Gima MIC Wi-Fi & USB CAPILLAROSCOPE, MIC Wi-Fi & USB POLARIZED DERMATOSCOPE User guide

Page 1
USER MANUAL
Wi-Fi - USB DIGITAL DEVICE
MDR 2017/745/UE
MEDICAL DEVICE IN CLASS 1
FEDMEDMICFI
MODELS
2020 Copyright, ITALECO S.R.L. All rights reserved.
Reproduction or duplication of this manual in any form is not permitted
or any part of it without prior permission from ITALECO S.R.L.
Last date manual revision: Rev. 15 April 2022
Page 2
WARNING! USERS SHOULD READ THIS MANUAL VERY CAREFULLY BEFORE USING THE DEVICE.
MANUFACTURER
ITALECO SRL
Via Pavia 105/i, 10098, Rivoli (To), Italy
Tel: +39-011-5690297
Fax: +39-011-5690298
http://www.mic-fi.com
ITALECO S.R.L. won’t be held responsible in case of damages or injuries due to an illegal or improper use of the product resulting from failure to follow instructions, warnings, notices or indications for use, published on this manual. “Mic-Fi” software/app, provided together with this device, is under the Copyright. All rights are reserved. The software is protected by the copyright law and the provisions of applicable internationals treaties. In accordance to these laws, the licensee is permitted to use a copy of the software embedded in it in the manner provided for the operation of the products in which it is incorporated. The software or copies of the software are not for sale, all right, title and ownership of the software belongs to manufacturer or its suppliers. For further information about the device, please contact:
2
Page 3
1. Summary
1. UNPACKING ..................................................................................................................... 6
2. PACKAGING CONTENTS .................................................................................................... 6
3. SYMBOLS ......................................................................................................................... 7
4. WARNINGS ...................................................................................................................... 9
5. INTRODUCTION.............................................................................................................. 10
5.1 DESTINATION OF USE ...................................................................................................................... 10
5.2 Technical features ........................................................................................................................... 11
6. GENERAL GUIDELINES FOR HANDLING ............................................................................ 12
7. GENERAL MAINTENANCE ............................................................................................... 13
8. INSTRUCTION FOR USE ................................................................................................... 14
8.1 DERMATOSCOPE .............................................................................................................................. 15
8.2 CAPILLAROSCOPE ............................................................................................................................ 16
8.3 OTOSCOPE ....................................................................................................................................... 18
8.4 DENTALSCOPE .................................................................................................................................. 19
8.5 CAMERA for OPTICAL DEVICES ........................................................................................................ 20
9. INSTALLATION AND USE OF THE SOFTWARE AND APP’s ................................................. 20
9.1 INFORMATION ................................................................................................................................. 20
9.2 WARNING! ....................................................................................................................................... 21
9.3 USE Mic-Fi Software and Apps ........................................................................................................ 21
9.3.1 Modalità Wi-Fi ......................................................................................................................... 21
9.3.2 Modalità USB ........................................................................................................................... 21
9.3.3 Simultaneous use of the device with others Wi-Fi net ........................................................... 22
9.4 ”Mic-Fi Medical” on iPad ................................................................................................................. 23
9.4.1 Insert a new patient................................................................................................................. 23
9.4.2 Analysis .................................................................................................................................... 24
9.4.3 Historical .................................................................................................................................. 25
9.5 “Mic-Fi MED Plus” on Android Tablet /Phone ................................................................................. 26
9.5.1 Add a patient ........................................................................................................................... 26
9.5.2 Analysis .................................................................................................................................... 27
9.5.3 Historical .................................................................................................................................. 28
9.5.4 Report ...................................................................................................................................... 28
9.6 “Mic-Fi Med” Wi-Fi and USB per Windows ..................................................................................... 29
9.6.1 Tool Bar .................................................................................................................................... 29
3
Page 4
9.6.2 Calibration ............................................................................................................................... 30
9.6.3 File. .......................................................................................................................................... 31
9.6.4 Setting ...................................................................................................................................... 32
9.6.5 Snapshot Enable ...................................................................................................................... 33
9.7 Mic-Fi Wi-Fi and USB for MAC ......................................................................................................... 34
9.7.1 Installation ............................................................................................................................... 34
9.7.2 Mode of use ............................................................................................................................. 35
9.7.3 Setting ...................................................................................................................................... 36
9.7.4 Snapshot .................................................................................................................................. 36
9.7.5 Calibration ............................................................................................................................... 37
9.8 Mic-Fi DERMA per Windows ........................................................................................................... 38
9.8.1 Installation ............................................................................................................................... 38
9.8.2 Patient management ............................................................................................................... 39
9.8.3 ANALYSIS .................................................................................................................................. 39
9.8.4 Calibration ............................................................................................................................... 40
9.8.5 Operations available on the image and save .......................................................................... 41
9.8.6 “Historical”: Image historical ................................................................................................... 42
9.8.7 REPORT .................................................................................................................................... 43
9.8.8 All Report ................................................................................................................................. 43
9.9 “Mic-Fi Derma” per MAC ................................................................................................................ 44
9.9.1 Installation ............................................................................................................................... 44
9.9.2 Add patient .............................................................................................................................. 45
9.9.3 Select Points ............................................................................................................................ 46
9.9.4 Analysis .................................................................................................................................... 46
9.9.5 Calibration ............................................................................................................................... 47
9.9.6 Operation available ................................................................................................................. 48
9.9.7 “Historical”: Image historical ................................................................................................... 49
9.9.8 Reports .................................................................................................................................... 49
9.9.9 All Reports ............................................................................................................................... 50
9.10 Mic-Fi Capillaries per Windows ....................................................................................................... 50
9.10.1 Installation ............................................................................................................................... 50
9.10.2 Add patient .............................................................................................................................. 52
9.10.3 Select Points ............................................................................................................................ 53
9.10.4 Analysis .................................................................................................................................... 53
9.10.5 Calibration procedure .............................................................................................................. 54
4
Page 5
9.10.6 Operation available ................................................................................................................. 55
9.10.7 “Historical”: Image historical ................................................................................................... 56
9.10.8 Reports .................................................................................................................................... 56
9.10.9 All Reports ............................................................................................................................... 56
9.11 Mic-Fi Capillaries per MAC .............................................................................................................. 57
9.11.1 Installation ............................................................................................................................... 57
9.11.2 Add patient .............................................................................................................................. 58
9.11.3 Select Points ............................................................................................................................ 59
9.11.4 Analysis .................................................................................................................................... 59
9.11.5 Calibration ............................................................................................................................... 60
9.11.6 Operation available ................................................................................................................. 61
9.11.7 “Historical” ............................................................................................................................... 62
9.11.8 Reports .................................................................................................................................... 62
9.11.9 All Reports ............................................................................................................................... 63
10. BATTERY CHARGER ........................................................................................................ 63
11. CLEANING THE DEVICE ................................................................................................... 64
12. DISPOSAL ....................................................................................................................... 65
13. WARRANTY .................................................................................................................... 66
5
Page 6
1. UNPACKING
We kindly remember that the packaging element (paper, cellophan, staples, tape and so on) can cut the skin or injure if handled careless. It is important to handle them carefully and by using proper equipments; do not let equipments and packaging elements near people who are not responsible. Once you have opened the packaging, please check properly its content to be sure that all parts are inside it and that they are in perfect condition.
2. PACKAGING CONTENTS
The devices belonging to the family Microscopes and digital camera Wifi-USB Mic-Fi are supplied to the user in a package, which contains:
One Mic-Fi device One CD with software + manual One Charger 5.0V/1A One USB - mini USB cable One quick guide Accessories: depending on the intended use
6
Page 7
3. SYMBOLS
WARNING This symbol shown on this manual indicates that some condition or
procedures, which are not respected, could cause injuries or damages to the user or to the patient if occurs a misuse of the device.
WARNING Draws attention to the label of the device and indicates to "read the instructions for use".
High
Fragile
Keep dry
Atmospheric pressure limits
Limits of atmospheric humidity
Recyclable
Logo of the manufacturer.
Electromedical class II device in accordance with the provisions of technical standard EN 60601­1:2006+A1:2012.
Brand medical products company Italeco Srl.
Electromedical device with type B applied parts in accordance with the provisions of technical standard EN 60601­1:2006+A1:2012.
The symbols shown on the following pages can be found on the body of the device, on the packaging, on the transport container or in this manual.
WARNING SYMBOLS
SYMBOLS FOR MAINTENANCE, TRANSPORT AND ENVIRONMENT
SYMBOLS OF DEVICE
7
Page 8
Indicates that the product meets the essential requirements of the MDR 2017/745/UE for medical devices
IP40
IP Code (International Protection) enclosure protected against solid bodies larger than 1mm and against access with a wire, but not protected from the entry of liquids.
Manufacturer. Indicates the manufacturer of the medical device.
Indicates compliance with the provisions of Directive 2015/863/EU, so-called Rohs III.
Medical device covered by Directive 2012/19/EU and s.m.i. so-called WEEE.
Spent batteries are considered special waste. Do not throw them into household waste.
Device in accordance with the provisions of the Federal Communications Commission (FCC). The device does not cause interference with other devices commonly present in the office or at home
Date of manifacture Indicates the batch code of the manufacturer, thus allowing the batch or lot to be identified.
Indicates the serial number of the device, so that it can be uniquely identified (alphanumeric code + date of manufacture)
IDENTIFICATION SYMBOLS
8
Page 9
4. WARNINGS
The device is suitable for external use only.
The device does not make diagnosis.
The size that can be viewed from Mic-Fi SW and App are indicative and must not to affect to the final judgment borne by the doctor. The Mic-Fi devices are not devices with a measuring function
DO NOT INSERT THE DE VICE IN HUMAN OR ANIMAL ’S ORIFICE. DANGER OF HEAVY INJURIES.
Keep out of reach of children.
The Mic-Fi Device is not sterile. Make sure that the covers is disinfected before and after each use.
Before using the digital device please wear disposable gloves in order to keep a suitable hygienic level.
Do not direct the beam of light toward the eyes.
Do not use the Mic-Fi device for capture images or videos of eyes.
Do not use the Mic-Fi device on open wounds or on painful parts of the body.
Do not use the device in close proximity to high heat, magnetic fields.
Do not use the device in contact with lenses, glasses or other surfaces that can change the perceptions of the images.
Being devices in support of diagnosis and not devices that perform some type of action against a specific disease, there are no contraindications from the use of medical devices belonging to the family Microscopes and digital camera Wi-Fi-USB Mic-Fi and related accessories. Italeco Srl is not responsible for the consequences of inappropriate use by the user, it assumes no responsibility in case of injury or illegal and improper use of the product resulting from a failure to comply with the instructions, warnings, notices or instructions for use published in this Manual.
READ THE MANUAL CAREFULLY BEFORE USING THE DEVICE!
9
Page 10
Do not use the device in Wi-Fi mode during the charging.
Please note that in case of a serious accident or failure to do so, the user is required to promptly inform the Manufacturer and the Competent Authority of the incident, without undue delay.
5. INTRODUCTION
Mic-Fi device despite being extremely easy to use is a medical device, and we therefore recommend reading this book before using the equipment.
The Mic-Fi device must be used exclusively by medical personnel in the medical office, surgeries and/or hospitals.
WARNING! THE DEVICE IS NOT INTENDED FOR CLINICAL INVESTIGATIONS
Thank you for purchasing the Mic-Fi device. The Mic-Fi device is able to combine visual analysis (magnification, photo, video) with extreme ease of handling and ease of use.
5.1 DESTINATION OF USE
The Mic-Fi Wifi-USB Digital Microscope devices are intended to be used to support the clinical expert in improving the visualization and saving of related images and videos, depending on the specific application and then the configuration of the device, the skin, the ear canal, oral and rectal. The fields of application of Mic-Fi Wifi-USB Digital Microscope devices, in fact, according to the accessories supplied and therefore to the device configuration, are: dermatology, rheumatology, otorhinolaryngology, endoscopy, proctology and dentistry.
The Mic-Fi Wifi-USB Digital Microscope devices are medical devices to support diagnosis, which allows to observe, according to the accessories supplied in the packaging or supplied to the clinical expert: capillaries, skin, hair, oral cord or ears. The device technology allows the display on a common PC/tablet and/or smartphone of what is observed from the integrated camera to the device, thanks to the installation of the appropriate software/Mic-Fi application on the PC/tablet and/or smartphone itself. The captured image is illuminated by a LED light source embedded inside the body of the Digital Microscope Wifi-USB Mic-Fi (excluding the Camera), whose operation is guaranteed through the Lithium battery power supply.
10
Page 11
The use of medical devices belonging to the family Microscopes and digital camera Wi-Fi-USB Mic-
GENERAL FEATURES
Operating System
Windows, OS X, IOS, Android
Battery
Li-ion, Continuos working time appx. 2 hours
Wall adaptors
DC5.0V/1A Please use ONLY usb wall adaptors that are certified according the CE and/or FCC regulations
Device class according to EN 60601­1:2006+A1:2012
Class II
Applied parts according to EN 60601­1:2006+A1:2012
Type B
Degree of protection (EN 60529:1991+A2:2013)
IP40 (casing protected against solids larger than 1mm and against access by a wire, but not protected against the ingress of liquids)
Mode of work
Continuous
Lifetime
5 years
Wi-Fi TRANSMISSION FEATURES
Wi-Fi standard
IEEE 802.11 b/g/n
Wi-Fi Wireless Transmission Distance
Up to 5m
Photo and Video Resolution
1.3Mpixel – 320(H)x240(V), 640(H)x 480(V), 1280(H)x1024(V)
Frame Rate
15fps – 1280x1024, 30fps – 640x480, 320x240
Wi-Fi Transmit Power
11n HT40 MCS7: +13dBm, 11bCCK: +18dBm, 11gOFDM: +15dBm
Data rate
802.11n: up to 150Mbps,
802.11b: 1, 2, 5,5, 11Mbps,
802.11g: 6, 9, 12, 18, 24, 36, 48, 54Mbps
Frequency Range
2.400÷2.4835GHz
Power Consumption
Max 2.5W
Frame Rate
10÷30fps
Fi and its accessories is suitable for all patients, or any person of age, gender and race or origin.
5.2 TECHNICAL FEATURES
The devices Microscope and digital camera Wifi-USB Mic-Fi are equipped with the technical characteristics shown in the following table:
11
Page 12
Photo and Video Resolution
1.3 or 5MPixel 1280x1024, 640x480, 320x240
Frame Rate
15fps (1280x1024), 30fps(640x480), 320x240
Image Format
.bmp, .jpg
PC interface
USB 2. 0
Mini USB cable
Depends on the device
DERMATOSCOPE FEATURES
Magnification
Depends on the device
Object distance
Depends on the device
Focus Range
Depends on the device
LED
Depends on the device
Lens
Depends on the device
Dimensions
Depends on the device
Weight
Depends on the device
CAMERA FEATURES
Camera
Color CMOS Camera
Sensor
Depends on the device
Optic
Exposure: auto Gamma: auto White balance: auto Reference: auto
Do not use the Mic-Fi device with wet hands.
Do not use the Mic-Fi device if damaged or if it does not work correctly.
Always ensure that the device is properly connected to the hardware on which you installed the software/application supplied.
Do not use the device during the charging of the battery.
Do not subject the Mic-Fi device to sudden shocks.
USB FEATURES
6. GENERAL GUIDELINES FOR HANDLING
We would always recommend wearing disposable gloves when using the device, however this can reduce the perception while operating with the device. Use alcohol disinfectants.
12
Page 13
7. GENERAL MAINTENANCE
Do not place the device and accessories in direct sunlight, high temperatures or in humid or dusty places.
Do not immerse the device in liquids.
DO NOT EVER DISASSEMBLE THE DEVICE.
Before use, the product must be checked in all its parts, even after cleaning. Before use, disinfect the device in whole or in part (the assessment of this need must be carried out by the doctor) and in any case clean at least the part in contact with the patient with alcohol or common disinfectant. After cleaning, use the device only after 15 minutes have elapsed. For more details, refer to paragraph 11. Cleaning the device.
The product is made of corrosion-resistant materials and the environmental conditions required for normal use, so it does not need specific attention; however it must be kept in a closed environment, clean and dry, avoiding exposure to light and atmospheric agents, taking care to protect it from dust in order to ensure hygiene conditions.
13
Page 14
8. INSTRUCTION FOR USE
The following are the indications for each button/button on the body of the device and for each light symbol:
Wi-Fi: Press for 2 seconds; the microscope will generate a Wi-Fi network Wi-Fi Indicator (Blue LED): Flashing light indicates data transmission; constant light indicates no data
transmission.
Power Indicator (Green LED): Lights up when Wi-Fi power is on or connected to PC or
AC adaptor via MiniUSB cable.
LED: Press for 2 seconds to switch the LED; press + and to increase or decrease the
intensity of light or to change LED if the device is provided.
P.S. this key does not work in the FEDMEDMICFIEYE/EYE5M models. Photo: Press to capture the picture directly from microscope. Snapshot Button: Every time after the successful connection of the Wi-Fi microscope, you
have to enable the Snapshot feature on the App (The LED illumination will be off and then lights up again). The Snapshot only works with the receiver device when the App snapshot feature has been enabled.
14
Page 15
8.1 DERMATOSCOPE
The device is intended for external use only. Do not enter the device in the oral cavity
Do not use the device in contact with wounds or painful parts of the body.
The part of the device that needs to come into contact with the patient is made up of a transparent / opaque or black cover that must be disinfected after each use. If other parts of the Mic-Fi device should come into contact with the patient, it is sufficient to disinfect them. The Mic-Fi device is built with biocompatible materials. MICFICL30 and MICFICLOSEDCAP are disposable covers.
a) Fit the specific cover according to the enlargement that you want to display. b) Connect the device to the computer/tablet or smartphone c) Launch the Mic-Fi software (previously installed) See paragraph 9 d) Put the part to be observed below and in the centre of the illuminated area of your Mic-Fi
device.
e) Turn the focus wheel to have a clear image. f) Click on the “capture” button to get a photo or video g) Remove the device from the patient after the viewing
With the 10mm cover you get two enlargements on contact: about 60x and 200x; with the 30mm cover you get a contact magnification of about 30x.
The FEDMEDMICFIP and FEDMEDMICFIP5M models allow to polarize the light: act on the transparent ring in order to obtain the desired visualization. The FEDMEDMICFIUVWP model also allows polarization: keep the button with the LED symbol pressed for two seconds.
Note: For optimal use of the device without support, grip the device between thumb and index accommodating it in the palm of the hand and using your thumb to rotate the wheel of focus. The tool can be rotated on its axis in order to achieve the rotation of the image on the screen.
Please ask to the patient to assume a comfortable position and not to move during the whole visit.
15
Page 16
The size that can be viewed from Mic-Fi SW and App are indicative and must not to affect to the final judgment borne by the doctor.
8.2 CAPILLAROSCOPE
Use without stand
a) Connect your device to your PC/Tablet or Smartphone. b) Start the software/application previously installed (see paragraph 9). c) Place a drop of cedar oil preferably or dermatological oil on the periungual skin. d) Put the specific cover depending on the magnification you want to display. e) Place the device on the part to be observed. f) Rotate the ring 2 slowly until a sharp image is obtained. g) Capture the image through the button on the device, from the Sw/App or from the pedal. h) Remove the device from the patient after the viewing.
With the 10mm cover you get two enlargements on contact: about 60x and 200x; with the 30mm cover you get a contact magnification of about 30x.
The FEDMEDMICFIP and FEDMEDMICFIP5M models allow to polarize the light: act on the transparent ring in order to obtain the desired visualization. The FEDMEDMICFIUVWP model also allows polarization: keep the button with the LED symbol pressed for two seconds. The FEDMEDMICFIT6 model allows viewing at high magnifications (about 500x). Use the flat 10mm, 4.5mm or the concave cover included in the package depending on your needs.
16
Page 17
Note: For optimal use of the device without support, grip the device between thumb and index
a) Put the device on the support. b) Remove the transparent 10mm cover. c) Connect your device to your PC/Tablet
or Smartphone. d) Start the software/application previously installed (see paragraph 9). e) Put a drop of cedar oil preferably or
dermatological oil on the periungual
skin. f) Place the part to be observed under the
of the device. g) Place the ring 2 at about 60x for a larger
field of view; rotate the ring 1 slowly
until you get a sharp image; capture the
image through the button on the
device, from the Sw/App or with the
pedal. h) Place the ring 2 at the maximum
magnification; rotate the ring 1 slowly
until you get a sharp image; capture the
image through the button on the
device, from the Sw/App or with the
pedal.
accommodating it in the palm of the hand and using your thumb to rotate the wheel of focus. The tool can be rotated on its axis in order to achieve the rotation of the image on the screen. Please ask to the patient to assume a comfortable position and not to move during the whole visit.
Use with stand
The FEDMEDMICFIP and FEDMEDMICFIP5M models allow to polarize the light: act on the transparent ring in order to obtain the desired visualization. The FEDMEDMICFIUVWP model also allows polarization: keep the button with the LED symbol pressed for two seconds. The FEDMEDMICFIT6 model allows viewing at high magnifications (about 500x). Use the flat 10mm, 4.5mm or the concave cover included in the package depending on your needs.
17
Page 18
The device is intended for external use only. Do not enter the device in the oral cavity
Do not use the device in contact with wounds or painful parts of the body.
The part of the device that needs to come into contact with the patient is made up of a transparent / opaque or black cover that must be disinfected after each use. If other parts of the Mic-Fi device should come into contact with the patient, it is sufficient to disinfect them. The Mic-Fi device is built with biocompatible materials. MICFICL30 and MICFICLOSEDCAP are disposable covers.
The size that can be viewed from Mic-Fi SW and App are indicative and must not to affect to the final judgment borne by the doctor.
Earscopic investigations must be achieved by keeping the device parallel to the floor of the room in order to minimize the risk to cause pain to the patient or eardrum perforation.
8.3 OTOSCOPE
This medical device is used in conjunction with a speculum (sizes 3.5, 4.3, 5, 6, 7mm) which allows you to perform visual examinations of the hearing canal outside of the walls of the eardrum through direct analysis of otoscopic photos or digital video caught by the device. In this configuration the device offers doctors a valuable method for improving the accuracy of the visual analysis. Mic-Fi otoscope requires the deployment of a speculum CE marked disposable. Put the speculum on the device before capturing pictures or video (simple joint) and ask the patient to take a comfortable position and to remain immobile for the duration of the visit
Connect the device to the computer/tablet or smartphone Launch the Mic-Fi software (previously installed) See paragraph 9 Place the device on the part to be observed Turn the dial to get a clear picture Choose the “capture” button to get a photo or video Move the device from the patient Tell the patient is able to move agai
18
Page 19
The device is intended for external use only. Visual surveys of the ear and the hearing canal must be conducted exclusively by using a speculum (CE marked)
The part of the device that needs to come into contact with the patient is the speculum that must be disinfected after each use. If other parts of the Mic-Fi device should come into contact with the patient, it is sufficient to disinfect them. The Mic-Fi device is built with biocompatible materials. FEDMEDSP3.5,4.3,5.0,6.0 and 7.0 are disposable speculum.
The size that can be viewed from Mic-Fi SW and App are indicative and must not to affect to the final judgment borne by the doctor.
The device is intended for external use only. Do not enter the device in the oral cavity
Make sure that the holder has been regularly cleaned and disinfected before use
8.4 DENTALSCOPE
In the dental field the device with the appropriate accessories holder for dental mirror may be used to perform visual examinations of the oral cavity, through direct analysis in real time on the monitor or later through the analysis of the video and images captured and stored via software. In this configuration the device facilitates the dentist during the inspection of the oral cavity allowing a better view and a better working posture and ergonomics. Put the mirror into the holder device and ask the patient to take a comfortable position and to remain immobile for the duration of the visit
Connect the device to the computer/tablet or smartphone Launch the Mic-Fi software (previously installed) ( See paragraph 9) Place the device on the part to be observed Turn the dial to get a clear picture Choose the “capture” button to get a photo or video Move the device from the patient Tell the patient is able to move again
19
Page 20
Do not to use the device simultaneously with the cleaning and dental descaling
The size that can be viewed from Mic-Fi SW and App are indicative and must not to affect to the final judgment borne by the doctor.
The part of the device that must come into contact with the patient consists of the dental mirror (accessory not included in the tools supplied by Italeco Srl) that if reusable, must necessarily be sterilized after each use according to practice. If other parts of the Mic-Fi device come into contact with the patient, disinfect the above. The Mic-Fi device is built with biocompatible materials.
The size that can be viewed from Mic-Fi SW and App are indicative and must not to affect to the final judgment borne by the doctor.
8.5 CAMERA FOR OPTICAL DEVICES
Connect the Mic-Fi medical device to your optical instrument. The connection can be direct (through the step C) or by C-coupler. Visit www.mic-fi.com for viewing various accessories.
Connect the device to the computer/tablet or smartphone. Launch the Mic-Fi software (previously installed) See paragraph 9. Place the device on the part to be observed. Choose the “capture” button to get a photo or video.
The part of the device that must come into contact with the patient consists of the optical instrument connected to the Camera (Ex.: endoscope, microscope, rectoscope, etc.) which must necessarily be sterilized after each use according to practice. If other parts of the Mic-Fi device come into contact with the patient, disinfect the above. The Mic-Fi device is built with biocompatible materials.
9. INSTALLATION AND USE OF THE SOFTWARE AND APP’s
9.1 INFORMATION
Mic-Fi software and applications have been developed for use in conjunction with Mic-Fi Microscope and Digital Camera Wifi-USB devices. It is recommended to keep update the electronic devices (PC/tablet/smartphone) on which they have been installed and to perform the appropriate updates in order to maintain unchanged performance. In the face of new operating systems that could lead to major changes, contact support to verify compatibility with Mic-Fi software and apps.
20
Page 21
9.2 WARNING!
Before disinstalling Mic-Fi Softwares and Apps please contact our technical support in order to recuperate pictures and video.
assistance@mic-fi.com
9.3 USE MIC-FI SOFTWARE AND APPS
1. Install the Mic-Fi App and Software: a) For iPad: Search “Mic-Fi Medical” in App Store to download and install it. b) For iPhone: Search “Mic-Fi” in App Store to download and install it. c) For Smart-Phone and Tablet Android: Search “Mic-Fi MED Plus” in Google Play to
download and install it.
d) For Windows and Mac PC: install theMic-Fi Med.exe” software (per Windows) o “Mic-
Fi.dmg” (MAC OS).
Software included per FEDMEDMICFIPSWD, FEDMEDMICFIP5MSWD, FEDMEDMICFIUVWPSWD devices:
Windows and Mac PC: install Mic-Fi Derma.exe” (per Windows) or Mic-Fi Derma.dmg
(MAC OS).
Software included per FEDMEDMICFISTDSWC, FEDMEDMICFIP5MSWC, FEDMEDMICFIT6PSWD devices:
Windows e Mac PC: install “Mic-Fi Capillaries.exe” (per Windows) or “Mic-Fi
Capillaries.dmg (MAC OS).
9.3.1 Modalità Wi-Fi
1. Press for 2 seconds, wait for ~20 seconds until the Wi-Fi indicator and the illumination LEDs light, then the Wi-Fi signal will be transmitting correctly.
2. Check the wireless connection on your Phone/Tablet or PC and a list of available networks will
appear. Select the SSID “Wi-Fi2SCOPE_xxxxxx” network from the list to connect; the network “Wi-Fi2SCOPE_xxxxxx” requires a password: enter the password “12345678” and click OK.
3. Once connected, please wait for about 5 seconds to open the “Mic-Fi” application to see your images. If there is no video display, please close and restart the “Mic-Fi” app again. When the
Wi-Fi indicator LED flashes the video stream signal should be displayed in the app. Video streaming can be broadcasted to many devices, but the frame rate of live video will decrease as the number of receiving devices increases.
9.3.2 Modalità USB
Connect the mini USB cable to the device and the USB to the PC without turning on the device. Open the Mic-Fi application or software and select the "USB" icon
21
Page 22
9.3.3 Simultaneous use of the device with others Wi-Fi net
1. To access the internet by Wi-Fi
connection at the same time the Wi-Fi microscope is working, open an internet browser, input
http://10.10.1.1 and click “enter”. A
pop-up window will require you to enter a username and password, please enter: admin and admin, and then this window should appear:
2. Click in the window to pop-up
the screen:
3. Select “Wireless AP Client Mode”,
click to pop-up the screen:
4. Click till your desired
network SSID appears, and click it to pop-up the screen:
1. Open an internet browser, input
http://10.10.1.1 and click “enter”. A
pop-up window will require you to enter a username and password, please enter: admin and admin, and then this window should appear:
2. Click in the window to pop-up the
You can use the Wi-Fi network of the microscope and that of the PC / SmartPhone / Tablet at the same time
9.3.3.1 Wi-Fi Security Setup
To change the Wi-Fi access password:
22
Page 23
screen:
3. Click “Network Security Settings” to
pop-up the screen on the right.
4. Enter your desired Network Name
(SSID) and password in “WPA-PSK Key” and “Re-Enter”, and click “Next”. The
setting is done, now you can re­connect the new SSID Wi-Fi by entering the new password.
9.4 ”MIC-FI MEDICAL” ON IPAD
Click on
Click on to insert a new
patient.
1 Search Mic-Fi Medical” app on Apple store and install it in your android device. 2 Mic-Fi Medical icon will appear on your device. 3 Follow the instructions given in paragraph 9.3.1 of this manual. 4 Open the Mic-Fi Medical app.
9.4.1 Insert a new patient
23
Page 24
Filling out all the fields and click
on to confirm .
9.4.2 Analysis
Select a patient.
Click on . Click on “Analysis Patient” to start
the analysis.
Move the magnification black
wheel of your device to obtain the clear image and click on
to capture photo.
Video capture
Snaposhot On/Off.
Setting: resolution, Wi-Fi
internet
Return to the patients list.
Historic images
24
Page 25
9.4.3 Historical
Click on Click on the record saved.
Select a picture to see it more big.
Enlarge the image.
Select to export the images desired.
Select to delete the images desired
Select to write a comment
and to save
Select to export the report
25
Page 26
9.5 “MIC-FI MED PLUS” ON ANDROID TABLET /PHONE
9.5.1 Add a patient
Click on to start viewing.
No saving will be possible.
Click on
Click on to insert a new
patient.
Filling out all the fields and click
on to confirm
1 Search Mic-Fi MED Plus” app on Google Play and install it in your android device. 2 Mic-Fi MED Plus icon will appear on your device. 3 Follow the instructions given in paragraph 9.3.1 of this manual. 4 Open the Mic-Fi MED Plus app.
26
Page 27
Click on the patient.
Click on to start the
analysis
P.S. Long click to edit
patient data
9.5.2 Analysis
Move the magnification black
wheel of your device to obtain the clear. image and click on
Enable snapshot.
Return to the patient
folder.
Folder of the pictures
captured.
Setting
Data recovery if updated to Android 11 or higher
Restore the microscope brightness and contrast to the default value
Change the video settings; then press "Camera Settings"
27
Page 28
9.5.3 Historical
The images are cataloged by date.
Click on the image and then on
one of the two panels to enlarge it and put it in comparison with another image.
Click on to write the
report.
Click on to save the report.
Click on from this desk to
see all the patient reports.
9.5.4 Report
28
Page 29
9.6 “MIC-FI MED” WI-FI AND USB PER WINDOWS
: Zoom Out: zoom out the view window. : Normal: Normal the actual size. : Zoom In: zoom in the view window.
: Undo: undo the last action.
: Delete: delete all measurement. : Add Text.
Install the software Mic-Fi Med.exe for Windows. Double-click on Mic-Fi icon after completing installation to pop-up the below screen:
9.6.1 Tool Bar
USB mode: Connect USB cable to microscope and USB port of your PC and click on ; please
do NOT turn On the microscope.
Wi-Fi mode: Connect to Wi-Fi your microscope as indicated in the paragraph 9.3.1 of this
manual and click on .
: Take a picture.
: Record a video: you can click to start and stop recording a video.
29
Page 30
: Line Measurement: The line
measurement allows you to measure the linear distance between two points. Simple left-click and drag to the desired length, and click again to finish.
: Polygon : Diameter of Circle
: Three Points Circle
: Gridlines: you can add gridlines to the live video and images. The pitch will match the magnification inputted.
: Circle Grid: you can add circle grid to the live video and images. The pitch will match the magnification inputted.
: Crosshair
: Scale Crosshair
: Negative.
: Red monochrome image.
: Green monochrome image.
: Blue monochrome image.
The dimensions that you can view from Mic-Fi software and App are purely indicative and therefore should not in any way affect the final judgment at the doctor’s expense.
Before making a measurement, calibrate.
Calibrate the device at each magnification.
9.6.2 Calibration
Select a known measurement standard as the to-be-observed object, such as a ruler, to be used for
calibration. Bring photo of the selected ruler into focus by turning the “Focus & Magnification adjustment wheel”, determine the magnification by taking a reading from the scale on the focus
dial as below:
1. In the live video window select the calibration drop down menu.
2. Select “New Calibration”.
30
Page 31
3. A small window will pop-up, please enter profile name magnification and the actual length.
4. Drag the mouse to draw a line which has actual length.
5. Press “OK” button to finish calibration.
Now you can mark measurements using the same focus. (Please, do not adjust the “Focus &
Magnification wheel!”). The measurement data can be saved on the photo image.
9.6.3 File.
New Folder: Display the “New Folder” window that allow you to create a new folder to save
pictures/videos.
Folder Manager: Display the “Folder Manager” window, which allows you create, open and delete the folders.
Folder Properties: Display the “Folder Property” window. Allow you to see the properties of the
current save folder and also allow changing the name and edit notes about the current save folder.
31
Page 32
Exit: Exits Mic-Fi software.
9.6.4 Setting
Toolbar and Docking windows: Check to show the tool bar. Uncheck to hide the tool bar Status bar: Check to show the status bar. Uncheck to hide the status bar. Language: The language selection allows you to select the Mic-Fi interface language. Picture Format: Change the default picture format from BMP to JPG
IF SOFTWARE REPORT ERRORS OR INTERRUPTS, PLEASE RESTORE SETTINGS.
Measure Setting Click to pop-up the measure setting: (If the "Line width" is 2 or 3 Pixel, the "Pen Type" must be Solid-Line).
32
Page 33
Video Setting
Select the both check boxes of “Private networks” and “Public networks”, and then click “Allow access” to allow “Snapshot” program
communicate on these networks, or the button
snapshot doesn’t work. If miss the Allow operation, please click “Control panel-
>Windows firewall->Restore defaults”, and restart Mic-Fi and “Snapshot Enable” to reset it.
Restore Settings: Restore default settings.
9.6.5 Snapshot Enable
Press "Settings" and then "Snapshot" or Snapshot One for Wi-Fi mode
Warm Tips: If the Windows Security Alert appears when enable snapshot (Windows 8 as example)
as below window:
Select more photos
CTRL + Click tasto SX del mouse
33
Page 34
9.7 MIC-FI WI-FI AND USB FOR MAC
9.7.1 Installation
Install the “Mic-Fixxx.dmg” software.
Copy the file on the “desktop” and open with “DiskImageMounter”
Drag the “Mic-Fi” software to the “Applications” folder to compete installation.
P.S When installing a Mac App, plug-ins and installation packages from outside the Mac App Store, macos may require software authentication. Before opening the downloaded software go to System Preferences > Security and Privacy, in the General tab and give consent to the use of the App.
After the installation, double click on this icon to open the SW.
34
Page 35
9.7.2 Mode of use
Connect USB cable to microscope and USB port of your PC and click on Connect to Wi-Fi your microscope as indicated in the point 9.3.1 of this manual and click on
1) Take picture
2) Take/Stop Video
1) Undo
2) Delete
3) Redo
1) Line
2) Continuous line
3) Point to line
4) Radius Circle
5) Diameter Circle
6) Three Points Circle
7) Three Points Angle
8) Four Points Angle
9) Polygon
10) Add Text
1) Gridlines
2) GridCircles
3) Crosshair
4) Scale lines
1) Calibration
1) Folders Image/video
35
Page 36
9.7.3 Setting
9.7.4 Snapshot
P.s. The snapshot works only in Wi-Fi mode.
36
Page 37
9.7.5 Calibration
The dimensions that you can view from Mic-Fi software and App are purely indicative and therefore should not in any way affect the final judgment at the doctor’s expense.
Before making a measurement, calibrate.
Calibrate the device at each magnification.
Select a known measurement standard as the to-be-observed object, such as a ruler, to be used for calibration. Bring photo of the selected ruler into focus by turning the “Focus & Magnification
adjustment wheel”, determine the magnification by taking a reading from the scale on the focus
dial as below(The default magnification is 60):
1) Click on
1) Fill the fields and click on OK
2) Drag the mouse to draw a line which has actual length.
3) Select the calibration from this menu
Now the microscope is calibrate and you can mark measurements using the same focus. (Please, do not adjust the “Focus & Magnification wheel!”).
37
Page 38
9.8 MIC-FI DERMA PER WINDOWS
Mic-Fi Derma is compatible with Windows 7,10
9.8.1 Installation
Install Mic-Fi Derma software Click “Yes” Click "Next", and will start the InstallShield wizard (on some systems you might see
"Open”).
Read the license agreement for the user. If you agree, press "Yes" to continue, or "No" to
stop the installation.
Fill the field and copy the code for installation (MIC xxxx xxx xxx); see the cover of the CD. Click "Next" Click "Next" Click "Install" Click"Finish" After installation icon will appear software Mic Derma-Fi on your desktop. Connect the device in the preferred mode (see paragraph 9.3.1 or 9.3.2)
Click on the DERMA Mic-Fi icon with the right mouse button: Properties > Compatibility > select: Run this program as administrator > OK
Open The Mic-Fi Derma Software
Select how to use (USB or Wi-Fi)
Select the language
Press: Patient
38
Page 39
9.8.2 Patient management
Click on the icon only if you
want to see without capturing photos or videos.
Press the icon to add a patient;
after filling out all the fields press the icon to confirm.
Select a patient; Press the icon
to delete the patient.
Select a patient; Press the icon
to change the patient data
Enter first letter of patient’s
name; Press the icon to search
Press the icon to see a list of all
patients
9.8.3 ANALYSIS
Select Select the points of interest
Press the icon to delete the
last point
Press the icon to delete all
points
Press the icon to rotate the
image
39
Page 40
Select the points of interest.
Click on the icon to start the
Dermatoscope.
Click on the icon to take a picture.
Click on the icon to take the
video.
USB MODE: Click on the icon to
enable the Dermatoscope snapshot to capture the photos.
WiFi MODE: Click on the icon and
select “Wifi Snapshot One” and wait
about 30 seconds to enable the snapshot.
Click on the icon to set the video.
9.8.4 Calibration
Calibrate the device at every magnification
The size that can be viewed from Mic-Fi SW and App are indicative and must not to affect to the final judgment borne by the doctor.
See paragraph 9.8.6
1. Select a known standard measurement. Use
for example the graduated screen included in the package, or a ruler.
2. Focus the image by turning the adjustment wheel
3. Read the magnification on the setup dial in focus (The default magnification is 60)
4. Take the picture.
5. Select the desired image and press the right mouse button: select “Measure”
40
Page 41
6. Press and select: “New Calibration”
7. Fill mane and magnification in the field
8. Select the first point with a click and drag the line until you get the desired distance.
8. Enter the distance value and press "OK"
From this moment the SW setted the calibration on the basis of the magnification and in the left box will appear the name of the calibration.
Now you can mark measurements using the same focus. (Please, do not adjust the “Focus & Magnification wheel!”). Now is possible to measure
9.8.5 Operations available on the image and save
Select a captured image and press the right mouse button and select _ Measure" to access the following functions:
Calibration
Line
Allows you to view the linear distance between two points. Simply click and drag to the desired length, and click again to finish.
Continued Line
Allows you to view the distance of continuous lines and the total. Click and drag the lines. After double clicking.
Polygon
Allows you to view the area of the polygon. Click and drag.
Circle from the diameter
Allows you to view the area, circumference and diameter. Click and drag.
Rectangle
Allows you to view the area of the rectangle. Click and drag.
Text
Allows you to enter text.
41
Page 42
Once the measurement has been carried out, press the right mouse button in the middle of the image to save or delete the measurements taken.
9.8.6 “Historical”: Image historical
Press the icon to view the
history of images.
Select the desired point to view the history of the captured photos in the selected point
Click on the picture you want and then on larger panels to enlarge it and to put it in comparison with other photos
42
Page 43
9.8.7 REPORT
Press to access for the preparation of the document to be issued for the current patient
You can enter up to 4
reference photo and write a comment
Press to save the
document
Press to print the
document
9.8.8 All Report
Press to view the entire list of saved reports
Select the desired report to open it
43
Page 44
9.9 “MIC-FI DERMA” PER MAC
Mic-Fi Derma Sw is compatible with MAC 10.8 and following
9.9.1 Installation
Install “Mic-Fi Derma (Vxx).pkg”. Copy on the “DESK” and open with “Installer”
Proceed as follows:
Open Continue Install Insert your password
The software will be automatically installed on your MAC. If the pop-up window appears on Security & Privacy proceed to unlock. Connect the device in the preferred mode (see paragraph 9.3.1 or 9.3.2). Open Mic-Fi Derma Software.
Click on “Join”.
Close the windows and re-start the Mic-Fi Derma software.
Fill all the fields and click on on
“Submit”.
New account is now created.
44
Page 45
Fill the fields with the details of the Account created and click
on “Login”.
PS. Did you forget the password? Click on “Join” and insert
your mail.
9.9.2 Add patient
Click on the icon
to add a patient; after filling out all the fields click on the icon to confirm.
Select a patient;
click on the icon to delete the patient.
Select a patient; click on the icon to change the patient data and after filling out all the fields click on the icon to confirm.
Enter first letter of
patient’s name; click on the icon to search.
Click on the
icon to select the points
45
Page 46
of the analysis.
9.9.3 Select Points
Select the points of interest.
Click on the icon to rotate the
image.
Click on the icon to delete the
last point.
Click on the icon to delete all
points.
9.9.4 Analysis
Click on “Analyze” icon. Select a point.
For more information about use mode
USB or Wi-Fi, please refer to the manual (Paragraph 9.3.1 or 9.3.2)
Connect the USB cable and click on.
Turn on the dermatoscope, connect to (WIFI2SCOPExxxx) Wi-Fi and click on.
Click to capture a photo.
46
Page 47
Click to start/stop a video.
Undo – Delete– Redo the last operation (ex. line, text)
Click Photo to view the photo captured;
Click Video to view the video captured; Click Setting to change the video setting of dermatoscope. In the setting is possible to change the resolution. In Wi-Fi mode the resolution availabe is only in Mjpeg, so the options are n. 4, 5, 6
9.9.5 Calibration
Calibrate the device at every magnification.
The size that can be viewed from Mic-Fi SW and App are indicative and must not to affect to the final judgment borne by the doctor.
1. Select a known standard measurement.
Use for example the graduated screen included in the package, or a ruler.
2. Focus the image by turning the
adjustment wheel.
3. Take the picture.
4. Select the picture and click on the right
key of the mouse
5. Select “Measure”.
6. Click on
7. Fill the fields (read the magnification on
the setup dial in focus).
8. Enter the distance value and click on "OK".
9. Select with a click on the first point and
47
Page 48
drag the line until you get the desired distance.
Calibration performed. You will see the name of the calibration in the right box and you will can make the measurement.
If the new object to be analyzed is not in focus, only adjust the distance between the microscope and the object until you get the image in focus without touching the magnification knob, otherwise the calibration data will be invalid.
: Calibration folder
9.9.6 Operation available
Double click on the photo captured to enlarge it. Select one photo and click on the right key mouse to view popup menu.
Line
Allows you to view the linear distance between two points. Simply Click on and drag to the desired length, and Click on again to finish.
Rectangle
Allows you to view the area of the rettangle. Click on and drag.
Continuos line
Allows you to view the distance of continuous lines and the total. Click on and drag the lines. After, double click.
Polygon
Allows you to view the area of the polygon. Click on and drag.
48
Page 49
Circle
Allows you to view the area, circumference and radius. Click on and drag.
Text
Allows you to enter text.
Insert viewfinder, grid.
PS: To save the changes click on the photo captured with the right mouse button “Save as”: overwrite the original image.
9.9.7 “Historical”: Image historical
Click on the icon
to view the history of images.
Click on icon “1”, for
example, to view the history of the photos captured at the selected point.
• Click on the desired photo and then on the larger boxes to enlarge it and to compare it with other photos.
9.9.8 Reports
Click on to access for the preparation of the document to be issued for the current patient.
You can enter up to two reference
photo and write a comment.
Click on to save the
document.
Click on to print the
document.
49
Page 50
9.9.9 All Reports
Click on to view the entire list of saved reports.
Select the desired report to view it.
9.10 MIC-FI CAPILLARIES PER WINDOWS
9.10.1 Installation
Mic-Fi Capillaries is compatible with Windows 7,10.
1. Install the Mic-Fi Capillaries software for windows.
2. Click on "Run".
3. Input the PSW “FEDxxxxxxxx” (Read the label on the cover of the CD).
4. Click on"Next".
5. Fill User Name and Company Name field.
6. Click on"Next".
7. Click on “Complete” for complete installation.
8. Click on"Next".
9.Click on"Install" ("Install").
10.Click on"finish" ("End").
11. After installation icon Mic-Fi Capillaries software will appear on your desktop.
Connect the device in the preferred mode (see paragraph 9.3.1 or 9.3.2) Open Mic-Fi Software Capillaries Select Language
50
Page 51
Click on “Register”.
Close the windows and re-start the Mic-Fi Capillaries software
Fill all the fields and click on “ok”. New account is now created.
Fill the fields with the details of the Account created and click
on “Login”.
PS. Did you forget your password? Click on “Remember PSW”.
51
Page 52
9.10.2 Add patient
Click on the icon only if
you want to see without capturing photos or videos.
Click on the icon to add a
patient; after filling out all the fields click on the icon to confirm.
Select a patient; click on
the icon to delete the patient.
Select a patient; click on the icon to change the patient data. After click on the icon to confirm.
Enter first letter of
patient’s name; click on the icon to search.
Click on the icon to see a
list of all patients.
Click on “SELECT POINTS
52
Page 53
9.10.3 Select Points
Select the points of interest.
Click on the icon to delete the
last point.
Click on the icon to delete all
points.
Maximum 8 points can be selected. Press "START ANALYSIS".
9.10.4 Analysis
Select the points of interest.
Click on the icon to start the
Capillaroscope.
Click on the icon to take a
picture.
Click on the icon to take the
video.
USB MODE: Click on the icon
to enable the Capillaroscope snapshot to capture the photos.
WiFi MODE: Click on the icon and
select “ Wifi Snapshot One” and
wait about 30 seconds to enable the snapshot.
Click on the icon to set the
video.
53
Page 54
9.10.5 Calibration procedure
Calibrate the device at every magnification.
The size that can be viewed from Mic-Fi SW and App are indicative and must not to affect to the final judgment borne by the doctor.
1. Select a known standard measurement. Use
for example the graduated ruler included in the package.
2. Focus the image by turning the adjustment
wheel.
3. Select the picture and click on the right key of
the mouse
4. Select “Measure”.
5. Click on ; See the paragraph 9.10.6
6. Fill the fields (read the magnification on the
setup dial in focus).
7. Select with a click on the first point and drag
the line until you get the desired distance. A window will open.
8. Enter the distance value and click on "OK".
Calibration performed.
If the new object to be analyzed is not in focus, only adjust the distance between the microscope and the object until you get the image in focus without touching the magnification knob, otherwise the calibration data will be invalid.
In the "Calibration Folder" the calibrations will be saved.
54
Page 55
9.10.6 Operation available
Double click on the photo captured to enlarge it. Select one photo and click on the right key mouse to view popup menu
Line
Allows you to view the linear distance between two points. Simply click on and drag to the desired length, and Click on again to finish.
Rectangle
Allows you to view the area of the rectangle. Click on and drag.
Continued Line
Allows you to view the distance of continuous lines and the total. Click on and drag the lines. After, double click.
Polygon
Allows you to view the area of the polygon. Click on and drag.
Circle from the diameter
Allows you to view the area, circumference and radius. Click on and drag.
Text
Allows you to enter text.
To save the changes made, click on the image with the right mouse button: overwrite the original image.
55
Page 56
9.10.7 “Historical”: Image historical
Select the affected point to view the
history of the captured photos in the selected point.
Click on the picture you want and then
on larger panels to enlarge it and to put it in comparison with other photos.
9.10.8 Reports
Click on “Medical Report”to access for the preparation of the document to be issued for the current patient.
You can enter up to 4
reference photo and write a comment.
Click on to save
the document.
Click on to print
the document.
9.10.9 All Reports
Click on “All Reports” to view the entire list of saved reports of the patient.
Select the desired report to view the contents.
56
Page 57
9.11 MIC-FI CAPILLARIES PER MAC
9.11.1 Installation
Insert the CD and install “Mic-Fi Capillaries.dmg” . Copy on the “DESK” and open with “DiskImageMounter”
Drag " Mic-Fi Capillaries " in the "Application" folder.
P.S. If the pop-up window appears on Security & Privacy proceed to unlock. After the installation, double click on this icon to open the SW.
Connect the device in the preferred mode (see paragraph 9.3.1 or 9.3.2) Open the Software
Click on “Join”.
57
Page 58
Close the windows and re-start the Mic-Fi Capillaries software.
Fill all the fields and click on on
“Submit”.
New account is now created.
Fill the fields with the details of the Account created and click on
“Login”.
PS. Did you forget the password? Click on “Join” and insert
your mail.
9.11.2 Add patient
Click on the icon to
add a patient; after filling out all the fields click on the icon to confirm.
Select a patient; click
on the icon to delete the patient.
Select a patient; click on the icon to change the patient data; after, click on the icon to confirm.
Enter first letter of
patient’s name; click on the
58
Page 59
icon to search.
Click on the icon to
select the points of the analysis.
9.11.3 Select Points
Select the points of interest.
Click on the icon to delete the
last point.
Click on the icon to delete all
points.
9.11.4 Analysis
Click on “Analyze” icon. Select a point.
Connect the USB cable and click on.
Turn on the capillaroscope, connect to (WIFI2SCOPExxxx) Wifi and click on.
Click to capture a photo.
59
Page 60
Click to start/stop a video.
Undo – Delete– Redo the last operation (ex. line, text)
Click Photo to view the photo captured;
Click Video to view the video captured; Click Setting to change the video setting of capillaroscope.
9.11.5 Calibration
Calibrate the device at every magnification.
The size that can be viewed from Mic-Fi SW and App are indicative and must not to affect to the final judgment borne by the doctor.
1. Select a known standard measurement. Use
for example the graduated screen included in the package, or a ruler.
2. Focus the image by turning the adjustment
wheel.
3. Take the picture.
4. Select the picture and click on the right key of
the mouse
5. Select “Measure”.
6. Click on
7. Fill the fields (read the magnification on the
setup dial in focus).
8. Enter the distance value and click on "OK".
9. Select with a click on the first point and drag
the line until you get the desired distance.
Calibration performed. You will see the name of the calibration in the right box and you will can make the measurement.
If the new object to be analyzed is not in focus, only adjust the distance between the microscope and the object until you get the image in focus without touching the magnification knob, otherwise the calibration data will be invalid.
: Calibration folder.
60
Page 61
9.11.6 Operation available
Double click on the photo captured to enlarge it. Select one photo and click on the right key mouse to view popup menu.
Linea
Allows you to view the linear distance between two points. Simply Click on and drag to the desired length, and Click on again to finish.
Rectangle
Allows you to view the area of the rettangle. Click on and drag.
Continued line
Allows you to view the distance of continuous lines and the total. Click on and drag the lines. After, double click.
Polygon
Allows you to view the area of the polygon. Click on and drag.
Circle
Allows you to view the area, circumference and radius. Click on and drag.
Text
Allows you to enter text.
Insert viewfinder, grid.
Calibration folder
PS: To save the changes made press on the captured photo with the right mouse button and click "Save as" to save overwriting the image
61
Page 62
9.11.7 “Historical”
Click on the icon to
view the history of images.
Click on icon “1”, for
example, to view the history of the photos captured at the selected point.
• Click on the desired photo
and then on the larger boxes to enlarge it and to compare it with other photos.
9.11.8 Reports
Click on to access for the preparation of the document to be issued for the current patient.
You can enter up to two reference
photo and write a comment.
Click on to save the
document.
Click on to print the
document.
62
Page 63
9.11.9 All Reports
Click on to view the entire list of saved reports.
Select the desired report to view it.
10. BATTERY CHARGER
1. Connect the DC Adapter (plug the AC adapter into an AC outlet) or PC’s USB cable, it charges
whether the power switch is on Wi-Fi or OFF/USB. We DO NOT recommend charging Wi-Fi
Microscope while in use.
2. Battery LED ( Orange LED or Red LED ): Red LED when the battery level is low; Orange LED
blinking when the battery is charging
3. The charge LED : Orange LED fix when the battery is at fully charged.
WARNING : IMPAIRMENT OF IMAGE OR SLOW VIDEO ARE THE RESULT OF A LOW BATTERY. BEFORE FIRST USE OF
THE MICROSCOPE, PLEASE CHARGE THE BATTERY FOR AT LEAST 12 HOURS.
63
Page 64
11. CLEANING THE DEVICE
The cleaning and disinfection operations should take place after each use.
DO NOT USE the device for at least 15 minutes after its cleaning operation.
Avoid touching the surface of the lenses When cleaning the lens, if necessary, use dry compressed air to remove dirt.
Do not use abrasive materials to clean the lens, it could cause permanent damage to it.
It is very important to avoid touching the lens of the device because it is very delicate. The lenses should be cleaned regularly, as particles of dust on them could have a negative impact on image quality. For this reason, it is recommended to protect the device with its protective casing when not in use. Avoid using any other objects or liquids that could easily scratch or damage the surface of the lens or the sensor. To clean the body of the device use a slightly dampened soft cloth. The product is designed to last 5 years average, and is resistant to most chemicals products except ketones MEK, solvents and substances used to remove the smalt.
64
Page 65
12. DISPOSAL
For the correct disposal of devices Microscope and digital camera Wifi-USB Mic­Fi it is necessary to follow the provisions provided by the obligations of DL n.49 of 14 March 2014, Implementation of Directive 2012/19/EU on waste electrical and electronic equipment (WEEE), as amended by Directive 2018/849/EU, amending Directives 2000/53/EC on end-of-life vehicles, 2006/66/EC on batteries and accumulators and waste batteries and accumulators and 2012/19/EU on waste electrical and electronic equipment. The user must, therefore, give the end-of-life equipment to the appropriate electronic and electrotechnical waste collection centres, or return it to the retailer when purchasing a new equipment of equivalent type, one by one. adequate separate collection for the subsequent start-up of discarded equipment for recycling, treatment and disposal which is environmentally compatible, helps to avoid possible adverse effects on the environment and health and promotes the reuse and/or recycling of the materials of which the equipment is composed. In any case, therefore, the device must not be dispersed into the environment after use. The misuse of the product by the user involves the application of administrative penalties provided for by the current legislation.
The disposal of batteries and accumulators is regulated by Directive 2013/56/EU repealing Directive 2006/66/EC, as amended by Directive 2018/849/EU, amending Directive 2000/53/EC on end-of-life vehicles, 2006/66/EC on batteries and accumulators and waste batteries and accumulators and 2012/19/EU on waste electrical and electronic equipment. ATTENTION! This directive stipulates that batteries and accumulators must be treated as special waste, so that they must not be disposed of in the household or in the environment after use, but must be disposed of free of charge at the appropriate points of sale or collection. The obligation to collect spent batteries free of charge applies to all points of sale or collection and is not conditional on the purchase of new batteries. If the symbol shows a chemical symbol, in compliance with the above mentioned Directive, this symbol indicates the presence of a heavy metal (Hg = Mercury, Cd = Cadmium, Pb = Lead) in the battery with a concentration level above an applicable threshold specified in the Directive. The misuse of the batteries by the user involves the application of administrative penalties referred to D. Lgs. n. 22/1997 (art. 50 and following of ... D.Lgs. 22/1997).
65
Page 66
13. WARRANTY
ITALECO SRL
Via Pavia 105/i, 10098 Rivoli (To), Italy
Tel: +39-011-5690297
Fax: +39-011-5690298
http://www.mic-fi.com
Congratulations for purchasing our product. This product HAS BEEN MANUFACTURED TO THE HIGHEST QUALITY AND STANDARDS. Italeco guarantees this product against any defects due to materials or workmanship for a period of 2 years from the date of original purchase. To take advantage of the guarantee, you must retain your purchase invoice. If the product is found to be defective within the warranty period it should be returned to the selling agent with proof of purchase. The warranty is null and void if the damage is resulting from carelessness or misuse. No compensation will be given by Italeco and its distributors if the product is returned during the warranty period. In addition, any replacement or repair carried out during the warranty period, will not prolong the duration of the guarantee.
The warranty lapses in the following cases:
The product has been tampered with or the label placed on the usb cable bearing the serial
number has been removed, damaged or rendered unreadable
The product has been damaged or has been misused The product has been opened or repaired by unauthorised personnel The product has been damaged due to non compliance with the instructions in the booklet
Products considered to be defective should be returned only to the dealer from which it was purchased accompanied by the invoice. The product will be repaired or replaced, under our sole discretion.
If you need any information and / or assistance, contact Mic-Fi Technical Support:
assistance@mic-fi.com
Tel +39 (0)11 5690295
66