Gima PEEP VALVE User guide

Page 1

- 34227 PEEP VALVE

- 34228 PEEP VALVE ADAPTOR

USE AND MAINTENANCE BOOK

Product Description:

PEEP Valve indicated as an accessory to provide positive and expiratory pressure breathing capabilities to manual resuscitators.

Specification:

1. Adjustable range: 2-10 cmH2O and 5-20 cmH2O (+/- 2cm H2O) flowing capacity adjustment 3LPM
2.
Adaptor: 3mm, 22/15mm outer dimension 22 mm and 30mm
3.
Weight: 160g (without mask)

Operation:

1. Attach the diverter to the patient’s outlet as showed in the drawing.
2. Turn the diverter toward the direction away from the patient or the emergency personnel’s position.
3. Compress the resuscitator a few times to make sure all functions are normal after assembly.
4. Choose the proper PEEP valve within the specification range. (2-10cmH2O or 5-20cmHO).
5. Turn the knob of the PEEP valve to the needed manometer indicated on the valve base.
6. As showed in the drawing, attach the PEEP valve to the diverter, connect the resuscitator
patient’s outlet to manometer and the breathing bag, press on the resuscitator for the air exchange motion of the breathing bag, and adjust for the proper PEEP valve pressure needed on the PEEP valve.

Caution:

Federal law restricts this device to sale by or on the order of a physician.
This product should be used by trained personnel.

Storage:

Storage and transport temperature: -20°C to +60°C ( -4°F to 140°F)
Store the device in a cool dry place out of direct sunlight before use.

Cleaning:

1. Please disassemble the PEEP valve into the three components as above. (Please do not
disassemble the VALVE Spring as disassembly will damage the parts.)
2. Please wash the three components thoroughly in warm water containing a mild detergent.
Ensure that the detergent is compatible with the components materials to avoid causing
damage.
3. Rinse all the components thoroughly in warm water to remove all traces of detergent.
Page 2

Sterilization:

1. Autoclaved sterilization max (120/248)
2. Ethylene oxide
3. Common disinfecting solutions which are used for equipment coming into contact with the patient. Rinse thoroughly with water after this kind of sterilization.

Visual inspection through package assembly:

1. After sterilization, wait until all parts are dry.
2. Inspect all components for wear or damage. Replace as necessary.
3. After assembling the PEEP valve, shall seal bag and record the sterilization date on the bag.

Test the PEEP valve as follows:

Connect the PEEP valve to the resuscitator using the flow diverter as illustrated in directions for
use. Connect the resuscitator to a manometer and test lung. Ventilate the test lung to assure proper
function of the PEEP valve.
Page 3
Simboli - Symbols
M34
Gima S.p.A. - Via Marconi, 1
BMIU-09-04
Dispositivo medico conforme alla Direttiva 93/42/CEE
Medical Device complies with Directive 93/42/EEC
Leggere le istruzioni per lʼuso
Consult instructions for use
Attenzione: Leggere e seguire attentamente le istruzioni (avvertenze) per l’uso
Caution: read instructions (warnings) carefully
Conservare al riparo dalla luce solare
Keep away from sunlight
Data di fabbricazione
Date of manufacture
Data di scadenza
Expiration date
Codice prodotto
Product code
Fabbricante
Manufacturer
Conservare in luogo fresco ed asciutto
Keep in a cool, dry place
Numero di lotto
Lot number
227-M-1-02.20
20060 - Gessate (MI) - Italy Made in Taiwan