Gima REUSABLE LARYNGEAL MASK 1,5 User guide

Page 1
Gima S.p.A.
Via Marconi, 1 - 20060 Gessate (MI) Italy gima@gimaitaly.com - export@gimaitaly.com
PROFESSIONAL MEDICAL PRODUCTS
www.gimaitaly.com
MASCHERA LARINGEA RIUTILIZZABILE LARYNGEAL AIRWAY MASK
Manuale d’uso - User manual
ATTENZIONE: Gli operatori devono leggere e capire completamente questo manuale prima di utilizzare il prodotto. ATTENTION: The operators must carefully read and completely understand the present manual before using the product.
34424 - 34425 - 34426 34427 - 34428 - 34429
M34424-M-Rev.4-11.18
0476
Page 2
11 ENGLISH
CONTENTS
1 Intended Use ........................................................................................12
2 Specification ........................................................................................12
2.1 Name of products parts ...................................................................12
3 Preparation for use ..............................................................................13
3.1 Cleaning and sterilization ................................................................13
3.1.1 Cleaning instruction ..................................................................13
3.1.2 Sterilization instruction .............................................................14
3.2 Functional testing ...........................................................................15
3.2.1 Test 1-Visual inspection ...........................................................16
3.2.2 Test 2-Ination/deation test ....................................................16
4 Insertion ...............................................................................................17
4.1 Pre-insertion preparation ................................................................17
4.2 Insertion ..........................................................................................17
4.3 Ination ...........................................................................................18
4.4 Usage .............................................................................................19
5 Removal procedure ............................................................................19
6 Adverse Effects ..................................................................................19
Page 3
12ENGLISH
1. Intended Use
• The Laryngeal airway is an oropharyngeal airway, it is a device inserted into a patients pharynx through the mouth to provide a patient airway.
• GIMA Laryngeal Airway is indicated for use as an alternative to the face­mask for achieving and marinating control of the airway during routine and emergency anesthetic procedures. It is not indicated for use as a replace­ment for the end tracheal tube and is best suited for use in elective surgical procedures where tracheal intubations are not necessary.
• The airway is also indicated in known or unexpected difcult airway si­tuation. It is also indicated as a method of establishing a clear airway du­ring resuscitation in the profoundly unconscious patient with absent gloss pharyngeal and laryngeal reexes who may need articial ventilation. In these cases, the airway should be used only when tracheal intubations is not possible.
2. Specification
Mask Size
Size #1,5 #2 #2,5 #3 #4 #5
Description Infant Infant Child Boy Adult Adult
Recommended weights 5-10 kg 10-20 kg 20-30 kg 30-50 kg 50-70 kg > 70kg
Cuff Ination 7 ml 10 ml 14 ml 20 ml 30 ml 40 ml
Maximum Cuff pressure 60 cmH2O
Table 1: Selection guildlines for GIMA Laryngeal Airway.
2.1 Name of product parts
Cuff
Air Connector
Airway Tube
Pilot Tube
Pilot Balloon
Ination Valve
Page 4
13 ENGLISH
3. Preparation for use
3.1 Cleaning and sterilization
Cleaning and sterilization as described below must be carried out for each subsequent use.
The GIMA Laryngeal Airway can be used a maximum of 40 times by atten­ding the recommended cleaning, sterilization and handling procedures. Pro­per cleaning and sterilization of the device is essential to ensure continued safe usage up to 40 times.
GIMA has validated the instructions provided below as being capable of preparing an GIMA Laryngeal Airway for re-use. It remains the responsibi­lity of the processor to ensure that processing as actually performed using equipment, materials and personnel in the processing facility achieve the desired result. This requires validation and routine monitoring of the pro­cess. Likewise any deviation by the processor from the instructions provi­ded should be properly evaluated for effectiveness and potential adverse consequences.
CAUTION
Handle the GIMA Laryngeal Airway carefully as it is made of silicone, which can be torn or punctured. Avoid contact with sharp or pointed objects.
WARNING
Do not replace or disassembly any components of the GIMA Laryngeal Airway as it may cause product failures.
3.1.1 Cleaning instruction
It is recommended to follow the described validated manual cleaning proce­dure to ensure proper cleaning.
CAUTION
Do not expose the ination valve to any liquid as it may cause premature ination valve failure.
The GIMA Laryngeal Airway should be kept moist between the time of use and subsequent cleaning. Remove excess soil with disposable cloth/paper wipe. Thoroughly rinse the GIMA Laryngeal Airway in cold running water to avoid protein coagulation.
Page 5
ENGLISH
14
Wash the device thoroughly in water using a mild detergent until all adhe­rent visible soil is removed. It is recommended to wash the GIMA Laryngeal Airway with Endozime manufactured by Ruhof. The working solution should be prepared by using 35 ml Endozime per 4 liter warm water (one ounce of Endozime to 1 gallon of warm water).
CAUTION
All detergents should be used in the recommended dilution/concentration, temperature, water quality (e.g., ph, hardness), and exposure (soak or con­tact) time, in accordance with the detergent manufacturer’s directions.
WARNING
Do not use germicides, disinfectants, or chemical agents such as glutaral­dehyde, ethylene oxide, phenol-based cleaners, iodine-containing cleaners or quaternary ammonium compounds to clean or sterilize the GIMA Laryn­geal Airway. The material absorbs such substances, resulting in exposure of the patient to potentially severe tissue burns and possible deterioration of the device. Do not use an GIMA Laryngeal Airway that has been exposed to any of these substances.
To ensure proper cleaning of the airway tube it is recommended to use a soft bristle brush. Brushes should have a tight t, but be able to move back and forth in the area being cleaned. Never use hard brushes or other materials, which might damage the silicone cuff or surface.
Remove all cleaning residues by thoroughly rinsing the cuff and airway tube in running tap water for 1 minute. Make sure that the water passes throu­gh the tube. Closely examine the GIMA Laryngeal Airway to ensure that all visible foreign matter has been removed. Repeat the above procedure as necessary.
CAUTION
Effective cleaning must be carried out to achieve proper sterilization.
3.1.2 Sterilization instruction
The only recommended sterilization method is steam autoclaving. Strictly adhere to the following validated instructions to ensure sterilization and to prevent damage.
Page 6
15
ENGLISH
Immediately prior to steam autoclaving, deate the cuff completely. Ensure that the valve and syringe used to deate the cuff are dry.
Caution: Any air or moisture left in the cuff will expand during autoclaving process and will cause irreparable damage to the cuff and /or pilot balloon.
If a deated mask immediately and spontaneously reinates (even slight­ly),do not autoclave or reuse this mask. This may indicate the leak or valve malfunctioning. The GIMA Laryngeal Airway should be placed in an appropriate steam au­toclave-proof bag before steam autoclaving. Ensure that the pack is large enough to contain the device without stressing the seals. It is recommended to use the following sterilization cycle according to AAMI TIR 12:2004
Item Exposure time
Drying time
At 135°C (275°F)
Wrapped utensils 10 min 30 min
Table 2. Parameters for gravity-displacement steam sterilization cycles
The record card should be completed each time the GIMA Laryngeal Airway is sterilized. After sterilization the GIMA Laryngeal Airway should be stored in accordance with accepted hospital practice. The GIMA Laryngeal Airway should not be exposed to direct sunlight or or elevated temperatures during storage. Store in unopened bags at temperature between 10°C/50°F and 25°C/77°F
3.2 Functional testing
Functional testing as described below must be carried out before using the device. The tests should be conducted in a manner consistent with accep­ted medical practice that minimizes contamination of the GIMA Laryngeal Airway prior to insertion.
WARNING
Do not use and destroy the device if any one test fails. Dispose of the GIMA Laryngeal Airway in a safe manner according to local guidelines of medical waste.
CAUTION
Always wear gloves during the preparation and insertion of the GIMA Laryn­geal Airway to minimize contamination.
Page 7
ENGLISH
16
3.2.1 Test 1-Visual inspection
Closely examine the surface of the GIMA Laryngeal Airway for any damage, perforation, scratches, etc. Do not use the GIMA Laryngeal Airway if it is damaged in any way. Check that the interior of the tube and cuff are free from blockage and any loose parts. Parts and blockages should be removed as these may prevent the device from functioning properly. Do not use the GIMA Laryngeal Airway if any of loose parts or blockages cannot be remo­ved. Check that the airway connector on the GIMA Laryngeal Airway is tted tightly to the airway tube. Ensure that it cannot easily be pulled off. Do not twist the connectors as this may break the seal. Closely examine the pilot balloon for any damages.
WARMING
Do not use the GIMA Laryngeal Airway if the mask connector does not t tightly into the outer end of the airway tube.
3.2.2 Test 2-Inflation/deflation test
Deate the cuff of the GIMA Laryngeal Airway completely. Once deated, check the cuff thoroughly for any wrinkles or folds. Over-inate the cuff to the appropriate volume as specied in Table 3. Check that the inated cuff is symmetrical and smooth. There should not be any bulge nor any sign of leakage in the cuff, pilot tubing or balloon.
WARNING
Do not use the GIMA Laryngeal Airway if there are any bulges on the cuff or if there are any signs of leakage.
Mask size
#1,5 #2 #2,5 #3 #4 #5
Over-ination
10 ml 15 ml 21 ml 30 ml 45 ml 60 ml
Cuff volumes
Table 3. Over-inflation volumes for the GIMA Laryngeal Airway.
CAUTION
The ination volumes specied in Table 3 are for testing purposes only. The­se volumes are not to be used during normal use of the device the recom­mended standard ination volumes can be found in Table 1.
Page 8
17
ENGLISH
4. Insertion
4.1 Pre-insertion preparation:
Before insertion of the GIMA Laryngeal Airway, the cuff should be comple­tely deated so that the cuff is at and free of wrinkles. Simply press the cuff down onto a at sterile surface (e.g. a piece of sterile gauze) while at the same time deating the device with a syringe. Complete deation results in a shape like the rim of a saucer, and facilitates insertion and correct positio­ning of the device. To further facilitate insertion into the patient, a sterile, wa­ter-based lubricant (e.g. K-Y Jelly®) should be applied to the distal posterior surface of the cuff (local anesthesia is not recommended.)
WARNING Lubricate only the posterior tip of the cuff to prevent blockage of the airway aperture or aspiration of the lubricant.
4.2 Insertion
Before insertion, it is essential that all clinicians using the GIMA Laryngeal Airway are familiar with the warnings, precautions and indications, found in these directions for use.
The following points are extremely important
- Check for correct deation and lubrication as described above.
- The size of the GIMA Laryngeal Airway must t the patient. Use the guide-
lines in Table 1 combined with clinical judgement to select the correct size.
- Always have a spare GIMA Laryngeal Airway ready for use.
- Pre-oxygenate and use standard ready for use.
- Check that the level of anesthesia (or unconsciousness) is adequate befo-
re attempting insertion.
- The head of the patient should be extended with exion of the neck in a
position normally used for tracheal intubation (i.e “the snifng position”)
- Never use excessive force.
Page 9
ENGLISH
18
4.3 Inflation
After insertion, the line on the airway tube should be oriented anteriorly toward the patient’s nose. Without holding the tube, inate the cuff with just enough air to obtain a seal, equivalent to intracuff pressures of approxima­tely 60 cmH2O. In many cases, only half of the maximum volume is sufcient to achieve a seal-please refer table 1 for maximum volume. Check the cuff pressure at start and periodically, either with a cuff pressure gauge or by feeling the tension in the pilot ballon. This is especially important when N2O gases are used.
Never over-inflate the cuff. Avoid prolonged intracuff pressures greater than 60 cmH2O. The initial cuff pressure varies according to patient, mask size, head position, and depth of anesthesia. Do not hold the tubing du­ring ination as this prevents the mask from seating itself correctly. A small outward movement of the tube is often seen as the mask settles into the hypopharynx. To avoid overination, it is very important to strictly adhere to the cuff-in­ation volumes stated in table 1. Over-ination can be entirely avoided by completely deating the cuff prior to insertion by withdrawing all of the air with a suitable syringe. This is the method recommended by GIMA. In instances where an alternative technique is adopted, for example if the cuff is inserted in a neutral or semi-inated state, there is a risk that the cuff may be over-inated. Extra care therefore has to be taken after insertion to compensate for the air already in the mask when subsequently inating the cuff. The maximum extra volume depends on mask size and initial volume of air in the mask when inserted.
WARMING
Never overinate the cuff after insertion. Looking for the following signs of correct placement: the slight outward movement of the tube upon cuff ination, the presence of a smooth oval swelling in the neck around the thyroid and cricoid area, or no cuff visible in the oral cavity.
Page 10
19
ENGLISH
4.4 Usage
5. Removal procedure
Removal should always be carried out in an area where suction equipment and the facility for rapid tracheal intubation are available. Do not fully deate the cuff until after its removal to avoid secretions entering the larynx and to prevent larygospasm. Alternatively, it may be removed moderately ina­ted to aid complete removal of secretions. If the mask is to be removed in the Post-Anesthesia Care Unit, recovery room staff should receive thorough training in all aspects of the GIMA Laryngeal Airway.
6. Adverse Effects
Use of the GIMA Laryngeal Airway may cause minor adverse effects (e.g., sore throat) and major adverse effects (e.g. aspiration).
Page 11
Codice prodotto Product code
Numero di lotto Lot number
Attenzione! Leggere le avvertenze d’uso Caution! Read instructions carefully
Fabbricante Manufacturer
Prodotto conforme alla Direttiva Europea n. 93/42/CE
0476
Product complies with European Directive no. 93/42/CEE Conservare in luogo fresco e asciutto
Keep in a cool, dry place Conservare al riparo da fonti di calore e luce solare
diretta Keep away from sunlight
Consultare le istruzioni prima dell’uso Read instructions carefully
Data di scadenza Expiration date
Data di fabbricazione Date of manufacture
Sterilizzato mediante ossido di etilene Sterilised using ethylene oxide
20
Loading...