Electromagnetic Compatibility Information
The device satises the EMC requirements of the international standard IEC 60601-1-2. The requirements are satised under the conditions described in the table below. The device is an electrical medical product and is subject to special precautionary measures with regard to EMC which must be pub­lished in the instructions for use. Portable and mobile HF communications equipment can affect the device. Use of the unit in conjunction with non-ap­proved accessories can affect the device negatively and alter the electro­magnetic compatibility. The device should not be used directly adjacent to or between other electrical equipment.
Table 1
Guidance and declaration of manufacturer-electromagnetic emis­sions
The device is intended for use in the electromagnetic environment speci­ed below. The customer or the user of the device should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment –
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions IEC 61000-3-2
Voltage uctuations/ icker emissions IEC 61000-3-3
Table 2
Guidance and declaration of manufacturer-electromagnetic immu­nity
The device is intended for use in the electromagnetic environment speci­ed below. The customer or the user of the device should assure that it is used in such an environment.
Immunity test IEC 60601
Electrostatic discharge (ESD) IEC 61000-4-2
IP27
-
WEEE
Electrostatic transient/burst IEC 61000-4-4
Surge IEC 61000-4-5
Voltage dips, short interrup­tions and voltage variations on power supply
input lines
IEC 61000-4-11
Power frequency (50/60 Hz) magnetic eld IEC 61000-4-8
Group 1 The device uses RF energy only for
Class B The device is suitable for use in all
N/A
N/A
test level
± 8 kV contact ±2 kV, ±4 kV, ±8 kV, ±15 kV air
± 2 kV for power supply
lines
± 1 kV for input/output
lines ± 1 kV differ-
ential
mode ± 2 kV com­mon mode
< 5% UT (>95% dip in UT) for 0.5 cycle 40% UT (60% dip in UT) for 5 cycle 70% UT (30% dip in UT) for 25 cycle <5% UT (>95% dip in UT) for 5 secretary
30 A/m; 50Hz or 60Hz
guidance
its internal function. Therefore, its emissions are very low and are not likely to cause any inter­ference in nearby electronic equip­ment.
establishments, including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes.
Compli­ance level
± 8 kV
contact
±2 kV, ±4 kV, ±8 kV, ±15 kV air
N/A
N/A
N/A
30 A/m; 50Hz or 60Hz
Electromagnetic envi­ronment - guidance
Floors should be wood, concrete or ceramic tile. If oors are covered with synthetic material, the relative humidity should be at least 30 %.
Power frequency mag­netic elds should be at levels charactertic of a typical location in a typical commercial or hospital environment.
Table 3
Guidance and declaration of manufacturer-electromagnetic immu­nity
The device is intended for use in the electromagnetic environment speci­ed below. The customer or the user of the device should assure that it is used in such an environment.
Immunity test
Conducted RF IEC 61000-4-6
Radiated RF IEC 61000-4-3
RF
Wireless Communi­cation
Equipment IEC 61000-4-3
IEC 60601 test
level
3 Vrms 150 kHz to 80 Mhz
10 V/m 80 MHz to 2.7 Ghz
380MHz, 27V /m
450MHz, 28V /m
710MHz,745 MHZ, 780MHz 9V/m
810MHz,870 MHZ, 930MHz 28V/m
1720MHz, 1845 MHZ, 1970MHz 28V/m
2450MHz, 28V /m
5240MHz, 5500 MHZ, 5785MHz 9V/m
Compliance Electromagnetic
N/A
10 V/m
380MHz, 27V /m
450MHz, 28V /m
710MHz, 745 MHZ, 780MHz 9V/m
810MHz, 870 MHZ, 930MHz 28V/m
1720MHz, 1845 MHZ, 1970MHz 28V/m
2450MHz, 28V /m
5240MHz, 5500 MHZ, 5785MHz 9V/m
environment – guidance
Portable and mobile RF communications equipment should be used no closer to any part of the device, including cables, than the recommended
separation distance
calculated from the equation applicable to the frequency of the transmitter.
Recommended separation distance
3.5
P
d=
1
E
80 MHz to 800 MHz
7
P
d=
1
E
800 MHz to 2.7 GHz
Where P is the max­imum output power rating of the transmitter in watts (W) accord­ing to the transmitter manufacturer and d is the recommended
separation distance in
metres (m). Field strengths from xed RF transmitters, as determined by an electromagnetic site survey, a should be less than the compli­ance level in each fre­quency range. Interference may occur in the vicinity of equipment marked with the following sym­bol:
25557/DMT-4220
JOYTECH Healthcare CO., Ltd. No.365, Wuzhou Road, Yuhang Economic Development Zone, Hangzhou City, 311100 Zhejiang, China Made In China
Table 4
Recommended separation distances between portable and mobile RF communications equipment and the device
The device is intended for use in an electromagnetic environment in which radiated therefore disturbances are controlled. The customer or the user of the device can help prevent electromagnetic interference by maintain­ing a minimum distance between portable and mobile RF communications equipment (transmitters) and the device as recommended below, accord­ing to the maximum output power of the communications equipment.
Rated maximum output power of transmitter W
0.01 0.12 0.23
0.1 0.38 0.73
1 1.2 2.3
10 3.8 7.3
100 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer. NOTE1 At 80 MHz and 800 MHz, the separation distance for the higer frequency range applies. NOTE2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reection from structures, ob­jects and people.
Separation distance according to frequency of transmitter m
80MHz to 800MHz 800MHz to 2.7GHz
d=
3.5 E
P
1
d=
7
P
1
E
0197
20537 Hamburg - Germany
Importato da / Imported by / Importado por / Importado por / Importé par / Importowane przez / Εισαγωγή από / :

Gima S.p.A.
Via Marconi, 1 - 20060 Gessate (MI) Italy
55°C
IP27
-20°C
95% 1060 hPa
15% 700 hPa