DESCRIPTION/INTENDED USE:
These devices are single-use supplied sterile. Check plug matching on generator
base and connect. Press CUT/yellow button for cutting and COAG/blue button for
coagulation.
CONTRAINDICATIONS:
These devices should never be used when:
• There is visible evidence of damage to the exterior of the device such as cracked
plastic or connector damage.
• These devices fail the inspection described herein.
• In the presence of ammable gases, liquids, and/or oxygen enriched environments.
SAFETY TIPS:
• Use lowest possible power setting on electrosurgical unit capable of achieving
desired surgical effect.
• Never allow the cables connected to these devices to be in contact with skin of the
patient or operator during electrosurgical activations.
• Do not permit the cables connected to these devices to be parallel and in close
proximity to the leads of other electrical devices.
• Always place unused electrosurgical accessories in a safe insulated location such
as a holster when not in use.
• Inspect and test each device before each use.
• Discard devices that have reached their life expectancy.

4
• Activation of an electrosurgical device when not in contact with target tissue or
in position to deliver energy to target tissue (fulguration) may cause capacitive
coupling.
INSPECTION:
These devices should be inspected before and after each use. Visually examine
the devices for obvious physical damage including:
• Cracked, broken or otherwise distorted plastic parts.
• Broken or signicantly bent connector contacts.
• Damage including cuts, punctures, nicks, abrasion, unusual lumps, signicant
discoloration.
• Tips for damage, corrosion or misalignment condition.
STORAGE & HANDLING:
These devices must be stored in a clean, cool and dry area. Protect from mechanical damage and direct sunlight. Handle with extreme care.
Keep away from
sunlight
Caution: read
instructions
(warnings) carefully
Manufacturer
Date of manufacture Expiration date
Disposable device,
do not re-use
Keep in a cool, dry
place
Medical Device
complies with
Directive 93/42/EEC
Consult instructions
for use
Sterilized using
ethylene oxide