Gima OXY-5 PAEDIATRIC FINGER OXIMETER User guide [ar]

Page 1
Gima S.p.A. Via Marconi, 1 - 20060 Gessate (MI) Italy gima@gimaitaly.com - export@gimaitaly.com
PROFESSIONAL MEDICAL PRODUCTS
www.gimaitaly.com
PULSOXIMETRO OXY-5 OXIMETER OXY-5 OXYMÈTRE DE POULS OXY-5 PULOXIMETER OXY-5 PULSIOXÍMETRO OXY-5 MEDIDOR DE OXI-PULSAÇÕES OXY-5 ΚΟΡΕΣΤΟΜΕΤΡΟ OXY-5 OKSYMETR OXY-5 OXY-5
Manuale d’uso e manutenzione Use and maintenance book Instructions de foncionnement et entretien Betriebs und wartungs anweisungen Manual de uso y mantenimiento Manual de uso e manutenção
Εγχειριδιο χρησης και συντηρησης Podręcznik eksploatacji i konserwacji
ATTENZIONE: Gli operatori devono leggere e capire completamente questo manuale prima di utilizzare il prodotto. ATTENTION: The operators must carefully read and completely understand the present manual before using the product.
AVIS: Les opérateurs doivent lire et bien comprendre ce manuel avant d’utiliser le produit. ACHTUNG: Diese Anleitung muss vor dem Einsatz des Produkts aufmerksam gelesen und
vollständig verstanden werden. ATENCIÓN: Los operadores tienen que leer y entender completamente este manual antes de utilizar el producto. ATENÇÃO: Os operadores devem ler e entender completamente este manual antes de usar o produto.
ΠΡΟΣΟΧΗ: Οι χειριστές αυτού του προϊόντος πρέπει να διαβάσουν και να καταλάβουν πλήρως τις οδηγίες του εγχειριδίου πριν από την χρήση του. UWAGA: Użytkownik powinien uważnie zapoznać się z tym podręcznikiem przed jego użyciem.
M34282-M-Rev.6-03.19
34282 - 34265
Gima S.p.A. Via Marconi, 1 - 20060 Gessate (MI) - Italy Made in China
0476
Page 2
ARABIC
84
Page 3
85
ARABIC
Page 4
ARABIC
86
Page 5
87
ARABIC
Page 6
ARABIC
88
Page 7
89
ARABIC
Page 8
ARABIC
90
Page 9
91
ARABIC
Page 10
ARABIC
92
Page 11
93
ARABIC
Page 12
ARABIC
94
Page 13
95
ARABIC
Page 14
ARABIC
BF
%SpO2
PI
hpm
(
/
)
96
Page 15
97
ARABIC
Page 16
APPENDIX II EMC
The equipment meets the requirements of IEC 60601-1-2:2014.
Table 1
Guidance and manufacturer’s declaration-electromagnetic emission
The Fingertip Oximeter is intended for use in the electromagnetic environ­ment specied below. The customer or the user of the Fingertip Oximeter should assure that it is used in such an environment.
Emission test Compliance Electromagnetic
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions IEC 61000-3-2
Voltage uctuations/ icker emission IEC 61000-3-3
Group 1 The Fingertip Oximeter uses RF
Class B The Fingertip Oximeter suitable
N/A
N/A
environment-guidance
energy only for its internal func­tion. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
for use in all establishments, including domestic establishments and those directly network that supplies buildings used for domestic purposes.
98
Page 17
99
Guidance and manufacturer’s declaration-electromagnetic emission
The Fingertip Oximeter is intended for use in the electromagnetic environment specied below. the customer or the user of the Fingertip Oximeter should assure that it is used in such an environment.
Immunity test IEC60601
Electrostatic discharge (ESD) IEC 61000-4-2
Electrical fast transient/ burst IEC 61000-4-4
Surge IEC 61000-4-5
Voltage dips, short interruptions and voltage variations on power supply input lines IEC 61000-4-11
Power frequency (50/60Hz) magnetic eld IEC 61000-4-8
test level
±6kV contact ±8kV air
±2kV for power supply lines ±1kV for input / output lines
±1kV line (s) to line(s) ±2kV line(s) to earth
<5% UT (>95% dip in UT) for 0.5 cycle <40% UT (60% dip in UT) for 5 cycles <70% UT (30% dip in UT) for 25 cycles <5% UT (>95% dip in UT) for 5 s
3A/m 3A/m Power frequency
Compliance level
±6kV contact ±8kV air
Electromagnetic environment-guidance
Floors should be wood, concrete or ceramic tile. If oor are covered with synthetic material, the relative humidity should be at least 30%.
N/A N/A
N/A N/A
N/A N/A
magnetic elds should be at levels characteris­tic of a typical location in a typical commercial or hospital environment.
NOTE: UT is the a.c. mains voltage prior to application of the test level.
Page 18
Guidance and manufacturer’s declaration – electromagnetic immunity
The Fingertip Oximeter is intended for use in the electromagnetic environ­ment specied below. The customer or the user of The Fingertip Oximeter should assure that it is used in such an electromagnetic environment.
Immunity test
Conducted RF IEC 61000-4-6
Radiated RF IEC 61000-4-3
IEC60601 test level
3 Vrms 150 kHz to 80 MHz
3 V/m 80 MHz to
2.5 GHz
Compliance level
N/A
3 V/m
Electromagnetic environment ­guidance
Portable and mobile RF com­munications equipment should be used no closer to any part of The Fingertip Oximeter, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
Recommended separation distance
d= 1.2 √P
80MHz to 800MHz
d= 1.2 √P
800MHz to 2.5GHz
d= 2.3 √P
Where P is the maximum output power rating of the transmitter in watts (W) according to the trans­mitter manufacturer and d is the recommended separation dis­tance in metres (m). Field strengths from xed RF transmitters, as determined by an electromagnetic site survey,a should be less than the com­pliance level in each frequency range. Interference may occur in the vi­cinity of equipment marked with the following symbol.
b
b
100
Page 19
101
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies. NOTE 2 These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reection from structures, objects and people.
a Field strengths from xed transmitters, such as base stations for radio (cel­lular / cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to xed RF trans­mitters, and electromagnetic site survey should be considered. If the meas­ured eld strength in the location in which The Fingertip Oximeter is used exceeds the applicable RF compliance level above, The Fingertip Oximeter should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating The Fingertip Oximeter. b Over the frequency range 150 kHz to 80 MHz, eld strengths should be less than 3V/m.
Page 20
Recommended separation distances between portable and mobile RF communication the equipment
The Fingertip Oximeter is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of The Fingertip Oximeter can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communi­cations equipment (transmitters) and the Fingertip Oximeter as recommend­ed below, according to the maximum output power of the communications equipment.
Rated maximum output power of transmitter W (Watts)
For transmitters rated at a maximum output power not listed above, the recommended separation distanced in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer. NOTE 1 At 80MHz and 800MHz, the separation distance for the higher fre­quency range applies. NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reection from structures, objects and people.
Separation distance according to frequency of transmitter M (Meters
150KHz to 80MHz 80MHz to 800MHz 800MHz to 2.5GHz
d= 1.2 √P
0,01 N/A 0.12 0.23
0,1 N/A 0.38 0.73
1 N/A 1.2 2.3
10 N/A 3.8 7.3
100 N/A 12 23
d= 1.2 √P d= 2.3 √P
102
Page 21
103