Gima EXACON TEMPERATURE SENSOR User guide

EXACON
Temperature Probes
Temperatursonden
Sondes de température
Temperatuursensoren Sonde di temperatura
Sondas de temperatura
Temperatursensorer
Instructions for use
Hinweise zum Gebrauch
Instructions d’utilisation
Algemene informatie
Instruzioni per l’uso
Instrucciones para su uso
Informação Geral
Brugsanvisning Bruksanvisning
Medical Devi ce Directiv e 93/42/EEC
Annex II, section 3.2 - Class IIb
0543
QMS: EN ISO9001 & EN46001
MAX. 80°C
D-RA5JK D-RB4JK D-OS4JK D-OS3JK D-S18JK D-S10JK
D-OS2A / JK / HP / SA
D-S06A / JK / HP / SA
D-TM1A / JK / HP / SA
ChemSter
MAX. 121°C
D-RA5A / HP / SA D-RB4A / HP / SA D-OS4A / HP / SA D-OS3A / HP / SA D-S18A / HP / SA D-S10A / HP / SA
ChemSter & Autoclave
"JK"
"A"
"HP"
"SA"
MDD: 93/42/EEC Annex II, Sec. 3.2 Class IIa QMS: EN ISO9001 & EN46001
Std: EN60601 / EN12470-4 Ref: 2252 ohm / 25°C Tol: 0-25°C: ±0,2°C; 25-50°C: ±0,1°C
Pwr: Max 100 µW RA5-RB4-OS4-OS3-S18-S10
Max 70 µW OS2 - S06 - TM1
PVC
0543 LATEX
50ºC
0ºC
Ordering information Bestellinformation Rérérences de commande Información para pedidos Codici d’ordine Bestillingsnumre
Ordering example:
Standard: D-OS3JK Autoclave: D-OS3A¹ HP: D-OS3HP Siemens¹: “A” SC6000: "SA" SC9000: "SA"
Rectal / Rektal / Rectale / Recto / Rettale
D-RA5”+” Adults
D-RB4”+” Infants
Central / Zentral / Centrale / Centrali
D-OS4”+” Adults
D-OS3”+” Infants
D-OS2”+” Neonates
D-TM1”+” Tympanic
Skin / Haut / Cutanée / Piel / Cutanee
D-S18”+” Adults
D-S10”+” Infants
D-S06”+” Neonates
Instructions for use
Skin and Extremity Temperature Probes
Attach probe to finger or toe, using adhesive tape. Take care not to compress blood vessels. Furthermore secure lead with adhesive tape. Readout stabilizes within approximately 3 min.
Central Temperature Probes
Introduce probe, e.g. into the esophagus, the nasopharynx, the auditory canal up to the tympanic membrane, or the rectum.
Use only water based cremes for introduction.
Esophageal application
Introduce the probe until it is situated in the inferior quarter of the esophagus.
Nasopharyngeal application
The correct position has been reached if you feel resistance when trying to advance the probe further. Caution! Introduction of the probe may induce nosebleeding
Rectal application
The probe is in correct position when it has passed the sphincter. Use disposeable protective sheet (PC1)
Aural application
Advance the probe until tip reaches the tympanic membrane. Wrap a layer of cotton wool around the cable to close the auditory canal.
Different parts of the body may differ significantly in temperature. Monitoring temperature from two different sites, e.g. oesophagus and extremity, provides indirect information on the blood flow.
Cleaning
Remove adhesive tape, use cleaning benzine to remove rests of adhesive. Clean with a cloth dampened with soap water. Caution! Do not immerse connector in the cleaning solution.
Disinfection
Suitable disinfectants: Ethanol and isopropyl alcohol 70%.
Sterilization:
Steam autoclave at 121ºC ~ 250ºF in 20 min.
Chemical with an ethyleneoxide mixture.
Do not apply harmed or defective probes, consult the supplier.
MAX. 121°C
MAX. 80°C
LOT
YYWW
PC1
Oesophagus Nasopharynx