ASPIRATORE CHIRURGICO HOSPI PLUS
HOSPI PLUS SUCTION ASPIRATOR
ASPIRATEUR HOSPI PLUS
CHIRURGISCHER ABSAUGER HOSPI PLUS
ASPIRADOR QUIRÚRGICO HOSPI PLUS
ASPIRATOR CHIRURGICZNY HOSPI PLUS
ASPIRATORUL CHIRURGICAL HOSPI PLUS
HOSPI PLUS SUCTION ASPIRATOR is a surgical aspirator power-fed at 230V ~ / 50Hz, to be used for suctioning body liquids (such as mucus, phlegm and blood) provided with 4 antistatic wheels, two of which with
braking device, and a pulling handle. This equipment is designed for easy transport and continuous utilization.
Thanks to these characteristics and to its functions, this device is particularly suitable for utilization in hospital
wards and operation theatres both for suctioning body liquids and for gynaecological and dermatological (liposuction) applications. It’s provided with a plastic body, with thermal and electrical isolation in compliance with
European safety standards, two complete suction tanks in polycarbonate suitable for sterilization, and a oat
valve, besides being tted with a suction regulator and a vacuum gauge on the front panel. Versions tted with
footswitch control and ux deviator are available on request. The electronic management system tted on the
front panel allows to perform suction by means of the footswitch control as well as to suction liquids in both
tanks provided without having to switch the equipment off to reconnect the second tank.
ENGLISH
GENERAL WARNING
Read instruction manual carefully before use
Only Higly qualied staff use reserved
The instrument must not be disassembled. For a technical service always contact GIMA
Keep off the reach of children or not capable people without supervision
Full containers must be handled with great care during transfer to the disposal areas, following the
local procedures and regulations
IMPORTANT SAFETY RULES
1. Check the condition of the unit before each use. The surface of the unit should carefully inspected for
visual damage. Check the mains cable and do not connect to power if damage is apparent;
2.
Before connecting the appliance always check that the electric data indicated on the data label and the type
of plug used, correspond to those of the mains electricity to witch it’s to be connected;
3.
Respect the safety regulations indicated for electrical appliances and particularly:
• Use original components and accessories provided by the manufacturer to guarantee the highest ef-
ciency and safety of the device.
• The device can be used only with the bacteriological lter.
• Never immerge the appliance into water.
• Position the device on stable and at surfaces in a way that the air inlets on the back aren’t obstructed.
• To avoid incidents, do not place the aspirator on unstable surfaces, which may cause it to accidentally fall
and lead to a malfunction and/or breakage. Should there be signs of damage to the plastic parts, which
may expose inner parts of the energised device, do not connect the plug to the electrical socket. Do
not attempt to make the device work before it has been thoroughly checked by qualied personnel and/
or the GIMA technical service department.
• Don’t use in the presence of inammable substances such as anaesthetic, oxygen or nitrous oxide.
• Don’t touch the device with wet hands and always prevent the appliance coming into contact with liquids.
• Don’t leave the appliance connected to the power supply socket when not in use.
• Don’t pull the power supply cable to disconnect the plug remove the plug from the mains socket correctly.
• Store and use the device in places protected against the weather and far from any sources of heat. After
each use, it is recommended to store the device in its own box away from dust and sunlight.
• In general, it is inadvisable to use single or multiple adapters and/or extensions. Should their use be
necessary, you must use ones that are in compliance with safety regulations, however, taking care not to
exceed the maximum power supply tolerated, which is indicated on the adapters and extensions.
4.
For repairs, exclusively contact technical service and request the use of original spare parts. Failure to
comply with the above can jeopardise the safety of the device.
5.
Use only for the purpose intended. Don’t use for anything other than the use dened by the manu-
facturer. The manufacturer will not be responsible for damage due to improper use or connection to an
electrical system not complying with current regulation.
6.
Instrument and accessory discharging must be done according to current regulations in the country of use.
7.
WARNING: Do not change this equipment without the permission of the manufacturer Gima S.p.A. None
of electric or mechanical parts have been designed to be repaired by customers or end-users. Don’t open
the device, do not mishandle the electric / mechanical parts. Always contact technical assistance.
Page 3
ENGLISH
8.
Using the device in environmental conditions different than those indicated in this manual may harm seri-
16
ously the safety and the technical characteristics of the same;
9.
The medical device is in contact with the patient by means of a disposable probe (not supplied with the
device). If this device must be used with a specic suction probe, the end user is responsible for making
sure it complies with the ISO 10993-1 rule;
10.
The product and its parts are biocompatible in accordance with the requirements of regulation EN 60601-1;
11.
Operation of the device is very simple and therefore no further explanations are required other than those
indicated in the following user manual.
12.
The medical device requires special precautions regarding electromagnetic compatibility and must be
installed and used in accordance with the information provided with the accompanying documents: the
HOSPI PLUS SUCTION ASPIRATOR device must be installed and used away from mobile and portable
RF communication devices (mobile phones, transceivers, etc.) that may interference with the said device.
Under certain failure conditions, the temperature of the casing may become hot and there may be
a risk of burns if you touch those parts. In any case, the temperatures do not exceed the limit of 105°C (ref.
Interpretation Sheet IEC 60601-1 / ISH May 2013).
The manufacturer cannot be held liable for accidental or indirect damages should the device be modied,
repaired without authorization or should any of its component be damaged due to accident or misuse. Any
minimal modication / repair on the device voids the warranty and does not guarantee the compliance with
the technical requirements provided by the MDD 93/42/EEC (and subsequent changes) and its normatives.
CONTRAINDICATIONS
• Before using the HOSPI PLUS SUCTION ASPIRATOR, consult the instructions for use: failure to read all
the instructions in this manual can be harmful for the patient.
• The device cannot be used to drain chest uids;
• The device must not be used for suction of explosive, corrosive or easily ammable liquids.
• HOSPI PLUS SUCTION ASPIRATOR is not suitable for MRI. Do not introduce the device in MRI environments.
TECHNICAL CHARACTERISTICS
TYPOLOGY (MDD 93/42/EEC)Class IIa Medical Decice
MODEL
UNI EN ISO 10079-1HIGH VACUUM / HIGH FLOW
POWER FEEDING230V ~ / 50Hz
POWER CONSUMPTION385 VA
FUSEF 1 x 4A L 250V
MAXIMUM SUCTION PRESSURE (without jar)-90kPa / -0.90 Bar / -675mmHg
To clean the device external parts always use a cotton cloth dampened with detergent. Don’t use abrasive or
solvent detergents.
ATTENTION: During cleaning make sure that liquids do not come into contact with the membrane keyboard
(only in versions with footswitch and ux deviator) and adjacent areas as this would damage the component,
with possible inltration of the liquid inside the device..
The symbol positioned in the casing near the membrane keyboard requires the reading of the user
instructions before each use.
PARTICULAR CARE SHOULD BE TAKEN TO ENSURE THAT THE INTERNAL PARTS OF THE
EQUIPMENT DO NOT GET IN TOUCH WITH LIQUIDS. NEVER USE LIQUIDS (e.g. DETERGENTS AND/
OR SANITISING SUBSTANCES) TO CLEAN THE MAIN UNIT (ESPECIALLY NEAR THE MEMBRANE
KEYBOARD) AS THEY MAY PENETRATE INSIDE THE DEVICE
During all clearing operations use protection gloves and apron (if need be, also wear a face mask and glasses)
to avoid getting in contact with contaminating substances (after each utilization cycle of the machine).
ACCESSORIES SUPPLIED
• N°2 COMPLETE ASPIRATION JAR
• CONICAL FITTING
• SILICON SET TUBES 8 mm x 14 mm
• ANTIBACTERIAL AND HYDROFOBIC FILTER
• SAFETY TRAP
• FOOTSWITCH CONTROL (for versions equipped with footswitch)
• EUROPEAN POWER SUPPLY CORD (H05VV-F - 2x0.75mm² - 2mt)
REPLACING THE ANTIBACTERIAL FILTER: The lter is made of hydrophobic material that stops the
passage of liquids into the same lter. If you suspect the lter may have been contaminated and/or got wet
or discoloured, always remove and replace the lter. If the equipment is to be used on patients with unknown
pathological conditions or should you evaluate the possibility of indirect contamination, remove and replace
the lter after each utilization. The lter is not designed for decontamination, disassembly and/or sterilization. If
you suspect the lter may have been contaminated and/or got wet or discoloured, always remove and replace
the lter. If the equipment is to be used on patients whose pathologies are known and not implying any indirect
contamination risks, we recommend to remove and replace the lter at the end of each work shift or else every
month, even if the equipment has not been used. 4000ml or 5000ml complete jar versions are available on
request. Versions tted with FLOVAC® 2000ml or 3000ml disposable collection systems (including a re-usable
rigid polycarbonate container and a disposable Liner) are also available on request.
SAFETY TRAP with 220 ml capacity to collect the liquid that could leak from the overow valve of the vessel.
This ensures additional protection of the lter and pump. The trap is completely removable and autoclavable.
Not available in versions equipped with FLOVAC disposable collection system.
WARNING: Suction tubes for insertion in the human body purchased separately from the machine should
comply with ISO10993-1 standards on material biocompatibility.
WARNING: The medical device is provided without a specic suction probe. If this device must be used with a
specic suction probe, the end user is responsible for making sure it complies with the EN 10079-1 regulation.
Aspiration jar: The mechanical resistance of the component is guaranteed up to 30 cycles of cleaning and
sterilization. Beyond this limit, the physical-chemical characteristics of the plastic material may show signs of
decay. Therefore, we recommend that you to change it.
Silicone tubes: the number of cycles of sterilization and/or cleaning is strictly linked to the employment of the
said tube. Therefore, after each cleaning cycle, it is up to the nal user to verify whether the tube is suitable
for reuse. The component must be replaced if there are visible signs of decay of the material constituting the
said component.
Page 5
ENGLISH
Conical tting: the number of cycles of sterilization and the number of cleaning cycles is strictly linked to the
employment of the said component. Therefore, after each cleaning cycle, it is up to the nal user to verify
whether the tting is suitable for reuse. The component must be replaced if there are visible signs of decay in
the material constituting the said component.
Service life of the device: More than 10000-12000 hours of operation (or 3 years) in accordance with the
standard conditions of testing and operation. Shelf life: maximum 5 years from the date of manufacture.
18
CLEANING ACCESSORIES AND INTERNAL PARTS
Before using the device, the manufacturer advises you to clean and/or sterilize the accessories. Washing and
/ or cleaning the autoclavable jar as to be carried out as follows:
• Wear protection gloves and apron (glasses and face mask if necessary) to avoid contact with contaminat-
ing substances;
• Disconnect the tank from the device and remove the said container from the support of the device.
• Separate all the parts of the cover (overow device, washer).
• Disconnect all tubes from the jar and the protection lter
• Wash each part of the container from secretions under cold running water and then clean every single
part in hot water (temperature not exceeding 60°C)
• Once again, carefully wash each single part using, if necessary, a non-abrasive brush to remove any
deposits. Rinse with hot running water and dry all parts with a soft cloth (non-abrasive). It is possible to
wash with commercial disinfectants by carefully following the instructions and dilution values supplied by
the manufacturer. After cleaning, leave the parts to dry in an open, clean environment.
• Dispose of the aspiration catheter according to that provided by local laws and regulations.
The silicone aspiration tubes and the conical tting may be carefully washed in hot water (temperature must
not exceed 60°C). After cleaning, leave the parts to dry in an open, clean environment. When cleaning is
complete, reassemble the container for liquid aspirations according to the following procedure:
• Place the overow valve into its seat in the cover (under VACUUM connector)
• Insert oating valve keeping the o-ring towards the opening of the cage
• Place the o-ring into its seat around the cover
• After completing assembling operations always make sue that cover seals perfectly to avoid vacuum
leackages or liquid exit
The jar and the cover can be autoclaved by placing the parts into the autoclave and running one sterilization
stem cycle at 121°C (1 bar relative pressure – 15 min) making sure that the jar is positioned upsidedown.
Mechanical resistance of the jar is guaranteed up to 30 cycles of sterilization and cleaning at the indicated
conditions (EN ISO 10079-1). Beyond this limit the physical-mechanical characteristics of the plastic may
decrease and replacement of the part is therefore recommended. After sterilization and cooling at environment
temperature of the parts make sure that these are not damaged and assemble the container for the aspirated
liquids. The aspiration tubes can be sterilized on autoclave using a sterilization cycle at 121°C (1 bar relative
pressure – 15 min). The conical connector can be sterilized on autoclave using a sterilization cycle at 121°C
(1 bar relative pressure – 15 min).
DO NOT WASH, STERILIZE OR PUT IN AUTOCLAVE THE ANTIBACTERIAL FILTER
Instruction for disposal Liner Flovac®:
If the device is equipped with disposable collection systems FLOVAC ® carry out the disposal of the bag as
follows: Turn off the Vacuum and remove all the tubes connected to the Liner, giving particular attention to
avoid accidental contamination. Fit the appropriate plugs to the “PATIENT” and “TANDEM” ports, pressing the
home rmly, taking care to avoid accidental contamination. Remove the liner bag from the rigid container and
transfer it to the waste disposal area, ensuring that all the openings are sealed, keeping in mind the product
is potentially infectious. This product must be disposed of in accordance with the current hospital regulations.
Page 6
19
ENGLISH
MAINTENANCE
The HOSPI PLUS SUCTION ASPIRATOR suction equipment does not need maintenance or lubrication. It is,
however, necessary to inspect the unit before each use. With regard to training, given the information contained
in the user manual and since it is easy to understand the said device, it doesn’t appear to be necessary. The
device must be checked before each use in order to detect malfunctions and / or damage caused by transport
and / or storage. Always check the integrity of the footswitch power cord. Connect cable to electrical network
and turn switch on. Close the aspiration outlet with your nger and with suction regulator in maximum vacuum
position check that the vacuum indicators reaches -90 kPa (-0.90 bar) maximum. Rotate the knob from right
to left and check the aspiration regulating control. The vacuum indicator should go down -40 kPa (-0.40 bar).
Verify that loud noises are not present, these can indicate wrong functioning. A protection fuses (F 1 x 4A L
250V) reachable from exterior and situated in the plug protects the instrument. For fuses replacing, always the
type and the range. Before changing the fuse, disconnect the plug from the power supply socket. Internally,
the device (only for devices tted with a circuit board) is protected by a fuse (F 500mA L 250V)
Fault typeCauseSolution
1. The suction unit doesn’t work
2. No aspirationJar Cap badly screwed downUnscrewed the cap, then rescrew it correctly
3. No aspirationLid seal not in its seat
4. The Vacuum power onthe
patient side is either very low
or absent
5. The oat doesn’t close
6. The oat doesn’t closeThe oat it’s covered by dirty material
7. Low suctionFoam inside the jarFill the jar to 1/3 full of ordinary water
8. No aspiration due to ow
leakage of material and uid
Faults 1 - 2 - 3 - 4 - 5 - 6 - 7 - 8
Cable is damaged
External power source failure
a) Vacuum regulator set to minimum
b) Protection lter blocked or damaged
c) Connection tubes blocked, kinked
or disconnected
d) Shut-off valve blocked or damaged
e) Pump motor damaged
If the cap has been washed, ensure
that the oat is not partially detached
Filter blocked Replace lter
None of the remedies has achieved
the desired results
Replace the cable Check the external power
source
Unscrew the cap and insert the seal properly
in its seat
a) Turn the vacuum regulator clockwise and
check the value of the vacuum on the gauge
b) Replace the lter
c) Replace or reconnect the tubes, check the
jar connections
d) Empty the jar, or disconnect the tube from
the jar andunblock the shut-off valve. The
unit twill only work in theupright position
e) Refer to authorised service personnel
Insert the oat into it’s place
Unscrewed the cap, leave the and put in on
autoclave
Contact the seller or GIMA After-sales
Assistance Service
If the overll security system it’s activated, don’t proceede with the liquid aspiration. If the overll security
system doesn’t work there are two cases:
1° case – If the overll security system doesn’t work the aspiration will be stopped by the bacteriological lter
who avoid the liquid penetration inside the device.
2° case – If both the security system doesn’t work, there is the possibility that liquid comes inside the device,
in this case return the device to GIMA technical service.
Gima S.p.A. will provide upon request electric diagrams, components list, descriptions, setting instructions and
any other information that can help the technical assistance staff for product repair.
BEFORE EVERY CHECKING OPERATION, IN CASE OF ANOMALIES OR BAD FUNCTIONING,
PLEASE CONTACT GIMA TECHNICAL SERVICE. THE MANUTACTURER DOES NOT GIVE GUARANTEE
IF INSTRUMENT, AFTER THE TECHNICAL SERVICE CHECKING, APPEARS TO BE TAMPERED.
Page 7
ENGLISH
20
INSTRUCTIONS
• The working position must be such as to allow one to reach the control panel and to have a good view of
the empty indicator, the jar and the antibacterial lter.
• If the device is to be transported from one place to another, to prevent the liquid collection jar from falling
and consequently theliquid from spilling, removing the jar from the device is recommended.
WARNING: For proper use, place the aspirator on a at, stable surface in order to have the full volume of use of
the jar and better efciency of the overow device. The vacuum jar, during use, must be used in vertical mode,
to prevent the action of the backow valve. If this protection is triggered, turn the device off and disconnect the
pipe connected to the vacuum jar (indicated with the word VACUUM) on its cover.
• Connect the short silicone tube to the antibacterial lter connector (check the lter assembly photo) while
the other end must be connected to the safety trap “IN” nozzle using a short silicone tube.
SAFETY TRAP: The two connections on the sides of the bar can be used to insert a safety trap in the BASIC
and FS versions and two traps in the FULL version. The safety trap is an additional protection for the overow
valve of the vessel. In the event that the liquid goes beyond the overow valve during the suction process, the
trap collects the liquid thus protecting the antibacterial lter and the internal motor.
• Connect the remaining short silicone tube to the safety trap “OUT” nozzle, while the other end must be
connected to the vessel cover nozzle bearing the word “VACUUM”, which is tted internally with the oat
(overow device). When the 90% of the volume of the jar is reached there is the activation of the security
oat (the oat close the aspiration connector on the jar) to avoid liquid penetration inside the device.
FILTER ASSEMBLING
Make sure the lter is assembled with the arrows on the side of the patient.
WARNING: The inside of the medical device must be regularly checked for the presence of liquids or other
visible contamination (secretions). If liquids or other visible contamination are present, the medical device must
be replaced immediately due to the risk of insufcient vacuum ow. These products have been designed, tested
and manufactured exclusively for “single use” and for a period of use not exceeding 24 hours unless stated below.
• Connect the long silicon tube with the lid union still free and marked as “PATIENT”.
• Connect the conical junction for probe insertion with the free end of the long silicon tube.
• Insert the plug of the equipment feeding cable into a power socket.
• Press the ON/OFF button to start the medical equipment.
• To deal with foam formation within the tank, unscrew the tank lid and ll 1/3 of the tank with water (to make
cleaning easier and speed up depression while operating the equipment), place the lid on the jar.
• While using the equipment, the suction tank should always be used vertically to avoid the intervention of
the antireux valve. In case of intervention of this protection, switch the device off and disconnect the tube
connected with the suction tank (the one marked as “VACUUM”) on the same lid.
• You can then detach all accessories and perform cleaning operations as described under “Cleaning acces-
sories and internal parts” below.
Page 8
21
ENGLISH
WARNING: The power supply cable plug is the element of separation from the electrical mains system: even
if the units equipped with a special on / off switch button, the power supply plug must be kept accessible once
the device is in use so as to allow a further method of disconnection from the mains supply system.
MULTIPURPOSE BAR- MPR SYSTEM
The device is equipped with a multipurpose bar to easily change accessories (such as rings of different diameters
for different collection vessel capacities, safety traps, cannula holders or standard 30x10 mm stainless steel
bar, on which any other accessory can be inserted using standard clamps).
USING THE FOOTSWITCH CONTROL:
Connect the footswitch control feeding cable with the plug marked as
After the device has been connected,all Leds are still off. When the ON/OFF button is pressed, all Leds are
activated at once for 1 second (autotest). At the end of the autotest cycle, the ON/OFF button led will ash.
Press the button marked as (---) to perform suction using the footswitch control and execute intermittence
work cycles. Press the button marked as (-) to perform remote suction using the footswitch control (continuous
suction). To stop suction just push on the footswitch control with strength.
USING THE FOOTSWITCH CONTROL AND THE FLOW DEVIATOR:
If using equipment tted with a ow deviator, users may direct suctioned liquids in any of the two collection tanks
provided. Flow deviator comes with two complete suction kits (2 sets of tubes, 2 antibacterial and hydrophobic
lters and two conical junctions). After the device has been connected, all Leds are still off. When the ON/
OFF button is pressed, all Leds are activated at once for 1 second (autotest). At the end of the autotest cycle,
the ON/OFF button led will ash. To decide which side to perform the suction from, press OUT LEFT or OUT
RIGHT and the selected button led will show a blue light.
Press the ON/OFF button again to start the suction cycle. If the device is set up for using the ow deviator,
ensure the antibacterial lter has been positioned on both sides. Connect the footswitch control feeding cable
with the plug marked as
Page 9
ENGLISH
Press the button marked as (---) to perform suction using the footswitch control and execute intermittence
work cycles. Press the button marked as (-) to perform remote suction using the footswitch control (continuous
suction). To stop suction just push on the footswitch control with strength. Press the ON/OFF button to stop
the medical equipment. Before removing the feeding plug, ensure autotest has been performed on the panel.
NEVER USE THE DEVICE WITHOUT JAR AND / OR PROTECTION FILTER
Using FLOVAC® disposable collection system: Before connecting the disposable collection system, remove
the blu ring tted on the tank holder for a more comfortable insertion of the same container.
• After opening the package, fully stretch the bag and then atten it concentrically to eliminate as much air
as possible.
• Insert the bag and apply the cover to an appropriately sized reusable rigid container by pressing rmly
around the entire perimeter. Make sure that the system is completely sealed.
• Close the connector marked as “TANDEM” with the lid provided.
• Connect the power source of the vacuum to the VACUUM port equipped with specic reusable conical
tting with “male” connection.
• Connect the patient tube to the PATIENT port of the cover
• Before use, check all closures and make sure there are no leaks, starting the aspiration source. If the bag
expands to fully adhere to the walls of the rigid container and the cover bends towards the inside of the
glass, the system is not leaking.
• Start the aspiration and periodically check the lling level of the container. The overow valve will cause
the interruption of aspiration if the aspirated uids have reached the maximum lling level of the device.
• When the oat valve intervenes signalling the device is too full, the suction source must be disconnected
within no more than 5 minutes.
22
RISK OF ELECTROMAGNETIC INTERFERENCE AND POSSIBLE REMEDIES
This section contains information regarding the conformity of the compliance with the EN 60601-1-2 (2015)
Standard.
CISPR group and category classication: group 1, category B.
The HOSPI PLUS surgical aspirator is an electro-medical device that requires particular precautions regarding
electro-magnetic compatibility and which must be installed and commissioned according to the electromagnetic compatibility information supplied. Portable and mobile radio communication devices (mobile
phones, transceivers, etc.) may interfere with the medical device and should not be used in close proximity
with, adjacent to or on top of the medical device.
If such use is necessary and unavoidable, special precautions should be taken so that the electro-medical
device functions properly in its intended operating conguration (for example, constantly and visually checking
for the absence of anomalies or malfunctions).
The use of accessories, transducers and cables different to those specied, with the exception of transducers
and cables sold by the appliance and system manufacturer as spare parts, can lead to an increase in emissions
or in a decrease of the immunity of the device or system. The following tables supply information regarding
the EMC (Electromagnetic Compatibility) characteristics of the electro-medical device .
Page 10
23
ENGLISH
Guidance and manufacturer’s declaration – Electromagnetic Emissions
The HOSPI PLUS SUCTION ASPIRATOR is intended for use in the electromagnetic environment specied below. The
customers or the user of the HOSPI PLUS SUCTION ASPIRATOR should assure that it’s used in such an environment.
Group 1The HOSPI PLUS SUCTION ASPIRATOR only used RF energy
only for its internal functioning. Therefore its RF emissions are very
low and are not likely to cause any interference in nearby electronic
equipment.
Irradiated / Conducted
emissions CISPR11
Harmonic emissions
EN 61000-3-2
Voltage uctuations /
Class [B]The appliance is suitable fopr use in alla establishments included
domestic establishments and those directly connected to the public
Class [A]
low-voltage power supply network that supplies buildings used for
domestic purposes.
Complies
icker emissions EN 61000-3-3
Guidance and manufacturer’s declaration – Immunity Emissions
The HOSPI PLUS SUCTION ASPIRATOR is intended for use in the electromagnetic environment specied below. The
customers or the user of the HOSPI PLUS SUCTION ASPIRATOR should assure that it’s used in such an envitonment.
synthetic material, the relative humidity
should be at least 30%.
Electrical fast
transient / burst
EN 61000-4-4
+/-2kV power supply
lines
+/-1kV for input / output
The device doesn’t
change its state
Mains power quality should be that of
a typical commercial or hospital enviro
ment.
lines
Surge
EN 61000-4-5
± 0,5kV
± 1,0kV differential mode
The device doesn’t
change its state
Mains power quality should be that of a
typical commercial or hospital environment.
Volgate dips, short
interruptions and
voltage variations on
power supply input
lines EN 61000-4-11
Magnetic eld
(50/60 Hz)
EN 61000-4-8
T (>95% dip UT )
<5% U
for 0,5 cycle
T (60% dip UT )
40 % U
for 5 cycle
T (30% dip UT )
70 % U
for 25 cycle
T (>95% dip UT )
<5 % U
for 5 sec
30A/mThe device doesn’t
- -Mains power quality should be that of a
typical commercial or hospital environment. If the user of the HOSPI PLUS
SUCTION ASPIRATOR requires continued operation during power mains interruptions, it is recommended that the device be powered from an uninterruptible
power supply or a battery.
The power frequency magnetic eld
change its state
should be measured in the intended installation location to assure that it’s suf-
ciently low.
Note UT is the value of the power supply voltage.
Page 11
ENGLISH
Guidance and manufacturer’s declaration – Immunity Emissions
The HOSPI PLUS SUCTION ASPIRATOR is intended for use in the electromagnetic environment specied below. The
customers or the user of the HOSPI PLUS SUCTION ASPIRATOR should assure that it’s used in such an environment.
Immunity TestLevel indicated by
Conducted
Immunity
EN 61000-4-6
Radiated
Immunity
EN 61000-4-3
Note 1: At 80 MHz and 800 MHz the interval with the highest frequency is applied.
Note 2: These guide lines may not be applicable in all situations. The electro-magnetic propagation is inuenced by the
absorption and by reection from buildings, objects and people.
a) The eld intensity for xed transmitters such as the base stations for radiotelephones (mobile and cordless) and ter-
restrial mobile radio, amateur radio devices, radio AM and FM transmitters and TV transmitters can not be theoretically
and accurately foreseen. To establish an electro-magnetic environment generated by xed RF transmitters, an electromagnetic study of the site should be considered. If the eld intensity measured in the place where the device will be used
surpasses the above mentioned applicable level of conformity, the normal functioning of the device should be monitored.
If abnormal performance arises, additional measures such as changing the device’s direction or positioning may be
necessary.
b) The eld intensity on an interval frequency of 150 kHz to 80 MHz should be less than 3 V/m.
the EN 60601-1-2
da 3Vrms 150kHz
a 80MHz
(for non life-supporting
devices)
da 3V/m 80MHz
a 2.7GHz
(for non life-supporting
devices )
Livello di
conformità
V1 = 3 V rms
E1 = 10 V / m
Electromagnetic environments - guidance
Portable and mobile RF communication equipment, including cables, should be used no closer to any part of
the device, than the recommended separation distance
calculated from the equation applicable to the frequency of the transmitter.
Recommended separation distance
3,5
√P
d=
1
V
12
from 80MHz to 800MHz
√P
d=
1
E
23
d=
1
E
Where P is the maximum output power rating of the
transmitter in Watt (W) according to the transmitter
manufacturer and is the recommended separation
distance in metres (m). Field strengths from xed RF
transmitters, as determined by an electromagnetic site
study of the site, should be less that the compliance
level in each frequency range. Interference
may occur in the vicinity of equipment
marked with the following symbol:
from 800MHz to 2,7GHz
√P
24
Page 12
25
ENGLISH
Recommended separation distance between portable and mobile radio-communication devices and the monitor
The surgical aspirator is intended for use in an electromagnetic environment in which radiated RF disturbances are con-
trolled. The customer or the user of the HOSPI PLUS SUCTION ASPIRATOR device can help prevent electromagnetic
interference by maintaining a minimum distance between portable and mobile RF communication equipment (transmit-
ters) and the HOSPI PLUS SUCTION ASPIRATOR device, as recommended below, according to the maximum output
power of the communications equipment.
Maximum nominal
output power of the
Transmitter W
0,010,120,120,23
0,10,380,380,73
11,21,22,3
103,83,87,3
100121223
For transmitters with a maximum nominal output power not shown above, the recommended separation distance in metres
(m) can be calculated using the equation applicable to the transmitter frequency, where P is the maximum nominal output
power of the transmitter in Watt (W) depending on the transmitter’s manufacturer.
Note 1: At 80 MHz and 800 MHz the interval with the highest frequency is applied.
Note 2: These guide lines may not be applicable in all situations. The electro-magnetic propagation is inuenced by the
absorption and by the reection from buildings, objects and people.
Separation distance from the frequency transmitter (m)
150KHz a 80MHz
3,5
√P
d=
1
V
80MHz a 800MHz
12
√P
d=
1
E
800MHz a 2,7GHz
23
√P
d=
1
E
SYMBOLOGIE
~
(—)
Class II applied
General warnings and/or
specications
Medical Device compliant with
Directive 93/42/CEE
Type B applied part
Keep away from sunlightKeep in a cool, dry place
Atmospheric pressure limitHumidity limit
Temperature limitWEEE disposal
ManufacturerDate of manufacture
Alternating current
Hz
Mains Frequency
On / OffFuse
Using the footswitch control
(for continuous suction)
(---)
Using the footswitch control
(for intermittence suction)
Page 13
ENGLISH
Product codeLot number
Serial numberFollow instructions for use
26
IPX1
(on the footswitch
Covering Protection rate
control label)
Please note technical specications may vary upon the manufacturer’s discretion!
Disposal: The product must not be disposed of along with other domestic waste. The users must
dispose of this equipment by bringing it to a specic recycling point for electric and electronic equipment.
GIMA WARRANTY TERMS
The Gima 12-month standard B2B warranty applies.
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.