begleitpapiere beachten!
Geräte der Schutzklasse II
Anwendungsteil Typ B
Schutzleiteranschluss
Das Gerät erfüllt die Anforderungen für die elektromagnetische
Verträglichkeit. Bitte beachten Sie, dass unter Einfluss ungünstiger
Feldstärken, z.B. bei Betrieb von Funktelefonen oder radiologischen
Instrumenten, Beeinträchtigungen der Funktion nicht auszuschliessen
sind.
Achtung!
Es besteht evtl. die Gefahr der Entzündung von Gasen, wenn das Gerät
in Anwesenheit von brennbaren Gemischen, von Arzneimittel mit Luft
bzw. mit Sauerstoff oder Lachgas betrieben wird!
Nehmen Sie das Gerät niemals auseinander!
Es besteht die Gefahr eines lebensgefährlichen elektrischen Schlages.
Stecken Sie das Gerät vor der Reinigung bzw. Desinfektion aus.
3. Zweckbestimmung
Das in dieser Gebrauchsanweisung beschriebene Wandgerät ri-former®
wurde zum Betrieb mit verschiedenen Instrumentenköpfen und modula
ren Bauteilen zur nichtinvasiven Diagnose hergestellt.
5
-
Page 6
4. Befestigung
a.) Bohranweisung/Bohrplan
Die Bohranweisung und der Bohrplan, liegen separat bei.
Folgen Sie der Bohranweisung, um die Bohrungen an der Wand anzubringen.
b.) Anbringen der Wandplatten
Sobald Sie die Bohrungen angebracht haben, nehmen Sie die mitgelieferten Dübel und stecken Sie sie bis zum Anschlag in die Bohrungen.
Nehmen Sie die Wandplatte und halten Sie sie so an die Wand, dass Sie
die Schrauben durch die Bohrungen an der Wandplatte in die Dübel stecken können. Drehen Sie jetzt mit Hilfe eines Schraubendrehers die
Schrauben, bis zum Anschlag ein.
c.) Befestigung der Diagnosestation
Wenn Sie alle Schrauben bis zum Anschlag eingedreht haben, nehmen
Sie die Diagnosestation und führen Sie die Schraubenköpfe durch die
Öffnung (1). Dann Drücken Sie die Diagnosestation bis sie einrastet
nach unten.
d.) Befestigung des Ausbaumoduls
Verbinden Sie die Diagnosestation und das Ausbaumodul mit Hilfe des
Verbindungskabels.
Um das Verbindungskabel einstecken zu können, entfernen Sie die
Schiebeabdeckung (2) an der Diagnosestation.
Schließen Sie die Gehäuseöffnung des Ausbaumoduls, die nicht benötigt wird, mit der Schiebeabdeckung (2).
Nehmen Sie das Ausbaumodul und führen Sie die Schraubenköpfe
durch die Öffnung (1). Drücken Sie dann das Ausbaumodul nach unten.
Achtung:
Achten Sie darauf, dass sich das Verbindungskabel nicht hinter dem
Ausbaumodul verklemmt.
Schieben Sie das Verbindungskabel in die Aussparung in der Rückwand
des Ausbaumoduls.
5. Inbetriebnahme und Funktion
Inbetriebnahme der Diagnosestation mit
oder ohne Ausbaumodul:
Stecken Sie den Stecker in die Steckdose. Die optionale Uhr fängt an zu
blinken. Mit der linken Taste mit der Markierung HR und der rechten
Taste mit der Markierung MIN, können Sie die örtliche Uhrzeit durch wiederholtes Drücken der Tasten Einstellen.
Entnehmen Sie den Handgriff (5) nach oben aus der Griffhalterung (7)
und bringen Sie den gewünschten Instrumentenkopf an, indem Sie ihn
so aufsetzen, dass die beiden hervorstehenden Führungsnocken des
Handgriffes aufsitzen.
Drücken Sie den Instrumentenkopf leicht auf den Handgriff und drehen
Sie den Handgriff in Richtung Uhrzeigersinn bis zum Anschlag.
Das Abnehmen des Instrumentenkopfes erfolgt durch Drehung entgegen dem Uhrzeigersinn.
6
Page 7
6. Ein- und Ausschalten
Schalten Sie das Gerät ein, indem Sie den Wippenschalter (3) betätigen.
Die grüne Kontrolllampe (4) im Wippenschalter (3) zeigt die
Betriebsbereitschaft des Gerätes an. Der einzelne Handgriff (5) ist automatisch betriebsbereit, sobald er aus der Griffhalterungen (7) entnommen wird.
Durch Drehen des Rings (6) im Uhrzeigersinn schaltet sich das Gerät
ein. Durch Drehen des Rings entgegen dem Uhrzeigersinn bis zum
Anschlag, schaltet sich das Instrument ab.
Der Handgriff (5) wird automatisch durch Einsetzen in die Griffhalterung
(7) abgeschaltet.
6.1Rheostat zur Regulierung
der Lichtintensivität
Anhand des Rheostats ist es möglich, die Lichtintensivität am Handgriff
einzustellen. Je nachdem, wie Sie den geriffelten Ring (6) entgegen dem
oder in Richtung Uhrzeigersinn drehen, ist die Lichtintensität schwächer
oder stärker. Sie haben die Möglichkeit, die gewünschte Lichtintensität
einzustellen und kontinuierlich beizubehalten, indem Sie einfach den
Handgriff (5) über die automatische Ein- und Ausschaltung im Griffhalter
handhaben.
Achtung!
Achten Sie darauf, dass nie mehr als 3 Handgriffe (5) gleichzeitig benutzt
werden!
Wenn Sie mehr als 3 Handgriffe gleichzeitig benutzen, kann es sein,
dass der Trafo im Gerät überlastet wird und abschält.
7. Reinigung bzw. Desinfektion
Die Diagnosestation ri-former®mit Ausbaumodul kann von aussen mit
einem feuchten Tuch gereinigt werden. Sie kann ferner von aussen, bis
auf die Uhrglasabdeckung (8), mit folgenden Desinfektionsmitteln desinfiziert werden:
Aldehyde (Formaldehyd, Glutaraldeyhd, Aldehydabspalter),Tenside oder
Alkohole.
Beachten Sie bei der Anwendung dieser Stoffe unbedingt die
Vorschriften des Herstellers.
Als Hilfsmittel zur Reinigung oder Desinfektion können ein weiches möglichst fusselfreies Tuch oder Wattestäbchen verwendet werden.
Achtung!
Wir empfehlen, das Gerät vor der Reinigung bzw. Desinfektion auszustecken.
Achten Sie bei der Reinigung und Desinfektion darauf, dass niemals
Flüssigkeit in das Innere des Gerätes gelangt!
Sterilisation
Nach geltender Lehrmeinung (Prüfzentrum für Medizinprodukte
Tübingen) ist Sterilisation nur bei operativen Eingriffen vorgeschrieben.
7
Page 8
8. Technische Daten
Modell:Spannungsversorgung
Anschluss:Netzanschluss siehe Hinweis
Ausgang:1 x 3,5 V
Arbeitstemperatur:0° C bis + 40° C
Ort der Aufbewahrung:-5° C bis + 50° C, bis zu 85 %
Abmessungen
Diagnosestation:200 x 180,5 x 75 mm
Ausbaumodul:200 x 100 x 75 mm
Gewicht Diagnosestation:1450 g
Gewicht Ausbaumodul:490 g
Diagnosestation ri-former®
Typenschild Rückseite.
relative Luftfeuchtigkeit
8
Page 9
1. Important information read prior to start-up
You have purchased a high quality RIESTER wall instrument, which has
been manufactured according to the Directive 93/42 EEC and is subject
to the strictest quality controls at all times.
Read these instructions for use carefully before putting the unit into operation and keep them in a safe place.
If you should have any questions, we are available to answer queries at
all times. Our address can be found in these instructions for use.
The address of our sales partner will be given upon request.
Please note that all instruments described in these instructions for use
are only to be used by suitably trained personnel.
The perfect and safe functioning of this instrument is only guaranteed
when original parts and accessories from Riester are used.
2. Safety information and electromagnetic
compatibility
Meaning of the symbol on the model identification plate
Attention, read accompanying papers!
Instruments of protective class II
Application part type B
Protective conductor connection
The instrument satisfies the requirements for electromagnetic compatibility. Please note that under the influence of unfavourable field strengths,
e.g. during the operation of wireless telephones or radiological instruments, adverse effects on function cannot be excluded.
Attention!
There is a possible danger of inflammation of gases, if the instrument is
operated in the presence of inflammatory mixtures or mixtures of pharmaceuticals and air or oxygen or laughing gas!
Never attempt to take the instrument apart!
There is a danger of
Unplug the instrument before cleaning or when disinfecting.
life-threatening electrical shock.
3. Intended use
The wall instrument ri-former®described in these instructions for use was
manufactured for use with various instrument heads and modular components for non-invasive diagnostics.
9
Page 10
4. Attachment
a.) Drilling instructions/drilling plan
The drilling instructions and the drilling plan are enclosed separately.
Follow the drilling instructions in order to drill the holes in the wall.
b.) Attaching the wall mounting plates
After you have drilled the holes, take the plugs supplied and push them
into the holes as far they will go.
Take the wall mounting plate and hold it onto the wall so that the screws
can be pushed through the holes of the mounting plate into the plugs.
Now screw in the screws with a screw driver, as far as they will go.
c.) Attachment of the diagnostic station
When all screws have been screwed in tightly, take the diagnostic station and guide the screw heads through the openings (1). Then press the
diagnostic station downwards until it snaps into place.
d.) Attachment of the extension module
Connect the diagnostic station and the extension module with the help of
the connecting cable.
In order to plug in the connecting cable, remove the sliding cover (2) of
the diagnostic station.
Close the casing opening of the extension module, which is not needed,
with the sliding cover (2).
Take the extension module and guide the screw heads through the openings (1). Then press the extension module downwards.
Attention:
Take care that the connecting cable does not get caught behind the
extension module.
Push the connecting cable into the groove provided on the reverse side
of the extension module.
5. Operation and function
Putting the diagnostic station into service
with or without extension module:
Put the plug into the electrical socket. The optional clock starts to blink.
You can adjust it to local time by repeatedly pressing the keys; with the
left key marked
Move the handle (5) upwards out of the handle holder (7) and attach the
desired instrument head by placing it with the two projecting guide cams
onto the handle. Press the instrument head lightly onto the handle and
turn the handle in a clockwise direction until it stops. Removal of the
instrument head is carried out by turning in a counter-clockwise direction.
HR and the right key marked MIN.
6. Switching on and off
Switch on the instrument by using the rocker switch (3). The green control lamp (4) in the rocker switch (3) indicates that the instrument is ready
to use. Each handle (5) is automatically ready to operate as soon as it is
taken out of the handle holders (7).
By turning the ring (6) in a clockwise direction, the instrument is switched
on. By turning the ring in a counter-clockwise direction until it stops, the
instrument can be switched off.
10
Page 11
The handle (5) is automatically switched off when replaced back into the
handle holder (7).
6.1Rheostat for regulating the light intensity
It is possible to adjust the light intensity on the handle with the rheostat.
Depending on whether the fluted ring (6) is turned in a clockwise or counter-clockwise direction, the light intensity becomes stronger or weaker.
You have the possibility to adjust the light to the desired intensity and to
maintain this continuously by simply operating the handle (5) via the
automatic on/off switch in the handle holder.
Attention!
Make sure that no more than 3 handles (5) are used at the same time! If
more than 3 handles are used at the same time, the transformer in the
instrument may become overloaded and switch itself off.
7. Cleaning and disinfection
The diagnostic station ri-former(r) with extension module can be cleaned
externally with a moist cloth. Furthermore, it can also be disinfected from
the outside, with the exception of the clock glass cover (8), using the following disinfectants:
Aldehyde (formaldehyde, glutaraldeyhde, aldehyde derivatives), surfactants or alcohols.
When using these substances, the manufacturer’s instructions must be
strictly complied with.
Means for cleaning or disinfection may be a soft, possibly lint-free cloth
or Q-tips.
Attention!
We recommend unplugging the instrument before cleaning or disinfection.
Take care while cleaning and disinfecting that no liquid enters inside the
instrument!
Sterilisation
According to the current school of thought (Test Centre for Medical
Devices in Tübingen), sterilisation is only prescribed in the case of ope-
rative procedures.
8. Technical data
Model:Voltage supply
Diagnostic station ri-former®
Connection:Mains supply, see Note
Model identification plate on reverse side.
Outlet:1 x 3,5 V
Working temperature:0° C to + 40° C
Storage location:-5° C to + 50° C, up to 85 % relative
humidity
Dimensions
Diagnostic station:200 x 180,5 x 75 mm
Extension module:200 x 100 x 75 mm
Wight
Diagnostic station:1450 g
Extension module:490 g
11
Page 12
1. Informations importantes à lire
attentivement avant la mise en service
Vous avez fait l'acquisition d'un appareil mural RIESTER haut
de gamme ; cet appareil est fabriqué conformément aux dispositions de
la norme 93/42 CEE et il est soumis à des contrôlesqualité stricts et continuels.
Avant la mise en service, veuillez lire attentivement ce mode d'emploi et
le conserver.
N'hésitez pas à nous contacter si vous avez des questions, nous nous
tenons à votre disposition. Vous trouverez notre adresse dans ce mode
d'emploi. Sur simple demande, nous vous communiquerons l'adresse de
nos partenaires.
Veuillez noter que l'ensemble des instruments décrits dans ce mode
d'emploi sont exclusivement destinés à être utilisés par des personnes
disposant de la formation et des qualifications adéquates.
Nous vous prions de bien vouloir noter que la fonctionnalité irréprochable et sûre de cet instrument ne peut être garantie qu'à la condition que
seuls les accessoires de la maison RIESTER soient utilisés, à l'exception de tous autres.
2. Consignes de s´écurité et compatibilité
électromagnétique
La signification du symbole se trouve sur la plaque
signalétique
Attention! Respecter les instructions des documents fournis!
Appareils de la classe de protection II
Pièce d'utilisation de type B
Raccordement à la terre
L'appareil satisfait aux exigences de compatibilité électromagnétique.
Toutefois, il faudra noter qu'il est impossible d'exclure toute influence
néfaste lorsque l'appareil est utilisé dans un milieu dans lequel règne
une intensité de champ magnétique désavantageuse, comme par exemple lorsque des téléphones sans fil, des téléphones mobiles ou des
instruments de radiologie sont utilisés.
Attention!
Lorsque l'appareil est utilisé en présence de mélanges gazeux inflammables, de médicaments mélangés à de l'oxygène ou à du gaz hilarant
(protoxyde d'azote), il existe un éventuel danger d'in flammation des gaz.
Ne démontez jamais l'appareil !
Il y a risque de choc électrique pouvant mettre la vie en danger.
Débranchez toujours l'appareil avant de le nettoyer ou de le désin fecter.
3. Usage prévu
L'appareil mural ri-former®décrit dans ce mode d'emploi a été fabriqué
pour être utilisé avec différentes têtes d'instrument et différents modules
de diagnostic non invasif.
12
Page 13
4. Fixation
a.) Instructions et gabarit de perçage
Les instructions et le gabarit de perçage sont compris dans la livraison
et se présentent sous la forme d'un document séparé.
Pour percer les trous dans le mur, reportez-vous aux instructions de perçage.
b.) Fixation des plaques murales
Après avoir percé les trous, prenez les chevilles livrées avec le produit
et introduisez-les dans chacun des trous en les enfonçant jusqu'à la
butée.
Prenez la plaque murale et maintenez-la plaquée au mur de façon à ce
que l'on puisse faire passer les vis par les trous (de la plaque) et les visser dans les chevilles. A l'aide d'un tournevis, vissez alors les vis jusqu'à
la butée.
c.) Fixation de la station de diagnostic
Une fois que vous avez serré toutes les vis jusqu'à la butée, prenez la
station de diagnostic et introduisez les têtes de vis dans les orifices prévus à cet effet (1). Appuyez ensuite (vers le bas) sur la station de diagnostic jusqu'à ce que celle-ci s'emboîte.
d.) Fixation du module d’extension
Reliez la station de diagnostic et le module d'extension entre eux à l'aide du c‚ble de raccordement. Pour pouvoir brancher le câble de raccordement, ôtez le capot coulissant (2) de la station de diagnostic.
A l'aide du capot coulissant, fermez l'ouverture du module d'extension
dont vous n'avez pas besoin.
Prenez le module d'extension et positionnez-le de façon à introduire les
têtes de vis dans les orifices prévus à cet effet (1). Poussez ensuite le
module d'extension vers le bas.
Attention:
Veillez à ce que le câble de raccordement ne se coince pas derrière le
module d'extension.
Faites glisser le câble de raccordement dans le logement prévu à cet
effet : ce logement se trouve dans le module d'extension, du côté du mur.
5. Mise en service et fonctionnement
Mise en service de la station de diagnostic,
avec ou sans module d’extension
Mise en service de la station de diagnostic, avec ou sans module d'extension
Enfoncez la fiche dans la prise de courant. L'horloge (en option) se met
à clignoter. Réglez l'horloge à l'heure locale. Pour cela, appuyez plusieurs fois sur la touche gauche portant la mention HR pour régler les
heures et sur la touche de droite portant la mention MIN pour régler les
minutes.
Soulevez la poignée (5) pour la retirer de son logement (7), et montez la
tête d'instrument souhaitée de façon à ce que les deux cames de guidage du manche (ces dernières dépassent légèrement) reposent sur le
support. Appuyez légèrement la tête d'instrument sur le manche, puis
faites pivoter le manche dans le sens de aiguilles d'une montre jusqu'à
la butée.
Pour enlever la tête d'instrument, faites pivoter celle-ci dans le sens
inverse de celui des aiguilles d'une montre.
13
Page 14
6. Allumage et mise hors tension
Allumez l'appareil à l'aide de l'interrupteur à bascule (3). Le voyant vert
(4) de l'interrupteur à bascule (3) indique l'état opérationnel de l'appareil.
Le manche individuel (5) est automatiquement opérationnel dès qu'on le
retire de son logement (7).
Pour allumer l'appareil, faire pivoter la bague (6) dans le sens des aiguilles d'une montre. Pour éteindre d'appareil, faire tourner la bague dans le
sens inverse de celui de aiguilles d'une montre jusqu'à la butee.
Le manche (5) s'éteint automatiquement dès sa mise en place dans le
logement qui lui est destiné (7).
6.1Le rhéostat, pour une régulation de l'intensité
lumineuse
Grâce au rhéostat, il est possible de régler l'intensité lumineuse depuis
le manche. Pour augmenter l'intensité lumineuse, il vous suffit de faire
tourner la bague cannelée dans le sens inverse de celui des aiguilles
d'une montre, et pour réduire l'intensité lumineuse, vous faites tourner la
bague dans le sens des aiguilles du monde. Vous avez la possibilité de
régler l'intensité lumineuse souhaitée et de conserver ce réglage sur une
durée illimitée : pour cela, il vous suffit tout simplement d'utiliser la fonction allumage et d'arrêt automatique du manche (5).
Attention!
Veillez à ne jamais utiliser plus de trois manches (5) simultanément ! Si
vous utilisez plus de trois manches en même temps, une surcharge du
transformateur de l'appareil peut survenir et entraîner la mise hors tension de l'appareil.
7. Nettoyage ou désinfection
La station de diagnostic ri-former®peut être nettoyée de l'extérieur à l'aide d'un chiffon humide. En outre, une désinfection extérieure (sauf pour
le verre de protection de l'horloge) est possible à l'aide des dés
infectants suivants :
aldéhydes (formaldéhyde, glutaraldéhyde, séparateur aldéhydique),
dérivés tensioactifs ou alcools.
Lors de l'utilisation de ces produits, respectez impérativement les prescriptions du fabricant.
Vous pouvez utiliser comme auxiliaire de nettoyage ou de désinfection
un chiffon peluchant le moins possible ou des cotons-tiges.
Attention!
Il est recommandé de toujours débrancher l'appareil avant de le nettoyer ou de le désinfecter.
Lors du nettoyage et de la désinfection, veillez toujours à ce qu'aucun
liquide ne pénètre à l'intérieur de l'appareil !
Stérilisation
Selon les directives officielles en vigueur (données par le centre
d'examen des produits médicaux de Tübingen, Allemagne (Centre de
contrôle des produits médicaux de
gatoire que pour les interventions chirurgicales.
Tübingen), la stérilisation n'est obli
14
-
Page 15
8. Fiche technique
Modèle:alimentation électrique de la station
de diagnostic ri-former®
Raccordement:reccordement secteur, voir les
information insecrites sur la plaque
signalétique (face arrière)
Sortie:1 x 3,5 V
Température de service:0° C à + 40° C
Lieu de stockage:-5° C à + 50° C,
jusqu'à 85 % d'humidité relative
Dimension de la
station de diagnostic:200 x 180,5 x 75 mm
Dimension de la
modules d'extension:200 x 100 x 75 mm
Poids
station de diagnostic:1450 g
modules d'extension:490 g
15
Page 16
1. Informaciones importantes que deben
tenerse en cuenta antes del uso
Ha adquirido usted un aparato mural RIESTER de alta calidad, cuya
fabricación se rige por la directiva 93/42/CEE y est· sometida constantemente a estrictos controles de calidad.
Lea cuidadosamente estas instrucciones antes de utilizar el termómetro
y consérvelas en lugar un seguro. En caso de dudas, estamos a su disposición en todo momento. Nuestra dirección figura en estas instrucciones. Si lo desea, le facilitaremos con mucho gusto la dirección de nuestro distribuidor.
Tenga en cuenta que todos los instrumentos descritos en estas instrucciones están destinados exclusivamente a su uso por personas debidamente formadas.
El buen funcionamiento y la seguridad de este aparato sólo están garantizados si se utilizan repuestos originales de Riester.
2. Indicaciones sobre seguridad y
compatibilidad electromagnética
Significado del símbolo en la placa de características
Atención a la documentación adjunta
Aparatos de la clase de protección II
Componente de aplicación de tipo B
Conexión de toma de tierra
El aparato cumple todas las exigencias respecto a compatibilidad electromagnética. Sin embargo, en caso de intensidades de campo elevadas
-generadas por ejemplo por teléfonos móviles o instrumentos radiológicos- no pueden descartarse fallos de funcionamiento.
Atención!
Puede existir riesgo de inflamación de gases si el aparato se utiliza en
presencia de mezclas explosivas de medicamentos con aire, oxígeno u
óxido nitroso.
No desmonte el aparato en ningún caso!
Existe riesgo de recibir una descarga eléctrica potencialmente mortal.
Desenchufe el aparato antes de su limpieza o desinfección.
3. Finalidad de uso
El aparato mural ri-former®descrito en estas instrucciones ha sido fabricado para su uso en diagnóstico no invasivo en combinación con dife
rentes cabezales de instrumento y componentes modulares.
16
-
Page 17
4. Fijación
a.) Instrucciones/esquema de taladros
El esquema de taladros y las correspondientes instrucciones se adjuntan por separado. Siga las instrucciones para realizar los taladros en la
pared.
b.) Montaje de las placas de pared
Una vez realizados los taladros, introduzca en ellos hasta el tope los
tacos que se adjuntan.
Sostenga la placa de pared de modo que pueda introducir los tornillos
en los tacos a través de los orificios de la placa.
A continuación, apriete los tornillos hasta el tope utilizando un destornillador.
c.) Fijación de la unidad de diagnóstico
Una vez apretados a fondo todos los tornillos, tome la unidad de diagnóstico y haga pasar las cabezas de los tornillos por la abertura (1).
Después, presione la unidad de diagnóstico hacia abajo hasta que encaje.
d.) Fijación del módulo de ampliación
Conecte la unidad de diagnóstico y el módulo de ampliación con ayuda
del cable correspondiente.
Para poder enchufar el cable de conexión, retire la tapa deslizable (2) de
la unidad de diagnóstico.
Tape la abertura que no se necesite del módulo de ampliación con la
tapa deslizable (2).
Tome el módulo de ampliación y haga pasar las cabezas de los tornillos
por la abertura (1). Después, presione hacia abajo el módulo de ampliación.
Atención!
Preste atención a que el cable de conexión no quede pillado detrás del
módulo de ampliación.
Introduzca el cable de conexión en la muesca existente en la pared
posterior del módulo de ampliación.
5. Puesta en marcha y funcionamiento Puesta
en marcha de la unidad de diagnóstico con
o sin módulo de ampliación
Enchufe la unidad a la red. El reloj (opcional) empezar· a parpadear.
Pulse repetidamente los botones izquierdo (HR) y derecho (MIN) para
ajustar la hora local.
Tire hacia arriba del mango (5) para sacarlo del soporte (7) y monte el
cabezal de instrumento deseado colocándolo de modo que encajen los
dos salientes del guía del mango. Presione levemente el cabezal contra
el mango y gire el mango en sentido de las agujas del reloj hasta el tope.
Para desmontar el cabezal, gírelo en sentido contrario a las agujas del
reloj.
17
Page 18
6. Encendido y apagado
Encienda el aparato accionando el interruptor basculante (3). El piloto
verde (4) del interruptor (3) indica que el aparato est· listo para funcionar. Cada mango individual (5) está listo para funcionar en cuanto se
extrae del correspondiente soporte (7).
El instrumento se enciende girando el anillo (6) en el sentido de las agujas del reloj. Al girar el anillo hasta el tope en sentido contrario a las agujas del reloj, el aparato se apaga.
El mango (5) se apaga automáticamente al colocarlo en el soporte (7).
6.1Reostato para regulación de la intersidad luminosa
La intensidad luminosa puede ajustarse mediante el reostato del mango.
La luz será más o menos intensa según gire el anillo estriado (6) en sentido de las agujas del reloj o en sentido contrario. Puede usted conservar la intensidad luminosa seleccionada si no mueve el anillo y utiliza el
mecanismo automático de encendido y apagado del mango (5) a través
del soporte.
Atención!
Asegárese de no utilizar nunca más de 3 mangos (5) al mismo tiempo!
Si se emplean más de 3 mangos simultáneamente, el transformador del
aparato puede sufrir una sobrecarga y desconectarse.
7. Limpieza/ desinfección
La unidad de diagnóstico ri-former®con módulo de ampliación puede
limpiarse externamente con un paño húmedo. Además, puede desinfectarse externamente -salvo el vidrio protector del reloj (8)- con los siguientes productos:
Aldehídos (formaldehído, aldehído glutárico, desdoblador de aldehídos),
tensoactivos o alcoholes. Por favor aténgase a las instrucciones del
fabricante cuando utilice estos productos.
Como medio auxiliar para la limpieza o desinfección, puede utilizar un
paño suave que no deje pelusa o bastoncillos de algodón.
Atención!
Recomendamos desenchufar el aparato antes de su limpieza o desinfección.
Asegúrese de que nunca penetre líquido en el aparato durante su limpieza o desinfección!
Esterilización
Según los criterios científicos actuales (Centro de ensayos de productos
sanitarios de Tubingen), la esterilización sólo es necesaria en intervenciones quirúrgicas.
18
Page 19
8. Datos técnicos
Modelo:Fuente de alimentación
Unidad de diagnóstico
ri-former
Conexión:Para los datos de conexión a la
red céase la place de características en la cara posterior
Salida:1 x 3,5 V
Temperatura de funcionamiento:0° C a + 40° C
Lugar de almacenamiento:-5° C a + 50° C, humedad
relatica del aire hasta 85%
Dimensiones
Unidad de diagnóstico:200 x 180,5 x 75 mm
Módulo de ampliación:200 x 100 x 75 mm
Peso
Unidad de diagnóstico:1450 g
Módulo de ampliación:490 g
®
19
Page 20
1. ажная информация – читать перед вводом в
эксплуатацию
ы приобрели высококачественный настенный прибор Riester,
изготовленный в соответствии с требованиями директивы 93/42 Э и
прошедший самый строгий контроль качества.
нимательно ознакомьтесь с инструкцией по использованию прибора
перед вводом в эксплуатацию и сохраните инструкцию.
случае возникновения вопросов с ашей стороны мы в любое время в
ашем распоряжении. аш адрес указан в данной инструкции по
использованию прибора. Адрес нашего партнёра мы готовы сообщить
ам по запросу.
Описанные в данной инструкции по использованию прибора
инструменты предназначены для использования соответствующе
подготовленным персоналом.
есперебойная и надёжная работа данного прибора может быть
гарантирована только при использовании оригинальных приборов и
запасных частей фирмы Riester.
2. нформация по безопасности и электромагнитной
совместимости
начение символа на идентификационной табличке
модуля
нимание, прочитайте указания, содержащиеся в
сопроводительной документации!
риборы класса защиты II
спользование класса B
рисоединение защитного провода
рибор отвечает требованиям к электромагнитной совместимости.
Обратите внимание на то, что под влиянием неблагоприятных факторов
полей, например, при работе радиотелефонов или радиологических
инструментов, не исключено отрицательное воздействие на работу
прибора.
нимание!
озможно возникновение опасности воспламенения газов, если прибор
эксплуатируется в присутствии горючих смесей, медикаментов с
применением воздуха или кислорода, или веселящего газа!
атегорически запрещается разбирать прибор!
уществует опасность опасного для жизни удара электрическим током.
еред чисткой или дезинфекцией отключите прибор.
20
Page 21
3. азначение
Описанный в данной инструкции по использованию прибора настенный
прибор ri-former® изготовлен для работы с различными головками
приборов и модульными элементами для неинвазивной диагностики.
4. онтаж
a.) нструкция по сверлению/схема сверления
нструкция по сверлению и схема сверления прилагаются отдельно.
ледуйте указаниям инструкции по сверлению для выполнения
отверстий в стене.
б.) онтаж пластины настенного крепежа
осле выполнения отверстий возьмите входящие в комплект поставки
дюбели и вставьте их в отверстия до упора.
озьмите пластину настенного крепежа и прислоните её к стене так,
чтобы можно было вставить винты через отверстия пластины в дюбели.
еперь вверните отвёрткой винты до упора.
в.) репление диагностической станции
осле завинчивания всех винтов до упора возьмите диагностическую
станцию и введите головки винтов через отверстия (1). осле этого
нажмите диагностическую станцию вниз до фиксации.
г.) репление дополнительного модуля
оедините диагностическую станцию и дополнительный модуль с
помощью соединительного кабеля.
тобы вставить соединительный кабель, отодвиньте крышку (2) на
диагностической станции.
акройте неиспользуемое отверстие в корпусе дополнительного модуля
отодвижной крышкой (2).
озьмите дополнительный модуль и введите головки винтов через
отверстия (1).
осле этого нажмите дополнительный модуль вниз.
нимание:
Обратите внимание на то, чтобы не заклинило соединительный кабель
за дополнительным модулем.
ставьте соединительный кабель в паз в задней стенке дополнительного
модуля.
5. Эксплуатация и принцип действия вод в
эксплуатацию диагностической станции с
дополнительным модулем или без него:
ставьте штекер в электрическую розетку. асы (если они входят в
комплект) начинают мигать. ри помощи левой кнопки с маркировкой
HR и правой кнопки с маркировкой MIN повторным нажатием кнопок
можно установить местное время. однимите рукоятку (5) вверх из
у прибора так, чтобы сверху
21
ловк
е и поверните ручк
ятк
о
у по
модуля (7) и у
находились оба выступающих направляющих кулачка рукоятки.
легка прижмите го
часовой стрелк
оловка прибора снимается поворотом против часовой стрелки.
становите нужную го
у прибора к рук
ловк
е до упора.
Page 22
6. ключение и выключение
ключите прибор переключателем (3). елёная контрольная лампа (4) в
перключателе (3) указывает на готовность прибора к работе. Отдельная
рукоятка (5) готова к работе автоматически, как только её вынимают из
модуля (7). рибор включается вращением кольца (6) по часовой
стрелке. ращением кольца
против часовой стрелки до упора прибор может быть отключен.
укоятка (5) отключается автоматически при установке обратно в
модуль (7).
6.1 еостат для регулирования интенсивности освещения
помощью реостата можно регулировать интенсивность освещения в
ручке. зависимости от того, вращаете ли вы рифлёное кольцо (6)
против или по часовой стрелке, интенсивность освещения уменьшается
или увеличивается. ы можете настроить нужную интенсивность
освещения и поддерживать её постоянно, просто помещая рукоятку (5) в
модуль, используя автоматический способ включения и выключения.
нимание!
Обратите внимание на то, что одновременно нельзя использовать более
3 рукояток (5)!
ри одновременном использовании более 3 рукояток может произойти
перегрузка и самопроизвольное отключение прибора.
7. истка и дезинфекция
иагностическую станцию ri-former®с дополнительным модулем можно
чистить снаружи влажной тканью. роме того, для её дезинфекции, за
исключением стеклянной крышки часов (8), можно использовать
следующие дезинфицирующие
средства:
альдегиды (формальдегид, глютаральальдегид, производные
альдегидов), поверхностно-активные или спирто-содержащие вещества.
ри использовании этих веществ строго соблюдайте требования
фирмыизготовителя.
качестве средства для чистки или дезинфекции можно использовать
мягкую безворсовую ткань или ватные палочки.
нимание!
ы рекомендуем отключить прибор перед чисткой или дезинфекцией.
ри чистке и дезинфекции следите за тем, чтобы жидкость не попала
внутрь прибора!
терилизация
соответствии с мнением учёных (сследовательский центр продукции
медицинского назначения в юбингене) стерилизация необходима
только в случае проведения оперативных вмешательств.
22
Page 23
8. Teхнические данные
одель: итание от напряжения
иагностическая станция ri-former
одключение: нформацию по подключению к сети см. с
обратной стороны на идентификационной
табличке.
ыход: 1 x 3,5
абочая температура: от 0° C до + 40° C
ранение: от -5° C до + 50° C, относительная влажность
воздуха до 85 %
абариты
иагностическая станция: 200 x 180,5 x 75 мм
ополнительный модуль : 200 x 100 x 75 мм
ес
иагностическая станция: 1450 г
ополнительный модуль: 490 г
®
23
Page 24
1. Importanti avvertenze da rispettare prima
dell'uso
Avete acquistato un prodotto a parete RIESTER di alta qualità, realizzato ai sensi della direttiva 93/42/CEE e sottoposto a rigorosi controlli di
qualità costanti.
Vi preghiamo di leggere attentamente le presenti istruzioni prima di mettere in funzione lo strumento e di conservarle con cura per futuro riferimento. Non esitate a contattarci in caso di dubbi o domande. Il nostro
indirizzo è riportato nelle presenti istruzioni per l'uso. L'indirizzo
del nostro partner potrà essere comunicato su richiesta.
Si ricorda che l'impiego di tutti gli strumenti descritti nel presente libretto
di istruzioni è destinato esclusivamente a personale opportunamente
addestrato.
Il corretto e sicuro funzionamento di questo strumento è garantito esclusivamente in caso d'impiego di accessori originali Riester.
2. Avvertenze di sicurezza e compatibilità
elettromagnetica
Significato del simbolo riportato sulla targhetta:
Attenzione, leggere attentamente la documentazione!
Apparecchi della classe di protezione II
Parte da applicare tipo B
Attacco conduttore di protezione
Lo strumento risponde ai requisiti di compatibilità elettromagnetica.Si
ricorda che, in caso di interferenze con intensità di campo sfavorevoli, ad
esempio per utilizzo di radiotelefoni o strumenti radiologici, non si possono escludere anomalie di funzionamento.
Attenzione!
In certi casi esiste il pericolo di combustione di gas, qualora lo strumento sia azionato in presenza di miscele infiammabili, di farmaci contenenti aria, ossigeno o protossido d'azoto!
Non smontare mai lo strumento! Rischio di folgorazione elettrica
grave pericolo di morte!
Prima di eseguire la pulizia o la disinfezione, staccare sempre lo strumento dalla presa di rete.
con
3. Uso previsto
Lo strumento a parete ri-former®descritto nelle presenti istruzioni per
l'uso è stato realizzato per essere utilizzato con diverse teste strumenti
e componenti modulari per la diagnostica non invasiva.
24
Page 25
4. Fissaggio
a.) Istruzioni /piano di foratura
Le istruzioni e il piano di foratura vengono forniti separatamente. Per
eseguire i fori sulla parete, rispettare le relative istruzioni di foratura.
b.) Installazione delle piastre per montaggio a parete
Subito dopo avere eseguito i fori, inserirvi fino in fondo i tasselli forniti in
dotazione.
Posizionare la piastra di montaggio sulla parete in modo da potere inserire le viti nei tasselli facendole passare attraverso i fori praticati in precedenza. A questo punto, con l'ausilio di un cacciavite avvitare le viti fino
all'arresto.
c.) Fissaggio della stazione diagnostica
Dopo avere eseguito questa operazione, prendere la stazione diagnostica e fare passare le teste delle viti attraverso le aperture (1). Quindi
spingere verso il basso la stazione diagnostica fino all'innesto.
d.) Fissaggio del modulo espandibile Collegare la stazione diagnostica e il modulo espandibile utilizzando il relativo cavo.
Per potere innestare il cavo d'allacciamento, rimuovere la copertura
scorrevole (2) dalla stazione diagnostica.
Con l'apertura scorrevole (2) chiudere l'apertura dell'alloggiamento del
modulo espandibile non utilizzato. Prendere il modulo espandibile e
inserire le teste delle viti attraverso le aperture (1). Quindi spingere il
modulo espandibile verso il basso.
Attenzione:
Controllare che il cavo d'allacciamento non si incastri dietro al modulo
espandibile. Fare scorrere il cavo nella scanalatura posta sulla parete
posteriore del modulo espandibile.
5. Messa in esercizio e funzionamento
Messa in esercizio della stazione diagnostica con o senza modulo espandibile:
Innestare il connettore nella presa. L'orologio opzionale inizia a lam peggiare. Premendo più volte il tasto sinistro contrassegnato da HR e il tasto
destro contrassegnato da MIN, è possibile impostare l'ora locale.
Estrarre il manico (5) dal supporto (7) portandolo verso l'alto e montare
la testa dello strumento desiderato, posizionandola in modo che le due
camme di guida sporgenti del manico si trovino in posizione. Premere
leggermente la testa dello strumento sul manico e ruotare quest'ultimo in
senso orario fino all'arresto. Per smontare la testa, ruotarla in senso
antiorario.
6. Accensione e spegnimento
Accedere l'apparecchio azionando l'interruttore a bilanciere (3). La spia
di controllo verde (4) nell'interruttore a bilanciere (3) indica la funzionalità dell'apparecchio. Il singolo manico (5) è automaticamente pronto al
funzionamento non appena viene estratto dal supporto (7).
Ruotare l'anello (6) in senso orario per accendere l'apparecchio.
Ruotare l'anello in senso antiorario fino all'arresto per spegnere lo strumento. Il manico (5) si spegne automaticamente inserendo il supporto
del manico (7).
25
Page 26
6.1Reostato per la regolazione dell'intensità luminosa
Con il reostato è possibile regolare l'intensità luminosa sul manico. A
seconda del senso di rotazione dell'anello zigrinato (6) (senso orario o
antiorario), l'intensità luminosa sarà più debole o più forte. Esiste la possibilità di regolare e mantenere l'intensità luminosa desiderata, azionando semplicemente l'accensione e lo spegnimento automatico del manico (5) sul supporto.
Attenzione!
Fare attenzione a non utilizzare mai più di 3 manici (5) contemporaneamente! Se si utilizzano più di 3 manici nello stesso tempo, può accadere che il trasformatore dell'apparecchio si sovraccarichi e si spenga.
7. Pulizia e disinfezione
La stazione diagnostica ri-former®con modulo espandibile può essere
pulita esternamente con un panno umido. E' inoltre possibile eseguire la
disinfezione sulla superficie esterna, fino vetro asportabile dell'orologio
(8), con i seguenti disinfettanti:
aldeide (formaldeide, glutaraldeide, separatore per aldeidi), tensioattivi o
alcol.
Attenersi rigorosamente alle prescrizioni del costruttore quando si utilizzano queste sostanze.
Per la pulizia o la disinfezione si possono utilizzare un panno morbido
privo di peli o tamponcini di ovatta.
Attenzione!
Prima di eseguire la pulizia o la disinfezione, raccomandiamo di staccare sempre lo strumento dalla presa di rete. Durante la pulizia e la disinfezione controllare che non penetri liquido all'interno dell'apparecchio!
Sterilizzazione
Secondo un autorevole parere (Centro sperimentale per prodotti medicali di Tübingen), la sterilizzazione è prescritta soltanto in caso di interventi operatori.
8. Dati tecnici
Modello:tensione di alimentazione
Stazione diagnostica ri-former
Allacciamento:per I'allacciamento all rete vedere le
avvertenze retro targhetta
Uscita:1 x 3,5 V
Temperatura di lavoro:0° C a + 40°C
Luogo di conservazione:-5° C a + 50° C,
fino all’85 % di umidatà relative
Dimensioni
Stazione diagnostica:200 x 180,5 x 75 mm
Modulo espandibile:200 x 100 x 75 mm
Peso
Stazione diagnostica:1450 g
Modulo espandibile:490 g
26
®
Page 27
Gebrauchte elektrische und elektronische Geräte sollten nicht in den
normalen Hausmüll gelangen, sondern gemäß nationaler bzw. EURichtlinien separat entsorgt werden.
Used electrical and electronic products are not to be disposed as unsorted municipal waste and are to be collected separately accordingly to
national/EU regulations.
Les dispositifs électriques et électroniques usagés ne doivent pas être
éliminés avec les déchets domestiques non triés et doivent être collectés séparément conformément à la réglementation nationale/européenne en vigueur.
Los productos eléctricos y electrónicos usados no pueden eliminarse
como basura general; deberán desecharse de forma separada de acuerdo con las regulaciones nacionales/UE.
спользованные электрические и электронные изделия нельзя
утилизировать как несортированный городской мусор, их следует
собирать в отдельном месте в соответствии с национальными
правилами и правилами .
Apparecchi elettronici ed elettrici usati nom vanno smaltiti nei rifiuti casalinghi. Questi derono essere smaltiti separatamente attenendosi à le
direttive nazionali risp. direttive UE.
27
Page 28
ri-scope® Otoskop
ri-scope® Ophthalmoskop
ri-scope® Retinoskop (Skiaskop) XL 3,5 V
3.
ri-derma® Dermatoskop XL 3,5 V
3.1
28
4.
2.
3.2
Page 29
ri-scope® F.O. Lampenträger XL 3,5 V
ri-scope® F.O. Nasenspekulum XL 3,5 V
3.
2b
2a
ri-scope® F.O. Zungenspatelhalter XL 3,5 V
ri-scope® Human-Operationsotoskop XL 3,5 V
ohne Trichter
5.
3.
2.
ri-scope® Veterinär-Operationsotoskop XL 3,5 V
ohne Trichter
3.
29
2.
Page 30
ri-scope® Instrumentenköpfe
rrii-
-
scope® L Otoskope
1. Zweckbestimmung
Das in dieser Gebrauchsanweisung beschriebene
RIESTER Otoskop wird zur Beleuchtung und Untersuchung des
Gehörganges in Kombination mit den RIESTER Ohrtrichtern produziert.
2. Aufsetzen und Abnehmen von Ohrtrichtern
Zur Bestückung des Otoskopkopfes können wahlweise EinmalOhtrichter von RIESTER (in blauer Farbe) oder wiederverwendbare
Ohrtrichter von RIESTER (in schwarzer Farbe) gewählt werden.
Die Größe des Ohrtrichters ist hinten am Trichter gekennzeichnet.
Otoskop L1 und L2
Drehen Sie den Trichter in Richtung Uhrzeigersinn bis ein Widerstand
spürbar wird. Um den Trichter abnehmen zu können, drehen Sie den
Trichter gegen den Uhrzeigersinn ab.
Otoskop L3
Setzen Sie den gewählten Trichter auf die verchromte Metallfassung des
Otoskopes bis er spürbar einrastet. Um den Trichter abnehmen zu können, drücken Sie die blaue Auswerfertaste. Der Trichter wird automatisch abgeworfen.
3. Schwenklinse zur Vergrößerung
Die Schwenklinse ist fest mit dem Gerät verbunden und kann um 360°
geschwenkt werden.
4. Einführen
von externen Instrumenten ins Ohr
Wenn Sie externe Instrumente ins Ohr einführen möchten (z.B.
Pinzette), müssen Sie die Schwenklinse (ca. 3-fache Vergrößerung),
welche sich am Otoskopkopf befindet, um 180° verdrehen. Sie können
jetzt die Operationslinse einsetzen.
5. Pneumatischer Test
Um den pneumatischen Test (= eine Untersuchung des Trommelfells)
durchführen zu können, benötigen Sie einen Ball, der im normalen
Lieferumfang nicht enthalten ist, aber zusätzlich bestellt werden kann.
Der Schlauch des Balles wird auf den Anschluss gesteckt. Sie können
nun die notwendige Luftmenge vorsichtig in den Ohrenkanal eingeben.
rrii-
-
scope® L Ophthalmoskope
1. Zweckbestimmung
Das in dieser Gebrauchsanweisung beschriebene
RIESTER Ophthalmoskop wird zur Untersuchung des Auges und des
Augenhintergrundes hergestellt.
2. Linsenrad mit Korrekturlinsen
Die Korrekturlinsen können am Linsenrad eingestellt werden. Es stehen
folgende Korrekturlinsen zur Aus-wahl:
Plus: 1-45 in Einzelschritten
Minus: 1-44 in Einzelschritten
Die Werte können im beleuchteten Sichtfeld abgelesen werden.
Pluswerte werden durch grün, Minuswerte durch rote Zahlen angezeigt.
3. Blenden
Über das Blendenstellrad können folgende Blenden gewählt werden:
Ophthalmoskop L1
Halbmond, kleine/mittlere/große Kreisblende, Fixierstern, Slit und
Rotfreifilter.
Ophthalmoskop L2
Halbmond, kleine/mittlere/große Kreisblende, Fixierstern und Slit.
Ophthalmoskop L3
Halbmond, kleine/mittlere/große Kreisblende, Fixierstern, Slit und Karo.
BlendeFunktion
Kleiner Kreis: zur Reflexminderung bei kleinen
Mittlerer Kreis: Pupillen
und Halbkreis:
Großer Kreis: für normale
Fundusuntersuchungen
Karo:zur topographischen Feststellung
von Netzhautveränderungen
Leuchtspalt:zur Bestimmung von
Niveauunterschieden
Fixierstern:zur Feststellung von zentraler
oder exzentrischer Fixation
4 Filter
Über das Filterrad können zu jeder Blende folgende Filter zugeschaltet
werden:
Ophthalmoskop L1
Instrumentenkopf L1 wird ohne Filterrad geliefert.
(Rotfreifilter ist im Blenderad enthalten)
Ophthalmoskop L2
Rotfreifilter, Blaufilter und Polarisationsfilter.
Ophthalmoskop L3
Rotfreifilter, Blaufilter und Polarisationsfilter.
FilterFunktion
Rotfreifilter:kontrastverstärkend zur Beurteilung
Polarisationsfilter:zur genauen Beurteilung der
feiner Gefäßveränderungen
z.B. Netzhautblutungen
Gewebefarben und zur V
derung von Hornhautreflektionen
ermin-
31
Page 32
laufilter:zur besseren Erkennung von
B
Bei L2 + L3 kann jeder Filter zu jeder Blende hinzugeschaltet werden.
5. Fokussiervorrichtung (nur bei L3)
Durch Drehen des Fokussierrades kann eine schnelle Feineinstellung
des zu betrachtenden Untersuchungsfeldes auf diverse Enfernungen
erreicht werden.
Gefäßanomalien oder Blutungen,
zur Fluoreszenz-Ophthalmologie
ri-derma® Dermatoskop XL 3,5 V
1. Zweckbestimmung
Das in dieser Gebrauchsanweisung beschriebene Dermatoskop riderma® wurde zur Früherkennung von pigmentierten
Hautveränderungen (malignen Melanomen) hergestellt.
2. Fokussierung
Fokussieren Sie die Lupe durch Drehen des Okularringes.
3. Hautaufsätze
Es werden 2 Hautaufsätze mitgeliefert:
1) Mit Skalierung von 0 - 10 mm zur Messung von pigmentierten
Hautveränderungen wie malignen Melanomen.
Artikelnummer 10969
2) Ohne Skalierung
Artikelnummer 10968
Beide Hautaufsätze sind einfach abnehm- und austauschbar.
ri-scope® F.O. Lampenträger XL 3,5 V
1. Zweckbestimmung
Der in dieser Gebrauchsanweisung beschriebene Lampenträger wurde
zur Beleuchtung der Mundhöhle und des Rachenraumes hergestellt.
ri-scope® F.O. Nasenspekulum XL 3,5 V
1. Zweckbestimmung
Das in dieser Gebrauchsanweisung beschriebene Nasenspekulum
wurde zur Beleuchtung und somit zur Untersuchung des Naseninneren
hergestellt.
2. Funktion
Zwei Bedienungsarten sind möglich:
a) Schnellspreizen
Drücken Sie die Stellschraube am Instrumentenkopf mit dem Daumen
nach unten. Bei dieser Einstellung kann die Position der Schenkels des
Spekulums nicht verändert werden.
32
Page 33
b) Individuelles Spreizen
Drehen Sie die Stellschraube in Richtung Uhrzeigersinn bis Sie die
gewünschte Spreizöffnung erreichen.
Die Schenkel schließen sich wieder wenn Sie die Schraube entgegen
dem Uhrzeigersinn drehen
3. Schwenklinse
Am Nasenspekulum befindet sich eine Schwenklinse mit einer ca. 2,5
fachen Vergrößerung, die auf Wunsch einfach herausgezogen bzw. wieder in die dafür vorgesehene Öffnung am Nasenspekulum gesteckt werden kann.
ri-scope® F.O. Zungenspatelhalter XL 3,5 V
1. Zweckbestimmung
Der in dieser Gebrauchsanweisung beschriebene Spatelhalter wurde zur
Untersuchung des Mund- und Rachenraumes in Kombination mit handelsüblichen Holz- und Kunststoffspateln hergestellt.
2. Funktion
Führen Sie einen handelsüblichen Holz- oder Kunststoffspatel in die Öffnung unterhalb des Lichtaustrittes bis zum Anschlag ein. Nach der
Untersuchung kann der Spatel leicht entfernt werden, indem man den
Auswerfer betätigt.
ri-scope® Human-Operationsotoskop XL 3,5 V
ohne Trichter
1. Zweckbestimmung
Das in dieser Gebrauchsanweisung beschriebene RIESTER
Operationsotoskop wurde zur Beleuchtung und Untersuchung des
Gehörganges sowie für kleinere Operationen im Gehörgang produziert.
2. Aufsetzen und Abnehmen von Ohrtrichtern für
Humanmedizin
Setzen Sie den gewünschten Trichter auf die schwarze Halterung am
Operationsotoskop , so auf dass die Aussparung am Trichter in die
Führung in der Halterung passt. Fixieren Sie den Trichter, indem Sie ihn
entgegen dem Uhrzeigersinn drehen.
3. Schwenklinse zur Vergrößerung
Am Operationsotoskop befindet sich eine kleine um 360° schwenkbare
Vergrößerungslinse mit einer ca. 2,5-fachen Vergrößerung.
4. Einführen von externen Instrumenten ins Ohr
Das Operationsotoskop ist so gestaltet, dass problemlos externe
Instrumente ins Ohr eingeführt werden können.
33
Page 34
ri-scope®
Veterinär-Operationsotoskop XL 3,5 V
ohne Trichter
1. Zweckbestimmung
Das in dieser Gebrauchsanweisung beschriebene RIESTER
Operationsotoskop wurde ausschließlich zur Anwendung an Tieren und
somit für die Veterinärmedizin produziert. Es kann zur Beleuchtung und
Untersuchung des Gehörganges sowie für kleinere Operationen im
Gehörgang eingesetzt werden.
2. Aufsetzen und Abnehmen von Ohrtrichtern für
Humanmedizin
Setzen Sie den gewünschten Trichter auf die schwarze Halterung am
Operationsotoskop , so auf dass die Aussparung am Trichter in die
Führung in der Halterung passt. Fixieren Sie den Trichter, indem Sie ihn
entgegen dem Uhrzeigersinn drehen.
3. Schwenklinse zur Vergrößerung
Am Operationsotoskop befindet sich eine kleine um 360° schwenkbare
Vergrößerungslinse mit einer ca. 2,5-fachen Vergrößerung.
Auswechseln der Lampe
Otoskop L1
Nehmen Sie die Trichteraufnahme vom Otoskop ab. Drehen Sie die
Lampe entgegen den Uhrzeigersinn heraus.
Drehen Sie die neue Lampe in Richtung Uhrzeigersinn fest und setzen
Sie die Trichteraufnahme wieder auf.
Otoskope L2, L3, ri-derma®, Lampenträger, Nasenspekulum und
Spatelhalter
Drehen Sie den Instrumentenkopf vom Batteriegriff ab.
Die Lampe befindet sich unten im Instrumentenkopf. Ziehen Sie die
Lampe mittels Daumen und Zeigefinger oder eines geeigneten
Werkzeuges aus dem Instrumentenkopf. Setzen Sie die neue Lampe
fest ein.
Ophthalmoskope
Nehmen Sie den Instrumentenkopf vom Batteriegriff ab.
Die Lampe befindet sich unten im Instrumentenkopf. Entnehmen Sie die
Lampe mittels Daumen und Zeigefinger oder eines geeigneten
Werkzeuges dem Instrumentenkopf. Setzen Sie die neue Lampe fest
ein.
Achtung: Der Stift der Lampe muss in die Führungsnut am
Instrumentenkopf eingeführt werden.
Operationsotokope Veterinär/Human
Drehen Sie die Lampe aus der Fassung im Operations-otoskop und drehen Sie eine neue Lampe wieder fest ein.
Instrumentenköpfe: Retinoskop Strich und Punkt
Nehmen Sie den Instrumentenkopf vom Batteriegriff ab.
Die Lampe befindet sich in einer Hülse unten im Instrumentenkopf.
Entnehmen Sie die Lampe mit Hülse mittels Daumen und Zeigefinger
oder eines geeigneten Werkzeuges dem Instrumentenkopf. Setzen Sie
die neue Lampe fest in die Hülse ein und setzen Sie Hülse mit Lampe
wieder so in den Instrumentenkopf ein, dass der Stift der Lampe in der
Nut am Instrumentenkopf geführt wird.
34
Page 35
Pflegehinweise, Reinigung bzw. Desinfektion
Alle RIESTER ri-scope® Instrumentenköpfe können außen mit einem
feuchten Tuch gereinigt werden. Es kann ferner mit folgenden
Desinfektionsmitteln desinfiziert werden:
Aldehyde (Formaldehyd, Glutaraldeyhd, Aldehydabspalter) oder Tenside
Alle Instrumententeile außer den Schwenklinsen, Lupen und den
Abdeckgläsern können darüberhinaus mit Alkoholen desinfiziert werden.
Beachten Sie bei der Anwendung dieser Stoffe unbedingt die
Vorschriften des Herstellers.
Als Hilfsmittel zur Reinigung oder Desinfektion können ein weiches möglichst fusselfreies Tuch oder Wattestäbchen verwendet werden.
Achtung
Legen Sie die Instrumentenköpfe niemals in Flüssigkeit. Achten Sie
darauf, dass keine Flüssigkeit in das Gehäuseinnere eindringt.
a) Sterilisation
Nach geltender Lehrmeinung (Prüfzentrum für Medizinprodukte
Tübingen) ist Sterilisation nur bei operativen Eingriffen vorgeschrieben.
b) Wiederverwendbare Ohrtrichter
Obwohl, wie unter a) beschrieben, eine Sterilisation nicht notwendig ist,
ist sie trotzdem möglich.
Die Wiederverwendbaren Ohrtrichter können bei 134°C und 10 Minuten
Haltezeit im Dampfsterilisator sterilisiert werden.
Inbetriebnahme der Instrumentenköpfe
Setzen Sie den gewünschten Instrumentenkopf so auf die Aufnahme am
Griffoberteil auf, dass die beiden Aussparungen des Unterteils des
Instrumentenkopfes auf die beiden hervorstehenden Führungsnocken
des Batteriegriffes aufsitzen. Drücken Sie den Instrumentenkopf leicht
auf den Griff und drehen Sie den Griff in Richtung Uhrzeigersinn bis zum
Anschlag. Das Abnehmen des Kopfes erfolgt durch Drehung entgegen
dem Uhrzeigersinn.
Inbetriebnahme der Diebstahlsicherung
Funktion
a
b
Drücken Sie den Instrumentenkopf leicht auf den Griff und drehen Sie
den Griff in Richtung Uhrzeigersinn bis zum Anschlag. Um die
Diebstahlsicherung zu aktivieren, drehen Sie mit Hilfe des
Inbusschlüssels (a) (beim Instrumentenkopf beiliegend) die
Inbusschraube (b) bis zum Anschlag ein. Der Instrumentenkopf kann
nun nicht mehr vom Grif
deaktivieren, muss die Inbusschraube (b) mit Hilfe des Inbusschlüssels
(a) wieder heraus gedreht werden.
Setzen Sie den gewünschten
Instrumentenkopf so auf die Aufnahme
am Griffoberteil auf, dass die beiden
Aussparungen des Unterteils des
Instrumentenkopfes auf die beiden her-
vorstehen den Führungsnocken des
Batteriegriffes aufsitzen.
f entfernt werden. Um die Diebstahlsicherung zu
35
Page 36
ri-scope® instrumen heads
ri-scope® L otoscope
1. Purpose
The RIESTER otoscope described in these Operating Instructions is produced for illumination and examination of the auditory canal in combination with
2 Fitting and removing ear specula
Either RIESTER disposable ear specula (blue colour) or reusable RIESTER ear specula (black colour) can be fitted to the otoscope head.
The size of the ear specula is marked at the back of the speculum.
L1 and L2 otoscopes
Screw the speculum clockwise until noticeable resistance is felt. To
remove the speculum, screw the speculum counter clockwise.
L3 otoscope
Fit the chosen speculum on the chrome-plated metal fixture of the otoscope until it locks into place.
To remove the speculum, press the blue ejection button. The speculum
is automatically ejected.
3. Swivel lens for magnification
The swivel lens is fixed to the device and can be swivelled 360°.
4. Insertion of external instruments into the ear
If you wish to insert external instruments into the ear (e.g. tweezers), you
have to rotate the swivel lens (approx. 3-fold magnification) located on
the otoscope head by 180°. Now you can use the operation lens.
5. Pneumatic test
To perform the pneumatic test (= examination of the eardrum), you require a ball, which is not included in the normal delivery package, but can
be ordered separately. The tube for the ball is attached to the connector.
Now you can carefully insert the necessary volume of air into the ear
canal.
RIESTER ear specula.
ri-scope® L ophthalmoscope
1. Purpose
The RIESTER ophthalmoscope described in these Operating
Instructions is produced for the examination of the eye and the eyeground.
2. Lens wheel with correction lens
The correction lens can be adjusted on the lens wheel. The following correction lenses are available:
Plus: 1-45 in single steps
Minus: 1-44 in single steps
The values can be read off in the illuminated field of view. Plus values
are displayed in green numbers, minus values with red numbers.
36
Page 37
3. Apertures
The following apertures can be selected with the aperture hand-wheel:
L1 ophthalmoscope
Semi-circle, small/medium/large circular aperture, fixation star, slit and
red-free filter.
L2 ophthalmoscope
Semi-circle, small/medium/large circular aperture, fixation star and slit.
L3 ophthalmoscope
Semi-circle, small/medium/large circular aperture, fixation star, slit and
grid.
Aperture function
Small circle:
Medium circle: pupils
Semi-circle:
Large circle: for normal examination results
Grid: for topographic determination of retina changes
Light slit: to determine differences in level
Fixation star: to ascertain central or eccentric fixation
4 Filters
Using the filter wheel, the following filters can be switched for each aperture:
L1 ophthalmoscope
The L1 instrument head is supplied without a filter wheel.
(the red filter is contained in the aperture wheel)
L2 ophthalmoscope
Red-free filter, blue filter and polarisation filter.
L3 ophthalmoscope
Red-free filter, blue filter and polarisation filter.
Filter function
Red-free filter: contrast enhancing to assess fine vascular
Polarisation filter: for precise assessment of tissue colours and to
Blue filter: for improved recognition of vascular abnormali-
to reduce reflection for small
changes, e.g. retinal bleeding
avoid retinal reflections
ties or bleeding, for fluorescence ophthalmology
For L2 + L3, every filter can be switched to every aperture.
5. Focussing device (only with L3)
Fast fine adjustment of the examination area to be observed is achieved
from various distances by turning the focussing wheel.
ri-scope® retinoscope (skiascope) XL 3,5 V
1. Intended use
The ri-scope®retinoscope Slit and ri-scope®retinoscope Spot described
37
Page 38
in these operating instructions (also called skiascopes) have been manufactured for examining the refraction of the eye (refractive error).
2. Function
Rotation and focusing of the slit and/or spot image may now be effected
by the knurled screw.
3. Rotation
The slit or spot image may be rotated by 360° by the control. Each angle
may be directly read from the scale on the retinoscope.
4. Fixation cards
Fixation cards are suspended and fixed on the object side of the retinoscope into the bracket for the dynamic skiascope.
5. Slit/Spot design
The slit retinoscope may be converted to a spot retinoscope by exchanging the slit lamp against a spot lamp.
ri-derma® Dermatoskop XL 3,5 V
1. Intended use
The ri-derma®dermatoscope described in these operating instructions
has been produced for early recognition of melanotic skin changes
(malign melanoma).
2. Focusing
Focus the magnifying glass by rotating the eyepiece ring.
3. Skin adapters
Two skin adapters are supplied:
1) Including a scale of 0 - 10 mm for measuring melanotic skin changes,
such as malign melanoma.
article number 10969
2) without scale
article number 10968
Both skin attachments can be removed easily and exchanged.
ri-scope® F.O. bent arm illuminator XL 3,5 V
1. Intended use
The bent arm illuminator described in these instructions for use was
manufactured for illuminating the mouth and pharynx.
ri-scope® F.O. nasal speculum XL 3,5 V
1. Intended use
The nasal speculum described in these instructions for use was manu
38
-
Page 39
factured for illumination and examination of the inside of the nose.
2. Function
Two types of operation are possible:
a) Quick retraction
Press down the adjusting screw on the instrument head with the thumb.
In this adjustment, the position of the shank of the speculum cannot be
changed.
)
Individual retraction
b
Turn the adjusting screw in a clockwise direction until you obtain the
desired speculum opening. The shanks close again when the screw is
turned in a counter-clockwise direction
3. Swivel lens
A swivel lens with an approx. 2.5-fold magnification is to be found on the
nasal speculum, which can be simply removed or replaced again in the
opening provided in the nasal speculum.
ri-scope® F.O. tougue blade holder XL 3,5 V
1. Intended use
The tongue blade holder described in these instructions for use was
manufactured for the examination of the mouth and throat in combination with commercially available wooden and plastic blades.
2. Function
Insert a commercially available wooden or plastic tongue blade through
the opening below the light outlet until it stops. After the examination, the
blade can be easily removed by pushing the ejector.
ri-scope®
Human operations otoscope XL 3,5 V
without speculum
1. Intended use
The RIESTER operation otoscope described in these instructions for use
was manufactured for the illumination and examination of the auditory
canal as well as for small operations in the auditory canal.
2. Attachment and removal of ear specula for human medi-
cine
Place the desired speculum onto the black holder of the operation otoscope so that the recess on the speculum fits into the guide of the holder.
Fix the speculum by turning it in a counter-clockwise direction.
3. Swivel lens for magnification
There is a small magnification lens which can be swivelled 360° on the
operation otoscope with approx. 2.5-fold magnification.
39
Page 40
4. Insertion of external insruments into the ear
The operation otoscope has been designed so that external instruments
can be inserted into the ear without a problem.
ri-scope®
Veterinary operation otoscope XL 3,5 V
without speculum
1. Intended use
The RIESTER operation otoscope described in these instructions for use
was manufactured solely for use in animals and thus for veterinary medicine. It can be used for illumination and examination of the auditory canal
as well as for small operations in the auditory canal.
2. Attachment and removal of ear specula for veterinary
medicine
Place the desired speculum onto the black holder of the operation otoscope so that the recess on the speculum fits into the guide of the holder.
Fix the speculum by turning it in a counter-clockwise direction.
3. Swivel lens for magnification
There is a small magnification lens which can be swivelled by 360° on
the operation otoscope with approx. 2.5-fold magnification.
13.Replacing the lamp
L1 otoscope
Remove the specula fitting from the otoscope. Screw out the lamp counter clockwise.
Screw in the new lamp clockwise and replace the specula fitting.
Screw the instrument head off the battery holder.
The lamp is located at the base of the instrument head.
Pull the lamp out of the instrument head with thumb and forefinger or a
suitable tool. Insert a new lamp.
Ophthalmoscopes
Remove the instrument head from the battery holder.
The lamp is located at the base of the instrument head.
Remove the lamp from the instrument head with thumb and forefinger or
a suitable tool. Insert a new lamp.
Caution: The pin on the lamp must be inserted into the guide groove on
the instrument head.
Veterinary/human operation otoscope
Screw the lamp out of the fixture in the operation otoscope and screw in
a new lamp.
Instrument heads: Retinoscope slit and spot
Remove the instrument head from the battery handle.
The lamp is located in a sleeve at the base of the instrument head.
Remove the lamp from the sleeve using the thumb and index finger or a
suitable tool. Insert the new lamp firmly into the sleeve and replace the
sleeve back into the instrument head so that the base of the lamp fits into
the slot on the instrument head.
40
Page 41
Information on care, cleaning and disinfection
All RIESTER ri-scope®instrument heads can be cleaned on the outside
with a moist cloth. Furthermore, the following disinfectants can be used:
Aldehydes (formaldehyde, glutaraldehyde, aldehyde fission products) or
tensides.
All parts of instruments with the exception of the swivel lens, magnifying
glass and the cover glass can also be disinfected with alcohols.
When using these substances it is absolutely essential to follow the
instructions of the manufacturer.
A soft and, as far as possible, lint-free cloth or cotton bud can be used
as an auxiliary aid for cleaning or disinfection.
Attention
Never place the instrument heads in liquid. Take care that no liquid
enters inside the casing.
a) Sterilisation
According to the current school of thought (Test Centre for Medical
Devices in Tübingen), sterilisation is only prescribed in the case of operative procedures.
b) Reusable ear specula
Although sterilisation is not necessary as described in a), it is nevertheless possible.
The reusable ear specula can be sterilised at 134°C and 10 minutes
retention time in a steam sterilizer.
Putting the instruments heads into operation
Place the desired instrument head onto the attachment on the handle so
that the two recesses on the lower part of the instrument head sit on top
of the two projecting guide cams of the battery handle. Press the instrument head lightly onto the handle and turn the handle in a clockwise
direction until it stops. To remove the head turn it in a counter-clockwise
direction.
Putting the anti-theft security into operation
1 Function
a
b
In order to activate the anti-theft security, turn the Allen screw (b) using
the Allen key (a) (included with the instrument head) until it stops.
The instrument head can now no longer be removed from the handle. In
order to deactivate the anti-theft security, the Allen screw (b) has to be
unscrewed again using the Allen key (a).
Place the desired instrument head onto
the attachment on the handle so that the
two recesses on the lower part of the
instrument head sit on top of the two pro-
jecting guide cams of the battery handle.
Press the instrument head lightly onto the
handle and turn the handle in a clockwise
direction until it stops.
41
Page 42
Têtes d'instrument'ri-scope®
Otoscope ri-scope® L
1. Destination
L’otoscope RIESTER décrit dans ce mode d’emploi sert à éclairer le conduit auditif et à l’examiner avec les spéculums auriculaires RIESTER.
2. Insertion et éjection des spéculums auriculaires
On peut adapter sur la tête de l’otoscope, au choix, des spéculums auriculaires jetables de RIESTER (de couleur bleue) ou des spéculums réutilisables de RIESTER (de couleur noire). La taille du spéculum auriculaire est indiquée à l’arrière du spéculum.
Otoscope L1 et L2
Tourner le spéculum dans le sens horaire jusqu'à ce que vous sentiez
une résistance. Pour éjecter le spéculum, le tourner dans le sens antihoraire.
Otoscope L3
Mettre le spéculum sur la monture métallique chromée de l’otoscope et
appuyer jusqu’à vous sentiez qu’il s’encliquète. Pour éjecter le spéculum, appuyer sur le bouton bleu. Le spéculum est éjecté automatiquement.
3. Lentille grossissante pivotante
La lentille pivotante est fixée sur l’instrument et peut être tournée de
360°.
4. Introduction d’instruments externes dans l’oreille
Si vous voulez introduire dans l’oreille des instruments externes (p. ex.
une pincette), vous devez faire pivoter de 180° la lentille grossissante
(grossissement env. x 3) qui se trouve sur la tête de l’otoscope. Ensuite,
vous pouvez mettre en place la lentille chirurgicale.
5. Otoscopie pneumatique
Pour pouvoir effectuer l’otoscopie pneumatique (= un examen du tympan), vous avez besoin d’une poire, qui n’est pas comprise dans la livraison standard, mais que vous pouvez commander à part. Le tuyau de
la poire est enfoncé sur le raccord. Vous pouvez maintenant insuffler
doucement la quantité d’air nécessaire dans le canal de l’oreille.
Ophtalmoscope ri-scope® L
1. Destination
L’ophtalmoscope RIESTER décrit dans ce mode d’emploi sert pour l’examen de l’oeil et du fond de l’oeil.
2. Roue à lentilles avec lentilles de
correction
Les lentilles de correction peuvent être réglées sur la roue à lentilles.
Vous avez le choix entre les lentilles de correction suivantes :
Ophtalmoscope L1 et L2
Plus : 1-10, 12, 15, 20, 40.
Moins :
1-10, 15, 20, 25, 30, 35.
42
Page 43
Ophtalmoscope L3
Plus : 1-45 par pas
Moins : 1-44 par pas
Lecture des valeurs dans l’afficheur à éclairage. Affichage des valeurs
positives en vert et des valeurs négatives en rouge.
3. Diaphragmes
La roue à diaphragmes permet de sélectionner les diaphragmes suivants :
Ophtalmoscope L1
Demi-lune, petit/moyen/grand spot, étoile de fixation, fente et filtre absorbant du rouge.
Ophtalmoscope L2
Demi-lune, petit/moyen/grand spot, étoile de fixation et fente.
Ophtalmoscope L3
Demi-lune, petit/moyen/grand spot, étoile de fixation, fente et grille.
Fonction diaphragme
Petit spot :
Spot moyen : pour la réduction des réflexes des petites pupil-
les
demi-lune :
Grand spot : pour les examens de fond d’oeil
Grille : pour la constatation topographique des modifi-
cations de la rétine
Fente : pour la détermination des différences de niveau
Étoile de fixation : pour la constatation des fixations centrale et
excentrée
4. Filtres
La roue à filtres permet d’utiliser les filtres suivants avec chaque diaphragme :
Ophtalmoscope L1
La tête de l’ophtalmoscope L1 est livrée sans roue à filtres. (Le filtre
absorbant du rouge est inclus dans la roue à diaphragmes).
Ophtalmoscope L2
Filtre absorbant du rouge, filtre bleu et filtre de polarisation.
Ophtalmoscope L3
Filtre absorbant du rouge, filtre bleu et filtre de polarisation.
Fonction des filtres
Filtre absorbant du rouge : accentue les contrastes pour l’évaluation des
Filtre de polarisation : pour l’évaluation chromatique exacte des tissus
Filtre bleu : pour une meilleure reconnaissance des anoma-
Pour les ophtalmoscopes L2 + L3, chaque filtre peut être utilisé avec
petites modifications vasculaires, par exemple,
saignements rétiniens
et pour une réduction des réflexions de la cor-
née
lies vasculaires ou des saignements, pour
l’ophtalmoscopie par fluorescence
chaque diaphragme.
43
Page 44
5. Dispositif de focalisation
(uniquement L3)
La roue de focalisation permet de régler rapidement et avec précision le
champ à examiner sur différentes distances.
Rétinoscopes ri-scope® XL 3,5 V
1. Usage prévu
Les rétinoscopes Slit et Spot (également appelés skiascopes) décrits
dans ce mode d'emploi ont été mis au point pour déterminer la réfraction (vision défectueuse) de l'œil.
2. Functionnemente
Les rétinoscopes Slit et Spot (également appelés skiascopes) décrits
dans ce mode d'emploi ont été mis au point pour déterminer la réfraction (vision défectueuse) de l'œil.
3. Rotation
L'élément de commande permet de faire tourner de 360° l'image à traits
ou à points. La valeur de l'angle peut être lue directement sur la graduation du rétinoscope.
4. Carte de fixation
Pour la skiascopie dynamique, les cartes de fixation sont sus pendues et
fixées côté objet du rétinoscope.
5. Version Slit / Spot
En remplaçant la lampe à traits (version Slit) par une lampe à points (version Spot), le rétinoscope à traits devient un rétinoscope à points.
Dermatoscope ri-derma® XL 3,5 V
1. Usage prévu
Le dermatoscope ri-derma®décrit dans ce mode d'emploi a été mis au
point pour le dépistage précoce de modifications pigmentées de la peau
(mélanomes malins).
2. Focalisation
Ajustez la focale de la loupe en faisant tourner l'anneau de l'oculaire.
3. Embouts pour la peau
2 embouts pour la peau sont fournis:
1) avec une graduation de 0 à 10 mm pour mesurer les modifications
pigmentées de la peau telles que les mélanomes malins.
référence 10969
2) sans graduation
référence 10968
Les deux embouts cutanés sont amovibles et interchangeables.
44
Page 45
Support de lampe à F.O. ri-scope® XL 3,5 V
1. Usage prévu
Le support de lampe décrit dans ce mode d'emploi est destiné à l'éclairage de la cavité buccale et de l'espace pharyngien.
Spéculum nasal à F.O. ri-scope® XL 3,5 V
1. Usage prévu
Le spéculum nasal décrit dans ce mode d'emploi est destiné à l'éclairage et donc à l'examen de l'intérieur du nez.
2. Fonctionnement
Il existe deux types d'utilisation :
a) Ecartement rapide
A l'aide du pouce, appuyez (vers le bas) sur la molette de réglage située
au niveau de la tête de l'instrument. Ce réglage empêche alors tout
changement de l'écartement du spéculum.
)
Ecartement individuel
b
Faites tourner la molette de réglage dans le sens des aiguilles d'une
montre jusqu'à l'obtention de l'écartement souhaité.
Pour réduire à nouveau l'écartement, faites tourner la molette dans le
sens inverse de celui des aiguilles d'une montre.
3. Lentille privotante
Une lentille pivotante dont le pouvoir grossissant est d'environ 2,5 est
montée sur le spéculum nasal. En cas de besoin, il suffit de la tirer hors
de son logement ou bien de l'y placer à nouveau.
Support pour abaisse-langue à F.O.
ri-scope® XL 3,5 V
1. Usage prévu
Le support pour abaisse-langue décrit dans ce mode d'emploi est destiné à être employé en combinaison avec des abaisse-langues classiques
en bois ou en plastique, pour l'examen de la cavité buccale et de l'espace pharyngien.
2. Fonctionnemente
Introduisez un abaisse-langue classique en bois ou en plastique dans le
logement situé en dessous de la sortie de lumière, et poussez-le jusqu'à
la butée. Après l'examen, la touche d'éjection permet de retirer l'abaisselangue aisément.
ri-scope® XL 3,5 V
Otoscope chirurgical pour humains
sans spéculum auriculire
1. Usage prévu
L'otoscope chirurgical RIESTER décrit dans ce mode d'emploi est destiné l'éclairage et l'examen du conduit auditif ainsi qu'aux petites interven
45
-
Page 46
tions chirurgicales réalisés sur le conduit auditif.
2. Mise en place et retrait des spéculums auriculaires en
médecine humaine
Placez le spéculum auriculaire adéquat sur le support noir de l'otoscope
chirurgical de façon à ce que le creux du spéculum se place dans la glissière du support. Fixez le spéculum en le faisant tourner dans le sens
inverse de celui des aiguilles d'une montre.
3. Lentille pivotante grossissante
Une petite lentille grossissante pivotant sur 360° et dont le facteur de
grossissement est d'environ 2,5 est montée sur l'otoscope chirurgical.
4. Introduction d'instruments externes dans l'oreille
L'otoscope chirurgical est conçu de façon à permettre une introduction
facile d'instruments externes dans l'oreille.
ri-scope® XL 3,5 V
Otoscope chirurgical à usage vétérinair
sans spéculum auriculire
1. Usage prévu
L'otoscope chirurgical RIESTER décrit dans ce mode d'emploi est destiné à un usage vétérinaire uniquement. Il peut Ítre utilisé pour l'éclairage
et l'examen du conduit auditif ainsi que pour de petites interventions chirurgicales réalisÈs sur le conduit auditif.
2. Mise en place et retrait des spéculums auriculaires en
médecine vétérinaire
Placez le spéculum auriculaire adéquat sur le support noir de l'otoscope
chirurgical de façon à ce que le creux du spéculum se place dans la glissière du support. Fixez le spéculum en le faisant tourner dans le sens
inverse de celui des aiguilles d'une montre.
3. Lentille pivotante grossissante
Une petite lentille grossissante pivotant sur 360° et dont le facteur de
grossissement est d'environ 2,5 est montée sur l'otoscope chirurgical.
13. Remplacement de la lampe
Otoscope L1
Détacher le porte-spéculum de l’otoscope.
Tourner la lampe dans le sens antihoraire pour la démonter.
Mettre la lampe neuve en place en la tournant à fond dans le sens horaire et remettre le porte-spéculum en place.
Otoscopes L2, L3, ri-derma®, support de lampe, spéculum nasal et
support d’abaisse-langue
Détacher la tête de l’instrument du manche à piles.
La lampe se trouve dans le bas de la tête de l’instrument.
Sortir la lampe de la tête de l’instrument en la tenant par le pouce et l’index ou en vous aidant d’un outil adapté. Introduire la lampe neuve dans
la tête et bien la serrer
Ophtalmoscopes
Détacher la tête de l’instrument du manche à piles. La lampe se trouve
.
46
Page 47
dans le bas de la tête de l’instrument. Sortir la lampe de la tête de l’instrument en la tenant par le pouce et l’index ou au moyen d’un outil
adapté. Introduire la lampe neuve dans la tête et bien la serrer.
Attention : La pointe de la lampe doit être enfoncée dans l’encoche
dans la tête de l’instrument.
Otoscopes chirurgicaux pour la médecine
vétérinaire/humaine
Extraire la lampe de la douille dans l’otoscope chirurgical, mettre en
place une lampe neuve et la serrer à fond.
Têtes d'instrument : rétinoscope à trait et à spot
Désolidarisez la tête d'instrument du manche à piles.
La lampe se trouve dans une douille, laquelle se trouve elle-même dans
la partie inférieure de la tête d'instrument. A l'aide pouce et de l'index ou
bien à l'aide d'un outil approprié, retirez la lampe et sa douille de la tête
d'instrument. Mettez une lampe neuve en place en veillant à ce qu'elle
soit bien fixée dans sa douille, puis replacez l'ensemble lampe-douille
dans la tête d'instrument en veillant à ce que le taquet de la lampe s'introduise dans la rainure de la tête d'instrument.
Instructions d'entretien, de nettoyage et de désinfection
Toutes les têtes d'instrument ri-scope®de RIESTER peuvent être nettoyées à l'aide d'un chiffon humide. En outre, il est possible d'utiliser les
désinfectants suivants :
aldéhydes (formol, glutaraldéhyde, produits aldéhydiques) ou agents
tensioactifs.
A l'exception des lentilles pivotantes, des loupes et des verres de protection, toutes les pièces d'instrument peuvent être désinfectées avec des
alcools.
Respectez impérativement les prescriptions du fabricant lors de l'utilisation de ces substances.
Pour le nettoyage et la désinfection, on peut utiliser un chiffon doux si
possible non pelucheux ou bien des cotons-tiges.
Attention
Ne jamais plonger les têtes d'instrument dans un liquide. Veiller à ce
qu'aucun liquide ne pénètre à l'intérieur du boîtier.
a) Stérilisation
Selon les directives officielles en vigueur (données par le centre d'examen des produits médicaux de Tübingen, Allemagne [Centre de contrôle des produits médicaux de Tübingen]), la stérilisation n'est obligatoire
que pour les interventions chirurgicales.
b) Spéculums auriculaires réutilisables
Comme cela est décrit en a), la stérilisation n'est pas nécessaire. Celleci est toutefois possible.
Les spéculums auriculaires réutilisables peuvent être stérilisés à 134° C
pendant 10 minutes dans un stérilisateur à vapeur.
Mise en service des tÍtes d'instrument
Placez la tête d'instrument souhaitée sur la partie supérieure du manche
de façon à de que les deux alvéoles de la partie inférieure de la tête d'instrument reposent sur les deux cames de guidage (dépassant légèrement) du manche à piles. Appuyez légèrement la tête d'instrument sur le
47
Page 48
manche, puis faites pivoter le manche dans le sens de aiguilles d'une
montre jusqu'à la butée. Pour enlever la tête, faire pivoter celle-ci dans
le sens inverse de celui des aiguilles d'une montre.
Mise en service de l'antivol
Fonctionnement
a
b
sur le manche, puis faites pivoter le manche dans le sens de aiguilles
d'une montre jusqu'à la butée. Pour activer l'antivol, vissez la vis à six
pans creux (b) jusqu'à la butée à l'aide de la clé hexagonale (a) (livrée
avec la tête d'instrument). A présent, il est impossible de désolidariser la
tête d'instrument de son manche. Pour désactiver l'antivol, dévisser la
vis à six pans creux (a) à l'aide de la clé hexagonale.
Placez la tête d'instrument souhaitée sur
la partie supérieure du manche de façon
à de que les deux alvéoles de la partie
inférieure de la tête d'instrument reposent
sur les deux cames de guidage (dépas-
sant légèrement) du manche à piles.
Appuyez légèrement la tête d'instrument
Cabezales de insrtumento ri-scope®
Otoscopio ri-scope® L
1. Uso previsto
El otoscopio RIESTER que se describe en el manual del operador sirve
para iluminar y examinar el conducto auditivo en combinación con los
espéculos auriculares RIESTER.
2. Montaje y desmontaje de los espéculos auriculares
Para el cabezal del otoscopio se pueden seleccionar opcionalmente
espéculos desechables RIESTER (azules) o espéculos reutilizables
RIESTER (negros).
El tamaño del espéculo auricular se indica en la parte
posterior del espéculo.
Otoscopio L1 y L2
Gire el espéculo en sentido horario hasta que note resistencia. Para
poder desmontar el espéculo, gírelo en sentido antihorario.
Otoscopio L3
Introduzca el espéculo seleccionado en el soporte metálico cromado del
otoscopio hasta que encaje perceptiblemente.
Para poder desmontar el espéculo, pulse la tecla de expulsión azul. El
espéculo se expulsa automáticamente.
3. Lente giratoria de ampliación
La lente giratoria está unida de forma fija al instrumento y se puede girar
360°.
4. Introducción de instrumentos externos en el oído
Si desea introducir instrumentos externos en el oído (p. ej. unas pinzas),
debe girar 180° la lente giratoria (aprox. 3 aumentos) que se encuentra
en el otoscopio. Ahora puede introducir la lente quirúrgica.
48
Page 49
5. Prueba neumática
Para poder realizar la prueba neumática (un examen del tímpano) necesita una pera que no se incluye en el volumen de suministro normal pero
que podrá encargar opcionalmente. El tubo de la pera se inserta en el
conector. Ahora puede introducir con cuidado la cantidad de aire necesaria en el conducto auditivo.
Oftalmoscopio ri-scope® L
1. Uso previsto
El oftalmoscopio RIESTER descrito en este manual del operador sirve
para examinar el ojo y el fondo del ojo.
2. Rueda de lentes con lentes de
corrección
Las lentes de corrección se pueden ajustar en la rueda de lentes. Puede
seleccionar las lentes de corrección siguientes:
D+: 1-45 en pasos individuales
D-: 1-44 en pasos individuales
Los valores se pueden leer en el campo de indicación iluminado. Los
valores positivos se indican con números verdes, los valores negativos
con números rojos.
3. Diafragmas
Con la ruedecilla de diafragmación puede seleccionar los diafragmas
siguientes:
Oftalmoscopio L1
Semicírculo, diafragma circular pequeño/mediano/grande, estrella de
fijación, rendija y filtro exento de rojo.
Oftalmoscopio L2
Semicírculo, diafragma circular pequeño/mediano/grande, estrella de
fijación y rendija.
Oftalmoscopio L3
Semicírculo, diafragma circular pequeño/mediano/grande, estrella de
fijación, rendija y rombo.
DiafragmaFunción
Medio círculo:
Círculo pequeño:
de Semi círculo:
Círculo grande:para exámenes normales del fondo del ojo
Rombo: para la determinación topográfica de altera-
Rendija iluminada: para determinar diferencias de nivel
Estrella de fijación: para la determinación de la fijación central y
para reducción del reflejo
pupilas pequeñas
ciones de la retina
excéntrica
49
Page 50
4 Filtros
Con la rueda de filtros puede añadir a cada diafragma los filtros siguientes:
Oftalmoscopio L1
El cabezal de instrumentos L1 se suministra sin rueda de filtros. (el filtro
exento de rojo está incluido en la rueda de diafragma)
Oftalmoscopio L2
Filtro exento de rojo, filtro azul y filtro de polarización.
Oftalmoscopio L3
Filtro exento de rojo, filtro azul y filtro de polarización.
Función de los filtros
Filtro exento de rojo: efecto intensificador del contraste, para la eva-
Filtro de polarización: para evaluar con exactitud los colores de los
Filtro azul: para una mejor detección de anomalías vascu-
En L2 + L3 se puede añadir cada filtro a cualquier diafragma.
5. Dispositivo de enfoque (sólo en L3)
Girando la rueda de enfoque puede obtener un ajuste preciso y rápido a
distancias diferentes del campo de exploración que desea visualizar.
luación de pequeñas alteraciones vasculares,
p.ej. hemorragias de retina
tejidos y reducir la reflexión en la córnea
lares o hemorragias, para oftalmología de fluorescencia
Retinoscopios ri-scope® XL 3,5 V
1. Finalidad de uso
Los retinoscopios Slit y retinoscopios Spot descritos en este manual de
instrucciones para el uso (conocidos también como esquiascopios), sirven para determinar la refracción ocular (falta de vista).
2. Functionamiento
El tornillo moleteado posibilita la rotación y el enfoque de la imagen de
raya o de punto.
3. Rotación
Con el elemento de mando podrá orientar la imagen de raya o de punto
entre 360º. El correspondiente ángulo, podrá leerlo directamente en la
escala del retinoscopio.
4. Carta de fijación
Para el esquiascopio dinámica, se cuelgan y se fijan las cartas de fijación por el lado del objetivo del retinoscopio.
5. Ejecución Slit/ Spot
Cambiando la bombilla de raya (ejecución Slit) por una bombilla de
punto (ejecución Spot), el retinoscopio de raya se convierte en un retinoscopio de punto.
Dermatoscopio ri-derma® XL 3,5 V
1. Finalidad de uso
El dermatoscopio ri-derma®descrito en este manual de instrucciones
50
Page 51
para el uso sirve para la detección temprana de alteraciones en la pigmentación cutánea (melanomas malignos).
2. Enfoque
Enfoque la lupa girando para ello el anillo del ocular.
3. Contactos sobre la piel
El instrumento se suministra con dos contactos sobre la piel:
1) Con graduación de 0-10 mm para medir cambios de pigmentación
cutánea como melanomas malignos.
Número de artículo 10969
2) Sin escala
Número de artículo10968
Ambas piezas de contacto con la piel pueden desmontarse y cambiarse
fácilmente.
Portalámparas ri-scope®F.O. XL 3,5 V
1. Finalidad de uso
El portalámparas descrito en estas instrucciones ha sido fabricado para
la iluminación de la cavidad bucofaríngea.
Espéculo nasal ri-scope® F.O. XL 3,5 V
1. Finalidad de uso
El espéculo nasal descrito en estas instrucciones ha sido fabricado para
la
iluminación y exploración de las fosas nasales.
2. Funcionamiento
Son posibles dos modos de funcionamiento:
a) Expansión rápida
Empuje hacia abajo con el pulgar el tornillo de ajuste del cabezal del
instrumento. En este caso no es posible modificar la posición de las
ramas del espéculo.
)
Expansión personalizada
b
Gire el tornillo de ajuste en el sentido de las agujas del reloj hasta alcanzar la expansión deseada. Las ramas se vuelven a cerrar girando el tornillo en sentido contrario a las agujas del reloj.
3. Lente pivotante
El espéculo nasal dispone de una lente pivotante de aproximadamente
2,5 aumentos, que si se desea puede extraerse y posteriormente volverse a introducir en la correspondiente abertura del espéculo.
51
Page 52
Soporte para depresor lingual
ri-scope® F.O. XL 3,5 V
1. Finalidad de uso
El soporte para depresor lingual descrito en estas instrucciones ha sido
fabricado para la exploración de la cavidad bucofaríngea en combinación con depresores linguales de madera o plástico habituales en el mercado.
2. Funcionamiento
Introduzca hasta el tope un depresor lingual de madera o plástico del
tipo habitual en la abertura situada bajo la salida de luz. Tras la exploración, el depresor puede retirarse fácilmente accionando el expulsor.
Otoscopio quirúrgico humano
ri-scope® XL 3,5 V sin espéculo
1. Finalidad de uso
El otoscopio quirúrgico RIESTER descrito en estas instrucciones ha sido
fabricado para la iluminación y exploración del canal auditivo y para
pequeñas intervenciones quirúrgicas en el mismo.
2. Colocación y extracción de los espéculos auriculares
en medicina humana
Coloque el espéculo deseado sobre el soporte negro del otoscopio quirúrgico de modo que la muesca del espéculo coincida con la guía del
soporte. Fije el espéculo girándolo en sentido contrario a las agujas del
reloj.
3. Lente de aumento pivotante
El otoscopio qurúrgico dispone de una pequeña lente de aproximadamente 2,5 aumentos que puede pivotarse 360°.
4. Introducción de instrumentos externos en el oído
El otoscopio quirúrgico está diseñado de modo que permite introducir sin
problemas instrumentos externos en el oído.
Otoscopio quirúrgico veterinario
ri-scope® XL 3,5 V, sin espéculo
1. Finalidad de uso
El otoscopio quirúrgico RIESTER descrito en estas instrucciones ha sido
fabricado para su uso exclusivo en medicina veterinaria, no estando
destinado al uso humano. Puede emplearse para la iluminación y exploración del canal auditivo y para pequeñas intervenciones quirúrgicas en
el mismo.
2. Colocación y extracción de los espéculos auriculares en
medicina veterinaria
Coloque el espéculo deseado sobre el soporte negro del otoscopio qui
rúrgico de modo que la muesca del espéculo coincida con la guía del
soporte. Fije el espéculo girándolo en sentido contrario a las agujas del
reloj.
52
-
Page 53
3. Lente de aumento pivotante
El otoscopio qurúrgico dispone de una pequeña lente de aproximadamente 2,5 aumentos que puede pivotarse 360°.
13. Cambio de la lámpara
Otoscopio L1
Desmonte el soporte del espéculo del otoscopio.
Gire la lámpara en el sentido contrario a las agujas del reloj.
Enrosque la nueva lámpara en sentido horario y vuelva a montar el
soporte del espéculo.
Otoscopios L2, L3, ri-derma®, portalámparas, espéculo
nasal y portaespátulas
Desenrosque el cabezal de instrumentos del mango de pilas. La lámpara se encuentra en la parte inferior del cabezal de instrumentos. Extraiga
la lámpara con el pulgar y el dedo índice o con una herramienta adecuada del cabezal de instrumentos. Introduzca firmemente la nueva lámpara.
Oftalmoscopios
Desmonte el cabezal de instrumentos del mango de pilas. La lámpara se
encuentra en la parte inferior del cabezal de instrumentos. Extraiga la
lámpara con el pulgar y el dedo índice o con una herramienta adecuada
del cabezal de instrumentos. Introduzca firmemente la nueva lámpara.
Atención: El borne de la lámpara se debe introducir en la ranura guía
del cabezal de instrumentos.
Desenrosque la lámpara del zócalo del otoscopio quirúrgico y enrosque
la nueva lámpara.
Cabezales de instrumento: retinoscopio de raya y de punto
Desmonte del mango el cabezal del instrumento.
La lámpara se encuentra en un collar en la parte inferior del cabezal del
instrumento. Extraiga la lámpara del cabezal, junto con el collar, utilizando el pulgar y el índice o una herramienta adecuada. Coloque la nueva
lámpara en el collar en la misma posición y vuelva a introducir el collar
con la lámpara en el cabezal del instrumento de modo que el saliente de
la lámpara se deslice a través de la muesca del cabezal.
Cuidado de los instrumentos, limpieza y desinfecciún
Todos los cabezales de instrumento RIESTER ri-scope®pueden limpiarse externamente con un paño húmedo. También pueden utilizarse los
siguientes desinfectantes:
aldehidos (formaldehido, glutaraldehido, productos liberadores de aldehido) o agentes tensioactivos.
Además, todos los componentes de los instrumentos menos las lentes
pivotantes, lupas y vidrios protectores pueden desinfectarse con alcoholes.
Al emplear estos productos es imprescindible seguir las normas del
fabricante.
Como ayuda para la limpieza o desinfección puede utilizarse un paño
suave que no suelte pelusa o un bastoncillo de algodón.
Atención
No sumerja nunca en un líquido los cabezales de instrumento. Tenga
cuidado de que no penetre líquido en el interior de la carcasa.
53
Page 54
a) Esterilización
Según los criterios cientÌficos actuales (Centro de ensayos de productos
sanitarios de Tubingen), la esterilización sólo es necesaria en intervenciones quirúrgicas.
b) Espéculos auriculares reutilizables
Aunque, como se indica en el punto a), la esterilización no es imprescindible, los espéculos auriculares reutilizables pueden esterilizarse.
Para ello, manténgalos 10 minutos en autoclave a 134°C.
Puesta en marche de los cabezales de
instrumento
Introduzca el cabezal deseado en el alojamiento de la parte superior del
mango de modo que las dos muescas de la parte inferior del cabezal
queden situadas sobre los dos salientes de guía del mango. Presione
levemente el cabezal contra el mango y gire el mango en sentido de las
agujas del reloj hasta el tope. Para desmontar el cabezal, gírelo en sen-
tido contrario a las agujas del reloj.
Dispositivo antirrobo
Funcionamiento
a
b
Presione levemente el cabezal contra el mango y gire el mango en
sentido de las agujas del reloj hasta el tope. Para activar el dispositivo
antirrobo, utilice la llave Allen (a) adjunta al cabezal de instrumento para
enroscar hasta el tope el tornillo Allen (b). De ese modo, el cabezal ya
no podr· desmontarse del mango. Para desactivar el dispositivo antirrobo, deberá desenroscarse de nuevo el tornillo Allen (b) con ayuda de la
llave Allen (a).
Introduzca el cabezal deseado en el alo-
jamiento de la parte superior del mango
de modo que las dos muescas de la parte
inferior del cabezal queden situadas
sobre los dos salientes de guía del
mango.
Головки приборов к рукояткам ri-scope®
3. Отоскоп ri-scope® L
1. азначение
Описываемый в данном руководстве отоскоп RIESTER предназначен для
освещения и исследования слухового прохода и используется в комбинации с
ушными воронками RIESTER.
2. Установка и снятие ушных воронок
оловка отоскопа может оснащаться одноразовыми воронками RIESTER (синего
цвета) или многоразовыми воронками RIESTER (чёрного цвета).
азмер указан сзади на ушной воронке.
Отоскопы L1 и L2
ращайте воронку по часовой стрелке, пока не почувствуете сопротивление. ля
снятия воронки поворачивайте её против часовой стрелки.
54
Page 55
Отоскоп L3
Установите выбранную воронку на хромированный металлический патрон
отоскопа. ля снятия воронки нажмите на синюю кнопку выброса. оронка
автоматически выбрасывается.
3. оворотная линза для увеличения
оворотная линза жёстко соединена с прибором и поворачивается на 360¼.
4. ведение внешних инструментов в ухо
ри введении внешних инструментов в ухо (напр., пинцета) следует повернуть
линзу (прибл. 3-кратное увеличение) на головке отоскопа на 180°. еперь можно
установить операционную линзу.
5. невматический тест
ля проведения пневматического теста (= исследования барабанной перепонки)
требуется шарик, который не входит в стандартный комплект поставки, но может
быть заказан дополнительно. ланг шарика надевается на наконечник. осле
этого можно осторожно подать необходимое количество воздуха в ушной канал.
Офтальмоскоп ri-scope® L
1. азначение
Описываемый в данном руководстве офтальмоскоп RIESTER предназначен для
исследования глаза и глазного дна.
2. олёсико с корректирующими линзами
орректирующие линзы можно регулировать на линзовом колёсике. ожно
выбрать следующие корректирующие линзы:
люс: 1-45 одиночными шагами
инус: 1-44 одиночными шагами
начения можно считывать в освещённом поле зрения. люсовые значения
отображаются зелёными, минусовые - красными числами.
3. ленды
помощью колёсика установки бленд можно выбрать следующие бленды:
Офтальмоскоп L1
олусегмент, маленький/средний/больший круг, фиксированная звездочка,
светящаяся щель и фильтр без красного спектра.
Офтальмоскоп L2
олусегмент, маленький/средний/больший круг, фиксированная звездочка и
светящаяся щель.
Офтальмоскоп L3
олусегмент, маленький/средний/больший круг, фиксированная звездочка,
светящаяся щель и ромб.
55
Page 56
ленда ункция
алый круг:
редний круг: для ослабления отражения у малых зрачки
и полукруг:
ольшой круг: для нормальных исследований дна
омб: для топографической фиксации
ветящаяся щель: для определения
икс. звёздочка: для обнаружения центральной или эксцентрической
4 ильтры
помощью колёсика фильтров можно подключить следующие фильтры для
каждой бленды:
Офтальмоскоп L1
оловка L1 поставляется без колёсика фильтров. (ильтр без красного спектра
входит в комплект колёсика для бленд)
Офтальмоскоп L2
ильтр без красного спектра, синий светофильтр и поляризационный фильтр.
Офтальмоскоп L3
ильтр без красного спектра, синий светофильтр и поляризационный фильтр.
ильтр ункция
ильтр без кр. цвета: усиление контраста для анализа мелких
оляризац. фильтр: для точной оценки цветов ткани и уменьшения
иний фильтр: для лучшего распознавания аномалий сосудов или
моделях L2 + L3 можно подключить каждый фильтр к любой бленде.
5. окусировочное устройство (только в L3)
ращением фокусировочного колеса можно быстро добиться точной регулировки
рассматриваемой рабочей зоны на различных расстояниях.
изменений сетчатки
разницы в уровне
фиксации
изменений сосудов напр., кровоизлияний в сетчатку
отражений роговицы
кровотечений, для флуоресцентной офтальмологии
3. ri-scope® ретиноскоп (скиаскоп) XL 3,5
1. азначение
Описанные в данной инструкции пользователя ретиноск
в виде щели и со светом в виде точки (называемые также скиаскопами)
изготовлены для определения рефракции (аметропии) глаза.
2. ункция
помощью винта с накатанной го
фокусировку щели или пятна.
ловк
56
ой мо
опы ri-scope® со светом
жно выпо
лнять вращение и
Page 57
3. ращение
ель или пятно можно поворачивать с помощью рабочего элемента на 360°.
оответствующий угол можно считывать непосредственно на шкале ретиноскопа.
4. арты фиксации
арты фиксации помещаются на ретиноскопе со стороны пациента в держатель
для динамической скиаскопии.
5. сполнение щель / пятно
етиноскоп со светом в виде щели может быть переделан в ретиноскоп со светом
в виде точки путем замены щелевой лампы на точечную лампу.
ri-derma® дерматоскоп XL 3,5
1. азначение
Описанный в данной инструкции пользователя дерматоскоп ri-derma® изготовлен
для раннего распознавания пигментных изменений кожи (злокачественных
меланом).
2. окусировка
фокусируйте увеличительную лупу поворотом кольца окуляра.
3. ожные насадки
объём поставки входят 2 кожные насадки:
1) с масштабной шкалой 0 - 10 мм для измерения пигментированных изменений
Описанный в данной инструкции пользователя назальный расширитель
изготовлен для освещения и исследования внутренней полости носа.
2. ункция
озможны два режима использования:
a) ыстрое расширение
ажмите вниз большим пальцем винт на головке прибора. ри этом положение
створок воронки изменить нельзя.
57
Page 58
б) онтролируемое расширение
оворачивайте винт по часовой стрелке до получения требуемого расширения
воронки.
творки снова смыкаются при повороте винта против часовой стрелки.
3. ращающаяся линза
а назальном расширителе имеется вращающаяся линза с 2,5-кратным
увеличением, которую по желанию можно просто вынуть или снова вставить в
предусмотренное для этого отверстие в назальном расширителе.
ri-scope® фиброоптический держатель шпателя
XL 3,5 В
1. азначение
Описанный в данной инструкции пользователя держатель шпателя изготовлен
для исследования полости рта и зева в комбинации с обычными деревянными и
пластмассовыми шпателями.
2. ункция
ведите до упора обычный деревянный или пластмассовый шпатель в отверстие
под выходом света. осле обследования шпатель можно легко убрать, нажав
выбрасыватель.
ri-scope®
операционный отоскоп (медицинский) XL 3,5 В
без воронки
1. азначение
Описанный в данной инструкции пользователя операционный отоскоп Riester
изготовлен для освещения и обследования слухового прохода, а также для несложных операций в слуховом проходе.
2. Установка и снятие ушных воронок медицинских
ставьте нужную воронку в чёрный держатель на операционном отоскопе так,
чтобы выемка на воронке совпала с направляющей в держателе. афиксируйте
воронку, повернув её против часовой стрелки.
3. оворотная увеличительная линза
а операционном отоскопе имеется небольшая увеличительная линза с
возможность поворота на 360° для 2,5-кратного увеличения.
4. ведение внешних инструментов в ухо
Операционный отоскоп сконструирован таким образом, чтобы можно было легко
вводить в ухо внешние инструменты.
ri-scope®
операционный отоскоп ветеринарный XL 3,5 В
без воронки
1. азначение
Описанный в данной инструкции по применению операционный отоскоп Riester
изготовлен исключительно для применения на животных и для ветеринарии. го
можно использовать для освещения и исследования слухового прохода, а также
для несложных операций в слуховом проходе.
58
Page 59
2. Установка и снятие ушных воронок при лечении животных
ставьте нужную воронку в чёрный держатель на операционном отоскопе так,
чтобы выемка на воронке совпала с направляющей в держателе. афиксируйте
воронку, повернув её против часовой стрелки.
3. оворотная увеличительная линза
а операционном отоскопе имеется небольшая увеличительная линза с возможность поворота на 360° для 2,5-кратного увеличения.
амена лампы
Отоскоп L1
нимите зажим для воронок с отоскопа. ыкрутите лампу против часовой
стрелки.
рикрутите новую лампу по часовой стрелке и установите на место зажим для
воронок.
Открутите головку инструмента с рукоятки.
ампа находится внизу в головке. ольшим и указательным пальцем или с
помощью специального инструмента выньте лампу из головки. Установите новую
лампу.
Офтальмоскопы
нимите головку инструмента с рукоятки. ампа находится внизу в головке.
ольшим и указательным пальцем или с помощью специального инструмента
извлеките лампу из головки. Установите новую лампу.
нимание: стержень лампы необходимо ввести в направляющий паз на головке
инструмента.
етеринарные/медицинские операционные отоскопы
ыкрутите лампу из патрона в операционном отоскопе и вкрутите новую лампу.
оловки приборов: ретиноскоп щелевой и точечный
нимите головку прибора с с батареечной рукоятки.
ампа находится в гильзе под головкой прибора. звлеките лампу из гильзы
большим и указательным пальцами или соответствующим инструментом из
головки прибора. ставьте новую лампу в гильзу снова в головку прибора так,
чтобы штифт лампы оказался в пазе головки прибора.
Указания по уходу, чистке и дезинфекции
се головки приборов Riester ri-scope® - можно чистить снаружи влажной
тканью.
роме того, для дезинфекции можно использовать следующие
дезинфицирующие средства:
альдегиды (формальдегид, глютаральальдегид, расщепитель альдегидов) или
поверхностно-активные вещества.
се детали приборов, кроме вращающихся линз, увеличительных линз и стеклянных крышек, можно также дезинфицировать спиртами.
ри использовании этих веществ строго соблюдайте инструкции фирмыизготовителя.
качестве вспомогательного средства для чистки или дезинфекции можно
использовать мягкую безворсовую ткань или ватные палочки.
нимание
икогда не помещайте головки приборов в жидкость. ледите за тем, чтобы
жидкость не попала внутрь корпуса.
a) терилизация
соответствии с актуальным мнением учёных (сследовательский центр
59
Page 60
продукции медицинского назначения в юбингене) стерилизация необходима
только при оперативном вмешательстве.
б) ногоразовые ушные воронки
отя, как сказано в п. a), в стерилизации нет необходимости, тем не менее она
возможна.
ногоразовые ушные воронки можно стерилизовать при температуре 134°C в
течение 10 минут в паровом стерилизаторе.
Ввод в эксплуатацию головок приборов
Установите нужную головку прибора на креплении в верхней части рукоятки так,
чтобы обе выемки нижней части головки прибора совпали с обеими
направляющими кулачками батареечной рукоятки. легка нажмите головку
прибора и поверните рукоятку до упора по часовой стрелке. оловка снимается
вращением против часовой стрелки.
Ввод в эксплуатацию системы защиты от кражи
ункция
a
b
ля активизации системы защиты от кражи поверните до упора с помощью
ключа с внутренним шестигранником (a) (прилагается к головке прибора) винт с
внутренним шестигранником (b).
еперь инструментальную головку нельзя отделить от ручки. ля отмены
функции защиты от хищения необходимо снова вывернуть винт с внутренним
шестигранником (b) с помощью ключа с внутренним шестигранником (a).
Установите нужную головку прибора на
креплении в верхней части рукоятки так,
чтобы обе выемки нижней части головки
прибора совпали с обеими направляющими
кулачками батареечной рукоятки. легка
нажмите головку прибора и поверните руко-
ятку до упора по часовой стрелке.
Teste ri-scope®
Otoscopio ri-scope® L
1. Uso previsto
L’otoscopio RIESTER descritto nel presente libretto di istruzioni è destinato all’illuminazione e all’indagine del condotto uditivo in combinazione
con gli specoli auricolari RIESTER.
2. Posizionamento e rimozione di specoli
auricolari
Per la dotazione dell’otoscopio è possibile scegliere a piacere tra specoli auricolari monouso RIESTER (in colore blu) oppure specoli auricolari
riutilizzabili RIESTER (in colore nero).
La misura dello specolo auricolare è riportata sul retro dello strumento.
Otoscopi L1 e L2
Ruotare lo specolo in senso orario fino ad avvertire una certa resistenza. Per smontare lo specolo, ruotarlo in senso antiorario.
Otoscopio L3
Posizionare lo specolo selezionato sul supporto di metallo cromato
dell’otoscopio fino ad avvertire chiaramente lo scatto d’innesto. Per
smontare lo specolo, premere il tasto blu di rilascio. Lo specolo si distacca automaticamente.
60
Page 61
3. Lente d’ingrandimento mobile
La lente d’ingrandimento mobile è fissata allo strumento e può essere
ruotata di 360°.
4. Inserimento di strumenti esterni nell’orecchio
Se si desiderano inserire strumenti esterni nell’orecchio (ad es. pinzette), ruotare di 180° la lente mobile (ingrandimento di circa 3 x), montata
sulla testa dell’otoscopio. A questo punto è possibile utilizzare la lente
d’ingrandimento operatoria.
5. Test pneumatico
Per potere eseguire il test pneumatico (= esame della membrana del timpano), occorre disporre di una monopalla non fornita nella normale dotazione, ma che è possibile ordinare separatamente. Inserire il tubo della
monopalla sull’attacco. Ora è possibile immettere con precauzione la
necessaria quantità d’aria nel canale uditivo.
Oftalmoscopio ri-scope® L
1. Uso previsto
L’oftalmoscopio RIESTER descritto nel presente libretto di istruzioni è
destinato all’indagine dell’occhio e del fondo oculare.
2. Ruota portalenti con lenti correttive
È possibile regolare le lenti correttive sulla ruota portalenti. Sono disponibili, a scelta, le seguenti lenti correttive:
Più: 1-45 in intervalli singoli
Meno: 1-44 in intervalli singoli
I valori sono visibili nel campo visivo illuminato. Le cifre positive sono
visualizzate in verde, quelle negative in rosso.
3. Diaframmi
Dalla ruota portadiaframmi è possibili selezionare i seguenti diaframmi:
Oftalmoscopio L1
Semicerchio, cerchio piccolo/medio/grande, stella di fissazione, fessura
e filtro privo di rossi.
Oftalmoscopio L2
Semicerchio, cerchio piccolo/medio/grande, stella di fissazione e fessura.
Oftalmoscopio L3
Semicerchio, cerchio piccolo/medio/grande, stella di fissazione, fessura
e reticolo.
61
Page 62
Diaframma Funzione
Cerchio piccolo:
cerchio medio:
semicerchio:
Cerchio grande: per normali indagini del fondo dell’occhio
Reticolo: per l’individuazione topografica di alterazio-
Fessura: per le determinazione di differenze di livello
Stella di fissazione: per la definizione della fissazione centrale o
4 Filtri
Con la ruota portafiltri, per ogni diaframma è possibile utilizzare i seguenti filtri:
Oftalmoscopio L1
La testa dello strumento L1 viene fornita senza ruota portafiltri. (Il filtro
privo di rossi è contenuto nella ruota portadiaframmi)
Oftalmoscopio L2
Filtro privo di rossi, filtro blu e filtro di polarizzazione.
Oftalmoscopio L3
Filtro privo di rossi, filtro blu e filtro di polarizzazione.
Filtro Funzione
Filtro privo di rossi: con effetto di intensificazione del contrasto per
Filtro di polarizzazione: per l’esatta valutazione del colore dei tessuti e
Filtro blu: per il riconoscimento di anomalie vascolari o
Con L2 + L3 è possibile utilizzare ogni filtro con ciascuna diaframma.
per la riduzione del riflesso
in pupille di piccolo diametro
ni della retina
eccentrica
la valutazione di minime alterazioni vascolari, ad
esempio emorragie retiniche
per la riduzione della riflessione sulla cornea
emorragie, per indagini oftalmologiche a fluorescenza
5. Dispositivo di messa a fuoco (solo per L3)
Ruotando la ghiera di messa a fuoco è possibile ottenere a diverse
distanze una rapida microregolazione del campo d’indagine da osservare.
Retinoscopi ri-scope® XL 3,5 V
1. Uso previsto
I retinoscopi Slit (a linee) e Spot a punti (detti anche schiascopi), descritti in queste istruzioni per l’uso, sono destinati alla rilevazione della
rifrazione (difetti di rifrazione) dell'occhio.
2. Funzionamento
Con la vite a testa zigrinata è possibile effettuare la rotazione e la messa
a fuoco dell'immagine a linee o a punti.
62
Page 63
3. Rotazione
L'immagine a linee o a punti può essere ruotata di 360° con l'elemento
di comando. Il rispettivo angolo è leggibile direttamente sulla scala graduata del retinoscopio.
4. Cartina di fissazion
Per la schiascopia dinamica le cartine di fissazione vengono agganciate
nel supporto e fissate sul lato dell'oggetto del retinoscopio.
5. Versione Slit (a linee) / Spot (a punti)
Sostituendo la lampadina a linee (versione Slit) con una lampadina a
punti (versione spot) il retinoscopio a linee viene trasformato in un retinoscopio a punti.
Dermatoscopio ri-derma® XL 3,5 V
1. Uso previsto
Il dermatoscopio ri-derma®descritto in queste istruzioni per l’uso è stato
realizzato per l’individuazione precoce di alterazioni cutanee dovute ai
pigmenti (melanomi maligni).
2. Massa a fuoco
Per mettere a fuoco la lente d'ingrandimento, ruotare l'anello oculare.
3. Inserti
Vengono forniti in dotazione 2 inserti:
1) con scala graduata 0 - 10 mm per la misurazione di alterazioni cuta-
nee dovute ai pigmenti, come melanomi maligni
articolo numero 10969
2) Senza scala
articolo numero 10968
Entrambi gli inserti possono essere smontati e sostituiti agevolmente.
Portalampadina a squadra a fibra ottica
ri-scope® XL 3,5 V
1. Uso previsto
Il portalampadina descritto nelle presenti istruzioni per l'uso è destinato
all'illuminazione della cavità orale e della faringe.
Specolo nasale a fibra ottica
ri-scope® XL 3,5 V
1. Uso previsto
Lo specolo nasale descritto nelle presenti istruzioni per l'uso è destinato
all'illuminazione e quindi all'indagine delle narici.
2. Funzionamento
Sono possibili due modalità di funzionamento:
63
Page 64
a) Divaricazione rapida
Con il pollice premere verso il basso la vite di regolazione presente sulla
testa dello strumento. Durante questa operazione non è possibile modificare la posizione delle valve dello specolo.
)
Divaricazione individuale
b
Ruotare la vite di regolazione in senso orario fino a raggiungere la posizione di divaricazione desiderata. Per chiudere le valve ruotare la vite in
senso antiorario.
3. Lente d’ ingrandimento mobile
Sullo specolo nasale è montata una lente orientabile, a ingrandimento di
circa 2,5 x, che a scelta può essere semplicemente estratta e poi reinserita nell'apertura appositamente prevista presente sullo specolo stesso.
Porta abbassalingua a fibra ottica
ri-scope® XL 3,5 V
1. Uso previsto
Il porta abbassalingua descritto nelle presenti istruzioni per l'uso è stato
realizzato per l'indagine della cavità orale e della faringe, in combinazione con abbassalingua in legno e plastica normalmente reperibili in commercio.
2. Funzionamento
Inserire fino all'arresto un abbassalingua in legno o plastica, normalmente reperibile in commercio, nell'apertura situata sotto il foro d'uscita della
luce. Dopo avere eseguito l'esame, è possibile rimuovere agevolmente
l'abbassalingua azionando l'espulsore.
Otoscopio operatorio ri-scope® per uso umano
XL 3,5 V senza specolo
1. Uso previsto
L'otoscopio operatorio RIESTER descritto nelle presenti istruzioni per
l'uso è stato realizzato per l'illuminazione e l'indagine del condotto uditivo, nonché per eseguire piccoli interventi in tale parte del corpo.
2. Posizionamento e rimozione di specoli auricolari per
uso umano
Posizionare lo specolo prescelto sul supporto nero dell'otoscopio operatorio, in modo che la scanalatura presente sullo specolo coincida con la
guida del supporto. Bloccare lo specolo ruotandolo il senso antiorario.
3. Lente d'ingrandimento mobile
Sull'otoscopio operatorio è montata una piccola lente d'ingrandimento
orientabile di 360° con possibilità di ingrandimento di circa 2,5 x.
4. Inserimento di strumenti esterni nell'orecchio
L'otoscopio operatorio è realizzato in modo da potere introdurre senza
problemi nell'orecchio strumenti esterni.
64
Page 65
Otoscopio operatorio ri-scope® per uso veterinario XL 3,5 V senza specolo
1. Uso previsto
L'otoscopio operatorio RIESTER descritto nelle presenti istruzioni per
l'uso è stato prodotto esclusivamente per essere utilizzato su animali, ed
è quindi destinato alla medicina veterinaria. Può essere utilizzato per l'illuminazione e l'indagine del condotto uditivo, nonché per eseguire piccoli interventi in tale parte del corpo.
2. Posizionamento e rimozione di specoli auricolari per
uso veterinario
Posizionare lo specolo prescelto sul supporto nero dell'otoscopio operatorio, in modo che la scanalatura presente sullo specolo coincida con la
guida del supporto. Bloccare lo specolo ruotandolo il senso antiorario.
3. Lente d'ingrandimento mobile
Sull'otoscopio operatorio è montata una piccola lente d'ingrandimento
orientabile di 360° con possibilità di ingrandimento di circa 2,5 x.
Sostituzione della lampadina
Otoscopio L1
Estrarre l’alloggiamento dello specolo dall’otoscopio. Svitare la lampadina ruotandola in senso antiorario. Avvitare la nuova lampadina ruotandola in senso orario e rimontare l’alloggiamento dello specolo.
Otoscopio L2, L3, ri-derma®, portalampadina, specolo nasale e
porta abbassalingua
Svitare la testa dello strumento dal manico portabatterie. La lampadina
si trova sotto la testa dello strumento. Estrarre la lampadina dalla testa
dello strumento trattenendola con pollice e indice oppure utilizzando un
attrezzo adatto. Montare la nuova lampadina.
Oftalmoscopi
Estrarre la testa dello strumento dal manico portabatterie. La lampadina
si trova sotto la testa dello strumento. Estrarre la lampadina dalla testa
dello strumento trattenendola con pollice e indice oppure utilizzando un
attrezzo adatto. Montare la nuova lampadina.
Attenzione: La spina della lampadina va inserita nella tacca di guida
sulla testa dello strumento.
Otoscopi operatori per uso veterinario/umano
Svitare la lampadina dal portalampadina dell’otoscopio operatorio e avvitarne una nuova.
Teste strumenti: Retinoscopio a linea e a punto
Estrarre la testa dello strumento dal manico portabatterie. La lampadina
si trova in una capsula sotto la testa dello strumento. Estrarre la lampadina con la sua capsula dalla testa dello strumento trattenendola con pollice e indice oppure utilizzando un attrezzo adatto. Inserire la nuova lampadina nella capsula e rimontare entrambe nella testa dello strumento in
modo da guidare la spina della lampadina nella tacca.
Avvertenze per la cura dello strumento, pulizia e disinfezione
Tutte le teste strumenti ri-scope® di RIESTER possono essere pulite
esternamente con un panno umido. E’ inoltre possibile eseguire la disinfezione con i seguenti disinfettanti:
65
Page 66
aldeide (formaldeide, glutaraldeide) o tensioattivi. Tutte le parti degli strumenti, escluse le lenti mobili, le lenti d’ingrandimento e i vetri asportabili, possono inoltre essere disinfettati con alcol. Durante l’impiego di questi prodotti, attenersi rigorosamente alle istruzioni per produttore. Per la
pulizia e la disinfezione dello strumento è possibile utilizzare un panno
morbido, possibilmente non sfilacciato, o bastoncini d’ovatta.
Attenzione
Non immergere mai le teste degli strumenti in liquidi. Evitare la penetrazione di liquidi all’interno dell’apparecchio.
a) Sterilizzazione
Secondo un autorevole parere (Centro sperimentale per prodotti medicali di Tubinga), la sterilizzazione è prescritta soltanto in caso di interventi operatori.
b) Specoli auricolari riutilizzabili
La sterilizzazione, pur non essendo necessaria come indicato al punto
a), è tuttavia possibile. Gli specoli auricolari riutilizzabili possono essere
sterilizzati nello sterilizzatore a vapore a 134°C per 10 minuti.
Messa in funzione delle teste strumenti
Montare la testa dello strumento desiderato sul supporto nella parte
superiore del manico, in modo che le due scanalature della parte inferiore della testa si trovino sulle due camme di guida del manico portabatterie. Premere leggermente la testa dello strumento sul manico e ruotare
quest’ultimo in senso orario fino all’arresto.
Per smontare la testa, ruotarla in senso antiorario.
Messa in funzione del dispositivo antifurto
Funzionamento
a
Montare la testa dello strumento deside-
rato sul supporto nella parte superiore del
manico, in modo che le due scanalature
b
della parte inferiore della testa si trovino
sulle due camme di guida del manico por-
tabatterie.
66
Page 67
67
Page 68
Riester bietet eine große Produktauswahl in
➔
den Bereichen
lutdruckmessgeräte l Instrumente für H.N.O.,
B
Ophthalmologische Instrumente l Dermatologische
Instrumente l Thermometer l Stethoskope l Stirnspiegel,
Stirnlampen, Untersuchungslampen l Laryngoskope l
Gynäkologische Instrumente l Perkussionshämmer l
Stimmgabeln l Produkte zur Blutstauung I
Lungendruckmessgeräte l Dynamometer
lDruckinfusionsgeräte l Veterinärmedizinische Instrumente l
Arztkoffer/ -taschen
Die detaillierten Beschreibungen der Produkte
finden Sie unter der jeweiligen Rubrik im
Gesamtkatalog
(Best. Nr. 51231-50). Oder gehen Sie online
unter www.riester.de.
Riester offers a large selection of products in
➔
the areas of
Blood pressure measuring devices I Instruments for ENT,
Ophthalmological instruments I Dermatological instruments I
Thermometers I Stethoscopes I Head mirrors, Head lights,
Examination lights I Laryngoscopes I Gynaecological instruments I Percussion hammers I Tuning forks I Products for
blood stasis I Pulmonary pressure measuring devices I
Dynamometers I Pressure infusion instruments I Veterinary
instruments I Doctor’s cases and bags
Detailed descriptions of the products can be
found in the respective sections of the omnibus
edition catalogue (Order No. 51232-50). Or
online under www.riester.de.
99203 Rev. B · Änderungen vorbehalten · Subject to alterations · Sous réserve de modifications · Sujeto a modificaciones · озможны изменения · Con riserva di apportare modifiche
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.