INSTRUCTIONS FOR USE
The operating and maintenance instructions found in this manual should be followed to ensure years of
reliable service.
Please read these instructions carefully before using the product and keep them in a safe place for
future or use reference.
INTENDED USE AND PRODUCT INFORMATIONS
The Otoscope has universal ttings, which allow a myriad of attachment to be used for aural and nasal
examination. These Otoscopes are extremely helpful in examining the external acoustic duct and the
tympanic membrane. This instrument recommended for use by medically qualied personnel. Otoscope
is equipped with a rheostat to adjust the intensity of light. The product is equipped with 3 autoclavable ear
cannulas Ø 2.5 - 3.5 - 4.5 mm and some disposable speculum Ø 2.5 - 4 mm (with an adaptor). An adapter
can be also supplied to install a disposable alligator speculum (cod. 31494), suitable to examine children
Note: The product or its components cannot be used for purpose different from the one specied in the
present manual. It is not conceived for the employment in operation room.
Due to its powerful luminous intensity, it is advisable not to directly point it at eyes.Do not use
the equipment in case it is damaged. Avoid precarious repairs. Repairs shall be carried out with original
spare parts only, which shall be installed according to the intended use.
The ophthalmoscope is designed for examination of the eye.Precision ground lenses provide 24 dioptre
settings, from +ve: 0, 1 to 6 then 8, 10, 12, 15, 20 and 40 to -ve: 0, 1 to 6 then 8, 10, 15, 20 and 25. The
May wheel containing the lenses can be rotated in either clockwise or anti clockwise direction.Before
using, thoroughly check the product and accessories (if provided) the same operation shall be carried
out after cleaning.
These products are equipped with an easy to use Bayonet Locking System.Both models are equipped
with a 2.5 V Vacuum lamp and are powered by two 1.5 V "C" type alkaline batteries.
INSTALLATION
- Before using it, it is necessary to insert the batteries into the Handle compartment.
Proceed as follows:
- Open the battery compartment. To do this, turn the end cap counter clockwise.
- Once opened, insert the batteries paying attention to the direction of poles (see picture).
- To close the battery compartment, screw down the cap clockwise and check that the batteries are in
contact with poles.
For Operating otoscope / ophtalmoscope with bayonet locking system proceed as follows:
- It is necessary to hold the handle and head in each hand and turn the head counter-clockwise until a
click is heard. At this point it is possible to take the head off.
- To re-attach the head, match the bayonet grooves on both the head and handle, push the head down
and turn the head clock wise.
The operation necessary to remove and replace the speculum is very simple:
- Turn the speculum counter clockwise and draw it out of its seat.
- To re-attach the speculum, insert it into the speculum mount by matching the grooves with the small
pivot and turn clockwise.
CURVED ILLUMINATOR
Various attachments such as mirrors and tongue depressors can be attached to the curved illuminator.
HANDLES
Battery handles contain Rheostats. To turn the power on / off, depress the white button and rotate the
Rheostat.
• Do not allow liquid to seep into the Handle heads, Ophthalmoscope, Otoscope heads and the
Laryngeal stems.
• Reusable speculums, Mirrors and Tongue Depressors can be washed with soapy water and thor-
oughly dried.
• Allow bulbs to cool down before replacing.
• Do not reuse the disposable speculum, because may cause patient-to-patient contamination.

7
ENGLISH
BATTERIES (OPTIONAL)
The Battery handles accept 2 x "C" batteries. We recommend "Alkaline" batteries for all our Diagnostic
systems for Optimum performance. Only use genuine accessories and bulbs. Alternatives may shorten
the life of your purchase
STORAGE
Since the product is made of corrosion-proof materials suitable for environmental conditions foreseen
for its normal use, it does not require special care. However it is necessary to store it in a closed place
making sure that is protected from dust and dirt to assure it's hygienic properties. Store in clean environment and preserve in normal temperature.
MAINTENANCE
If the light is intermittent or in case it does not turn on, check the bulb, the batteries and the electric
contacts. In case the lamp stops working, it is possible to replace it with original spare lamp to replace
the lamp, follow lamp replacement instructions see Fig. 1 & 2.
Periodically check the battery condition, making sure that no sign of corrosion or oxidation is present,
in case they have to be replaced, it is highly recommended to use new Alkaline ones. Carefully handle
the batteries as the liquids they contain can irritate skin and eyes.
Before using, thoroughly check the product and accessories. The same operation shall be carried out
after cleaning. Check that the connection between the head and the handle is perfect and that the rheostat works correctly (for the otoscope). If the light is intermittent or in case it does not turn on, check the
bulb, the batteries and the electric contacts.
In case the bulb stops working, it is possible to replace it with original spare lamp. (For the otoscope) to
replace the bulb, Simply turn the bulb counter-clockwise(see picture). It is advisable to use only original
bulbs, with correct voltage (2.5/3.5V) with regard to your handle.
Before removing the bulb, make sure that the instrument has been turned off for some minutes, otherwise you run the risk of being burnt.
It is advisable to always have new bulbs and batteries at disposal for the maximum light.
however they cannot be dipped into liquids.
The reusable speculum can be sterilized in boiling water or in an autoclave (in this case the temperature
shall not be over 121°C). Alternatively, cold disinfectants can be used with the same concentration used
to sterilize medical instruments.
Symbols
Caution: read instructions
(warnings) carefully
Manufacturer Date of manufacture Lot number
Keep away from sunlight Keep in a cool, dry place
Type B applied part
Disposal: The product must not be disposed of along with other domestic waste. The users
must dispose of this equipment by bringing it to a specic recycling point for electric and electronic equipment.