DESCRIPTION/INTENDED USE:
These devices are single-use supplied sterile. Check plug matching on generator base and connect. Press CUT/yellow button for cutting and COAG/blue button for coagulation.
CONTRAINDICATIONS:
These devices should never be used when:
There is visible evidence of damage to the exterior of the device such as cracked
plastic or connector damage.
These devices fail the inspection described herein.
In the presence of ammable gases, liquids, and/or oxygen enriched environments.
SAFETY TIPS:
Use lowest possible power setting on electrosurgical unit capable of achieving
desired surgical effect.
Never allow the cables connected to these devices to be in contact with skin of the
patient or operator during electrosurgical activations.
Do not permit the cables connected to these devices to be parallel and in close
proximity to the leads of other electrical devices.
Always place unused electrosurgical accessories in a safe insulated location such
as a holster when not in use.
Inspect and test each device before each use.
Discard devices that have reached their life expectancy.
4
Activation of an electrosurgical device when not in contact with target tissue or
in position to deliver energy to target tissue (fulguration) may cause capacitive coupling.
INSPECTION:
These devices should be inspected before and after each use. Visually examine the devices for obvious physical damage including:
Cracked, broken or otherwise distorted plastic parts.
Broken or signicantly bent connector contacts.
Damage including cuts, punctures, nicks, abrasion, unusual lumps, signicant
discoloration.
Tips for damage, corrosion or misalignment condition.
STORAGE & HANDLING:
These devices must be stored in a clean, cool and dry area. Protect from mechan­ical damage and direct sunlight. Handle with extreme care.
Keep away from sunlight
Caution: read instructions (warnings) carefully
Manufacturer
Date of manufacture Expiration date
Disposable device, do not re-use
Keep in a cool, dry place
Medical Device complies with Directive 93/42/EEC
Consult instructions for use
Sterilized using ethylene oxide