ENGLISH
Description/Intended Use
These devices are single-use supplied sterile.
Contraindications
These devices should never be used when:
• There is visible evidence of damage to the exterior of the device such as cracked plastic or connector damage.
• These devices fail the inspection described herein.
• In the presence of ammable gases, liquids, and/or oxygen enriched environments.
Safety Tips
• Use lowest possible power setting on electrosurgical unit capable of achieving desired surgical effect.
• Never allow the cables connected to these devices to be in contact with skin of the patient or operator during
electrosurgical activations.
• Do not permit the cables connected to these devices to be parallel and in close proximity to the leads of other
electrical devices.
• Always place unused electrosurgical accessories in a safe insulated location such as a holster when not in use.
• Inspect and test each device before each use.
• Discard devices that have reached their life expectancy.
• Activation of an electrosurgical device when not in contact with target tissue or in position to deliver energy to
target tissue (fulguration) may cause capacitive coupling.
Inspection
These devices should be inspected before and after each use. Visually examine the devices for obvious physical
damage including:
• Cracked, broken or otherwise distorted plastic parts.
• Broken or signicantly bent connector contacts.
• Damage including cuts, punctures, nicks, abrasion, unusual lumps, signicant discoloration.
• Tips for damage, corrosion or misalignment condition
Storage & Handling
Electrodes (single use) must be stored in a clean, cool and dry area. Protect from mechanical damage & direct sun
light. Handle with extreme care.
2
Keep in a cool,
dry place
Manufacturer
Product code Lot number
Consult instructions
for use
Temperature limit
GIMA WARRANTY TERMS
The Gima 12-month standard B2B warranty applies.
Keep away
from sunlight
Date of
manufacture
Disposable device,
do not re-use
Expiration date
30420 - 30422 - 30425 - 30429 - 30678
Tecno Instruments Pvt. Ltd.
316C Small Industrial Estate
Sialkot, 51340 (PAKISTAN)
Importato da / Imported by:
Gima S.p.A.
Via Marconi, 1 - 20060 Gessate (MI) Italy
gima@gimaitaly.com - export@gimaitaly.com
www.gimaitaly.com
M30420-M-Rev.0-07.20
Obelis s.a.
Bd. General Wahis 53
1030 Brussels (BELGIUM)
5°C
Sterilized using
ethylene oxide
Caution: read instructions
(warnings) carefully
Authorized representative
in the European community
Medical Device complies
with Directive 93/42/EEC
1639
30°C