Manuale MCIN UK_ED2013 APRILE MFAP Definitivo.indd 113/06/2013 16:52:34
Introduction 1
The system 1
The meter 2
The display 3
The mode 3
The measurement 4
Coding the meter 4
How to obtain a drop of blood 6
Application of the blood 7
Procedure to test glucose 7
Procedure to test cholesterol and triglycerides 8
Warnings 9
The strips 10
The control solutions 11
Memory functions 12
General settings 13
Modication of the settings 13
Setting the acoustic signal 14
Setting the alarms 14
Enabling data transfer to a computer 15
Manteinance and Preservation conditions 16
Care and manteinance 16
Preservation and measuring conditions 16
Manuale MCIN UK_ED2013 APRILE MFAP Definitivo.indd 213/06/2013 16:52:35
The system
Thank you for choosing multiCare-in system.
multiCare-in is a new system, useful for the monitoring of glucose,
cholesterol and triglycerides in the blood.
multiCare-in has two different technologies:
amperometric, with multiCare-in in glucose electrodes strips.
•
reectometric, with multiCare-in cholesterol and triglycerides
•
strips.
Self –monitoring of glucose, cholesterol and triglycerides could
be helpful, but the treatment to be applied has to be anyway
agreed with the physician.
Information about multiCare-in system line
The multiCare-in system line includes all is the necessary for the analytes
check:
multiCare-in meter
•
multiCare-in strips
•
lancing device
•
sterile lancets
•
multiCare-in control solutions
•
multiCare-in data manager software
•
Read carefully the instruction for use inside the box of multiCare-in strips
that you want to use and read carefully the instruction for use of the lancing
device that are inside the multiCare-in meter box.
Manuale MCIN UK_ED2013 APRILE MFAP Definitivo.indd 113/06/2013 16:52:35
1
The meter
1. Display – Shows test results, messages, and
results stored in memory.
2. Power Button – Press this button to:
a. Turn on the meter or turn off the meter.
b. Enter the SET mode in the STANDBY mode.
c. Go to the next step in the SET mode.
d. Exit the SET mode and enter the STANDBY mode.
3. “+” Button -- Press this button to:
a. Enter the MEMORY mode.
b. Recall previous stored results in the MEMORY mode.
c. Change parameters displayed and increase
values in the SET mode.
4. “-” Button – Press this button to:
a. Enter the MEMORY mode.
b. Recall next stored results in the MEMORY mode.
c. Change parameters displayed and decrease
values in the SET mode.
d. Reset the test results.
5. Test Strip Slot – Insert test strip here.
The meter automatically enters the TEST mode
when a test strip is inserted.
6. PC Connector – Transfers data from the meter to a
computer.
7. Cover – Open this for cleaning.
8. Battery Cover.
Put your nger on this to let the battery cover slide.
9. Data-chip Slot – Insert the data-chip.
10. Ejector – Press to discharge the used strip.
Front
Rear
Manuale MCIN UK_ED2013 APRILE MFAP Definitivo.indd 213/06/2013 16:52:35
2
The display
1. Data-chip inserting.
2. Meter’s mode.
3. Low battery warning.
4. Test results and messages.
5. Unit of measurement.
6. Test analytes.
7. Abnormal temperature warning.
8. Time.
9. Day and month.
10. Alarm.
The mode
The multiCare-in has three modes.
You can check the present mode of the meter by seeing the top of
display.
1. Standby Mode
Press and release the power button.
A strip-shaped symbol will twinkle and the meter will display the data
chip number, time and date.
2. Memory Mode
Press and release the “+” or “-” button in the standby mode.
The meter will display a “MEM” symbol on the top of the display.
3. Set Mode
Press and release the power button in the standby mode.
Press again the power button for 3 seconds.
The meter will display a “SET” symbol on the top of the display.
3
Manuale MCIN UK_ED2013 APRILE MFAP Definitivo.indd 313/06/2013 16:52:35
Coding the meter
The multiCare-in meter uses a data-chip that is inside the strips
box. It will be necessary to change the data-chip in the following
cases:
a. before using a new strips boxb. when you want to change the analyte
1. Remove the data-chip from the strips box.
2. Insert the data-chip in the proper space in the lateral
part of the meter.
3. Gently press the data-chip up to the complete insertion.
Manuale MCIN UK_ED2013 APRILE MFAP Definitivo.indd 413/06/2013 16:52:36
4
4. Press power button and check the three digit number is the same of
the one printed on the strips vial.
5. If the data-chip is not inserted, it is impossible to perform the
measurement.
6. Analytes:
GLC EL=GLUCOSE ELECTRODES (BLUE)
CH = CHOLESTEROL (GREEN)
TGL=TRIGLYCERIDES (YELLOW)
Manuale MCIN UK_ED2013 APRILE MFAP Definitivo.indd 513/06/2013 16:52:37
5
How to obtain a drop of blood
To perform a measurement with multiCare-in System, it is necessary to
withdraw a drop of capillary blood using the following procedure:
• Wash the hands with hot water to increase the blood circulation in
the ngers to make easier the withdrawal of blood.
• Dry well the hands.
• Check that the nger to be pricked is completely dried.
• Prepare the lancing device that is inside the multiCare-in kit
by inserting the sterile lancet.
Read carefully the instruction for use of the lancing device contained
in the meter kit.
Use a new, not damaged lancet for each prick and use the lancing device
only on yourself.
Before the inserting, verify that the lancets and the protection of the
lancets are intact.
Only a lancet protection intact can guarantee sterile lancet.
Store the protection of the lancet in order to use it to seal the lancet after
use.
Manuale MCIN UK_ED2013 APRILE MFAP Definitivo.indd 613/06/2013 16:52:38
6
Application of the blood
Read carefully the instruction for use of the strips that you want to use to
perform the measurement.
1. Press the Power button after inserted
data-chip.
The number that appears on the display
has to match the one printed on the strips
vial.
If it does not match, contact Customer
Service.
2. Take a strip from the vial and insert it in the meter with the printed or
coloured part on the upper side. You will hear a click and the meter will
emit an acoustic signal.
Procedure to test glucose:
3a. Perform the prick on the nger with the
lancing device.
Apply a drop of blood on the part of the
strips that comes outside the meter.
Let the strip aspirate the blood up to see
the moving segments on the display.
If the volume is not sufcient the measurement does not start.
Manuale MCIN UK_ED2013 APRILE MFAP Definitivo.indd 713/06/2013 16:52:38
7
Procedure to test cholesterol and triglycerides
-
-
-
3.b.1 Sit down for one minute with the arm along the body.
Massage the nger to be pricked in order to increase the blood ow.
Prick the nger , remove the rst drop with a sterile gauze and gently
press the nger to obtain a second drop. Without touching and
pressing the nger on the strip, apply the drop on the white circular
area of the strip that comes out of the meter.
Wait until moving segments will appear on the display.
3.b.2 After insertign the strip, if “F” (Female)
on the left and “M” (Male) on the right
will appear on the display. Press the
corresponding button (“- “ for”F” or “+” for
“M”) to select the gender. After about two
seconds, “b” (before meal) on the left and
“a” (after meal) on the right will appear on
the display. Press the corresponding button
+
+
(“- “ for”A” or “+” for “P”) to select the letter.
The selected letter will remain two seconds
on the display. After that, the code and the
blinking drop will appear on the display,
indicating that the meter is ready to receive
the blood.
Follow 3.b.1 for the blood application.
4. The meter will emit an acoustic signal and the result will be visualised
on the display and automatically stored in the memory of the meter
(refers to the instruction for use, that you are using to know the reading
time).
5. Press the ejector key to remove the strip that you used, pointing the
Manuale MCIN UK_ED2013 APRILE MFAP Definitivo.indd 813/06/2013 16:52:38
meter down for the low.
8
Warnings
If the results obtained do not correspond to the well being
of the user, proceed with a check up of the multiCare-in
instrument and contact your physician.
If extreme results are obtained, too high (Hi) or too low (Lo),
contact your physician immediately.
When the measuring is nished, it is possible to turn OFF the multiCare-in
instrument. The instrument will shut down automatically after 60 seconds
if not used for other operations.
The measured value, along with date and time, will be saved in position 1
in the memory of the instrument.
The values previously measured will be saved in positions 2, 3 etc. starting
from the most recent. How to recall from the memory a measured value
and how to erase memorized values is described on page 12.
+
Manuale MCIN UK_ED2013 APRILE MFAP Definitivo.indd 913/06/2013 16:52:38
The strips and the used materials have to be
properly disposed, avoiding contamination. The
meter has to be disposed following the Electric and
Electronic Waste law.
In a laboratory and in a hospital, the procedures
of biological safety must be followed when
handling potentially infected material, in particular
when measurements are performed on different
patients.
To avoid infections clean the instrument regularly
and dispose of the used strips in a way that there is
no risk of infections for other people.
9
The strips
The correct use of the multiCare-in strips is very
important for obtaining accurate results. Therefore
it is important to keep in mind the following:
• The strips are sensitive to light and humidity:
keep the strips only in their original vial.
• Protect the vial from direct sunlight and keep at
a temperature between 5 and 30 °C (41-86°F).
• Use the strips immediately after removing from the
vial and close the vial immediately with its cap.
• The strips must not be used after the expiry date.
• When rst opening, write the date on the vial:
the strips must be used within 90 days after the
vial has been opened.
• Do not use strips that are damaged.
• Carefully read the instructions contained in the
package of strips.
•Take measurements only with original
multiCare-in strips.
Manuale MCIN UK_ED2013 APRILE MFAP Definitivo.indd 1013/06/2013 16:52:39
10
The Control Solutions
Read carefully the instruction for use of the control solution that
you want to use.
It is possible to check the working of the multiCare-in instrument
using the multiCare-in control solutions.
The control solutions can be obtained as a separate accessory
through specialized retailers.
The control solution is applied on the strip just like blood.
When testing is done with the control solution, the obtained result
must be within the interval as shown on the label of the strip’s vial.
In this way, it is possible to learn how to use the multiCare-in properly
and at the same time, check if it is working.
Write the date of opening on the vial when the solution is opened for
the rst time.
The control solution must be used within 90 days from rst opening,
if the expiry date is valid.
When to carry out a working check up using the control solutions:
• When suspecting a possible defect of the multiCare-in instrument.
• When suspecting the possibility of a malfunction or wrong
measurement.
The result obtained with the control solutions has to be deleted by
the memory of the meter in order to not interfere with the average
calculation of the glycaemia.
Manuale MCIN UK_ED2013 APRILE MFAP Definitivo.indd 1113/06/2013 16:52:39
11
Memory functions:
recalling and deleting result from the
memory
The multiCare-in meter stores up to 500 test results for
all analytes, letting you review them in order from the
recent (“1”) to the oldest (“500”).
If you insert the data-chip of glucose into the meter,
you allow to recall only previous glucose test results.
It’s all the same for other analytes.
1. Press and release the “+” or “-” button in the
STANDBY mode.
The meter will enter MEMORY mode and the
average of test results is displayed.
2.
While pressing the “+” or “-” button, memory number
will be displayed. If you hold down the “+” or “-”
button, the meter scrolls quickly and only memory
positions are displayed. A result is not shown until
you release the button. When release the “+” or “-”
button, the corresponding result will be displayed.
3. Repeat step 2 to see the remaining stored results.
4. Keep the power button pressed and then press the
“-” button for 1 second to delete one test result
stored in the meter.
Manuale MCIN UK_ED2013 APRILE MFAP Definitivo.indd 1213/06/2013 16:52:39
5. Press “-” button for 5 seconds together with the
power button to delete all test results stored in
the meter.
12
Modicationofsettings
1. Press and release the power button to select the
month.
Press and release the “+” or “-” button up to
reach the correct month.
2. Press and release the power button to select the
day.
Press and release the power button to change.
3. Press and release the power button to choose
the date format.
Press and release the “+” or “-” button to choose
between month-day (mm-dd) or day-month (ddmm).
4. Press and release the power button to change
the year. Press and release the “+” or “-” button
up to reach the actual year.
The meter allows to change the year from 2007
to 2017.
5. Press continuously the power button for 3
seconds. The meter will enter in the SET mode
and it will be possible to change the hour.
Press and release the “+” or “-” button up to
reach the correct hour.
Manuale MCIN UK_ED2013 APRILE MFAP Definitivo.indd 1313/06/2013 16:52:39
6. Press and release the power button to change
the minutes. Press and release the “+” or “-”
button up to reach the correct minutes.
13
7. Press and release the power button to change the
hour format. Press and release the “+” or “-” button to
choose between 12 or 24 hours.
8. Press and release the power button to
change the unit of measurement.
Press and release the “+” or “-” button to
choose between mg/dL and mmol/L.
9. Press and release the power button to select how many
days you want the average of results. Press and release
the “+” or “-” button up to reach the wished number of
days (0,7,14,21,28 days-“0” means that the meter will
not show any average of the measurements).
Setting the acoustic signal
1. Switch on the meter and press continuously
the power button to enter in the SET
mode.
Press and release the power button up to
see “bp” xed and “On” blinking on the
display.
2. Press the “+” or “-” button to activate or
not the acoustic signal.
Setting the alarms
Follow these instructions to set the alarms:
1. Press and release the power button of the meter.
Press the power button continuously to enter in the
SET mode. Press and release up to see the icon of the
alarm xed and “0” blinking.
Manuale MCIN UK_ED2013 APRILE MFAP Definitivo.indd 1413/06/2013 16:52:39
14
2. Press the “+” or “-” button to select the number of the
alarm to be set (0, 1, 2, or 3).
3. Press the power button again to select the hour
and the minutes related to the alalrms that you are
setting.
4. You can exit by the SET mode in every moment by
pressing continuously the power button for three
seconds.
The alarms are not activated if they are set on 12.00
or 0.00.
Enabling data transfer to a computer
All the test results stored in the multiCare-in
can be downloaded to a computer.
1. Install the software according to the manufacturer’s instructions.
2. Insert the cable into the USB port of the computer and the jack of the
cable into the meter.
3. Press and release the power button. “PC” will appear in the meter’s
display.
4. Set up your computer software to receive information from the meter.
After the data has been transferred, you can turn off the meter by
pressing the power button and remove the jack cable from the port of
the meter.
Manuale MCIN UK_ED2013 APRILE MFAP Definitivo.indd 1513/06/2013 16:52:39
15
Care and maintenance
For glucose measurement it is necessary a particular cleaning
procedure only if blood or other material (debris, dust, etc.)
come into direct contact with the plastic support of the reader. If
so, use only a wet cloth, moistened by water or alcohol solution
of 70%, to clean the external parts of the instrument.
Since for cholesterol or triglycerides measurement the
instrument uses an optical measuring method, its components
must be kept clean. To clean the optical components, lift the
front cover and use a wet cloth on the glass surface of the
optics. After the cleaning, dry the plastic support of the reader,
without scratching it.
For taking measurements on different people, like in a hospital
for example, clean the instrument after every measurement
with a solution 70% alcohol before proceeding with the next
measurement.
Preservation and measuring conditions
To assure the accuracy of each measurement, it is important to meet some
conditions.
The measurement with multiCare-in strips has to be performed between 10
and 40 °C (50-104°F). The multiCare- in kit (meter and strips) has not to be
stored at temperature lower than 5°C(41°F) and over 30°C(86°F).
Manuale MCIN UK_ED2013 APRILE MFAP Definitivo.indd 1613/06/2013 16:52:39
If the instrument is too cold or too hot (ex. after being in a car at very
Note
low or high temperatures), allow it to cool down or to heat up slowly
at room temperature (about 20 minutes).
16
Manuale MCIN UK_ED2013 APRILE MFAP Definitivo.indd 1713/06/2013 16:52:44
17
Manuale MCIN UK_ED2013 APRILE MFAP Definitivo.indd 1813/06/2013 16:52:47
18
Manuale MCIN UK_ED2013 APRILE MFAP Definitivo.indd 1913/06/2013 16:52:47
19
The multiCare-in instrument houses two lithium batteries 3V
Substituting the batteries
(type CR 2032).
If the battery icon is blinking, the charge is low and the battery
must be replaced before taking a measurement.
Always insert a new battery and not a used one.
Shut down the instrument before removing the battery.
Check out the new battery to see if it is of the right type before
placing.
How to change the battery:
• Open the sliding rear panel of the instrument (1).
• Remove the battery from the plastic holder (2).
• Place the new battery so that the type is readable. The negative side
of the battery must be at the top (3).
• Close the back cover until a click is heard.
Flat batteries can be disposed of through recycling centers.
Manuale MCIN UK_ED2013 APRILE MFAP Definitivo.indd 2013/06/2013 16:52:48
If the battery is replaced or simply removed for more than 60 seconds,
Note
it is necessary to set time and date again.
20
Information on the conditions
of warranty
Biochemical Systems International Srl grants to the customer
that the instrument will be free of material defects and
manufacturer handling for 3 years starting from the purchasing
date. During such period, Biochemical Systems International
Srl should replace the defected instrument with an equivalent
version or the same model after verifying the defect.
Such warranty is subject to exception and limitation as
described below. The warranty is limited to the replacement
due to defects in some parts and handling. Instruments that
present defects due to damages caused from improper use,
alteration, negligence, technical assistance performed from not
authorised personnel, accident, abuse or by using strips which
are different from multiCare-in will not be replaced.
Biochemical Systems International Srl reserves the right to apply
technical modications to the instrument multiCare-in without
the obligation to insert such specication to those products
already introduced in the market.
Any other warranty will be granted for this product unless the
replacement option described above.
Every responsibility for special, indirect or consequential
damages is excluded.
Manuale MCIN UK_ED2013 APRILE MFAP Definitivo.indd 2113/06/2013 16:52:48
21
✂
Manuale MCIN UK_ED2013 APRILE MFAP Definitivo.indd 2213/06/2013 16:52:48
Manuale MCIN UK_ED2013 APRILE MFAP Definitivo.indd 2313/06/2013 16:52:48
23
Manuale MCIN UK_ED2013 APRILE MFAP Definitivo.indd 2413/06/2013 16:52:51
24
Manuale MCIN UK_ED2013 APRILE MFAP Definitivo.indd 2513/06/2013 16:52:51
multicarein@biosys.it
0344
ED2013 MAGGIO-E
Manuale MCIN UK_ED2013 APRILE MFAP Definitivo.indd 2613/06/2013 16:52:51