Gima DefiMonitor XD DEFIBRILLATOR manual User guide

DefiMonitor XD
Operating instructions
24050 | GB | I
10/2021
www.primedic.com

Masthead

Publisher
Metrax GmbH Rheinwaldstr. 22 D-78628 Rottweil Germany Phone: +49 741 257-0 Fax: +49 741 257-235
www.primedic.com
info@primedic.com Revision: I Date of issue: 10/2021
Proprietary note
Metrax GmbH reserves all rights to these instructions for use. These instructions for use may not be duplicated or made accessible to third parties without the approval of Metrax GmbH. The same ap­plies to individual parts and excerpts of these instructions for use.
Non-compliance with this gives rise to a right to claim damages and can have consequences under criminal law.
We reserve the right to make amendments to these instructions for use.
Metrax GmbH Rheinwaldstr. 22 78628 Rottweil Germany
Table of contents
DefiMonitor XD | 24050 | GB | I
3 / 108
Table of contents
Masthead 2
1 Glossary 7
2 Introduction 7
2.1 Foreword 7
2.2 Validity 8
2.3 Disclaimers 8
2.4 Symbols used in these instructions for use 8
2.5 Pictograms 9
3 Purpose 12
3.1 Functionality 12
3.2 Operational principle 12
3.3 Intended use 14
3.3.1 Medical indications 14
3.3.2 Medical contraindications 15
3.3.3 Intended patient group 15
3.3.4 Probable body part 16
3.3.5 Intended use environment 16
3.3.6 Intended user profile 16
4 General safety advice 16
5 Description of device 17
5.1 General description 17
5.2 Scope of delivery 18
5.3 Description of device details 19
5.3.1 Controls 22
5.3.2 Monitor display 23
5.3.3 Monitor symbols 24
6 Preparatory measures before (initial) start-up 25
6.1 Power supply 26
6.1.1 Removing the energy module 27
6.1.2 Activate the AkuPak LITE XD 27
6.1.3 Inserting the energy module 28
6.1.4 Charging the AkuPak LITE XD 28
6.1.5 Battery capacity display on the monitor 29
6.1.6 Connection to the supply network 29
6.1.7 Separation from the supply network 30
7 Using the DefiMonitor XD 30
7.1 Switching on / off 32
7.1.1 Switching on 32
7.1.2 Switching off 32
7.2 Device self-test 32
7.2.1 Self-test after switching on 33
7.2.2 Automatic, periodic self-tests 33
Table of contents
4 / 108
DefiMonitor XD | 24050 | GB | I
7.2.3 DefiMonitor XD status display 33
7.2.4 Internal error 35
7.3 Settings 35
7.3.1 Changing settings – time 35
7.3.2 Basic settings 36
7.4 Positioning the electrodes on the patient 37
7.4.1 Undressing the patient 38
7.4.2 Removing chest hair 38
7.4.3 Drying the skin 38
7.4.4 SavePads 38
7.4.5 Paddles 39
7.4.6 Positioning the SavePads and/or paddles on adults 40
7.4.7 Positioning the SavePads and/or paddles on children 41
7.4.8 Opening and attaching SavePads 42
7.4.9 Positioning the ECG adhesive electrodes 43
7.5 Removing the electrodes from the patient 44
7.6 Fitting an SpO2 sensor 44
7.7 Removing an SpO2 sensor 45
7.8 Alarms and alarm limits 45
7.8.1 Alarms with high priority 45
7.8.2 Alarms with medium priority 46
7.8.3 Alarms with low priority 47
7.8.4 Informative messages 47
7.8.5 Messages in AED-Mode 53
7.8.6 Alarm volume 54
7.8.7 Alarm limits 54
7.8.8 Muting the alarm 55
7.9 ECG monitoring (ECG MON) 55
7.9.1 Connecting the ECG electrodes 56
7.9.2 Changeover of ECG source 56
7.9.3 Switching the leads on the monitor 57
7.10 Manual asynchronous / synchronous defibrillation (MAN / SYNC) 57
7.10.1 Energy selection 58
7.10.2 Charging 58
7.10.3 Triggering defibrillation 59
7.11 SpO2 measurement 62
7.11.1 General SpO2 safety advice 63
7.11.2 Shown on the monitor 64
7.12 Pacer (PACE) 64
7.12.1 Setting the pacer modes 66
7.12.2 Setting stimulation frequencies 66
7.12.3 Setting stimulation intensities 66
7.12.4 Starting and stopping stimulation in pacer mode (PACE) 67
7.12.5 Defibrillation during pacing / Defibrillation via multifunction electrodes 68
7.13 AED-Mode (AED) 68
7.13.1 Voice messages 69
7.13.2 Carrying out ECG analysis in AED-Mode 69
7.13.3 Defibrillation required 70
7.13.4 Defibrillation not required 71
7.14 Keeping the defibrillator ready for use 71
Table of contents
DefiMonitor XD | 24050 | GB | I
5 / 108
7.14.1 MMI test (Man-Machine-Interaction) 72
7.15 Event button 72
7.16 Operating the printer 73
7.16.1 Inserting paper into the printer 73
7.16.2 Automatic self-test printout 73
7.16.3 ECG signal log 75
7.16.4 Automatic printout after each shock 76
7.16.5 Printing out the event memory 76
8 SaveCard data management 76
8.1 Inserting / replacing the SaveCard 77
8.1.1 Inserting SaveCard 77
8.1.2 Changing SaveCard 78
9 Cleaning, maintenance and shipping 78
9.1 Cleaning and disinfection 78
9.2 Servicing 79
9.3 Shipping 79
10 Disposal 79
11 Technical Data 80
11.1 Monitor 80
11.2 Alarms 80
11.3 Physical properties 80
11.4 Printer 80
11.5 Memory 81
11.6 Voltage supply 81
11.7 Environmental conditions 82
11.8 Sound definitions 83
11.9 Monitoring 83
11.9.1 ECG 83
11.9.2 SpO2 85
11.10 Treatment parameters 86
11.10.1 Biphasic curve properties 86
11.10.2 Manual defibrillation (asynchronous / synchronous) 88
11.10.3 Pacer 88
11.10.4 AED-Mode 89
11.11 Multifunction electrodes (SavePads) 89
12 Appendix 91
12.1 Illustration of time-current curves 91
12.2 General advice for using pulse oximeters 94
12.2.1 Data Update Period, Data Averaging, and Signal Processing 94
12.2.2 Functional versus Fractional Saturation 94
12.2.3 Performance Considerations 95
12.2.4 Nellcor™ Pulse Oximetry Sensors 97
12.2.5 Functional Testers and Patient Simulators 99
12.3 Rhythm detection system in AED-Mode 99
12.4 Electromagnetic compatibility 100
12.5 Optional accessories 104
Table of contents
6 / 108
DefiMonitor XD | 24050 | GB | I
Glossary
DefiMonitor XD | 24050 | GB | I
7 / 108

1 Glossary

Term / abbreviation
Description
AED
Automated external defibrillator
AHA
American Heart Association
AkuPak LITE XD
Rechargeable energy module
bpm
“beats per minute”
EAR
Used Electronic Appliances Register
ECG
Electrocardiogram
ElektroG
German Electrical and Electronic Equipment Act
ERC Guidelines
European Resuscitation Council guidelines on cardiopulmonary re­suscitation (CPR)
HF
High frequency
CPR
Cardiopulmonary resuscitation
Internal buffer battery
Internal buffer battery for continued running of DefiMonitor XD real­time clock if no energy module is inserted.
MDD
Medical Device Directive
MIT
Massachusetts Institute of Technology
ms
millisecond
ÖRE
Public law
Patient impedance
Patience resistance between the SavePads
SaveCard
Memory card for data transfer
s
second
SavePads
Self-adhesive multifunction electrodes for defibrillation, stimulation, monitoring, cardioversion
WEEE
Waste of Electrical and Electronic Equipment

2 Introduction

2.1 Foreword

Dear User, You are faced with the task of using the DefiMonitor XD on human beings in a medical emergency! To ensure that you can react quickly and correctly in this special situation and can optimally use the
options given with the device, it is necessary for you to read through these operating instructions in your own time beforehand to familiarise yourself with the device, its functions and the areas of applica­tion.
Keep these instructions for use near the device so that you consult them for any queries which may arise.
If you have any questions regarding the start-up, use or maintenance of the DefiMonitor XD please do not hesitate to contact us.
In case of unexpected device behaviour or events, please contact us. The instructions given on the device are no substitute for reading these operating instructions. Serious incidents related to the defibrillator must be reported. If the defibrillator has not performed as
expected, contact the manufacturer and the appropriate local authority.
Introduction
8 / 108
DefiMonitor XD | 24050 | GB | I
A "serious incident" means an event that has had, could have had, or may have had, directly or indi­rectly, any of the following consequences
the death of a patient, user or other person
the temporary or permanent serious deterioration of the health status of a patient, user or other
person
a serious risk to public health. You will find our contact address on the masthead at the start of these instructions for use.

2.2 Validity

The descriptions in these operating instructions refer to all models in the DefiMonitor XD range of de­fibrillators made by Metrax GmbH.
These operating instructions describe the software version ARM 1.XX, DSP 4.XX. Please note that, depending on the model, your DefiMonitor XD may not have all the features de-
scribed in these operating instructions. Refer to the start dialogue of your DefiMonitor XD to determine which model you have.
The DefiMonitor XD is run using an AkuPak LITE XD. This item will also be referred to as an energy module in this document.
SavePads are self-adhesive multifunction electrodes which can be used for defibrillation, stimulation, monitoring and cardioversion. For statements in these operating instructions concerning all SavePads models, they will be referred to simply as SavePads. If there are specifics for using individual models, the item description will be given in full (e.g. SavePads Connect).
The content of this document can be changed without prior notice.

2.3 Disclaimers

Liability claims for personal injury and damage to property are excluded if they are attributable to one or more of the following causes:
Using the device in a manner for which it was not intended.
Improper use and maintenance of the device.
Operating the device with the protective covers removed or if there is obvious damage to the ca-
bles and/or electrodes.
Non-observance of the advice given in these instructions for use with regard to operation, mainte­nance and repair of the device.
Customer intervention, repairs or constructional changes to the device.
Autonomous exceeding of the performance limits.
Lack of monitoring of parts that are subject to wear and tear.
Treating patients without prior indication.

2.4 Symbols used in these instructions for use

Note
This symbol indicates text which contains important information.
Follow the instructions in the order in which they are given in the instructions for use. The instructions are structured as follows:
Procedure:
First instruction
Second instruction
Introduction
DefiMonitor XD | 24050 | GB | I
9 / 108
Possible interim result
etc. Possible final result
This symbol indicates a list.
Numbers in brackets, e.g. (3), refer to items in figures. Voice messages or voice instructions are in bold in these instructions for use. In some instances, they
will be shown as text messages on the monitor at the same time. Settings options will be placed in “ “.

2.5 Pictograms

Symbol
Meaning
Serial number
CE symbol for the notified body
CE symbol
HIBC / UDI code (example)
GOST R symbol
EAC symbol
Medical device
Applied part, defibrillation-proof, type CF
Protection class II
Do not dispose of device in household waste
IP33
Protected against solid foreign objects larger than 2.5 mm and against spraying water
Manufacturer and production date if necessary YYYY-MM-DD
Follow the instructions for use
"General warning” safety symbol – the individual meanings are explained in the operating instructions
Dangerous electrical voltage (high voltage)
Introduction
10 / 108
DefiMonitor XD | 24050 | GB | I
Symbol
Meaning
Caution
Service life of the internal battery YYYY-MM-DD
Do not short-circuit energy module
Do not open or disassemble energy module
Protect energy module from fire
For adults and children more than 8 years old and with a bodyweight greater than 25 kg
For children aged 1 – 8 and with a bodyweight up to a maximum of 25 kg
Infants up to the age of 1
SavePads Connect can only be used in conjunction with a coded Save­Pads Connect cable
Contains no hazardous substances and can be recycled
Open packaging here Do not use if packaging is damaged
Do not bend or fold the electrodes and do not store beneath heavy items
Recycling code for low-density polyethylene
Do not reuse
non sterile
Can be used a maximum of 24 hours after opening
Observe the instructions for use
Not made from natural latex
Introduction
DefiMonitor XD | 24050 | GB | I
11 / 108
Symbol
Meaning
Batch code
Article number
Global Trade Item Number
Can be used until YYYY-MM-DD
Maximum frequency of use for 1 pair of multifunction electrodes (Save­Pads) for defibrillation
Maximum duration of use for 1 pair of multifunction electrodes (SavePads) for monitoring
Maximum duration of use for ECG electrodes
Maximum duration of use for 1 pair of multifunction electrodes (SavePads) for pacing
Positioning and handling the SavePads
1. Dry skin
2. Open the packaging of the electrodes
3. Pull off protective foil
4. Attach SavePads and smooth out
Positioning and handling the SavePads Mini
1. Dry skin
2. Open the packaging of the electrodes
3. Pull off protective foil
4. Attach SavePads and smooth out Prescription only
Silver / Silver Chloride
Remove chest hair
Dry skin
Connect the ECG cable to the ECG electrodes
Purpose
12 / 108
DefiMonitor XD | 24050 | GB | I
Symbol
Meaning
Remove protections foil
Attach the ECG electrode to the skin and press down with circular move­ments
Store remaining ECG electrodes in their original packaging. Close the packaging by folding the opening twice
This side up
Fragile goods
Protect from heat (solar radiation)
Protect from moisture
Air pressure in hPa
Humidity specification in %
Permissible temperature range in Celsius and Fahrenheit

3 Purpose

3.1 Functionality

The DefiMonitor XD is a portable defibrillator / monitor that is intended for use of trained medical per­sonnel in indoor and out-of-doors emergency care setting with environmental condition specified in the technical data. DefiMonitor XD is designed to be used during ground transportation. DefiMonitor XD is battery powered and can be powered by mains. The battery has to be inserted for operation. DefiMon­itor XD may only be used on one patient at a time.
For additional intended use information see below.
Function
Availability
ECG monitoring
Standard
Async manual defibrillation
Standard
Sync manual defibrillation
Standard
AED-Mode
Option
Pacing
Option
Pulse oximetry (SpO2)
Option

3.2 Operational principle

ECG
The electrocardiogram (ECG) records the electrical activity of the heart. The ECG allows the interpre­tation and identification of cardiac rhythm, dysrhythmias and calculation of heartrate. The ECG is
Purpose
DefiMonitor XD | 24050 | GB | I
13 / 108
captured by multifunction electrodes or ECG electrodes on the patient's skin what allows the electrical activity to be monitored and recorded. ECG monitoring is used in addition to assess the patient, do not rely solely on the ECG monitor, care should be taken to assess the patient at all time.
Asynchronous Defibrillation
DefiMonitor XD applies a high-intensity current pulse to the patient to treat life-threatening cardiac ar­rhythmia.
The high-intensity current pulse is called defibrillation. DefiMonitor XD will apply high-intensity current pulse in biphasic waveform. In adult mode, DefiMonitor XD will apply up to 360J, in pediatric mode up to 100J.
DefiMonitor XD DefiMonitor XD provides asynchronous manual defibrillation using multifunction elec­trodes or paddles.
Synchronous Defibrillation (Cardioversion)
DefiMonitor XD applies a high-intensity current pulse synchronized to R-wave of the ECG to the pa­tient to treat cardiac arrhythmia.
The high-intensity current pulse is called defibrillation. DefiMonitor XD will apply high-intensity current pulse in biphasic waveform. In adult mode, DefiMonitor XD will apply up to 360J, in pediatric mode up to 100J.
DefiMonitor XD DefiMonitor XD provides synchronous manual defibrillation using multifunction elec­trodes or paddles.
Pacemaker therapy
Pacemaker therapy mode generates electrical impulses delivered by multifunction electrodes on the patient’s chest to contract the heart muscle and regulate the electrical conduction system of the heart. By this the heart rate of the patient can be controlled by the pacemaker therapy.
The electrical stimulus is applied from one electrode to the other electrode.
AED-Mode
AED-Mode is intended for treating patients with symptoms of sudden cardiac arrest. AED-Mode pro­vides treatment protocol using voice and visual guidance to operator. After the multifunction electrodes
are attached the patient’s chest or upper back DefiMonitor XD analyses the patient’s heartbeat. Defi-
Monitor XD will deliver a high-intensity current pulse to the patient. The high-intensity current pulse is delivered by the multifunction electrodes. The high-intensity current pulse is delivered when the opera­tor presses the shock button
The high-intensity current pulse is called defibrillation. DefiMonitor XD will apply high-intensity current pulse in biphasic waveform. In adult mode, DefiMonitor XD will apply up to 360J, in pediatric mode up to 100J.
SpO2
SpO2 Monitoring uses pulse oximetry to measure functional oxygen saturation in the blood. Pulse oxi-
metry works by applying a Nellcor™ sensor to tissue regions with rich presence of capillaries and arte-
rioles, such as a finger or toe. The sensor contains a dual light source and a photodetector. Bone, tissue, pigmentation, and venous vessels normally absorb a constant amount of light over time.
The vascular bed normally pulsates and absorbs variable amounts of light during the pulsations. The ratio of light absorbed is translated into a measurement of functional oxygen saturation (SpO2).
Ambient conditions, sensor application, and patient conditions can influence the ability of the monitor­ing cable to accurately measure SpO2.
Pulse oximetry is based on two physical principles: oxyhemoglobin and deoxyhemoglobin differ in their absorption of red and infrared light (measured using spectrophotometry), and the volume of arterial blood in tissue (and hence, light absorption by that blood) changes during the pulse (registered using plethysmography). A monitoring system determines SpO2 by passing red and infra-red light into a vas­cular bed and measuring changes in light absorption during the pulsatile cycle. Red and infrared low­voltage light-emitting diodes (LED) in the sensor serve as light sources; a photo diode serves as the photo detector.
Since oxyhemoglobin and deoxyhemoglobin differ in light absorption, the amount of red and infra-red light absorbed by blood is related to hemoglobin oxygen saturation.
Purpose
14 / 108
DefiMonitor XD | 24050 | GB | I
The monitoring cable uses the pulsatile nature of arterial flow to identify the oxygen saturation of arte­rial hemoglobin. During systole, a new pulse of arterial blood enters the vascular bed, and blood vol­ume and light absorption increase. During diastole, blood volume and light absorption reach their low­est point. The monitoring cable bases its SpO2 measurements on the difference be-tween maximum and minimum absorption (measurements at systole and diastole). By doing so, it focuses on light ab­sorption by pulsatile arterial blood, eliminating the effects of non-pulsatile absorbers such as tissue, bone, and venous blood.

3.3 Intended use

ECG
Monitor the electrical activity of the heart.
Asynchronous Defibrillation
Treatment of life-threatening cardiac arrhythmia applying high-intensity current pulse to patient’s chest or upper back.
Synchronous Defibrillation (Cardioversion)
Treatment of cardiac arrhythmia applying high-intensity current pulse to patient’s chest or upper back.
Pacemaker therapy
Treatment of cardiac arrhythmia applying an electrical stimulus to patient’s chest or upper back.
AED-Mode
Treatment of life-threatening cardiac arrhythmia applying high-intensity current pulse to patient’s chest or upper back.
SpO2
SpO2 monitoring included in DefiMonitor XD is a portable pulse oximeter intended as a continuous non-invasive monitor of arterial oxygen saturation (SpO2) and pulse rate of adult, pediatric, and neo­natal patients during both no motion and motion conditions, and for patients who are well or poorly perfused. The monitoring system is intended for use in hospitals, hospital-type facilities, and during intra-hospital transport.

3.3.1 Medical indications

ECG
ECG monitoring is indicated for recognition of the heart rhythm and to monitor the heart rate of the pa­tient.
Asynchronous defibrillation
Manual asynchronous defibrillation is indicated for the termination of certain potentially fatal arrhyth­mias, such as ventricular fibrillation and ventricular tachycardia.
Synchronous defibrillation (cardioversion)
Manual synchronous defibrillation (Cardioversion) is indicated for the treatment of atrial fibrillation, atrial flutter paroxysmal supraventricular tachycardia and in relatively stable ventricular fibrillation.
Pacemaker therapy
DEMAND / FIX: Pacemaker therapy in DEMAND or FIX pacer mode is intended for treatment of symp­tomatic bradycardia with pulse.
OVERDRIVE: Pacemaker therapy in OVERDRIVE pacer mode is intended for suppression of tachy­cardia.
Purpose
DefiMonitor XD | 24050 | GB | I
15 / 108
AED-Mode
AED-Mode should be used when the patient has all of the following symptoms:
Unconsciousness
Absence of normal breathing
Absence of signs of circulation
SpO2
Nellcor OxiMax™ technology allows the use of different types of SpO2 sensors. The indications of the
sensors are listed in the instructions for use of the sensors.

3.3.2 Medical contraindications

ECG
No contraindications known.
Asynchronous defibrillation
Manual asynchronous defibrillation is contraindicated in the treatment of Pulseless Electrical Activity (PEA) and in the treatment if asystole. Defibrillation is unintended for patient who show signs of circu­lation.
Synchronous defibrillation (cardioversion)
Synchronous defibrillation (Cardioversion) is contraindicated in the treatment of Pulseless Electrical Activity (PEA), such as idioventricular or ventricular escape rhythms, and in the treatment of asystole.
AED-Mode
AED-Mode should not be used if the patient shows one of following symptoms:
Consciousness
Breathing
Signs of circulation
Pacemaker therapy
DEMAND / FIX: Pacemaker therapy in DEMAND or FIX pacer mode is unintended for treatment of ventricular fibrillation or asystole.
OVERDRIVE: Pacemaker therapy in OVERDRIVE pacer mode is unintended for treatment of ventricu­lar fibrillation or asystole.
SpO2
Nellcor OxiMax™ technology allows the use of different types of SpO2 sensors. The contraindications are listed in the instructions for use of the sensors.

3.3.3 Intended patient group

ECG
Patients older than one year where heart rate and heart rhythm recognition is indicated.
Asynchronous defibrillation / synchronous defibrillation (Cardioversion)
Patients where asynchronous or synchronous defibrillation is indicated. Patients older than one year but with less body weight than 25kg shall be treated with SavePads Mini
in pediatric mode or child paddles and a maximum energy of 100J.
Pacemaker therapy
Patients older than one year where pacemaker therapy is indicated.
SpO2
Nellcor OxiMax™ technology allows the use of different types of SpO2 sensors. The intended patient
group is specified in the instruction for use if the SpO2 sensor.
General safety advice
16 / 108
DefiMonitor XD | 24050 | GB | I
AED-Mode
Patients older than one year that show symptoms of sudden cardiac arrest. Patients older than one year but with less body weight than 25kg shall be treated with SavePads Mini
in pediatric mode.

3.3.4 Probable body part

ECG electrodes
ECG electrodes are attached to the patient’s chest.
Multifunction electrodes
SavePads Connect, SavePads Mini Connect, SavePads, SavePads PreConnect, SavePads Mini are attached to the patient’s chest or upper back.
Paddles
Adult / Pediatric paddles are attached to the patient’s chest.
SpO2
Nellcor OxiMax™ technology allows the use of different types of SpO2 sensors. The applied part de-
pends on the sensor used. The applied part is defined in the instructions for use of the sensor itself.

3.3.5 Intended use environment

The DefiMonitor XD will be used on scene of emergency. The DefiMonitor XD can be used in road am­bulances as well. The limitations for temperature, humidity and air pressure are specified in chapter
11.7.

3.3.6 Intended user profile

The DefiMonitor XD is intended to be used by trained medical personnel who are familiar with basic monitoring, vital sign assessment, resuscitation situations and trained on the use of DefiMonitor XD. While using the DefiMonitor XD, the patient has to be attended by trained medical personnel con­stantly.

4 General safety advice

Read the operating instructions carefully before using the DefiMonitor XD for the first time. Only use the DefiMonitor XD as described in the instructions for use.
Consider the environmental conditions mentioned in the technical specifications when storing and op­erating the device.
Install the DefiMonitor XD so that you can separate the device from the supply network at any time. Follow the instructions of the DefiMonitor XD. Use the DefiMonitor XD on a non-conductive base only. Do not use the DefiMonitor XD in standing
water or in the rain. Do not use the DefiMonitor XD in the presence of flammable materials. The DefiMonitor XD and its accessories are safe when used as intended and when following the de-
scriptions and information detailed in these operating instructions. Despite this, if used incorrectly, the DefiMonitor XD and its accessories can be dangerous to the user, the patient or third parties!
Do not touch the contacts on the DefiMonitor XD or on the energy module. The use of several medical devices simultaneously may present a danger to the patient as a result of
the cumulation of patient currents. Keep the device away from children! Only the use of original accessories specified in these instructions for use provides defibrillation pro-
tection. The warnings in the following chapters indicate dangers and must be heeded to avoid the dangers.
They are divided into various escalation levels, see the following warnings.
Description of device
DefiMonitor XD | 24050 | GB | I
17 / 108
DANGER
Texts marked DANGER indicate an extraordinarily serious immediate danger which will defi­nitely lead to serious injury or even death if no preventative measures are adopted! It is im­perative that you pay attention to these texts!
WARNING
Texts marked WARNING indicate an extraordinarily serious potential danger which could lead to serious injury or even death if no preventative measures are adopted! It is imperative that you pay attention to these texts!
CAUTION
Texts marked CAUTION indicate a potentially dangerous situation which could lead to minor injuries! It is imperative that you pay attention to these texts!

5 Description of device

5.1 General description

The DefiMonitor XD is an external defibrillator with integrated 6-channel ECG. The ECG can be rec­orded either via the SavePads, the defibrillation paddles or the four-pin patient cable with self-adhe­sive ECG electrodes. The DefiMonitor XD is available in various models. The respective model will be shown on the monitor when the defibrillator is turned on.
The DefiMonitor XD is available in the following models:
Model
Product designations so far
Basic
device
AED
option
PACER
option
SPO2 option
DefiMonitor XD
DefiMonitor XD1 / DefiMoni­tor XD1xe
X
DefiMonitor XD SPO2
DefiMonitor XD3 / DefiMoni­tor XD3xe
X
X
DefiMonitor XD PACER
DefiMonitor XD10 / Defi­Monitor XD10xe
X X
DefiMonitor XD PACER, SPO2
DefiMonitor XD30 / Defi­Monitor XD30xe
X X
X
DefiMonitor XD AED
DefiMonitor XD100 / Defi­Monitor XD100xe
X X
DefiMonitor XD AED, PACER
DefiMonitor XD110 / Defi­Monitor XD110xe
X X X
DefiMonitor XD AED, SPO2
DefiMonitor XD300 / Defi­Monitor XD300xe
X X
X
DefiMonitor XD AED, PACER, SPO2
DefiMonitor XD330 / Defi­Monitor XD330xe
X X X
X
For all models, energy is provided by the AkuPak LITE XD or via the mains. For further information re­garding the AkuPak LITE XD, please refer to the separate operating instructions.
A wide range of accessories are available. The handling of the accessories is described in separate operating instructions in-part.
The lifetime of DefiMonitor XD is 10 years.
Description of device
18 / 108
DefiMonitor XD | 24050 | GB | I

5.2 Scope of delivery

After delivery, first of all check the packaging and the device for transport damage. If you notice any damage to the device, immediately contact your transport company or dealer, or di-
rectly contact the technical service at Metrax GmbH, stating the device number and describing the damage to the device.
Satisfy yourself that the scope of delivery is complete in accordance with the enclosed delivery note (standard accessories).
Article
Article no.
Further information
DefiMonitor XD
model dependent
language de-
pendent
AkuPak LITE XD
97311
Power supply cord M290 SK II EU black 2,5 m or
Power supply cord M290 SK II USA black 2,5 m
23955 24026
only for Latin America SavePads Connect cable coded
97384
Cable length 3.6 m
SavePads Connect (1 pair)
96516
Paddle set XD
96591
ECG patient cable 4-pole coded IEC
97386
Cable length 3.6 m
ECG electrodes, 1 pack = 30 pcs.
-
Printer paper, 1 roll
-
Contact gel, 1 tube
-
SaveCard
20770
Instructions for useDefiMonitor XD
language de-
pendent
Instructions for use AkuPak LITE XD
language de-
pendent
only for devices with SPO2 option:
Nellcor SpO2 finger sensor FLEXMAX
97802
reusable, patient weight >20 kg
Nellcor SPO2 interface cable DOC10
97221
Cable length 3 m
Subject to change without notice.
Description of device
DefiMonitor XD | 24050 | GB | I
19 / 108

5.3 Description of device details

Fig. 1 DefiMonitor XD – front view
1
Carry handle
2
Paddle
3
Paddle cable, detachable
4
Membrane keypad (option-dependent, here: DefiMonitor XD AED, PACER, SPO2)
5
Monitor
6
Status display
Fig. 2 DefiMonitor XD – rear view (similar to illustration)
1
Opening for single-hand release hook
2
Identification plate
3
Quality seal
Description of device
20 / 108
DefiMonitor XD | 24050 | GB | I
Fig. 3 DefiMonitor XD – bottom view
1
SaveCard cover
2
Release button (to remove the energy module)
3
Energy module AkuPak LITE XD
Fig. 4 DefiMonitor XD – left-hand side view
1
Slot for paddle
2
Loudspeaker
3
Socket for electrode connector
4
Socket for paddle connector
5
Socket for SpO2 sensor (optional)
6
Attachment point for bag
Description of device
DefiMonitor XD | 24050 | GB | I
21 / 108
Fig. 5 DefiMonitor XD – right-hand side view
1
Slot for paddle
2
Release lever for printer cover
3
Printer cover
4
Power supply socket
5
Attachment point for bag
Fig. 6 Paddle set
1
STERNUM paddle button
2
APEX paddle button
3
Paddle connector
Description of device
22 / 108
DefiMonitor XD | 24050 | GB | I

5.3.1 Controls

Fig. 7 DefiMonitor XD controls
1
Pacer on/off (only for the Pacer option)
2
Pacer mode DEMAND / FIX / OVERDRIVE (only for the Pacer option)
3
Stimulation frequency +/- (only for the Pacer option)
4
Pacer Start/Stop with LED (only for the Pacer option)
5
ECG source
6
Stimulation intensity +/- (only for the Pacer option)
7
Alarm acknowledgement button
8
AED button with LED (only for the AED option)
9
Settings keys
Key to scroll upwards in Settings, to increase parameters, or to select upper ECG curve
lead
Enter key to select or confirm
Key to scroll downwards in Settings, to decrease parameters, or to select lower ECG
curve lead
10
On / Off button
11
SYNC button
12
Charging button for use with the multifunction electrodes (SavePads)
13
Energy level in joules
14
Event button
15
Paper feed
16
Printer button
17
Shock button for use with the multifunction electrodes (SavePads)
18
LED display - battery charging
19
LED display - connected to mains
Description of device
DefiMonitor XD | 24050 | GB | I
23 / 108

5.3.2 Monitor display

Fig. 8 Monitor display
1
Period since device started / energy level / progress of charging process / pacer
2
Mode: AED / MAN / SYNC / PACE / ECG MON
3
ECG channel display / number of shocks
4
Heart rate and alarm limits display
5
Plethysmogram and alarm limits display
6
Status bar for displaying SaveCard storage capacity, patient contact status, time, pediatric mode, pacer status, printer status, paper status, battery capacity
7
Calibration peak, height corresponds to 1 mV
8
ECG channels display (max. 2)
9
SpO2 curve, notes, information (SpO2 option)
Description of device
24 / 108
DefiMonitor XD | 24050 | GB | I
Fig. 9 Monitor area
1
Information column
2
Display area
3
Information area 1
4
Information area 2
5
Information area 3
6
Information area 4
7
Information area 5
8
Status bar

5.3.3 Monitor symbols

Symbol
Meaning
Pacer passed self-test
Pacer failed self-test
No paper
Printer error
Paddle
Heart rate
Alarm limits / Alarm activated
Alarm deactivated
Preparatory measures before (initial) start-up
DefiMonitor XD | 24050 | GB | I
25 / 108
Symbol
Meaning
Pediatric mode activated Charge level of energy module, see chapter 6.1.5
Patient contact status
The following tables show the patient contact status. The symbols are displayed in the status bar. Display when using ECG electrodes
ECG lead I
ECG lead II
Display
valid
valid
II / I
valid
invalid
- / I
invalid
valid
II / -
invalid
invalid
- / ­with ECG patient cable, 4-pin, coded (97386)
Display when using SavePads or Paddles
Patient contact
Display
valid
II
invalid
-
Note
You can find more information about ECG leads in chapter 7.9.

6 Preparatory measures before (initial) start-up

Note
Should the device have been stored or transported under extreme environmental con­ditions (see Chapter 11.7), allow the device acclimatise for at least 2 hours before turning it on.
Prior to initial use, the device must be fully commissioned.
Daily testing of the DefiMonitor XD
Note
We recommend carrying out visual inspection of the DefiMonitor XD, its accessories and the MMI test daily.
Conduct the daily testing of the DefiMonitor XD according to the following checklist.
Check expiration dates
Check the expiration date
Of the SavePads Of the ECG electrodes Of the electrode gel Of the AkuPak LITE XD
Replace the parts if necessary!
Preparatory measures before (initial) start-up
26 / 108
DefiMonitor XD | 24050 | GB | I
Check device
Check whether
The status display “OK” is shown You can switch on the device The device automatically carries out the self-test after being switched on The slot for the energy supply is clean The device is fully equipped The cables and plugs show no signs of damage
Check the cleanliness of paddles and children’s paddles.
Carry out the MMI test, see Chapter 7.14.1.
Pay attention to the following
Check whether the SavePads’ case is undamaged. Replace the SavePads if necessary.
If parts of the housing, the insulation or accessories are damaged, they must be repaired or re-
placed immediately.
If parts of the housing, the insulation or accessories are damaged, do not commission the device or switch it off immediately!
Testing the AkuPak LITE XD every 4 - 6 weeks
Disconnect the defibrillator from the mains.
Switch the defibrillator on in battery mode.
Allow the defibrillator to operate for approx. 5 minutes.
Reconnect the defibrillator to the mains power network.
The charge level of the AkuPak LITE XD is reassessed and displayed.

6.1 Power supply

WARNING
The device can only be used if the AkuPak LITE XD is inserted
Treatment or monitoring not possible
Make that sure than an AkuPak LITE XD has been inserted. If the device reports during use that AkuPak LITE XD needs to charged, charge the AkuPak
LITE XD.
If you are unable to charge the AkuPak LITE XD, the device can be operated until it switches itself off.
Regularly check whether the AkuPak LITE XD has sufficient charge. Charge the AkuPak LITE
XD if necessary.
DANGER
Defective cells in the energy module
Danger due to the emission of fluids, gases or flying parts
Do not damage battery cells mechanically. Wear personal protective equipment when handling burst cells. Do not inhale any fumes which arise. Do not inhale any gas which is emitted. Keep the energy module away from sources of ignition.
The DefiMonitor XD has two power supply options:
Energy module AkuPak LITE XD energy module (see attached operating instructions)
Running from the mains with a power cable.
Preparatory measures before (initial) start-up
DefiMonitor XD | 24050 | GB | I
27 / 108
Note
When the AkuPak LITE XD is delivered, the energy module is in “shipping mode” and
must be activated before the first use. The activation process is described in Chapter
6.1.2 .

6.1.1 Removing the energy module

Note
When the DefiMonitor XD is delivered, there is a film between the energy module con­tacts and the device contacts. This film must be removed before use.
Note
Only change the power module when the device is switched off and the mains plug has been disconnected.
Remove the electrode connector before changing the energy module. Wait at least 5 seconds after switching off before removing the energy module.
Fig. 10 DefiMonitor XD – removing the energy module Procedure:
Lay the DefiMonitor XD on its back.
Press the release mechanism button, fig. 10 (1) to the right until the energy module (2) is released
and protrudes from the slot.
Pivot the energy module (2) to the front and remove it.

6.1.2 Activate the AkuPak LITE XD

The AkuPak LITE XD is put into a special energy saving mode – "shipping mode” – before being des­patched, ensuring maximum storage time for the battery. You therefore need to take the AkuPak LITE XD out of this mode before operating a defibrillator with it for the first time.
Fig. 11 AkuPak LITE XD charge level display
1
Button to activate charge level display
2
Charge level display
Preparatory measures before (initial) start-up
28 / 108
DefiMonitor XD | 24050 | GB | I
Procedure, Fig. 11:
Press the charge level display button (1) for several seconds.
Watch the charge level display (2)
If the LEDs start to light up from the middle LED and then also go out again starting from the
middle, the AkuPak LITE XD has been activated.
Checking the AkuPak LITE XD charge level, Fig. 11
If an AkuPak LITE XD is stored outside the device, its charge level can be checked by briefly pressing the button (1).
Charge level dis­play
Meaning 81 % - 100 % charged
61 % - 80 % charged
41 % - 60 % charged
21 % - 40 % charged
1 % - 20 % charged
0 % charged

6.1.3 Inserting the energy module

Fig. 12 DefiMonitor XD – inserting the energy module Procedure:
Lay the DefiMonitor XD on its back.
Push the energy module, fig. 12 (2) into the device in the direction of the arrow until it rests against
the stop position (1) in the energy module slot.
Press the energy module (2) into the energy module slot until it audibly clicks into place and it is flush with the device housing.
The DefiMonitor XD switches on automatically when the AkuPak LITE XD is inserted and per-
forms an automatic self-test.
If the status display shows the message "OK", the device is ready to use.
Switch the DefiMonitor XD off (if required) by pressing the On/Off button, fig. 7 (10).

6.1.4 Charging the AkuPak LITE XD

Procedure
Insert the AkuPak LITE XD into the DefiMonitor XD (see chapter 6.1.3).
Connect the DefiMonitor XD to the mains using the power cable.
The LED to indicate the device is connected to the mains lights up, Fig. 7 (19). It signals that
the DefiMonitor XD has been connected to the mains.
The battery charging LED (18) on the front of the equipment signals that the battery is charg-
ing. It goes out when the AkuPak LITE XD is fully charged.
Preparatory measures before (initial) start-up
DefiMonitor XD | 24050 | GB | I
29 / 108
Note
The AkuPak LITE XD can not be charged with the Charger Basis or the ClipCharger as there are no contacts on the bottom.
The AkuPak LITE XD must be charged before being used for the first time. The process for this is as described above.

6.1.5 Battery capacity display on the monitor

On the DefiMonitor XD the charge status of the energy module is displayed as a bar chart in the status bar on the monitor. Examples of what might be displayed are explained below:
Shown on the monitor
Meaning
100 % charged
80% charged
60% charged
40% charged
20% charged
0 % charged (device will keep going until it is forced to shut down)
Communication with AkuPak LITE XD interrupted or AkuPak LITE XD has reached the end of its service life.
To ensure optimum capacity, the AkuPak LITE XD is monitored using electronic charge regulation. In addition to this display, the DefiMonitor XD also plays a voice message if the charge level of the re-
chargeable battery is too low and it is about to shut down imminently. This voice message will be re­peated every three minutes during operation.
Voice message
Display on monitor
Charging status battery low, please recharge
Charging status battery low, please recharge
Note
When the voice message Charging status battery low, please recharge is played for the first time, the capacity in the energy module will still be sufficient for at least 3
defibrillations at full energy or 90 minutes’ monitoring (ECG and SpO2) or 60 minutes’
pacer operation with maximum output.
Note
Make sure to protect the AkuPak LITE XD from moisture as soon as you have re­moved it from the defibrillator. Otherwise moisture can penetrate inside the battery via the interface and cause damage.

6.1.6 Connection to the supply network

Procedure:
Insert the power cable connector into the power supply socket, Fig. 5 (4) at the device.
Plug the mains plug into an socket with the correct voltage.
The LED, Fig. 7 (19) lights up. This signals that the DefiMonitor XD has been connected to the
mains power supply.
Using the DefiMonitor XD
30 / 108
DefiMonitor XD | 24050 | GB | I
Note
Ensure that the AkuPak LITE XD is inserted in the energy module slot. This is vital to ensure use of the DefiMonitor XD using the mains power supply. The power cable pro­vided must be used.
Should the DefiMonitor XD monitor show a broken energy module and the mains plug be removed, operation of the DefiMonitor XD cannot be guaranteed.

6.1.7 Separation from the supply network

Procedure:
Separate the power cable connector from the power supply socket, Fig. 5 (4) at the device.
Separate the mains plug from the socket.
The LED, Fig. 7 (19) goes out. This signals that the DefiMonitor XD has been separated from
the supply network.

7 Using the DefiMonitor XD

DANGER
Damaged device or accessories
Treatment not possible, injury to patient, user or third party due to electric shock
Do not use the device or its accessories if it is damaged. Check the status display before using the device.
DANGER
Danger of electric shock and too little energy for the patient
Triggering cardiac arrhythmia and burns caused by electric shock
Do not touch the patient during defibrillation. Warn third parties about the dangers of defibrillation. Do not touch any conductive items (metal, blood, water, other liquids, etc.) connected to the pa-
tient during defibrillation.
WARNING
Incorrect use of the device
Treatment with incorrect level of energy, defibrillation at the wrong point in time, no treatment possi­ble, not identifying critical condition of a patient
Make sure that they meet the qualifications for proper use. Make sure that they have been trained in using the device before using the device. Read the operating instructions carefully before using the device for the first time. Monitor the patient continuously while the device is in use. Only actuate the device if the charging process is complete.
WARNING
Defibrillation in environments where there is a risk of fire or explosion
Danger of fire or explosion, burn injuries
Do not use the defibrillator in areas where there is a risk of fire. Do not use the defibrillator in the presence of flammable substances Do not use the defibrillator in areas where there is a risk of explosion Do not use the defibrillator in oxygen-enriched atmospheres.
Using the DefiMonitor XD
DefiMonitor XD | 24050 | GB | I
31 / 108
WARNING
Accessories from third party providers
It may be the case that the device specifications (e.g. the energy emitted does not correspond the energy selected, measurement inaccuracies, electromagnetic radiation) are not adhered to if a non­original accessory is used.
Only use original accessories.
WARNING
Interference from external influences
Defibrillator may not be working as intended Do not use the defibrillator for therapy at the same time as high-frequency surgical devices
(manual asynchronous/synchronous defibrillation, AED-Mode, pacer).
Do not use the defibrillator in areas in which high levels of electromagnetic interference occur
(e.g. near MRI scanners).
Do not use portable high-frequency communication devices (radio equipment and accessories)
closer than 30 cm to the defibrillator or accessories connected to the defibrillator.
WARNING
Injury due to incorrectly secured defibrillator
Injuries to patients, users or third parties
Secure the DefiMonitor XD to the wall bracket during transport. Secure the accessories before transport.
WARNING
DefiMonitor XD can not be turned on
No use of the DefiMonitor XD is possible After use, check the power supply contacts of the energy module and the defibrillator for dam-
age.
WARNING
Incorrect cable laying
Strangulation of the patient Lay the cables on the patient in such a way that it cannot result in strangulation.
WARNING
Other devices interfering with the defibrillator
Potential incorrect operation
Avoid using the defibrillator in the immediate vicinity of other devices. Do not use the defibrillator stacked on top of other devices.
If it is nevertheless necessary to use the device in this way, this device and the other de­vices should be monitored to make sure that they are working properly.
Using the DefiMonitor XD
32 / 108
DefiMonitor XD | 24050 | GB | I
WARNING
Improper lifting of DefiMonitor XD
DefiMonitor XD could fall down.
Do not lift DefiMonitor XD by the cables. Do not lift DefiMonitor XD by the paddles. Carry DefiMonitor XD only on the carry handle.
WARNING
The alarm volume is quieter than the ambient noise
A critical patient condition may not be recognised. Select an alarm volume which is audible despite the ambient noise
WARNING
Improper use of accessories designed for single use
Reduced liability, transmission of pathogens between patients
Do not use disposable accessories if they have expired or are damaged. Do not use disposable accessories when they have dried out. Use disposable accessories only once.
Select your position to the DefiMonitor XD so that you can see the monitor at all times and operated the device.

7.1 Switching on / off

7.1.1 Switching on

The device is switched on by pressing the On / Off button, Fig. 7 (10). Directly after switching it on, an internal self-test is carried out to check important functions and signal devices. The device will go into manual mode after passing a self-test.
Standby is confirmed by a beep.

7.1.2 Switching off

You can switch off the device in different ways:
By pressing the On / Off button for approx. 2 seconds, Fig. 7 (10). A continuous warning beep will sound simultaneously.
The device switches off automatically if it does not detect any signal source for 30 minutes, or if no key is pressed over that same period. A continuous warning beep will sound simultaneously.
If the device detects an error, it will automatically switch off to avoid injuries. A continuous warning beep will sound simultaneously.

7.2 Device self-test

WARNING
Reduced charge level of the energy module due to repeated extended self-tests
The device may not be ready for use.
Regularly check the status display. Resolve technical faults if possible.
Using the DefiMonitor XD
DefiMonitor XD | 24050 | GB | I
33 / 108
WARNING
Only check pacer functionality if the device if switched on
Pacer may not be ready for use. Check the information in the status bar.
WARNING
Only check SpO2 module functionality if the device is switched on
SpO2 module may not be ready for operation Check the display on the monitor
Note
Pacer and SpO2 module functionality are not part of the self-test and must be checked by the user.

7.2.1 Self-test after switching on

The self-test is activated by turning the DefiMonitor XD on or by inserting the AkuPak LITE XD into the device. The self-test checks all important functions and signalling devices.
If an error is detected, an extended self-test (LONG) is automatically carried out. If the error is not re­solved, all subsequent self-tests will be extended (LONG). This can lead to the charge level of the AkuPak LITE XD decreasing more quickly than expected.

7.2.2 Automatic, periodic self-tests

The device carries out automatic self-tests at 8:00 pm device time to ensure that it is always ready for operation. Self-testing is only possible if the DefiMonitor XD is either connected to the mains or if a charged battery has been inserted.
Frequency
Self-test
Scope of test
Daily
SHORT
Software, membrane keypad, ECG calibration, clock, inter­nal voltage supply and HV unit at 0 V, impedance meas­urement
First day of the month
MEDIUM
Software, membrane keypad, ECG calibration, clock, inter­nal voltage supply and HV unit at 300 V, impedance meas­urement
On 1st July and 1st January every year
LONG
Software, membrane keypad, ECG calibration, clock, inter­nal voltage supply and HV unit at 1600 V, impedance measurement
Note
Automatic printout of the results of the self-test can be activated. For more information on this, see chapter 7.16.2

7.2.3 DefiMonitor XD status display

Note
The status display is always viewable even if the device is switched off. Regular checking of the display status is vital for ensuring that the DefiMonitor XD is ready for operation.
Using the DefiMonitor XD
34 / 108
DefiMonitor XD | 24050 | GB | I
Fig. 13 Status display Ensuring readiness for operation
Check the status display, Fig. 13 regularly.
The following table lists the possible displays in the status display and their meanings.
Energy module sta­tus display
Status display DefiMonitor XD
Energy module meaning
Meaning DefiMonitor XD
Availability for use
Measure
Energy module ca­pacity is sufficient
Self-test passed
DefiMonitor XD Ready for use
None
Energy module dis­charged
Self-test passed
DefiMonitor XD Ready for limited use
Charge or replace the energy module
Energy module use-by date exceeded
Self-test passed
DefiMonitor XD Ready for limited use
Check use-by date, replace the energy module.
No energy module in­serted
Self-test passed
DefiMonitor XD Not ready for use
Insert energy module
Battery symbol flashes dur­ing opera­tion
Internal buffer bat­tery empty
Self-test passed
DefiMonitor XD Ready for limited use
Contact the technical services team for re­placement of the inter­nal buffer battery
Energy module ca­pacity is sufficient
Self-test failed
DefiMonitor XD Not ready for use
Carry out a major self­test by reinserting the energy module or by switching the Defi­Monitor XD on again. If the status remains, contact the technical services team
Communi­cation error between DefiMonitor XD and the energy module
Self-test failed
DefiMonitor XD Not ready for use
Carry out a major self­test by reinserting the energy module or by switching the device on again
Charge or replace the energy module
Using the DefiMonitor XD
DefiMonitor XD | 24050 | GB | I
35 / 108
Energy module sta­tus display
Status display DefiMonitor XD
Energy module meaning
Meaning DefiMonitor XD
Availability for use
Measure
Energy module deeply dis­charged
If the status remains, contact the technical services team

7.2.4 Internal error

If the DefiMonitor XD detects an internal error, voice message Internal error will be played. The error code will be displayed on the monitor while the voice message is playing. The DefiMonitor XD will shut itself down independently immediately afterwards.
The error code will be stored on the SaveCard and can be read for servicing purposes.
Note
It may be the case that this error is only temporary or that it is reversible. After receiv­ing the Internal error error message you should turn the device back on after waiting for about 30 seconds and await the results of the self-test started up after the device was turned on. If this is completed successfully, the unit can be used without any problems. If the error continues to occur, contact the technical service.

7.3 Settings

Note
You can close the settings menu at any time by changing the mode or by selecting an energy level, Fig. 7.
The device is configured in the factory. You can change certain parameters in the settings menu. The new configuration remains stored until it
is changed again, irrespective of whether the device is switched off or the power module replaced.
General navigation, Fig. 7 (9):
To open the settings menu during operation, press the enter key .
Press the ▲ (up) key or the ▼ (down) key to navigate in the settings menu and to increase or de-
crease a selected parameter.
Press the key to select a parameter and to confirm the changed value.

7.3.1 Changing settings – time

Procedure:
To open the settings menu during operation, press the enter key , Fig. 7 (9).
Press the ▲ key until the field "To page 2" is highlighted.
Change the page by pressing the key several times until “Settings page 3” is displayed.
Move the cursor upwards by pressing the ▲ key until "Time" is highlighted.
Select the highlighted parameter “Time” by pressing the key.
The hour value is highlighted.
Change the hour by pressing the ▲ key or ▼ key.
Confirm the correct value by pressing the key.
The minute value is highlighted.
Change this entry as described above. Confirm the selected value by pressing the key.
The parameter "Time" is highlighted.
Using the DefiMonitor XD
36 / 108
DefiMonitor XD | 24050 | GB | I
If required, you can change the other parameters in the same way.
To exit the settings menu, use the ▲ key or ▼ key to move the cursor to “Exit” and confirm by
pressing the key. The device is now ready for operation again.

7.3.2 Basic settings

The following table shows the settings of a fully-equipped DefiMonitor XD with AED, PACER and SPO2 options. With the other device variants only entries corresponding to existing options are dis­played.
Settings
Value
Basic setting
SpO2 alarm limits
lower alarm limit: 70 - 99 % upper alarm limit: 71 - 100 %
85 100 SpO2 alarm volume
0 %, 25 %, 50 %, 75 %, 100 %
100 %
ECG alarm limits
lower alarm limit: 30 - 150 bpm upper alarm limit: 31 - 300 bpm
50 100 ECG alarm volume
0 %, 25 %, 50 %, 75 %, 100 %
100 %
ECG sensitivity
5 mm/mV , 10 mm/mV , 15 mm/mV
10 mm/mV
AED rescue breaths
On, Off
On
AED pediatric mode
15:2, 30:2
15
To page 2
-
End
Settings page 2
Value
Basic setting
Print format
1-channel, 3-channel, 6-channel
3-channel
Print speed
25 mm/s, 50 mm/s
25 mm/s
Print on shock
On, Off
Off
Self-test report
On, Off
Off
Print events
- QRS volume
0 %, 25 %, 50 %, 75 %, 100 %
25 %
Metronome volume
0 %, 25 %, 50 %, 75 %, 100 %
100 %
Master volume
0 %, 25 %, 50 %, 75 %, 100 %
100 %
To page 3
-
End
Settings page 3
Value
Basic setting
Date
-
DD/MM/YYYY
Time
-
hh:mm
MMI Test
-
Language
-
depends on lan­guage package
Contrast
60 - 180
120
To page 4
-
End
Settings page 4
Value
Information
ARM SW
-
Version
- Checksum
- Date
Using the DefiMonitor XD
DefiMonitor XD | 24050 | GB | I
37 / 108
Settings page 4
Value
Information
DSP SW
-
Version check­sum
- Date
MSP SW
-
Version check­sum
- Date
ULF
-
Checksum
To page 5
-
End
Settings page 5
Value
Information
BQ type
-
Model
BQ SN
-
Serial number
Ext. MSP SW
-
Version check­sum
Ext. MSP HW
-
Version
SpO2 module
-
PMB05N
Pacer SW
-
Version check­sum
Pacer HW
-
Version check­sum
Device SN
-
Serial number
To page 1
-
End
Note
The values "SpO2 alarm volume" and "ECG alarm volume" are always set to 100 % after the device is switched on.

7.4 Positioning the electrodes on the patient

WARNING
Placing multifunction electrodes or paddles above active implants
Faulty readings due to active implants or damage to the active implants
Make sure that there is no active implant in the path of the current. Do not stick the multifunction electrodes directly over an active implant (pacemaker or similar). Do not place the paddles directly over an active implant (pacemaker or similar).
Make sure that the conductive parts of the electrodes and the connectors of the parts being used are not in contact with conductive parts or the earthing.
Note
Use the electrodes on intact (uninjured) skin.
Note
The electrodes may cause redness in the area to which they are applied.
Using the DefiMonitor XD
38 / 108
DefiMonitor XD | 24050 | GB | I

7.4.1 Undressing the patient

Undress the patient’s upper body so that you can place the multifunction electrodes (SavePads).

7.4.2 Removing chest hair

Remove any chest hair present in those areas where the multifunction electrodes (SavePads) are to be placed.

7.4.3 Drying the skin

If the patient’s chest is damp, dry the skin in those areas where the multifunction electrodes (Save­Pads) are to be placed. You can use the patient’s clothing as a drying material.

7.4.4 SavePads

WARNING
Using SavePads Mini / child paddles on adults
Energy output too low for patients with a bodyweight >25 kg For patients with a bodyweight >25 kg, use the multifunction electrodes for adults or the adult
paddles.
WARNING
Improper usage of multifunction electrodes meant for one-off usage
Irritation or burning of the skin in the area to which the multifunction electrodes are applied, dis­rupted ECG analysis, reduced treatment functions due to multifunction electrodes drying out, trans­mission of pathogens between patients
Do NOT use the SavePads if they are out-of-date or damaged. Do NOT use the SavePads if they have dried out. Use the SavePads only once.
SavePads are self-adhesive multifunction electrodes which can be used for defibrillation, stimulation, monitoring and cardioversion.
SavePads are available in various models, see chapter 12.5. The DefiMonitor XD is delivered with SavePads Connect as standard. The SavePads Connect are connected to the DefiMonitor XD using the SavePads Connect cable.
Note
Heed the use-by date. The SavePads are to be replaced once they have expired.
Fig. 14 SavePads Connect (unpacked)
Using the DefiMonitor XD
DefiMonitor XD | 24050 | GB | I
39 / 108
7.4.4.1 SavePads Mini / Pediatric mode
When the DefiMonitor XD detects that SavePads Mini are connected, the energy is limited to max. 100 joules. The DefiMonitor XD is in Pediatric Mode. This applies for asynchronous and synchronous de­fibrillation. Also in AED mode the energy is automatically reduced. The ratio of chest compression to ventilation is adjusted to the setting.
TheDefiMonitor XD plays the voice prompt Pediatric Mode. As long as the DefiMonitor XD is in Pedi­atric Mode, the status bar shows the icon .
When the SavePads Mini are removed from DefiMonitor XD, the voice prompt Adult Mode will sound.

7.4.5 Paddles

DANGER
Damaged device due to defibrillation with short-circuited paddles
Treatment not possible, injury to patient, user or third party due to electric shock Do not trigger defibrillation if the electrode areas of the two paddles are touching (short circuit)
WARNING
Short-circuit via electrode gel
Treatment not possible, injury to patient, user or third party due to electric shock
Make sure that no short-circuiting can occur via a gel bridge. Make sure that the user is not connected to the patient via electrode gel.
CAUTION
High contact resistance between paddles and patients
Irritation or burning of the skin in the area the paddles are applied, disrupted ECG analysis, reduced treatment functions
Apply a sufficient amount of electrode gel to the contact surfaces of the paddles before usage.
Note
If the paddles are used, the contact surfaces must be furnished with sufficient elec­trode gel.
The DefiMonitor XD has paddles with integrated child paddles for the defibrillation of children. There is a button for triggering defibrillation on each paddle.
To treat patients aged 1-8, proceed as follows:
Unscrew the large electrodes on both paddles by turning them anticlockwise.
You now have a reduced electrode surface suitable for children.
Note
Deliver a maximum of 100 joules using the reduced electrode surface!
Using the DefiMonitor XD
40 / 108
DefiMonitor XD | 24050 | GB | I
Fig. 15 Locking the adult paddles in place over the child paddles Locking the adult paddles in place over the child paddles
Place an adult paddle on top of a child paddle.
Tighten the adult paddle by turning it clockwise.
Repeat the process for the second paddle.

7.4.6 Positioning the SavePads and/or paddles on adults

WARNING
Increased patient resistance
Danger of skin burns and too-low energy output
Remove chest hair in those areas where the electrodes are to be positioned. If necessary, dry the skin before attaching the electrodes. Attach the electrodes directly to the skin. Remove any plasters or anything similar. Do not contact any metallic parts with the multifunction electrodes which are connected to the
patient.
Keep a distance between multifunction electrodes and other electrodes.
Fig. 16 anterior - anterior position for adults
For the anterior - anterior position, the SavePads / paddles are positioned:
on the right side of the chest, below the collar bone (1)
on the left side of the chest, above the apex of the heart on the axillary line (2)
Using the DefiMonitor XD
DefiMonitor XD | 24050 | GB | I
41 / 108
Fig. 17 anterior - posterior position for adults
For the posterior - anterior position, the SavePads are positioned:
at the back left of chest, between tip of shoulder blade and spine (1)
at the front left of chest, between sternum and left nipple (2)
Incorrectly positioned electrodes can lead to faulty readings.
Note
Plasters must be removed before using the SavePads.

7.4.7 Positioning the SavePads and/or paddles on children

WARNING
Using adult electrodes on children
Energy output too high for patients aged 1 – 8 (bodyweight <25 kg)
For patients aged 1 – 8 (bodyweight <25 kg), use the SavePads Mini. If no SavePads Mini are available, the device can be used on patients aged 1 – 8 (bodyweight
<25 kg) by using multifunction electrodes for adults.
Do not delay the treatment due to not knowing the patient’s exact age or weight.
Fig. 18 anterior - anterior position for children
For the anterior - anterior position, the SavePads / paddles are positioned:
on the right side of the chest, below the collar bone (1), red SavePad
on the left side of the chest, above the apex of the heart on the axillary line (2) green SavePad
Using the DefiMonitor XD
42 / 108
DefiMonitor XD | 24050 | GB | I
Fig. 19 anterior - posterior position for children
For the posterior - anterior position, the SavePads are positioned:
at the back left of chest, between tip of shoulder blade and spine (1), red SavePad
at the front left of chest, between sternum and left nipple (2) green SavePad
Incorrectly positioned electrodes can lead to faulty readings.
Note
Plasters must be removed before using the SavePads.

7.4.8 Opening and attaching SavePads

Fig. 20 Removing the protective foil from the electrodes (illustration similar)
1
Protective foil
2
Electrode with a layer of gel
Note
The SavePads themselves are coloured so that they can be placed in their correct po­sitions more easily.
Proceed as follows to attach the SavePads to the patient, Fig. 20:
Open the SavePads bag by tearing along the red-coloured groove.
First, remove the protective foil (1) from one of the electrodes (2) and then place the electrode in
the position specified previously (see chapter 7.4.6 and 7.4.7)
Remove the protective foil from the second electrode and place it in its position.
Smooth the electrodes onto the patient ensuring there are no pockets of air under the electrodes.
If you are using the SavePads Mini , insert the electrode connector into the socket of the DefiMon-
itor XD.
If you are using the SavePads Connect or SavePads Mini Connect, it is imperative that you fol- low the procedure described below.
Using the DefiMonitor XD
DefiMonitor XD | 24050 | GB | I
43 / 108
Note
Do not touch the floor, objects, clothing or other body parts with the opened Save­Pads. This could remove the conductive layer of gel on the electrodes.
Connecting SavePads Connect / Mini Connect
Fig. 21 Coded SavePads Connect cable
1
Connector
2
Connecting clips for SavePads Connect / Mini Connect
Pay attention to the following sequence for attaching the SavePads Connect / Mini Connect, Fig. 21 :
Attach the SavePads Connect / Mini Connect as in Chapter 7.4.6 or 7.4.7 described onto the up- per body of the patient.
Connect the SavePads Connect cable to the SavePads Connect / Mini Connect attached to the patient. To do this, open the connecting clips (2) in turn and plug the multifunction electrode con­nection tabs into the respective slit on the connecting clips.
Make sure that the coloured clips are facing upwards.
Snap the upper part of the clips back in place to fix the connection.
Plug the connector of the SavePads Connect cable (1) into the socket on the switched-on Defi-
Monitor XD. Make sure that the markings match up.

7.4.9 Positioning the ECG adhesive electrodes

Fig. 22 ECG patient cable, 4-pin IEC
1
Connector
2
Electrode clips (green, black, red, yellow)
Using the DefiMonitor XD
44 / 108
DefiMonitor XD | 24050 | GB | I
Fig. 23 Positions of the ECG adhesive electrodes on the patient
The positions of the ECG adhesive electrodes are:
(1) red (R)
directly below the centre of the right collar bone (medioclavicular)
(2) yellow (L)
directly below the centre of the left collar bone (medioclavicular)
(3) green (F)
directly below the left chest muscle on the medioclavicular line
(4) black (N)
directly below the right chest muscle on the medioclavicular line

7.5 Removing the electrodes from the patient

Proceed as follows to separate the patient from the DefiMonitor XD:
Remove the connector for the SavePads or ECG electrodes from the DefiMonitor XD.
Slowly remove the electrodes from the patient’s skin.
Dispose of the electrodes in the residual waste.

7.6 Fitting an SpO2 sensor

The approval of the DefiMonitor XD was carried out using the Nellcor™ SpO2 finger sensor FLEXMAX and the Nellcor™ interface cable DOC10. Nonetheless, the option exists to insert all Nellcor™ sensors with OxiMax™ technology. Information regarding the use, warnings, precautionary measures and fur­ther information can be found in the operating instructions for the respective Nellcor™ sensor.
To attach the FLEXMAX sensor models, proceed as follows:
Insert the patient’s index finger into the sensor
Note
If possible, use the sensor on the patient’s index finger. Depending on its size, the
sensor can also be attached to a different finger (with the exception of the thumb).
Ensure that the fingertip has contact with the inner end of the sensor. Guide the sensor cable along the surface of the hand.
Note
Ensure that the sensor is attached the right way round and that the finger doesn’t pro-
trude out of the sensor.
Connect the sensor cable to interface cable DOC10.
Insert the interface cable connector into the socket, Fig. 4 5) of the DefiMonitor XD.
The measurement values are displayed on the DefiMonitor XD monitor.
Note
Check the sensor measurement site at least every 6 hours. Pay attention to pressure injuries and the skin condition. If necessary, place the sensor elsewhere.
Using the DefiMonitor XD
DefiMonitor XD | 24050 | GB | I
45 / 108

7.7 Removing an SpO2 sensor

Removing sensor model FLEXMAX
Press the sides of the sensor together to open the sensor.
Remove the sensor from the finger.
Disconnect the sensor cable from the interface cable.
Remove the interface cable plug from the DefiMonitor XD.
Clean and disinfect the sensor. Please also observe the operating instructions for the sensor.
Store the sensor until it is next used.
The expected shelf life of the sensor is 1 year.

7.8 Alarms and alarm limits

WARNING
Extremely set alarm limits
No alarm in case of a patient's critical state. The patient’s critical state is not recognised. Select purposeful alarm thresholds.
The DefiMonitor XD has an intelligent alarm system. High, medium and low priority alarms exist and also informative notifications and error notifications. If several alarms are triggered simultaneously, all the alarms are displayed on the monitor. However, only the heart alarm signal (heart frequency and VF/VT) is played back. Alarm signalling takes place for all alarms in less than 10 seconds.
Physiological alarms
Physiological alarms are triggered by patient monitoring. Different alarm signals exist for alarms which are triggered via heart monitoring (heart frequency and VF/VT) and for alarms which are triggered via the SpO2 monitoring.
Technical alarms
Technical alarms are triggered by the monitoring of the DefiMonitor XD, e.g. battery charge level low. These alarms are signalised acoustically and optically.
Informative notifications and error notifications
Informative notifications and error notifications are signalised acoustically and optically. If your Defi­Monitor XD has SpO2, the notifications will be displayed instead of the SpO2 curve. The notifications will be shown until the cause is resolved.

7.8.1 Alarms with high priority

A high-priority alarm indicates that it is necessary for the user to react immediately.
Parameter
Status
Notification
Mode
ECG, SpO2 (physiological alarm)
Upper alarm limit for heart rate exceeded
Blinking heart rate value Blinking alarm bell Blinking alarm LED Alarm signal
ECG MON MAN SYNC PACE
Blinking heart rate value Blinking crossed-out
alarm bell
AED Fallen below lower alarm limit for heart rate
Blinking heart rate value Blinking alarm bell
ECG MON MAN
Using the DefiMonitor XD
46 / 108
DefiMonitor XD | 24050 | GB | I
Parameter
Status
Notification
Mode
Blinking alarm LED Alarm signal: High-prior-
ity alarm
SYNC PACE
Blinking heart rate value Blinking crossed-out
alarm bell
AED
VF/VT (physiological alarm)
A life-threatening cardiac arrhythmia has been de­tected
Alarm signal: High-prior­ity alarm
Text message: VF/VT possible, check patient,
Blinking alarm LED
ECG MON
SpO2 (physiological alarm)
Upper ECG alarm limit exceeded for SpO2
Blinking SpO2 value Blinking alarm bell Blinking alarm LED Alarm signal: High-prior-
ity alarm
ECG MON MAN SYNC PACE
Upper ECG alarm limit exceeded for SpO2
Blinking SpO2 value Blinking crossed-out
alarm bell
AED
Fallen below lower SpO2 alarm limit
Blinking SpO2 value Blinking alarm bell Blinking alarm LED Alarm signal: High-prior-
ity alarm
ECG MON MAN SYNC PACE
Fallen below lower SpO2 alarm limit
Blinking SpO2 value Blinking crossed-out
alarm bell
AED
SpO2 sensor no longer attached to the patient (physiological alarm)
The SpO2 sensor was correctly connected but is no longer attached to the patient.
If the condition persists for longer than 10 sec­onds:
Alarm signal Blinking alarm LED Text notification Check
SpO2 Sensor (omitted when in AED mode)
- - -” instead of the SpO2 value
No SpO2 curve No pulse amplitude dis-
play No SpO2 alarm bell No SpO2 alarm limits The following symbol is
displayed in the SpO2 area:
ECG MON MAN SYNC PACE AED
Check the SpO2 sensor’s connection to the patient. Correct it if necessary.

7.8.2 Alarms with medium priority

The following table provides an overview of the alarms with medium priority.
Using the DefiMonitor XD
DefiMonitor XD | 24050 | GB | I
47 / 108
Parameters
Status
Notification
Mode
Battery charge level low
(technical alarm)
The AkuPak LITE XDs charge level is low.
The following icon is dis­played in the status line:
The following icon is dis­played in the status dis­play:
ECG MON MAN SYNC PACE AED
Voice message Charg-
ing status battery low, please recharge After
each CPR cycle
AED
Voice message Charg-
ing status battery low, please recharge
Every 2 minutes
ECG MON MAN SYNC PACE
Charge the AkuPak LITE XD as soon as possible. Connect the DefiMonitor XD to the mains if pos- sible.

7.8.3 Alarms with low priority

The following table provides an overview of the alarms with low priority.
Parameters
Status
Notification
Mode
Time interval for data update is longer than 30 s
(technical alarm)
The measurement of the SpO2 value may be in­correct.
“?” displayed next to the
SpO2 value
ECG MON MAN SYNC PACE AED
Contact the technical services team if necessary.

7.8.4 Informative messages

The following table provides an overview of the informative notifications.
Parameters
Status
Notification
Mode
Key is disabled
ECG electrodes are used and an attempt to change to a therapy mode is made.
Information signal
ECG MON
Shock method via SavePads:
Connect the patient to the DefiMonitor XD using SavePads.
Shock method via paddles:
Select ECG lead via paddles to switch to MAN mode.
SpO2 module notifies faulty SpO2 signal
SpO2 or pulse rate may be incorrect
The measurement of the SpO2 value may be in­correct.
“?” displayed next to the
SpO2 value
ECG MON MAN SYNC PACE AED
Using the DefiMonitor XD
48 / 108
DefiMonitor XD | 24050 | GB | I
Parameters
Status
Notification
Mode
Check the SpO2 sensor.
Check the SpO2 sensor’s connection to the patient
Ensure that the SpO2 sensor is safely connected to the DefiMonitor XD.
SpO2 sensor not con­nected to the DefiMon­itor XD
The SpO2 sensor is not connected to the Defi­Monitor XD.
Text notification Check SpO2 Sensor
(omitted when in AED mode)
- - -” instead of the SpO2 value
No SpO2 alarm bell No SpO2 alarm limits No SpO2 curve No pulse amplitude dis-
play The following symbol is
displayed in the SpO2 area:
ECG MON MAN SYNC PACE AED
SpO2 sensor not at­tached to the patient
The SpO2 sensor has not been attached to the patient.
- - -” instead of the SpO2 value
No SpO2 alarm bell No SpO2 alarm limits
ECG MON MAN SYNC PACE AED
If the condition persists for longer than 10 sec­onds:
Text notification Check SpO2 Sensor
(omitted when in AED mode)
The following symbol is displayed in the SpO2 area:
No SpO2 curve No pulse amplitude dis-
play
ECG MON MAN SYNC PACE AED
SpO2 sensor no longer attached to the patient
The SpO2 sensor was correctly connected but is no longer attached to the patient.
Text notification Check SpO2 Sensor
(omitted when in AED mode)
- - -” instead of the SpO2 value
No SpO2 curve, no empty pulse amplitude display
No SpO2 alarm bell No SpO2 alarm limits
ECG MON MAN SYNC PACE AED
Using the DefiMonitor XD
DefiMonitor XD | 24050 | GB | I
49 / 108
Parameters
Status
Notification
Mode
The following symbol is displayed in the SpO2 area:
If the condition persists for longer than 10 sec­onds:
Alarm signal Blinking alarm LED Text notification Check
SpO2 Sensor (omitted when in AED
mode) - - -” instead of the SpO2
value No SpO2 curve No pulse amplitude dis-
play No SpO2 alarm bell No SpO2 alarm limits The following symbol is
displayed in the SpO2 area:
ECG MON MAN SYNC PACE AED
Check the SpO2 sensor’s connection to the patient. Correct it if necessary.
Pacer treatment was stopped
Pacer treatment was stopped as the energy specified differs from the set value.
Voice message: Pacing stopped
Information signal
PACE
The DefiMonitor XD cannot achieve the set parameter in PACE mode and automatically switches to MAN mode. Pacer treatment is interrupted.
Check the multifunction electrodes.
Pacer treatment was stopped
The pacer treatment was
stopped as the patient’s
connection to the Defi­Monitor XD was inter­rupted.
Voice message: Pacing stopped
Information signal
PACE
The patient cannot be treated using PACE mode as the patient’s connection to the DefiMonitor XD was interrupted.
Check the connection between the patient and the DefiMonitor XD.
Re-establish the connection between the patient and the DefiMonitor XD if necessary.
Pacer self-test not passed
PACE mode is unavaila­ble.
PACE mode cannot be activated.
The following icon is dis­played in the status line:
PACE
Using the DefiMonitor XD
50 / 108
DefiMonitor XD | 24050 | GB | I
Parameters
Status
Notification
Mode
ECG interrupted dur­ing rhythm analysis
The patient moved during the rhythm analysis.
Voice message: Patient movement detected
Restart the rhythm analy­sis
AED
Ensure that the patient is not moved during the rhythm analysis.
Cease chest compressions during rhythm analysis.
Information signal, that the mode has changed from MAN to SYNC or vice-versa.
Mode change from MAN to SYNC.
Mode change from SYNC to MAN.
Information signal
MAN SYNC
The DefiMonitor XD is in manual asynchro­nous or manual syn­chronous mode and the energy for defibril­lation has been fully charged. An attempt is made to change be­tween these two modes.
The energy required for defibrillation is fully charged. No change from MAN to SYNC and vice­versa possible
Information signal Notification: No shock
delivered
MAN SYNC
The DefiMonitor XD is in manual asynchro­nous or manual syn­chronous mode and the energy for defibril­lation has been fully charged. Defibrillation should be made via the paddles.
No valid patient connec­tion at the point in time of defibrillation.
Information signal Notification: No shock
delivered
MAN SYNC
DefiMonitor XD is switched on.
DefiMonitor XD is switched on and ready for operation.
Information signal
ECG MON MAN SYNC AED
DefiMonitor XD is switched off.
DefiMonitor XD is switched off
Information signal
ECG MON MAN SYNC AED
Error Messages
The following table provides an overview of the error notifications.
Internal error
An internal error has been detected.
The DefiMonitor XD is not ready for use.
Voice message: Internal error
ECG MON MAN SYNC PACE AED
The DefiMonitor XD has detected an internal error and is not ready for operation. The DefiMonitor XD turns itself off independently.
Switch the DefiMonitor XD back on. If the self-test is successful, the DefiMonitor XD can be used. If not, please contact the tech-
nical services team.
Using the DefiMonitor XD
DefiMonitor XD | 24050 | GB | I
51 / 108
Parameters
Status
Notification
Mode
ECG in invalid range
The ECG is in the satura­tion region or cannot be displayed.
Dashed line instead of ECG curve
ECG MON MAN SYNC PACE AED
Wait until the ECG signal returns to a valid region.
DefiMonitor XD is ready for defibrillation.
The energy required for defibrillation is fully charged to the DefiMoni­tor XD.
Information signal
MAN AED
Shock via SavePads
Press the Shock button to trigger defibrillation.
Shock via paddles
Press the paddles on the patient’s chest and press both paddle buttons simultaneously.
Internal discharge after 15 seconds
If defibrillation is not triggered within 15 seconds, the energy is discharged internally.
DefiMonitor XD is ready for defibrillation.
The energy required for defibrillation is fully charged to the DefiMoni­tor XD.
Information signal
SYNC
Shock via SavePads
Press the Shock button to trigger defibrillation. The pitch changes during synchronisation.
Shock via paddles
Press the paddles on the patient’s chest and press both paddle buttons simultaneously. The pitch changes during synchronisation.
Internal discharge after 15 seconds
If defibrillation is not triggered within 15 seconds, the energy is discharged internally.
No connection be­tween the patient and the DefiMonitor XD
The ECG electrodes are not attached to the pa­tient or the ECG cable is not connected to the DefiMonitor XD.
The ECG electrodes used are defective.
The ECG electrodes have short-circuited.
Voice message: Plug in
electrode cable Apply electrodes one
after the other to pa­tient's bare chest
ECG MON
Check the ECG electrodes and ensure there is a connection between the patient and the Defi- Monitor XD.
No connection be­tween the patient and the DefiMonitor XD
The connection between the patient and the Defi­Monitor XD has been in­terrupted.
Voice message: Check
electrodes Apply electrodes one after the other to pa­tient's bare chest
AED MAN SYNC PACE
Using the DefiMonitor XD
52 / 108
DefiMonitor XD | 24050 | GB | I
Parameters
Status
Notification
Mode
Check the SavePads and ensure there is a connection between the patient and the DefiMonitor XD.
Printer not ready for operation
An internal error has been detected in the printer.
The following icon is dis­played in the status line:
ECG MON MAN SYNC PACE AED
Please contact the technical services team.
Printer not ready for operation
There is not enough printer paper in the printer.
The following icon is dis­played in the status line:
ECG MON MAN SYNC PACE AED
Refill with printer paper.
Data recording not possible
The SaveCard in the DefiMonitor XD is full.
“CF:FULL” displayed in
the status bar
ECG MON MAN SYNC PACE AED
Archive the data on the SaveCard externally.
Insert an empty SaveCard into the DefiMonitor XD.
Data recording not possible
There is no SaveCard in the DefiMonitor XD.
The SaveCard in the DefiMonitor XD is defec­tive.
“NO CF” displayed in the status bar
ECG MON MAN SYNC PACE AED
Insert a SaveCard into the DefiMonitor XD.
Error in the SpO2 module
An error has been de­tected in the SpO2 mod­ule.
No SpO2 curve - - -” instead of the SpO2
value Text message:SpO2 er-
ror
ECG MON MAN SYNC PACE AED
Error in the SpO2 sen­sor
An error has been de­tected in the SpO2 sen­sor.
- - -” instead of the SpO2 value
Text message:Replace SpO2 Sensor
The following symbol is displayed in the SpO2 area:
ECG MON MAN SYNC PACE AED
Please contact the technical services team.
Using the DefiMonitor XD
DefiMonitor XD | 24050 | GB | I
53 / 108

7.8.5 Messages in AED-Mode

The DefiMonitor XD issues voice messages. The corresponding text version of the message will be shown in the lower area of the monitor. For devices with SpO2 option, the plethysmogram will not be shown in AED mode.
Status / Action
Notification
Voice message
The patient must be connected to the DefiMonitor XD.
Apply electrodes one af­ter the other to patient's bare chest
Apply electrodes one af­ter the other to patient's bare chest
An internal error has been detected. The DefiMonitor XD is not ready for use.
Internal error
Internal error
The patient must not be touched.
Stand clear of patient
Stand clear of patient
The energy required for defibrillation has been charged. Trigger defibrilla­tion by pressing the shock button.
Deliver shock now
Deliver shock now
The ECG analysis has revealed that there is no heart rhythm requiring de­fibrillation.
No shock advised
No shock advised
The ECG analysis has revealed that there is a heart rhythm requiring defib­rillation.
Shock advised
Shock advised
Carry out cardiopulmonary resuscita­tion.
Cardio pulmonary resus­citation
Cardio pulmonary resus­citation
No correct electrode contact. Check whether there are any pockets of air between the electrodes and the skin.
Check electrodes
Check electrodes
The patient may neither be touched nor moved.
Do not touch the patient
Do not touch the patient
An ECG analysis is being carried out
Analysing rhythm
Analysing rhythm
Message from the MMI test. This noti­fication will not appear while the de­vice is being used on the patient.
If you hear this message, press the shock button
If you hear this message, press the shock button
Call emergency services.
Call emergency services
Call emergency services
Ventilate the patient twice.
Give 2 rescue breaths
Give 2 rescue breaths
Perform 30 chest compressions.
Give 30 chest compres­sions
Give 30 chest compres­sions
Perform 15 chest compressions.
Give 15 chest compres­sions
Give 15 chest compres­sions
Insert the SavePads connector or the coded SavePads Connect cable.
Plug in electrode cable
Plug in electrode cable
The AkuPak LITE XD charge level is low. Charge the AkuPak LITE XD if possible.
Charging status battery low, please recharge
Charging status battery low, please recharge
Use the SavePads.
Electrodes not suitable. Please change to defibril­lation electrodes.
Electrodes not suitable. Please change to defibril­lation electrodes.
Switch to AED mode or assess the ECG. There may be a heart rhythm requiring defibrillation.
VF/VT possible, check pa­tient
VF/VT possible, check pa­tient
Using the DefiMonitor XD
54 / 108
DefiMonitor XD | 24050 | GB | I
Status / Action
Notification
Voice message
Perform chest compressions.
Chest compressions
Chest compressions
Patient movement has been detected. Ensure that the patient is not moved.
Patient movement de­tected
Patient movement de­tected
The device is in adult mode.
Adult Mode
Adult Mode
The device is in Pediatric Mode.
Pediatric Mode
Pediatric Mode
The electrodes seem to be out of or­der technically. Replace the elec­trodes if possible.
Invalid electrodes. Please use different electrodes
Invalid electrodes. Please use different electrodes
No defibrillation was delivered to the patient.
No shock delivered
No shock delivered

7.8.6 Alarm volume

The alarm volume can be adjusted as follows:
Parameter
Area
Resolution
Basic setting
ECG alarm volume
0-100%
25%
100%
SpO2 alarm volume
0-100%
25%
100%
Changing the alarm volume
To start the settings menu during operation, press the enter key , Fig. 7 (9).
Move the cursor upwards by pressing the ▲ key several times until the desired parameter is high-
lighted.
Select the highlighted parameter “ECG alarm volume” or “SpO2 alarm volume” by pressing the key.
Change the alarm volume by pressing the ▲ key or ▼ key.
Confirm the selected value by pressing the key.
To exit the settings menu, use the ▲ key or ▼ key to move the cursor to “Exit” and press the
key. The device is now ready for operation again.
The volume of the acoustic alarm can be changed as described in the table.

7.8.7 Alarm limits

The alarm limits can be adjusted as follows:
Parameter
Area
Resolution
Basic setting
ECG alarm limits Lower alarm limit
30 – 300 bpm
1 bpm
50 bpm
ECG alarm limits Upper alarm limit
Lower alarm limit - 300
bpm
1 bpm
100 bpm
SpO2 alarm limits Lower alarm limit
70 - 99 %
1 %
85 %
SpO2 alarm limits Upper alarm limit
Lower alarm limit -
100 %
1 %
100 %
Changing the alarm limits
To start the settings menu during operation, press the enter key , Fig. 7 (9).
Move the cursor upwards by pressing the ▲ key several times until the desired parameter is high-
lighted.
Using the DefiMonitor XD
DefiMonitor XD | 24050 | GB | I
55 / 108
Select the highlighted parameter “ECG alarm limits” or “SpO2 alarm limits” by pressing the key. The lower alarm limit is highlighted.
Change the lower alarm limit by pressing the ▲ key or ▼ key.
Confirm the selected value by pressing the key.
The upper alarm limit is highlighted.
Change this entry as described above. Confirm the selection by pressing the key.
The parameter "ECG alarm volume" is highlighted.
To exit the settings menu, use the ▲ key or ▼ key to move the cursor to “Exit” and press the key.
The device is now ready for operation again.

7.8.8 Muting the alarm

The alarms are always activated when the DefiMonitor XD is switched on. The alarm limits can be ad­justed in the settings menu.
If an alarm occurs, it can be muted for 60s using the alarm acknowledgement button, Fig. 7 (7). The visual alarms will continue to be displayed during this time.
Check the patient and initiate the appropriate measures.

7.9 ECG monitoring (ECG MON)

The ECG shows the electrical activity of the patient’s heart. An ECG can be used to determine the pa­tient’s heart rate or heart rhythm, and detect any arrhythmias. For an ECG, electrodes are placed on the patient’s skin to determine the ECG.
The ECG signal may be briefly disrupted by defibrillation. Make sure that neither the ECG electrodes nor the conductive parts of the electrode clips contact
other conductive parts or the ground. The DefiMonitor XD can record the patient ECG using the following sensors:
4-pin ECG cable
SavePads
Paddles
ECG leads I, II, III, aVR, aVF and aVL can be represented by the 4-pin ECG cable. Lead II can be rep­resented by the SavePads and the paddles.
Lead
Lead formation
I
R - L
II
R - F
III
L - F
aVR
aVF
aVL
It is not possible to switch off pacemaker rejection. If the patient has a pacemaker implanted, the stimulation signal will be shown as follows:
Using the DefiMonitor XD
56 / 108
DefiMonitor XD | 24050 | GB | I
Fig. 24 Display of the suppression of pacemaker pulses

7.9.1 Connecting the ECG electrodes

Note
Do not use any dried-out or out-of-date electrodes.
Note
Attach all four ECG electrodes to the patient's chest to perform ECG monitoring, see chapter 7.4.9.
Connection for ECG adhesive electrodes Fig. 4 (3)
Insert the 4-pin ECG cable connector into the socket, Fig. 4 (3)
Connect the ECG electrodes with the electrode clips.

7.9.2 Changeover of ECG source

The ECG can be recorded via the paddles or via the adhesive electrodes. Pressing the ECG source button, Fig. 7 (5) alternates between the paddles and multifunction electrodes.
Using the DefiMonitor XD
DefiMonitor XD | 24050 | GB | I
57 / 108

7.9.3 Switching the leads on the monitor

Fig. 25 Display leads
If the 4-pin ECG cable is used, the leads of the ECG curves can be selected independent of each other.
By pressing the ▲ (up) key, fig. 7 (9) several times, the upper ECG curve lead (1) can be changed (I, II, III, aVR, aVF, aVL) during operation. The selected lead is displayed in information area 3 (2).
By pressing the ▼ (down) key (9) several times, the second ECG curve lead (4) can be changed (I, II, III, aVR, aVF, aVL) during operation. The selected lead is displayed in information area 3 (3).

7.10 Manual asynchronous / synchronous defibrillation (MAN / SYNC)

Side-effects
The following side-effects may be incurred during or after defibrillation: Frequently
Muscle contractions
Likely
Irritation or burning of the skin in the area of the electrodes
Occasionally
Cardiac arrhythmia (atrial fibrillation or atrial flatter)
Damage to the heart muscle
Chest pain
Note
Separate the patient from other electrical medical devices which are not defibrillation­proof for the defibrillation.
Note
If the settings view is active, you can switch to MAN mode directly by pressing one of the energy levels or the ECG source button.
Using the DefiMonitor XD
58 / 108
DefiMonitor XD | 24050 | GB | I

7.10.1 Energy selection

DANGER
Energy too high for children
Skin burns, current density too high Select 100 joules at most for treating children (bodyweight < 25 kg).
WARNING
Unintentional energy selection
Treatment with incorrect energy Check the energy selection before triggering defibrillation.
Various energy levels are available for manual defibrillation. In adult mode, the following energy levels are avaiable by button selection: 50 J, 70 J, 100 J, 150 J,
200 J, 250 J, 300 J, 360 J. In pediatric mode, the following energy levels are avaiable by button selection: 50 J, 70 J, 100 J. By pressing the arrow button, in adult and pediatric mode, you can select the energy levels 2 J, 5 J, 7
J, 10 J, 20 J, 30 J. The selection is repeated from the beginning again after 2 J. The selected energy level is shown in information area 1 in Fig. 9.

7.10.2 Charging

First choose the shock method.
Note
It is possible to correct the selected energy level using the keypad, Fig. 7 (13).
Shock via paddles
Charge up the selected energy by pressing the paddle button, Fig. 6.
The energy charging progress is shown on the monitor. The energy level shown on the monitor will quickly be available to be delivered as defibrilla-
tion.
If the charging process is complete, a warning sound will be played. The selected charge is available for 15 seconds. If defibrillation does not occur during this pe-
riod, the energy will be discharged internally.
Shock via SavePads
Charge up the selected energy by pressing the charge button, Fig. 7 (12).
The energy charging progress is shown on the monitor. The energy level shown on the monitor will quickly be available to be delivered as defibrilla-
tion.
If the charging process is complete, the green shock button (17) lights up and a warning
sound will be played.
The selected charge is available for 15 seconds. If defibrillation does not occur during this pe-
riod, the energy will be discharged internally.
If SYNC mode was activated beforehand, the defibrillator switches to MAN mode from SYNC
mode.
The energy can also be discharged internally by pressing the charge button (12) again.
Using the DefiMonitor XD
DefiMonitor XD | 24050 | GB | I
59 / 108

7.10.3 Triggering defibrillation

Note
The DefiMonitor XD carries out automatic impedance measurement. If it doesn’t de-
tect the patient, the electrodes and the adhesive contact on the skin need to be checked. Defibrillation can only be carried out if the DefiMonitor XD detects the pa­tient.
Note
Triggering defibrillation can temporarily disrupt the ECG and SpO2 measurement.
7.10.3.1 MAN (manual mode)
After switching the device on and successfully carrying out the self-test, all DefiMonitor XD models will automatically go into manual mode (MAN).
“MAN” mode will be shown in information area 2 of the monitor.
Shock via paddles
Select energy level and charge up as described in chapter 7.10.1 and 7.10.2.
Press the paddles on to the patient’s chest with at least 60 N (6 kg) of force.
Wait until the ECG is visible on the monitor.
Press the two paddle buttons simultaneously to trigger defibrillation, which occurs immediately af-
ter the button is pressed.
Hold the buttons down until the shock has been administered.
Avoid contact with the sockets on the device during defibrillation.
The number of shocks applied is shown in information area 3 of the monitor for 8 seconds.
Shock via SavePads
Select energy level and charge up as described in chapter 7.10.1 and 7.10.2.
Press the lit-up, green shock button, Fig. 7 (17) to trigger defibrillation, which occurs immediately
after the button is pressed.
Hold the button down until the shock has been administered.
Avoid contact with the sockets on the device during defibrillation.
The number of shocks applied is shown in information area 3 of the monitor for 8 seconds,
Fig. 9.
If the defibrillation energy is not delivered to the patient, the corresponding message will be shown in display area 3. (see Fig. 9)
Using the DefiMonitor XD
60 / 108
DefiMonitor XD | 24050 | GB | I
When defibrillation is triggered, the display on the monitor looks like this:
Fig. 26 Manual defibrillation MAN
7.10.3.2 SYNC mode (cardioversion)
WARNING
Incorrect R waves detection
No treatment possible, delivering a shock at the wrong time may lead to ventricular fibrillation Make sure that the R wave detection is correct before charging up the energy.
WARNING
Pressing button for too short a time when triggering synchronous defibrillation
No treatment possible Shock via paddles: Press the paddle buttons on both paddles until synchronous defibrillation
has been carried out.
Shock via SavePads: Press the shock button until synchronous defibrillation has been carried
out.
Synchronisation only occurs via ECG lead II and is conducted via the selected shock method if the DefiMonitor XD detects the R waves and marks them with triangles.
SYNC mode can only be activated via MAN mode. The delay time between detecting a QRS complex (synchronous pulse) and the energy transfer is less
than 60 ms.
Defibrillation in SYNC mode:
Press the SYNC button to get to SYNC mode from MAN mode.
Information area 2 on the monitor now shows SYNC.
Shock method via paddles:
Select the energy level as described in Chapter 7.10.1 .
Press the paddles on to the patient’s chest with at least 60 N (6 kg) of force.
Wait until the ECG is visible on the monitor.
The DefiMonitor XD now highlights the R waves of the ECG.
Make sure that the ECG is free of artefacts.
Charge up the energy level as described in Chapter 7.10.2 .
Using the DefiMonitor XD
DefiMonitor XD | 24050 | GB | I
61 / 108
As long as there is disposition to shock, a beep sounds.
Trigger defibrillation: Keep both paddle buttons pressed until the defibrillation output at the next marked QRS complex is completed.
Until defibrillation is triggered, a changed beep will sound. If the energy is delivered to the patient, the DefiMonitor XD displays this.
Fig. 27 Display for synchronous defibrillation (SYNC) via paddles Shock method via SavePads:
Wait until the ECG is visible on the monitor.
The DefiMonitor XD now highlights the R waves of the ECG.
Make sure that the ECG is free of artefacts.
Select energy level and charge up as described in Chapters 7.10.1 and 7.10.2 .
As long as there is disposition to shock, a beep sounds.
Trigger defibrillation: Press and hold the shock button until defibrillation is carried out.
Until defibrillation is triggered, a changed beep will sound. If the energy is delivered to the patient, the DefiMonitor XD displays this.
Note
If no R waves have been found for synchronising within 15 seconds, the DefiMonitor XD will discharge the energy internally.
Using the DefiMonitor XD
62 / 108
DefiMonitor XD | 24050 | GB | I
When synchronous defibrillation is triggered, the monitor looks like this:
Fig. 28 Display for synchronous defibrillation (SYNC) via SavePads
Note
The DefiMonitor XD automatically switches back to MAN mode following synchronous defibrillation. This is displayed on the monitor accordingly.
Changing from SNYC mode into MAN mode:
Press the SYNC button once again to get to MAN mode.
Information area 2 on the monitor shows MAN. The DefiMonitor XD is in manual mode (MAN).

7.11 SpO2 measurement

WARNING
Only check SpO2 module functionality if the device is switched on
SpO2 module may not be ready for operation Check the display on the monitor
WARNING
Incorrect use of a sensor
Tissue damages
Observe the instructions for use of the sensor. Inspect the measuring point of the sensor regularly. Pay attention to the maximum duration of use of the sensor.
WARNING
Strong magnetic fields
Mutual interference from MRI equipment Do not use the device near MRI equipment
Using the DefiMonitor XD
DefiMonitor XD | 24050 | GB | I
63 / 108
WARNING
Ambient light
Inaccurate measurement of the SpO2 value
Check the environmental conditions. Cover the sensor, if necessary. Check sensor is connected to the patient correctly.
WARNING
Moving the patient, medical measures, external influences such as dysfunctional haemoglo­bin, arterial dyes, low perfusion, dark pigmentation and dyes applied externally such as nail polish or pigment cream
Impairment of pulse oximetry signal quality Eliminate the cause if possible.
WARNING
Constriction of the extremity the SpO2 sensor is on
Potential incorrect measurement due to blood stasis Do not use blood pressure cuffs or other constricting instruments on the extremities the SpO2
sensor is placed on.
WARNING
Use of incompatible accessories
Injury to the patient, no SpO2 measurement possible, incorrect SpO2 measurement results Check the compatibility of the DefiMonitor XD, the interface cable and the sensor.
CAUTION
Do not attach any cable intended for computer use to the sensor port connector.
Note
Only the pulse amplitude display is active in AED mode, in case of high-priority and technical alarms. The SpO2 curve is not displayed.
Note
Triggering defibrillation can temporarily disrupt SpO2 measurement.
SpO2 measurement can be used to determine functional oxygen saturation. The sensor can also be used to determine the pulse rate. If the patient is not connected to the DefiMonitor XD with ECG elec­trodes or via SavePads, the pulse rate will be displayed on the monitor instead of the heart rate. As soon as the heart rate can be determined via the ECG signal (ECG electrodes or SavePads), the source automatically switches back to ECG.
Delivery includes a reusable Nellcor™ SpO2 finger sensor FLEXMAX, which is connected to the Defi­Monitor XD via the Nellcor™ SpO2 interface cable DOC10.

7.11.1 General SpO2 safety advice

Please observe the instructions for use for the SpO2 sensor, including all warnings, risks and in- structions.
Do not use damaged sensors or cables. Do not use sensors whose optical components are ex- posed.
While using the SpO2 sensor, check the signal path on the monitor or the pulse amplitude display before considering the measurement values as correct measurement data.
Using the DefiMonitor XD
64 / 108
DefiMonitor XD | 24050 | GB | I
Do not use blood pressure cuffs or other constricting instruments on the arm the SpO2 sensor is placed on.
Make sure that you do not touch the connector system and the patient at the same time.
Do not use the SpO2 measurement equipment near high-frequency surgical equipment.
Regularly check the sensor and cable for damage. Only use undamaged cables and sensors.
Long cables (sensor or extension cables) may lead to strangulation if they are laid improperly. SpO2 measurement signal quality may be affected during defibrillation or by other electromagnetic in-
terference.

7.11.2 Shown on the monitor

Fig. 29 SpO2 signal display
1
Pulse amplitude display A bar showing the pulse rate and relative pulse amplitude. If the detected pulse becomes stronger, the bar is filled further.
2
SpO2 curve, plethysmographic (pleth) curve This unnormalized curve uses real-time sensor signals which reflect the relative pulse strength of the input signals.

7.12 Pacer (PACE)

DANGER
Too great a deviation (± 30 %) from the selected intensity, treatment automatically stopped
Selected treatment is stopped
Check the multifunction electrodes being used If necessary, replace the multifunction electrodes
Using the DefiMonitor XD
DefiMonitor XD | 24050 | GB | I
65 / 108
WARNING
Inappropriate stimulation intensity or stimulation frequency
Ineffective or faulty stimulation
Regularly check whether the stimulation is effective. Do not leave patient unattended with pacer activated. Comply with the time limit for pacemaker operation on the SavePads packaging.
WARNING
Only check pacer functionality if the device if switched on
Pacer may not be ready for use. Check the information in the status bar.
Side-effects
The following side-effects may be incurred during or after defibrillation: Frequently
Muscle contractions
Likely
Irritation or burning of the skin in the area of the electrodes
Occasionally
Cardiac arrhythmia (atrial fibrillation or atrial flatter)
Damage to the heart muscle
Chest pain
Note
Defibrillation has priority ahead of treatment with a transcutaneous pacemaker. If an energy level is selected during active pacer treatment and the energy has been charged up, pacemaker treatment is stopped and the DefiMonitor XD switches to MAN mode.
Pacer mode can only be activated when the DefiMonitor XD is in MAN mode and multifunction elec­trodes are connected.
Positioning the multifunction electrodes is described in chapter 7.4 . When the pacer is switched-on, information areas 1 and 2 (see Fig. 9) may show “Pacemaker Init” to-
gether with a progress bar. This indicates that the Pacer is currently performing an internal self-test routine. After a few seconds, the Pacer is ready for configuration.
The result of the pacer self-test is shown in the status bar as follows:
Pacer passed self-test
Pacer failed self-test
Note
If it should become necessary to defibrillate the patient during stimulation (pacing), de­fibrillation can be carried out via the SavePads multifunction electrodes.
Note
If more than 3 minutes elapse without the pacer being operated, it switches off auto­matically.
The pacer has three modes:
Using the DefiMonitor XD
66 / 108
DefiMonitor XD | 24050 | GB | I
DEMAND (basic setting)
FIX
OVERDRIVE
Mode
Meaning
DEMAND (basic setting)
Stimulation is only carried out “as required”. I.e. only if the spontaneous heart
rate falls below the set DEMAND rate.
FIX
Fixed-frequency stimulation. A fixed heart rate is imposed independently of the spontaneous heart rate.
OVERDRIVE
Overstimulation of the heart with high-frequency fixed-rate stimulation (max. 250 1/min) to stop, for example, ventricular tachycardia.
DEMAND is automatically activated when the pacer is switched on. The mode is displayed as text on the monitor.
The default settings are reactivated after the pacer is turned off and on again.

7.12.1 Setting the pacer modes

Switching mode:
Push the MODE button, Fig. 7 Press the MODE button, Fig. (2) until the desired mode is dis- played on the monitor.
No stimulation pulse will be administered during mode selection.
Note
The mode cannot be changed while the stimulation is being administered via the pacer. To change the mode, the pacer must first be stopped.
Note
Limited duration of stimulation in OVERDRIVE mode: To prevent dangerous, excessively long stimulation in OVERDRIVE mode, stimulation time without intervention by the user is limited to 15 seconds.

7.12.2 Setting stimulation frequencies

Various stimulation frequencies (number of pacer pulses per minute) are available depending on the pacer mode:
DEMAND, FIX
30 to 180 1/min (beats per minute)
OVERDRIVE
30 to 250 1/min
The following frequency values are preset when the respective modes are activated:
DEMAND, FIX
70 1/min
OVERDRIVE
200 1/min
Procedure:
Using the RATE ppm ▲ (+) and ▼ (-) keys, Fig. 7 (3), the stimulation frequency can be increased or decreased by increments of 5 units (5 1/min) each time the key is pressed. In the ranges shown above.
The stimulation frequency can also be changed during pacing.

7.12.3 Setting stimulation intensities

Various stimulation intensities (current strengths) are available depending on the pacer mode:
DEMAND, FIX, OVERDRIVE
10 to 180 mA
The following current strengths are preset when the respective modes are activated:
Using the DefiMonitor XD
DefiMonitor XD | 24050 | GB | I
67 / 108
DEMAND, FIX, OVERDRIVE
10 mA
Procedure:
Using the OUTPUT mA ▲ (+) and ▼ (-) keys, Fig. 7 (6), the stimulation intensity can be increased or decreased by increments of 5 mA each time the key is pressed.
The stimulation intensity can also be changed during pacing.
The level of stimulation intensity depends on the patient's physical constitution. The current selected should result in the effect of the stimulation being clearly visible on the monitor.
Pacing results in a contraction of the skeletal muscles. This is not a sign of effective heart stimulation.
Watch the ECG on the monitor and the patient’s reaction to the treatment to determine when the inten-
sity is sufficient.

7.12.4 Starting and stopping stimulation in pacer mode (PACE)

Pacer mode must be selected before pacing is started. The mode cannot be changed once pacing has started. The set mode is displayed on the monitor.
Starting stimulation, fig. 7:
Press the Start/Stop button (4). Stimulation is started with the preset values.
A beep is issued to confirm this. The mode display text on the monitor starts to flash. Delivered stimulation pulses are shown
by the LED next to the Start/Stop button (4) lighting up. The stimulation pulses are illustrated as follows:
Fig. 30 Display of activated pacer with stimulation Stopping stimulation, fig. 7:
Press the Start/Stop button (4). This is confirmed by an acoustic signal.
The display text stops flashing.
Stimulation in overdrive mode ends after 15 seconds if neither the pacer intensity (6) nor the pacer rate (3) are changed. If one of these keys is pressed, the 15 seconds are restarted.
Stimulation can be stopped before the 15 seconds have elapsed by pressing the Start/Stop button (4).
Using the DefiMonitor XD
68 / 108
DefiMonitor XD | 24050 | GB | I
7.12.5 Defibrillation during pacing / Defibrillation via multifunction
electrodes
Pacer treatment is automatically interrupted when selecting the energy level. Proceed as described in chapter 7.10. Defibrillation can be performed without modifying the SavePads. Alternatively, the pad­dles can also be used to carry out defibrillation.

7.13 AED-Mode (AED)

DANGER
Disrupted ECG analysis due to movement of the patient
Defibrillation for a non-shockable ECG or no shock recommendation for a shockable ECG
Ensure that the patient is not touched during the ECG analysis. If an ECG analysis is being conducted in transit, the vehicle must be stopped and the motor
switched off.
Switch off the chest compression device for the ECG analysis if one is being used. Cease chest compressions during ECG analysis.
DANGER
Danger of electric shock and too little energy for the patient
Triggering cardiac arrhythmia and burns caused by electric shock
Do not touch the patient during defibrillation. Warn third parties about the dangers of defibrillation. Do not touch any conductive items (metal, blood, water, other liquids, etc.) connected to the pa-
tient during defibrillation.
WARNING
Using adult electrodes on children
Energy output too high for patients aged 1 – 8 (bodyweight <25 kg)
For patients aged 1 – 8 (bodyweight <25 kg), use the SavePads Mini. If no SavePads Mini are available, the device can be used on patients aged 1 – 8 (bodyweight
<25 kg) by using multifunction electrodes for adults.
Do not delay the treatment due to not knowing the patient’s exact age or weight.
Side-effects
The following side-effects may be incurred during or after defibrillation: Frequently
Muscle contractions
Likely
Irritation or burning of the skin in the area of the electrodes
Occasionally
Cardiac arrhythmia (atrial fibrillation or atrial flatter)
Damage to the heart muscle
Chest pain
Note
Separate the patient from other electrical medical devices which are not defibrillation­proof for the defibrillation.
In AED-Mode, the ECG is analysed via an implemented algorithm. This is only possible when using the SavePads.
Using the DefiMonitor XD
DefiMonitor XD | 24050 | GB | I
69 / 108
If heart rhythms potentially requiring defibrillation are detected, the device recommends defibrillation and generates the necessary electric shock for resuscitation when enabled by the user. An electrical shock is not generated if the device does not detect a rhythm requiring defibrillation. The device rec­ommends cardiopulmonary resuscitation.
The resuscitation procedure should be carried out in accordance with the current guidelines of the Eu­ropean Resuscitation Council (ERC) or of the American Heart Association (AHA).
If you have a device with AED-Mode, you must – after switch-on, fig. 7 (10) – start the AED-Mode us­ing the AED button (8). This state is indicated by the illuminated LED above the button.
In AED-Mode, the DefiMonitor XD only enables defibrillation to be triggered after it has detected a shockable rhythm.
In adult mode the energy levels 290J, 340J, 360J are delivered, in pediatric mode the energy levels 50J, 70J, 100J.

7.13.1 Voice messages

You will be asked to examine the patient while the voice messages are played. After the device has successfully carried out the self-test and switched over to the AED-Mode, the fol-
lowing instructions will be given:
Call emergency services Apply electrodes one after the other to patient's bare chest Plug in electrode cable
The message Plug in electrode cable is only played if the electrode connector is not plugged in. The last two voice instructions are repeated for a period of one minute. If the device cannot recognise
a patient impedance/patient at that time, the device will give instructions for cycle of cardiopulmonary resuscitation:
Adult mode
Pediatric mode
Give 30 chest compressions
Give 15 chest compressions
Give 2 rescue breaths
Give 2 rescue breaths
Afterwards, the device will give instructions for attaching the electrodes for a maximum of one minute. This procedure will continue until the device recognises a valid patient impedance / patient and begins the rhythm analysis.
Note
If the patient is connected to the DefiMonitor XD , an ECG analysis will be carried out immediately. In this case, the other voice messages will be skipped.
In basic state, a dashed line appears on the display and the acoustic message Check electrodes is played if the electrodes are not connected. As soon as a circuit is made between the electrodes, the ECG signal appears on the monitor.

7.13.2 Carrying out ECG analysis in AED-Mode

DANGER
Disrupted ECG analysis due to movement of the patient
Defibrillation for a non-shockable ECG or no shock recommendation for a shockable ECG
Ensure that the patient is not touched during the ECG analysis. If an ECG analysis is being conducted in transit, the vehicle must be stopped and the motor
switched off.
Switch off the chest compression device for the ECG analysis if one is being used. Cease chest compressions during ECG analysis.
Using the DefiMonitor XD
70 / 108
DefiMonitor XD | 24050 | GB | I
If the SavePads (multifunction electrodes) have been applied correctly, the device will start the first ECG analysis automatically. Automatic analysis can only be carried out using the multifunction elec­trodes
The patient must now be put in an immobile position and may no longer be touched. The device notifies Do not touch the patient, Analysing rhythm. The DefiMonitor XD will then analyse the ECG. If the device detects a cardiac rhythm requiring defibril-
lation, it will recommend defibrillation. No further ECG analysis is made during energy charging. If the device detects a cardiac rhythm not requiring defibrillation, it will recommend no defibrillation.
The ECG analysis is repeated after 2 minutes of cardiopulmonary resuscitation.

7.13.3 Defibrillation required

WARNING
Pressing the shock button early results in the energy being discharged internally
Delayed treatment Only trigger defibrillation when the shock button lights up and the device prompts you to do so.
Note
In AED mode, the energy for defibrillation is charged up automatically.
If the device detects a heart rhythm requiring defibrillation, it will recommend defibrillation, for which automatic preparations are made inside the device.
The device announces:
Shock advised Chest compressions Metronome
To reduce the time without chest compressions, the metronome is activated during the charging phase. This time span may vary depending on the charge level of the energy module. Carry out the chest compressions until the metronome tone stops.
Once the capacitor is charged internally, power for the defibrillation pulse is available for 15 seconds. This is signalled by a continuous acoustic warning, the voice message
Stand clear of patient Deliver shock now
a continuous tone and the shock button lighting up in "green".
Options for leaving the status of readiness to shock
If the DefiMonitor XD looses the connection to the patient in the status of readiness to shock, the en­ergy is discharged internally. The status of readiness to shock will remain for 15 seconds. If no shock delivery takes place within this period, the energy is discharged internally.
Triggering defibrillation:
Warn those around you loudly before applying the defibrillation!
Press the shock button – lit up green – to apply the shock.
Note
Triggering defibrillation can temporarily disrupt the ECG and SpO2 measurement.
The number of defibrillations carried out will be shown in information area 3 for 10 seconds. Defibrillation and cardiopulmonary resuscitation (CPR) will be repeated on an alternating basis. The charge time of the capacitor for defibrillation depends on the available battery capacity. Charging
may take longer if the power module is partly discharged. The device notifies:
Using the DefiMonitor XD
DefiMonitor XD | 24050 | GB | I
71 / 108
Give 30 chest compressions Give 2 rescue breaths
Furthermore, during the chest compressions, you will be supported by a metronome which will give you the correct frequency for the chest compressions (100 compressions/min).
Note
Once the CPR time has expired (2 mins.), the device returns to ECG analysis.
If the shock is not triggered within 15s,
An internal safety discharge of the defibrillation energy takes place.
the DefiMonitor XD outputs the message No shock delivered
the DefiMonitor XD instructs for cardiopulmonary resuscitation.

7.13.4 Defibrillation not required

If the device cannot find a rhythm requiring defibrillation, it will recommend cardiopulmonary resuscita­tion (CPR).
No shock advised Cardio pulmonary resuscitation Give 30 chest compressions Give 2 rescue breaths
Furthermore, during the chest compressions, you will be supported by a metronome which will give you the correct frequency for the chest compressions (100 compressions/min).
Note
Once the CPR time has expired (2 mins.), the device returns to ECG analysis.

7.14 Keeping the defibrillator ready for use

DANGER
Damaged device or accessories
Treatment not possible, injury to patient, user or third party due to electric shock
Do not use the device or its accessories if it is damaged. Check the status display before using the device.
WARNING
Defibrillator contamination
Defibrillation not possible, patient infection
Clean the defibrillator after every use. Clean the accessories after every use. Disinfect the defibrillator and accessories if necessary. Clean the paddles after use.
WARNING
No treatment / monitoring possible
Empty or faulty energy module
Check the status display regularly. Do not use any faulty or deeply discharged energy modules.
Using the DefiMonitor XD
72 / 108
DefiMonitor XD | 24050 | GB | I
Procedure:
After each use, check the DefiMonitor XD and accessories for damage.
Clean the DefiMonitor XD and accessories after each use.
Disinfect the DefiMonitor XD and the accessories if there is a risk of infection, see chapter 9.1.
Replace disposable accessories.
Check the expiry date of the disposable accessories and replace them if necessary.
Charge or replace the power module if necessary.
Check the expiry date of the internal battery (sticker in the power module slot).
Contact the technical service for replacement.
Perform the MMI test to check the visual and audible alarm signals, see chapter 7.14.1.
In the event of any faults or anomalies, contact our technical service as soon as possible.

7.14.1 MMI test (Man-Machine-Interaction)

The MMI test checks the function of the loudspeaker, the alarm LED and the buttons on the mem­brane keypad of the DefiMonitor XD.
Procedure:
Start the MMI test via the settings on page 3, see paragraph 7.3.2.
Follow the instructions on the monitor.
After the MMI test is successfully completed, the DefiMonitor XD switches back to the settings.

7.15 Event button

By pressing the event button, Fig. 7 (14), a mark is placed on the ECG which causes the ECG to be stored for the 5 seconds before and the 5 seconds after the mark. This ECG sequence can then be printed out later from the event memory.
The printing format for the ECG sequence is always 2-channel print. The signal curves on the monitor at the time of the event are printed out.
Using the DefiMonitor XD
DefiMonitor XD | 24050 | GB | I
73 / 108

7.16 Operating the printer

7.16.1 Inserting paper into the printer

Fig. 31 DefiMonitor XD – right-hand side view
1
Slot for paddle
2
Release lever for printer cover
3
Printer cover
4
Power supply socket
5
Attachment point for bag
Procedure, Fig. 31:
Push down the release lever (2).
Open the printer lid (3) forwards.
Remove the adhesive strip on the roll of paper.
Unwind the paper by approx. 5 cm.
Insert the roll of paper into the printer slot with the chequered side facing up.
Close the printer lid (3).
Note
The printer can only be operated or used in manual mode.

7.16.2 Automatic self-test printout

The self-test printout contains the following parameters:
Model, serial number
Date, time
The selection (off / short / detailed)
Result of the test
If the test was completed successfully, the test will be receive the grade PASS. If the test was not completely successfully, the DefiMonitor XD will turn itself off automatically. No
printout will be possible.
Using the DefiMonitor XD
74 / 108
DefiMonitor XD | 24050 | GB | I
Fig. 32 Self-test printout (similar to illustration)
Fig. 33 Detailed self-test printout (similar to illustration)
Using the DefiMonitor XD
DefiMonitor XD | 24050 | GB | I
75 / 108

7.16.3 ECG signal log

The DefiMonitor XD has a printer. ECG printout of 1 to 6 channels simultaneously is possible. Feed speeds of 25 and 50 mm/s can be selected.
Logging the ECG curve during monitoring:
Pressing the printer button, Fig. 7 (16), causes the log printout to start.
The log printout is stopped by pressing the printer button (16) again.
ECG printout is made with the parameters selected in the settings. The following settings can be se­lected:
Printing parame­ter
Meaning
ECG sensitivity
Print format: 1-channel
Prints the upper ECG channel displayed on the monitor.
5, 10, 15 mm/mV
Print format: 3-channel
Prints the two ECG channels displayed on the monitor. If an SpO2 sensor is connected to the patient, the plethys-
mogram is also printed.
10 mm/mV
Print format: 6-channel
Prints the leads I, II, III, aVR, aVL, aVF simultaneously, de­pending on the electrodes attached, with a maximum of 3 leads, 5 seconds before and the 5 seconds after pressing the button.
5 mm/mV Printing parame-
ter
Meaning
25 mm/s printout speed
Printout is produced at 25 mm/s.
50 mm/s printout speed
Printout is produced at 50 mm/s.
The following relevant parameters are printed out in a header:
Date, time
Speed
Scale
Heart rate
Energy (joules)
Mode
SpO2 value (option: SpO2)
There is a time lag of 7 seconds between the monitor display and the printout, i.e. events which oc­curred before activation of the printout function can be displayed. If the printout is stopped, printout also ends with data recorded 7 seconds previously.
Use the integrated cutting edge on the printer cover to tear off the ECG log printout. Tear off the strip upwards and to the side.
Note
Every printout disrupts energy charging up before defibrillation.
If the ECG lead is changed during printout:
The printing process stops immediately.
The header is rewritten.
The printout then continues.
SaveCard data management
76 / 108
DefiMonitor XD | 24050 | GB | I
Data still stored at the time of the changeover is discarded, the new printout starts at the time of the changeover.

7.16.4 Automatic printout after each shock

The DefiMonitor XD allows you to record the event automatically each time defibrillation or cardiover­sion is performed. Data from 5 seconds before and 5 seconds after the shock was administered is rec­orded.
The “printout on shock” feature can be turned on or off on page 2 of the settings. When the device is
delivered, the function is switched off. The selection remains active after the device is switched off or the battery is changed.

7.16.5 Printing out the event memory

DefiMonitor XD automatically stores the last 30 defibrillations / cardioversions / events in an event memory. For this purpose, the ECG (the 5 seconds before and the 5 seconds after each shock) and the following parameters are stored.
Date, time
Speed
Scale
Heart rate
Energy (joules)
Mode
SpO2 value (option: SpO2)
The contents of the memory will be printed out, beginning with the last event logged.
Procedure, Fig. 7:
Select the parameter “Print events” on page 2 in the settings menu.
Press the enter key (9).
The Memo Print function is activated.
To stop the printout, press the printer button (16).
Printout is produced at 25 mm/s.
The data remains in the event memory after printout and after switching off the DefiMonitor XD. It can be printed out as often as required.
Error messages on the monitor:
Symbol on the monitor
Cause
No paper
Printer error

8 SaveCard data management

The DefiMonitor XD all usage data on a removable SaveCard. The saved data can be displayed using a PC / laptop.
Internal errors are stored on the SaveCard in the file “syserr.txt.”. The file is also available after switch-
ing the DefiMonitor XD off.
Note
If the memory space on the SaveCard is full, no more data can be stored on the Save­Card.
SaveCard data management
DefiMonitor XD | 24050 | GB | I
77 / 108
The device can be operated with full memory space and without a SaveCard.
The data saved on the SaveCard should be externally archived after every use.
Delete the data every time you archive the SaveCard.
The SaveCard supplied with the device is already formatted and can be used straight away. If you have problems with the SaveCard, or with new memory cards, it must be formatted using the FAT 16 or FAT 32 file system.
Proceed as follows for Windows 2000, Windows XP, Windows Vista, Windows 7, Windows 8 and Windows 8.1:
Insert the SaveCard into the memory card slot on the PC / laptop.
Start a command line window using "Start->Run" and enter "cmd.exe” in the entry field.
The command line window will then open.
Now enter the following:
for SaveCards up to 2 GB= format f: /U /FS:FAT /X /V:savecard for SaveCards above 2 GB= format f: /U /FS:FAT32 /X /V:savecard where f: stands for the drive letter of the card reading device – you may need to adjust this.
When starting the device for the first time after formatting a SaveCard, it will take considerably longer because the device performs various self-tests. The device then resumes normal operation again.

8.1 Inserting / replacing the SaveCard

Fig. 34 DefiMonitor XD – open bottom view
1
Slot for SaveCard
2
SaveCard eject button
3
Cover

8.1.1 Inserting SaveCard

Procedure inserting SaveCard, Fig. 34:
Lay the device on its back.
Remove the energy module if inserted. See chapter 6.1.1.
Open the slot by sliding the cover down towards the energy module slot (3).
Push the SaveCard as far as possible into the slot (1).
The ejector (2) sticks out of the opening.
Close the slot by sliding the cover (3) over the shaft until the cover audibly clicks into place.
Insert the energy module again.
The device starts and performs a self-test.
Check the status bar.
The message "CF:" indicating the card capacity in % is displayed:
The SaveCard is inserted correctly.
Cleaning, maintenance and shipping
78 / 108
DefiMonitor XD | 24050 | GB | I
The message "NO CF" is displayed:
The SaveCard probably has no contact with the unit.
In this case, press the ejector (2) and remove the SaveCard.
Insert the SaveCard again as described above.
If the message "NO CF" is still displayed, contact the technical service.

8.1.2 Changing SaveCard

Procedure, Fig. 34:
Lay the device on its back.
Remove the energy module, see chapter 6.1.1.
Open the slot by sliding the cover towards the energy module slot (3).
To remove the SaveCard, press the eject button (2) in fully, which makes the SaveCard protrude
from the slot (1) a little. The SaveCard can now be removed.
Push the SaveCard as far as possible into the slot (1). The ejector (2) sticks out of the opening.
Close the slot by sliding the cover (3) over the shaft until the cover audibly clicks into place.
Insert the energy module again.
The device starts and performs a self-test.
Check the status bar.
The message "CF:" indicating the card capacity in % is displayed:
The SaveCard is inserted correctly.
The message "NO CF" is displayed:
The SaveCard probably has no contact with the unit.
In this case, press the ejector (2) and remove the SaveCard.
Insert the SaveCard again as described above.
If the message "NO CF" is still displayed, contact the technical service.

9 Cleaning, maintenance and shipping

9.1 Cleaning and disinfection

WARNING
Warning: physical harm to user
Risk of electrocution
Only clean the device when switched off. Clean the accessories after every use. Clean the child paddles before screwing on the adult paddles. Use damp cloths to clean.
Clean the device and all its accessories with soap and water solution. Use a slightly damp, clean cloth. Use isopropyl alcohol to disinfect it. For cleaning and disinfecting the SpO2 sensor, please observe the separate instructions for use for
the sensor.
Disposal
DefiMonitor XD | 24050 | GB | I
79 / 108

9.2 Servicing

WARNING
Defibrillator not ready for use due to permanent connection to the mains
Status display indicates device is ready for use even though it is not
Disconnect the defibrillator from the mains. Switch the defibrillator on in battery mode. Allow the defibrillator to operate for approx. 5 minutes. Reconnect the defibrillator to the mains.
The charge level of the AkuPak LITE XD will be re-evaluated and displayed.
Repeat this test every 4 – 6 weeks.
Note
During service or maintenance, the DefiMonitor XD must not be used on a patient.
Note
If fluid leaks from the housing, do not touch the fluid.
Note
The device does not have any parts which can be modified by the user.
Testing after each use
Check your defibrillator, energy module and accessories for damage.
Please contact the technical service for further maintenance.

9.3 Shipping

Observe the current valid dangerous goods guidelines for shipping lithium batteries. If you are shipping the DefiMonitor XD together with the energy module, break the contact between
the energy module and the DefiMonitor XD for shipment. This prevents the device from switching on during transport.

10 Disposal

In accordance with the founding principles of Metrax GmbH, your product has been developed and made using high-quality materials and components which are recyclable.
At the end of its service life, recycle the device via your public waste disposal company. Proper dis­posal of this product is in the interest of protecting the environment.
Through the registration of Metrax GmbH with the responsible authorities, we ensure that the disposal and utilisation of electronics devices brought to the market by us is secure in accordance with the EU directive on the disposal of electronic and electrical equipment (WEEE-directive).
In Germany, in accordance with legislation on the sale, return and environmentally friendly disposal of electrical and electronic devices (Elektro- und Elektronikgerätegesetz – ElektroG), Metrax is registered with EAR under the number 73450404.
For business customers in the European Union Please contact your dealer or supplier if you want to dispose of electrical and electronic equipment.
Technical Data
80 / 108
DefiMonitor XD | 24050 | GB | I

11 Technical Data

11.1 Monitor

Model
LCD monitor
Dimensions
115 x 86 mm (diagonal 144 mm, 5.7“)
Resolution
320 x 240 pixels
Displays
Heart rate, SpO2

11.2 Alarms

LED alarm indicator
Alarm category
Colour
Frequency
Duty cycle
High priority
red
2 Hz
56:44
Alarm signal delay
Signal
Delay
High-priority alarm
< 10 s
Alarms with medium priority
< 10 s
Information signal
< 10 s
Sound level range of the audible alarm signals
Volume
Sound level range
25 %
52 ± 6 dBA
100 %
71 ± 6 dBA

11.3 Physical properties

Dimensions
33 x 16 x 29 cm (W x D x H)
Weight
approx. 5.3 kg (without energy module) approx. 5.8 kg (with energy module)
Protection class of applied parts
defibrillation-proof, type CF
Protection class of casing
Ingress of foreign matter
IP3X protection against solid foreign matter with diameter of 2.5 mm
Ingress of liquid
IPX3 protection against spraying water
Operating mode
Continuous
Classification
Class IIb (MDD Annex IX Rule 09)

11.4 Printer

Model
Thermal printer
Number of channels
1 to 3 channels
Type of paper
Thermal paper
Paper width
58 mm
Printing speed
25 mm/s, 50 mm/s
Technical Data
DefiMonitor XD | 24050 | GB | I
81 / 108

11.5 Memory

Model
Compact Flash Card 2GB

11.6 Voltage supply

Built-in power supply
110 - 240 V, 50 - 60 Hz
Power consumption
110W
Protection class
II for mains usage
AkuPak LITE XD
Lithium ion battery internally provided with current
Lithium-ion technology
Lithium iron phosphate (LiFePO4)
Nominal operating voltage
13.2 V DC
Charging voltage
14.4 V
Max. charging current
5.1 A
Charging cycles
> 1000 (100% SOC)
Periodic replacement of battery
4 years from date of manufacture
Maximum storage period before re­charging
Outside the DefiMonitor XD: 3 months Inserted in the DefiMonitor XD: maximum 1 month
Nominal capacity
2500 mAh
Energy content (100 % SOC)
33 Wh
Charge time (0 – 90% SOC)
Approx. 3h in the DefiMonitor XD All details refer to an ambient temperature of 20°C ± 5°C
Number of shocks at 200J
160
Number of shocks at 360J
95
Operating time (monitoring)
at least 5:00 h (for AkuPak LITE XD in a DefiMonitor XD after three initial shocks in AED mode, ECG/SpO2 moni­toring until device is switched off at 20°C ± 5°C)
Operating time (pacing)
at least 3:20 h hours 20 minutes (for AkuPak LITE XD in a DefiMonitor XD after three initial shocks in AED mode, ECG/SpO2 monitoring and pacing in FIX mode at 70 ppm/100 mA until device is switched off at 20°C ± 5°C)
Note
Please note that the storage conditions have a direct effect on the potential service life of the AkuPak LITE XD and make a considerable contribution to the decision when it needs to be recharged in order to avoid deep discharge.
If possible, store the AkuPak LITE XD at a temperature range of 15°C 35°C. These conditions will provide for maximum storage capacity and service life.
Always be sure to charge the AkuPak LITE XD fully after longer storage before possible use.
All details are for a fully-charged, new AkuPak LITE XD and a temperature of 20 degrees Celsius ± 5 degrees Celsius.
AkuPak LITE XD: The service life of an AkuPak LITE XD is usually up to 4 years or 1,000 charge cy­cles – depending on which occurs first – if the following conditions are met: The AkuPak LITE XD is inserted into the device, the device is only ever in standby mode and is not used, only the regular self­tests recommended by Metrax are carried out, and the ambient temperature is consistently around 23 degrees Celsius (± 2 degrees Celsius). Storing the device outdoors significantly reduces the service
Technical Data
82 / 108
DefiMonitor XD | 24050 | GB | I
life of the AkuPak LITE XD. As a wide variety of factors can influence the service life of the AkuPak, Metrax accepts no liability for the service life of the AkuPak LITE XD.

11.7 Environmental conditions

Operating conditions
Conditions during continuous operation
Temperature
DefiMonitor XD with energy mod­ule
0 °C to +45 °C +32 °F to +113 °F
SavePads
0°C to +50°C +32 °F to +122 °F
ECG electrodes
5°C to +30°C +41 °F bis +86 °F
Printer paper
0°C to +40°C +32 °F to +104°F
SpO2
0 °C to +40 °C +32 °F to +104°F
Humidity
15 % to 95 % non-condensing
Air pressure
620 hPa to 1060 hPa
Transient operating conditions
The DefiMonitor XD can be operated for at least 20 minutes with the conditions specified as follows.
Temperature
DefiMonitor XD with energy mod­ule
-20 °C to +50 °C
-4 °F to +122 °F Humidity
15 % to 95 % non-condensing
Storage conditions
Temperature
-20 °C to +50 °C
-4 °F to +122 °F
Humidity
15 % to 95 % non-condensing
Air pressure
620 hPa to 1060 hPa
Transport conditions (max. 10 days)
Temperature
-25 °C to +50 °C
-13 °F to +122°F
Humidity
15 % to 95 % non-condensing
Air pressure
500 hPa to 1060 hPa
Environmental conditions Nellcor™ SpO2 finger sensor FLEXMAX
Operating temperature range
0 °C to +40 °C 32 °F to 104 °F
Transient operating conditions
The sensor can be operated for 20 minutes at tempera­tures from -20 °C (-4 °F) and +50 °C (122 °F).
Temperature range for storage and transport
-40 °C to +70 °C (-40 °F to 158 °F) Humidity
15 % to 95 % non-condensing
Air pressure
620 hPa to 1060 hPa
Technical Data
DefiMonitor XD | 24050 | GB | I
83 / 108
Stabilisation time (from storage to op­eration)
Up to 20 minutes

11.8 Sound definitions

Alarm signal for high-priority alarms
Volume levels
Can be set to 0 %, 25 %, 50 %, 75 %, 100 %
Pitch
398 Hz – 796 Hz
Pulse width
120 ms
Number of pulses
10 pulses every 2.5 s
Repeated
Every 15 s
Information signal
Volume levels
Can be set to 0 %, 25 %, 50 %, 75 %, 100 %
Pitch
696 Hz
Pulse width
75 ms – 1000 ms
Number of pulses
1 or 2 pulses
Repeated
Not repeated

11.9 Monitoring

11.9.1 ECG

Heart rate
Measurement range
From 30 to 300 bpm
Resolution
1 bpm
Valid QRS amplitude range
0.5 mV to 5 mV
Valid QRS duration range
40 ms to 120 ms
Note: No differentiation is made between adults and children for QRS recognition. The DefiMonitor XD displays a valid heart frequency if the QRS complexes have a duration of 10ms and an ampli­tude of 1mV.
Display update rate
1 s
Accuracy
± 10 % or ± 5 bpm, depending on which is larger
Suppression of large T waves
maximum T wave amplitude 5 mV
Heart rate average determination
The heart frequency displayed on the monitor is an aver­age which is based on the time from a QRS complex peak to the next peak. Normally, this average is based on the data from the previous 10 seconds, however at higher heart frequencies only the pervious 10 times between the QRS peak values are taken into account. The initial heart frequency value appears after some 5 seconds (a maxi­mum of 10 seconds) after the ECG signal is available. The heart frequency is updated after each new QRS complex, however not more often than every 0.5s.
SpO2 value average determination
7 to 20s
Accuracy of the heart frequency dis­play and behaviour in case of irregular rhythms (IEC 60601-2-27:2011+Cor.:2012)
The following values will be displayed after 20 seconds: A1, ventricular bigeminies: 80 bpm A2, slowly changing ventricular bigeminies: 60 bpm
Technical Data
84 / 108
DefiMonitor XD | 24050 | GB | I
A3, fast changing ventricular bigeminies: 118 bpm A4, bidirectional systoles: 90 bpm
Time until alarm in the event of tachy­cardia Upper alarm limit set to 100 bpm Lower alarm limit set to 60 bpm (IEC 60601-2-27:2011+Cor.:2012)
B1, ventricular tachycardia 1 mV peak to valley: 4 s B1, ventricular tachycardia 2 mV peak to valley: (doubled
amplitude): 4 s B1, ventricular tachycardia 0.5 mV peak to valley (halved
amplitude): 6 s B2, ventricular tachycardia 2 mV peak to valley: 3 s B2, ventricular tachycardia 4 mV peak to valley (doubled
amplitude): 3 s B2, ventricular tachycardia 1 mV peak to valley (halved
amplitude): 5 s
Response time for heart rate display after change in heart rate
HR change from 80 to 120 bpm: 9 s HR change from 80 to 40 bpm: 13 s
ECG signal
Leads
I, II, III, aVR, aVL, aVF
Impedance
500 to 2500 Ohms
Power output for the measurement of electrodes which have fallen off
4 µA RMS, 30 kHz, sinusoidal Detection of electrodes that fell off
Detected and shown
Pacemaker suppression region
Effective pacemaker pulses For individual pacemaker pulses and a normally stimulated
QRS and T peak, pulses between ±10 mV and ±700 mV amplitude, with a pulse width between 0.1ms and 2ms and overshooting from 0 to 100ms.
For double pacemaker pulses with 150 ms gaps and a nor­mally stimulated QRS and T peak, pulses between ±10 mV and ±700 mV amplitude, with pulse widths between 0.1 ms and 2 ms and overshooting from 0 to 20ms.
For double pacemaker pulses with 250 ms gaps and a nor­mally stimulated QRS and T peak, pulses between ±10 mV and ±700 mV amplitude, with pulse widths between 0.1ms and 2 ms and overshooting from 0 to 4 ms.
Pacemaker pulses with a stimulated QRS pattern inef­fectively stimulated by the pacemaker
For pacemaker pulses with an ineffective QRS pattern, pulses between ±10 mV and ±700 mV amplitude, with pulse widths between 0.1 ms and 2 ms and overshooting from 0 to 10ms.
For pacemaker pulses and double pacemaker pulses with gaps of 150 ms and 250 ms and an ineffectively simulated QRS pattern, pulses between ±10 mV and ±700 mV ampli­tude, with pulse widths between 0.1 ms and 2 ms and overshooting from 0 to 4 ms.
Pacemaker pulses without a QRS For single pacemaker pulses and double pacemaker
pulses with gaps of 150 ms and 250 ms alone, pulses be­tween ±10 mV and ±700 mV amplitude, with pulse widths between 0.1 ms and 2 ms and overshooting from 0 to 10 ms.
Input
Dynamic input range
± 5 mV AC, ± 300 mV DC
Technical Data
DefiMonitor XD | 24050 | GB | I
85 / 108
Voltage range for detecting QRS com­plexes
± 0.5 mV ~ ± 5 mV QRS complex signal width
40 to 120 ms (Q to S)
Output
Frequency response (monitor)
0.67 to 40 Hz
ECG sensitivity (monitor)
5, 10, 15 mm/mV
Spot velocity
25.0 mm/s
Pacing pulse detection
On
Alarm for electrode separation
Voice message
ECG/paddle input classification
CF, defibrillation-proof
Signal recording via
ECG electrodes, multifunction electrodes or paddles

11.9.2 SpO2

Measurement Range
Pulse oximetry saturation SpO2
1 % - 100 %
Pulse rate measurement range
Source SpO2
20 – 250 bpm
Wavelength red
660 nm
Output power1
< 5 mW
Wavelength infrared
885 nm focus
Output power1
< 5 mW
Power dissipation
52,5 mW
SpO2 input classification
CF, defibrillation-proof
SpO2 measurement accuracy
Adult
2, 3
70 - 100 % ± 2 digits
Adult and Neonate Low Sat
2, 3, 4
60 - 80 % ± 3 digits
Neonate
4, 5
70 - 100 % ± 2 digits
Low Perfusion6
70 - 100 % ± 2 digits
Adult and Neonate with Motion
2, 7
70 - 100 % ± 3 digits
Pulse rate measurement accuracy
Adult and Neonate
2, 3, 4
20 - 250 ± 3 bpm
Low Perfusion 6
20 - 250 ± 3 bpm
Adult and Neonate with Motion
2, 7
48 - 127 ± 5 bpm
1
Saturation accuracy varies by sensor type. Refer to the Sensor Accuracy Grid at
www.covidien.com/rms.
2
Accuracy specifications were validated using measurements of healthy non-smoking adult volunteers during controlled hypoxia studies spanning the specified saturation ranges. Subjects were recruited from the local population and comprised both men and women ranging in age from 18-50 years old, and spanned a range of skin pigmentations. Pulse oximeter SpO2 readings were compared to SaO2 values of drawn blood samples measured by hemoximetry. All accuracies are expressed as ±1 SD. Because pulse oximeter equipment measurements are statistically distributed, about two-thirds of the measurements can be expected to fall in this accuracy (ARMS) range (refer to the Sensor Accuracy Grid for more details).
Technical Data
86 / 108
DefiMonitor XD | 24050 | GB | I
3
Adult specifications are shown for OxiMax MAXA and MAXN sensors with the Nellcor™ Bedside Respiratory Patient Monitoring System.
4
Neonate specifications are shown for OxiMax MAXN sensors with the Nellcor™ Bedside Respiratory Patient Monitoring System.
5
Clinical functionality of the MAXN sensor has been demonstrated on a population of hospitalized ne­onate patients. The observed SpO2 accuracy was 2.5% in a study of 42 patients with ages of 1 to 23 days, weight from 750 to 4,100 grams, and 63 observations made spanning a range of 85% to 99% SaO2.
6
Specification applies to Nellcor™ Bedside Respiratory Patient Monitoring System oximeter perfor- mance. Reading accuracy in the presence of low perfusion (detected IR pulse modulation amplitude
0.03% - 1.5%) was validated using signals supplied by a patient simulator. SpO2 and pulse rate val­ues were varied across the monitoring range over a range of weaksignal conditions and compared to the known true saturation and pulse rate of the input signals.
7
Motion performance was validated during a controlled hypoxia blood study. Subjects performed rub­bing and tapping movements 1-2 cm in amplitude with aperiodic intervals (randomly changing) with a random variation in frequency between 1-4 Hz. Applicability: OxiMax MAXA, MAXAL, MAXP, MAXI, and MAXN sensors.

11.10 Treatment parameters

Charging time (in accordance with IEC 60601-2-4)
Manual mode
Maximum time until ready for defibril­lation of 360J
12 ± 3 s when run at rated voltage from mains 12 ± 3 s when operating at 90 % rated voltage 12 ± 3 s when run with fully charged, new AkuPak LITE XD 12 ± 3 s when run with an AkuPak LITE XD after discharge
of energy for 15 shocks
Maximum time from switching on until ready for defibrillation of 360J
≤ 25 s when operated at rated voltage from the mains ≤ 25 s when operated at 90 % of rated voltage ≤ 25 s when operated with a fully charged, new AkuPak
LITE XD
≤ 25 s when operated with an AkuPak LITE XD after dis-
charge of energy for 15 shocks
AED-Mode
Maximum time between the start of analysis and readiness for defibrilla­tion of 360J
≤ 30 s when operated at rated voltage from the mains ≤ 30 s when operated at 90 % of rated voltage ≤ 30 s when operated with a fully charged, new AkuPak
LITE XD ≤ 30 s when operated with an AkuPak LITE XD after dis-
charge of energy for 15 shocks
Maximum time between switching on and readiness for defibrillation of 360J
≤ 33 s when operated at rated voltage from the mains ≤ 33 s when operated at 90 % of rated voltage ≤ 33 s when operated with a fully charged, new AkuPak
LITE XD
≤ 33 s when operated with an AkuPak LITE XD after dis-
charge of energy for 15 shocks

11.10.1 Biphasic curve properties

Pulse length
Positive phase 11.25 ms, negative phase 3.75 ms
Pulse shape
Biphasic, current-regulated
Technical Data
DefiMonitor XD | 24050 | GB | I
87 / 108
Output energy in AED mode (adult mode) for
Patient im­pedance
1st stage
2nd stage
3rd stage
Tolerance 25 Ohm
150J
220J
290J
± 15 %
50 Ohm
290J
340J
360J
± 15 %
75 Ohm
330J
340J
340J
± 15 %
100 Ohm
320J
320J
320J
± 15 %
125 Ohm
296J
296J
296J
± 15 %
150 Ohm
274J
274J
274J
± 15 %
175 Ohm
250J
250J
250J
± 15 %
Output energy in AED mode (pae­diatric mode) for
Patient im­pedance
1st stage
2nd stage
3rd stage
Tolerance
25 Ohm
41J
55J
81J
± 15 %
50 Ohm
50J
70J
100J
± 15 %
75 Ohm
49J
64J
96J
± 15 %
100 Ohm
44J
60J
89J
± 15 %
125 Ohm
42J
56J
83J
± 15 %
150 Ohm
39J
51J
77J
± 15 %
175 Ohm
36J
48J
71J
± 15 %
Output energy in manual mode (adult)de­pending on pa­tient im­pedance
Energy
to:
25 Ω
50 Ω
75 Ω
100 Ω
125 Ω
150 Ω
175 Ω
Toler-
ance
2J
2J
2J
2J
2J
2J
2J
1J
0,5J -
5J
5J
4J
5J
5J
4J
4J
4J
4J
± 3 J
7J
6J
7J
7J
6J
6J
5J
5J
± 3 J
10J
8J
10J
9J
9J
8J
8J
7J
± 3 J
20J
16J
20J
19J
18J
17J
15J
14J
± 3 J
30J
25J
30J
29J
27J
25J
23J
21J
± 15 %
50J
41J
50J
49J
44J
42J
39J
36J
± 15 %
70J
55J
70J
64J
60J
56J
51J
48J
± 15 %
100J
81J
100J
96J
89J
83J
77J
71J
± 15 %
150J
122J
150J
143J
134J
123J
115J
106J
± 15 %
200J
165J
200J
192J
179J
166J
153J
143J
± 15 %
250J
205J
250J
239J
224J
208J
192J
178J
± 15 %
300J
244J
300J
287J
268J
249J
230J
214J
± 15 %
360J
288J
360J
337J
315J
291J
269J
250J
± 15 %
Technical Data
88 / 108
DefiMonitor XD | 24050 | GB | I
Output energy in man­ual mode (paedi­atric) de­pend­ing on patient imped­ance
Energy
to:
25 Ω
50 Ω
75 Ω
100 Ω
125 Ω
150 Ω
175 Ω
Toler-
ance
2J
2J
2J
2J
2J
2J
2J
1J
0,5J -
5J
5J
4J
5J
5J
4J
4J
4J
4J
± 3 J
7J
6J
7J
7J
6J
6J
5J
5J
± 3 J
10J
8J
10J
9J
9J
8J
8J
7J
± 3 J
20J
16J
20J
19J
18J
17J
15J
14J
± 3 J
30J
25J
30J
29J
27J
25J
23J
21J
± 15 %
50J
41J
50J
49J
44J
42J
39J
36J
± 15 %
70J
55J
70J
64J
60J
56J
51J
48J
± 15 %
100J
81J
100J
96J
89J
83J
77J
71J
± 15 %
Note: With SavePads Mini you cannot select more than 100J.

11.10.2 Manual defibrillation (asynchronous / synchronous)

Impedance range
23 Ω – 200 Ω
Measurement frequency impedance
30 kHz
Energy levels adult mode
2 J, 5 J,7 J, 10 J, 20 J, 30 J, 50 J, 70 J, 100 J, 150 J, 200 J, 250 J, 300 J, 360 J
Energy levels pediatric mode
2 J, 5 J,7 J, 10 J, 20 J, 30 J, 50 J, 70 J, 100 J
Time until internal discharge
15 s
Maximum delay time between syn­chronisation pulse and energy output This delay time is the time from the R­wave to the onset of the defibrillation impulse.
60 ms
Time until charged up to 360 J
12 s
Shock method
Paddles or multifunction electrodes SavePads (Connect),
for pediatric patients: multifunction electrodes SavePads Mini

11.10.3 Pacer

Impedance range
23 Ω – 200 Ω
Measurement frequency impedance
30 kHz
Modes
FIX, DEMAND, OVERDRIVE
Stimulation frequency Fix, Demand Overdrive
30 ppm – 180 ppm 30 ppm – 250 ppm
Stimulation frequency accuracy
± 0.5 %
Energy output via
Multifunction electrodes
Stimulation intensity
10 mA – 180 mA
Stimulation intensity accuracy
± 10 % or + 3/-1 mA
Pulse width
20 ms
Technical Data
DefiMonitor XD | 24050 | GB | I
89 / 108
Pulse width accuracy
± 100 µs
Refractory period
340 ms for a stimulation frequency < 100 bpm 240 ms for a stimulation frequency ≥ 100 bpm

11.10.4 AED-Mode

Impedance range
23 Ω – 200 Ω
Measurement frequency impedance
30 kHz
Shock method
Multifunction electrodes for adults or children
Asystole threshold
≥ 200 µV.
Analysis duration
4 - 20 s
Adult mode energy stages to 50 Ω
290J, 340J, 360J
Pediatric mode energy stages to 50 Ω
50J, 70J, 100J
Sensitivity
> 90 %
Specificity
> 95 %
Real predictive value
> 90 %
False positive rate
< 5 %
ECG rhythm for determining whether a shock should be delivered
Ventricular fibrillation at an amplitude greater than or equal to 0.2 mV
Ventricular tachycardia at a heart rate greater than or equal to 160 bpm
Subject to change without notice.

11.11 Multifunction electrodes (SavePads)

SavePads Connect
Max. 50 shocks at 360 J
Max. 24 hours of monitoring
Max. 1 hour of pacing at 140 mA / 120 ppm (pulse duration 20 ms)
Max. 8 hour of pacing at 70 mA / 60 ppm (pulse duration 20 ms)
Check the multifunction electrodes every 30 minutes
Electrode shape
Rectangular
Total surface area
Approx. 125 cm²
Adhesive surface area
Approx. 121 cm²
Gel surface area / active surface area
Approx. 87 cm²
Gel thickness
0.60 ± 0.10 mm
Pieces / bag
1 set (2 pieces)
Carrier material
Adhesive PE foam
Conductive material
Tin
Gel
Adhesive hydrogel
Separating film
Siliconised PET film
Packaging material
PET, Al, PE
Cable length
N/A
Technical Data
90 / 108
DefiMonitor XD | 24050 | GB | I
SavePads Mini Connect
Developed for patients with a maximum bodyweight of 25 kg and maximum age of 8 years.
Max. 25 shocks at 100 J
Max. 8 hours of monitoring
Max. 1 hour of pacing at 140 mA / 140 ppm (pulse duration 20 ms)
Check the multifunction electrodes every 30 minutes
Electrode shape
Rectangular
Total surface area
Approx. 80 cm²
Adhesive surface area
Approx. 75 cm²
Gel surface area / active surface area
Approx. 42 cm²
Gel thickness
0.60 ± 0.10 mm
Pieces / bag
1 set (2 pieces)
Carrier material
Adhesive PE foam
Conductive material
Tin
Gel
Adhesive hydrogel
Separating film
Siliconised PET film
Packaging material
PET, Al, PE
SavePads Mini Connect cable length
N/A
SavePads PreConnect
Max. 50 shocks at 360 J
Max. 24 hours of monitoring
Max. 1 hour of pacing at 140 mA / 120 ppm (pulse duration 20 ms)
Max. 8 hour of pacing at 70 mA / 60 ppm (pulse duration 20 ms)
Check the multifunction electrodes every 30 minutes
Electrode shape
Rectangular
Total surface area
Approx. 148 cm²
Adhesive surface area
Approx. 145 cm²
Gel surface area / active surface area
Approx. 87 cm²
Gel thickness
0.60 ± 0.10 mm
Pieces / bag
1 set (2 pieces)
Carrier material
Adhesive PE foam
Conductive material
Tin
Gel
Adhesive hydrogel
Separating film
Siliconised PET film
Packaging material
PET, Al, PE
SavePads PreConnect cable length
2 m
SavePads Mini
Developed for patients with a maximum bodyweight of 25 kg and maximum age of 8 years.
Appendix
DefiMonitor XD | 24050 | GB | I
91 / 108
Max. 25 shocks at 100 J
Max. 8 hours of monitoring
Max. 1 hour of pacing at 140 mA / 140 ppm (pulse duration 20 ms)
Check the multifunction electrodes every 30 minutes
Electrode shape
Oval
Total surface area
Approx. 75 cm²
Adhesive surface area
Approx. 74 cm²
Gel surface area / active surface area
Approx. 43 cm²
Gel thickness
0.60 ± 0.10 mm
Pieces / bag
1 set (2 pieces)
Carrier material
Adhesive PE foam
Conductive material
Tin
Gel
Adhesive hydrogel
Separating film
Siliconised PET film
Packaging material
PET, Al, PE
SavePads Mini cable length
Approx. 1.2 m

12 Appendix

12.1 Illustration of time-current curves

The following diagrams show the graphs for the defibrillation pulse displayed depending on the load resistance:
Maximum output
Fig. 35 Defibrillation with a maximum output of 360 joules
Appendix
92 / 108
DefiMonitor XD | 24050 | GB | I
AED adult mode
Fig. 36 First defibrillation in AED adult mode
Fig. 37 Second defibrillation in AED adult mode
Appendix
DefiMonitor XD | 24050 | GB | I
93 / 108
Fig. 38 Third and subsequent defibrillations in AED adult mode
AED pediatric mode
Fig. 39 First defibrillation in AED pediatric mode
Appendix
94 / 108
DefiMonitor XD | 24050 | GB | I
Fig. 40 Second defibrillation in AED pediatric mode
Fig. 41 Third and subsequent defibrillations in AED pediatric mode

12.2 General advice for using pulse oximeters

12.2.1 Data Update Period, Data Averaging, and Signal Processing

The advanced signal processing of the OxiMax™ algorithm automatically extends the amount of data
required for measuring SpO2 and pulse rate depending on the measurement conditions. The Oxi­Max™ algorithm automatically extends the dynamic averaging time required beyond seven (7) sec­onds during degraded or difficult measurement conditions caused by low perfusion, signal artifact, am­bient light, electrocautery, other interference, or a combination of these factors, which results in an in­crease in the dynamic averaging to 20 seconds.

12.2.2 Functional versus Fractional Saturation

This monitoring system measures functional saturation where oxygenated hemoglobin is expressed as a percentage of the hemoglobin that can transport oxygen. It does not detect significant amounts of dysfunctional hemoglobin, such as carboxy-hemoglobin or methemoglobin. In contrast, hemoximeters such as the IL482 report fractional saturation where oxygenated hemoglobin is expressed as a per­centage of all measured hemoglobin, including measured dysfunctional hemoglobins. To compare
Appendix
DefiMonitor XD | 24050 | GB | I
95 / 108
functional saturation measurements to those from a monitoring system that measures fractional satu­ration, fractional measurements must be converted using the listed equation.
Φ functional saturation
η % carboxyhemoglobin
φ fractional saturation
Λ % methemoglobin

12.2.3 Performance Considerations

This section contains information for optimizing the performance of the monitoring system. Prior to initial installation in a clinical setting, have a qualified service technician verify the performance
of the monitoring system per the Service Manual.
Patient Conditions
Application issues and certain patient conditions can affect the measurements of the monitoring sys­tem and cause the loss of the pulse signal.
Anemia Anemia causes decreased arterial oxygen content. Although SpO2 readings may ap-
pear normal, an anemic patient may be hypoxic. Correcting anemia can improve arterial oxygen content. The monitoring system may fail to provide an SpO2 reading if hemoglobin levels fall be­low 5 gm/dl.
Dysfunctional hemoglobins Dysfunctional hemoglobins such as carboxyhemoglo-bin, methemo-
globin, and sulphemoglobin are unable to carry oxygen. SpO2 readings may appear normal; how­ever, a patient may be hypoxic because less hemoglobin is available to carry oxygen. Further as­sessment beyond pulse oximetry is recommended.
Additional possible patient conditions may also influence measurements.
Poor peripheral perfusion Excessive patient movement Venous pulsations Dark skin pigment Intravascular dyes, such as indocyanine green or methylene blue Externally applied coloring agents (nail polish, dye, pigmented cream) Defibrillation
Sensor Performance Considerations Inaccurate Sensor Measurement Conditions
A variety of conditions can cause inaccurate Nellcor™ pulse oximetry sensor mea-surements.
Incorrect application of the pulse oximetry sensor
Placement of the pulse oximetry sensor on an extremity with a blood pressure cuff, arterial cathe-
ter, or intravascular line
Ambient light
Failure to cover the pulse oximetry sensor site with opaque material in high ambient light condi-
tions
Excessive patient movement
Dark skin pigment
Intravascular dyes or externally applied coloring, such as nail polish or pigmented cream
Appendix
96 / 108
DefiMonitor XD | 24050 | GB | I
Signal Loss
Loss-of-pulse signal can occur for several reasons:
Pulse oximetry sensor applied too tightly
Inflation of a blood pressure cuff on the same extremity as the attached pulse oximetry sensor
Arterial occlusion proximal to the pulse oximetry sensor
Poor peripheral perfusion
Recommended Usage
Select an appropriate Nellcor™ pulse oximetry sensor, apply it as directed, and observe all warnings
and cautions presented in the Instructions for Use accompanying the sensor. Clean and remove any substances such as nail polish from the application site. Periodically check to ensure that the sensor remains properly positioned on the patient.
High ambient light sources such as surgical lights (especially those with a xenon light source), bilirubin lamps, fluorescent lights, infrared heating lamps, and direct sunlight can interfere with the performance of a Nellcor™ pulse oximetry sensor. To prevent interference from ambient light, ensure that the sen­sor is properly applied, and cover the sensor site with opaque material.
If patient movement presents a problem, try one or more of the following remedies to correct the problem:
Verify the Nellcor™ pulse oximetry sensor is properly and securely applied.
Move the sensor to a less active site.
Use an adhesive sensor that improves patient skin contact.
Use a new sensor with fresh adhesive backing.
Keep the patient still, if possible.
If poor perfusion affects performance, consider using the Nellcor™ forehead SpO2sensor (MAXFAST).
12.2.3.1 Sensor Performance Considerations
Inaccurate Sensor Measurement Conditions
A variety of conditions can cause inaccurate Nellcor™ pulse oximetry sensor measurements.
Incorrect application of the pulse oximetry sensor
Placement of the pulse oximetry sensor on an extremity with a blood pressure cuff, arterial cathe-
ter, or intravascular line
Ambient light
Failure to cover the pulse oximetry sensor site with opaque material in high ambient light condi-
tions
Excessive patient movement
Dark skin pigment
Intravascular dyes or externally applied coloring, such as nail polish or pigmented cream
Signal Loss
Loss-of-pulse signal can occur for several reasons:
Pulse oximetry sensor applied too tightly
Inflation of a blood pressure cuff on the same extremity as the attached pulse oximetry sensor
Arterial occlusion proximal to the pulse oximetry sensor
Poor peripheral perfusion
Recommended Usage
Select an appropriate Nellcor™ pulse oximetry sensor, apply it as directed, and observe all warnings and cautions presented in the Instructions for Use accompanying the sensor. Clean and remove any substances such as nail polish from the application site. Periodically check to ensure that the sensor remains properly positioned on the patient.
Appendix
DefiMonitor XD | 24050 | GB | I
97 / 108
High ambient light sources such as surgical lights (especially those with a xenon light source), bilirubin lamps, fluorescent lights, infrared heating lamps, and direct sunlight can interfere with the performance of a Nellcor™ pulse oximetry sensor. To prevent interference from ambient light, ensure that the sen­sor is properly applied, and cover the sensor site with opaque material.
If patient movement presents a problem, try one or more of the following remedies to correct the problem:
Verify the Nellcor™ pulse oximetry sensor is properly and securely applied.
Move the sensor to a less active site.
Use an adhesive sensor that improves patient skin contact.
Use a new sensor with fresh adhesive backing.
Keep the patient still, if possible.
If poor perfusion affects performance, consider using the Nellcor™ forehead SpO2sensor
(MAXFAST).
12.2.3.2 Oximetry Considerations
Pulse Rate
The monitoring system only displays pulse rates between 20 and 250 bpm. Detected pulse rates above 250 bpm appear as 250. Detected pulse rates below 20 appear as a zero (0).
Saturation
The monitoring system displays saturation levels between 1% and 100%.

12.2.4 Nellcor™ Pulse Oximetry Sensors

When selecting a Nellcor™ sensor, you should take the patient’s weight and activities, the appropri­ateness of the perfusion and the available sensor locations, the necessity of sterility and the likely du­ration of monitoring into account. Use the recommended sensor operating instructions in order to sim­plify the selection of the sensor or contact Covidien or a local Covidien representative. Reference Chap. 12.2.3. The Nellcor™ SPO2 interface cable DOC10 connects the monitoring system to the Nellcor™ sensor. Do not connect any cable to the sensor connection which is intended for computer use. Only use sensors and interface cables authorised by Covidien when connecting to the sensor connection.
Nellcor™ pulse oximetry sensor models and patient weight
Nellcor™ pulse oximetry sensor
Article description
Patient weight
Nellcor™ SpO2 sensor for premature babies,
non-adhesive (for single-patient use only)
SC-PR
<1.5 kg
Nellcor™ SpO2 sensor for newborn babies, non­adhesive (for single-patient use only)
SC-NEO
1.5 to 5 kg
Nellcor™ SpO2 sensor for adults, non-adhesive (for single-patient use only)
SC-A
>40 kg
Nellcor™ SpO2 sensor with a poultice for adults
and newborn infants (reusable with adhesives)
OXI-A/N
<3 or >40 kg
Nellcor™ SpO2 sensor with a poultice for adults
and newborn infants (reusable with adhesives)
OXI-P/I
3 to 40 kg
Nellcor™ SpO2 sensor for children, two-part (sterile, not reusable)
P
10 to 50 kg
Nellcor™ SpO2 sensor for newborn infants and
adults, two-part (sterile, not reusable)
N
<3 or >40 kg
Nellcor™ SpO2 sensor for adults, two-part (ster­ile, not reusable)
A
>30 kg
Appendix
98 / 108
DefiMonitor XD | 24050 | GB | I
Nellcor™ pulse oximetry sensor
Article description
Patient weight
Nellcor™ SpO2 sensor for newborn infants and
adults (sterile, not reusable)
MAXN
<3 or >40 kg
Nellcor™ SpO2 sensor for babies (sterile, not
reusable)
MAXI
3 to 20 kg
Nellcor™ SpO2 sensor for children (sterile, not reusable)
MAXP
10 to 50 kg
Nellcor™ SpO2 sensor for adults (sterile, not re-
usable)
MAXA
>30 kg
Nellcor™ SpO2 sensor for adults with a long ca-
ble (sterile, not reusable)
MAXAL
>30 kg
Nellcor™ SpO2 nasal sensor for adults (sterile,
not reusable)
MAXR
>50 kg
Nellcor™ SpO2 forehead sensor (sterile, not re-
usable)
MAXFAST
>10 kg
Nellcor™ SpO2 sensor for adults, reusable (not
sterile)
DS-100A
>40 kg
Nellcor™ SpO2 sensor for adults and children
(except for babies and newborn infants), reusa­ble (not sterile)
FLEXMAX Large
>20 kg
Nellcor™ SpO2 sensor for adults and children
(except for babies and newborn infants), reusa­ble (not sterile)
FLEXMAX-P Small
>20 kg
Nellcor™ SpO2 sensor for various locations, re-
usable (not sterile)
D-YS
>1 kg Nellcor™ SpO2 ear sensor clip (not sterile)
D-YSE
>30 kg
Nellcor™ SpO2 sensor clip for children, reusa-
ble (not sterile)
D-YSPD
3 to 40 kg
Contact Covidien or a local Covidien representative for a Nellcor™ Oxygen Saturation Accuracy Spec-
ification Grid, which lists all the Nellcor™ sensors used with the monitoring system. Covidien retains a soft copy under www.covidien.com.
Note
Physiological conditions such as exaggerated patient movement, medical processes or external substances such as dysfunctional haemoglobin, arterial dyestuffs, low per­fusion, dark pigment and externally applied dyestuffs such as nail polish, dyestuffs or pigmented cream may influence the capability of the monitoring system to recognise and show the capability of the monitoring system.
Nellcor™ sensor functions
The Nellcor™ sensor functions vary with sensors which have different revision statuses and by sensor type (adhesive, recycled and reusable). A sensor’s revision status can be found on the sensor con-
nector.
Biocompatibility test
The Nellcor™ pulse oximetry sensors were subjected to a biocompatibility examination in accordance
with ISO 10993-1 (Biological evaluation of medical devices, part 1: Evaluation and testing). The pulse oximetry sensors successfully passed the recommended biocompatibility tests and thus meet the re­quirements of ISO10993-1.
Appendix
DefiMonitor XD | 24050 | GB | I
99 / 108

12.2.5 Functional Testers and Patient Simulators

Some models of commercially available bench top functional testers and patient simulators can be used to verify the proper functionality of Covidien Nellcor™ monitoring systems, sensors, and cables. Reference the individual testing device's operator's manual for the procedures specific to the model of tester used. While such devices may be useful for verifying that the sensor, cabling, and monitoring system are functional, they are incapable of providing the data required to properly evaluate the accu­racy of a system's SpO2 measurements.
Fully evaluating the accuracy of the SpO2 measurements requires, at a minimum, accommodating the wavelength characteristics of the sensor and reproducing the complex optical interaction of the sensor
and the patient’s tissue. These capabilities are beyond the scope of known bench top testers. SpO2
measurement accuracy can only be evaluated in vivo by comparing monitoring system readings with values traceable to SaO2 measurements obtained from simultaneously sampled arterial blood using a laboratory CO-oximeter.
any functional testers and patient simulators have been designed to interface with the monitoring sys­tem's expected calibration curves and may be suitable for use with monitoring systems and/or sen-
sors. Not all such devices, however, are adapted for use with the OxiMax™ digital calibration system.
While this will not affect use of the simulator for verifying system functionality, displayed SpO2 meas­urement values may differ from the setting of the test device. For a properly functioning monitoring system, this difference will be reproducible over time and from monitoring system to monitoring system within the performance specifications of the test device.

12.3 Rhythm detection system in AED-Mode

The DefiMonitor XD rhythm detection system analyses the patient's ECG and supports it if the device detects a rhythm requiring defibrillation or not.
The device’s rhythm detection system comprises:
Evaluating the electrode contact
Automatic evaluation of the ECG
Operator control of the defibrillation shock therapy
The transthoracic impedance of the patient is measured by the defibrillation electrodes. If the imped­ance measured is not within the valid range, the device establishes whether the electrodes have suffi­cient contact with the patient or are short-circuited. ECG analysis is hindered. The voice prompt "Check electrodes" follows if the electrode contact is inadequate.
Automatic interpretation of the ECG
The rhythm detection system of the device is designed to recommend a defibrillation shock when the system is has been connected up to a patient and the system detects a rhythm which requires defibril­lation. For all other ECG rhythms, including fine ventricular fibrillation, asystole and normal sinus rhythms, the rhythm detection system does not recommend defibrillation.
Operator control of the output of defibrillation shocks
The device's rhythm detection system triggers an automatic power charge if the device detects a car­diac rhythm which requires defibrillation. The defibrillator does not analyse the cardiac rhythm when charging energy and during readiness to shock. Optical and acoustic messages are generated to show the user that the device recommends giving a defibrillation shock. If a defibrillation shock is rec­ommended, the user must decide whether and when the shock is to be given.
The algorithm
Monitors the ECG rhythm during a continuous recording of up to 20 seconds
Filters interference and measures artefacts
Calculates several ECG signal parameters including frequency and morphological parameters -
rejects implantable pacemaker artefacts
Measures the QRS rate Based on the parameters collected, the algorithm decides whether the rhythm required defibrillation.
The first 4 seconds are used for initial diagnosis. If no rhythm has been established within the first 4
Appendix
100 / 108
DefiMonitor XD | 24050 | GB | I
seconds which requires defibrillation, the analysis is continued up to 16 further seconds and a decision taken each second.
Adult mode
The following databases are used for validation purposes: AHA and MIT. When calculating the characteristic section values in the ECG data sets for the above-named data-
bases which are marked with the PhysioBank annotation code as requiring shocks, we view these as cardiac rhythms requiring defibrillation. These sections also contain ventricular tachycardias which, however, are not annotated extra and cannot therefore be part of the statistics.
Data sets with a length of 20 seconds without rhythm changes and artefacts were used for validation purposes. The database incorporates 1369 data sets not requiring defibrillation and 185 which require defibrillation. These sections also contain ventricular tachycardias. However, they are not separately annotated and cannot therefore be shown in the statistics.
The performance results meet the requirements of IEC 60601-2-4:2010:
Sensitivity
> 90 %
Specificity
> 95 %
Real predictive value
> 90 %
False positive rate
< 5 %
Paediatric Mode
The following database was used for validation purposes: Development and validation dataset of the Physical-Technical Federal Institute (PTB) Berlin. This data was collected by the PTB within the scope of the research project MNPQ 07/09 carried out by the German Federal Ministry of Economics and Technology.
The cardiac rhythms not requiring defibrillation cover bundle branch blocks and supraventricular tachy­cardia, as well as normal sinusoidal rhythms. These sections also contain ventricular tachycardias which, however, are not separately evaluated and cannot therefore be included in the statistics.
The performance results meet the requirements of IEC 60601-2-4:2010:
Sensitivity
> 90 %
Specificity
> 95 %
Real predictive value
> 90 %
False positive rate
< 5 %

12.4 Electromagnetic compatibility

Guidelines and manufacturer's declaration on electromagnetic emissions
The DefiMonitor XD is designed for use in an environment like the one described below. The cus­tomer or user of the DefiMonitor XD should ensure that it is used in an environment of this kind.
Emitted interfer­ence measure­ments
Conformance
Electromagnetic environment – code of practice
HF emissions as per CISPR 11
Group 1 Class B
The DefiMonitor XD only uses HF energy for its internal function. This means that its HF emission is very low and it is unlikely that equipment in the vicinity will be disrupted.
Emission of har­monics as per IEC 61000-3-2
Class A
The DefiMonitor XD is suitable for use in all facilities (“pro-
fessional healthcare”), including residential areas (“home healthcare”) and those directly connected to a public sup-
ply network which also supplies buildings used for residen­tial purposes.
Emission of volt­age fluctuations / flickers as per IEC 61000-3-3
Compliant
Loading...