METRAX GmbH reserves all rights to these operating instructions. These operating
instructions may not be duplicated or made accessible to third parties without the approval of
METRAX GmbH. The same applies to individual parts or excerpts of these operating
instructions.
Non-compliance gives grounds to a right to claim damages and can have consequences under
criminal law (refer to DIN 34).
We reserve the right to make amendments to these operating instructions.
Table of contents
Table of contents
1 Glossary 5
2 Introduction 6
2.1 Foreword 6
2.2 Validity 6
2.3 Guarantee 6
2.4 Disclaimers 7
2.5 Symbols used in these operating instructions 8
2.6 Pictogrammes 9
2.7 Summarized operating instructions 11
3 Intended use 12
3.1 Indication/Contraindication for Defibrillation 13
3.1.1 Indications 13
3.1.2 Contraindications 13
4 Safety information 14
4.1 General safety advice 14
5 Description of device 15
5.1 General description 15
5.2 Description of device details 16
5.3 Status display 21
5.4 Capacity display on the monitor 22
5.5 Data management 23
5.6 Accessories 24
5.6.1 Standard accessories 24
5.6.2 Optional accessories 24
6 Preparatory measures before (initial) start-up 25
6.1 Unpacking 25
6.2 Inserting / Replacing the SaveCard 26
6.3 Power module 26
6.3.1 Insert energy module. 27
6.3.2 Remove the power module 28
6.4 PRIMEDIC™ Battery 28
6.5 PRIMEDIC™ AkuPak LITE 29
6.6 Connecting the PRIMEDIC™ PowerLine (mains unit) (optional accessory) 30
7 HeartSave self-tests 31
7.1 Self-test after switching the HeartSave on 31
7.2 Automatic, periodic self-tests 31
7.3 Tests during equipment operation 31
8 Configuration 32
8.1 Simple change of configuration – example: Time 33
8.2 Changing the PIN 34
8.3 Calling up/activating a profile 34
8.4 Saving menu parameters in a profile 34
8.5 Ratio chest compression:ventilation in Pediatric Mode 35
9 Operating the HeartSave and sequence of resuscitation 36
9.1 Switching the HeartSave on 36
9.2 Examining and preparing the patient 36
9.3 Defibrillation 37
III
Table of contents
9.3.1 Defibrillation in Adult Mode 38
9.3.2 Defibrillation in Pediatric Mode 39
9.4 Opening SavePads and placing electrodes 41
9.5 Plug in electrode cable 42
9.6 Check electrodes 43
9.7 Carrying out the ECG analysis 44
9.8 Defibrillation required 44
9.9 Defibrillation not required 45
9.10 Switching HeartSave off 46
9.11 Keeping the defibrillator ready for use 46
9.12 Monitoring the patient 46
10 Cleaning, maintenance and despatch 47
10.1 Cleaning 47
10.2 Servicing 47
10.2.1 Servicing check list 48
10.3 Sending the HeartSave 48
11 Disposal 49
12 List of error codes 50
13 Technical Data 55
14 Warranty conditions 58
15 Depiction of the current time functions 59
15.1 Adult Mode 59
15.2 Pediatric mode 61
16 Rhythm detection system 63
16.1 Adult mode 64
16.2 Pediatric mode 64
17 Guidelines and manufacturer's declaration – Electromagnetic emissions 66
18 Safety checks 71
19 Appendix 72
19.1 Index of diagrams 72
IV
Glossary
Operating instructions
HeartSave AED-M 23702 / GB / A
5 / 74
Term / abbreviation
Description
AED
Automated external defibrillator
AHA
American Heart Association
Biphasic impulse
The current flow of the defibrillator changes its direction during shock
appliance its direction
BLS
Basic measures of resuscitation / Cardio pulmonary resuscitation
(Basic Life Support)
CPR
Cardio pulmonary resuscitation
EAR
Used Electronic Appliances Register
ECG
Electrocardiogram
ElektroG
German Electrical Equipment Act
ERC Guidelines
European Resuscitation Council guidelines
on Cardio Pulmonary Resuscitation (CPR)
EU
European Union
CPR
Cardio pulmonary resuscitation
Medical devices log
Documentation of all data for a medical devices log according to § 7
of the Ordinance on the Operation and Use of Medical Devices
(MPBetreibV)- to be maintained by operator, including serial number,
test data, instructions, safety checks.
Metronome
Metronome for chest compressions
MDD
Medical Device Directive
MIT
Massachusetts Institute of Technology
MPBetreibV
Medical Device Operator Ordinance
MPG
Medical Devices Act
ÖRE
Public law
Patient impedance
Patience resistance between the SavePads
PTB
Physikalisch-Technische Bundesanstalt (Federal German Physical
Technical Institute)
SaveCard
Memory card for data transfer
SavePads
Defibrillation electrode
WEEE
engl. Waste of Electrical and Electronical Equipment
(in German: Elektro- und Elektronikgeräte-Abfall)
1 Glossary
Introduction
6 / 74
HeartSave AED-M 23702 / GB / A
Operating instructions
2 Introduction
2.1 Foreword
Dear User,
You are preparing to face the task of using the PRIMEDIC™ HeartSave AED/AED-M in a
medical emergency on human beings!
So that you react quickly and properly in these special circumstances and make optimal use of
the opportunity the device provides you with , it is necessary for you to take your time carefully
to read through these operating instructions beforehand, thus familiarising yourself with the
device, its functions and applications.
Keep these operating instructions near the device so that you consult them for any queries
which may arise.
If you have any questions regarding the device or other PRIMEDIC™ products, we are glad to
be at your disposal.
You will find our contact address on the masthead at the start of these operating instructions.
The instructions given on the device are no substitute for reading these operating instructions.
2.2 Validity
The descriptions in these operating instructions refer to the PRIMEDIC HeartSave AED-M
made by METRAX GmbH. Both units are referred to as HeartSave in the following operating
instructions.
The content of this document can be changed without prior notice.
2.3 Guarantee
The warranty period is 24 months and starts on the day of delivery. The guarantee conditions
and additional information can be found at www.primedic.com
Introduction
Operating instructions
HeartSave AED-M 23702 / GB / A
7 / 74
2.4 Disclaimers
Liability claims in the event of damages to people or property are excluded if they are based
on one or more of the following reasons:
Using the device in a manner for which it was not intended.
Improper use and maintenance of the device.
Operating the device with the protective covers removed or if there is obvious
damage to cables and/or electrodes.
Non-compliance with the instructions in these operating instructions with regard to
operation, maintenance and repair of the equipment.
Using accessories and spare parts made by other manufacturers.
Autonomous intervention, repairs or constructional changes to the device.
Autonomous exceeding of the performance limits.
Lack of monitoring of parts that are subject to wear and tear.
Treating patients without prior indication.
Introduction
8 / 74
HeartSave AED-M 23702 / GB / A
Operating instructions
DANGER
Texts marked DANGER indicate an extraordinarily serious, current danger which will
definitely lead to serious injury or even death if no preventative measures are adopted.
You must follow these instructions!
WARNING
Texts marked WARNING indicate extraordinarily serious possible dangers which, should no
preventative measures be taken, may lead to serious injury or even death.
You must follow these instructions!
CAUTION
Texts marked with CAUTION indicate a possible dangerous situation which could lead to
minor injuries.
You must follow these instructions!
ATTENTION
Texts marked with ATTENTION indicate possible property damage.
You must follow these instructions!
Note
This symbol indicates text which contains important advice / comments or tips.
First instruction
Second instruction
etc. This point marks lists
(3)
Numbers in brackets refer to items in diagrams.
< ... >
Texts set in angle brackets denote acoustic information / instructions for the device
which are shown simultaneously on the monitor, depending on the device model.
2.5 Symbols used in these operating instructions
The instructions are described in the following manner. Follow the instructions in the order in
which they are described in the instructions.
Introduction
Operating instructions
HeartSave AED-M 23702 / GB / A
9 / 74
Certification authority
Protection against contact and dust deposits on the inside and against
water spray (jet) from any angle. Details on the unit only valid with
energy module fitted.
Protect against contact and dust deposition inside and against falling
spray water up to 60° from a vertical direction. Details on the energy
module is for this one alone.
Follow the operating instructions
Safety symbol "General warning symbols"
The individual meanings are explained in the operating instructions
Do not dispose of device in domestic refuse.
Dangerous electric voltage (high voltage)
Degree of protection BF
Durability of the internal battery MM/YYYY
Protect battery from fire
Do not charge the battery
Do not reuse
Observe the operating instructions
not sterile
Good for 1 day after opening
2.6 Pictogrammes
Introduction
10 / 74
HeartSave AED-M 23702 / GB / A
Operating instructions
Storage temperature in Celsius and Fahrenheit
Protect from sunlight
Store in a dry place
Latex-free
Remove protective film from adhesive electrodes
Batch code
Manufacturer
Order number
Can be used until YYYY/MM
At the end of its life, send the power module for recycling. Device
contains Lithium-Ion cells.
Introduction
Operating instructions
HeartSave AED-M 23702 / GB / A
11 / 74
Only open package If unconscious and
not breathing!
Only press pediatric
button with patients
< 25 kg
Apply electrodes
Shave patients with
hair
Remove protective
film
Apply electrodes
Press shock button
when instructed to
do so by device!
SavePads PreConnect AED
2.7 Summarized operating instructions
The brief instructions can be found on the utensils carrier and helps you with the use of the
HeartSave.
Intended use
12 / 74
HeartSave AED-M 23702 / GB / A
Operating instructions
Note
HeartSave defibrillators may only be used as described and under the
conditions detailed in these operating instructions!
3 Intended use
The PRIMEDIC™ HeartSave AED-M is intended for use by expert medical practitioners
working under the instructions issued by a doctor, and for doctors who frequently need an AED
for emergency situations that they encounter in the course of their activities.
The integrated voice messages make it possible for the PRIMEDIC™ HeartSave AED-M to be
used by laymen. They must be instructed on the unit and be trained in the execution of lifesaving measures (BLS), but whose level of knowledge is unknown at the time of the event.
The PRIMEDIC™ HeartSave AED-M is suitable for domestic use and in medical premises.
The device is intended for use on patients with symptoms of a sudden heart death who are
unconscious (do not respond to speech) who are not breathing.
The user is guided by the device with acoustic notices (voice message) and optical indications
as well as by the device identification, so that the defibrillation electrodes are fixed on the body
of the patient and the BLS measures, chest compressions and rescue breathing according to
the current recommendations of the ERC or AHA can be carried out. The first aider is
requested to step back from the patient in order to perform the analysing rhythm and apply a
shock. The unit monitors and analyses the cardiac rhythm of the patient, loads the capacitor
according to patient impedance and delivers the energy with a constant-current biphasic shock
when the user presses the trigger button. The first 3 shocks are based on the shock strategy
with the shock steps 20A (281J at 50 Ohm), 25A (350J at 50 Ohm) and 30A (360J at 50 Ohm)
From the third shock on, all further shocks are delivered with the 30A shock step (360J at 50
Ohm). In pediatric defibrillation mode, the defibrillation energy is reduced to 50 J (1st shock),
70J (2nd shock) and 90 J (3rd and subsequent shocks) at 50 Ohm. For safety reasons, no
shock is delivered with asystole, as no therapeutic effect is to be expected. Controlled
ventricular electrical activity caused by supraventricular tachycardia such as atrial fibrillation,
atrial flutter, ventricular extra-systoles and idioventricular rhythms do not lead to a shock being
delivered.
In combination with the ECG monitoring cable, it is possible to carry out restricted
rhythmological monitoring of the patient for a short period (a few hours) in the presence of
medically-trained personnel.
The PRIMEDIC™ HeartSave AED-M is designed for treating adult patients in combination with
the PRIMEDIC™ SavePads PreConnect, PRIMEDIC™ SavePads C or PRIMEDIC™
SavePads Connect disposable electrodes. Children aged 8 years and above and/or with a
weight over 25 kg are treated as adults.
Using the coded defibrillation electrodes PRIMEDIC™ SavePads mini, the PRIMEDIC™
HeartSave AED-M can also be used on children aged 1-8 weighing less than 25 kg. By coding
the defibrillation electrodes, the maximum defibrillation energy is reduced to a maximum of 90
J according to the current escalation levels. If in an actual emergency these electrodes are not
available, the user can switch the defibrillator into the Child defibrillations mode manually. Here
too, the maximum defibrillation energy is limited to 90 J even if the defibrillation electrodes for
adults are connected.
Intended use
Operating instructions
HeartSave AED-M 23702 / GB / A
13 / 74
DANGER
Warning: physical harm
Risk of heart arrhythmia which may lead to death
Only use HeartSave as intended
3.1 Indication/Contraindication for Defibrillation
3.1.1 Indications
Only use the HeartSave when the patient:
is unconscious and
not breathing
and is older than 1 year
3.1.2 Contraindications
The HeartSave must not be used if the patient:
is conscious or
is breathing normally or
is a child less than one year old
Safety information
14 / 74
HeartSave AED-M 23702 / GB / A
Operating instructions
4 Safety information
4.1 General safety advice
Read the operating instructions carefully before using HeartSave for the first time. Only use
HeartSave as described in the operating instructions.
Note the ambient conditions in the technical specifications when storing and operating the
device.
Always follow the commands issued by HeartSave!
Use HeartSave on a non-conductive base. Do not use HeartSave in standing water or rain.
Do not use HeartSave in the presence of flammable materials.
Both in conjunction with its accessories and the optional accessories, and also individually,
HeartSave fulfils the currently applicable safety standards and complies with the provisions of
the medical products regulations.
HeartSave and its accessories are safe when used as intended and when following the
descriptions and information detailed in these operating instructions.
Despite this, if used incorrectly, the HeartSave and its accessories can be dangerous to the
user, the patient or third parties.
Keep the device away from children!
Applicable for Europe:
HeartSave satisfies Medical Device Directive - MDD 2007/47/EU.
For Germany and Austria, the following is also applicable:
HeartSave complies with the Medical Devices Law (MPG) and is subject to the
Ordinance on the Operation and Use of Medical Devices (MPBetreibV).
According to the Ordinance on the Operation and Use of Medical Devices
(MPBetreibV), the device must be subjected to the regular checks explained in the
appendix.
According to the Ordinance on the Operation and Use of Medical Devices
(MPBetreibV), a medical devices log needs to be kept for the device. Regular
checks of the equipment are to be documented there.
For the other states in the European Community, national regulations for operating medical
devices apply.
Description of device
Operating instructions
HeartSave AED-M 23702 / GB / A
15 / 74
Note
The wall bracket and accessories are described in separate operating
instructions.
5 Description of device
5.1 General description
The PRIMEDIC™ HeartSave is an automatic external defibrillator (AED) with an integrated
Single Channel ECG.
The ECG is recorded using the PRIMEDIC™ SavePads. HeartSave can detect potentially fatal
heart arrhythmia. HeartSave generates the electric shock needed to bring the patient back to
consciousness (defibrillation). This method is the generally recognized therapy.
The PRIMEDIC™ HeartSave product family has been designed to be safe and quick to use in
an emergency. All functional units and operating elements are subject to the following
principles:
Clear organisation of functional units
Reduction of functions to those necessary
Intuitive and logical operator guidance
Clear, self-explanatory operating elements
Ergonomic layout.
The defibrillator unit has been optimised to be safe and very quickly ready to use. The loading
time for a defibrillation is approx. 12 seconds with a battery capacity of approx. 90 % of the
rated value.
Description of device
16 / 74
HeartSave AED-M 23702 / GB / A
Operating instructions
Fig. 1: Front view with cover
(1) Status display
(2) Strap to pull the cover off the device (with expiry date of SavePads)
(3) Carry handle
(4) Cover of the device
Fig. 2: Rear View
(1) Identification plate
(2) Fixings for wall mounting
5.2 Description of device details
Description of device
Operating instructions
HeartSave AED-M 23702 / GB / A
17 / 74
Fig. 3: View from below (without energy module)
(1) Contacts for energy module
(2) Slot for SaveCard
(3) Release button, SaveCard
(4) Release button, energy module
Description of device
18 / 74
HeartSave AED-M 23702 / GB / A
Operating instructions
Fig. 4: PRIMEDIC™ HeartSave AED-M front view
(1) Pediatric button
(2) Jack for electrode plugs
(3) Connector symbol with LED
(4) Monitor
(5) On/Off switch
(6) Button to navigate upwards or to increase parameters
(7) Select/ confirm button
(8) Button to navigate downwards or to reduce parameters
(9) Loudspeaker
(10) Shock button (shock button for defibrillation)
Note
You can change the language in the Setup Menu on Page 3. You can choose
from 4 languages. The individual language is quoted briefly after activation.
When switched on, the device will start with the language that was active when
it was switched off previously.
Description of device
Operating instructions
HeartSave AED-M 23702 / GB / A
19 / 74
Illustration similar
Fig. 5: Monitor illustration
(1) Status line to show CF card capacity, patient impedance, time of day, microphone,
battery capacity
(2) Heart rate
(3) Number of defibrillations
(4) Number of detected VFs
(5) Display of switch-on duration / time sequence CPR cycle
Voice instructions will be issued at regular intervals while HeartSave is in
operation. The battery symbol in the status display is displayed.
5.3 Status display
In the table below is a list of the possible things displayed in the status display and their
meanings.
The battery is monitored using an electronic charge balance process.
Description of device
22 / 74
HeartSave AED-M 23702 / GB / A
Operating instructions
100 % charged
80 % charged
60 % charged
40 % charged
20 % charged
0% (device runs on until charge is exhausted)
Fault in the device or service life of the power module has expired
Audible warning.
Display on monitor
AkuPak LITE
< Charging status battery low,
please recharge >
Charging status battery low, please
recharge
Battery
< Battery low, please replace battery
>
Battery low, please replace battery
5.4 Capacity display on the monitor
With the HeartSave AED-M, the battery charge of the battery / AkuPak appears on the display.
The different images that may be displayed have the following meaning:
The optional AkuPak LITE and the battery are monitored by means of electronic charge
measurement to ensure the most accurate capacity display possible. In addition to this display,
all HeartSave units issue a warning if the battery is about to be exhausted.
If the device is being used, the corresponding voice prompt will be repeated regularly in the
selected language.
The battery symbol in the status display is displayed.
Description of device
Operating instructions
HeartSave AED-M 23702 / GB / A
23 / 74
Note
HeartSave automatically records all deployment data (ECG, environmental
noise etc.) onto a removable SaveCard.
Note
The data saved on the SaveCard should be archived after every deployment if
possible. Delete the data from the SaveCard after archiving.
5.5 Data management
The stored data can be viewed using a PC / laptop and the software PRIMEDIC ™ ECG
Viewer (optional accessory). However, this data may not be used for diagnostic purposes or
for therapy for the patient. It should only be used for administrative or legal purposes. The
software has a deployment log into which additional patient data can be entered.
Once the storage capacity or the maximum number of files of the SaveCard is exhausted, no
further data will be saved.
Once the storage capacity of the SaveCard is exhausted, no further data will be saved. The
device remains ready for operation even if the memory is exhausted and even without a
SaveCard.
Operating the software is described separately.
The SaveCard supplied with the device is already formatted and can be used straight away. If
you have any problems with the available SaveCard or new CF cards, you have to reformat
them with the FAT16 file system. Therefore, when formatting the SaveCard, ensure that you
do not accidentally transfer the FAT32 file system.
Proceed as follows:
For Windows Vista, Windows 7, Windows 8, Windows 8.1
Start a command line window using "Start->Execute" and in the entry field, enter
"cmd.exe". The command line window will then open.
There you enter the following: format f: /U /FS:FAT /X /V: (where f: stands for the
drive letter of the CF card reading device which you may have to adjust).
Description of device
24 / 74
HeartSave AED-M 23702 / GB / A
Operating instructions
Fig. 7: PRIMEDIC™ SavePads PreConnect (unpacked)
(1) Defibrillation electrodes with protective film
(2) Electrode plug
5.6 Accessories
The accessories need to be secured appropriately before being transported.
5.6.1 Standard accessories
Batterie 6, Order No. 97641 / AkuPak LITE, Order No. 97196 (depending on version)
SavePads PreConnect, Order No. 97085
5.6.2 Optional accessories
PRIMEDIC™ SavePads mini, Order No.: 97534
PRIMEDIC™ SavePads Connect Cable 12, Order No.: 97384
PRIMEDIC™ SavePads Connect (1 Pair), Order No. 96516
PRIMEDIC™ SavePads Connect (5 Pairs), Order No. 96710
ECG electrodes, Order No. 96592
Carrying case with storage compartments, Order No.: 96379
Wall-mounting box SaveBox, Order No.: 96740
Wall-mounting box with Alarm SaveBox Advanced, Order No.: 96776
Wall-mounting bracket with unlocking facility, Order No.: 96378
Defibrillator Signs Set 1 Order No.: 97016
Monitoring cable, 2-core 12, Order No. 97385
Subject to change without notice.
Preparatory measures before (initial) start-up
Operating instructions
HeartSave AED-M 23702 / GB / A
25 / 74
DANGER
Danger caused by damaged device
Risk of burns and heart arrhythmia as the result of electric shock
Do not use damaged devices
6 Preparatory measures before (initial) start-up
6.1 Unpacking
After delivery, first of all check the packaging and the device for transport damage.
If you notice any damage to the device, immediately contact your transport company, dealer or
directly contact technical services at METRAX GmbH, stating the serial number and describing
the damage to the device.
Remove the insulation film between the energy module and the unit. Proceed as described in
Chapter 0.
Convince yourself that the scope of delivery is complete in accordance with the enclosed
delivery note.
Preparatory measures before (initial) start-up
26 / 74
HeartSave AED-M 23702 / GB / A
Operating instructions
Fig. 8: Inserting / replacing the SaveCard
Note
The data saved on the SaveCard should be archived externally after every
deployment if possible. Once the storage capacity of the SaveCard is
exhausted, no further data will be saved. The device remains ready for
operation even if the memory is exhausted and even without a SaveCard.
To read out the saved data, you can use the software PRIMEDIC™ ECG
Viewer which is available as an optional accessory.
Note
For the first message "Battery low, please replace battery / Charging status
battery low, please recharge" there are still at least 3 energy discharges at
max. energy available. The energy module should be replaced if this message
appears.
Note
Check the battery level after each use. If necessary, the battery should be
replaced by a new one.
6.2 Inserting / Replacing the SaveCard
To remove the SaveCard or to change it, firstly remove the power module.
Procedure:
Press the button (2) in fully – this pushes the SaveCard (1) slightly out of its holder.
Completely remove the SaveCard from the device and transfer the data (if
applicable) onto a PC and insert this card, or a new one, in the device with the plug
end first.
Gently press the card in until the button (2) projects slightly out of the device.
Finally insert the power module into the device again.
6.3 Power module
Before using the HeartSave for the first time, you must insert the power module in the specially
designed slot.
Preparatory measures before (initial) start-up
Operating instructions
HeartSave AED-M 23702 / GB / A
27 / 74
Fig. 9: Insert energy module
Note
If the battery has been installed correctly, the device will start independently
once the cover of the housing has been removed and it will run a self-test. Now
follow the acoustic instructions from the device and then switch it off. Now the
device is ready to use.
ATTENTION
Danger: faulty device
Device is not operational
Only use the device when the status display indicates an 'OK'
6.3.1 Insert energy module.
Procedure:
Lay the device on its back.
Push the (new) battery (1) in the direction of the arrow (3) into the device until it
reaches its end position as shown in the diagram.
Then press the battery in the direction of the arrow (4) into the power module slot
until the release button (2) locks the power module tongue securely into position.
Press the battery completely into the device until you hear the "click" when it slots
into place and the battery
is flush with the outside edge of the device.
After this, the device will carry out a self-test and is ready to use.
If the status display does not indicate an 'OK', proceed as follows:
Switch on the device and wait for the result of the self-test.
Preparatory measures before (initial) start-up
28 / 74
HeartSave AED-M 23702 / GB / A
Operating instructions
Note
Only replace the energy module when the device is switched off and the
defibrillation electrodes plug is disconnected.
Fig. 10: Removing the energy module
WARNING
Do not charge the battery
Risk of explosion
Replace the flat battery
ATTENTION
Note the battery expiration date
Device is not operational
Replace batteries after passing their expiration date
6.3.2 Remove the power module
Procedure:
Lay the device on its back.
Press the unlocking button (2) to the right until the tongue on the power module is
released and the power module (1) snaps out of the slot slightly.
Twist the power module slightly in the direction of the arrow (4) and then pull it in the
direction of the arrow (3) out of the device.
6.4 PRIMEDIC™ Battery
The battery is a disposable lithium battery. It is fully charged when delivered. This type of
battery is state-of-the-art and was selected due to its extremely long service life and energy
storage.
Preparatory measures before (initial) start-up
Operating instructions
HeartSave AED-M 23702 / GB / A
29 / 74
Note
If the device has to be sent away to technical services, remove the battery
before sending it and put some adhesive insulation tape over its contacts.
Observe the separate shipping regulations when sending the battery.
Fig. 11: PRIMEDIC™ AkuPak LITE battery charge indicator
(1) Push-button to activate battery charge indicator
(2) Battery charge indicator
81% - 100 % charged
41% - 60% charged
1% - 20% charged
Note
When charging using the Charger Basic, the PRIMEDIC™ AkuPak is
automatically charged up again fully if the level falls below 80% of the charge
capacity.
This 80%-limit can temporarily be made inoperative by pressing 3 seconds on
push button (2), i.e., the PRIMEDIC™ AkuPak can be recharged even before
reaching this limit. This, for example, is practical if you wish to fully recharge
PRIMEDIC™ AkuPak prior to next use, independent of its current charge
status. After recharging, the programmed 80%-limit is operative for the next
automatic full recharge.
Heed the documentation enclosed with the battery and keep it safe with these operating
instructions.
6.5 PRIMEDIC™ AkuPak LITE
Battery charge indicator (1) means:
Preparatory measures before (initial) start-up
30 / 74
HeartSave AED-M 23702 / GB / A
Operating instructions
Note
Charging the PRIMEDIC™ AkuPak LITEs outside the operating temperatures
quoted in the appendix can cause damage to the rechargeable battery.
Fig. 12: PRIMEDIC™ PowerLine
Note
METRAX GmbH recommends that you keep the HeartSave unit, with the
PRIMEDIC™ PowerLine inserted, continuously connected to the mains, so that
the equipment self-tests are carried out automatically.
A completely discharged battery must be charged for at least 2 hours. If the charging time is
too short, incorrect interpretation of the rechargeable battery charge status may occur.
Trouble-free functioning of the equipment cannot be assured. Charging the PRIMEDIC™AkuPak is interrupted at temperatures above 45°C.
6.6 Connecting the PRIMEDIC™ PowerLine (mains unit) (optional
accessory)
Procedure:
Insert the PRIMEDIC™ PowerLine following the instructions in chapter 6.3.1.
Then plug the mains plug of the mains cable into a socket in the vicinity of the patient.
The PRIMEDIC™ HeartSave carries out a self-test and is then ready for use.
HeartSave self-tests
Operating instructions
HeartSave AED-M 23702 / GB / A
31 / 74
Frequency
Test coverage
SHORT
Daily
Software, operating membrane, ECG calibration, clock,
internal power supply and HV part at 0 V, impedance
measurement
MEDIUM
First day of the
month
Software, operating membrane, ECG calibration, clock,
internal power supply and HV part at 300 V, impedance
measurement
LONG
On the 1st. July
and on the 1st.
January every
year
Software, operating membrane, ECG calibration, clock,
internal power supply and HV part at 1600 V, impedance
measurement
Note
Under certain circumstances this error will only be present temporarily, or it
may be reversible, and for this reason you should always switch the unit on
again after this message appears and after waiting for approx. 30 seconds, and
then wait for the result of the internal switch-on self-test. If this is successful,
you can continue to use the unit without any problems. If the error remains,
please send the unit to our service department for a more accurate analysis.
7 HeartSave self-tests
7.1 Self-test after switching the HeartSave on
The HeartSave is switched on either by opening the device cover, pressing the On/Off switch
or by inserting the battery when the device cover is removed. Afterwards, the HeartSave runs
a device self-test to check all the important functions and signal mechanisms.
If a fault is detected, the large self-test (LONG) is carried out automatically.
7.2 Automatic, periodic self-tests
The HeartSave carries out automatic self-tests to ensure that it is always ready for operation.
7.3 Tests during equipment operation
The HeartSave monitors the most important equipment and safety functions permanently
during operation. If a fault is detected during one of the many internal self-tests and this fault
no longer ensures safe operation of the unit, the unit will switch off, the spoken message <
Internal Error > will be issued, and the service symbol appears in the status display.
Configuration
32 / 74
HeartSave AED-M 23702 / GB / A
Operating instructions
Parameters
Selection options
Basic
[Active/ ]
Profile1
[Active/ ]
Profile 2
[Active/ ]
Profile 3
[Active/ ]
Page 2:
Microphone:
[On/Off]
BLS information:
[On / Off]
HLW sound
[0 % / 25% / 50% / 75% / 100%]
CPR cycles adult
1-15
CPR cycles child
1-15
CPR Pediatric Mode
[15:2 / 30:2]
Systole sound:
[0 % / 25% / 50% / 75% / 100%]
Volume:
[25% / 50% / 75% / 100%]
Page 3:
Language:
(depends on installed language packages)
Date:
in format DD/MM/YYYY
Time:
00:00 in 24 h format
Mains filter
[50Hz/60Hz/Off]
Display:
[0 degrees / 180 degrees]
8 Configuration
The PRIMEDIC™ HeartSave has been configured in the factory. In the setup menu (displayed
on the monitor) you can change certain parameters. You can save different configurations in a
total of four profiles for different user groups. To activate a profile see Chapter 8.3. The device
always starts in the profile „Basic“, independently of which changes have been made to the
configuration before switching off or removing the power module.
General navigation:
Press the selection / confirmation key during operation to open the setup menu .
Press the button (up) or the button (down) to navigate in the menu and to
increase or respectively decrease a selected parameter
Press the button to select a parameter and to confirm the changed value.
Configuration
Operating instructions
HeartSave AED-M 23702 / GB / A
33 / 74
Parameters
Selection options
Contrast:
from 60 to 180
Page 4:
New PIN
0000-9999
Repeat PIN
0000-9999
Change PIN
[OK, fault]
Save to profile
[OK, fault]
PIN entry
0000-9999
Profile selection
Basic/Profile 1/Profile 2/Profile 3
Page 5:
ARM SW:
x.xx(Version number) xxxx (Check sum 8-digits)
Date (e.g. Jul 11 2005)
DSP SW:
x.xx(Version number) xxxx (Check sum 4-digits)
Date (e.g. Jul 11 2005)
Note
If no key is pressed for one minute, the device automatically leaves the Setup
Menu and goes back to standby.
8.1 Simple change of configuration – example: Time
To change the time, proceed as follows:
Use the key to change into the setup menu
Navigate the cursor with the button to the menu point "to Page 2"
Actuate the button to go to Page 2 of the menu.
Switch to Page 3 with the button
Navigate to "Time of day" with the button and confirm the input with the button
Set the hours and minutes using the buttons and .
Confirm the input with the button
Move the cursor by pressing the key to the menu item Time. Select the highlighted
menu item Time by pressing the key. The hour is then highlighted.
Configuration
34 / 74
HeartSave AED-M 23702 / GB / A
Operating instructions
Note
When the device is first delivered, the PIN is always set in the factory to 0000.
Note
Please note that your profile selection is only active until the device is switched
off. The devices always starts with the profile „Basic“
8.2 Changing the PIN
The PIN is used to save profiles. Entering a PIN is absolutely necessary. If you want to change
the PIN you will always need to know the old PIN.
proceed as follows:
Use the key to change into the setup menu
Switch to Page 5 of the menu.
Navigate with the key to the entry < PIN > and confirm your selection with the key
Enter the current PIN as follows: Using the keys you can increase or decrease a
digit. With the key you can change to the next digit. After the fourth digit, it jumps
back to the menu item
Navigate to the entry < New PIN > and enter your new PIN as described above.
Navigate to the entry < Repeat PIN > and enter your PIN again.
Select the menu parameter < Change PIN > and confirm your new PIN with the key
.
On the right next to the cursor the entry < OK > should appear. This means your new
PIN is active.
8.3 Calling up/activating a profile
Certain parameter settings for the menu can be summarised into profiles.
Saved profiles can be called up as follows:
Use the key to change into the setup menu
Select your required profile using the keys and confirm it with the key
The selected profile is active
8.4 Saving menu parameters in a profile
Certain parameter settings for the menu can be saved as profiles. The profiles Basic, Profile 1,
Profile 2, and Profile 3 are available.
If you want to save parameters in a profile or want to change a profile, proceed as follows:
Use the key to change into the setup menu
Change the required parameters from the various pages of the menu to suit your
needs.
Change to page 4 of the menu.
Configuration
Operating instructions
HeartSave AED-M 23702 / GB / A
35 / 74
Note
The parameter „Network filter“ can only be changed temporarily while operating
the device. After starting the device, the network filter is always switched off
initially.
Setting
CPR cycles child
CPR pediatric mode
(1)
5 cycles
30:2
(2)
7 cycles
15:2
Note
In the standard delivered condition, the ratio between chest compression and
rescue breathing is 30:2.
Use the key to navigate to the entry < Profile selection > and confirm your selection
with the key
Use the keys to select the required profile which is to be used to save the menu
parameters previously selected. Confirm this with the key
Navigate with the key to the entry < PIN > and confirm your selection with the key
Enter the current PIN as follows: Using the keys you can increase or decrease a
digit. With the key you can change to the next digit. After the fourth digit, it jumps
back to the menu
Change to the menu entry < Save to profile > and confirm your selection with the key
On the right next to the cursor the entry < OK > should appear. This means the profile
is saved.
Now leave the menu by using the button to navigate to the menu item < End > and
confirm this with the button
If you want to change the configuration that your device starts up with when it is switched on,
you have to save your changed menu parameters in the "Basic" profile.
8.5 Ratio chest compression:ventilation in Pediatric Mode
The ratio between chest compressions and rescue breathing can be changed in the Pediatric
mode as follows.
(1) Give 30 chest compressions : Give 2 rescue breaths
(2) Give 15 chest compressions : Give 2 rescue breaths
In order to set the desired ratio between the chest compressions:ventilation, you need to
adjust the following parameters as shown in the table:
Operating the HeartSave and sequence of resuscitation
36 / 74
HeartSave AED-M 23702 / GB / A
Operating instructions
Note
The sequence of the reanimation is realised in the device according to the
recommended guidelines of the European Resuscitation Council (ERC
Guidelines 2015). We recommend the user has carried out before using
HeartSave.
9 Operating the HeartSave and sequence of
resuscitation
9.1 Switching the HeartSave on
HeartSave is automatically activated by removing its cover on the device. If the device is not
switched on automatically, switch it on by pressing the On/Off button. After this, all buttons are
unlocked, except the shock button. Triggering the defibrillation is only released after detecting
the ventricular fibrillation.
Directly after switching it on, an internal self-test is carried out to check important functions and
signal devices. Standby is confirmed by a beep. It is important to ensure that the loudspeaker
is working.
9.2 Examining and preparing the patient
Check to see whether the patient is unconscious and is not breathing normally. Proceed as
follows:
Please talk to patient and touch him/her to see whether he or she is still conscious.
Make sure the emergency services have been notified.
If there is no response, hyperextend the patient's head and check that he/she is
breathing.
If the patient is breathing normally, bring him/her into a stable position on his/her
side and continue to treat him/her.
If the patient is not breathing normally, expose his/her breast area to attach the
defibrillations electrodes. Make sure the patient is lying on a hard surface in order to
ensure the chest compressions are effective. If HeartSave is not already available,
make sure someone collects it in order to carry out further treatment.
Using the supplied razor, remove breast hair from the area where the defibrillation
electrodes are to be attached.
If the surface of the skin is damp, dry the skin at the spots where the defibrillation
electrodes are to be attached to improve the adhesion.
Operating the HeartSave and sequence of
resuscitation
Operating instructions
HeartSave AED-M 23702 / GB / A
37 / 74
DANGER
Danger of damage to health of user or a third party
Triggering heart arrhythmia and burns caused by electrocution
Do not touch the patient during defibrillation
Warn third parties about the dangers of defibrillation
Do not touch any conductive materials (metal, blood, water, other liquids, etc.)
during defibrillation
DANGER
Warning of explosion
Risk of burns
Do not use the device in potentially explosive areas
Do not use the device in oxygen-enriched atmospheres
Do not use the device in the presence of flammable materials
DANGER
Warning: potential malfunctions
Active implants may lead to a false diagnosis
Do not stick the defibrillation electrodes directly over a pacemaker or similar.
WARNING
Warning: physical harm
Risk of skin burns
Remove heavy build-up of hair at the electrode positions
Where necessary, dry the skin before attaching the electrodes
ATTENTION
Material damage to other devices
Remove all devices at risk from the defibrillation from patients before defibrillation.
Do not attach the defibrillation electrodes directly over a pacemaker or similar.
9.3 Defibrillation
Operating the HeartSave and sequence of resuscitation
38 / 74
HeartSave AED-M 23702 / GB / A
Operating instructions
Note
The defibrillator automatically starts in Adult Mode.
Note
Follow the voice commands issued by HeartSave!
Note
For this, you will need to take the disposable gloves out of the cover of the
device and put them on.
Defibrillation with HeartSave can be performed on children or adults. Use the Pediatric Mode
for patients who are younger than 8 years or weigh less than 25 kg. Use the Adult Mode for
patients who are older than 8 years or weigh more than 25 kg.
The therapy should not be delayed in order to determine the precise age or weight of the
patient.
9.3.1 Defibrillation in Adult Mode
After the self-test has been successfully finished by the device the following BLS voice
instructions (BLS= the basic measures of cardio pulmonary resuscitation) are given.
<Adult mode>
< Call emergency services >
< Apply electrodes one after the other to patient's bare chest >
< Plug in electrode cable >
The last two spoken instructions are repeated for a period of one minute. If the device cannot
recognise a patient impedance at that time, the device will give instructions for one cycle of
cardio pulmonary resuscitation:
< Give 30 chest compressions >
< Give 2 rescue breaths >
Afterwards the device will give instructions to attach the electrodes for maximum one minute.
This procedure will go on until the device recognises a valid patient impedance and begins
with the analysing rhythm.
Operating the HeartSave and sequence of
resuscitation
Operating instructions
HeartSave AED-M 23702 / GB / A
39 / 74
Fig. 13: Position of electrodes on adults
Note
In order to use adult electrodes (SavePads PreConnect or SavePads C) in
Pediatric Mode, proceed as follows:
Open the cover on the unit / switch HeartSave on
Insert the electrode plug into the socket on HeartSave
Push the pediatric button
Stick the electrodes onto the naked upper torso
Follow the voice prompts issued by HeartSave
The positions of the electrodes are:
On the right chest area, below the collar bone (1) and
On the left side of the chest, above the apex of the heart on the axillary line (2).
9.3.2 Defibrillation in Pediatric Mode
If the patient is younger than 8 years old or weighs less than 25 kg, please use the SavePads
mini. When these electrodes are inserted, HeartSave will automatically switch to Pediatric
Mode. If you do not have any SavePads mini available, you can switch to pediatric mode by
pressing the pediatric button with the SavePads PreConnect.If HeartSave is in Pediatric Mode,
the indicator LED next to the pediatric button will illuminate.
The Pediatric Mode has been especially developed for the needs of children. In Pediatric
Mode, the HeartSave supplies less energy than in Adult Mode.
<Pediatric mode>
< Call emergency services >
< Apply electrodes one after the other to patient's bare chest >
The last two spoken instructions are repeated for a period of one minute. If the device cannot
recognise a patient impedance at that time, the device will give instructions for one cycle of
cardio pulmonary resuscitation:
< Give 30 chest compressions >
< Give 2 rescue breaths >
Operating the HeartSave and sequence of resuscitation
40 / 74
HeartSave AED-M 23702 / GB / A
Operating instructions
Fig. 14: Position of electrodes on a child
Fig. 15: alternative position of electrodes on a child
Afterwards the device will give instructions to attach the electrodes for maximum one minute.
This procedure will go on until the device recognises a valid patient impedance and begins
with the analysing rhythm.
The positions of the electrodes are:
On the right chest area, below the collar bone (1) and
On the left side of the chest, above the apex of the heart on the axillary line (2).
The positions of the electrodes are:
(1) in the middle of the chest
(2) on the back at heart level
Attach both electrodes so that the patient's heart lies in between them.
Operating the HeartSave and sequence of
resuscitation
Operating instructions
HeartSave AED-M 23702 / GB / A
41 / 74
WARNING
Damage to gel layer on defibrillation electrodes
Skin burns
Be careful not to touch the gel layer before attaching the electrodes
Fig. 16: Removing the film from the electrodes
(1) Protective film on electrodes
(2) SavePads defibrillation electrodes
9.4 Opening SavePads and placing electrodes
HeartSave will give a voice prompt for you to apply the defibrillation electrodes to the patient.
< Apply electrodes one after the other to patient's bare chest >
Procedure:
Open the defibrillation electrodes bag by tearing open the protective cover along the
tear strip.
Remove the protective film (1) from one of the electrodes (2) and then immediately
place the electrode on the position you had ascertained previously.
Proceed to remove the protective film from the second electrode and place it in its
position.
Smooth the electrodes onto the patient ensuring there are no air bubbles under the
electrodes!
Operating the HeartSave and sequence of resuscitation
42 / 74
HeartSave AED-M 23702 / GB / A
Operating instructions
Note
If you have already inserted the SavePads, HeartSave will skip this step and
begin analysing the patient's cardiac rhythm
Fig. 17: Plug in electrode cable
(1) Socket
(2) Connector symbol
(3) Electrode plug
Note
Once the electrodes are connected to the patient and the electrode plug is
plugged in, the BLS instructions are automatically interrupted.
9.5 Plug in electrode cable
Procedure:
After hearing the voice prompt
<Plug in electrode cable> ,
insert the connector (3) of the electrode cable into the socket (1) on HeartSave as
shown above.
Make sure the red dot points to the front.
The red "plug symbol LED" (2) on the device should go out.
Operating the HeartSave and sequence of
resuscitation
Operating instructions
HeartSave AED-M 23702 / GB / A
43 / 74
9.6 Check electrodes
If the device reports the fault < Check electrodes >, this can be for several reasons:
Electrode plug connector not plugged in. This will be signalled by flashing LEDs in
the electrode plug connector symbol and on the electrodes positions on the front
film.
Too little resistance between electrodes (e.g. electrodes fixed too close to each
other). The LEDs blink on the electrodes positions on the front foil.
Too high resistance between the electrodes (e.g. chest hair not removed from the
patient). The LEDs blink on the electrodes positions on the front foil.
Air pockets between skin and defibrillation electrodes create a bad contact. The
LEDs blink on the electrodes positions on the front foil.
Dried out electrodes. The LEDs blink on the electrodes positions on the front foil.
The device repeats the following voice messages:
< Check electrodes >
< Apply electrodes one after the other to patient's bare chest >
If the plug on the PRIMEDIC™ SavePads has not been inserted in the unit, the following
additional instruction appears
< Plug in electrode cable >
These spoken instructions are repeated for a period of one minute. If the device cannot
recognise a patient impedance at that time, the device will give instructions for five cycles of
cardio pulmonary resuscitation:
< Give 30 chest compressions >
< Give 2 rescue breaths >
Afterwards, the device will once again give instructions to attach the electrodes for a maximum
period of one minute. This procedure will go on until the device recognises a valid patient
impedance and begin with the analysing rhythm.
It is very important to remedy the cause of the fault!
Operating the HeartSave and sequence of resuscitation
44 / 74
HeartSave AED-M 23702 / GB / A
Operating instructions
DANGER
Danger of damage to health of user, patient or a third party
Triggering heart arrhythmia
Do not touch the patient during defibrillation
Warn third parties about the dangers of defibrillation
Do not touch any conductive materials (metal, blood, water, other liquids, etc.)
during defibrillation
If the patient wakes up during reanimation, stop defibrillation
Note
If the ECG analysis is conducted in a vehicle, the engine will need to be
switched off for analysis in order not to distort the results.
Note
Pressing the shock key during power charging (before it turns green) does not
result in release of shock, it rather leads to internal safety discharge.
Note
Defibrillation may cause muscle contractions in the patient.
9.7 Carrying out the ECG analysis
If the defibrillation electrodes have been applied, the device will automatically start the
analysis.
The patient must now be put in an immobile position and must no longer be touched. The
device prompts:
< Do not touch the patient, analysing rhythm >
and the "do not touch patient" zone flashes on the membrane keyboard.
The algorithm of the device programme will now check the ECG for ventricular fibrillation. This
process takes approx. 7 - 12 seconds. If the device identifies a ventricular fibrillation, it will
recommend defibrillation.
9.8 Defibrillation required
If the device clearly identifies VF, then it will recommend defibrillation which is automatically
prepared inside the device.
The device prompts:
< Shock advised >
< Device is charging >
< Chest compressions >
< Metronome >
Operating the HeartSave and sequence of
resuscitation
Operating instructions
HeartSave AED-M 23702 / GB / A
45 / 74
Note
Once the CPR time has expired (2 mins.), the device returns to ECG analysis.
To reduce the time without chest compressions, the metronome is activated during the
charging phase. This time may vary depending on the charge level of the battery.
Carry out the chest compressions until the metronome tone stops.
Once the capacitor is charged internally, power for the defibrillation pulse is available for 15
seconds. This is signalled by a continuous acoustic warning, the voice message
< Stand clear of patient >
< Press lit shock button now >,
a continuous tone and the shock button lighting up in "green".
Warn those around you loudly before applying the defibrillation!
Actuate the green illuminated shock button to apply the shock.
If you do not defibrillate within 15 secs, an internal safety discharge will follow and the ECG will
be analysed again.
Defibrillation and cardio-pulmonary resuscitation (CPR) are repeated according to the
directives of the ERC Guidelines 2015.
The charge time of the capacitor for defibrillation depends on the available battery capacity.
Charging may take longer if the power module is partly discharged.
If an error should occur during charging, an intermittent warning beep will sound.
9.9 Defibrillation not required
If the device cannot find a shockable rhythm, then it recommends cardio-pulmonary
resuscitation (CPR).
< No shock advised >
< Cardio pulmonary resuscitation >
< Give 30 chest compressions >
< Give 2 rescue breaths >
Furthermore, during the chest compressions, you will be supported by the installed metronome
function which will give you the correct frequency for the chest compressions (100
compressions/min). Be sure that you keep to the given rhythm. The artificial respiration will
also be supported by two acoustic outputs. From the second to fifth CPR cycle, only these
sound signals are emitted. For your support, the correct measures for cardio pulmonary
resuscitation are illustrated in pictograms on the utensils carrier.
Carry out cardio pulmonary resuscitation until the emergency services arrive. Once the patient
is conscious again, lay him or her down and monitor him/her until the emergency services
arrive.
Operating the HeartSave and sequence of resuscitation
46 / 74
HeartSave AED-M 23702 / GB / A
Operating instructions
Note
If, when the device is switched on, no ECG is recorded for 10 minutes or no
button is pressed, the device automatically switches off. Approx. 30 seconds
before the switch-off this is signalized by an interrupted warning tone. Pressing
any button or any other activity will interrupt the switching off process.
9.10 Switching HeartSave off
HeartSave can be switched off in various ways:
By pressing the on/off button for approx. 3 seconds. A warning beep will sound
simultaneously. This time has been chosen to avoid it being switched off
accidentally.
By closing the cover of the device.
If the device does not recognise a signal for 10 minutes and if no button is
pressed, it will switch off automatically.
If HeartSave detects a fault, it will switch off automatically to prevent any injuries.
9.11 Keeping the defibrillator ready for use
Check HeartSave for damage after each use.
Clean HeartSave and accessories after each use. Disinfect HeartSave and
accessories if there is a risk of infection, see section 10.1.
Renew the SavePads and replace or charge the energy module, if necessary, so
that the HeartSave is ready for use as quickly as possible.
If any malfunctions or noticeable problems occur, please contact your nearest
service station.
9.12 Monitoring the patient
After successful defibrillation, the patient can be monitored by the monitoring function using
the SavePads electrodes already used whilst they are being transferred to hospital. The Lead
II (Einthoven) is available. If in this situation, ventricular fibrillation is detected again, renewed
resuscitation can be carried out very quickly. For this purpose, the vehicle must be stopped
and the engine switched off to ensure the analysis is correct.
If you need to monitor the ECG of a patient in other situations, please use the 2-core ECG
patient cable.
This cable is only used for ECG monitoring of a patient. This cable cannot be used for
defibrillation. If the automatic background analysis carried out by the HeartSave detects a
heart rhythm that is in need of defibrillation, the following verbal message is issued:
< Analysis Recommended. Use SavePads >
In order to be able to carry out defibrillation, the ECG cable must be removed and replaced by
the SavePads.
Cleaning, maintenance and despatch
Operating instructions
HeartSave AED-M 23702 / GB / A
47 / 74
WARNING
Warning: physical harm to user
Risk of electrocution
Only clean the device when switched off
Do not immerse the device in liquids
Use damp cloths to clean
ATTENTION
Warning: property damage
The device does not have any parts which can be modified by the user
Do not carry out any repairs to the device
Do not carry out any modifications to the device
Do not dismantle HeartSave
Only use genuine accessories!
10 Cleaning, maintenance and despatch
10.1 Cleaning
Clean HeartSave and all its accessories, such as the wall bracket, with commercially available
household cleaners.
Use a slightly damp, clean cloth. Use ordinary surface disinfectants to disinfect (e.g. Gigasept
FF, Bacillol or Spitacid).
10.2 Servicing
Regardless of how often HeartSave is used, we recommend you carry out a regular visual
check on HeartSave and accessories at least once per week.
Make sure that the housing, cable, SavePads and all the other accessories are undamaged.
Cleaning, maintenance and despatch
48 / 74
HeartSave AED-M 23702 / GB / A
Operating instructions
DANGER
Risk of fire due to short circuit
Before sending, protect the contacts with insulating adhesive tape.
10.2.1 Servicing check list
Check the expiration date
of the SavePads and
the battery / AkuPak LITE
if necessary replace the parts.
Check whether
the status display "OK" is shown!
the device is fully equipped.
Contact service if there is a fault on the device.
10.3 Sending the HeartSave
Use the original box where possible.
If the original box is no longer available, use suitable packaging materials to protect HeartSave
from impact and damage.
Pay attention to national and international shipping regulations concerning the transport of
Lithium batteries.
Disposal
Operating instructions
HeartSave AED-M 23702 / GB / A
49 / 74
CAUTION
Warning: physical harm
Risk of acid burns
Dispose of the device and its individual parts according to local regulations
Fig. 18: Disposal
11 Disposal
In accordance with the founding principles of the company Metrax GmbH, your product has
been developed and made using high quality materials and components which are recyclable.
At the end of its service life, recycle the device via disposal companies registered under public
law (council recycling facilities). Proper disposal of this product helps with environmental
protection.
Through the registration of Metrax GmbH with the responsible authorities, we ensure that the
disposal and utilisation of electronics devices brought to the market by us is secure in
accordance with the EU directive on the disposal of electronic and electrical equipment
(WEEE-directive).
In Germany, in accordance with the law on bringing electrical and electronic equipment onto
the market, taking it back and disposing of in an environmentally friendly manner
(Electrical and Electronic Equipment Act– ElektroG) Metrax is registered with EAR (register of
old electronic equipment) under the number: 73450404 .
For business customers in the European Union
Please contact your dealer or supplier if you want to dispose of electrical and electronic
equipment. He will have further information on this for you.
Information for disposal in countries outside the European Union
This symbol is only applicable within the European Union.
List of error codes
50 / 74
HeartSave AED-M 23702 / GB / A
Operating instructions
Error
code
Error text
Cause(s)
User action
50
HV, Internal FSM error
Error during the shock
appliance
- return the device for servicing
51
HV, CPLD error
Error during the shock
appliance
- return the device for servicing
52
HV, +5V voltage failed
Error in the voltage supply,
possibly due to an empty
accumulator/battery
- charge and/or replace the
energy module
- if the error persists: Return
the device for servicing
53
HV, +5VSW voltage failed
Error in the voltage supply,
possibly due to an empty
accumulator/battery
- charge and/or replace the
energy module
- if the error persists: Return
the device for servicing
54
HV, +24V voltage failed
Error in the voltage supply,
possibly due to an empty
accumulator/battery
- charge and/or replace the
energy module
- if the error persists: Return
the device for servicing
55
HV, VREF voltage failed
Error in voltage reference
- return the device for servicing
56
HV, HVPWR voltage failed
Error in the voltage supply,
possibly due to an empty
accumulator/battery
- charge and/or replace the
energy module
- if the error persists: Return
the device for servicing
57
HV, RLPWR voltage failed
Error in the voltage supply,
possibly due to an empty
accumulator/battery
- charge and/or replace the
energy module
- if the error persists: Return
the device for servicing
58
HV, Current setup DAC failed
Error when setting the current
specifications - defect in the
high voltage section
- return the device for servicing
59
HV, RTCCLK error was
detected
Error in generation of system
clock
- restart the device
- if the error persists: Return
the device for servicing
60
HV, MCLK error was detected
Error in system clock
generation
- restart the device
- if the error persists: Return
the device for servicing
61
HV, HVHALT event was
detected
Emergency shutdown during
the shock application
- return the device for servicing
62
HV, HVHALT net error was
detected
Error during the test of the
emergency shutdown
- return the device for servicing
12 List of error codes
The following table lists all messages that the unit emits when it switches off because of a
severe error. This switch-off is carried out with the spoken message "Internal Error"; the cause
of the error is displayed with the error code number from the first column and the text from the
second column. The other columns list the possible causes of the error and the remedial
measures that can be carried out by the user themselves.
List of error codes
Operating instructions
HeartSave AED-M 23702 / GB / A
51 / 74
Error
code
Error text
Cause(s)
User action
63
HV, Disarm cannot be
performed
Error during the internal
discharge, not possible to
discharge the condenser
- return the device for servicing
64
HV, HV capacitor overvoltage
was detected
Error during the high voltage
load, voltage at the condenser
too high
- return the device for servicing
65
HV, HV charger error was
detected
Error during the high voltage
load, impossible to load the
high voltage
- return the device for servicing
66
HV, Error of HV measurement
circuit
Error during measurements in
high voltage section
- return the device for servicing
67
HV, HV charge time is too big
Error during the high voltage
load, impossible to load the
high voltage
- return the device for servicing
68
HV, Too big current was
detected
Current too high during shock
delivery
- return the device for servicing
69
HV, Too low current at shock
was detected
Current too low during shock
delivery, possibly due to fault in
final stage
- return the device for servicing
70
HV, Current sensor error
Error in the current regulation
- return the device for servicing
71
HV, SHKEN net error
Error during the safety test
before the shock delivery
- return the device for servicing
72
HV, START net error
Error during shock release
- restart the device
- if the error persists: Return
the device for servicing
75
HV, HV hardware protection
error
Error during shock release
- restart the device
- if the error persists: Return
the device for servicing
76
HV, Residual voltage after
disarm error
Residual current too high at the
condensor after the discharge
- restart the device
- if the error persists: Return
the device for servicing
79
HV, Operator warning error
Error during the shock
preparation, defective output
for the user
- return the device for servicing
80
HV, Shock mode error
Defective switching between
synchronous and
asynchronous shock modality
- restart the device
- if the error persists: Return
the device for servicing
84
PRWIN fail > 100 ms
Error in the voltage supply,
possibly due to a flat
accumulator/battery
- charge and/or replace the
energy module
- if the error persists: Return
the device for servicing
87
Program update error
Error during program update,
the update will be interrupted
- return the device for servicing
List of error codes
52 / 74
HeartSave AED-M 23702 / GB / A
Operating instructions
90
Tasks start/run error
Error during internal processing
- restart the device
- if the error persists: Return
the device for servicing
91
Menu init error
Defective display in the Setup
menu
- return the device for servicing
92
Shock key error
Error during the check of
critical input elements (e.g.
shock button)
- restart the device
- if the error persists: Return
the device for servicing
93
Charge time-out error
The high voltage charge lasts
too long, possibly due to a fault
in the final stage
- return the device for servicing
94
Disarm time-out error
Impossible to discharge the
residual energy in the
condensor, possibly due to a
fault in the final stage
- return the device for servicing
96
AVDD exceed
Error in the voltage supply,
possibly due to an empty
accumulator/battery
- charge and/or replace the
energy module
- if the error persists: Return
the device for servicing
97
AVDD drop
Error in the voltage supply,
possibly due to an empty
accumulator/battery
- charge and/or replace the
energy module
- if the error persists: Return
the device for servicing
98
CVDD exceed
Error in the voltage supply,
possibly due to an empty
accumulator/battery
- charge and/or replace the
energy module
- if the error persists: Return
the device for servicing
99
CVDD drop
Error in the voltage supply,
possibly due to a flat
accumulator/battery
- charge and/or replace the
energy module
- if the error persists: Return
the device for servicing
100
DVDD exceed
Error in the voltage supply,
possibly due to an empty
accumulator/battery
- charge and/or replace the
energy module
- if the error persists: Return
the device for servicing
101
DVDD low
Error in the voltage supply,
possibly due to an empty
accumulator/battery
- charge and/or replace the
energy module
- if the error persists: Return
the device for servicing
102
5V exceed
Error in the voltage supply,
possibly due to an empty
accumulator/battery
- charge and/or replace the
energy module
- if the error persists: Return
the device for servicing
List of error codes
Operating instructions
HeartSave AED-M 23702 / GB / A
53 / 74
103
5V drop
Error in the voltage supply,
possibly due to an empty
accumulator/battery
- charge and/or replace the
energy module
- if the error persists: Return
the device for servicing
104
24V exceed
Error in the voltage supply,
possibly due to an empty
accumulator/battery
- charge and/or replace the
energy module
- if the error persists: Return
the device for servicing
105
24V drop
Error in the voltage supply,
possibly due to an empty
accumulator/battery
- charge and/or replace the
energy module
- if the error persists: Return
the device for servicing
106
PWRIN exceed
Error in the voltage supply,
possibly due to an empty
accumulator/battery
- charge and/or replace the
energy module
- if the error persists: Return
the device for servicing
107
PWRIN drop
Error in the voltage supply,
possibly due to an empty
accumulator/battery
- charge and/or replace the
energy module
- if the error persists: Return
the device for servicing
108
Battery exceed
Internal battery voltage too high
- restart the device
- if the error persists: Return
the device for servicing
110
Temperature exceed
Operating temperature too high
- operate the device in a cooler
environment
111
Temperature drop
Operating temperature too low
- operate the device in a
warmer environment
116
HDQ interface fail
Error communicating with
energy module
- restart the device
- if the error persists: Return
the device for servicing
117
CYCLE time error
Error during the shock delivery
- return the device for servicing
118
VREF exceed
Error in the voltage supply,
possibly due to a flat
accumulator/battery
- charge and/or replace the
energy module
- if the error persists: Return
the device for servicing
119
VREF drop
Error in the voltage supply,
possibly due to an empty
accumulator/battery
- charge and/or replace the
energy module
- if the error persists: Return
the device for servicing
120
Battery/accumulator fail
Energy module voltage too low
- charge and/or replace the
energy module
121
DSP timeout error
Internal communication error
- restart the device
- if the error persists: Return
the device for servicing
122
Previous error detected need full test
General description of error
detected during self test
- restart the device
- if the error persists: Return
the device for servicing
List of error codes
54 / 74
HeartSave AED-M 23702 / GB / A
Operating instructions
123
Programme code area corrupt
Error in the programme
memory
- return the device for servicing
124
Keyboard error
Error in the membrane
keyboard
- restart the device
- if the error persists: Send the
unit in to the Service
Department
- do not push any key during
equipment test!
125
ECG calibration error
Error during the ECG
calibration
- restart the device
- if the error persists: Return
the device for servicing
127
Clocks (32 kHz or
3.6864MHz) error
Error in system clock
generation
- restart the device
- if the error persists: Return
the device for servicing
129
Wrong hardware platform
Error during the hardware
detection
- return the device for servicing
138
Protected variable error
Defective storage of important
variables
- return the device for servicing
140
Impedance measurement
error
Error during the impedance
measurement
- return the device for servicing
150
Aura LED Test error
Error during the test of the LED
aura
- return the device for servicing
151
Electrode coding error
Defective electrode type
- use different electrodes
and/or different electrode
cables
- check the electrode plug
housing
Technical Data
Operating instructions
HeartSave AED-M 23702 / GB / A
55 / 74
Defibrillation
Operating modes:
Asynchronous, external
Patient impedance:
23 – 200 Ohm
Impulse shape:
Biphasic, current regulated (CCD)
Output energy in Adult
mode:
Patient impedance
1st stage
2nd. stage
3rd stage
25 Ohm
165 J
254 J
310 J
50 Ohm
298 J
348 J
360 J
75 Ohm
336 J
346 J
346 J
100 Ohm
320 J
320 J
320 J
125 Ohm
296 J
296 J
296 J
150 Ohm
274 J
274 J
274 J
175 Ohm
236 J
236 J
237 J
Output energy in pediatric
mode:
Patient impedance
1st stage
2nd. stage
3rd stage
25 Ohm
37 J
53 J
70 J
50 Ohm
48 J
68 J
87 J
75 Ohm
48 J
66 J
84 J
100 Ohm
45 J
62 J
79 J
125 Ohm
41 J
57 J
73 J
150 Ohm
38 J
53 J
68 J
175 Ohm
35 J
49 J
63 J
Accuracy:
All data is subject to a tolerance of +/- 15%
Impulse length:
Positive phase 11.25 ms, negative phase 3.75 ms
ECG
Line:
Einthoven II
Heart frequency:
30 – 300 min-1 (Accuracy +/- 1/min, 1%)
Input:
Class BF, for 2-pin patient cable, defibrillation protected
Input resistance:
> 5 MOhm @ 10 Hz
CMRR:
> 85 dB
Input DC voltage:
± 0.5 V
13 Technical Data
Technical Data
56 / 74
HeartSave AED-M 23702 / GB / A
Operating instructions
Bandwidth:
0.5 – 40 Hz (- 3 dB) SR = 101 samples/s
Impedance measurement
Defibrillation:
23 ... 200 Ohm (accuracy +/- 20%)
Measurement frequency:
30 kHz
Analysis
Analysis recognition:
Ventricular fibrillation (VF)
Analysis duration:
Approx. 7 s until VF is recognized
Period from start of the
analysis until the end of
high voltage charge (with
fully charged battery /
after six shocks / after 15
shocks)
27 s / 27 s / 27 s
Period from switching on
until the end of high
voltage load (with fully
charged battery / after six
shocks / after 15 shocks)
40 s / 40 s / 40 s
Monitor
Type:
High-resolution LCD monitor, 95 x 72 mm (diagonals 120 mm, 4.7
inch)
Resolution:
320 x 240 pixels (pixel size: 0.36 x 0.36 mm)
Displays:
Heat frequency, number of defibrillations, number of VFs
identified, resuscitation time, date, time, power capacity, ECG
graph
Presentation:
X 25 mm/sec, Y 10 mm/mV
Data storage
Memory type:
CompactFlashCard 2 GB
Technical Data
Operating instructions
HeartSave AED-M 23702 / GB / A
57 / 74
Safety
Classification:
Medical product in class IIb, Device with internal power supply,
Type BF, Defi-resistant,
Identification:
The device is a medical product and complies with EC Directive 93
/ 42 / EEC
Other
Operating conditions:
0 ... 55 °C, 30 ... 95 % rel. humidity, but without condensation
700 hPa ... 1060 hPa continuous mode
Storage conditions:
- 20 ... 70 °C, 20 ... 95 % rel. humidity, but without condensation
500 hPa ... 1060 hPa
Dimensions:
28 x 25 x 9 cm (W x H x D)
Weight:
approx. 2.0 Kg (without energy module)
approx. 2.5 kg (with energy module)
Standards applied
Standards (for licensing in the EU, the corresponding harmonised
European standards EN were used instead of the IEC standards):
The warranty period is 24 months and starts on the day of purchase. Please keep the invoice
as proof of purchase.
Within this time period, METRAX GmbH will remedy any defects in the device free of charge if
they are based on material or manufacturing faults. The device can be reinstated to its original
condition as selected by METRAX GmbH either by repair or replacement.
A claim made under warranty does not extend the original warranty period.
Guarantee and also legally entitled warranty claims are not applicable if the usefulness of the
device is only negligibly affected, or in the case of normal wear and tear (e.g. consumables
such as AkuPak) or damage caused after transfer of risk as a result of incorrect or negligent
handling, excessive wear or are caused by special external influences which are not provided
for according to the contract. The same applies if inappropriate changes or incorrect repair
work is carried out by the buyer or by a third party.
All other claims on METRAX GmbH are excluded, unless such claims are based on intent or
gross negligence or compulsory legal liability standards.
Warranty claims made by the buyer against the seller (dealer) are not affected by this
guarantee.
In the case of a warranty claim, please return the device, with proof of purchase (e.g. invoice)
stating your name and address, to your dealer or to METRAX GmbH.
METRAX After-Sales Service is glad to be at your disposal, even after the warranty period has
expired.
Depiction of the current time functions
Operating instructions
HeartSave AED-M 23702 / GB / A
59 / 74
First shock
Second shock
Third shock
25 Ohm
Current [A]
Time [ms]
Time [ms]
Time [ms]
50 Ohm
Current [A]
Time [ms]
Time [ms]
Time [ms]
75 Ohm
Current [A]
Time [ms]
Time [ms]
Time [ms]
100 Ohm
Current [A]
Time [ms]
Time [ms]
Time [ms]
15 Depiction of the current time functions
The following diagrams show the graphs for the defibrillation pulse displayed depending on the
load resistance.
15.1 Adult Mode
Depiction of the current time functions
60 / 74
HeartSave AED-M 23702 / GB / A
Operating instructions
First shock
Second shock
Third shock
125 Ohm
Current [A]
Time [ms]
Time [ms]
Time [ms]
150 Ohm
Current [A]
Time [ms]
Time [ms]
Time [ms]
175 Ohm
Current [A]
Time [ms]
Time [ms]
Time [ms]
195 Ohm
Current [A]
Time [ms]
Time [ms]
Time [ms]
Depiction of the current time functions
Operating instructions
HeartSave AED-M 23702 / GB / A
61 / 74
First shock
Second shock
Third shock
25 Ohm
Y=2A/Div; X=2ms/Div
Y=5A/Div; X=2ms/Div
Y=5A/Div; X=2ms/Div
50 Ohm
Y=5A/Div; X=2ms/Div
Y=5A/Div; X=2ms/Div
Y=5A/Div; X=2ms/Div
75 Ohm
Y=5A/Div; X=2ms/Div
Y=5A/Div; X=2ms/Div
Y=5A/Div; X=2ms/Div
100 Ohm
Y=5A/Div; X=2ms/Div
Y=5A/Div; X=2ms/Div
Y=5A/Div; X=2ms/Div
15.2 Pediatric mode
Depiction of the current time functions
62 / 74
HeartSave AED-M 23702 / GB / A
Operating instructions
First shock
Second shock
Third shock
125 Ohm
Y=5A/Div; X=2ms/Div
Y=5A/Div; X=2ms/Div
Y=5A/Div; X=2ms/Div
150 Ohm
Y=2A/Div; X=2ms/Div
Y=2A/Div; X=2ms/Div
Y=2A/Div; X=2ms/Div
175 Ohm
Y=2A/Div; X=2ms/Div
Y=2A/Div; X=2ms/Div
Y=2A/Div; X=2ms/Div
Rhythm detection system
Operating instructions
HeartSave AED-M 23702 / GB / A
63 / 74
16 Rhythm detection system
The rhythm detection system on the HeartSave analyses the patient's ECG and supports it if
the unit detects a shockable or non-shockable rhythm.
Operator control of the defibrillation shock therapy:
Ascertaining the electrode contact
Automatic evaluation of the ECG
Operator control of the defibrillation shock therapy
The transthoracic impedance of the patient is measured by the defibrillation electrodes. If the
baseline impedance is greater than the maximum critical value, then the device determines
whether the electrodes are not in good enough contact with the patient or if they are not
connected properly to the device. ECG analysis and dispensation of defibrillation shocks are
therefore prevented. The voice output says "Check electrodes" if the contact of the electrodes
is insufficient.
Automatic Interpretation of the ECG
The rhythm detection system of the device is designed to recommend a defibrillation shock
when the system is has been connected up to a patient and the system detects a rhythm
which requires defibrillation
With all other ECG rhythms, including asystolia and normal sinus rhythms, the HeartSave
rhythm detection system does not recommend defibrillation.
Operator control of the output of defibrillation shocks
The device's rhythm detection system triggers an automatic power charge if the device detects
a cardiac rhythm which requires defibrillation. Optical and acoustic messages are emitted to
show you that the device recommends giving a defibrillation shock. If a defibrillation shock is
recommended, you decide whether and when the shock is to be given.
The Algorithm:
Observes the ECG rhythm over a continuous recording history of 10 seconds, of
which 7 seconds have been used for an initial diagnosis or to display the message
"Shock advised".
Measures symmetry and energy content of the signal
Filters and measures artefacts and interference
Detects pacemakers
Measures the QRS rate
Rhythm detection system
64 / 74
HeartSave AED-M 23702 / GB / A
Operating instructions
16.1 Adult mode
For validation the following databases have been used: AHA and MIT
Performance results (weighted average, rhythms identified in the databases as VF are
evaluated as requiring defibrillation):
Sensitivity 99.30%
Specificity 99.88%
False positive rate 0.04 %
Real predictive value 97.93 %
The databases in use have a total length of 10,004 minutes. The calculation was made in
accordance with IEC60601-2-4:2010.
As cardiac rhythms requiring defibrillation, in the calculation we look at the characteristic
values of the sections in the ECG datasets above databases , which are marked with the
PysioBank Annotationscode for ventricular flutter ("[" Begin, "]" End; also refer
www.physionet.org
) using the PhysioBank Annotationscode.
These sections also contain ventricular tachycardia that however are not separately annotated
and therefore cannot form part of the statistic.
The rhythm detection system using these data thus meets the requirements of IEC 60601-24:2010 (sensitivity > 90%, specificity > 95%).
16.2 Pediatric mode
For validation the following database has been used: Development and validation dataset of
the Physical-Technical Federal Institute (PTB) Berlin. These data were collected by the PTB
as part of research project MNPQ 07/09 of the German Federal Ministry of Economics and
Technology.
Performance results:
Sensitivity 90.9%
Specificity 99.6%
False positive rate 0.4 %
Real predictive value 90.9 %
The PTB dataset consisted of 529 records which were almost evenly split into a development
and a validation data set (265/264). The development data set can be made accessible to
manufacturers. The validation dataset on the other hand must remain confidential in order to
prevent the rhythm detection system from being aligned too closely to the data. This
methodology meets the recommendations of IEC 60601-2-4:2010.
Cardiac rhythms not requiring defibrillation are found in 509 of the 529 records; cardiac
rhythms requiring defibrillation appear in just 20 of the records as these a very rare
occurrences in children. The cardiac rhythms not requiring defibrillation cover thigh blockages
and supraventricular tachycardia, as well as normal sinusoidal rhythms.
The rhythm detection system using these data thus meets the requirements of IEC 60601-24:2010 (sensitivity > 90%, specificity > 95%).
Guidelines and manufacturer's declaration – Electromagnetic
emissions
66 / 74
HeartSave AED-M 23702 / GB / A
Operating instructions
The PRIMEDIC™ HeartSave is designed for use in an environment as described below. The
customer or user of the PRIMEDIC™ HeartSave should ensure that the device is only used in
an environment of this kind.
Emitted interference
measurements
Conformance
Electromagnetic environment - code of practice
HF emissions according to
CISPR 11
Group 1
The PRIMEDIC™ HeartSave only uses HF
energy for its internal function. This means that
its HF emission is very low and it is unlikely that
electronic devices in the vicinity are disrupted.
HF emissions according to
CISPR 11
Group 2
The PRIMEDIC™ HeartSave must emit
electromagnetic energy to carry out its intended
function efficiently. Neighbouring
electromagnetic devices could be influenced.
HF emissions according to
CISPR 11
Class B
The PRIMEDIC™ HeartSave is suitable for use
in all facilities, including residential areas and
those directly connected to a public supply
network which also supplies buildings used for
residential purposes.
Emission of harmonics
according to IEC 61000-32
n/a for battery
/
PRIMEDIC™
AkuPak
Transmission of voltage
fluctuations / Flicker
according to IEC 61000-33
n/a for battery
/
PRIMEDIC™
AkuPak
17 Guidelines and manufacturer's declaration –
Electromagnetic emissions
for PRIMEDIC™ HeartSave AED/AED-M (in the following referred to as PRIMEDIC™
HeartSave)
Guidelines and manufacturer's declaration –
Electromagnetic emissions
Operating instructions
HeartSave AED-M 23702 / GB / A
67 / 74
The PRIMEDIC™ HeartSave is designed for use in an environment as described below. The
customer or user of the PRIMEDIC™ HeartSave should ensure that the device is only used in
an environment of this kind.
Interference
immunity test
IEC 60601 test
level
Level of
conformance
Electromagnetic environment guidelines
Discharge of
static electricity
(ESD) according
to IEC 61000-4-2
± 6 kV contact
discharge
± 8 kV air discharge
± 6 kV
± 6 kV air
Floor should be made of wood or
concrete or be tiled with ceramic
tiles. If the floor is covered with a
synthetic material, the relative air
humidity should be at least 30%.
Rapid transient
electrical
disturbances/
bursts according
to IEC 61000-4-5
± 2 kV for AC
power lines
± 1 kV for incoming
and outgoing lines
n/a for battery /
PRIMEDIC™
AkuPak
The quality of the supply voltage
should correspond to that of a
typical commercial or hospital
environment.
Surges
according to IEC
61000-4-5
± 1 kV normal
mode voltage
± 2 kV common
mode voltage
n/a for battery /
PRIMEDIC™
AkuPak
The quality of the supply voltage
should correspond to that of a
typical commercial or hospital
environment.
Voltage dips,
short breaks and
fluctuations in
the supply
voltage
according to IEC
61000-4-11
< 5% Ut (> 95% dip
in Ut) for ½ period
40% UT (60% dip in
UT)for 5 periods
70% UT (30% dip in
UT)for 25 periods
< 5% Ut (> 95% dip
in Ut) for 5s
n/a for battery /
PRIMEDIC™
AkuPak
The quality of the supply voltage
should correspond to that of a
typical commercial or hospital
environment. If the user of the
PRIMEDIC™ HeartSave requires
continued functioning even when
disruptions in the power supply
occur, it is recommended that the
PRIMEDIC™ HeartSave is fed
from a power supply free of
disruptions or a battery.
Magnetic field at
the supply
frequency (50/60
Hz) according to
IEC 61000-4-8
3 A/m
3 A/m
Magnetic fields at the mains
frequency should correspond to
the typical values found in a
commercial or hospital
environment.
Note: UT is the mains AC before applying the impulse test level.
Guidelines and manufacturer's declaration – Electromagnetic
emissions
68 / 74
HeartSave AED-M 23702 / GB / A
Operating instructions
The PRIMEDIC™ HeartSave is designed for use in an environment as described below. The
customer or user of the PRIMEDIC™ HeartSave should ensure that it is used in an
environment of this kind.
Immunity to
interference testing
IEC
60601
test level
Level of
conformance
Electromagnetic environment - guidelines
Conducted HF
interference
according to IEC
61000-4-6
3 Veff
150 kHz
to 80
MHz
outside
the ISM
bands a
3 Veff
150 kHz
to 80
MHz
outside
the ISM
bands a
Not
applicable
for battery
n.a.
Portable and mobile radio transceivers
should not be used any closer to the
PRIMEDIC™ HeartSave, including its
cables, than the recommended protective
distance which is calculated according to
the equation applicable for the
transmission frequency.
Recommended protective distance:
P
V
d
1
5,3
P
V
d
2
12
P
E
d
1
12
for 80 MHz to 800 MHz
P
E
d
1
23
for 800 MHz to 2.5 GHz
With P as the maximum power rating of
the transmitter in Watts (W) in accordance
with information provided by the
manufacturer of the transmitter and d as
the recommended protective distance in
metres (m). b
The field strength of stationary
transmitters in accordance with an
investigation on locations should be lower
that the level of conformance.
Interference is possible
in the vicinity of devices
which have the following
pictogram.
Radiated HF
disturbances
according to IEC
61000-4-3
10 V/m
80 MHz
to 2 GHz
10 V/m for
battery
Note 1: At 80 MHz and 800 MHz the higher frequency range applies.
Note 2: These guidelines may not be applicable in all cases. The spread of electromagnetic
factors is affected by absorption and reflection from buildings, objects and people.
a The ISM frequency ranges (for industrial, scientific and medical applications) between 150
kHz and 80 MHz are 6.765 MHz to 6.795 MHz; 13.553 MHz to 13.567 MHz; 26.957 MHz to
Guidelines and manufacturer's declaration –
Electromagnetic emissions
Operating instructions
HeartSave AED-M 23702 / GB / A
69 / 74
27.283 MHz and 40.66 MHz to 40.70 MHz.
b The conformance levels in the ISM frequency bands between 150 kHz and 80 MHz and in
the frequency band from 80 MHz to 2.5 GHz are defined to reduce the probability that
mobile/portable communication devices can cause interference, if they are unintentionally
brought into the vicinity of the patient. For this reason the additional factor of 10/3 is applied
when calculating the recommended safety distance in these frequency ranges.
c The field strength of stationary transmitters, such as base stations of wireless telephones
between 150 kHz and 80 MHz and mobile field radio transmitters, amateur radio stations, AM
and FM radio and television transmitters can theoretically not be precisely determined in
advance. To determine the electromagnetic environment with regards to the stationary
transmitters, a study of the location should be considered. If the field strength at the location
where the PRIMEDIC™ HeartSave AS is being used exceeds the upper conformance level,
the PRIMEDIC™ HeartSave must be observed to prove that it is functioning properly. If
unusual performance characteristics are observed, then it may be necessary to take additional
measures, such as changing the orientation or the location where the PRIMEDIC™ HeartSave
is being used.
Guidelines and manufacturer's declaration – Electromagnetic
emissions
70 / 74
HeartSave AED-M 23702 / GB / A
Operating instructions
Recommended protective distances between portable and mobile HF telecommunication
devices and the PRIMEDIC™ HeartSave
The PRIMEDIC™ HeartSave is designed for use in an electromagnetic environment in which
the HF interference is controlled. The customer or user of the PRIMEDIC™ HeartSave can
help avoid electromagnetic interference by maintaining the minimum distance between
portable and mobile HF telecommunication devices (transmitters) and the PRIMEDIC™
HeartSave – independently of the output power of the communication device, as shown below.
Power rating of transmitter
W
Protective distance depends on the transmission frequency
m
80 MHz to 800 MHz
P
E
d
1
12
800 MHz to 2.5 GHz
P
E
d
1
23
0.01
0.12
0.23
0.1
0.32
0.73
1
1.2
2.3
10
3.8
7.3
100
12
23
For transmitters with a maximum power rating that is not given in the table above, the distance
can be determined by using the equation that belongs to the respective column, whereby P is
the maximum power rating of the transmitter in Watts (W) according to the manufacturer of the
transmitter.
NOTE 1 At 80 MHz and 800 MHz the higher frequency range is applicable.
NOTE 2 The ISM frequency ranges (for industrial, scientific and medical applications) between
150 kHz and 80 MHz are 6.765 MHz to 6.795 MHz; 13.553 MHz to 13.567 MHz; 26.957 MHz
to 27.283 MHz and 40.66 MHz to 40.70 MHz.
NOTE 3 The purpose of the concordance levels in the ISM frequency bands between 150 kHz
and 80 MHz and 2.5 GHz is to reduce the probability of mobile/portable communication
devices causing interruptions if they are unintentionally brought into the vicinity of the patient.
For this reason the additional factor of 10/3 is applied when calculating the recommended
safety distance in these frequency ranges.
NOTE 4 These guidelines may not be applicable in all cases. The spread of electromagnetic
factors is affected by absorption and reflections from buildings, objects and people.
Safety checks
Operating instructions
HeartSave AED-M 23702 / GB / A
71 / 74
18 Safety checks
We recommend a regular technical check of our defibrillators in accordance with our
stipulations, at intervals of 24 months. The national stipulations must be observed.
Appendix
72 / 74
HeartSave AED-M 23702 / GB / A
Operating instructions
19 Appendix
19.1 Index of diagrams
Fig. 1: Front view with cover .......................................................................................... 16
METRAX GmbH
Representative office
Ul. Vavilowa 5, corpus 3
Office 406-3
119334 Moscow
Russia
Tel.: +7 495 722 1705
www.primedic.com.ru
info@metrax.ru
METRAX GmbH
Представительство в
странах СНГ
119334 Москва Россия
ул.Вавилова, д.5, корп.3,
офис 406-3
тел.: +7 495 722 1705
www.primedic.com.ru
info@metrax.ru
About Us
For 40 years, Metrax GmbH has been at
the service of medical technology and
produces professional and automated
external defibrillators for professionals and
laymen of compromise-free high quality.
Well-engineered and safe technology,
simple operation, high quality and absolute
reliability under extreme conditions. These
are the distinctive characteristics of
PRIMEDIC defibrillators.
OSI Systems took over Metrax GmbH in
October 2014. With the PRIMEDIC
defibrillators, Spacelabs Healthcare
(medical technology division of OSISystems) supplements its product palette
in the field of patient monitoring and
cardiology products.
Your dealers
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.